CAUTIONS... 4 INTRODUCTION...

Similar documents
Weighing Indicator 1WMPD

CI-2400BS OWNER S MANUAL

Jewelry Scale HT-500GD 1WMPD

BALANCES.COM. Contents

Trade Approved Version NMI: S520

Cautions for Your Safety

WM+PD Weighing Indicator

Multi-Function Smart Indicator User Manual

! User Manual Rev. 1

AD Weighing Indicator INSTRUCTION MANUAL 1WMPD A

BALANCES.COM. Contents

DWP-98 Series. Counting scale. User s Guide PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE ATTEMPT TO OPERATE THE INSTRUMENT

EXCELL PRECISION CO., LTD.

D-500s Indicator D LED. User Instructions. English Issue September SangR. Page 2. User s manual of Sang D-500s LED

MI101 / MI102 / MI102S INDICATOR

Table of Contents. 3-1 Specifications & Features Front Panel Display Keyboard. 3-3 Rear Panel. 3-4 Power Supply. 4.

CONTENTS. 1. Preface Features Technical Specifications Application of Dummy Plate Usage of Connector & Roller...

BW/BWS Weighing Indicator

NOTE. This is warning & caution mark. This is hazard alert mark. This is useful information mark

PS-IN202 Weighing Indicator User Manual

WM+PD Weighing Indicator

Table of Contents 1. Introduction Features & Main function Technical Specification Measure of Appearance...

UWE-1707 SERIES CONTENTS OPERATION MANUAL ELECTRONIC WEIGHING INDICATOR 1. SPECIFICATIONS 2. INSTALLATION

DWP-PC SERIES PRICE COMPUTING SCALE User s Guide

TRWS series. Waterproof indicator User s guide UGTRWS-E1.1

SAFETY PRECAUTIONS WARNING. Repairs Do not disassemble the scale. Contact your local A&D dealer if your balance needs service or repair.

KW serial. User s guide. Bench Scales UGKW-E0302

The contents of this manual and the specifications of the instrument covered by this manual are subject to change for improvement without notice.

TWP series. Weighing Indicator Service Manual

COMPLETE MANUAL. WEIGHING INDICATOR TI-1200 and TI-1200-S TI-1200 with rechargeable battery.

Weighing Indicator Service Manual V1.24X

TWP-W. Weighing Indicator Service Manual

USER S GUIDE. Weighing Indicator. globalindustrial.com. Use of RS232 Port Using Windows for the Printer Driver, Adjust Computer Settings as Follows:

PRECAUTIONS...4 PREFACE...6 NAMES AND FUNCTIONS...6 INSTALLATIONS...8 OPERATIONS...11 ERROR MESSAGE...13 SPECIFICATIONS...13

Digital Weight Indicator Setup / Operation Manual

MWP SERIES. High Precision

1. Introduction Page 2. Specifications Page Installation Page Key Functions Page Operations Page 6. Check Weighing Page 6

Weighing Indicator USER S GUIDE. globalindustrial.com. globalindustrial.com. 11 Harbor Park Drive Port Washington, NY 11050

PW-II Service Manual PW-II SERVICE MANUAL

CONTENTS. PRECAUTIONS Introduction Installation... 7


Multi Function Weighing Indicator

WUNDER SA.BI.SRL EN USER MANUAL. MULTIFUNCTIONS ELECTRONIC SCALE MOD. W100 LED. User Manual

MC Piece Counting Scale

Optima OP-902 User Manual

part of this manual may be photocopied, reproduced, or translated into another language without the prior written consent of the A&D Company.

GMC-206/215/230 Digital Desktop Weighing Scale Operation Manual

DL Series NEWDOLBO OWNER S MANUAL. Simple Weighing Scale 9002-DL

USER INSTRUCTION MANUAL FOR LOADCELL TRANSMITTER MODEL TDC/I/0550 (SOFTWARE: VER2A) INDEX

Digital Bench Scale. Revision 1.2 September 14, 2000 Contents subject to change without notice.

