Operating Instructions PRO-AB /6.15. Leakage Current Measuring Adapter

Similar documents
Operating Instructions SECUTEST SI /12.10

Operating Instructions. LSP-KONSTANTER Series 32K Laboratory Power Supply 8/10.13

Operating Instructions. SECU-cal 10. Calibration Adapter for Testers per DIN VDE / /5.10

Energy /10.13

Installation Instructions VIPA FM355 / R355. 4/8-Channel Controller Module for Siemens S7-300 and Vipa System 300V /10.

PROFITEST PV Peak Power Measuring Instrument and Curve Tracer for PV Modules and Strings (Measurement at Capacitive Load)

PROFITEST Tester for DIN EN and VDE 0113

SECUTEST S2 N+w Test Instrument with Automatic Test Sequences for IEC and Draft IEC 62638

PROFITEST Tester for DIN EN and VDE 0113

PROFITEST INSTALLATION

SECUTEST S2 N+w Test Instrument with Automatic Test Sequences for IEC and Draft IEC 62638

PROFITEST Tester for DIN EN and VDE 0113

PROFITEST. Testing Protective Measures MASTER TESTER INSTALLATION

Demonstration Case BENNING DB 1

TETRIS 2500 High Impedance Active Probe. Instruction Manual

CLAMP MULTIMETER METRACLIP 85 ENGLISH. Start Guide

TG basic 1. Appliance tester for testing protective measures in accordance with DIN VDE (DIN EN 62638)

PROFITEST 204 For testing the safety of electrical machinery equipment

DINALOG A 144 x 36 A1400 Light-Strip Indicator

7032 Digital-Analog Multimeter

SYSKON P1500 Computer Controlled Laboratory Power Supply

SECULIFE ST and SECULIFE ST HV Test Instrument for DIN EN 60601/60335/60950/61010, DIN VDE 0700/ and IEC (VDE )

PROFITEST 204 Tester for DIN EN and VDE 0113

SECULIFE ST and SECULIFE ST HV Test Instrument for DIN EN 60601/60335/60950/61010, DIN VDE 0700/ and IEC (VDE )

Solar Installation PV100 Operating Instructions

MC 11 EB-2 Power supply cabinet with external bus, AC version

PHOENIX CONTACT - 07/2006

700 Series 200 Amp Clamp Meters

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL:

A2000 Multifunctional Power Meter

Firmware Update Tool Software Program for Updating the Firmware for the SECUTEST BASE, BASE10 and PRO Test Instruments

KONSTANTER SPL Series Programmable Electronic Load

PCT2000. User Manual. Precision current transducer 2000 A

SECUTEST SII + Test Instrument for Measurements per BGV A3 and Operational Safety Regulation

PKP Prozessmesstechnik GmbH. Borsigstrasse 24. D Wiesbaden-Nordenstadt. Tel: / Fax: / Operating manual PSA06

PCI-PIO Digital I/O card (PCI)

LED Driver Linear / area dimming

PHOENIX CONTACT - 01/2007

Model INSTRUCTION MANUAL DIGITAL MULTIMETER

MINI-PS AC/10-15DC/8

Extract from the online catalog

LED Driver Linear / area dimming

TELARIS 0701/0702 easy TELARIS 0701/0702

LED Driver Linear / area fixed output

LED Driver Linear / area fixed output

S-14 S-14. Compact Digital Multimeter. Compact Digital Multimeter

Electronic Temperature Controller. Instruction Manual Version

DranISO 5000D. Advanced Test Equipment Rentals ATEC (2832)

KHT 6000 C HV-Probe Calibrator. Instruction Manual

LED Driver Compact fixed output

GMH Operating Manual for Digital Quick-Response Thermometer

C.A 6165 Appliance multitester

PHOENIX CONTACT - 01/2010. Features. DANGER OF EXPLOSION! Only remove equipment when it is disconnected and not in the potentially explosive area.

