Installation Guide MODELS 522 / 526. Pressure Transducers TOLL FREE FAX Web Site.

Similar documents
GEMS SENSORS & CONTROLS SERIES 35XX

Model 206 Operating Instructions. Setra Systems, Inc. 159 Swanson Road, Boxborough, MA

Installation Guide. Setra Systems Model 3100/3200 Pressure Transducer

P027 Series Pressure Transducer Installation & Operation Manual

YT-870 / 875 Series. Rotork YTC Limited VERSION 1.06

PRESSURE TRANSMITTER

January, 2019 Price List

Operating Instructions

Operation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter

Pressure transmitter for highest pressure applications Up to 217,500 psi (15,000 bar) Model HP-2

KROHNE 12/ GR Installation and Operating Instructions BM 500

TECHNICAL MANUAL PRESSURE TRANSDUCER TYPE MLP

MODEL 245 / 345 High Range

LCI Load Cell Junction Box with Fault Monitor

CombiTemp TFRN/TFRH. Operating instructions. Temperature sensor. Operating Instructions: EN.

FA-2448 SIX POSITION Filter Wheel

HI ma Simulator

Installation, Operation & Maintenance Instructions / Technical Document

OPERATION MANUAL MODEL TT-LP

Druck DPI 620 advanced modular calibrator

LCI User Manual mantracourt.com

OPERATING INSTRUCTIONS 7 SERIES STATIC GENERATORS

Series ULT Ultrasonic Level Transmitter

I/O SIGNAL CONDITIONER

HI ma Simulator

Operating Instructions Differential Pressure Transducer PS27

Inductive sensor with analog output BI15-M30-LI-EXI

HIGH ACCURACY PRESSURE TRANSDUCERS

5504 Thermocouple Analog Input Module

Cable entry systems for potentially explosive atmospheres

NT PRESSURE TRANSDUCER MODELS 4100, 4210, 4130, 4230

EX- Documentation VA 450 / VA 452. thermal mass flow sensor

MODEL 8000MP LEVEL SENSOR

SLA Battery Capacity Analyzer

VDI Pro Voltage & Dry Contact Interface Installation and programming Guide MODEL VDI MK2

Magnetic field sensor Magnetic-inductive Proximity Sensor BIM-M12E-Y1X

Installation & Operation Manual

7032 Digital-Analog Multimeter

Pressure sensor For highest pressure applications to 15,000 bar Model HP-2

Level sensor Stainless steel version Model RLT-1000, for industrial applications

MicroLink Installation Operation & Specifications Manual. General Description

Digital Lighting Systems, Inc.

Cable Entry Systems for Potentially Explosive Atmospheres

Digi-Stem Digital Temperature Indicating Devices for Food Processing

esense Duct (-Disp) Conduit

MAXIMA+ Series Rotary Level Indicator

Operating Manual / IEC. Explosion-Proofed Junction Boxes, Polyester. Wallücker Bahndamm 7 DE Kirchlengern

LS 6600 Technical Datasheet

Description, continued

Manual of Electrical Installation for Pressure Probes KS and KS-SIL2 series

English. Operating Manual. Conductive point level switch MLR120. Retain for later reference. Companies / brands of GHM

Table of Contents. Introduction... 2 Description...2 Specifications...3 Installation.. 5 Calibration.. 6 General Maintenance...7

MODEL SST3-MV INSTALLATION INSTRUCTIONS FOR FEDERAL MODEL SST3-MV SELECTABLE SIRENTONE-MULTI VOLT

PRO I Series CAPACITANCE PROBE

BA505C 2-wire 4/20mA manual setpoint station Issue 6

TFR5. Room or outdoor temperature sensor. Operating instructions for temperature sensor. English page 1...5

Operating Instructions. Power Supply & Isolation Relay for Switches Model: RL manual_rl-5900_0215

Electronic pressure switch with display For sanitary applications Model PSA-31

QSB (Q-Star) Electronics for QSE Electromagnetic Meter. Owner s Manual TABLE OF CONTENTS