1. On / Off. 2. Initial reset to zero

Technical Parameter... 4 Installation and Connection... 5

This document hosted by: Extension Load Cell Cable (OP-02)

AD-8923-BCD. Remote Controller (BCD) INSTRUCTION MANUAL 1WMPD

EB300 Weighing Scale Operation Manual

Totalcomp TLI Indicator. Operation and Maintenance Manual Manual

Please keep the following using conditions certainly

Pennsylvania Scale Company Model 400B Battery Powered Digital Indicator Operation & Calibration Manual

Oxford E Series. Instruction Manual

EHC-WF. User/Technical Manual. Contents subject to change without notice

Digital Bench Scale. Revision 1.2 August 24, Contents subject to change without notice.

TI-1600 Series. Digital Indicator. Setup / Operation Manual. Revision 1.5 June 5, 2006

Copyright Western Scale Co. Limited. All rights reserved.

Multi-Function Indicator. Operation Manual (Full Version) PLEASE READ THIS MANUAL VERY CAREFULLY BEFORE OPERATING THIS INSTRUMENT

SY021 Portable Load/Force Meter User instructions

CONTENTS PRECAUTIONS... 3 INSTALLATION... 4 ZERO... 9 TARE PRE-TARE AUTO TARE SELECTION OF WEIGHING UNIT... 16

PRECAUTIONS... 4 PREFACE... 6 NAMES AND FUNCTIONS... 6 OPERATIONS... 8

Weighing Indicator M701. Handrail scales. REV: M1, April,05,2014 (v )

AELP PALLET BEAMS SCALES

Digital Weighing Indicator XK3190-A12 User Manual

SERVICE MANUAL iforks

9. FUNCTION MENU SETTINGS -TION FUNCTION DESCRIPTION DEFAULT. VALUE F0 H-L SEt Lo Set a value for the Low

Calibration & Connectivity

PRW series. Bench Scale. User s guide UGPRW-E1.01

PS-C60K-D. Counting Scale Service Manual

Digital Bench Scale. Revision 8.93 August 12, 1993

DMC-688. When Accuracy Counts

DCS Series. Digital Counting Scale. Setup Manual

OCEAN WATERPROOF WEIGHING SCALE USER MANUAL

Model S200 Series. Revision 1.0 July, Contents subject to change without notice.

ES210. Animal & Vet Scale User s Manual (V1611) Anyload Transducer Co. Ltd

QUICK START GUIDE PACKAGING. 1 x 220 V/12 VDC 1A mains adapter. Accompanying documentation. 1 x K3 weight indicator

BW/BWS/CW/VW series. Bench Scales User s guide UGBW/BWS/C W/VW-E1.17

CONTENTS 1.0 CAUTIONS AND PREPARATION 2.0 DISPLAY AND KEY: 7.0 DEALER S SETTING CAUTIONS PREPARATION DISPLAY

CONTENTS. PRECAUTIONS 4 PREFACE 6 NAMES AND FUNCTIONS 6 OPERATIONS 8 General Weighing 8 Weighing with Tare 10 Counting Mode 10 % Mode 12 Setup Mode 13

CONTENTS. PRECAUTIONS... 4 PREFACE... 6 OVERALL VIEW... 6 DISPLAY & KEYBOARD... 7 OPERATIONS I. Programming Mode II. Sales Mode...

Optima OP-900-LD Score Board Indicator (with Remote Control) User Manual

Digital Measurement Metrology, Inc PRECISION IS OUR VISION

805HP. Handheld Digital Weight Indicator Operations Manual (V1612) Anyload Transducer Co. Ltd Website:

Chapter III Operation Instruction I. Start II. Key Operation III. Weighing Operation

Instruction Manual FH EXT

Load Line 2 Weight Indicator User Manuel

Digital Weighing Indicator SI 4630 (For both External Display and Indicator)

Instruction Manual FH EXT

uline.com OVERVIEW OF CONTROLS. k g. Tare Menu. Exit. Print. Units PRINT

Cảm ơn các bạn đã tham khảo tài liệu cân điện tử từ CÂN ĐIỆN TỬ GIA PHÁT!

