NI 6703/6704 Calibration Procedure for Traditional NI-DAQ

Similar documents
NI 6703/6704 Calibration Procedure for NI-DAQ mx

SCXI Contents CALIBRATION PROCEDURE

CALIBRATION PROCEDURE NI PXIe Channel, Isolated Voltage/Current Analog Output Module

Contents. Software Requirements. CALIBRATION PROCEDURE NI PXIe-4357

Contents. Software. CALIBRATION PROCEDURE NI 9216/ Channel, 24-Bit, 100 Ω/1000 Ω RTD Analog Input Module

CALIBRATION PROCEDURE NI channel, ±2- ma/±10 V, 24-bit Analog Input Module. ni.com/manuals

SCXI Introduction CALIBRATION PROCEDURE. What Is Calibration? Why Should You Calibrate? How Often Should You Calibrate?

CALIBRATION PROCEDURE SCXI Contents. Software Requirements. ni.com/manuals

CALIBRATION PROCEDURE PXIe channel, 16-bit, 400 ks/s/ch, Ch-Ch Isolated Analog Input Module. ni.com/manuals

Contents. Software. CALIBRATION PROCEDURE NI Channel, ±5 V, 24-Bit Software-Selectable IEPE and AC/DC Analog Input Module

Contents. Software. CALIBRATION PROCEDURE NI Channel, 300 Vrms, 24-Bit, Simultaneous, Channel-to-Channel Isolated Analog Input Module

FP-RTD-122. Introduction CALIBRATION PROCEDURE. Version 2.0

SCXI Equipment and Other Test Requirements CALIBRATION PROCEDURE. Test Equipment. Software and Documentation

CALIBRATION PROCEDURE CAL Contents. Software

The following conventions appear in this document:

NI 9213 Calibration Procedure

NI PCIe-784xR, NI PXI-784xR, NI PXIe-784xR, NI USB-784xR, NI PCIe-785xR, NI PXI-785xR, NI PXIe-785xR, NI USB-785xR, and NI PXIe-786xR

NI R Series. Contents. Software CALIBRATION PROCEDURE

SCXI -1104/C. Contents CALIBRATION PROCEDURE. For NI-DAQmx

SCXI -1102/B/C. Contents CALIBRATION PROCEDURE. For NI-DAQmx

±10 V, 20 MS/s/ch, 14-Bit, 4-Channel C Series Digitizer Module

E/S/M/B Series Calibration Procedure for NI-DAQ mx

Calibrating FieldPoint Analog Input Modules with Calibration Executive

ESERIES CALIBRATION FIXTURE INSTALLATION GUIDE

Contents. Software CALIBRATION PROCEDURE NI TC-4353

Contents. Software Requirements CALIBRATION PROCEDURE NI TB-4353

Where to Start with NI ELVIS

To calibrate the NI 5105, install the following versions of NI-SCOPE. NI-SCOPE Version

NI PXI Contents CALIBRATION PROCEDURE

NI Introduction CALIBRATION PROCEDURE. What Is Calibration? External Calibration

IMAQ Vision Deployment Engine Note to Users

IMAQ Vision Deployment Engine Note to Users

SCC-AO10 Isolated Analog Output Module

NI 446X Calibration Procedure

NI PXI Contents CALIBRATION PROCEDURE

LabVIEW Real-Time Module for Mac OS X User Manual Addendum

The NI TB-2648/2649/2650/2651 terminal blocks configure your NI PXI/PXIe-2531 (NI 2531) in one of the following topologies. 1-wire 8 64 matrix

Programming for the LabVIEW Real-Time Module Using LabWindows/CVI

SCB-68 User Guide. 68-Pin Shielded Desktop Connector Block

Where to Start with NI ELVIS

NI CB-37F-LP. Introduction INSTALLATION GUIDE. Connector Block for 37-Pin D-Sub Devices

SCXI -1313A Terminal Block

Getting Started with R Series Intelligent DAQ

R Series Digital I/O Module for PCI Express, 128 DIO, 512 MB DRAM, Kintex-7 160T FPGA. This document describes how to begin using the PCIe-7821R.

