SCXI Introduction CALIBRATION PROCEDURE. What Is Calibration? Why Should You Calibrate? How Often Should You Calibrate?

Similar documents
SCXI Contents CALIBRATION PROCEDURE

SCXI Equipment and Other Test Requirements CALIBRATION PROCEDURE. Test Equipment. Software and Documentation

NI 6703/6704 Calibration Procedure for Traditional NI-DAQ

CALIBRATION PROCEDURE SCXI Contents. Software Requirements. ni.com/manuals

The following conventions appear in this document:

SCXI -1104/C. Contents CALIBRATION PROCEDURE. For NI-DAQmx

NI PCIe-784xR, NI PXI-784xR, NI PXIe-784xR, NI USB-784xR, NI PCIe-785xR, NI PXI-785xR, NI PXIe-785xR, NI USB-785xR, and NI PXIe-786xR

SCXI -1102/B/C. Contents CALIBRATION PROCEDURE. For NI-DAQmx

NI R Series. Contents. Software CALIBRATION PROCEDURE

NI 6703/6704 Calibration Procedure for NI-DAQ mx

FP-RTD-122. Introduction CALIBRATION PROCEDURE. Version 2.0

Contents. Software Requirements. CALIBRATION PROCEDURE NI PXIe-4357

CALIBRATION PROCEDURE NI PXIe Channel, Isolated Voltage/Current Analog Output Module

CALIBRATION PROCEDURE CAL Contents. Software

Contents. Software. CALIBRATION PROCEDURE NI Channel, ±5 V, 24-Bit Software-Selectable IEPE and AC/DC Analog Input Module

Contents. Software. CALIBRATION PROCEDURE NI 9216/ Channel, 24-Bit, 100 Ω/1000 Ω RTD Analog Input Module

CALIBRATION PROCEDURE NI channel, ±2- ma/±10 V, 24-bit Analog Input Module. ni.com/manuals

Contents. Software. CALIBRATION PROCEDURE NI Channel, 300 Vrms, 24-Bit, Simultaneous, Channel-to-Channel Isolated Analog Input Module

CALIBRATION PROCEDURE PXIe channel, 16-bit, 400 ks/s/ch, Ch-Ch Isolated Analog Input Module. ni.com/manuals

NI 9213 Calibration Procedure

E/S/M/B Series Calibration Procedure for NI-DAQ mx

±10 V, 20 MS/s/ch, 14-Bit, 4-Channel C Series Digitizer Module

Calibrating FieldPoint Analog Input Modules with Calibration Executive

NI PXI Contents CALIBRATION PROCEDURE

Contents. Software Requirements CALIBRATION PROCEDURE NI TB-4353

Contents. Software CALIBRATION PROCEDURE NI TC-4353

NI Introduction CALIBRATION PROCEDURE. What Is Calibration? External Calibration

SCXI SHIELDED MULTI-CHASSIS CABLE ADAPTER

ESERIES CALIBRATION FIXTURE INSTALLATION GUIDE

To calibrate the NI 5105, install the following versions of NI-SCOPE. NI-SCOPE Version

SCXI OFFSET-NULL AND SHUNT-CALIBRATION HIGH-VOLTAGE TERMINAL BLOCK INSTALLATION GUIDE

E Series Multifunction I/O 1.25 MS/s, 12-Bit, 16 or 64 Analog Inputs

SCXI -1313A Terminal Block

SCC-AO10 Isolated Analog Output Module

Related Information Refer to the NI Switches Help for detailed information about features and programming options for your NI switch product.

Chapter 1 Introducing the OM-USB Functional block diagram... 5

SCXI-1345 Shielded Cable

Low-Cost Multifunction DAQ for USB

Install the Software and Hardware

Contents. CALIBRATION PROCEDURE NI PXIe-5624R 2 GS/s, 12-Bit IF Digitizer

TBX-1316 High-Voltage Attenuator Terminal Block

NI PXI Contents CALIBRATION PROCEDURE

OPERATING INSTRUCTIONS AND SPECIFICATIONS. NI 9208E 16-Channel, ±20 ma, 24-Bit Analog Input Module

24-Channel, ±24 V Precision PXI Source Measure Unit

High-Speed Voltage Output up to 1 MS/s/Channel, up to 16 Bits, up to 32 Channels

12-Channel, ±24 V Precision PXI Source Measure Unit

Ethernet/Serial Interfaces for Compact FieldPoint

Strain gauge Measuring Amplifier GSV-1A8. Instruction manual GSV-1A8, GSV-1A8USB, GSV-1A16USB