POCKET MULTIMETER Model No: MM18

Model Number Structure

Operating Instructions

SECULIFE ST and SECULIFE ST HV Test Instrument for DIN EN 60601/60335/60950/61010, DIN VDE 0700/ and IEC (VDE 0751)

Operating instructions AS-i SmartLine module AC3200 AC /00 06/2016

LOW WEIGHT, SMALL SIZE,

Dual Channel Monitor for Measurement of Speed and Ratio with SIL1 requirements Series D224.xx

PRODUCT SPECIFICATION. PC Board Relays. Series.

OPERATING INSTRUCTIONS 7 SERIES STATIC GENERATORS

TETRIS 1000 / High Impedance Active Probe Order-No: Features:

Power supply module, bus connection module

METREL test and measurement accessories: 3-phase AktivGT / Machine adapter A1322 Instruction manual Version 1.0, Code no.

IDEAL INDUSTRIES INC. TECHNICAL MANUAL MODEL:

User's Guide. MiniTec TM Series Model MN25 MultiMeter

ILBPB24DO32. Inline Block IO Module for PROFIBUS With 32 Digital Outputs. AUTOMATIONWORX Data Sheet 6889_en_04. Description

ILBIB24DO16-DSUB. Inline Block IO Module for INTERBUS With 16 Digital Outputs; Bus Connection via D-SUB Connectors

Operating manual. GTL - Configuration tool. Please keep the manual for future use.

PROFiTEST C Test Instrument per DIN VDE 0100

USB-PIO Digital I/O interface (USB)

* * Agilent Power Distribution Unit (PDU) Installation Guide

MINI MCR-SL-UI-2I(-SP)(-NC)

OPERATOR S INSTRUCTION MANUAL DIGITAL MULTIMETER

Operating Instructions Electronic Circuit Protector ESX10-TC-DC 12 V

SECUSTAR FM / FM+ and SECULIFE SB Modular Test System

QUINT-BUFFER/24DC/24DC/40

PGT 120. Personnel Grounding Tester. Wrist Strap and Footwear Tester. Users's Manual

PROFITEST 204 +, 204 L+, 204HP, 204HV Test Instrument for EN / DIN VDE 0113

MARTINDALE INSTRUCTIONS MM35 DIGITAL MULTIMETER ELECTRIC. Trusted by professionals

Operating Manual. English. Digital-Analog Converter DACON8-UMB.

PROFiTEST PSI-BC / SI-BC

Industriefunkuhren. Technical Manual. Signal Converter. for DIN Rail Mounting Series 4800xx-yy ENGLISH

Power Supply, Primary Switch Mode, Narrow Design MINI-PS AC/24DC/1

Digital Multimeter User's Manual

Extension module for lighting control

PHOENIX CONTACT - 08/2009. Features. DANGER OF EXPLOSION! Only remove equipment when it is disconnected and not in the potentially explosive area.

SECUTEST BASE(10) and PRO Tester for Measuring the Electrical Safety of Devices per VDE , IEC and IEC

SPL KONSTANTER Programmable Electronic Load

LED Driver Linear / area fixed output

LED Driver Linear / area fixed output

ATS-16 HV USER MANUAL. Automatic Transfer Switch 16A / 230Vac V090318

Instruction Manual IF2004/USB. 4-Channel RS422/USB Converter

Operating instructions. Switching amplifier DN0210 DN / / 2015

EARTH RESISTANCE & RESISTIVITY TESTER

PCT200. User Manual. Precision current transducer 200 A

Operating instructions. Standstill monitor A / / 2011

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER

MTP INSTRUCTION MANUAL

Transcription:

Operating Instructions PRO-AB Leakage Current Measuring Adapter 3-349-621-15 3/6.15

