MPB SERIES Electronic Balances. r1 v1.09. Operation Manual

Similar documents
NB/NHB Precision Balance

Userʼs Guide. 1200g High Precision Scale. 1200g High Precision Scale. Legal for Trade. Includes RS-232 interface port. UG Version 1.

9. FUNCTION MENU SETTINGS -TION FUNCTION DESCRIPTION DEFAULT. VALUE F0 H-L SEt Lo Set a value for the Low

INTRODUCTION INSTALLATION

CONTENTS INTRODUCTION 1 INSTALLATION 1 KEYS 3 MAXIMUM DISPLAY & DISPLAY RESOLUTION 4 BATTERY STATUS INDICATION 5 POWER ON SETUP 5

BW/BWS/VW/CW/NTW/NSW. Weighing Indicator

Oxford E Series. Instruction Manual

KW serial. User s guide. Bench Scales UGKW-E0302

CONTENTS. PRECAUTIONS Introduction Installation... 7

Sartorius MP8 Series Balances and Scales

MWP SERIES. High Precision

User Manual. Models: BL 303 BL 602 BL 2002 BL 3001 BL 6001 BL 30K1

ANALYTICAL / SEMI-MICRO BALANCES

ANALYTICAL/SEMI-MICRO BALANCES

LOW COST HIGH PERFORMANCE WEIGHING SCALE

Laboratory Balances. Standard. Option (addit. price) KERN Your DKD Calibration Partner. Your specialist dealer

Operating instructions METTLER TOLEDO BD balances

PRECAUTIONS... 4 PREFACE... 6 NAMES AND FUNCTIONS... 6 OPERATIONS... 8

STANDARD OPERATING PROCEDURE ACB SERIES ACB-600H PRECISION BALANCE

Trade Approved Version NMI: S520

Scout Resource Guide Scout Portable Balances

Eclipse EBL Series (P.N , Revision 1.00, Effective Aug 2015) Operating Manual For internal ( i ) and external ( e ) calibration models

Adventurer TM Pro Analytical and Precision Balances

AL-Series Analytical and Toploading Balances

Compact Balances. EK-610j EK-6100j. in an Authorized A&D Dealer

Table of Contents. Laboratory Balances. Instruction Manual - TORBAL AGC Series

Adam Equipment HIGHLAND SERIES SERVICE MANUAL. (Software rev. 2.23)

VWR BALANCES. A-Series Balances for High Precision with Advanced Features. B2-Series Balances are Rugged, Affordable, and Reliable

VWR BALANCES. A-Series Balances for High Precision with Advanced Features. B2-Series Balances are Rugged, Affordable, and Reliable

TRWS series. Waterproof indicator User s guide UGTRWS-E1.1

EW EG-(N) Laboratory Balance. Standard. Option (addit. price) KERN Your DKD Calibration Partner. Your specialist dealer

Adventurer TM Pro. The OHAUS Adventurer TM Pro The Most Complete Balance in Its Class! Analytical and Precision Balances

OCEAN WATERPROOF WEIGHING SCALE USER MANUAL

Adventurer TM Pro Analytical and Precision Balances

OHAUS SCOUT BALANCE SCALE

Pro Balances. Frameless Draughtshield With Removable Doors Provides Convenient Access

Adventurer TM Pro. The OHAUS Adventurer TM Pro The Most Complete Balance in Its Class! Analytical and Precision Balances

AELP PALLET BEAMS SCALES

Adventurer TM Pro Analytical and Precision Balances

1. Introduction Page 2. Specifications Page Installation Page Key Functions Page Operations Page 6. Check Weighing Page 6

EHC-WF. User/Technical Manual. Contents subject to change without notice

Adventurer Pro Analytical and Precision Balances

1. INTRODUCTION This manual contains installation, operation and maintenance instructions. Please read the manual completely before using the balance.

Adventurer TM Pro Analytical and Precision Balances

What s In The Box. Key Descriptions. - ON/OFF - Turns the scale MOXIE ON or OFF. PCS - Enter counting mode. Hold to set sample quantity for average

PLATINUM PRO1001DXP DIGITAL COUNTERTOP / PORTABLE BALANCE

CONTENTS. PRECAUTIONS 4 PREFACE 6 NAMES AND FUNCTIONS 6 OPERATIONS 8 General Weighing 8 Weighing with Tare 10 Counting Mode 10 % Mode 12 Setup Mode 13

Pioneer. Analytical and Precision Balances

The all-new OHAUS EXPLORER. Intelligent. Intuitive. Ingeniously Practical.

