AquaTroll 600 Configuration and MultiLogger Integration

Similar documents
Integrating the Yieldpoint d-exto MPBX

Integrating HART Sensors with the Campbell CR1000 Datalogger

Integrating the MDT SMART Link-485 Sensor Application Note #32

Integrating the Measurand Shape Accelerometer Arrays (SAA) Sensor Application Note #19

Using the Lantronix SGX 5150 Serial Device Server

MultiLogger Software

Integrating the RXTX Telependulum with the CR800/CR1000 Sensor Application Note #17

Using the Lantronix WiBox with the CR800 or CR1000

Integrating the White Mini-Seis III Seismic Recorder with MultiLogger Suite

Integrating SISGEO Digitized MEMs IPI s with the CR800 or CR1000 Sensor Application Note #21

Using the RAVEN II CDPD Modem Application Note #12

Integrating the Kinemetrics Obsidian Seismic Recorder with MultiLogger Suite

MST Database Integration Database Application Note #10

Using Campbell AVW2xx Vibrating Wire Interfaces MultiLogger Application Note #21

CANARY. Acquisition Solutions. What Do We Do? Software Systems. Network Management. Sensors Input. Dataloggers Measure. Communications Transfer

Technical Note: Interfacing a C-sense pco 2 Sensor to a Campbell Scientific CR1000 Datalogger

Preparing Chart Aggregations with Custom Report

All About the MultiLogger Setup File Application Note #7

Using the MultiLogger Language Manager Software Application Note #18 Revision

YSI EXO Series Multi-Parameter Water Quality Sondes

Sensor Mux Wiring Diagrams MultiLogger Application Note #17 Revision March 2008

RST INSTRUMENTS LTD.

Modbus Signal Output Adapter Quick Start Guide

Temperature-Humidity Sensor Configuration Tool Rev. A 1/25/

Campbell Based Multi-Channel Data Logger. User Manual. Man /08/2014 Final David Farnham. Chris Rasmussen. Chris Rasmussen. Manual No.

Geokon Model D Pendulum Operation Notes Sensor Application Note #16

Product User Guide PR2000. PR2000 Pressure Data Logger with LCD

Using the VWDSP Interface Application Note #11

Thermo Scientific Orion Star A329 Portable ph/ise/conductivity/rdo/do Meter. Instruction Sheet

How to Integrate Digiquartz Intelligent Products with Campbell Scientific CR1000 Series Data Loggers

HI 9828 Multiparameter Water Quality Meter...

HI 9828 Multiparameter Water Quality Meter...

User Manual. Thermo Scientific Orion

Thermo Scientific Orion Star A326 Portable ph/rdo/do Meter. Instruction Sheet

Docking Stations DS-U1, DS-U2, DS-U4, DS-U4-WL and DS-U Adjustment and configuration procedure for the analog inputs

CAMPBELL BASED MULTI-CHANNEL DATA LOGGER LITE USER MANUAL

YOUR EYES IN THE FIELD

MultiMux Multiplexer

41126 Cognento (MODENA) Italy Via Bottego 33/A Tel: +39-(0) Internet: Fax: +39-(0)

Reading Aquistar Smart Sensors with Campbell Loggers

MultiLogger Suite Installation Guide 2016

Aqua TROLL 600 Multiparameter Sonde

Fox Thermal Instruments, Inc.

FIELD OPERATION GUIDE YSI 610 DISPLAY/LOGGER

Docking Station DS-U4WEB with web server version 1 Instruction Manual

ph and Electrochemistry

Water Quality Meter. Water Quality Meter Kit. Instruction Manual

Campbell Based Multi-Channel Data Logger Lite. User Manual. Man158a /05/2015 Final Ben Scott James Kwist Philip Day. Manual No.

Fox Thermal Instruments, Inc.

Diver-Office Premium DEMONSTRATION EXCERCISE

Multiparameter Waterproof Meters. Multiparameter measurement in compact, IP67 rated waterproof enclosures

AANDERAA. DATALOGGERS 3634 and 3660

RST INSTRUMENTS LTD.