Table of Contents Page 1 Scope of delivery...2 2 Safety Precautions...2 3 General...3 4 Preparing for Measurement...4 5 Performing Measurement...5 6 Self-Test Function for Testing the Adapter... 6 6.1 Executing the Test Function at the Instrument...7 7 Measuring Instrument Requirements...8 8 Remedying Malfunctions...8 9 Characteristic Values...8 10 Return and Environmentally Sound Disposal...9 11 Repair and Replacement Parts Service Calibration Center and Rental Instrument Service...10 12 Product Support...10 1 Scope of delivery 1 Leakage current measuring adapter 1 Set of operating instructions 2 Safety Precautions The leakage current measuring adapter is a measuring device in accordance with IEC 62353 (VDE 0750, part 1). It may only be used for its intended purpose! Permissible values for continuous leakage and patient auxiliary current must be taken into consideration. For safety reasons, the adapter must be checked for correct functioning before and after measurements (e.g. by means of the self-test function included in the test instrument, or perhaps by measuring ohmic input and output impedance). 2 GMC-I Messtechnik GmbH

Input impedance With open output When fuse is OK: 1020 Ω ±3% When fuse if blown: infinite Output impedance Range: 10:1 With open input: 10.1 kω With closed input: 10 kω Range: 1:1 With open input: 11.0 kω With closed input: 10 kω Opening of Adapter / Repair The adapter may be opened only by authorized service personnel to ensure the safe and correct operation of the equipment and to keep the warranty valid. Even original spare parts may be installed only by authorized service personnel. In case the adapter was opened by unauthorized personnel, no warranty regarding personal safety, measurement accuracy, conformity with applicable safety measures or any consequential damage is granted by the manufacturer. 3 General In order to measure contact voltage in accordance with DIN VDE 0107, part 10, as well as continuous leakage and patient auxiliary current per IEC 62353 (VDE 0750, part 1) / IEC 601-1 / EN 60 601-1:2006 (Medical electrical equipment General requirements for basic safety), the PRO-AB leakage current measuring adapter must be used as an accessory for the PROFITEST MXTRA or SECULIFE IP test instrument. When performing leakage current measurements, the adapter is plugged into the measurement inputs at the left-hand side of the PROFITEST MXTRA or SECULIFE IP (2-pole current clamp input and probe input) as shown in figure 1. As specified in the standards listed above, current values of up to 10 ma may be measured with this adapter. In order to be able to fully cover this measuring range using the measurement input provided on GMC-I Messtechnik GmbH 3

the test instrument (2-pole current clamp input), the measuring instrument is equipped with range switching between transformation ratios of 10:1 and 1:1. In the 10:1 range, voltage dividing takes place at the same ratio. The internal measurement resistor (1.0 kω) is protected by an upstream device fuse in order to prevent overloading (see Characteristic Values on page 8. The adapter s frequency response is determined by a low-pass filter (10 kω / 0.015 μf) with a cut-off frequency of 1kHz (20dB attenuation). 4 Preparing for Measurement Before performing measurements, the adapter s measurement outputs must be plugged into the measurement inputs at the left-hand side of the PROFITEST MXTRA or SECULIFE IP (2-pole current clamp input and probe input). Figure 1: Leakage Current Measuring Adapter at the PROFITEST MXTRA Either of the leakage current measuring adapter s inputs is connected to reference earth (e.g. safe earth electrode / equipotential bonding) via a measurement cable. The metallic housing (accessible part) of the device under test is contacted with a test probe or alligator clip which is connected to the other input by means of a second measurement cable. 4 GMC-I Messtechnik GmbH

Device Under Test Leakage Current Measuring Adapter Ground / Equipotential Bonding PROFITEST MXTRA or SECULIFE IP Figure 2: Test Setup for Leakage Current Measurement 5 Performing Measurement Measurement is performed with the rotary switch in the EXTRA position and with EXTRA measurements selected as an option in the submenu. The leakage current measuring adapter is equipped with a 1 kω measurement resistor and a low-pass filter with a cut-off frequency of 1 khz. The measuring range is selected by means of a switch for the 10:1 and 1:1 ranges. The filter function is always active and cannot be deactivated with the current device design. After completing the test setup as described in section 4, measurement can be started at the test instrument.! Attention! The test plug should be located in the storage slot during leakage current measurement. Under no circumstances may the test plug be connected with any system components (including PE / ground potential). First of all, the rotary selector switch at the test instrument is set to the EXTRA ' measuring function. The TYPE softkey is then pressed after which a selection menu appears, by means of which the Leakage Current Measurement sub-function is selected with the cursor keys and selection is acknowledged by pressing the enter key. The menu screen for leakage current measurement then appears. The measurement can be started or stopped by pressing the START button. Leakage current measurement is a long-term measurement, GMC-I Messtechnik GmbH 5

