GREISINGER electronic GmbH D Regenstauf, Hans-Sachs-Straße 26 phone: / , fax: / ,

Similar documents
GREISINGER electronic GmbH D Regenstauf, Hans-Sachs-Straße 26 phone: / , fax: / ,

Manometer. (Over- /Under- And Difference Pressure) Operating Manual GDH

GREISINGER electronic GmbH D Regenstauf, Hans-Sachs-Straße 26

Digital Quick-Response Thermometer

GREISINGER electronic GmbH D Regenstauf, Hans-Sachs-Straße 26

GMH Operating Manual for Digital Quick-Response Thermometer

GMH to 1300 mbar absolute Version 4.7

Hygro-/Thermometer. Operating Manual GFTH 95

GREISINGER electronic GmbH. D Regenstauf, Hans-Sachs-Straße 26. T-Logg 100. T-Logg 100 E

GREISINGER electronic GmbH D Regenstauf, Hans-Sachs-Straße 26

Universal interface adapter for GMSD / GMxD - sensor

Operating Manual Digital Quick Response Thermometer GMH 3210 For Thermocouple Probes Type J, K, N, S, T from Version 1.0

GREISINGER electronic GmbH. D Regenstauf, Hans-Sachs-Straße 26. T-Logg 120 W -... T-Logg 120 K -...

Hand-held measuring device with alarm function

GHM Messtechnik GmbH Standort Greisinger Hans-Sachs-Str. 26 D Regenstauf +49 (0) 9402 / (0) 9402 /

GMH B to mbar relative -7.5 to mbar relative Version 6.5

Precision thermometer

Digital Precision Hand-held Measuring Device

Hand-Held Pressure Indicator Model CPH6200-S1 Model CPH6200-S2 (2-channel version)

GREISINGER electronic GmbH D Regenstauf, Hans-Sachs-Straße 26

Hand-Held Pressure Indicator

Precision digital pressure gauge Model CPG1500

Portable low-pressure controller Model CPC2000

Manual Differential pressure meter PCE-910 / PCE-917

Precision hand-held pressure indicator Model CPH6400

Operating Instructions Differential Pressure Transducer PS27

Intrinsically safe Hand-Held Pressure Indicator (ATEX-version) Model CPH62I0

Instruction Manual P82R Differential Pressure Transducer with Root- Extracted Characteristic Curve

PRESSURE. Handheld instrument. Handheld instrument. Display / Controller. Logger / EASYBus. Transmitter. Temperature probe. Alarm / Protection

SX100A. Features. Typical Applications. Description

CalibratorUnit Models CPU6000-W, CPU6000-S, CPU6000-M

Operating Instructions for Hand-held Conductivity Measuring Units. Model: HND-C110

SX15GD2. Features. Typical Applications. Description

High-Precision Pneumatic Pressure Controller Series CPC8000-L/-X

SX01DN. Features. Typical Applications. Description

ProcessCalibrator Model CPH6000

Operating manual. GTL - Configuration tool. Please keep the manual for future use.

Precision digital pressure gauge Model CPG1500

Operating manual GTL 142 GTL 152. Temperature probe. Please keep the manual for future use.

Digital Precision Hand-held Measuring Device

Pneumatic precision pressure controller Model CPC6000

Hand-Held Thermometer Model CTH 6200

Operating Instruction PPC-04 /2

Precision pressure indicator 1- or 2-channel version Model CPG2500

GIR 230 TC Version 1.1

Pneumatic precision pressure controller Model CPC6000

Instruction Manual PU/PI and PS10 Differential Pressure Transducers

Pneumatic High-Speed Pressure Controller Model CPC3000

T2314, T2414. Instruction Manual

M203 Altimeter / Indicated Air Speed Tester USER S MANUAL

USER GUIDE. Heavy Duty Differential Pressure Manometer. Model HD700

Pneumatic precision pressure controller Model CPC6000

Hand-held thermometer Model CTH6200

Precision digital gas density indicator Model GDI-100-D

CONTENT. Check up. Brief. 1.Before use. 2.Operation. 3.Others

Operating Instructions for Handheld Pressure Measuring Devices with Integrated Pressure Sensors. Model: HND-P236

350 SMART MANOMETER OPERATING INSTRUCTIONS

ProcessCalibrator Model CPH6000

Portable pressure calibrator Model CPH7600

Instruction manual Air sampling device

Pressure Transmitter and Reader of the PT-RF Series Manual

Manometers 2PSI, 5PSI, 15PSI, 30PSI, 100PSI , , ,

T2314, T2414. Instruction Manual

Precision thermometer

DS 96/48 PK. Digital indicator. Manual 42/33-90 EN Rev. 00 DIM

GB/UK. Operating manual. Data logger EBI USB

DMT PERFORMING UNDER PRESSURE DMT. Applications. Special features. Description 1/9