User Manual. Thermo Scientific Orion Star A320 Series

The Art of Measuring. User Manual English. Paraly SW 112. Latest Product Information:

Revision: 10/26/2018 Copyright Campbell Scientific

Using the Hotwire Anemometer

Parallel Printers Set-Up

RST INSTRUMENTS LTD.

Thermo Scientific Orion Star A325 Portable ph/conductivity Meter. Instruction Sheet

Thermo Scientific Orion Star A223 and Star A323 Portable RDO/DO Meter. Instruction Sheet

D8000 SERIES QUICK START GUIDE

Safety. 2. Configuration - Operation 2.1 RTD Selection 2.2 Driver Test Header. 3. Diagnostics 3.1 ph Input 3.2 Temperature Input

GRYF. Measuring Instruments. measuring electrochemical and physical values. monitoring the environment. controlling technological processes

VAN WALT Monitoring your needs

Datalogging Conductivity/ TDS Meter. Instruction Manual

Read Temperature Data

Contents. Contents... 1 Introduction... 4 An Overview of HOBOware... 4

Contents. Onset Computer Corporation 470 MacArthur Blvd. Bourne, MA Contents Introduction... 3 An Overview of HOBOware...

Datalogging ph Meter

Thermo Scientific Orion Star A222 and Star A322 Portable Conductivity Meter

Enhancement Improved Database import and storage for DXF sources. Bug Update incorrect CR1000 gage types. Bug Update to MDT SmartLink Gage File

B: Boiler Condensate Conductivity

See below for software screens to guide you through the configuration and calibration of your sensor

Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips

Hanna Instruments HI9819X Multiparameter series

TS110 DISPLAY MANUAL Pro Refrigeration, Inc

Benchtop Multi-parameter Water Quality Meter

Reference Manual. Rohrback Cosasco Systems, Inc. Tel: +1 (562) East Smith Avenue Fax: +1 (562)

19.Configure HMI as a MODBUS Server

Color iqc and Color imatch. Getting Started and Spectrophotometer Setup Guide

Getting Started: Using HOBOware with HOBO Data Loggers

FT2 View Instruction Manual

DATA LOGGER MODELS DL-1080/1081

BST-DL114 / BST-DL115/BST-DL116/BST-DL117 Wireless Monitor Temperature Humidity Data Logger. Manual V1.0

Temperature Datalogger

Configuring LabVIEW 2009 to Communicate with Acromag Busworks and Etherstax Modbus I/O Modules via an OPC Server

Instruction Manual XLComm Data Acquisition Software For XL Series Meters

ShineNet. Growatt Version Date Notes New. ShineNet V1.0

Synergy Application Software

and Multiparameter Meters

Level TROLL 100 and Level TROLL 200 Operator s Guide

Datalogger and RS232 Interface

HIGH-SPEED DOWNLOAD PROCEDURE. EDGE / IoE2 Modules

Course Code: ServiceMaxx Overview and Snapshot Analysis Study Guide

9936A LogWare III. User s Guide. Revision

Quick Guide Flex Gateway G2

Electrochemistry Meters. Series SD (IP 67 waterproof) Electrochemistry Meters

GRYF. Measuring Instruments. For Fisheries. Practice and research

Onset Computer Corporation 470 MacArthur Blvd. Bourne, MA Part #: MAN-BHW-GS Doc #: B. 28 HOBOware Getting Started Guide

Transcription:

5 Gould Road, PO Box 2155 New London, NH 03257 USA Voice: (603) 526-9800 info@canarysystems.com www.canarysystems.com AquaTroll 600 Configuration and MultiLogger Integration Sensor Application Note #31 Revision 06/12/2018 Overview The AquaTroll 600 is a rugged multiparameter sonde that can be deployed in a variety of settings. It can measure water level (pressure), temperature, barometric pressure and water quality data, to name a few. The AquaTroll 600 can be integrated with Canary System's MLSuite, allowing for automated data acquisition. The Canary Systems MLDAQ-WD includes an RE Smith signal converter capable of converting the AquaTroll RS485 output to RS232, and a Campbell Scientific CR1000 datalogger. This Sensor Application Note covers the setup of a single probe on the datalogger. Contact Canary Systems if your application requires multiple probes per datalogger. (Figure 1 The In-Situ AquaTroll 600) Related documentation: The AquaTROLL 600 Operator's Manual specifies installation and calibration of the instrument. The MultiLogger Software User's Guide details the use of MultiLogger and its features. AquaTroll is a registered trademark of In-Situ Inc. Canary Systems, MultiLogger and MLSuite are registered trademarks of Canary Systems, Inc. Campbell Scientific is a registered trademark of Campbell Scientific, Inc. AquaTroll 600 Configuration and MultiLogger Integration 1

