Quick Reference. 5973N and 5973 inert Mass Selective Detector Local Control Panel (LCP)

Similar documents
Agilent GC, MS, and ALS

Agilent G1732AA MSD Security ChemStation

Getting Started. Environmental Analysis Software

Agilent G1701DA MSD Productivity ChemStation Software

Agilent 5975 Series MSD

Agilent 5977 MSD Overview

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes. System Information. Agilent Technologies

Agilent 7890GC. Firmware Bulletin. Revision B and A External use only

Agilent G1732AA MSD Security ChemStation

Waters Empower Software A1100 System Configuration Instructions for Ethernet Control

Agilent G1701DA MSD ChemStation

Agilent G1732BA MSD Security ChemStation

Setting up a LAN Instrument Network. Installation Guide

5973N and 5973 inert Mass Selective Detector

Agilent G1701DA MSD Productivity ChemStation Software

SIM Direct Inlet Probe. Agilent 5975 MSD. Agilent 7000 Triple Quadrupole

5973 Network Mass Selective Detectors

Agilent MassHunter GCMS Data Acquisition for the 5977 Series MSD System

Agilent OpenLAB CDS. ChemStation Edition Instrument Configuration. Guide. Agilent Technologies

Agilent G1732AA MSD Security ChemStation

Integration of GC/MSD ChemStation with OpenLAB ECM

Retention Time Locking with the MSD Productivity ChemStation. Technical Overview. Introduction. When Should I Lock My Methods?

Cyclone Microsystems PCIe Chassis Monitor User s Manual P/N CM026

SCI Testers and Remote Communication: Using HyperTerminal

For the configuration of the I/O-modules in a control system an EDS-file is required. The names of the files are as follows:

DART SVP. Software Manual For Web Based User Interface And For Apple ipod touch User Interface. Software Version 4.0.x Document #: 7.5.

SCN 848 Release Notes

Table Of Contents. 1. Introduction... 1

Agilent 6100 Series Quad LC/MS System

Citrix GoToAssist Express User Guide

Administrative Settings

ION MILLING USING PVD 75 SOP. January 2014 START UP

Agilent ChemStation for GC Systems

UHC-HD Software Update Instructions

Windows 2000 Safe Mode

Agilent 7890 Series Gas Chromatograph

50PMA-XXX Sample Test Software for Programmable Attenuator Assemblies

4.2 NORMAL OPERATION NORMAL START UP PROCEDURE

HP G1701BA Productivity ChemStation Software. Installation Manual

AquaConnect. Troubleshooting Guide Residential. TSG-ACHN110d. Copyright 2017 Hayward Industries Inc.

Welcome. Unleash Your Phone

Agilent OpenLAB Chromatography Data System A SR1 - Updates

5975 inert MSD. Chris Sandy European GC-MS Product Specialist Agilent Technologies

CLEAR MODEM (SERIES G) USER GUIDE

Agilent GC/MSD Instructions

Warning 101 Invalid heater power for front detector, inlet, and aux 1 Warning 102 Invalid heater power for back detector, inlet, and aux 2

Agilent 6400 Series Triple Quadrupole LC/MS System

Bridge Cable User s Guide

User s Manual PowerPanel Shutdown Service Graceful Shutdown and Notification service to ensure power protection of your computer

User Manual PDUTracker

Operating Manual. High Performance Liquid Chromatograph. Scientific Equipment Center, Prince Of Songkla University

Using GIGABYTE Notebook for the First Time

Hardware Guide. AB SCIEX QTRAP 5500 System

82341C HPIB Installation and Troubleshooting Agilent ChemStations on Windows 95 and Windows NT

Programming through Web browser:

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes. Agilent 6230B Series TOF LC-MS System Installation Checklist

Fujitsu LifeBook A Series

T E C H N O L O G I E S ENT ESC. User Guide. 1:4 CF or CFAST Duplicator PRO (CFASTD4-2S / CFD4-2S)

Building Agilent GC/MSD Deconvolution Reporting Libraries for Any Application Technical Overview