Product specifications are subject to change without any obligation on the part of the manufacture.

The contents of this manual and the specifications of the instrument covered by this manual are subject to change for improvement without notice.

GMC-206/215/230 Digital Desktop Weighing Scale Operation Manual

OHAUS Asia-Pacific Dealer Meeting March 23-25, 2010 Hua Hin, Thailand Compact & Counting Scales

QB Series. Operation Manual. Precision Balance. To be the best by every measure

Sartorius Element Series

BW/BWS/CW/VW series. Bench Scales User s guide UGBW/BWS/C W/VW-E1.17

Discovery. Ohaus Discovery. Semi-Micro and Analytical Balances. The Professional s Choice For Analytical Applications!

This document hosted by: High-Precision Tuning Fork Electronic Balance. SJ-E Series. Operation Manual IMPORTANT

DWP-PC SERIES PRICE COMPUTING SCALE User s Guide

DWP-98 Series. Counting scale. User s Guide PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE ATTEMPT TO OPERATE THE INSTRUMENT

The all-new OHAUS EXPLORER. Intelligent. Intuitive. Ingeniously Practical.

Built-in Fanless Ionizer! BM-20. BM-22 (Smart range) BM-252. Series. Micro Analytical Balances BM-200 BM-300 BM g 0.

WARNING CAUTION. This is a hazard alert mark.

EB300 Weighing Scale Operation Manual

Nimbus NBL Series (P.N , Revision 3.00, Effective Aug 2017)

IDS-801 Series Compact Bench Counting Scale. User Manual

Scout Pro. The Latest in the Scout Line of Portable Balances The Ohaus Scout Pro. Portable Electronic Balances

Lab Catalog. Balances and Scales. Perfect Balance - Speed, Performance, Value

USER MANUAL. Industrial balance MARK series KL. USER_MANUAL_SERIE_KL_ (EN)_REV2_0.doc 01 Sep. 09

SCS3MT080, SCS3MT100, SCS3MT120, SHDHS1500 & SPLS1000 User Manual

Ingeniously Practical. Adventurer Product Replacement Cross Reference Analytical and Precision Balances. Adventurer Pro to New Adventurer AX

Advanced Test Equipment Rentals ATEC (2832)

Nimbus NBL Series External Calibration Models (NBL XXXe) (P.N , Revision A1, Jul 2014)

The Only Scale in its Class with Touchless Sensors That Free Up Your Hands!

High-Precision Tuning Fork Electronic Balance. SJ-CE Series. Operation Manual IMPORTANT. To ensure safe and proper use of the balance, please read

1.0 Description. 2.0 Unpacking. 3.0 Installation

AJ Series. Intelligent-Lab NTEP Centigram Balance. Made in Japan

Explorer Pro. The Ohaus Explorer Pro Series, the new standard for performance and value in laboratory balances!

Counting Scales. HC-i Series

Adventurer TM Pro Analytical and Precision Balances

IQ6200/6500 Counting Scales. Installation Manual

VA-6 kg, VA-12 kg, VA-30 kg

WM-25 USER S MANUAL. Manufacturer s Declaration of Conformity

uline.com OVERVIEW OF CONTROLS. k g. Tare Menu. Exit. Print. Units PRINT

Operation Manual. Function Zero/Tare IMPORTANT. To ensure safe and proper use of the balance, please read this manual carefully.

The OHAUS Explorer Pro Series, the new standard for performance and value in laboratory balances!

Operating instruction Precision balances

Discovery. The Professional s Choice For Analytical Applications! Semi-Micro and Analytical Balances

Operating Instructions. ALC Models. Electronic Analytical Balances and Precision Scales

A&D WEIGHING FX-i Series Toploader Balances Capacity: 120 g-3,200 g Readability: g-0.01 g

MI101 / MI102 / MI102S INDICATOR

OHAUS Discovery. The Professional s Choice For Analytical Applications! Semi-Micro and Analytical Balances

The Economical Choice of Balance for Routine Weighing Applications

TWP series. Weighing Indicator Service Manual

Explorer. Analytical, Precision, and High Capacity Balances. Standard Features Include:

GP SERIES GP-12K/GP-20K/GP-22K/GP-30K GP-40K/GP-60K/GP-100K/GP-102K GP-30KS/GP-100KS

Sartorius Economy Series Sartorius Express Series

Laboratory Precision (0.1 g) Balances

Sartorius Element Series

Pro Analytical Balances

Transcription:

MPB SERIES Electronic Balances r1 v1.09 Operation Manual

CONTENT 1. Introduction 1 2. Key and display 2.1 Key Function 1 2.2 Display 1 3. Operation 3.1 Before use 2 3.2 Zeroing the display 2 3.3 Taring 2 3.4 Percent weighing 2 3.5 Parts counting 3 3.6 Accumulation 3 3.7 Calibration 3 4. Battery operation 4 5. Parametering setting 5.1 Enter parameter setting 4 5.2 Setting the weighing unit 4 5.3 Setting the backlight 4 5.4 Setting the communication mode 4 6. Calibration 4 7. Weighing unit listing 5

1. INTRODUCTION The MPB series of electronic balances offers an accurate, fast and versatile range of general purpose balances with counting, and % weighing functions. There are 6 models within the MPB series, with capacities up to 6000g. They have stainless steel weighing platforms placed on an ABS base assembly. The keypads are light touch switches and the displays are large and easy to read liquid crystal type displays (LCD). The LCD has a backlight. All units include automatic zero tracking, automatic tare, and an accumulation facility that allows the count to be stored and recalled as an accumulated total. 2. KEY AND DISPLAY Symbol Key Function ZERO Set the zero point for all subsequent weighing. The display shows zero ENTER Enter when setting parameters or other functions. TARE Tares the scale. Stores the current weight in memory as a tare value, subtracts the tare value from the weight and shows the results. This is the net weight. % PERCENT SMPL MODE To enter percent weighing function. When in percentage mode, press this key to return normal weighing mode To enter Counting Mode. Also use it to return normal weighing mode when counting mode. Allows for selection of different weighing units, such as; kilograms, pounds, ounce or other units can change the current value for parameters or other functions When in counting mode, it allows the weight, unit weight, and count to be seen when parts counting PRINT ON/OFF To print the results to a PC or printer using the optional RS-232 interface. It also adds the value to the accumulation memory if the accumulation function is not automatic Turn on or turn off the power.

MPB Description Stable indicator. Ensure it stable indicator is on before pressing tare, sample, percentage, or accumulation. Low battery; charge immediately. >T >0 < Tare indicator when in net weight mode. Zero indicator 3. OPERATION 3.1 Before use To ensure that measurements are precise, turn on the scale 10 minutes prior to use. Recalibrate the balance when the operat+ing environment has been changed. 3.2 Zero the display You can press the ZERO key at any time to set the zero point from which all other weighing and counting is measured. This will usually only be necessary when the pan is empty. When the zero point is obtained the zero indicator will show. The scale has an automatic re-zeroing function to account for minor drifting or accumulation of material on the platform. If small amounts of weight are shown when the platform is empty, press the ZERO key to re-zero the scale. 3.3 Taring Ensure the display shows zero, and that the zero indicator is on. If needed, press the ZERO key. Place a container on the platform, its weight will be displayed. Press the TARE key to tare the scale. The weight that was displayed is stored as the tare value and that value is subtracted from the display, leaving zero on the display. The "TARE" indicator will be on. As items are added only the weight of the product will be shown. When the container is removed a negative value will be shown. If the scale was tared just before removing the container this value is the gross weight of the container plus all product that was removed. The zero indicator will also be on because the platform returns to the condition it was when the ZERO key was last pressed. 3.4 Percentage weighing The scale will show a sample weight as 100%. Any other weight placed on the scale after, will be displayed as a percentage of the original. sample. E.g. If a 350g object is placed on the scale and the % key is pressed the display will show as 100.00%. If you remove the 350g weight and put a 300g weight on the scale the display will show 85.71%, as 300g is 85.71% of 350g. -2-