System Wiring and Configuration A converter is typically required to convert the RS-485 output of the AquaTroll to RS-232. The example wiring in Figure 2 outlines connecting an AquaTroll to a Campbell Scientific CR1000 using the RSCOM2A converter from RE Smith. The RSCOM2A provides a low-cost, highly configurable signal converter. (Figure 2 Connecting an AquaTroll to a CR1000 using the RSCOM2A) AquaTroll 600 Configuration and MultiLogger Integration 2

Function Test Once wiring is complete, functionality of AquaTroll sensors should be tested by connecting them directly to a PC (e.g. with a TrollCOM USB adapter). Install the Win-Situ 5 Plus software from the software disc included with the AquaTroll sensor. Once installed, verification of the readout from the sensor, and Modbus configurations should be performed. If a multi-drop Modbus installation is necessary, some additional configuration of the Modbus settings is required. (Figure 3 Select No after launching Win-Situ) After launching the Win-Situ software, the above dialog box should appear. Select No. (Figure 4 The Communication Settings dialog of Win-Situ) The Communication Settings dialog box should appear next. Once configuration for the chosen method of communications is set, the device address can be found by selecting Search For Devices. After all settings are completed, select the check mark button in the bottom right-hand corner. AquaTroll 600 Configuration and MultiLogger Integration 3

(Figure 5 The configuration tab of the Win-Situ data monitoring window) The main data monitoring window should appear, with the AquaTroll model and serial number displayed. To verify the AquaTroll serial communications settings, and/or to configure for multi-drop Modbus, select the configuration tab. (Figure 6 AquaTroll settings and status in the configuration tab, Modbus Setup button highlighted) The above should be displayed in the main window. If device status appears to be OK, select Modbus Setup. AquaTroll 600 Configuration and MultiLogger Integration 4

This will bring up the Serial Communication Settings window. Verify that the settings match the settings above. If using multiple sensors on one station, assign the device addresses (from 1 to 4) in the Device Address box, and set the baud rate to 19200. Select the check mark icon in the bottom right when complete. (Figure 7 The Serial Communication Settings window, where Device Address and baud rate can be set) The AquaTroll is now ready for MultiLogger integration. See the following section. AquaTroll 600 Configuration and MultiLogger Integration 5

MultiLogger Configuration The steps below are necessary for establishing communications between the AquaTroll and MLSuite and collecting data: 1. Connect the AquaTroll's datalogger to the network. 2. Configure the AquaTroll sensors. 3. Configure data collection and enable automatic database import. Refer to the MultiLogger Software User's Guide for more information on MultiLogger. Connect the AquaTroll to the MultiLogger Network In order to add the AquaTroll to the network, a node representing the datalogger it is connected to must be added to the network tree within MultiLogger. From the initial Network Manager form of MultiLogger, press Add and add a Gateway to the network tree, then save changes, or select an existing Gateway. (Figure 8 A blank MultiLogger Network Manager configuration after adding a Gateway, Add button highlighted) AquaTroll 600 Configuration and MultiLogger Integration 6

Next, add a TCP/IP Socket and name it as necessary. Save the changes. (Figure 9 The new Socket added to the Gateway in MultiLogger) Left-click on the TCP/IP socket so that it is selected, press Add and select the datalogger model that the AquaTroll is connected to. Check the inside of the MLDAQ to confirm the model. This example uses a CR1000 from the Campbell Current Loggers group. Press Accept. (Figure 10 Selecting a CR1000 datalogger from the Add new Gateway or Device form) AquaTroll 600 Configuration and MultiLogger Integration 7

