Transmitter / Controller Series J dtrans ph 01, Rd 01, Lf 01, Rw 01, Az 01 with C-UL-US approval PR 20004 GB ph mv ms/cm µs/cm M xcm
Transmitter / Controller Series J dtrans ph 01, Rd 01, Lf 01, Rw 01, Az 01 completely in microprocessor technology simple operation from membrane keys simultaneous indication of process value and temperature on easily read LED display up to two isolated analog outputs, for use as process value output or proportional controller (freely configurable) temperature monitoring is available sensor monitoring is available as standard two freely programmable relays for limit monitoring or control, also one alarm output up to four relay outputs can be implemented pulse duration or pulse frequency controller with P, PI, PD and PID action programmable modulating controller function compact size 96 x 48 x 110mm (W x H x D) protection IP65 front only two power supply variants from 24V DC to 240V AC interface RS422 / 485 (option) PROFIBUS-DP surface-mounting or outdoor case IP65 available manual function (pulsing and switching) simulated process output available, e.g. for dry start-up Az 01 version standard signal input 0 / 4-20mA universal calibration adaptation supply voltage for 2-wire transmitters Operation Operating level Parameter level Configuration level The controller parameters and configuration data are arranged on several levels for easy programming and operation. The levels are protected against unauthorized access by codewords. Membrane keys provide simple and user-friendly operation. The two LED displays indicate parameter symbols and corresponding values. J SVS 2000N Process visualization software with batch-related data reporting and evaluation Simple and rapid linking of Series J dtrans ph 01, Rd 01, Lf 01, Rw 01, Az 01 microprocessor transmitter/controller to the SVS-2000N visualization software. (You can request further information on the SVS-2000N using the fax reply on the back page.) *relays are also programmable as limit comparators for temperature
J dtrans ph 01 for ph Complete measurement chain from one supplier J dtrans Rd 01 for redox potential ph and metal electrodes, conductivity cells, fittings and accessories are also available J dtrans Lf 01 for electrolytic conductivity J dtrans Rw 01 for high-purity water wall-mounting housing (option)
Simple operation... e.g. calibration of ph electrode 1. Enabling the operating level Instrument in measurement mode (Fig. 1) protected against unauthorized operation by codeword Abb. 1 Press key PGM,hold down until Cod.E appears Using the arrow keys set codeword for operating level and enter with PGM (Fig. 2) Fig. 2 Press EXIT key instrument jumps back to measurement mode 2. Calibration with 2 buffer solutions Start calibration by pressing and PGM simultaneously Fig. 3 Set temperature of buffer solution with arrow keys (manual temperature compensation), or indication of measured buffer temperature (automatic temperature compensation) (Fig. 3) Enter with PGM Fig. 4 CAL 1: immerse electrode in first buffer solution. As soon as reading is steady, adjust indication with arrow keys to buffer ph, enter with PGM (Fig. 4) CAL 2: use buffer 2, repeat as for CAL.1 (Fig. 5) Fig. 5 Instrument jumps back to measurement mode (Fig. 1)
measurement and control range ph 1.00 to +14.00 can be switched from ph to redox (ORP) with redox: display in %, without dimensional unit (can be freely assigned) electrode connection through BNC connector suitable for all conventional ph and redox electrodes (combination electrodes and separate measurement / reference electrodes) input for antimony electrodes provided version for JUMO 2-wire ph transmitter available measurement and control range 1999 to +1999 mv, freely adjustable display in %, without dimensional unit (can be freely assigned) suitable for all conventional redox electrodes (metal combination electrodes or separate metal and reference electrodes) temperature indication can be switched on version for JUMO 2-wire redox transmitter available freely adjustable ranges from 0-0.5µS / cm to 0-200mS / cm in one instrument selection without jumpers or solder links can be connected to electrolytic 2-electrode cells with cell constants 0.01 / 0.1 / 1.0 / 3.0 / 10.0 (also intermediate values) integrated calibration for temperature coefficient and cell constant cable break monitor freely programmable dual-slope calibration of process value output for electrolytic 2-electrode cells with cell constant 0.01 and 0.1 indication can be switched from MOhm cm to µs / cm temperature compensation to ASTM D-1125-95 meets the requirements of USP several compensation curves are stored: acidic (HCl), alkaline (NaOH), high-purity water (NaCl); temperature compensation linear or switched off cell constant programmed manually
Fax reply (for fax number see address below) PR 20004 GB Transmitter / Controller Series J dtrans ph 01, Rd 01, Lf 01, Rw 01, Az 01 Please send the following documentation: Data Sheet 20.2530 (ph 01) Data Sheet 20.2535 (Rd 01) Data Sheet 20.2540 (Lf 01) Data Sheet 20.2545 (Rw 01) Data Sheet 20.2550 (Az 01) Sectional Catalog Analytical Measurement Broschure 70052 GB SVS-2000N Send me your Product Guide (PR 00046 GB) Sender: Name Company Industry Address Send me a quotation the latest version of your free CD Products Please arrange a personal discussion with... Phone Fax phone... e-mail time... Date Stamp and signature JUMO GmbH & Co. KG JUMO Instrument Co. Ltd. JUMO PROCESS CONTROL INC. Street address: JUMO House 885 Fox Chase, Suite 103 Moltkestraße 13-31 Temple Bank, Riverway Coatesville PA 19320, USA 36039 Fulda, Germany Harlow, Essex CM 20 2TT, UK Phone: 610-380-8002 Postal address: Phone: +44 12 79 63 55 33 1-800-554-JUMO 36035 Fulda, Germany Fax: +44 12 79 63 52 62 Fax: 610-380-8009 Phone: +49 661 6003-0 e-mail: sales@jumo.co.uk e-mail: info@jumousa.com Fax: +49 661 6003-607 Internet: www.jumo.co.uk Internet: www.jumousa.com e-mail: mail@jumo.net Internet: www.jumo.net 6.04 / 00366374