User s Manual PowerPanel Shutdown Service Graceful Shutdown and Notification service to ensure power protection of your computer

ilo MP Utilities ilo MP

Device Manager. Managing Devices CHAPTER

Silex SD-320AN-03 Serial Server

GC-MS ChemStation Reference Guide. Last Updated July 2014

System Software Update Instructions

CCNA 1 Chapter 2 v5.0 Exam Answers 2013

Service Bulletin SB685. Date: 8/18/2017 TriPac EVOLUTION Communications Update Bulletin Location: TSA Info Central\Service Bulletins

1. Turn TigerStop ON.

NOTICE: This document contains references to Varian. Please note that Varian, Inc. is now part of Agilent Technologies. For more information, go to

Agilent 7820A Gas Chromatograph

T E C H N O L O G I E S. User Guide. 1:1 Duplicator (HDUSAS)

Table of Contents pg " Display pg Cruise Mode pg Map Screen pg Stereo Screen pg Depth Screen pg.

T E C H N O L O G I E S. User Guide. 1:1 HDD Duplicator PRO (HDUSI325)

Agilent G1369B LAN Interface

How to Use imageprograf Firmware Update Tool (Version or later) (Mac OS) Canon Inc.

SonicWALL strongly recommends you follow these steps before installing Global VPN Client (GVC) 4.0.0:

LAN-505 Ethernet Communication

Using the Hart 1620 and Fluke 5020A Thermo-Hygrometers with MET/CAL

Contents for How to Use the Simulator Program

User Manual Electronic Systems Protection, Inc. / Technical Support: / espei.com

Agilent G1701DA GC/MSD ChemStation

WSDA User Guide

How to Run the CASPiE GCMS

Installing Your Multifunction to Your Network via the WPS Push Button Method. PIXMA MG5720 Mac OS

Setting Up Hardware Failover

User s guide for T5_R5 esuf

The Agilent 5975 inert MSD: New Tools for the Forensic Analyst

ReadyNAS OS 6 Desktop Storage Systems Hardware Manual

Agilent OpenLAB CDS. Version 2.2. Release Notes. OpenLAB CDS Release Notes 1

T E C H N O L O G I E S. User Guide. 1:5 Duplicator (HDU5SAS)

Installing the Printer to Your Wireless Network. imageprograf PRO Mac OS

ToroPCTM User s Manual

Residual Gas Analyzers XT Series

IPGpulser. IPGpulser Overview

Agilent Instrument Control Framework (ICF) Update A Driver Update 1 (DU1) Hot Fix 2 (HF2)

Sound Manager 2 User Guide

PF31XX FAMILY FIRMWARE UPDATE GUIDE. v1.2 Revised Nov 29, 2016

SOUND MANAGER 2.0 USER GUIDE SOUND MASKING PAGING AUDIO MASS NOTIFICATION REV

Agilent MassHunter Drug Analysis Mode Using Quantitative Analysis Workflow Guide

Agilent MassHunter Workstation Software 7200 Accurate-Mass Quadrupole Time of Flight GC/MS

Transcription:

Quick Reference 5973N and 5973 inert Mass Selective Detector Local Control Panel (LCP)

Using the Local Control Panel The Local Control Panel (LCP) lets you determine the readiness and performance of the MSD without using the With the 5973N or 5973 inert MSD, the ChemStation may be located anywhere on the site LAN (local area network) and so might not be near the instrument itself. And because the LCP communicates with the ChemStation via the LAN, you can access ChemStation software functions such as tuning and starting a run right from the MSD. Note Only certain functions are available from the LCP; the ChemStation is the full-featured controller for most instrument control operations. Modes of Operation The LCP has two modes of operation: Status and Menu. Status mode requires no interaction and simply displays the current status of the MSD instrument or its various communication connections. If you select [Menu] then [No / Cancel], you will be returned to the Status mode. Menu mode allows you to query various aspects of the GC/MSD and to initiate some actions like running a method or sequence or preparing to vent the system. To access a particular menu option: Press the Menu key until the desired menu appears. Press the Item key until the desired menu item appears. Use one or more of the following keys as appropriate to respond to prompts or select options: Use the Up arrow key to increase the displayed value or to scroll up (such as in a message list). Use the Down key to decrease the displayed value or to scroll down (such as in a message list). Use the Yes / Select key to accept the current value. Use the No / Cancel key to return to the Status mode. After you make your selection or if you cycle through all available menus, the display automatically returns to Status mode. Pressing [Menu] then [No / Cancel] will always display the Status mode. 2