A prompt will display for selecting the Logger Configuration File. Select Assign a new Configuration File and press Accept. (Figure 11 Assigning a new Logger Configuration File to the CR1000 datalogger) A datalogger node will be added to the tree. Select the device, and it can be renamed by slow double-clicking its node. When selected, open the Settings tab to enter the IP address and TCP port number of the device. An example is shown in Figure 12 below. Save the configuration. (Figure 12 Updating the IP address of the AquaTroll with that of the installed and configured WiFi adapter) The AquaTroll sensors can now be configured. AquaTroll 600 Configuration and MultiLogger Integration 8

Configure the AquaTroll Sensors Once the AquaTroll is configured on the network, its sensors can be configured. This is done through the MultiLogger Logger form, which is opened by selecting the Logger icon at the top of the MultiLogger Network Manager screen, as shown in Figure 13: (Figure 13 The location of the Logger form icon in the MultiLogger Network Manager screen) From the Program tab of the Logger form, click the Direct Connect button. This opens the Configure Direct Connect Channels form, as shown in Figure 14. Each AquaTroll sensor is represented by a channel, each of which can be configured from this form. (Figure 14 The Direct Connect button in the Logger form Program tab) AquaTroll 600 Configuration and MultiLogger Integration 9

Initially, the form displays options for the first channel. Note the CHANNEL selector on the left of the form, from which the channel being configured can be selected. (Figure 15 The Configure Direct Connect Channels form, with a Depth sensor configured for channel 1) When configuring channels, Digital must be selected as the Gage Type in the Measurements group, and In- Situ must be selected as Make. Once Gage Type and Make have been set, an option can be selected from the Model drop-down that represents the specific sensor being configured. A complete list of AquaTroll sensors and their associated Model selection is available below: ModBus COM1 Actual Conductivity ModBus COM1 Barometric Pressure ModBus COM1 Density of Water ModBus COM1 Depth ModBus COM1 External Battery ModBus COM1 ORP mv ModBus COM1 Percent Battery ModBus COM1 ph ModBus COM1 ph mv ModBus COM1 Pressure ModBus COM1 Resistivity ModBus COM1 Salinity ModBus COM1 Specific Conductivity ModBus COM1 Surface Elevation ModBus COM1 Temperature ModBus COM1 Total Dissolved Solids Configure the Units Conversion, if desired. See Section 4.3 of the MultiLogger User's Guide for more information on the Channel Configuration form. AquaTroll 600 Configuration and MultiLogger Integration 10

Once all channels are configured press Accept. The Logger form will display. The program must be now compiled and downloaded to the logger. Press the Start (or Update) button on the Logger form toolbar to compile the program and download it to the logger. You will be asked to confirm. (Figure 16 The location of the Start/Update button in the toolbar of the Logger form) After download completes, the Monitor Mode will start, which can be used to verify correct gage measurements. Data Collection settings must now be configured. Proceed to the next section. AquaTroll 600 Configuration and MultiLogger Integration 11

Configure Data Collection and Import To configure the system to automatically collect the data from the AquaTroll and import it into the MultiLogger database, data collection for the device must be configured. Exit the Logger form. The Network Manager will display. Click on the Gateway level in the Network Configuration. Notice the Database Alias drop-down list, shown in Figure 17. Select the database into which the data will be imported. Note If the edit is disabled, the Gateway List is not configured properly. See Section 3.2 of the MultiLogger User's Guide for more information on the Database Alias and the Data Collection tab. (Figure 17 The MultiLogger Network Manager with Database Alias drop-down highlighted) AquaTroll 600 Configuration and MultiLogger Integration 12

Additional data collection settings are configured from the Data Collection tab. Be sure to check Enable Schedule and Automatic Database Import, which will enable scheduled collection of data from the device and automatic import of the collected data into the selected database. Then configure the Interval for collection and make sure Resume Schedule After Delayed Retries is checked, so that data collection will continue even if a previously scheduled collection failed. (Figure 18 The Data Collection tab for the AquaTroll's datalogger in the MultiLogger Network Manager) Press Save once all settings are configured. MultiLogger can be closed and the data collection will be performed autonomously. Collection status can be viewed at any time by clicking the Data Collection Status tab of the Network Manager. AquaTroll 600 Configuration and MultiLogger Integration 13