Retrospective Spiral Respiratory Correlated Imaging with Varian RPM

Similar documents
Brilliance CT Big Bore.

Helical 4D CT pitch management for the Brilliance CT Big Bore in clinical practice

Laser Systems for Patient Alignment

Artefakt-resistente Bewegungsschätzung für die bewegungskompensierte CT

Virtual Phantoms for IGRT QA

How To Title: Scan and File into ICS Purpose: To scan and file information into the patient s medical record Audience: All Staff Date: June 21, 2013

Image Guidance and Beam Level Imaging in Digital Linacs

AAPM Standard of Practice: CT Protocol Review Physicist

An investigation of temporal resolution parameters in cine-mode four-dimensional computed tomography acquisition

41126 Cognento (MODENA) Italy Via Bottego 33/A Tel: +39-(0) Internet: Fax: +39-(0)

icatvision Quick Reference

Optical Guidance. Sanford L. Meeks. July 22, 2010

Motion artifact detection in four-dimensional computed tomography images

3T/7T MRI FACILITY. 3T MRI System Start-Up and Shutdown

XRADIA microxct Manual

File Management Utility User Guide

Deviceless respiratory motion correction in PET imaging exploring the potential of novel data driven strategies

DDX-R Quick Start Guide

7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT

SecureGuard VMS. SecureGuard All-In-One Quick Install Guide V. 2.5

Moving Metal Artifact Reduction for Cone-Beam CT (CBCT) Scans of the Thorax Region

CT Protocol Review: Practical Tips for the Imaging Physicist Physicist

CT Basics Principles of Spiral CT Dose. Always Thinking Ahead.

JAZZ HARMONY User Manual

Sabre for Windows v4.5g INSTALLATION GUIDE

Philips SPECT/CT Systems

Thoracic target volume delineation using various maximum-intensity projection computed tomography image sets for stereotactic body radiation therapy

VITALSTIM PLUS Software Tutorial

Image Quality Assessment and Quality Assurance of Advanced Imaging Systems for IGRT. AAPM Penn-Ohio Chapter Sep 25, 2015 Soyoung Lee, PhD

Guidelines of Ethernet for Using a Star Dvice. - Windows - Rev. 2.0

PURE. ViSION Edition PET/CT. Patient Comfort Put First.

Initial Clinical Experience with 3D Surface Image Guidance

SICAT SUITE VERSION 1.0. Instructions for use SIDEXIS XG English

Soarian Quick Start Guide for Office Staff

URGENT IMPORTANT FIELD SAFETY NOTIFICATION

Annexure XII SPECIFICATIONS FOR A NEW STATE OF ART 16 SLICE ALL PURPOSE C. T. SCANNER

Analysis of Acquisition Parameters That Caused Artifacts in Four-dimensional (4D) CT Images of Targets Undergoing Regular Motion

epact2 User Guide Accessing and Using Dashboards, Pre-built Reports & Quick Search Accessing the Dashboards... 1 Navigating the Dashboards...

Chapter 1 General Information

Using New Features in CDR DICOM 3.5 Software

Brown County Avatar User Guide V1.0 Chapter 10 Order Entry

USING SOUNDMINER V4PRO - V4.5

1. Login as an administrator

Application Notes for Avaya IP Office Applications and CA Business Protection Suite Issue 1.0

SICAT SUITE VERSION 1.0. Instructions for use English

SICAT SUITE VERSION 1.1. Instructions for use English

RaySafe Xi View. User Manual

Carestream Vita user quick guide. Software version 3.2 From April 2012

Patterson Imaging Multi-User Installation Instructions

CTA HEAD Perfusion AqONE without and with IV Contrast

Mobile Workforce. Quick Start Guide

Image-In Direct Capture user guide

Rev: August 29, CIRS Motion Control User Manual

Development and Optimization of Fourdimensional Magnetic Resonance Imaging (4D-MRI) for Radiation Therapy

UvA-DARE (Digital Academic Repository) Motion compensation for 4D PET/CT Kruis, M.F. Link to publication

Installing Cash Register Express

Copyright 2012 Pulse Systems, Inc. Page 1 of 29

Manual. User Reference Guide. Analysis Application (EMG) Electromyography Analysis

IntelliVue Information Center

AN EVALUATION OF IMSIMQA AS A TOOL FOR COMMISSIONING 4DCT

DRCR.NET Image Acquisition Protocol. Optical Coherence Tomography Using: Zeiss Cirrus OCT

PACS ADMIN. Quick Reference Guide

OUTPUTS AND INTERFACES CARDIOSCAN HOLTER ECG. VERSION 76a

Automatic Intrinsic Cardiac and Respiratory Gating from Cone-Beam CT Scans of the Thorax Region