LCP Status Messages The following messages can be displayed on the Local Control Panel to help you quickly determine the status of the MSD system. If the LCP is currently in Menu mode, cycle through the menus to return to Status mode. Note No messages will be displayed if an online instrument session is not currently running on the Press Sideplate A reminder during startup to press the MSD sideplate to ensure an adequate vacuum seal. Loading OS The operating system of the instrument controller is being initialized. Loading MSD Firmware The MSD s firmware is being initialized. The following messages alternately appear on the LCP if the MSD does NOT complete its bootup sequence properly: Server not Found Check LAN Connection Seeking Server Bootp Query xxx These messages indicate that the MSD has not received its unique IP address from the CAG Bootp Server. If the messages persist after you have logged onto your account on the ChemStation, consult the Troubleshooting section of the Software Installation manual. ChemStation Loading <timestamp> The MSD Productivity ChemStation software is starting up. MS locked by <computer name> MS parameters can only be changed from the Executing <type>tune A tuning procedure is in progress (type = QuickTune or Autotune). Loading Method <method name> Method parameters are being sent to the MSD. Method Loaded <method name> Method parameters have been sent to the MSD. Run: <method> Acquiring <datafile> A run is in progress; data is being acquired to the designated data file. <method> Complete <timestamp> The run and subsequent data processing are done. The same message appears even if the run was terminated prematurely. Instrument Available <timestamp> The MSD Productivity ChemStation software is not running. To view system status during startup 1 The following messages are displayed on the Local Control Panel display during startup. Press sideplate Loading OS Press sideplate Loading MSD Firmware 2 Continue to press the sideplate of the MSD until the MSD Ready message appears. This helps the instrument to pump down more quickly. 3

LCP Menus To access a particular menu option, press [Menu] until the desired menu appears, then press [Item] until the desired menu item appears. Note Many menu items, especially on the ChemStation, MS Parameters, and Maintenance menus, have no effect when the instrument is acquiring data. ChemStation Menu Run Method Run Sequence Run Current Tune Displays the current method name and starts an analysis. Displays the current sequence and starts a sequence. Displays the current tune file and starts an autotune (EI mode only, CI tune must be started from the ChemStation). # of messages Displays the number of messages and the text of the most recent message. Use the arrow keys to scroll through previous messages (up to 20). Release ChemStation Connection status Disassociates the ChemStation from the MSD. Displays the LAN connection status for the MSD. Remote = connected to a ChemStation online session Maintenance Menu Prepare to vent Pump down MS Parameters Menu Turbo pump speed Pump status Ion source temp, o C Mass filter temp, o C Electron multiplier Electron energy, ev Emission current, µa Reminds you to shut down the GC then vents the instrument when [Yes] is pressed. Initiates a pumpdown sequence. Displays the turbo pump speed Displays the diffusion pump status (on/off) Displays and sets the ion source temperature Displays and sets the mass filter temperature Displays and sets the EM voltage Displays and sets the electron energy (ev) Displays and sets the emission current (µa) Note MS parameters can t be set from the LCP while an online session is running on the Name of instrument Local = not connected to a ChemStation online session Displays the name of the instrument if connected to a ChemStation online session 4

Network Menu MSD IP Gateway IP IP ChemStation MAC MS controller Ping ChemStation Ping gateway IP GC Ping GC Displays the IP address for the MSD. Displays the gateway IP address for the MSD. Displays the IP address for the Displays the address of the LAN/MS control card. Checks communication with the Checks communication with the gateway. Displays the IP address for the GC. Checks communication with the GC. Controller Menu Self-test, short Self-test, long Reboot controller Display diagnostics HTTP check Initiates a short version of the self-test (will fail if an online ChemStation is running). Initiates a comprehensive self- test (will fail if an online ChemStation is running). Starts the LAN/MS control card. Initiates a diagnostic test of the two line display. Checks the status of the HTTP server. Version Menu Control firmware Operating system Front panel Displays the MSD firmware revision. Displays the MSD operating system revision. Displays the revision of the local control panel (LCP). 5