Pneumatic Precision Pressure Controller Model CPC6000

English. Instruction and operation manual S 505. Dew point meter (portable)

User Guide. Heavy Duty Differential Pressure Manometer. Model HD750

Operating instructions. Standstill monitor A / / 2011

High-end pressure controller Model CPC8000

Air data test indicator Model CPA2501

Operating Instructions Bedienungsanleitung Mode d emploi Gebruiksaanwijzing Istruzioni per l uso Instrucciónes para el manejo HPM110

Differential pressure measuring instrument

testo 175 Data loggers Instruction manual

DMT PERFORMING UNDER PRESSURE DMT. Applications. Special features. Description 1/9

GREISINGER electronic GmbH D Regenstauf, Hans-Sachs-Straße 26

Pneumatic precision pressure controller Model CPC6000

User Manual. Heavy Duty Differential Pressure Manometer. Model SDL720. Additional User Manual Translations available at

Intrinsically safe batch controller Batching Master 110i

SC SENSOR COMMUNICATOR MANUAL (ENGLISCH)

InBin-P Pressure switches from 5 Pa Pa

GHM Messtechnik GmbH Standort Greisinger Hans-Sachs-Str. 26 D Regenstauf +49 (0) 9402 / (0) 9402 /

CMC III Temperature/Humidity Sensor

Hand-held thermometer Model CTH6200

Operating Instructions PRO-AB /6.15. Leakage Current Measuring Adapter

Operating instructions AS-i SmartLine module AC3200 AC /00 06/2016

LEX 1 Ei. II 2 G EEx ia II C T6 PTB 05 ATEX 2012 X. Highly Precise, Intrinsically Safe Digital Manometer for use in Hazardous Applications.

Serviceman SensoControl

Original operating instructions Fail-safe inductive sensor GF711S / / 2013

Assembly and Operating Instructions Door Control

GHM Messtechnik GmbH Standort Greisinger Hans-Sachs-Str. 26 D Regenstauf +49 (0) 9402 / (0) 9402 /

353 PRECISION ALTIMETER TESTER OPERATING INSTRUCTIONS

TA400 OPERATING MANUAL DYNAMIC PRESSURE ANEMOMETER TRT-BA-TA400-TC-002-EN

M-610 USER MANUAL PLEASE TAKE THE TIME TO READ THESE INSTRUCTIONS BEFORE STARTING TO USE THE SCALES

Operating Instructions for Differential Pressure Sensor. Model: PMP

Electronic Temperature Controller. Instruction Manual Version

Digital thermo-hygro-barometer with external probe Manual for use

Industrial pressure controller Model CPC4000

Transcription:

H55.0.31.6C-04 Operating Manual Vacuum- / Baro- / Manometer from Version 1.1 GDH 200-14 GREISINGER electronic GmbH D - 93128 Regenstauf, Hans-Sachs-Straße 26 phone: +49 9402 / 9383-0, fax: +49 9402 / 9383-33, email: info@greisinger.de

H55.0.31.6C-04 Operating Manual GDH 200-14 page 2 of 8 Content 1 INTENDED USE...3 2 GENERAL ADVICE...3 3 SAFETY INSTRUCTIONS...3 4 OPERATION NOTES...3 5 OPERATION...4 5.1 DISPLAY ELEMENTS...4 5.2 PUSHBUTTONS...4 6 STARTING...4 7 PRESSURE CONNECTION...4 8 MIN-/MAX- VALUE MEMORY...5 9 ZERO-FUNCTION (RELATIVE MEASURING)...5 10 GENERAL INFORMATION ON ABSOLUTE PRESSURE MEASUREMENT...5 11 CONFIGURATION OF THE DEVICE...6 12 ADJUSTING OF THE INSTRUMENT...7 13 NOTES TO THE CALIBRATION SERVICES...7 14 ERROR AND SYSTEM MESSAGES...7 15 SPECIFICATION...8 16 DISPOSAL NOTES...8