M Series USB-6221/6225/6229/6251/6255/6259 OEM Devices

OPERATING INSTRUCTIONS AND SPECIFICATIONS. NI 9208E 16-Channel, ±20 ma, 24-Bit Analog Input Module

TBX-1316 High-Voltage Attenuator Terminal Block

Contents. CALIBRATION PROCEDURE NI PXIe-5624R 2 GS/s, 12-Bit IF Digitizer

NI USB-622x/625x/628x OEM

12-Channel, ±24 V Precision PXI Source Measure Unit

NI-Industrial Communications for EtherCAT

NI USB-6211/6218 OEM

SH Cable Specifications

24-Channel, ±24 V Precision PXI Source Measure Unit

FieldPoint. FP-1000/1001 FieldPoint Quick Start Guide. Install the Network Module Mount the module on a DIN rail. What You Need to Get Set Up

NI-DAQ mx for USB Devices

Electrical. SPECIFICATIONS SH Cable

High-Precision Data Acquisition

E Series Multifunction I/O 1.25 MS/s, 12-Bit, 16 or 64 Analog Inputs

Getting Results with CompactRIO and LabVIEW

NI REM Remote I/O Overview DATASHEET. Analog Output Module for Remote I/O

SCXI OFFSET-NULL AND SHUNT-CALIBRATION HIGH-VOLTAGE TERMINAL BLOCK INSTALLATION GUIDE

R Series for PXI Express Digital RIO with Kintex-7 325T FPGA. ni.com/manuals. This document explains how to install and configure the NI PXIe-7822R.

R Series for USB Multifunction RIO with Kintex-7 70T FPGA. ni.com/manuals

NI PXIe-4080/4081/4082

Getting Started with R Series Multifunction RIO

R Series for USB Multifunction RIO with Kintex-7 160T FPGA. ni.com/manuals

LabVIEW Application Builder User Guide

NI Contents. Hardware Overview INSTALLATION GUIDE. NI 8221 Features

NI USRP-2900/2901. Contents. Electromagnetic Compatibility Guidelines GETTING STARTED GUIDE. Universal Software Radio Peripheral

NI USB Introduction USER GUIDE AND SPECIFICATIONS. 4-Channel, 0 20 ma, 16-Bit Analog Current Output Module

Getting Started with the LabVIEW Real-Time Module

Getting Started with the NI 783x R

R Series for USB Multifunction RIO with Kintex-7 160T FPGA. ni.com/manuals. Electromagnetic Compatibility Guidelines

PXIe Contents. Verifying the System Requirements GETTING STARTED GUIDE. 32-Channel Digital Pattern Instrument

NI USB Contents. Electromagnetic Compatibility Guidelines GETTING STARTED GUIDE. 10 MHz to 18 GHz RF Power Meter

USER MANUAL NI 9755 NI Powertrain Controls CompactRIO NOx Sensor Module Kit

USER GUIDE USB-634x/635x/636x OEM X Series USB-6341/6343/6356/6361/6363/6366 OEM Devices

Related Information Refer to the NI Switches Help for detailed information about features and programming options for your NI switch product.

NI-DAQ mx Base 1.x. Contents GETTING STARTED GUIDE

NI-DAQmx: Reduce Development Time and Improve Performance

NI PXI Contents GETTING STARTED GUIDE. 8 GHz RF Amplifier

NI 653x Cable Adapter

This installation guide describes how to install and connect signals to the NI TBX-2808 screw terminal accessory for use with NI SwitchBlock cards.

NI PXIe-1062Q Power Supply Shuttle

GETTING STARTED GUIDE

NI PXIe-4138/4139. Contents. Electromagnetic Compatibility Guidelines GETTING STARTED GUIDE. Single-Channel System SMU

VXI-MXI-Express Series

NI PXI/PXIe/PCI Contents GETTING STARTED GUIDE. 14-Bit, 100 MS/s PXI Oscilloscope

PXI/PCI Contents GETTING STARTED GUIDE. 2 GS/s Oscilloscope

Introduction... 1 System Diagram... 2 Hardware... 2 Powering the Hardware... 3 Wiring Harness... 5

GETTING STARTED GUIDE NI AO, 0 ma to 20 ma, 16 Bit Simultaneous

To access NI 5622 documentation, navigate to Start»All Programs»National Instruments» NI-SCOPE»Documentation.

