Important Medical Device Advisory NM Implantable Pulse Generator Inaccurate Elective Replacement Indicator

Similar documents
Important Medical Device Advisory NM Implantable Pulse Generator Inaccurate Elective Replacement Indicator

Important Medical Device Advisory

URGENT FIELD SAFETY NOTICE

Freelink Remote Control

CELLULAR TELEPHONE EQUIPMENT AND SERVICES POLICY

Medical records movement

URGENT - Medical Device Correction IntelliVue MX40 Missing Warnings in IFU

Bionic Navigator 1.2 Software Guide

WASHINGTON UNIVERSITY HIPAA Privacy Policy # 7. Appropriate Methods of Communicating Protected Health Information

Physical Security Reliability Standard Implementation

Vercise DBS Remote Control 3 Handbook

This document contains important information for the continued safe and proper use of your equipment

The Fairfield Schools Infinite Campus Portal: Parent Guide

SAFE USE OF MOBILE PHONES AT WORK POLICY

GMSS Information Governance & Cyber Security Incident Reporting Procedure. February 2017

BionicNavigator 1.2 Software Guide

Unofficial Comment Form Project Real-time Monitoring and Analysis Capabilities IRO and TOP-010-1

URGENT Field Safety Notice Philips HeartStart MRx Monitor/Defibrillator

URGENT FIELD SAFETY NOTICE Reserve battery and new instructions for users of the PAD 300/PAD 300P (Public Access Defibrillator)

Updating MACS documentation

March 20, Division of Dockets Management (HFA-305) Food and Drug Administration 5630 Fishers Lane, Room 1061 Rockville, MD 20852

PRODUCT SAFETY ALERT

This document contains important information for the continued safe and proper use of your equipment

Cyber Security Supply Chain Risk Management

IDENTITY THEFT PREVENTION Policy Statement

INFORMATION ASSURANCE DIRECTORATE

ACEDS. Certified E-Discovery Specialist RECERTIFICATION APPLICATION ACEDS.ORG. A BARBRI Professional Association

Standard Operating Procedure. Preparation, Review and Approval of Clinical Research Standard Operating Procedures and Work Instructions

FIELD CORRECTIVE ACTION NOTICE

Cellphone Provision Policy

Manage Users. Step 1 Navigate to the Manage Users tab on the dashboard and select the Invite New Manager page from the top navigation bar.

NHSmail Skype for Business

IATF Transition Strategy Presenter: Cherie Reiche, IAOB

Project Cyber Security - Order No. 791 Identify, Assess, and Correct; Low Impact; Transient Devices; and Communication Networks Directives

Clinician Programmer Print Tool For use with Prometra Programmable Infusion Systems Software Version

Risk: Security s New Compliance. Torsten George VP Worldwide Marketing and Products, Agiliance Professional Strategies - S23

ERO Enterprise Strategic Planning Redesign

Medical records movement

Volume Licensing Service Center User Guide MICROSOFT VOLUME LICENSING

Purpose This document defines the overall policy, principles, and requirements that govern the mybyu Portal.

Volume Licensing Service Center User Guide

Individual Funding Requests (IFR) Guidance notes for referrers (GPs and Referring Clinicians)

NAFSA Annual Conference and Expo. SEVIS Technical Functionality Session. May 30, 2017

Development Authority of the North Country Governance Policies

Privacy Breach Policy

It applies to personal information for individuals that are external to us such as donors, clients and suppliers (you, your).

Policy & Procedure Privacy Policy

Article I - Administrative Bylaws Section IV - Coordinator Assignments

1. Post for 45-day comment period and pre-ballot review. 7/26/ Conduct initial ballot. 8/30/2010

Ministry of Government and Consumer Services. ServiceOntario. Figure 1: Summary Status of Actions Recommended in June 2016 Committee Report

Major Appliance Recycling Roundtable ONLINE REGISTRATION & REPORTING SYSTEM INSTRUCTIONS

CONFIRM Rx TM INSERTABLE CARDIAC MONITOR WORLD S FIRST AND ONLY SMARTPHONE-COMPATIBLE ICM CONVENIENT. CONNECTED. CONTINUOUS.

CYBER SECURITY POLICY REVISION: 12

CARELINK NETWORK UPDATES SUMMARY OF ENHANCEMENTS May 2017

EEI Fall 2008 Legal Conference Boston, Massachusetts Stephen M. Spina November 1,

Vercise DBS Programming Manual

NIHB Claims Services Provider Website

Visit Mon General Registration

Office Properties Income Trust Privacy Notice Last Updated: February 1, 2019

Table of Contents. Policy Patch Management Version Control

Procedure for Handling Third Party Requests to Access Data on eportfolio

2016 Survey: A Pulse on Mobility in Healthcare

Policies and Procedures Date: February 28, 2012

COLLECTION & HOW THE INFORMATION WILL BE USED

April 28, Division of Dockets Management (HFA-305) Food and Drug Administration 5630 Fishers Lane, Room 1061 Rockville, MD 20852

Document Version: 1.0. Purpose: This document provides an overview of IBM Clinical Development released by the IBM Corporation.