High-Speed Voltage Output up to 1 MS/s/Channel, up to 16-Bit, up to 32 Channels

Multifunction DAQ Accessory and Cable Selection Guides

Where to Start with NI ELVIS

NI 446X Calibration Procedure

High-Speed M Series Multifunction DAQ 16-Bit, up to 1.25 MS/s, up to 80 Analog Inputs

SCB-68 User Guide. 68-Pin Shielded Desktop Connector Block

GE Fanuc IC694ALG221. Rx3i PacSystem

Data Acquisition Card or DMM: Which is Right for Your Application? By John Tucker and Joel Roop Keithley Instruments, Inc.

SCXI User Manual. 32-Channel Isolated Digital Output Module. September 1994 Edition Part Number

DAQ Accessories Selection Guide

First Automated and Self Calibrating Voltage Divider

High-Speed M Series Multifunction DAQ 16-Bit, up to 1.25 MS/s, up to 80 Analog Inputs

Features. High-precision Analog input board (Low Profile size) for PCI Express AI-1616L-LPE AI-1616L-LPE 1. Ver.1.01

High-Accuracy M Series Multifunction DAQ 18-Bit, up to 625 ks/s, up to 32 Analog Inputs

A variety of ECONseries modules provide economical yet flexible solutions. Waveform Generation

Developing Networked Data Acquisition Systems with NI-DAQ

SCXI User Manual

Electrical Metrology Applications of LabVIEW Software

Portable E Series Multifunction DAQ 12 or 16-Bit, up to 1.25 MS/s, 16 Analog Inputs

VXI-8340 SERIES INSTALLATION

A 40-pin connector provides access to 24 DIO connections.

Model 2700 Multimeter/Data Acquisition System Service Manual

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

BC-9000 Field Calibration Procedure CF1_FIELDCAL_BC9000 Revised 03/16/2013

VXI-SC-1000 CARRIER MODULE

ACTIONI/Q Q498. DC Powered DC Input Field Configurable Isolator with Math Functions

Temperature measurement board, optically isolated, 16/8/4 channels for thermocouples, Pt100, RTD, 18-bit

NI CB-37F-LP. Introduction INSTALLATION GUIDE. Connector Block for 37-Pin D-Sub Devices

Strain gauge Measuring Amplifier GSV-1A8. Instruction manual GSV-1A8, GSV-1A8USB, GSV-1A16USB

High-Density Industrial Digital I/O for USB 96-Channel, 5 V TTL/CMOS

High-Precision Data Acquisition

BC-7000 Field Calibration Procedure 55 AMP

Low-Cost E Series Multifunction DAQ. 12 or 16-Bit, 200 ks/s, 16 Analog Inputs

OP5332 USER GUIDE Isolated Digital to Analog Converter Module

LGR-5325 Specifications

SCXI User Manual. 32-Channel Isolated Digital Input Module. March 1994 Edition Part Number

USB-201-OEM. Analog and Digital I/O. User's Guide

Register Map and Descriptions

SCXITM. SCXI-1125 User Manual. SCXI-1125 User Manual. October C-01

LabVIEW Real-Time Module for Mac OS X User Manual Addendum

Getting Started with the NI 783x R

The NI TB-2648/2649/2650/2651 terminal blocks configure your NI PXI/PXIe-2531 (NI 2531) in one of the following topologies. 1-wire 8 64 matrix

NI PXI Contents GETTING STARTED GUIDE. 8 GHz RF Amplifier

UNIVERSAL MOTION INTERFACE (UMI) ACCESSORY

BC-5000 Field Calibration Procedure 55 AMP

Isolated, Process Current Output 7B39 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

PCI-DAS6402/16 Specifications

8-Slot PXI Express Chassis for PXI and PXI Express Modules

Isolated Process Current Input 7B32 FEATURES APPLICATIONS PRODUCT OVERVIEW FUNCTIONAL BLOCK DIAGRAM

Isolated Process Current Input with Loop Power 7B35

INSTALLATION INSTRUCTIONS NI TB-2630B Terminal Block for the NI PXI-2530B. Contents. Introduction

NI USB-9201/9221. Introduction USER GUIDE AND SPECIFICATIONS. 8-Channel, 12-Bit Analog Input Devices

Transcription:

CALIBRATION PROCEDURE SCXI -1125 Introduction What Is Calibration? Why Should You Calibrate? How Often Should You Calibrate? This document contains information and step-by-step instructions for verifying and calibrating the National Instruments SCXI-1125 signal conditioning module. Calibration is a procedure of reading offset and gain errors from a device and updating special analog calibration circuitry that corrects these errors. National Instruments calibrates every SCXI-1125 module at the factory. During the factory calibration process, the calibration constants are stored in the non-volatile memory of the device. These values are loaded from memory and used as needed by the device. Offset and gain errors drift with time and temperature changes, which could invalidate the factory-set calibration of a device and increase measurement uncertainty. Calibration restores the device to its specified accuracy. The measurement accuracy requirements of your application determine how often you should calibrate your SCXI-1125 module. National Instruments recommends you perform a complete calibration at least once every year. You can shorten this interval to six months or 90 days, based on the demands of your application. CVI, LabVIEW, National Instruments,ni.com,NI-DAQ, and SCXI are trademarks of National Instruments Corporation. Product and company names mentioned herein are trademarks or trade names of their respective companies. 370271A-01 Copyright 2000 National Instruments Corp. All rights reserved. December 2000

Equipment and Other Test Requirements This section describes the equipment, software, documentation, and test conditions required for calibrating the SCXI-1125. Test Equipment Calibration requires a high-precision voltage source with at least 50 ppm accuracy, a multiranging 5 1/2 digit digital multimeter (DMM) with 15 ppm accuracy, and a National Instruments E Series DAQ device. National Instruments recommends the following instruments for calibrating the SCXI-1125: Calibrator Fluke 5700A DMM NI 4060 or HP 34401A National Instruments E Series DAQ device If these instruments are not available, use the accuracy requirements listed above to select a substitute calibration standard. Note If you do not have custom connection hardware, you will need a connector block such as the National Instruments SCXI-1320, a shielded 68-pin connector cable, a 50-pin ribbon cable, a 50-pin breakout box, and an SCXI adapter. These components give easy access to the individual pins on the SCXI-1125 front and rear connectors. Software and Documentation You will need the following software and documentation to calibrate the SCXI-1125 signal conditioning modules. You can find these on the National Instruments Web site at ni.com/calibration: The latest version of the National Instruments NI-DAQ driver NI-DAQ Function Reference Online Help file NI-DAQ User Manual for PC Compatibles Software The SCXI-1125 calibration procedure requires the latest version of the National Instruments NI-DAQ driver on the calibration system. The driver supports a number of programming languages, including LabVIEW, LabWindows/CVI, Microsoft Visual C++, Microsoft Visual Basic, and Borland C++. When you install the driver, you only need to install support for the programming language you will use. SCXI-1125 Calibration Procedure 2 ni.com

Documentation The NI-DAQ Function Reference Online Help file and NI-DAQ User Manual for PC Compatibles containdetailedinformationonusingthe NI-DAQ driver. The function reference online help file includes detailed information on the driver functions. You can access the online help by clicking Start»Programs»National Instruments DAQ»NI-DAQ Help. The user manual provides instructions on installing and configuring National Instruments DAQ devices. This manual also includes detailed information on creating applications that use the NI-DAQ driver. These are your primary references for writing your calibration utility. For further information on the products you are calibrating, refer to your device user manuals. Test Conditions Follow these guidelines to optimize the connections and the environment during calibration: Keep connections to the SCXI module short. Long cables and wires act as antennae, picking up extra noise and thermal offsets that can affect measurements. Use shielded copper wire for all cable connections to the device. Use twisted-pair wire to eliminate noise and thermal offsets. Maintain the temperature between 18 and 28 C. Keep relative humidity below 80%. Allow a warm-up time of at least 15 minutes for the SCXI module and 30 minutes for the E Series device to ensure the measurement circuitry is at a stable operating temperature. Calibration The calibration process consists of the following steps: 1. Setting up the device for testing. 2. Verifying the existing operation of the device to determine whether it is operating within its specifications. 3. Adjusting the device with respect to a known voltage source. 4. Verifying that the device is operating within its specifications after adjustments. National Instruments Corporation 3 SCXI-1125 Calibration Procedure