NI WSN-9791 Ethernet Gateway

Electromagnetic Compatibility Guidelines

You need the following to set up and use your terminal block:

NI REM Remote I/O Overview DATASHEET. Bus Coupler for Remote I/O

NI PXI-2567 Specifications

R Series Reconfigurable I/O Module (AI, AO, DIO) for PXI Express, 6 AI, 18 AO, 48 DIO, 1 MS/s AIO, 512 MB DRAM, Kintex-7 160T FPGA

NI USB Introduction USER GUIDE AND SPECIFICATIONS. 4-Channel, ±10 V, 16-Bit Analog Voltage Output Module

Transcription:

NI 6703/6704 Calibration Procedure for Traditional NI-DAQ Introduction This document contains instructions for calibrating the NI 6703/6704 for PCI/PXI/CompactPCI using Traditional NI-DAQ. Use this calibration procedure with the ni6704cal.dll file, which contains specific functions required for calibrating the NI 6703/6704. The NI 6703/6704 should be calibrated at a regular interval as defined by the measurement accuracy requirements of your application. National Instruments recommends that you perform a complete calibration at least once every year. You can shorten this interval to 90 days or six months. Note Refer to ni.com/support/calibrat/mancal.htm for a copy of the ni6704cal.dll file. Calibration Options: External Versus Internal NI 6703/6704 products have two calibration options, an internal or self-calibration, and an external calibration. Internal Calibration The NI 6703/6704 is continuously applying corrections to the output levels based on previous calibration results. Both the NI 6703 and the NI 6704 have 16 voltage outputs and two voltage calibration channels. The NI 6704 also has 16 current outputs and two current calibration channels. All the output and calibration channels are scanned and refreshed together. The outputs of the calibration channels for the NI 6703/6704 are continuously compared to the onboard reference, and the offset and gain of the 16-bit digital-to-analog converter (DAC) are adjusted to minimize the error of the calibration channels. Because the NI 6703 and NI 6704 have excellent channel-to-channel matching of offset and gain errors, minimizing the error on the calibration channels also minimizes the error for all of the voltage and current outputs.

External Calibration External calibration requires a high-precision digital multimeter (DMM) to measure error on the NI 6703/6704 outputs. The error is then corrected with a set of calibration constants that adjust the offset and gain of the 16-bit DAC to ensure that the output voltages and currents are within the device specifications. Because of the NI 6703/6704 calibration circuitry, the only calibration adjustment you must make is to the values of the calibration channels to account for time- or temperature-related drift of the onboard reference. Equipment and Other Test Requirements This section describes the equipment, test conditions, documentation, and software you need to calibrate the NI 6703/6704. Test Equipment To calibrate the NI 6703/6704, you need a high-precision DMM that is at least 10 ppm (0.001%) accurate. NI recommends that you use the Hewlett-Packard 3458A DMM for calibration. If you do not have a Hewlett-Packard 3458A DMM, use the accuracy specifications to select a substitute calibration standard. If you do not have custom connection hardware, you may need a connector block such as the NI CB-68 and a cable such as the SH6868-D1. These components give you easy access to the individual pins on the 68-pin I/O connector. Test Conditions Follow these guidelines to optimize connections and test conditions during calibration: Keep connections to the NI 6703/6704 short. Long cables and wires act as antennae, picking up extra noise that can affect measurements. Use shielded copper wire for all cable connections to the device. Use twisted-pair wire to eliminate noise and thermal offsets. Maintain a temperature between 18 and 28 C. To operate the module at a specific temperature outside this range, calibrate the device at that temperature. Keep relative humidity below 80%. Allow a warm-up time of at least 15 minutes to ensure that the measurement circuitry is at a stable operating temperature. NI 6703/6704 Calibration Procedure for Traditional NI-DAQ 2 ni.com

Software Because the NI 6703/6704 is a PC-based measurement device, you must have the proper device driver installed in the calibration system before attempting calibration. For this calibration procedure, you need the latest version of NI-DAQ installed on the calibration computer. NI-DAQ, which configures and controls the NI 6703/6704, is available at ni.com/downloads. NI-DAQ supports a number of programming languages, including LabVIEW, LabWindows /CVI, Microsoft Visual C ++, Microsoft Visual Basic and Borland C ++. When you install the driver, you only need to install support for the programming language that you intend to use. You also need a copy of the ni6704cal.dll, ni6704cal.lib, and ni6704cal.h files. The DLL provides calibration functionality that does not reside in NI-DAQ, including the ability to protect the calibration constants, update the calibration date, and write to the factory calibration area. The factory calibration area and the calibration date should only be modified by a metrology laboratory or another facility that maintains traceable standards. You can access the functions in this DLL through any 32-bit compiler. Configuring the NI 6703/6704 in NI-DAQ The NI 6703/6704 must be configured in NI-DAQ, which automatically detects the device. The following steps briefly explain how to configure NI-DAQ. Refer to the PCI/PXI-6703/6704 User Manual for detailed installation instructions. You can install this manual when you install NI-DAQ. 1. Power off the computer. 2. Install the NI 6703/6704 in an available slot. 3. Power on the computer. 4. Launch Measurement & Automation Explorer (MAX). 5. Configure the NI 6703/6704 device number. 6. Click the Test Resources button to ensure that the NI 6703/6704 is working properly. The NI 6703/6704 is now configured. Note After a device is configured in MAX, the device is assigned a device number. This number is used in each of the function calls to identify which DAQ device is to be calibrated. National Instruments Corporation 3 NI 6703/6704 Calibration Procedure for Traditional NI-DAQ