RelayHealth Legal Notices

IATF Transition Strategy Presenter: Mrs. Michelle Maxwell, IAOB

Personal Information. New Profile Icon

Accreditation & Certification Supplier Guide

Identity Theft Prevention Policy

JUNIPER NETWORKS PRODUCT BULLETIN

Implementation of UDI In the Medical Device Industry

Last updated: July 7th, 2008 Microsoft Security Software Advisor Fee Initiative Guide ("SSA Program Guide")

Precision Spectra Remote Control Handbook, , Rev C

IAF Informative Document. Information on the Transition of Management System Accreditation to ISO/IEC :2015 from ISO/IEC 17021:2011

ICD-10 Compliance Project November 2013 update

Use of Synthetic Data in Live Environments

MNsure Privacy Program Strategic Plan FY

Brian Russell, Chair Secure IoT WG & Chief Engineer Cyber Security Solutions, Leidos

Internal Audit Follow-Up Report. Multiple Use Agreements TxDOT Office of Internal Audit

What s My Profile? REVISION 1. JANUARY

Welcome to the Hyster Training Knowledge Center

International Standard ISO/IEC 17799:2000 Code of Practice for Information Security Management. Frequently Asked Questions

Customer Service Excellence Response Audit (QM054c) Department: Quality Assurance. Summary by: Nassir Ahad (Quality Improvement Auditor)

Report (CCR) Rule Retrospective Review: Electronic Delivery

Urgent Medical Device Correction LIFEPAK 1000 Defibrillator

Steffanie Hall, RHIA HIM Director/Privacy Officer 1201 West 12 th Emporia, Kansas ext

INTERNATIONAL CIVIL AVIATION ORGANIZATION ASIA and PACIFIC OFFICE ASIA/PAC RECOMMENDED SECURITY CHECKLIST

2019 NCA Examination Schedule

NCDB Vendor Webinar: NCDB Call for Data April 2019 and RQRS Updates

1- How do you register for an account with Alberta Transportation s Online Services?

Relevant certificates must be shown to the primary trainer at commencement of clinical attachment.

Ticket to Work Helpline WIPA Referral Protocol

Army CCQAS 2.8 Quick Guide: Registering for a New User Account Reviewers, Evaluators, and Privileging Authorities

What is New in MyChart? My Medical Record Health Preferences Settings Appointments and Visits Visits Schedule an Appointment Update Information

Gain Control Over Your Cloud Use with Cisco Cloud Consumption Professional Services

Privacy Policy. I. How your information is used. Registration and account information. March 3,

SAS 70 revised. ISAE 3402 will focus on financial reporting control procedures. Compact_ IT Advisory 41. Introduction

7 Processing Applications for Interfacility Credentials Transfer Briefs (ICTBs)

Transcription:

Important Medical Device Advisory NM Implantable Pulse Generator Inaccurate Elective Replacement Indicator 12 September 2017 Dear Physician, In an effort to keep you continually informed, and to ensure optimal care of your patients, Abbott is advising physicians that our company is aware of instances in which the elective replacement indicator (ERI) in some devices has triggered earlier than intended. In these cases, the ERI alerts were triggered early due to an error in how the device calculates the actual remaining battery life in the impacted implantable pulse generators (IPG). As you know, ERI alerts estimate battery life based on programmed device parameters and patient usage and are unique to each patient. Importantly, you should be aware that this issue is an error in software calculation and is not an indication of the devices actual battery performance. Currently, all implanted IPGs within the Proclaim family are affected by this advisory. As of July 11, 2017, among the approximately 21,208 Proclaim Elite Spinal Cord Stimulation (SCS) and Proclaim Dorsal Root Ganglion (DRG) IPG devices sold worldwide, ERI messages have occurred earlier than intended in approximately 1.5% of devices. To date, of the devices that have received early ERI messages, 100 devices (0.5%) have been explanted and replaced. In addition, there is a low risk that the ERI alert will not trigger prior to a device's end of battery life, though this has not been observed to date. We are contacting physicians to provide details regarding risk and patient management recommendations, thereby enabling physicians to make a determination with their patients on whether they should consider an elective device replacement in these circumstances. Identifying Systems Subject to Inaccurate Battery Indications Currently, the patient controller (PC) battery capacity indicator may display less than full upon first use due to this calculation error. In this situation, the battery continues to function normally and no action is recommended. When the clinician programmer (CP) or PC display an ERI message, physicians are advised to follow patient management recommendations outlined below. Pictures of the current ERI messages in both the PC and CP can be found in Appendix A. Patient Management Recommendations Abbott is planning to address inaccurate longevity information in two phases. The first phase requires an individual assessment of the device prior to any replacement decisions when an ERI message is received. The second phase is a software update that addresses the error in ERI calculation, and will be deployed upon approval by local governing agencies. Phase I: ERI Assessment If a patient reports that an ERI message is displayed on the PC, please contact your Abbott Representative to obtain the device-specific generator logs and to conduct the ERI Assessment using the steps in Appendix B prior to making an elective replacement decision.