SCS3MT080, SCS3MT100, SCS3MT120, SHDHS1500 & SPLS1000 User Manual

I. PANEL DESCRIPTION... 1

TI-500 Plus DIGITAL WEIGHT INDICATOR Setup / Operation Manual

WM-25 USER S MANUAL. Manufacturer s Declaration of Conformity

Transcription:

CONTENTS CAUTIONS... 4 INTRODUCTION... 6 FEATURES & MAIN FUNCTIONS... 6 SPECCIFICATION... 7 PART NAME & KEY FUNCTION... 8 SYSTEM MODE...11 GENERAL FUNCTION & DESCRIPTION... 16 CONVERSION MODE... 18 PRINT FORMAT... 20 RS-232C COMMUNICATION... 21 INSTALLING & CONNECTION... 22 USING & BATTERY... 24 TEST MODE... 26 CALIBRATION MODE... 30 GRAVITY COMPENSATION... 33 SEALING & OPTIONAL DRAWING... 35 ERROR MESSAGE DESCRIPTION & MANAGMENT... 36

CAUTIONS

INTRODUCTION We greatly appreciate for you to purchase BI-SERIES of CAS. This goods has the excellent performance and splendid properties through strict test under severe quality management. It is recommended to read this manual in full before using BI-SERIES for good function application. FEATURES Simple outdoor using in use of battery Using general (Mn, Alkaline) battery or rechargeable battery (NiCd) Easy operation and various function SPAC single pass automatic span calibration Water proof (IP65) High precise resolution Liquid crystal display of 6 place weight (LCD) Using long-time battery MAIN FUNCTIONS Hold Function COUNT Function (Unit weight setting, unit weight setting by sample) Low / High Weight Comparing Serial Communication Interface (RS-232) Serial Printer Connection Self Hardware Test Independent Zero Calibration Gravity Compensation

SPECIFICATION Load Cell Excitation Voltage DC 5 V (350Ω x 4EA) Temperature Range -10 ~ +40 Zero Adjustment Range Re-zero Range Initial Zero Range Input Sensitivity 0.05mV ~ 5mV ±2 of Max. Capacity ±10 of Max. Capacity 2 / D A/D Internal Resolution 1 / 100,000 A/D External Resolution Display Below Zero Tare Subtraction Weight Display Type Character Display 1 / 10,000(Non OIML), 1/5,000(OIML) - Minus signal Full capacity LCD (6 digit + Sign, 24mm Height) Stable ( ), Zero ( ), Tare ( ), Net ( ),Hold ( ), Low battery ( ), Weight unit (kg), High, OK, Low ( ),Count( ) Power Consumption Approx. 0.25W Product Size Product Weigh Serial Communication RTC 220(W) x 205(D) x 88(H)mm about 1.8Kg RS-232C Real Time Clock Option BATTERY 1. DC 7.2V(6EA SIZE " AA " ) 2500mAh Rechargeable battery 2. DC 9V(6EA SIZE " C " ) Dry Battery 3. DC 12V 1.25A AC Adaptor

PART NAME & KEY FUNCTION 1. Part Name 2. DISPLAY

LAMP SYMBOL DESCRIPTION STABLE ON when current weight is stable ZERO ON when current weight is 0 NET ON at displaying the pure weight HOLD TARE ON at weighing the moving object ON when tare is activated. HI OK LOW ON when current weight is more than high limit value. ON when current weight is right weight band. ON when current weight is less than low limit value. BAT Display the battery recharging time (rechargeable battery) Display the battery replacing time (dry battery) - about 6.9V remain PCS ON when current mode is count mode. CHARGE When charging battery, charging lamp (RED) is ON. When it is completed, charging lamp (GREEN) is ON. POWER ON at connecting the DC 12V adapter. Minus ON when weight is - Kg Weight unit

