Validation of a digital mammography image simulation chain with automated scoring of CDMAM images

Similar documents
Monte Carlo simulation of scatter field for calculation of contrast of discs in synthetic CDMAM images

Detection of microcalcification clusters by 2D-mammography and narrow and wide angle digital breast tomosynthesis

Effective detective quantum efficiency (edqe) and effective noise equivalent quanta (eneq) for system optimization purposes in digital mammography

Nicholas Marshall Department of Radiology, University Hospitals Leuven, Herestraat 49, 3000 Leuven, Belgium

Linköping University Post Print A MONTE CARLO-BASED MODEL FOR SIMULATION OF DIGITAL CHEST TOMOSYNTHESIS

Digital Scatter Removal in Mammography to enable Patient Dose Reduction

Foveal Algorithm for the Detection of Microcalcification Clusters: a FROC Analysis

Investigating Oblique Reconstructions with Super-Resolution in Digital Breast Tomosynthesis

Noise power spectrum and modulation transfer function analysis of breast tomosynthesis imaging

Power Spectrum Analysis of an Anthropomorphic Breast Phantom Compared to Patient Data in 2D Digital Mammography and Breast Tomosynthesis

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

Simulation of Internal Backscatter Effects on MTF and SNR of Pixelated Photon-counting Detectors

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images

Digital breast tomosynthesis: comparison of different methods to calculate patient doses

Scatter reduction for grid-less mammography using the convolutionbased image post-processing technique

A Method for Producing Simulated Mammograms: Observer Study

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

A fast scatter field estimator for Digital Breast Tomosynthesis

Design Considerations in Optimizing a Breast Tomosynthesis System

Background 8/2/2011. Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging. Stephen J.

An Automated Image-based Method for Multi-Leaf Collimator Positioning Verification in Intensity Modulated Radiation Therapy

Simple method for modulation transfer function determination of digital imaging detectors from edge images

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

Detailed analysis of scatter contribution from different simulated geometries of X-ray detectors

A Software Tool for Measurement of the Modulation Transfer Function

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE /

Monte Carlo Simulation for Neptun 10 PC Medical Linear Accelerator and Calculations of Electron Beam Parameters

Application of Tatian s Method to Slanted-Edge MTF Measurement

Medical Physics Research Center, Mashhad University of Medical Sciences, Mashhad, Iran.

Towards Breast Anatomy Simulation Using GPUs

Assessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT

1. Deployment of a framework for drawing a correspondence between simple figure of merits (FOM) and quantitative imaging performance in CT.

Spatial Resolution Properties in Penalized-Likelihood Reconstruction of Blurred Tomographic Data

Phase-Contrast Imaging and Tomography at 60 kev using a Conventional X-ray Tube

Detector Noise evaluation by means of Continue Wavelet Transform. Comparison with Fourier Transform methods

Projection and Reconstruction-Based Noise Filtering Methods in Cone Beam CT

Edge and local feature detection - 2. Importance of edge detection in computer vision

Partial Volume Simulation in Software Breast Phantoms

Iterative CT Reconstruction Using Curvelet-Based Regularization

8/1/2017. Current Technology: Energy Integrating Detectors. Principles, Pitfalls and Progress in Photon-Counting-Detector Technology.

Investigation of the effect of varying scatter-toprimary ratios on nodule contrast in chest tomosynthesis

IOSR Journal of Electronics and communication Engineering (IOSR-JCE) ISSN: , ISBN: , PP:

Acknowledgments and financial disclosure

Quantitative Modelling of Microcalcification Detection in Digital Mammography

Quality control phantoms and protocol for a tomography system

Digital Image Processing ERRATA. Wilhelm Burger Mark J. Burge. An algorithmic introduction using Java. Second Edition. Springer

Methodology of NEQ (f) analysis for optimization and comparison of digital breast tomosynthesis acquisition techniques and reconstruction algorithms

Registration concepts for the just-in-time artefact correction by means of virtual computed tomography

Developments in Dimensional Metrology in X-ray Computed Tomography at NPL

Classification of Hyperspectral Breast Images for Cancer Detection. Sander Parawira December 4, 2009