Writing the Calibration Procedure The calibration procedure in the Calibrating the NI 6703/6704 section provides instructions for calling the appropriate calibration functions. These calibration functions are C function calls from NI-DAQ that are also valid for Visual Basic programs. Although LabVIEW VIs are not discussed in this procedure, you can program in LabVIEW using the VIs that are named similarly to the NI-DAQ function calls in this procedure. Refer to the Flowcharts section for illustrations of the code used at each step of the calibration procedure. Often you must follow a number of compiler-specific steps to create an application that uses NI-DAQ. Refer to the NI-DAQ User Manual for PC Compatibles at ni.com/manuals for details about the required steps for each of the supported compilers. Many of the functions listed in the calibration procedure use variables that are defined in the nidaqcns.h file. To use these variables, you must include the nidaqcns.h file in the code. If you do not want to use these variable definitions, you can examine the function call listings in the NI-DAQ documentation and the nidaqcns.h file to determine what input values are required. Documentation For information about NI-DAQ, refer to the following documentation: Traditional NI-DAQ Function Reference Help (Start»Programs» National Instruments»NI-DAQ»Traditional NI-DAQ Function Reference Help) NI-DAQ User Manual for PC Compatibles at ni.com/manuals Calibrating the NI 6703/6704 These two documents provide detailed information about using NI-DAQ. The function reference help includes information about the functions in NI-DAQ. The user manual provides instructions on installing and configuring DAQ devices and detailed information about creating applications that use NI-DAQ. These documents are the primary references for writing the calibration utility. For further information about the products you are calibrating, you may also want to install the PCI/PXI-6703/6704 User Manual. To calibrate the NI 6703/6704, complete the following steps: 1. Verify the performance of the NI 6703/6704. This step, which is covered in the Verifying the Performance of the NI 6703/6704 section, confirms whether the device is in specification prior to adjustment. NI 6703/6704 Calibration Procedure for Traditional NI-DAQ 4 ni.com

2. Adjust the NI 6703/6704 calibration constants with respect to a known voltage source. This step is described in the Adjusting the NI 6703/6704 section. 3. Reverify the performance to ensure that the NI 6703/6704 is operating within its specifications after adjustment. Verifying the Performance of the NI 6703/6704 Verification determines how well the device is meeting its specifications. The verification procedure is divided into the major functions of the device. Throughout the verification process, refer to the tables in the Specifications section to see if the device needs to be adjusted. Verifying Voltage Output This procedure verifies whether the output of voltage channel 0 is within specifications. Because of the channel-to-channel matching of the NI 6703/6704 voltage channels, if voltage channel 0 is operating within specifications, all the voltage channels should be operating within specifications. If you want to verify the channels individually, this procedure can be easily expanded to verify all the voltage channels. Perform the procedure once for each value of the channel parameter from 0 to 15. Each time you repeat the procedure, you also need to connect the appropriate channel to the DMM. Refer to the PCI/PXI-6703/6704 User Manual at ni.com/manuals for the pin numbers corresponding to each channel. To verify the voltage output, read the Equipment and Other Test Requirements section, and complete the following steps: 1. Call the Get_Cal_Date function contained in ni6704cal.dll to determine the previous calibration date of the NI 6703/6704. 2. Connect the DMM to the NI 6703/6704 as shown in the following table. Signal Name Description NI 6703/6704 Output DMM Input VCH0 Voltage Channel 0 Pin 34 High AGND0 Analog Ground 0 Pin 68 Low 3. Place the DMM in voltage mode. 4. Call AO_VWrite to generate 9.9 V on voltage channel 0. Set the chan: 0 voltage: 9.9 National Instruments Corporation 5 NI 6703/6704 Calibration Procedure for Traditional NI-DAQ