3. KEY Part KEY SYMBOL DESCRIPTION ON / OFF Use to power ON / OFF (Press for approx 3 seconds) ZERO / MODE Use in case of rezero (within 2% of max. capacity), To enter the system mode by pressing for 2 seconds TARE * Use at weighing in use of TARE Enter the TIME SETTING MODE by pressing for 2 seconds Use to turn ON / OFF Back Light (Option) or to print out in the case that F-2 is set to 4. Enter the CONVERSION MODE by pressing for 2 seconds. Hi/Low Comparing Mode key or Weighing Mode key NET / GROSS HOLD Use to display NET weight or GROSS weight. Enter the TEST MODE by pressing for 2 seconds Use to weight the moving object (Manual, Automatic) Counting Mode Key U.W. Use to display the unit weight (for approx. 5 seconds) W. Use to display the current weight (for approx. 5 seconds)

SYSTEM MODE Simple weight mode Display the weight after weighing. High / Low comparing mode Determine that the weight is high, normal or low value and perform the display. Count mode Display the quantity by setting the unit or sample weight. Quantity comparing mode Perform the display after discrimination by setting High / Low quantity. 1. MODE CONVERSION Press Zero-Mode key for 2~3 sec. to display the following message. Execute depending on the below mode conversion and setting method. 2. DIGIT INPUT METHOD Input the digit in use of the below keys. STAR key : Use to set the input value to G/N key : Use to increase the first place of value by on. TARE key : Use to move the current place to the left by one place. ENTER key : Use to move the current place to the right by one place. ZERO/MODE key : Use to complete the input.

Ex) For 20.5 input Display KEY DESCRIPTION STEP 1 Press once (Set to 0 ) STEP 2 STEP 3 STEP 4 STEP 5 STEP 6 Press five times. Press once (Move to the left by one place) Press once (Decimal point input) Press twice times (input 2 as the 3th place value) Input completion 3. SIMPLE WEIGHING MODE CONVERSION Display KEY PLATFORM DESCRIPTION STEP 1 STEP 2 Empty Press once Press once. Generally operated 4. HIGH / LOW COMPARING MODE CONVERSION In case of no resetting the High / Low value Display KEY PLATFORM DESCRIPTION STEP 1 STEP 2 Empty Press twice Press once. Generally operated

In case of resetting the High / Low value Display KEY PLATFORM DESCRIPTION STEP 1 STEP 1 Press twice Press once (Low weight input) STEP 1 INPUT Refer to 'Digit input' In page 11 STEP 1 High value input STEP 1 INPUT Refer to 'Digit input' In page 11 STEP 1 Empty Generally operated 5. COUNT MODE CONVERSION In case of no setting the unit / sample weight Display KEY PLATFORM DESCRIPTION STEP 1 STEP 2 Empty Press three times Press once Generally operated In case of setting the unit weight (Min. range of unit weight ; Min. Division) Display KEY PLATFORM DESCRIPTION STEP 1 STEP 2 STEP 3 Press three times Press once Display the set weight STEP 3 STEP 5 Input the unit weight Digit input in page 11 Press 'Mode' key to round & display unit weight. Generally operated Step 5. Ex) Min. division is 0.1 Input the unit weight as 0.532 and round off the second place of decimal to 0.5

In case of setting the unit weight by sample (Samples : 10-200, 10 unit) Display KEY PLATFORM DESCRIPTION STEP 1 Press three times STEP 2 Press once STEP 3 Press once STEP 4 STEP 5 Empty Press once Press five times. Simple count input(50pcs) STEP 6 Empty Press once [ENTER] STEP 7 STEP 8 STEP 9 Sample Press once with empty [ENTER] key Unit weight display. Press once STEP 10 Empty Generally operated Note : For setting the unit weight by sample count, the min. unit weight shall be more than min. division