Setting Up Your Module Verifying Gain and Offset Errors Complete the following steps to set up your module for testing: 1. Install the SCXI-1125 in slot 1 of the SCXI chassis. 2. Connect a 68-to-68-pin cable between the SCXI module and the E Series DAQ device installed in the host computer. 3. Configure the hardware with Measurement & Automation Explorer (MAX). For more information on configuring the system, refer to the E Series device and SCXI module documentation. The verification procedure determines how well the SCXI-1125 module is meeting its specifications. You can use this information to select the appropriate calibration interval for your application. Complete the following steps to verify gain and offset errors in the SCXI-1125: 1. Read the Test Conditions section earlier in this document. 2. Refer to the specifications in Table 1 at the end of this document. 3. Connect the E Series device to the SCXI-1125. 4. Call Calibrate_E_Series to minimize the uncertainty associated with the E Series device measurements. Set the following parameters: devicenumber The value assigned by MAX calop ND_SELF_CALIBRATE setofcalconstants ND_USER_EEPROM_AREA calrefvolts 0.0 5. Call AI_Configure to put the E Series device in bipolar mode. Set the following parameters: devicenumber The value assigned by MAX chan 0 inputmode 0 inputrange 0 polarity 0 driveais 0 6. Call MIO_Config to enable dithering on the E Series device. Set the following parameters: devicenumber The value assigned by MAX dither 1 useamux 0 SCXI-1125 Calibration Procedure 4 ni.com

7. Connect the calibrator to the appropriate analog input, starting with channel 0. Refer to Figure 1 to determine which pins on the 96-pin front connector correspond to the positive and negative inputs of the specified channel. For example, the positive input for channel 0 (CH0+) is pin A32. The negative input for channel 0 (CH0 ) ispin C32. If you are using an SCXI-1320 terminal block connected to the SCXI-1125, connect the calibrator to the CH0+ and CH0 inputs. 8. Call SCXI_Single_Chan_Setup to configure the module for single-channel measurements. Set the following parameters: SCXIchassisID The value assigned by MAX moduleslot 1 modulechan The same channel number from step 7 DAQdeviceNumber The device number assigned by MAX for the E Series DAQ device 9. Call SCXI_Set_Gain to set the gain you want to test on the module. Set the following parameters: SCXIchassisID The ID obtained when the chassis is configured with MAX moduleslot 1 channel The channel number from step 7 gain The first gain value you want to test you can find this gain value in Table 1. 10. Call SCXI_Configure_Filter to configure the filter setting on the module. Set the following parameters: SCXIchassisID The ID obtained when the chassis is configured with MAX moduleslot 1 channel The channel number from step 7 filtermode 1 cutofffrequency 10,000 11. Set the calibrator to output the test point voltage from Table 1. 12. Call DAQ _Op. Set the following parameters: devicenumber ThedevicenumberoftheDAQdevice channel 0 gain 1 for 16-bit E Series devices or 1 for 12-bit E Series devices count 5,000 samplerate 10,000 National Instruments Corporation 5 SCXI-1125 Calibration Procedure

13. Call SCXI_Scale. Set the following parameters: SCXIChassisID The value assigned by MAX moduleslot 1 channel The same channel number from step 7 SCXIgain The same gain value from step 9 TBgain 1 DAQboard The device number for the DAQ device DAQChannel 0 DAQgain 1 for 16-bit devices, 1 for 12-bit devices numpoints 5.000 binarray The array returned from the DAQ_Op function The result is a set of scaled voltages that you read from the SCXI module. 14. Average the results returned from the SCXI_Scale function. Compare the averaged result to the upper and lower limits in Table 1. If the result falls within these limits, your module has passed this portion of the test. If it does not, your module needs to be adjusted using the steps in the next section, Adjusting for Gain and Offset Errors. 15. Repeat steps 8 through 14 for the remaining gains. 16. Repeat steps 7 through 15 for the remaining channels. You have now finished verifying your SCXI-1125. Note If all channels fall within the upper and lower limits, your device does not need to be adjusted. Adjusting for Gain and Offset Errors The following two sections explain how to determine the gain and offset errors of your SCXI-1125 and how to adjust the calibration constants to correct for these errors. Determining Gain and Offset Errors Complete the following steps to determine the gain and offset errors of the SCXI-1125: 1. Read the Test Conditions section. 2. Connect the E Series device to the SCXI-1125. 3. Call Calibrate_E_Series to minimize the uncertainty associated with the E series device measurements. Set the following parameters: devicenumber The value assigned by MAX SCXI-1125 Calibration Procedure 6 ni.com