LCP Tasks To query system status 1 Make sure that an online session is running on the 2 Press [Menu] on the Local Control Panel (LCP) until the ChemStation menu appears. 3 Press [Item] on the LCP until the Messages item appears. This item displays the total number of messages and the text of the most recent message. 4 Use the arrow keys to scroll through previous messages. Up to 20 messages can be displayed. Note No messages will be displayed if an online instrument session is not currently running on the To start an autotune 1 Make sure that an online session is running on the 2 Press [Menu] on the Local Control Panel (LCP) until the ChemStation menu appears. 3 Press [Item] on the LCP until the Run Tune item appears. After a few seconds, the name of the current tune file is displayed. 4 Press [Yes]. Note The autotune will not be performed if an online instrument session is not currently running on the To start a run 1 Make sure that an online session is running on the 2 If a run is already in progress, wait until it is done or stop it from the If you are not sure, look for the Run:<method> message as described in To query instrument status. 3 Press [Menu] on the Local Control Panel (LCP) until the ChemStation menu appears. 4 Press [Item] on the LCP to display the Run Method item. After a few seconds, the name of the current method is displayed. 5 Press [Yes]. If automatic injection is selected in the method, the sample is injected when the GC temperatures are stable. If manual injection is selected in the method, wait until the Not Ready light on the GC goes off, then inject the sample and press [Start] on the GC. Note The run will not start if an online instrument session is not currently running on the The current data file name should have a number in it (e.g., mydata001.d) to prevent subsequent overwrites. The number in the file will be incremented for the next injection. To start a sequence 1 Make sure that an online session is running on the 2 If a run is already in progress, wait until it is done or stop it from the If you are not sure, look for the In Run message as described in To query system status. 3 Press [Menu] on the Local Control Panel (LCP) until the ChemStation menu appears. 4 Press [Item] on the LCP until the Run Sequence item appears. After a few seconds, the name of the current sequence is displayed. 5 Press [Yes] to start the current sequence. Note The sequence will not start if an online instrument session is not currently running on the 6

To prepare to vent the instrument 1 Press [Menu] on the Local Control Panel (LCP) until the Maintenance menu appears. 2 Press [Item] on the LCP to display the Prepare to vent item. You are prompted to turn off the GC. Refer to the 5973N or 5973 inert Hardware Manual for complete instructions on venting the MSD. 3 After you have turned off the GC as described in the 5973N or 5973 inert Hardware Manual, press [Yes] on the Local Control Panel on the MSD to vent the instrument. Note You cannot initiate the vent procedure when an online session is running on the Either exit the online session first, or disassociate the instrument from the ChemStation using the Release ChemStation item on the Controller menu on the LCP. If an online session is running on the ChemStation, the message the instrument is remote is displayed along with the name of the instrument. If no online session is running, the message the instrument is local is displayed. You can view the IP addresses of the MSD, the GC, and the gateway. The IP address of the ChemStation will only be reported if an online session is running on the You can verify communication with the GC, the ChemStation, and the gateway with the Ping menu items. The ChemStation will appear not to be connected if no online ChemStation session is running. To pump down the instrument 1 Press [Menu] on the Local Control Panel (LCP) until the Maintenance menu appears. 2 Press [Item] on the LCP to display the Pumpdown item. 3 Press [Yes] on the Local Control Panel on the MSD to pump down the instrument. Note You cannot initiate the pumpdown procedure when an online session is running on the Either exit the online session first, or disassociate the instrument from the ChemStation using the Release ChemStation item on the Controller menu on the LCP. To query network status 1 Press [Menu] on the Local Control Panel (LCP) until the Network menu appears. 2 Press [Item] on the LCP repeatedly to scroll through the various network connections. To reboot the MSD Use this procedure to restart the MSD; do not use the on/off switch on the MSD unless specifically instructed to do so in a maintenance procedure. 1 Press [Menu] on the Local Control Panel (LCP) until the Controller menu appears. 2 Press [Item] on the LCP until the Release ChemStation item appears. 3 Press [Yes]. 4 Press [Menu] on the LCP until the Controller menu appears. 5 Press [Item] on the LCP until the Reboot controller item appears. 6 Press [Yes]. 7

Document History 7/03 Copyright 2003 Agilent Technologies Printed in U.S.A. 7/03