After the ERI Assessment is completed, follow-up communication from your Abbott representative will be provided in order to make appropriate decisions. The response will be one of two possible responses, specific to therapy model: o For the Proclaim DRG model only: 1. ERI is valid at this time for this device, follow existing information provided by IFU and CP for the elective replacement window for this device. 2. The device has the appropriate level of battery voltage to provide the existing therapy until the next assessment, which should occur no later than March 2018. During this period, the ERI message will be displayed on the PC at the start of each session. This message may be dismissed to enter the session on the PC. o For Proclaim SCS models only: 1. ERI is valid at this time for this device, follow existing information provided by IFU and CP for the elective replacement window for this device. 2. The device has the appropriate level of battery voltage to provide the existing therapy until the next assessment. Schedule a follow up appointment to re-perform the Elective Indication Assessment in approximately two (2) months. During this period, the ERI message will be displayed on the PC at the start of each session. This message may be dismissed to enter the session on the PC. Phase II: ERI Software Update Abbott will deploy a software upgrade, upon approval by local governing agencies that addresses the errors in calculation causing the inappropriate ERI message. When this software update is deployed, the Phase I ERI assessment process is no longer needed. After the CP is updated, replacement decisions can be made following the revised instructions for use, which will be available at manuals.sjm.com. Once approved for use in your area, notification of the availability of this software update will occur through the Apple Public App Store for the PC and through the SJM App Catalog for the CP. Should you have questions about patient management, including questions regarding error messages related to IPGs, please contact your local Sales Representative or Abbott Support at 1-800-727-7846 (Opt3) (U.S) and +46 8 474 4147 (Outside the U.S.). We apologize for any difficulties this may cause you and your patients. Abbott is committed to providing the highest quality products and support. Thank you for your continued support. Sincerely, Susan Jezior Slane Divisional Vice President, Quality Assurance and Compliance Abbott Cardiovascular and Neuromodulation Attachments

APPENDIX A: Replacement Indication Messages The images below show the current Patient Controller (PC) displays for Elective Replacement Indicator (A) and End-of-Service (B). APPENDIX B: Steps to Retrieve and Submit Generator Logs for ERI Assessment At the patient evaluation visit, conduct these steps prior to any programming assessment or changes to programming from that which has been used resulting in ERI message displayed on the Patient Controller (PC). It is important to assess the as used state of the device prior to changes for ERI assessment. Extract the generator log and submit the generator log for ERI assessment. Environment Requirements in order to execute the extraction steps: SJM Clinician Programmer Email Account on the Clinician Programmer Network Connectivity

Connect to the IPG with the Clinician Programmer and press the IPG icon in the header to view the generator information screen and press the Generator Log option NOTE: this screen content will vary slightly depending on therapy Select the Email Basic Generator Log option and wait for process to complete

After completion of the Email Basic Generator Log option an email is automatically generated. Press Send to email the log files to NMDTechnicalSupport@sjm.com The extracted and submitted logs will be associated to a record ID for the assessment of the ERI event internally by Abbott NOTE: the device does not have to be on a network to complete these steps After analysis by Abbott, if the response to ERI assessment is that the ERI is valid, follow existing Clinician Programmer guidance for remaining elective replacement window. This response will be provided to your Abbott representative. After analysis by Abbott, if response to the ERI assessment is that a future assessment, after a provided period of time is recommended; please schedule a follow up with the patient to conduct that assessment. DRG: The ERI assessment is that the device has the appropriate level of battery voltage to provide the existing therapy until the next assessment; please schedule a follow up with the patient to conduct that assessment no later than March 2018 SCS: The ERI assessment is that follow-up is recommended approximately 2 months from the date generator log files were obtained to re-conduct the ERI assessment. Instructions for Use can also be located at manuals.sjm.com for additional information regarding CP / PC messages and programming limits. Frequently Asked Questions is available on www.sjm.com/notices that you may use or provide to your patients to assist them with information regarding this issue, and the process to evaluate the ERI messages.