calop ND_SELF_CALIBRATE setofcalconstants ND_USER_EEPROM_AREA calrefvolts 0.0 4. Call AI_Configure to put the E Series device in bipolar mode. Set the following parameters: devicenumber The value assigned by MAX chan 0 inputmode 0 inputrange 0 polarity 0 driveais 0 5. Call MIO_Config to enable dithering on the E Series device. Set the following parameters: devicenumber The value assigned by MAX dither 1 useamux 0 6. Call SCXI_Single_Chan_Setup to configure the module for single-channel measurements. Set the following parameters: SCXIchassisID The value assigned by MAX moduleslot 1 modulechan The channel you want to adjust DAQdeviceNumber The device number assigned by MAX for the E Series DAQ device 7. Call SCXI_Set_Gain. Set the following parameters: SCXIchassisID The ID of the SCXI chassis obtained when you configure the chassis with MAX moduleslot 1, because the SCXI-1125 should be installed in slot 1 channel The same channel number from step 6 gain The gain value you want to adjust. Refer to Table 1 for acceptable gain values 8. Call SCXI_Configure_Filter to configure the filter setting on the module. Set the following parameters: SCXIchassisID The ID of the SCXI chassis obtained when you configure the chassis with MAX moduleslot 1 channel The same channel number from step 7 National Instruments Corporation 7 SCXI-1125 Calibration Procedure

filtermode 1 cutofffrequency 10,000 9. Connect the calibrator to the analog input you want to adjust. Refer to Figure 1 to determine the pins on the 96-pin front connector that correspond to the positive and negative inputs of the specified channel. For example, the positive input for channel 0 is pin A32, which is labeled CH0+. The negative input for channel 0 is pin C32, which is labeled CH0. If you are using an SCXI-1320 terminal block connected to the SCXI-1125, connect the calibrator to the CH0+ and CH0 inputs. 10. Connect the DMM to the positive output on pin 3 (MCH0+) and negative output on pin 4 (MCH0 ) on the rear panel connector. Figure 2 shows the 50-pin rear panel connector. Tip For easy access to individual pins, use a screw terminal adapter connected to a 50-pin terminal block. 11. Set the calibrator voltage to the positive test point value in Table 1. 12. Read the voltage from the DMM. Record the actual DMM reading, which is output1, and the calibrator output voltage, which is volt1,for later use. 13. Set the calibrator to the negative test point value for the same gain. Skip any input limits that are specified as 0.0 V. You need only upper and lower limits for adjustment. 14. Read the voltage from the DMM. Record the actual DMM reading, output2, and the calibrator output voltage, voltage2. You now have two pairs of data points, (volt1, output1) and (volt2, output2), where volt is the calibrator reading and output is the DMM reading. 15. Convert the DMM voltage readings (output1 and output2)tobinary readings (binary1 and binary2) by using one of the following equations: binary reading for 12-bit E Series products (binary) = output ---------------- 2 20 12 output binary reading for 16-bit E Series products (binary) = ---------------- 2 20 16 For example, using a 12-bit product such as a PCI-MIO-16E-1 and obtaining a DMM reading of 9.9 V, you would have the following results: binary reading =( 9.9/20) 2 12 = 2027.52 SCXI-1125 Calibration Procedure 8 ni.com

16. Record binary1 with volt1 and binary2 with volt2 for later use. 17. Repeat steps 7 through 16 for the remaining gain values you want to adjust on this channel. 18. Repeat steps 6 through 17 for the remaining channels you want to adjust. You have determined the new calibration constants. Use these values in the the following section, Adjusting Calibration Constants. Adjusting Calibration Constants Complete the following steps to adjust the calibration constants on the SCXI-1125: 1. Call SCXI_Cal_Constants to create and store the new calibration constants in the SCXI-1125 memory. Set the following parameters: SCXIchassisID TheIDassignedbyMAX ModuleSlot 1, because the SCXI-1125 should be installed in slot 1 channel The channel on the SCXI-1125 you want to adjust OpCode 2 calibrationarea 0 rangecode 0 (not used for the SCXI-1125) SCXIgain The gain you want to adjust DAQboard The device number assigned by MAX DAQChan 0 DAQGain 1 for 16-bit devices, 1 for 12-bit devices Tbgain 1.0 volt1 and volt2 The first and second voltage readings for the gain values that you are about to save on the SCXI-1125 binary1 and binary2 The first and second binary readings for the gain values that you are about to save on the SCXI-1125 calconst1 and calconst2 These are return values 2. Repeat step 1, changing calibrationarea to 1. 3. Repeat steps 1 and 2 for the next channel. You have now finished adjusting the calibration constants of your SCXI-1125. National Instruments Corporation 9 SCXI-1125 Calibration Procedure