ProcessCalibrator Model CPH6000

Operation Manual CX10 1/8 DIN Microcomputer Based Conductivity/Resistivity Controller

Operating Instructions for Low Volume Rotating Vane Flow Meter. Model: DTK

SA SERIES Arlingate Lane, Columbus, Ohio (614)

USER MANUAL. WBV412U01 AC voltage transducer

Magnetic Field Sensor Magnetic-inductive Proximity Sensor BIM-EG08-Y1X

Magnetic Field Sensor Magnetic-inductive Proximity Sensor BIM-M12E-Y1X-H1141

ADAMCZEWSKI. Ordering data AD-STVEX 710 GVD Option: Factory-software configuration as per customer specifications.

Operating Instructions for Rotating Vane Flow Meter. Model: DRH-...

KOBOLD PDA Series Digital Pressure Transmitter

Installation instructions Inclination sensor EC2019 EC / / 2013

KOBOLD TDA Series Digital Temperature Transmitter

Inductive sensor NI3-EG08K-Y1-H1341

PRO II X Series CAPACITANCE PROBE

PanelView Plus/VersaView CE Terminals and Display Modules

Model SPDT Switch Module

DR7100 SERIES Supplementary instructions

HART Density Gauge Datasheet

12-Bit 0-5 Volt Input Adapter (Part # S-VIA-CM14)

Continuous level measurement Ultrasonic transducers

MDM 011-Z1 Regen Resistor

Inductive sensor BI5-EM18-Y1X-H1141

USER MANUAL. WBI414F21 AC current transducer

TEMPERIX S Clamp. FAFNIR GmbH Schnackenburgallee 149 c Hamburg Germany

INSTRUCTION MANUAL FOR THE MC-01 CONTROL UNIT

USER MANUAL. WBI412F41 AC current transducer

D Issue A. Instruction Manual. Active Strain Gauge

Electronic pressure switch with display For sanitary applications Model PSA-31

MAXIMA + Series ROTARY LEVEL CONTROL

TETRIS 2500 High Impedance Active Probe. Instruction Manual

RTD SAMPLER OPERATOR S MANUAL

Model 675 DREXELBROOK. Shark Cage Submersible Level Transmitter DATA SHEET. Description. Features. Applications.

PHOENIX CONTACT - 12/2007

MODEL: 6DV-B. [2] TERMINAL BLOCK 0: None (Euro terminal block at the rear of indicator module)

CERTIFICATE of EU TYPE-APPROVAL

MODEL ATV 4/6 BATTERY POWERED DIGITAL VACUUM INSTRUMENT

Limit Switch Box YT-870/875 SERIES

Operating Instructions for Level Switch for Liquids. Model: RFS

Instruction Manual. T3110 T3110L T3110Ex

P30-2. Overview. Pressure Transmitter Datasheet. *High Quality* 3 Years Warranty. At a Glance. Function. Application

Installation and Maintenance Instructions for GC55 Differential Pressure Transducer Version 1.0 5/11

Transcription:

Installation Guide MODELS 522 / 526 Pressure Transducers 1-800-257-3872 TOLL FREE 1-978-264-0292 FAX www.setra.com Web Site 560550-0087 Rev A

Setra offers a complete line of products for these industries: Industrial HVAC Test & Measurement Barometric Ultra High Purity / Sanitary