H55.0.31.6C-04 Operating Manual GDH 200-14 page 3 of 8 1 Intended use The device measures the absolute pressure of air or non corrosive / non ionizing gases. Either the ambient air pressure can be measured or the measuring pressure is connected with a suitable tube to the connecting nozzle. Relative measurements can also be realized via the integrated Zero-function. Field of application Barometric measurements (weather) Vacuum measurement (down to 1 mbar abs) Overpressure measurements pneumatics, tyre pressure etc. The pressure connection has to be done adequately! At higher pressures the connection tube has to be locked. Please refer to chapter Pressure Connection. The safety instruction (see chapter 3) have to be observed. The device must not be used for purposes and under conditions for that the device had not been designed. The device must carefully dealt with and has to be used according to the specifications (do not throw, knock, etc.). It has to be protected against dirt. 2 General advice Read through this document attentively and make yourself familiar to the operation of the device before you use it. Keep this document in a ready-to-hand way in order to be able to look up in the case of doubt. 3 Safety instructions This device has been designed and tested in accordance to the safety regulations for electronic devices. However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures and special safety advises given in this manual will be adhered to when using it. 1. Trouble-free operation and reliability of the device can only be guaranteed if it is not subjected to any other climatic conditions than those stated under Specification. Transporting the device from a cold to a warm environment condensation may result in a failure of the function. In such a case make sure the device temperature has adjusted to the ambient temperature before trying a new start-up. 2. Whenever there may be a risk whatsoever involved in running it, the device has to be switched off immediately and to be marked accordingly to avoid re-starting. Operator safety may be a risk if: - there is visible damage to the device or the device is not working as specified - the device has been stored under unsuitable conditions for a longer time In case of doubt, please return device to manufacturer for repair or maintenance. 3. Warning: Do not use this product as safety or emergency stop device, or in any other application where failure of the product could result in personal injury or material damage. Failure to comply with these instructions could result in death or serious injury and material damage. 4 Operation notes If the symbol "LOBAT is displayed at the left side of display, the battery is weak, measuring can be continued for a short period. If BAT is displayed in the main display the battery is used up and needs to be replaced. Measuring is no more possible. The battery has to be removed, when storing device above 50 C. Hint: We recommend removing the battery if device is not used for a longer period of time! Risk of leakage!

H55.0.31.6C-04 Operating Manual GDH 200-14 page 4 of 8 5 Operation 5.1 Display elements 5.2 Pushbuttons 1: Main display Display of current measuring value or of minimal measured value ( Lo blinking) or maximal value ( Hi blinking) LOBAT Indicates low battery on/off: mode: zero: on/off switch, press key long to switch device off press short: changes between actual measuring value, the minimal measured value ( Lo blinking) and maximal measured ( Hi blinking) value. press >2s: reset Min and Max value press >2s: activates / deactivates the Zero-function 6 Starting Switch the device on with button. After the segment test ( ) the device displays some information to its current configuration:, if offset correction is active, if scale corrections active Unit of meas. value: nbar (=mbar) bar PSI nnhg (=mmhg), if sea-level correction is active P.OFF, if auto power-off is active Afterwards the device is ready for measuring. 7 Pressure Connection The pressure is connected to the instrument via the front side pressure port. This is optimised for the connection of 6x1 mm (4 mm inner diameter) tubes. Suitable tubes are: PVC up to 5 bar rel. PE up to 10 bar rel. PUR up to 9 bar rel. At pressures above 2 bar rel. the tube should be locked e.g. by means of a cable strap or a hose clamp. PUR tube and hose clamp Up to 2 bar rel. Above 2 bar rel: lock with clamp

H55.0.31.6C-04 Operating Manual GDH 200-14 page 5 of 8 8 Min-/max- value memory Display MIN-value (Lo): Press button mode shortly Display changes between Lo and min-value Display MAX-value (Hi): Press mode shortly again Display changes between Hi and max-value Display actual value again: Press mode shortly again Actual value is displayed Delete MIN/MAX: Press button mode for 2 s MIN&MAX are deleted. CLr (Clear) is displayed shortly The Min-/max- display is automatically set to actual value display after 10 s. The Min-/max- values are deleted if the device is switched on after it had been turned off. 9 Zero-function (relative measuring) With help of the Zero-function relative measurements can be realized: press zero -button for 2s. NULL is displayed shortly and the displayed value is set to 0. Press again zero for 2s to display the absolute value again. Note: Setting and resetting the Zero function will delete your min-/max- value memory. Field of application: e.g. tyre pressure measurement, pneumatic measuring up to 10 bar rel. 10 General information on absolute pressure measurement The device measures the absolute pressure. This must not be confused with the pressure at sea level display by weather stations. There the pressure loss due to height above sea level is deducted. The device is also able to perform this sea-level correction on the displayed air pressure. Therefore (Sea Level correction) has to be activated (= on ) and the current height above sea level (ALTI = Altitude in [m]) has to be entered at the configuration.