6. COUNT COMPARING MODE CONVERSION In case of no setting the High / Low quantity Display KEY PLATFORM DESCRIPTION STEP 1 STEP 2 Press four times Press once. Generally operated In case of no setting the High / Low quantity Display KEY PLATFORM DESCRIPTION STEP 1 STEP 2 STEP 3 STEP 4 STEP 5 STEP 6 Press four times Press once. Low value setting Low value input Digit input in page 11 Press once High value input High value input Digit input in page 11 Press once. Generally operated

GENERAL FUNCTION & DESCRIPTION 1. REZERO FUNCTIOIN (Use at change zero point) Rezero band range : Within 2% of Maximum capacity Zero point is changed. Press 'Zero' key. Zero lamp is ON. 2. TARE FUNCTIOIN (Use at weighing in use of TARE) Maximum TARE range : Maximum capacity Put the TARE on platform0 (TARE weight : 10kg) Press 'TARE' key. (The TARE weight is saved) Put the object on platform (object weight : 20kg) Note : The weight including TARE weight can't exceed the maximum capacity In case to want know the Gross weight Press 'NET/GROSS' key (object + TARE weight display) In case to want know the NET weight Press 'NET/GROSS' key (object weight display) Remove the TARE and object form platform to display the saved TARE weight In case to remove the Tare weight Remove the TARE and object from load cell(up figure) and press TARE key(weight is 0)

3. HOLD FUNCTIOIN (Use at weighing the moving object) Auto hold function (Hold function operated whenever weighing) Empty the platform Press 'Hold' key 'Auto hold' message is displayed for 1 sec. Hold lamp is ON When the weight change is stable after putting the Object(20kg) on platform, 'Hold' message is displayed The weight is averaged for approx. 3~5 sec. The hold weight is displayed. Hold start is more than 5.5 division and 10% of maximum capacity after displayed hold weight. To release the hold weight Empty the platform or press 'Hold' key and return to the normal operation Auto hold function release Empty the platform Press 'Hold' key Auto hold release message is shown for 1 sec. And then normally operated. 'Hold' lamp is OFF. Manual hold function (Executed at pressing 'Hold' key) Put the object (10kg) on platform Press 'Hold' key. 'Hold' message is displayed for 1 sec. The weight is averaged for displayed. 3-5 sec. The hold weight is displayed. Empty the platform or press 'Hold' key to release the hold weight and perform the normal operation. 'Hold' lamp is OFF If change weight is more than 10% of maximum capacity after displayed hold weight, hold function is cancelled

CONVERSION MODE 1. ENTER method If you press the * key for three seconds to normal operation state, after display "SET" Message for one second and then enter. 2. KEY using Move to the next menu. Increase the set point. Return to the normal operation at completion of setting :[ENTER] key. 3. CONVERSION menu (F01 - F06) F - 1 : Auto power OFF setting. F - 2 : RS-232C communication setting. F - 3 : Select print format type. F - 4 : Stable condition F - 5 : Display Time Control F - 6 : Hold Time Control F - 1 : Auto power OFF setting Auto power OFF function is used to spare the battery by means of power off when this scale is not used. Set point F 1 0 Auto power OFF is not used F 1 1~9 Meaning Automatically power off when the weight is not change or no key is used for 1 ~ 9 minutes

F - 2 : RS-232C communication setting Set point Meaning F 2 0 No transmit F 2 1 - F 2 2 Transmit at stable weight F 2 3 Transmit at data request. (Data request signal D ) F 2 4 Use for print mode F - 3 : Select print format type Set point Meaning F 3 0 Select print format 1 F 3 1 Select print format 2 F 3 2 Select print format 1 + Current Time F 3 3 Select print format 2 + Current Time F - 4 : Stable Condition Set point F 4 1~9 Stable condition Level 1~9 Meaning F - 5 : Display Time Control Set point Meaning F 5 1~9 Display Time Control (1 : Past, 9 : Slow) F - 6 : Hold Time Control Set point Meaning F 6 1~7 Hold Time Control (1 : Past, 7 : Slow)