Please Note: Series 522 / 526 are designed and manufactured in accordance with sound engineering practice as defined by the Pressure Equipment Directive 97/23/EC. This product must not be used as a safety accessory as defined by the Pressure Equipment Directive, Article 1, Paragraph 2.1.3. The CE Mark on the unit does not relate to the Pressure Equipment Directive. INTRODUCTION The Series 522 / 526 pressure transducers use molecularly bonded high output strain gauges to provide 100mV output for full range pressure, when used with a 10V d.c. power supply. Series 522 / 526 high output pressure transducers and transmitters are fitted with an ASIC providing various optional (at time of order) voltage outputs, and a 4-20mA current output capable of being used in control and indicating loops without further amplification. Series 522 / 526 with the CE Mark conform with the essential protection requirements of the EMC Directive 89/33/EEC amended by certified type testing to EN 50082-2 and EN 50081-1. HAZARDOUS PRODUCTS The Consumer Protection Act of 1987, Section 6 of the Heath and Safety at Work Act 1974 and the Control of Substances Hazardous to Health Regulations 1988 require that we advise recipients and users of our products of any potential hazards associated with their storage, handling or use. The products which our Company supplies may be classified as Electrical, Electro-Mechanical and Electronic equipment. These products are tested and supplied in accordance with our published specifications or individual special requirements that are agreed in writing at time of order. They are constructed so as not to affect adversely the safety of persons and property when properly installed, maintained and used by qualified personnel, in the applications for which they were designed and manufactured. Conformity with the requirements of the CE mark only applies when the installation conditions described in these instructions have been met. For units supplied without a cable assembly connection to the transducer must be accomplished using Setra Systems approved cable. See APPROVED CABLE section. GENERAL * Transducer should not be subjected to greater than the maximum allowable pressure listed on label and the following operating temperature specifications: Compensated Temperature Range: Operable Temperature Range: Electrical Connection Code E1 Electrical Connection Code N1 Electrical Connection Code NA -5 F to +180 F (-20 C to +80 C) -40 to +260 F (-40 C to +125 C) -5 F to +180 F (-20 C to +80 C) -5 F to +125 F (-20 C to +50 C) * Transducer should not be subjected to mechanical impact. * In the event of fire the end user must ensure that the system pressure is vented to a safe area.

* The effects of decomposition of unstable fluids should be considered by the user when placing this device in service. * The pressure transducer has no means of draining or venting, this must be performed by another component in the end users system. * Pressure range must be compatible with the maximum pressure being measured. * Pressure media must be compatible with the transducer/transmitter wetted parts listed in these instructions. * Liquid must not be allowed to freeze in the pressure port. * The gasket must be fitted under the electrical connector. MECHANICAL INSTALLATION Mounting: Pressure Transducer is designed to be attached by the coupling thread only. Omnidirectional, self supported directly into the pipework. Use a 19mm AF (3/4") wrench on the hexagon provided to apply a maximum torque of 11.6 lb.ft (15.8Nm). The Customer must ensure that the pressure seal is suitable for the application. If in doubt contact Setra Systems. ELECTRICAL INSTALLATION All types with the CE Mark include suppression devices providing transient protection to EN 61000-4-2 and EN 61000-4-4. Conformity with the requirements of the CE mark only applies when connection is made with Setra Systems approved cable, see APPROVED CABLE section, and is connected as shown below:- Millivolt Output Transducers: Four wire shielded cable should be used with the cable shield connected to the instrumentation ground. The cable shield should not be connected to the transducer body. High Output Transducers: The shield of the cable must be grounded at the instrumentation end. If a ground loop problem is encountered when the body of the unit is grounded by the piping it is permissible to remove the cable shield from the instrumentation ground. APPROVED CABLE Setra Systems uses cables comprising 2, 3 or 4 color-coded wires, enclosed by an aluminised polyester shield where the shield is in contact with a separate drain wire. The outer shield is chrome PVC and overall diameter is approximately 0.20in. OPERATION Having installed the transducer as instructed, it is ready for use. The transducer should not be removed while the system is at pressure. Before applying power, check that the correct polarity and excitation levels are being applied. See Table 1 for electrical connections. CALIBRATION Transducers are calibrated at time of order: Code A Absolute datum Code G Gauge datum vented to atmosphere via the electrical connector or cable Code C Compound datum