H55.0.31.6C-04 Operating Manual GDH 200-14 page 6 of 8 11 Configuration of the device Follow this instructions to configure the functions of the device: Switch the device off. Press mode button and keep it pressed. Turn the device on. Release the mode button after the segment test ( ) finished. The first parameter P.OFF is displayed. If a parameter should be changed press key up or down ( parameter will be shown and can be changed via up/down keys. ) the present setting of the The parameter setting is confirmed by pressing. Jump to the next parameter with. Parameter value information button buttons P.OFF UNIT ALTI INIT Auto Power-Off (turn-off delay) factory setting: 20 min. Auto Power-Off (turn-off delay) in minutes. If no key is pressed for the 1 120 time adjusted in this parameter, the device is automatically switched off to safe battery power. (adjustable range 1 to 120 min) OFF automatic power-off is deactivated (continuous operating) Unit and range of display factory setting: nbar=mbar PSI 0 160.00 PSI, resolution 0.02 PSI nnhg nbar bar 0 8250 mmhg, resolution 1mmHg 0 11000 mbar, resolution 1 mbar 0 11.000 bar, resolution 0.001 bar Sea-level correction factory setting: off Sea-level correction off OFF ON Sea-level correction on Altitude input for sea-level correction factory setting: 340-500 9000-500 9000 m selectable Restore to factory settings Settings are kept NO YES Restoring device to factory settings Pressing Please note: again stores the settings, the instruments restarts (segment test) If there is no key pressed within the menu mode within 2 minutes, the configuration will be cancelled, the entered settings are lost!

H55.0.31.6C-04 Operating Manual GDH 200-14 page 7 of 8 12 Adjusting of the instrument To adjust the instruments according to your needs proceed like follows: Switch instrument off. Press zero button and keep it pressed. Turn the device on. Release the mode button after the segment test ( ) finished. The first parameter SCL is displayed. If a parameter should be changed press key up or down ( parameter will be shown and can be changed via up/down keys. ) the present setting of the The parameter setting is confirmed by pressing. Jump to the next parameter with. Parameter value information button buttons Offset of sensors factory setting: off OFF -5.0 +5.0 Selectable -5.0 +5.0 Scale correction factory setting: off = 0% OFF -5.00 +5.00 Adjustment in % Pressing Please note: again stores the settings, the instruments restarts (segment test) If there is no key pressed within the menu mode within 2 minutes, the adjusting will be cancelled, the entered settings are lost! 13 Notes to the calibration services Calibration certificates - DKD-certificates - other certificates: If device should be certificated for its accuracy, it is the best solution to return it with the referring sensors to the manufacturer. (please state desired test values, e.g. 1bar abs) Only the manufacturer is capable to do efficient recalibration if necessary to get results of highest accuracy! 14 Error and system messages ERR. 1 measuring range has been exceeded ERR. 2 measured value has fallen below permitted range ERR. 3 display range has been exceeded (>19999) ERR. 4 measuring value has fallen below displayable range (< -1999) ERR. 7 System fault - the device has detected a system fault (defective or far outside allowable ambient temperature range) --- Sensor error or value could not be calculated If the symbol LOBAT is displayed at the left side of display, the battery is weak, measuring can be continued for a short time. If BAT is displayed in the display the battery is used up and needs to be replaced. Measuring is no more possible.

H55.0.31.6C-04 Operating Manual GDH 200-14 page 8 of 8 15 Specification Measuring range air pressure: Measuring media: Max. overload: Accuracy: (±1 digit) Pressure connection: 0... 11000 mbar abs., resolution 1 mbar 0... 11.000 bar abs., resolution 0.001 bar 0... 8250 mmhg abs., resolution 1 mmhg 0... 160.00 PSI abs., resolution 0.02 PSI non corrosive gases 13 bar abs. +/- 3 mbar or +/- 0.10 % typ., whichever is higher (at nominal temperature 25 C) +/- 0.30 % FS temperature influence 0 to 50 C connecting nozzle made of nickel-plated brass on top of the device, for tubes 6x1mm (4mm innerø), approx. 11mm projecting 1 measurement per second digital offset and scale adjustment for air pressure measurement Measuring frequency: Offset and scale: Min-/max-value memory: min- and max- value are stored Sea level correction: The displayed value can be referenced to sea level. Therefore the current height above sea level has to be entered. Zero-function: relative measurement: the displayed value is set to zero Display: approx. 13 mm high, 4½-digit LCD display Pushbuttons: 3 membrane keys for on/off, min-/max-value request, offset adjustment Working conditions: -25 to 50 C; 0 to 80 % RH (non condensing) Storage temperature: -25 to 70 C Power supply: 9V-battery, type 6F22 (included in scope of delivery) Current consumption: ca. 60 µa (Operating time with standard zinc-carbon battery >7500 hours Display for weak battery: automatically if battery consumed "BAT", warning: "LOBAT" Auto-off function: Housing: Dimensions: Weight: EMC: If activated: device will be automatically switched off if not operated for longer time (selectable 1..120min) impact resistant ABS housing, front side IP65 approx. 106 x 67 x 30 mm (H x W x D), without connecting nozzle approx. 135 g incl. battery The device corresponds to the essential protection ratings established in the Directives of the European Parliament and the council on the approximation of the laws of the member states relating to the electromagnetic compatibility (2004/108/EG). Additional error: <1% 16 Disposal notes Dispense exhausted batteries at destined gathering places. This device must not be disposed as residual waste. To dispose this device, please send it directly to us (adequately stamped). We will dispose it appropriately and environmentally friendly.