Modelling Flat Spring Performance Using FEA

8/3/2016. Outline. The EPID Strikes Back: Future EPID Technology and Applications. Active Matrix Flat-Panel Imagers (AMFPIs)

Structural Similarity Optimized Wiener Filter: A Way to Fight Image Noise

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

DUAL energy X-ray radiography [1] can be used to separate

Imaging-based Classification Algorithms on Clinical Trial Data with Injected Tumour Responses

Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers

An Improvement of the Occlusion Detection Performance in Sequential Images Using Optical Flow

Breast Image Registration Using Non-Linear Local Affine Transformation

DIGITALIZATION OF A MAMMOGRAPHIC PHANTOM VIEW THROUGH A MONTE CARLO SIMULATION

Convolution-Based Truncation Correction for C-Arm CT using Scattered Radiation

CT Systems and their standards

Generalized Filtered Backprojection for Digital Breast Tomosynthesis Reconstruction

Contrast Enhancement with Dual Energy CT for the Assessment of Atherosclerosis

Spectral analysis of non-stationary CT noise

Energy resolved X-ray diffraction Cl. J.Kosanetzky, G.Harding, U.Neitzel

Limitations of Projection Radiography. Stereoscopic Breast Imaging. Limitations of Projection Radiography. 3-D Breast Imaging Methods

Improvement of Efficiency and Flexibility in Multi-slice Helical CT

SPECT QA and QC. Bruce McBride St. Vincent s Hospital Sydney.

Computer Vision and Graphics (ee2031) Digital Image Processing I

Digital Image Processing COSC 6380/4393

CoE4TN4 Image Processing. Chapter 5 Image Restoration and Reconstruction

Virtual Clinical Trials for the Assessment of Novel Breast Screening Modalities

Respiratory Motion Estimation using a 3D Diaphragm Model

Support Vector Machine Density Estimator as a Generalized Parzen Windows Estimator for Mutual Information Based Image Registration

Development and Evaluation of 2D and 3D Image Quality Metrics. Simon Nicholas Murphy. Graduate Program in Medical Physics Duke University

Industrial Radiography Simulation by MCNPX for Pipeline Corrosion Detection

Adaptive Power Blurring Techniques to Calculate IC Temperature Profile under Large Temperature Variations

Cluster-oriented Detection of Microcalcifications in Simulated Low-Dose Mammography

Set-Permutation-Occurrence Matrix based Texture Segmentation

Detection of Sub-resolution Dots in Microscopy Images

Interpolating Silicon Photomultipliers

Computer-aided detection of clustered microcalcifications in digital mammograms.

Noise filtering for television receivers with reduced memory

Application of Characteristic Function Method in Target Detection

then assume that we are given the image of one of these textures captured by a camera at a different (longer) distance and with unknown direction of i

Motivation. Gray Levels

Motivation. Intensity Levels

Scatter Correction Methods in Dimensional CT

ALGORITHM OF PRELIMINARY SESSILE DROP PROCESSING FOR NOISE SUPPRESSION X

Frequency-based method to optimize the number of projections for industrial computed tomography

Fits you like no other

Accuracy of parameterized proton range models; a comparison

ABSTRACT 1. INTRODUCTION

Robust Ring Detection In Phase Correlation Surfaces

7/13/2015 EVALUATION OF NONLINEAR RECONSTRUCTION METHODS. Outline. This is a decades-old challenge

SINGLE-PHOTON emission computed tomography

Sampling-Based Ensemble Segmentation against Inter-operator Variability

Basic Algorithms for Digital Image Analysis: a course

8/2/2016. Acknowledgement. Common Clinical Questions. Presumption Images are Good Enough to accurately answer clinical questions

Transcription:

Validation of a digital mammography image simulation chain with automated scoring of CDMAM images Mary Yip 1, Abdulaziz Alsager 2, Emma Lewis 1, Kevin Wells 1, and Kenneth C. Young 3 1 Centre for Vision, Speech and Signal Processing, Faculty of Engineering and Physical Sciences, University of Surrey, Guildford, GU2 7XH m.yip@surrey.ac.uk 2 Department of Physics, Faculty of Engineering and Physical Sciences, University of Surrey, Guildford, GU2 7XH 3 National Coordinating Centre for the Physics of Mammography, Royal Surrey County Hospital, Guildford, GU2 7XX Abstract. A wide variety of digital mammography systems are available for breast cancer imaging, each varying in physical performance. However, the relationship between physical performance assessment and clinical outcome is not clear. Thus, a means of simulating technically and clinically realistic images from different systems would represent a first step towards elucidating the impact of physical performance on clinical outcome. To this end, a framework for simulating technically realistic images has been developed. A range of simulated test objects, including CDMAM have been used to determine whether the simulation chain correctly reproduces these objects thus validating the simulation framework. Results evaluated for two digital mammography systems have been promising, with simulated images proving similar to experimental images for Modulation Transfer Function and Normalised Noise Power Spectrum measurements differing by approximately 3%. Key words: Digital mammography, simulation, CDMAM phantom, validation 1 Introduction With the growing availability of commercial digital mammography technology, it is becoming increasingly important to be able to assess the physical performance of these systems in terms relevant to impact on clinical outcome, i.e. cancer detection. Since extensive clinical trials to examine any one particular technology are prohibitively expensive, a method of synthesising technically and clinically realistic images would be highly relevant. This could be used to assess the performance of radiologists and CAD systems in detecting cancers when different systems, beam qualities, radiation doses and image processing is employed. Therefore, a simulation framework has been developed to generate images typical

2 Mary Yip et al. of specific digital mammography systems. This framework uses measurements of physical performance parameters such as Modulation Transfer Function (MTF) and Normalised Noise Power Spectrum (NNPS). However, before simulating clinically realistic mammographic images, we have attempted to simulate a range of common test objects, including the CDMAM to examine whether our simulated images correctly represent the experimental data, and thereby validate the simulation framework. The CDMAM phantom is a contrast-detail test object widely used to evaluate the performance of mammography systems in terms of their ability to demonstrate very fine and low contrast details. The test object is made up of cells with two identical gold discs: one at the centre, and a second disc located in one of the four corner quadrants within the cell. The task of an observer is to attempt to locate the corner disc in each cell. As this is a time consuming and subjective task, automated scoring software has been developed [1],[2],[3],[4] and used as part of the validation process. Results for two digital mammography systems are presented here, one with a selenium detector and the other with a CsI phosphor detector system. 2 Methodology Images of three types of test phantom were simulated. Uniform images of a 45mm thick block of PMMA (to compare the NNPS of real and simulated images) Images of a straight edge for MTF measurement (to compare the MTF of real and simulated images) Images of the CDMAM phantom (to assess the contrast degradation of small discs and contrast detail detection curves for real and simulated images) The image simulation chain to simulate the CDMAM test object is described below. Similar steps were used albeit simplified to generate synthetic images for assessing the MTF and NNPS. This simulation chain comprised of two parts: the first part creates a noise free image blurred according to the particular MTF or impulse response and the second creates a noise image associated with a particular dose level for a specific imaging technology. This is illustrated in Fig. 1. The noise free blurred image was created from a binary template representing the background structures in the CDMAM test object (cell mesh, manufacturer icons and numerical labelling). The binary template was created by thresholding an actual high-dose X-ray image of the CDMAM phantom. The pixel values of the background and mesh were assigned as measured experimentally for a specific dose. A heel effect mask was applied at this point to account for residual artefacts left after the manufacturer s flat-field algorithm had been applied to the detector images. This was derived through median filtering of an X-ray image of the CDMAM phantom, acquired on the system to be simulated, and normalised. This idealised template is upsampled to 10µm per pixel so that the image simulation chain may be applied to any digital mammography detector