i.e. it continues until it is stopped by the user. The momentary measured value is display continuously during measurement. In order to perform a measurement, the self-test must be deactivated in the menu (set the TEST ON/OFF function key to OFF ). Always start with the large measuring range (10:1), unless there s no doubt that small measured values can be expected, in which case the small measuring range can be used (1:1). The measuring range must be selected at the measuring adapter, as well as in the menu using the corresponding function key ( RANGE ). It must be assured that the range settings at the adapter and at the test instrument are always identical, in order to prevent any distortion of measurement results. Depending on the magnitude of the measured values, the range setting can, or must (in the case of overranging), be manually corrected at the measuring adapter and the test instrument. Individual limit values can be adjusted after pressing the Limits function key. Exceeded limit values are indicated by the red limit value LED at the test instrument. The menu includes a self-test function for testing the adapter, which can be selected as needed via the parameter settings. For testing as well as for performing measurements, the adapter is plugged into the sensor sockets at the PROFITEST MXTRA or SECULIFE IP test instrument. In addition to this, the two test probes at the 2-pole adapter equipped with test plug must be inserted into the measurement sockets at the leakage current measuring adapter. Polarity is irrelevant. The adapter s status is displayed by means of an appropriate message after the test routine has been completed ( OK, Blown Fuse or Interruption ). A detailed description of the test routing is included in section 6. 6 Self-Test Function for Testing the Adapter A self-test which checks the adapter s essential functions is included in the test instrument, in order to be able to assure that the leakage current measuring adapter is functioning correctly. During this self-test, the integrated fuse (see Characteristic Values on page 8) is tested for continuity and the measurement resistor is checked for a correct value of 1.0 kω within the permissible tolerance range. Two additional tests are then executed which check response characteristics, and in particular test for a correct divider ratio after switchover. The self-test is executed automatically at the test instrument, i.e. the individual tests described above are executed sequentially and the results are displayed after testing has been completed. 6 GMC-I Messtechnik GmbH

6.1 Executing the Test Function at the Instrument The performance test for the leakage current measuring adapter executed at the test instrument is conducted in accordance with the steps described below: 1 Plug the leakage current measuring adapter into the test instrument as shown in figure 1. 2 Use the 2-pole adapter at the test instrument (test plug insert), and plug the PE test probe and the L test probe into the two input jacks at the leakage current measuring adapter (polarity is irrelevant). 3 Within the selected measuring function (EXTRA / IL): Activate the test function by pressing the function key (TEST ON / OFF) a pop-up window appears containing the message Test 10:1, after which the range selector switch at the leakage current measuring adapter must be set to the 10:1 position (if this is not already the case). 4 Start the test sequence by acknowledging with the START key at the test instrument. The test sequence is then executed automatically. 5 After about 1 second, a pop-up window appears containing the message Test 1:1. The range selector switch at the leakage current measuring adapter must then be set to the 1:1 position, and testing is continued automatically after once again acknowledging with the START key (in this case in order to acknowledge that the range has been switched). In the event that an error should occur at the leakage current measuring adapter, this is also indicated by means of an appropriate pop-up window ( Blown Fuse replace the fuse and repeat self-testing / Test Error faulty test execution or defective, unusable leakage current measuring adapter). 6 After the second test phase has been completed (after having switched to the 1:1 range), another pop-up window appears after about 1 second in which the test results are displayed ( Test OK test passed, leakage current measuring adapter can be used / Test Error faulty test execution or defective, unusable leakage current measuring adapter, repair or replacement required). GMC-I Messtechnik GmbH 7