Print Format Weight Mode - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - < Format 1 > < Format 2 > No. 001 : 1.00 kg 001 : Net 1.00 kg Tare 0.00 kg Gross 1.00 kg < Format 1 + Time > < Format 2 + Time > 2005 03. 17 11:23:46 2005 03. 17 11:23:46 No. 001 : 1.00 kg 001 : Net 1.00 kg Printed weight is net weight. Tare 0.00 kg Gross 1.00 kg PCS Mode - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - < Format 1 > < Format 2 > No. 001 : 1.00 pcs 001 : Net 1.00 pcs Tare 0.00 pcs Gross 1.00 pcs < Format 1 + Time > < Format 2 + Time > 2005 03. 17 11:23:46. 2005 03. 17 11:23:46 No. 001 : 1.00 pcs 001 : Net 1.00 pcs Printed count is net count. Tare 0.00 pcs Gross 1.00 pcs

RS-232C communication 1. Port connection method Connect the computer serial port with RS-232C port in the back panel of indicator as follows to communicate with computer (2)TXD (1)RXD (5)GND CN 1 of Indicator ( male ) - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3 Receive Data 2 Transmit Data 7 Signal Ground 4 Request to Send 5 Clear to Send 6 Data Set Ready 8 Carrier Detect 20 Data Terminal Ready Computer serial 25 pin port(female) (2)TXD (1)RXD (5)GND CN 1 of Indicator ( male ) - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 2 Receive Data 3 Transmit Data 5 Signal Ground 1 Carrier Detect 4 Data Terminal Ready 6 Data Set Ready 7 Request to Send 8 Clear to Send Computer serial 9 pin port (female) 2. RS-232C Output data format Baud rate : 9600bps Data bit : 8, Stop bit : 1, Parity bit : None Code : ASCII

3. Transmit data format Total 22 Bytes,,, Data(8 bytes) K G CR LF SOH Lamp em US (Unstable) GS(Gross W.) Status Pty ST (Stable) NT(Net W.) Byte Weight OL (Overload) Unit Lamp status byte : Display the current state (ON/OFF)of indicator lamp (Low Enable) Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0 1 STABLE HIGH LOW HOLD GROSS TARE ZERO Data (8 bytes) Weight data including a decimal point - and. That is, for -13.5kg: Transmit ASCII code 8 bytes Corresponding to -,,,, 1, 3, 5, Data request mode If F_2 is set to 3, the indicator outputs the specified data when ASCII code corresponding to D is sent to it. INSTALLATION & CONNECTION 1. Load cell connection - Load cell connector with the load cell port of indicator back panel. Load cell output to Resolution and Pin specification 5V impression to load cell Max. load cell output Recommended resolution 2 mv 1/2,000 (Max.) 5 mv 1/5,000 (Max.) 10 mv 1/10,000 (Max.)

Connecting method Wire color can be different depending on the load cell's manufacturer or it's model. 2. Load cell color table PIN COLOR PIN COLOR 1 EXC RED 5 SHIELD SHIELD 2 EXC WHITE 6 SEN BROWN 3 SIG GREEN 7 SEN BLACK 4 SIG BLUE If 1(EXC+) connect with 6(SEN+) and 2(EXC-) connect with 7(SEN-), this is 4-wire system.

USING Replace the general(mn, Alkaline) battery or charge the rechargeable battery in full at using after long-time storage Power on by pressing ON/OFF key in the front panel BATTERY 1. Battery replacement or charging time Power is off in a specific time after lamp is on in the right top of display part. The battery lamp is on when the amount of battery is about 6.9 V and power is off When the amount of battery is about 6.5V automatically. When the lamp is on, replace the dry or charge the rechargeable battery. 2. General (Mn, Alkaline) battery using and replacing Open the top cover and make as follows in the left bottom of cover at using the general (Mn, Alkaline) battery. Insert the battery into the battery case after checking the battery polarity.