5. Compare the resulting value shown by the DMM to the upper and lower limits listed in Table 1 in the Specifications section. If the value is between these limits, the device passes the test. 6. Call AO_VWrite to generate 0.0 V on voltage channel 0. Set the chan: 0 voltage: 0.0 7. Compare the resulting value shown by the DMM to the upper and lower limits listed in Table 1 in the Specifications section. If the value is between these limits, the device passes the test. 8. Call AO_VWrite to generate 9.9 V on voltage channel 0. Set the chan: 0 voltage: 9.9 9. Compare the resulting value shown by the DMM to the upper and lower limits listed in Table 1 in the Specifications section. If the value is between these limits, the NI 6703/6704 passes voltage verification. 10. Disconnect the DMM from the NI 6703/6704. You have now verified the voltage output of the NI 6703/6704. Verifying Current Output (NI 6704 Only) This procedure verifies that the current output of channel 16 is within specifications. Because of the channel-to-channel matching of the NI 6704 current channels, if current channel 16 is operating within specifications, all the current channels should be operating within specifications. However, you can verify the channels individually for each current channel. Perform the procedure once for each value of the channel parameter from 16 to 31. You also need to connect the appropriate channel to the DMM each time you repeat the procedure. Refer to the PCI/PXI-6703/6704 User Manual for the pin numbers corresponding to each channel. To verify current output, complete the following steps: 1. Call the Get_Cal_Date function to determine the previous calibration date of the NI 6704. 2. Connect the DMM to the NI 6704 as shown in the following table. Signal Name Description NI 6704 Output DMM Input ICH16 Current Channel 16 Pin 67 High AGND16 Analog Ground 16 Pin 68 Low NI 6703/6704 Calibration Procedure for Traditional NI-DAQ 6 ni.com

Adjusting the NI 6703/6704 3. Place the DMM in current mode. 4. Call AO_VWrite to generate 0.1 ma on current channel 16. Set the chan: 16 voltage: 0.1 5. Compare the resulting value shown by the DMM to the upper and lower limits listed in Table 2 in the Specifications section. If the value is between these limits, the device passes the test. 6. Call AO_VWrite to generate 20.0 ma on current channel 16. Set the chan: 16 voltage: 20.0 7. Compare the resulting value shown by the DMM to the upper and lower limits listed in Table 2 in the Specifications section. If the value is between these limits, the device passes current verification. 8. Disconnect the DMM from the NI 6704. You have now verified the current output of the NI 6704. At the end of each calibration procedure, new constants are stored in the user calibration area of the NI 6703/6704, ensuring that the device stores a permanent record of the calibration constants. Storing the calibration constants here also ensures that the constants automatically load when the NI 6703/6704 is powered on. Adjusting Voltage Output For the voltage outputs, you need to make offset and gain adjustments. Adjust offsets first, followed by gains. Adjust offset and gain again to ensure accuracy. This second set of offset and gain adjustments is necessary because a large initial change in one of the calibration channels can affect the accuracy of the other calibration channel. The following table lists the offset and gain channel numbers. Calibration Channel NI 6703 Channel Numbers NI 6704 Channel Numbers Voltage Offset 16 32 Voltage Gain 17 34 National Instruments Corporation 7 NI 6703/6704 Calibration Procedure for Traditional NI-DAQ

To measure voltage outputs, complete the following steps: 1. Call Set_DAQ_Device_Info to configure the NI 6703/6704 so that you can access the calibration channels. Set the infotype: ND_CALIBRATION_ENABLE infovalue: ND_YES 2. Connect the DMM to the NI 6703/6704 as shown in the following table. Signal Name Description NI 6703/6704 Output DMM Input VCH0 Voltage Channel 0 34 High AGND0 Analog Ground 0 68 Low 3. Place the DMM in voltage mode. 4. Call AO_VWrite to generate 0.0 V on voltage channel 0. Set the chan: 0 voltage: 0.0 5. Call AO_VWrite with the channel set to the voltage offset channel. Adjust the voltage parameter until the DMM reads as close to 0.0 V as possible. The value you write to the voltage offset channel is close to 0.0 V. Increasing the voltage value lowers the magnitude of the voltage on voltage channel 0. Decreasing the voltage value increases the magnitude of the voltage on voltage channel 0. 6. Call AO_VWrite to generate 10.0 V on voltage channel 0. Set the chan: 0 voltage: 10.0 7. Call AO_VWrite with the channel set to the voltage gain channel. Adjust the voltage parameter until the DMM reads as close to 10.0 V as possible. The value you write to the voltage gain channel is close to 10.0 V. Increasing the voltage value lowers the magnitude of the voltage on voltage channel 0. Decreasing the voltage value increases the magnitude of the voltage on voltage channel 0. You have now adjusted the voltage output. Adjusting Current Output (NI 6704 Only) For the current outputs, you need to adjust offset and gain. Adjust offset first, followed by gain. Adjust both offset and gain again to ensure accuracy. This second set of offset and gain adjustments is necessary NI 6703/6704 Calibration Procedure for Traditional NI-DAQ 8 ni.com