7 Measuring Instrument Requirements At the moment, this leakage current measuring adapter can only be used in combination with the PROFITEST MXTRA and SECULIFE IP from GMC-I Messtechnik GmbH. 8 Remedying Malfunctions In the event that the self-test conducted with the test instrument of the manual inspection of the leakage current measuring adapter reveals a defect, either the internal fuse must be replaced or the adapter must be sent to the service department for repair, depending upon the type of defect. A blown fuse can be replaced by the user. Proceed as follows in the event that this should become necessary: Remove the adapter from the test instrument and disconnect the measurement cables from the adapter. Remove the screw-on lid from fuse holder at the adapter. Carefully remove the fuse and insert a new one (use correct type only).! Attention! Use the specified fuse type only (see Characteristic Values on page 8). After replacing the fuse, the leakage current measuring adapter must be tested for correct functioning as described above. 9 Characteristic Values Input Input current Input impedance Measuring impedance Nominal insulation voltage Fuse 0 to 10 ma 1020 Ω ±3% (±5% according to EN60601-1) 1000 Ω ±3% (±5% according to EN60601-1) 250 V GMC-I type G fuse 5 x 20 mm, T 50 ma / 250 V (article no. 3-578-292-01), DIN 41571-3 Output 10:1 range Output voltage: 0 to 1 V (0.1 V/mA) 1:1 range Output voltage: 0to10V (1V/mA) Output impedance 10 kω 8 GMC-I Messtechnik GmbH

Frequency Response Frequency response per EN 60601-1 Low-pass filter with cut-off frequency of 1 khz Temperature Ambient temperature 0 to +40 C Storage temperature 25 to +70 C Ambient Conditions Operating temperature 0 to + 40 C Storage temperature 25 to + 70 C Relative humidity Max. 75%, no condensation allowed Elevation Max. 2000 m Electromagnetic Compatibility (EMC) Interference emission EN 61326-1:2006, class B Interference immunity EN 61326-1:2006 Mechanical Design Dimensions W x H x D: 80 x 27 x 71 mm Weight Approx. 100 g Protection Housing: IP 40 per EN 60529 / DIN VDE 0470, part 1 10 Return and Environmentally Sound Disposal The instrument is a category 9 product (monitoring and control instrument) in accordance with ElektroG (German electrical and electronic device law). This device is subject to the RoHS directive. Furthermore, we make reference to the fact that the current status in this regard can be accessed on the Internet at www.gossenmetrawatt.com by entering the search term WEEE. In accordance with WEEE 2012/19/EU and ElektroG, we identify our electrical and electronic devices with the symbol in accordance with DIN EN 50419 which is shown at the right. Devices identified with this symbol may not be disposed of with the trash. Please contact our service department regarding the return of old devices (see address in section 11). GMC-I Messtechnik GmbH 9

11 Repair and Replacement Parts Service Calibration Center* and Rental Instrument Service If required please contact: GMC-I Service GmbH Service-Center Thomas-Mann-Str. 20 90471 Nürnberg, Germany Phone: +49 911 817718-0 Fax: +49 911 817718-253 E-mail service@gossenmetrawatt.com www.gmci-service.com This address is only valid in Germany. Please contact our representatives or subsidiaries for service in other countries. * DAkkS Calibration laboratory for measured electrical quantities, D-K-15080-01-01, accredited in accordance with DIN EN ISO/IEC 17025:2005 Accredited quantities: direct voltage, direct current value, direct current resistance, alternating voltage, alternating current value, AC active power, AC apparent power, DC power, capacitance, frequency, temperature 12 Product Support If required please contact: GMC-I Messtechnik GmbH Product Support Hotline Phone: +49 911 8602-0 Fax: +49 911 8602-709 E-mail: support@gossenmetrawatt.com Edited in Germany Subject to change without notice PDF version available on the Internet GMC-I Messtechnik GmbH Südwestpark 15 90449 Nürnberg, Germany Phone +49 911 8602-111 Fax +49 911 8602-777 E-mail info@gossenmetrawatt.com www.gossenmetrawatt.com