Note: the scale may jump by large numbers unexpectedly if small weights are used to set the 100% level. For example if only 23.5g is on a scale with 0.5g increments and the scale is set to 100%, the display will show 100.00%, however a small change of weight will cause the display to jump to 102.13% as one scale division (0.5g) increase to 24.0g will be equivalent to a 2.13% increase. Pressing the % key will return the scale to weighing. 3.5 Parts counting Tare the weight of any container that will be used, leaving the empty container on the scale. Place the number of samples on the scale. The number should match the options for parts counting, 10, 20, 50, 100 or 200 pieces. When the scale is showing weight, pressing the SMPL key will start the parts counting function. Press the SMPL key to begin. The scale will show "SP 10" asking for a sample size of 10 parts. You can press MODE key to select sample quantity: 10, 20, 50, 100, 200 and back to 10. Press the ZERO key when the number shown matches the number of parts used for the sample. As more weight is added the display will show the number of parts (pcs). Press the MODE key to display unit weight (g/pcs), total weight (kg) or the count (pcs). Press the SMPL key to exit parts counting mode. 3.6 Accumulation The weight displayed will be stored in memory when the PRINT key is pressed and the weight is stable, the display will show "ACC 1", then the total in memory for 2 seconds before returning to normal. If the RS-232 interface is installed, the weight will be sent to a printer or PC. Remove the weight, allowing the scale to return to zero and put a second weight on. Press PRINT, the display will show ACC 2 and then show the new total. To view the totals in memory, press the PRINT key when the scale shows zero. The display will show the total number of items "ACC xx" and the total weight before returning to zero. The totals will also be printed via the RS-232 interface. To erase the memory press PRINT to view the totals, followed by the SMPL key to clear the memory. 3.7 Calibration In normal weighing mode, empty the pan, hold the MODE key for 3 second, display will show unload. Wait for several seconds until the display shows LOAD. Put the calibrate weight on the pan. After STABLE indicator turns on, the window will show pass. After selfchecking, the calibration will be complete. Note: you can use half capacity or full capacity calibrate weight to calibrate the balance. You don t need key in the weight value, as the balance will automatically detect the weight. -3-

4. BATTERY OPERATION When the battery is low, the indicator will light up. Change the battery or use an AC adapter after it turns on. When battery indication show, charge the battery immediately. The battery life is approximately 40 hours. To charge the battery simply plug it into the mains power. The balance does not need to be turned on. The battery should be charged for 12 hours for full capacity. The battery indicator flickers when it is charging. When the battery is full, the light stops flickering. Over time, the battery may fail to hold a full charge. If the battery life becomes unacceptable then contact your distributor. 5.1 PARAMETER SETTING 5.1 Enter parameter setting Press MODE key during self-checking to enter parameter setting mode, after enter setting mode, press MODE key to change setting, press ZERO key to confirm setting. Press PRINT key to exit mode 5.2 Setting the weighing unit After entering parameter setting, the display will show f1 unt. Press ZERO key to enter setting. Press ZERO key to set the weighing unit. To return back to F1 UNT, press PRINT key to exit mode. 5.3 Setting the backlight After entering parameter setting, the display will show f1 unt. Press MODE key until display show F2 El. Press ZERO key to enter setting. Press MODE key to select backlight mode, then press ZERO key to confirm selection. To exit mode, press the PRINT key. AU: auto backlight, when load weight is on the pan, backlight will turn on ON: always turn on OFF: always turn off Note: when battery is low, the backlight will not function 5. 4 Setting the communication mode After entering parameter setting, display shows f1 unt, press MODE key until display show F3 com. Press ZERO key to enter setting. Press MODE key to select communication mode, press ZERO key to confirm selection. To exit mode, press the PRINT key. 6. CALIBRATION To enter calibration mode, press the MODE key while it is self-checking. The display will show the first function, "F1 UNT", press the MODE key until display show TECH, press ZERO key to enter, display will show PIN, press SMPL, PRINT, TARE key, and press ZERO key to enter technical parameters setting mode. Press MODE key until display show P2 CAL, press ZERO, the display will show "unload". Remove all weight from the pan and then press the ZERO key to set the initial zero point. Then put the calibration weigh on pan, after stable, calibration finished. -4-

7. Weight Unit Listing Unit mark Unit name Conversion(g) 1 gram g 2 g gram (hi resolution) 3 ct net carat =0.2g 4 lb lb =453.59237g 5 oz ounce =28.349523125g 6 d dram =1.7718451g 7 gn* grain =0.06479891g 8 ozt troy ounce =31.1034768g 9 dwt penny weight =1.55517384g 10 mom momme =3.749996g 11 tl.t tael twn =37.49995g 12 tl.h tael chn =37.799375g 13 Tlj troy tael =37.4290018g 14 t tola =11.6638039g 15 bt Bangladeshi tola 1bt=11.6638039g 1bt=16Ana=96Roti 16 N newton 1N=101.916g