2015 Optima Healthcare Solutions Page 2

IntelliVue Information Center

Real World Experience: Developing Dose and Protocol Monitoring from Scratch

Secure Access Authentication / Restriction of workflows based on User

Lower Dose with Confidence

Routine excellence. Philips Ingenuity Core specifications

Cardiac Dual Energy CT: Technique

Hitachi Medical Systems America, Inc. Hitachi Medical Corporation

Mobile POS Tablet. Barcode Scanner DATA CAPTURE MODULE OPERATION GUIDE. Peripheral Scanning Tool. To Connect Barcode Scanner Module

VivoSense. User Manual - LifeShirt Import Module. Vivonoetics, Inc. San Diego, CA, USA Tel. (858) , Fax. (248)

Mobile POS Tablet Barcode Scanner

EagleSoft Set-up Instructions for Denoptix Version: Date Created: 6/10/05 Last Modified: 10/24/05 Filename: Denoptix ES Setup Instructions

Tutorial: set up of measurement recipe with database

The Near Future in Cardiac CT Image Reconstruction

HESKA DCU (DATA CAPTURE UTILITY) VERSION 2010 AND HIGHER

Getting Started with Crazy Talk 6

Abstract. Apteryx Imaging (XrayVision, XVLite ) Installation Guide. Abstract Abstract

FOREWORD TO THE SPECIAL ISSUE ON MOTION DETECTION AND COMPENSATION

Abstract. Apteryx Imaging (XrayVision, XVLite ) Installation Guide. Abstract Abstract

D1.5a Create Access Master Files

NVMS User Manual. Version 2.1.0

Under SOAP Notes icon/ Progress Notes icon

Four-Dimensional MVCT Imaging and Spatial Translation of Helical Tomotherapy Dose Distributions Using Sinogram Modification

State-of-the-Art IGRT

Zmachine Data Viewer Software Quick Start

An automated method for comparing motion artifacts in cine four-dimensional computed tomography images

8/7/2017. Disclosures. MECT Systems Overview and Quantitative Opportunities. Overview. Computed Tomography (CT) CT Numbers. Polyenergetic Acquisition

Hardware: Acquiring Data and Communicating with Instruments

Using Scantron ParLAN 6.5 for the First Time:

Questionmark Assessments Taking an Assessment

Tumor motion during liver SBRT

ATTACHING MULTIPLE IMAGES TO A DOCUMENT POWERING SUCCESSFUL PRACTICES TM VETERINARY SOLUTIONS

7/31/ D Cone-Beam CT: Developments and Applications. Disclosure. Outline. I have received research funding from NIH and Varian Medical System.

ABAXIS VETSCAN CHEMISTRY VERSION AND HIGHER

PVDOMICS. Chest CT Scanning and Data Transmission. Jason Lempel, MD Lung Imaging Core Data Coordinating Center Cleveland Clinic Nov.

Web Forms: Erasmus+ & ESC How to complete the form

Transcription:

Retrospective Spiral Respiratory Correlated Imaging with Varian RPM This is a Quick Step Guide for Retrospective Spiral Respiratory Correlated Imaging (4D CT) using the Varian RPM device v1.7 with the Brilliance CT Big Bore v2.2.2 system and the Brilliance 16-64 v2.2.5. For more detailed information, please refer to the Varian RPM operator manual and reference guide. Before starting, verify the following items are set correctly on the CT scanner and the RPM workstation. 1. Connect the Varian RPM Sensor Device to the gantry. 2. Ensure that the Varian RPM computer is turned on, is connected to the hospital network, and is fully booted up before logging in to the CT Host. 3. Wait at least 5 seconds after the login prompt appears before logging in to the CT Host. Note Restart the CT Host daily. 4. On the CT scanner, Select Preferences in the Directory tool panel. 5. Select Scanner from menu. 6. In the Pulmo Sensor Device option, select Varian. 7. Click OK to accept. 8. Place the marker block on the patient s chest or abdomen. 9. Review the procedure with the patient emphasizing the importance of maintaining a consistent breathing pattern. Refer to the Varian RPM reference guide for information about usage of Audio or Visual coaching. 1

Start Study on the CT scanner 1. Click Start Study to begin a new study. 2. Complete the patient data form. 3. Click Exam Protocol Groups. 4. Select the Thorax Protocol group. 5. Select Pulmo Retrospective Spiral Protocol. Launch the Varian RPM workstation 1. Login to the Varian RPM Workstation. 2. Double click the RPM Respiratory Gating Icon. 3. Close the Patient List. 4. Proceed with System Configuration. 5. Click on the View Menu. 6. Select System Configuration. 2