because a large initial change in one of the calibration channels can affect the accuracy of the other calibration channel. Refer to the offset and gain channel numbers in the following table. Calibration Channel To adjust current output, complete the following steps: 1. Call Set_DAQ_Device_Info to configure the NI 6704 so you can access the calibration channels. Set the infotype: ND_CALIBRATION_ENABLE infovalue: ND_YES NI 6704 Channel Numbers Current Offset 33 Current Gain 35 2. Connect the DMM to the NI 6704 as shown in the following table. Signal Name Description NI 6704 Output DMM Input ICH16 Current Channel 16 Pin 67 High AGND16 Analog Ground 16 Pin 68 Low 3. Place the DMM in current mode. 4. Call AO_VWrite to generate 0.1 ma on current channel 16. Set the chan: 16 voltage: 0.1 5. Call AO_VWrite with the channel set to the current offset channel. Adjust the voltage parameter until the DMM reads as close to 0.1 ma as possible. The value you write to the current offset channel is close to 0.01 ma. Increasing the voltage value lowers the magnitude of the current on current channel 16. Decreasing the voltage value increases the magnitude of the current on current channel 16. 6. Call AO_VWrite to generate 20.0 ma on current channel 16. Set the chan: 16 voltage: 20.0 National Instruments Corporation 9 NI 6703/6704 Calibration Procedure for Traditional NI-DAQ

7. Call AO_VWrite with the channel set to the current gain channel. Adjust the voltage parameter until the DMM reads as close to 20.0 ma as possible. The value you write to the current gain channel is close to 19.608 ma. Increasing the voltage value lowers the magnitude of the current on current channel 16. Decreasing the voltage value increases the magnitude of the current on current channel 16. You have now adjusted the current output. Saving Calibration Values After you have adjusted the calibration constants, you must store the new calibration constants in the user calibration area of the NI 6703/6704 EEPROM to ensure that the new calibration values are permanently saved and automatically loaded when the NI 6703/6704 is powered on. If the calibration is being performed by a metrology laboratory or another facility that maintains traceable standards, the calibration values should also be stored in the factory calibration area. Calibration constants stored in the factory calibration area cannot be overwritten by the user. To save the calibration constants and update the calibration dates, complete the following steps: 1. Call AO_Calibrate to store the new calibration constants in the user calibration area of the EEPROM. Set the operation: 2 calibrationarea: 1 2. Call Set_DAQ_Device_Info to configure the NI 6703/6704 so it is no longer possible to access the calibration channels. Set the following parameters: infotype: ND_CALIBRATION_ENABLE infovalue: ND_NO 3. If the calibration is being performed by a metrology laboratory or another facility that maintains traceable standards, call the Copy_Cal function in the ni6704cal.dll. Calling this function copies the calibration values to the factory calibration area and updates the calibration date to match the system clock of the calibration computer. The NI 6703/6704 is now adjusted. Verifying the New Calibration Constants You can verify the new calibration constants by repeating the Verifying the Performance of the NI 6703/6704 section. NI 6703/6704 Calibration Procedure for Traditional NI-DAQ 10 ni.com