Warning Open the cover and check the battery type. If the battery is not the rechargeable battery (this is : Mn, Alkaline), Never connect the AC adaptor jack in the back panel 3. Rechargeable battery using and recharging Open the top cover and make as follows in the left bottom of cover at using the rechargeable battery. Open the jack protecting cap in the back panel and connect the adapter as follows. When the adapter is connected with indicator, RED(POWER) and RED(CHARGE) lamp are ON. When the charging is completed, CHARGE lamp is changed to GREEN. It takes approximately 4-5 hours to charge in full.

4. Battery using period Conditions BACK LIGHT(E.L) is not used BACK LIGHT(E.L) is always used Using Period Recharging Battery : Approx. 70 hours Mn Battery : Approx. 100 hours Alkaline Battery : Approx. 250 hours Recharging Battery : Approx. 30 hours Mn Battery : Approx. 45 hours Alkaline Battery : Approx. 90 hours Battery using time can change according to battery using period or loadcell number. TEST MODE How to enter test mode 1. Open the top cover of indicator and power on with pressing CAL switch in the left top. At that time, LCD screen shows ZEro = test, tare = CAL message and pressing Zero key moves to test 1. 2. Press G/N key for approx 3 seconds. Test menu (test 1~7) TEST 1 : Liquid Crystal Display test TEST 2 : Load cell and A/D conversion test TEST 3 : Key test TEST 4 : RS-232C communication test TEST 5 : Temperature test TEST 6 : Battery test TEST 7 : RTC test TEST 1 (Enter ' TEST ' mode to start TEST 1) FUNCTION : LCD TEST KEY LCD SCREEN DESCRIPTION [ ENTER Key] Press once : Test running One more time : Next menu [ * Key] : Back light test Display TEST 1 state

TEST 2 FUNCTION : A/D Conversion TEST [ ENTER Key] Press once : Test running One more time : Next menu KEY LCD SCREEN DESCRIPTION Display TEST 2 state Display the digital value Corresponding to current weight TEST 3 FUNCTION : Key test [ ENTER Key] Press once : Test running One more time : Next menu KEY LCD SCREEN DESCRIPTION Display TEST 3 state Pressing the key to test shows the key code. Key name code Key name code. 1 ZERO 2 TARE 3 GROSS /NET 4 TEST 4 FUNCTION : RS-232C communication test (Computer link test) [ ENTER Key] Press once : Test running One more time : Next menu KEY LCD SCREEN DESCRIPTION Display TEST 4 state Receive : " 1 " Transmit : " 2 " Reference 1. Connect the computer COM port with RS-232C connector of indicator before running this test. (Possible only at installing RS-232C option) Reference 2. Press " 1 " in computer keyboard and check that " 1 " is shown in LCD screen of indicator and Check that the next ASCII code value, " 2 " is inputted in the computer.

TEST 5 FUNCTION : Temperature test KEY LCD SCREEN DESCRIPTION [ ENTER Key] Press once : Test running One more time : Next menu Display TEST 5 state for temperature Current temperature is shown TEST 6 FUNCTION : Battery test KEY LCD SCREEN DESCRIPTION [ ENTER Key] Press once : Test running Display TEST 6 state for battery voltage value Current battery voltage is shown. TEST 7 FUNCTION : RTC test KEY LCD SCREEN DESCRIPTION [ ENTER Key] Press once : Can see the current date & time Press twice : Exit Test Mode [ * Key ] Can see date & time Alternately Display TEST 7 state for current date & time

Current date & time setting method In the normal mode, Press the TARE Key long and can see the message and. After this step, You can input the current date and time as follows How to set the current date and time KEY LCD SCREEN DESCRIPTION [TARE] Key : Shift one digit to the left [STAR] Key : Shift one digit to the right [ZERO/MODE] Key : Reset the input value [ N / G] Key : Increase input value Date & time setting mode Initial value : increase 5 [Year : 2005] : increase 2 [Month : 02] : increase 1 : increase 7 [Day : 17] : increase 1 : increase 8 [Hour : 18] : increase 1 [ENTER] Key : Exit this input mode & : increase 3 [Minute] Go to normal mode 2005/02/17 18:13 If you input wrong date, this scale display "out" message and doesn't save the date. Load Cell test method Power off before testing the load cell output voltage and check that the load cell output voltage (SIG+, SIG-) is same as the specified output voltage with ON/OFF key in the front panel pressed.