Validation of digital mammography image simulation chain 3 Fig. 1. Flow chart of image simulation chain currently available. Discs are added with contrasts based initially on the theoretical contrast-thickness relationship for the beam quality and phantom geometry used, and adjusted to match empirically measured disc contrasts on real images. The presampled MTF of the detector is measured experimentally and modelled in 2D to produce a Fourier filter [5]. This is multiplied by the Fourier transform of the idealised image to produce the noise free blurred image, I B, representing an idealised (noise free) image of the particular imaging system under examination. In addition to the presampled MTF, the geometric MTF, due to the finite size of the focal spot, was also modelled based on Sandborg s work [6]. It is assumed that the focal spot intensity distribution follows a box function, where there is uniform emission distribution within the box with sides f 0 and no emission outside of the box. Therefore, the MT F geo can be modelled as: MT F geo = sin (πf 0 v (m 1)) /πf 0 v (m 1) (1) where v is the spatial frequency and m is the magnification of the system for the images of the gold discs in the CDMAM. The geometric MTF is also applied to the Fourier transformed I B and inverse Fourier transformed before it is downsampled to the detector pixel size of the particular system to be simulated, using neighbourhood averaging methods, leaving the blurred resampled image, I BR. To create a noise image, a Fourier noise filter is derived based on the NNPS of the system. Each dose was simulated with the use of NNPS measured from a uniform image taken at a dose similar to the experimental CDMAM image. This is measured in 2D with a Hann window [5] and used to filter a Gaussian noise

4 Mary Yip et al. field in Fourier space. This is inverse Fourier transformed to create the noise image, I N. Flat field correction algorithms are used in digital mammography sets to correct pixel intensity values due to device defects and the heel effect. However, this does not correct for the non-uniform variance across the image due to structural and geometric variations photon statistics across the detector face. Therefore, a variance map [7] was incorporated to model this effect as shown in equation 2: I NV (x, y) = I V (x, y).i N (x, y) (2) The noise image, I NV, is scaled on a pixel by pixel basis in relation to the resampled blurred image, I BR. This is denoted in equation 3 by σ 2 (I BR (x, y)), an empirically measured mean-variance relationship for a particular detector[8]: I NS (x, y) = I NV (x, y). σ 2 (I BR (x, y)) (3) As dose increases, the noise (measured as variance) also increases. Finally the simulated image is created by adding this scaled noise image to the blurred resampled image as illustrated in Fig 1. 2.1 Validation Validation of this simulation chain was achieved by comparing the simulated images with the real images based on image quality metrics such as MTF and NNPS as shown in Fig 2(a) and 2(b). The contrast degradation factor (CDF) [6] was calculated for different disc diameters. As the discs were added on a contrast-thickness relationship, the contrast for each disc is known, therefore the effects of MTF modelling could be assessed. The peak contrasts were measured for each disc using equation 4 C peak = (P b φ) (P disc φ) (P b φ) (4) where P b is the mean background pixel value within the cell, P disc is the peak pixel value in the disc and φ is the pixel value offset for the detector. The pixel values are linear with the absorbed radiation dose to the detector. The peak contrast of each disc was measured in the disc input iamge, C input, (before the MTF blurring) and blurred resampled image, C BR, after the MTF blurring and resampling stage, (but before noise is added) in order to remove any discrepancies due to noise, using the following CDF equation: CDF MT F = C BR C input (5) It is expected that the CDF will be 1.0 for the larger discs and fall below 1.0 for the smaller discs due to contrast losses due to MTF-induced blurring. The simulated images of the CDMAM were automatically analysed using a CDMAM scoring software tool, CDCOM, to predict the threshold gold thickness for a typical observer as outlined in previous papers[1],[3],[9],[4] and compared

Validation of digital mammography image simulation chain 5 with results of a similar analysis of experimental CDMAM images. Sets of eight images of the CDMAM test object were simulated at three doses (1.56mGy, 3.21mGy, 6.46mGy) for a selenium detector and contrast-detail curves plotted. Results were also obtained for a system with a CsI/amorphous silicon detector at one dose (6.46mGy). 3 Results Experimental and simulated straight edge and uniform images for the system were used to derive the MTF and NNPS curves, and these are compared in Figures 2 and 3, respectively. The experimental and simulated MTF profiles for the selenium system differed by 4.5% compared with 3.2% for the CsI system. Both systems differed by approximately 3% for the radial NNPS profile. The (a) (b) Fig. 2. Comparison of real and simulated images using (a) MTF and (b) radial NNPS for the selenium detector (a) (b) Fig. 3. Comparison of real and simulated images using (a) MTF and (b) radial NNPS for the CsI phosphor detector