Specifications The following tables are the specifications to use when calibrating the NI 6703/6704. The tables show the specifications for 1-year and 24-hour calibration intervals. Using the Tables The following definitions describe how to use the specification tables in this section. Test Point Value (V) Test Point The test point is the voltage value that is input or output for verification purposes. 24-Hour Ranges The 24-Hour Ranges column contains the upper limits and lower limits for the test point value. If the NI 6703/6704 has been calibrated in the last 24 hours, the test point value should be between the upper and lower limit values. 1-Year Ranges The 1-Year Ranges column contains the upper limits and lower limits for the test point value. If the NI 6703/6704 has been calibrated in the last year, the test point value should be between the upper and lower limit values. Table 1. Voltage Output Specifications 24-Hour Ranges 1-Year Ranges Lower Limit (V) Upper Limit (V) Lower Limit (V) Upper Limit (V) 9.9 9.8991019 9.9008981 9.8989435 9.9010565 0.0 0.0007100 0.0007100 0.0007100 0.0007100 9.9 9.9008981 9.8991019 9.9010565 9.8989435 National Instruments Corporation 11 NI 6703/6704 Calibration Procedure for Traditional NI-DAQ

Test Point Value (ma) Flowcharts Voltage Output Verification Table 2. Current Output Specifications (NI 6704 Only) Lower Limit (ma) 24-Hour Ranges Upper Limit (ma) Lower Limit (ma) 1-Year Ranges Upper Limit (ma) 0.1 0.0985616 0.1014384 0.09855 0.10145 20 19.997885 20.002115 19.995565 20.004435 The following flowcharts show the appropriate NI-DAQ function calls for verifying and adjusting the NI 6703/6704. Refer to the Calibrating the NI 6703/6704 section of this document, the Traditional NI-DAQ Function Reference Help, and the NI-DAQ User Manual for PC Compatibles for additional information about the software structure. Refer to Table 1, Voltage Output Specifications, to determine the appropriate ranges for the voltages and currents to verify. Get_Cal_Date (devicenumber)(optional) Connect the DMM to voltage channel 0. AO_VWrite(deviceNumber,0,voltage) Verify that the voltage read by the DMM is within range. Figure 1. Voltage Output Verification NI 6703/6704 Calibration Procedure for Traditional NI-DAQ 12 ni.com

Current Output Verification (NI 6704 Only) Refer to Table 2, Current Output Specifications (NI 6704 Only), to determine the appropriate ranges for the voltages and currents to verify. Connect the DMM to current channel 16. AO_VWrite(deviceNumber,16,voltage) Verify that the current read by the DMM is within range. Figure 2. Current Output Verification (NI 6704 Only) National Instruments Corporation 13 NI 6703/6704 Calibration Procedure for Traditional NI-DAQ

Voltage Output Adjustment Set_DAQ_Device_Info(deviceNumber,ND_CALIBRATION_ENABLE,ND_YES) Connect the DMM to voltage channel 0. AO_VWrite(deviceNumber,0,0.0) NI 6703: AO_VWrite(deviceNumber,16,voltage) Adjust voltage until the DMM reads 0.0 V. NI 6704: AO_VWrite(deviceNumber,32,voltage) Adjust voltage until the DMM reads 0.0 V. AO_VWrite(deviceNumber,0,10.0) NI 6703: AO_VWrite(deviceNumber,17,voltage) Adjust voltage until the DMM reads 10.0 V. NI 6704: AO_VWrite(deviceNumber,34,voltage) Adjust voltage until the DMM reads 10.0 V. Figure 3. Voltage Output Adjustment NI 6703/6704 Calibration Procedure for Traditional NI-DAQ 14 ni.com

Current Output Adjustment (NI 6704 Only) Set_DAQ_Device_Info(deviceNumber,ND_CALIBRATION_ENABLE,ND_YES) Connect the multimeter to current channel 16. AO_VWrite(deviceNumber,16,0.1) AO_VWrite(deviceNumber,33,voltage) Adjust voltage until the DMM reads 0.1 ma. AO_VWrite(deviceNumber,16,20.0) AO_VWrite(deviceNumber,35,voltage) Adjust voltage until the DMM reads 20.0 ma. Figure 4. Current Output Adjustment (NI 6704 Only) National Instruments Corporation 15 NI 6703/6704 Calibration Procedure for Traditional NI-DAQ

Saving Calibration Values AO_Calibrate(deviceNumber,2,1) Set_DAQ_Device_Info(deviceNumber,ND_CALIBRATION_ENABLE,ND_NO) Copy_Cal(deviceNumber) Figure 5. Saving Calibration Values CVI, LabVIEW, National Instruments, NI, ni.com, and NI-DAQ are trademarks of National Instruments Corporation. Product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your CD, or ni.com/patents. 2002 2004 National Instruments Corp. All rights reserved. *370463B-01* 370463B-01 Apr04