Verifying Adjusted Values After completing the adjustments to your SCXI-1125 module, it is important that you verify the analog input operation by repeating the steps listed in the Verifying Gain and Offset Errors section of this document. Reverifying after adjustments ensures that your SCXI-1125 module is now operating within its specifications. SCXI-1125 Calibration Procedure 10 ni.com

Front and Rear Signal Connector Diagrams A B C CH0+ 32 32 CH0 CH1+ 30 30 CH1 28 28 CH2+ 26 26 CH2 CH3+ 24 24 CH3 22 22 CH4+ 20 20 CH4 CH5+ 18 18 CH5 16 16 CH6+ 14 14 CH6 CH7+ 12 12 CH7 10 10 RSVD 8 8 RSVD RSVD 6 6 RSVD +5 V 4 4 CJTEMP CHSGND 2 2 RSVD This symbol indicates that no physical pin is present Figure 1. Front Signal Connector Pin Assignments National Instruments Corporation 11 SCXI-1125 Calibration Procedure

AOGND MCH0+ MCH1+ MCH2+ MCH3+ MCH4+ MCH5+ MCH6+ MCH7+ OUTREF SERDATIN DAQD*/A SLOT0SEL* DIGGND SERCLK 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 AOGND MCH0 MCH1 MCH2 MCH3 MCH4 MCH5 MCH6 MCH7 DIGGND SERDATOUT SCALK SY Figure 2. Rear Signal Connector Pin Assignments SCXI-1125 Calibration Procedure 12 ni.com

Specifications The following table contains the specifications you should use when verifying and calibrating your SCXI-1125 module. If your device has been calibrated within the last year, the test point value should fall between the Upper Limit and Lower Limit values. Table 1. SCXI-1125 Specifications at 25 C ±10 C Gain Range Test Point Upper Limit (V) Lower Limit (V) 1 5 4.950000 4.9623181 4.9376819 1 0 0.000000 0.005002 0.005002 1 5 4.950000 4.9376819 4.9623181 2 2.5 2.475000 2.4811601 2.4688400 2 0 0.000000 0.002502 0.002502 2 2.5 2.475000 2.4688400 2.4811601 5 1 0.990000 0.9924652 0.9875348 5 0 0.000000 0.001002 0.001002 5 1 0.990000 0.9875348 0.9924652 10 0.5 0.495000 0.4962336 0.4937664 10 0 0.000000 0.000502 0.000502 10 0.5 0.495000 0.4937664 0.4962336 20 0.25 0.247500 0.24811781 0.24688220 20 0 0.000000 0.000252 0.000252 20 0.25 0.247500 0.24688220 0.24811781 50 0.1 0.099000 0.09924832 0.09875168 50 0 0.000000 0.000102 0.000102 50 0.1 0.099000 0.09875168 0.09924832 100 0.05 0.0495000 0.04962516 0.04937484 100 0 0.000000 0.000052 0.000052 100 0.05 0.049500 0.04937484 0.04962516 200 0.025 0.024750 0.02481358 0.02468642 200 0 0.000000 0.000027 0.000027 National Instruments Corporation 13 SCXI-1125 Calibration Procedure

Table 1. SCXI-1125 Specifications at 25 C ±10 C (Continued) Gain Range Test Point Upper Limit (V) Lower Limit (V) 200 0.025 0.024750 0.02468642 0.02481358 250 0.02 0.019800 0.01985126 0.01974874 250 0 0.000000 0.000022 0.000022 250 0.02 0.019800 0.01974874 0.01985126 500 0.01 0.009900 0.00992663 0.00987337 500 0 0.000000 0.000012 0.000012 500 0.01 0.009900 0.00987337 0.00992663 1000 0.005 0.00495 0.00496432 0.00493568 1000 0 0.000000 0.000007 0.000007 1000 0.005 0.004950 0.00493568 0.00496432 2000 0.0025 0.002475 0.00248316 0.00246684 2000 0 0.000000 0.000005 0.000005 2000 0.0025 0.002475 0.00246684 0.00248316 Limits Valid for 4 Hz or 10 khz Filter Settings Valid when using an MIO device after performing an MIO internal calibration. Valid for both filter settings with 5.000 point averaging at 10,000 samples per second. SCXI-1125 Calibration Procedure 14 ni.com