6 Mary Yip et al. CDFs for different detail diameters are shown in Fig. 4, demonstrating the effect of the presampled MTF on disc contrast in simulated images. The CDF for the CsI system degraded smaller discs earlier than for the selenium system. Contrast Fig. 4. Contrast degradation factor for various disc diameters. Error bars indicate 2 standard errors of mean. detail curves were fitted to predicted human results[1][3][9] of the experimental and simulated CDMAM image sets are plotted in Fig. 5 and Fig. 6. As a result of on-going work, we have found that NNPS is critically dependent on dose. Therefore, only one dose level has been successfully simulated for the system with a CsI detector. 4 Discussion MTF and NNPS curves for the real and simulated images provided good matches for all spatial frequencies. The average difference between experimental and simulated data for MTF and radial NNPS profiles was approximately 3% illustrating the effectiveness of the simulation framework.. The CDF curve (Fig. 4) showed the effect of the MTF curve applied in our framework on the contrasts of the smaller discs. Both systems exhibited a CDF close to one for large details and was much smaller for details with diameters below 0.4mm. The lower MTF of the system with a CsI detector caused a lower CDF for the smaller discs when compared to the system with the selenium detector as expected. The contrast detail curves for simulated images of the CDMAM test object were successfully obtained as shown in Fig. 5 and 6. Results for the simulated data for the selenium system are comparable to that obtained experimentally. As mentioned in the results, we have found that NNPS is critically dependendent on dose, and as such only one dose was successfully simulated for the CsI detector. This is currently under development. This validation has shown that our simulation framework largely addresses our initial aims, but some further refinement is needed to improve the match to real images from this system.

Validation of digital mammography image simulation chain 7 (a) (b) Fig. 5. Contrast detail curves for (a) experimental and (b) simulated CDMAM images at three doses for the selenium system. Error bars indicate 2 standard errors of mean.

8 Mary Yip et al. Fig. 6. Contrast detail curve for experimental and simulated CDMAM images at one dose for the CsI system. Error bars indicate 2 standard errors of mean. References 1. Young, K.C., Cook, J.J.H., Oduko, J.M.: Automated and human determination of threshold contrast for digital mammography systems. In: Astley, S.M., Brady, M., Rose, C., Zwiggelaar, R. (eds.) IWDM 2006. LNCS, vol. 4046, pp 266 272. Springer, Heidelberg (2006) 2. Visser, M., Karssemeijer, N.: Manual CDCOM version 1.5: software for automated readout of CDMAM 3.4 images, www.euref.org 3. Young, K.C., Cook, J.J.H., Oduko, J.M., Bosmans, H.: Comparison of software and human observers in reading images of the CDMAM test object to assess digital mammography systems. In: Flynn, M.J. and Hsieh, J. (eds.)medical Imaging 2006: Physics of Medical Imaging. SPIE, vol. 6142, San Diego (2006) 4. Young, K.C., Alsager, A., Oduko, J.M., Bosmans, H., Verbrugge, B., Geertse, T., van Engen, R.,:Evaluation of software for reading images of the CDMAM test object to assess digital mammography systems. In: Hsieh, J. and Samei, E. (eds.) Medical Imaging 2008: Physics of Medical Imaging. SPIE, vol. 6913, (2008) 5. Saunders, R.S., Samei, E.,: A method for modifying the image quality parameters of digital radiographic images. Medical Physics, 30(11), 3006 3017, 2003. 6. Sandborg, M., Dance, D.R., Alm Carlsson, G.: Calculation of contrast and signalto-noise degradation factors for digital detectors in chest and breast imaging. Department of Radiation Physics, Linkoping University, 2003. ISSN 1102-1799. 7. Marshall, N.W.: A comparison between objective and subjective image quality measurements for a full field digital mammography system. Phys. Med. Biol. 51, 2441 2463 (2006) 8. Carton, A-K., Vandenbroucke, D., Struye, L., Maidment, A.D.A., Kao, Y-H., Albert, M., Bosmans, H., Marchal, G.,: Validation of MTF measurement for digital mammography quality control. Medical Physics, 32(6), 1684 1695 (2005) 9. Karssemeijer, N., Thijssen, M.A.O.: Determination of contrast detail curves of mammography systems by automated image analysis. In: Doi, K. and Giger, M.L. and Nishikawa, R.M. and Schmidt, R.A. (eds.). IWDM 1996. pp 155 160. Elsevier, Amsterdam, (1996)