LOAD CHARACTERISTICS (4-20mA Current Output) The total resistive load in the loop (to include all the cable resistance) can be from 'zero to 50 x (supply volts -7) ohms' e.g. with a 24V d.c. supply the permissible load is from zero up to 850 ohms. OPERATIONAL LIFE Limited to 100M cycles to maximum allowable pressure. WARRANTY We guarantee this instrument against faulty workmanship and material for a period of two years from date of delivery. The Company undertake to repair, free of charge, ex-works any instrument found to be defective within the specified period providing the instrument has been used within the specification in accordance with these instructions and has not been misused in any way. Detailed notice of such defects and satisfactory proof thereof must be given to the Company immediately after the discovery and the goods are to be returned free of charge to the Company, carefully packed and accompanied by a detailed failure report. See "RETURN TO FACTORY". WETTED PARTS 316 and 17-4 PH stainless steel plus Nickel Braze to BS1845 : NK3/HTN2. SERVICING The transducer cannot be repaired locally and if damaged should be returned to Setra Systems with the following:- 1. Name and phone number of person to contact 2. Shipping and billing information 3. Full description of malfunction 4. Remove any pressure fittings and plumbing that you may have installed and enclose any required mating electrical connectors and wiring diagrams. Please ship prepaid to Setra Systems at the address shown below to to your local Setra Distributor when a replacement/repair is required: United States Setra Systems Inc 159 Swanson Road Boxborough, MA 01719 Attn: Repair Department Outside The United States contact your local Setra Distributor RETURN TO FACTORY PLEASE NOTE: To comply with Health and Safety requirements, the instrument must be clean and safe to handle and accompanied by a formal statement to that effect duly signed by an authorised officer of the Company.

Any instrument returned without certification will be quarantined and no action will occur until cleared. It may ultimately be returned to you and subject to a transportation charge. MAINTENANCE Routine Inspection: Not required except for periodic inspection of the cable and connector to ensure that these are neither damaged nor softened by incompatible liquid.

TABLE 1 CONNECTIONS 522 / 526 SERIES TYPE NUMBER CE CONNECTOR mv 4-20mA VOLTAGE IN+ OUT+ OUT- IN- EARTH IN+ IN- EARTH IN+ COM OUT+ EARTH 522BP-E1 YES MINI DIN STYLE 1 2 3 E 522BP-NA YES CABLE WEATHER PROOF R Y BL G S22BP-N1 YES CABLE USA R W G BK 522BP-E1 YES MINI DIN STYLE 1 2 E 1 2 3 E 522XX-NA YES CABLE WEATHER PROOF R BK DRAIN R BK W DRAIN 522XX-N1 YES CABLE USA R BK DRAIN R BK W DRAIN 526BP-B1 YES 8-4P A B C D 526BP-C2 YES 10-6P A B C D 526BP-E2 YES LARGE DIN 1 2 3 E 526BP-UA YES MOLDED IMMERSIBLE R Y B W DRAIN 526BP-UA YES MOLDED IMMERSIBLE R Y B W DRAIN 526BP-A2 YES CONDUIT LEADS R W G BK 526XX-B1 YES 8-4P A B D A C B D 526XX-B3 YES 10-6P A B E A C B E 526XX-C2 YES LARGE DIN 1 2 E 1 2 3 E 526XX-UA YES MOLDED IMMERSIBLE R BL DRAIN R W Y DRAIN 526XX-UA YES MOLDED IMMERSIBLE R BL DRAIN R W Y DRAIN 526XX-A1 YES CONDUIT LEADS R BK G R BK W G OUTPUT SUPPLY VOLTAGE OUTPUT SUPPLY VOLTAGE BP 0 to 100mV 10V d.c. 27 1 to 5V 6.5 to 35V d.c. 11 4 to 20mA 7 to 35V d.c 24 0.5 to 5.5V 7 to 35V d.c. 28 1 to 6V 7.5 to 35V d.c 2B 0 to 5V 6.5 to 35V d.c. 2R 1 to 11V 12.5 to 35V d.c 2C 0 to 10V 11.5 to 35V d.c. 159 Swanson Road, Boxborough, MA 01719-1304 Tel: 800-257-3872 / 978-263-1400 OP522/526 Rev A 11/03