7. Click on the CT Interface Tab. CT Interface Tab 8. Select Trigger based on phase value. 9. Set Trigger At: to 0 degrees. 10. Set CT Trigger Length: to 0.50 sec. 11. Select Sensed via Digital I/O. 12. Set Binary Value indicating X-ray on: to [1] 13. Select No new pulse after and set to 180 sec. 14. Click File Export Tab 3

File Export Tab 1. Select Prompt to export a respiration data file after recording is finished. 2. Select Recalculate phase before respiration data export. Note Select Recalculate phase before respiration data export, only if you choose to recalculate phase before exporting. 3. Export filename format must be set to Patient ID_Exam Number_Series Number.vxp. 4. Export file location must be set to C:\Documents and Settings\VARIAN\Desktop\Gating\Breath (unless instructed otherwise by service or hospital personnel). 5. Click Chart Tab. Chart Tab 1. Set Recording Length to no higher than 9 minutes. Note Values greater than 9 minutes not supported. 2. Select Auto scale. 3. Click Tracking Options Tab. 4

Tracking Options Tab Note Options to be determined at installation by the field engineer (refer to the Varian manual for more information). 1. Once the System Configuration settings are confirmed click OK. Patient List 1. On the RPM System Click File. 2. Select Patient, patient list will open. 3. Click New Patient. 5

4. Complete the Patient Data Form, to match the CT scanner patient data form. 5. Click OK when complete. 6. Highlight the Patient Name. 7. Click Select. 8. In the New Treatment Field window, fill in the Treatment Field Label. 9. Select Appropriate Treatment Field. 10. Click OK. 6

Note A message will display: Do you want to create a reference session for selected option? 11. Click Yes. 12. RPM screen will display with Track Enabled. Learning Phase on the RPM System During the Learning Phase the RPM system will track the patient s respiratory motion and attempt to learn the patient s respiratory motion. Once the system has a good signal and learned the respiratory motion, the blue indicator bar at top left of the screen should display one or two bars. Acquiring Surview on the CT scanner On the CT scanner, acquire the Surview and any other preliminary scans desired. Note Some customers prefer a routine chest scan before the retrospective pulmonary acquisition. These must be completed before pressing the record button on the Varian RPM. 7

1. Confirm scan parameters. 2. Do not use Auto Voice. 3. Click Go to start the Surview. 4. On the CT set Scan Parameters. 5. On the Varian RPM computer Press Record. 6. On the RPM computer Click Enable Gating Button. Note Pulses will start to display on the RPM workstation after the enable gating is activated. 7. Use the table below to determine the pitch and rotation speed. Breath Rate (in breaths per minute) For 0.5 sec Rotation Time, use a pitch no higher than: For 0.44 sec Rotation Time, use a pitch no higher than: 20 0.15 0.12 15 0.11 0.1 14 0.105 0.09 13 0.09 0.085 12 0.09 0.08 11 0.08 0.07 10 0.075 0.065 8. Once the parameters are set on the CT scanner click Go to start the scan. 8

9. Press Enable. 10. When patient s breathing is steady, press Manual to start the scan. Note The waveform displayed on the Varian system will turn green when x-rays are on and the Radiation On symbol will appear. Note Monitor the patient s breathing during the scanning process. You may choose to stop the scan if breathing is irregular through the area of interest. Note On the CT scanner you will be prompted to import the Varian File when the scan is completed. 11. Do Not click OK keep window open and proceed to the next step. Note The file has not been saved for export, it is still recording. 12. On the RPM Computer click the Stop button to save the file. 9

13. On the RPM Computer Type in 1 in the Exam Number field. 14. On the RPM Computer Type in 1 in the Series Number field. 15. On the RPM Computer click OK when complete. 16. On the RPM Computer click OK to New Export Step Verification. Note On the RPM Computer Respiration data file was exported successfully will display on the screen. 17. On the RPM Computer click OK. [again?] 18. On the CT scanner click OK on the Import Varian File window to complete the file import. 10

19. On the CT scanner review the wave form that was imported, and edit tags if desired. Note Tags represent where the reconstruction will define the beginning of the breath cycle (0%). Please be aware that there is no opportunity to edit tags sent from RPM TM during gated treatment. 20. On the CT scanner click OK to start the image reconstruction. Note May take up to 90 seconds for the first image 21. On the Varian RPM workstation click Close Patient. Philips Medical Systems is part of Royal Philips Electronics Philips Medical Systems Cleveland, Ohio 440 483 3000 Philips Medical Systems All rights reserved. (2007) Reproduction in whole or in part Is prohibited without prior consent of the copyright holder. Philips Medical Systems Printed in U.S.A. 11