. CALIBRATION MODE How to enter calibration mode Open the top cover of indicator and power on with pressing CAL switch in the left top. At that time, LCD screen shows " ZEro = test, tare = CAL " message and pressing ' Tare ' key moves into CAL 1. Calibration menu (CAL1~CAL5) CAL 1 : Maximum Capacity Set CAL 3 : Span Weight Setting CAL 5 : Span Calibration CAL 2 : Minimum Division Set CAL 4 : Zero Calibration Digit input method Increase the first place value by 1. Move to the left by 1 place. Set the inputted value to " 0 " Move to the next menu. Ex) 150 Input Key : Press once (LCD screen : 0) Key : Press once (LCD screen : 1) Key : Press once (LCD screen : 10) Key : Press fifth (LCD screen : 15) Key : Press once (LCD screen : 150) CAL 1 FUNCTION : Maximum Capacity Setting (Range : 1kg ~ 999,999kg) KEY LCD SCREEN DESCRIPTION Input by digit input method ENTER key : Move to the next menu 50kg 2000kg

CAL 2 FUNCTION : Minimum Division Setting (Range : 0.001kg ~ 50kg) KEY LCD SCREEN DESCRIPTION Choose 0.001kg or 50kg using Tare or C/N key (Reference the table below) ENTER key : Move to the next menu 0.001kg 0.5kg Min. division Min. division Min. division Min. division Min. division 1 0.001 kg 4 0.01 kg 7 0.1 kg 10 1 kg 13 10 kg 2 0.002 kg 5 0.02 kg 8 0.2 kg 11 2 kg 14 20 kg 3 0.005 kg 6 0.05 kg 9 0.5 kg 12 5 kg 15 50 kg CAL 3 FUNCTION : Span Weight Setting (Range : 1kg ~ 999,999kg) KEY LCD SCREEN DESCRIPTION Input by digit input method ENTER key : Move to the next menu Set the span weight to 50 kg Set the span weight to 2000Kg Reference 1. The span weight should be within the 10 % to 100 % of maximum capacity. The initial span weight is 100% of max. capacity but if the current span weight is different from it, input the desired weight again. Reference 2. The span weight should be above 10% of the maximum capacity. If not so, an error message (Err 23) is shown Reference 3. Never install the span weight higher than maximum capacity because an " Err 22 " is shown.

CAL 4 FUNCTION : Zero Calibration KEY LCD SCREEN DESCRIPTION ENTER key : Move to the next menu after zero calibration Empty the platform and when LCD screen shows " UnLoAd ", press ENTER ' key. Then move into span calibration after counting up to 4. Reference 1. Zero calibration is performed automatically by program. Load cell output is automatically adjusted within 0.05mV and 5mV CAL 5 FUNCTION : Span Calibration KEY LCD SCREEN DESCRIPTION ENTER key : Move into the Initial menu after span calibration Put the weight set in CAL 3 on platform and press ' ENTER ' key Span calibration is terminated Error occurrence. If not error, Move into the initial menu after counting up to 4. Reference 1. Display " End " message and move into the initial menu at completion of span calibration. Reference 2. At occurrence of Err 24, reset by lowering the resolution.

GRAVITY COMPENSATION How to enter gravity compensation Open the top cover of indicator and power on with pressing CAL switch in the left top. LCD screen shows " ZEro = test, tare = CAL " message and pressing ' * ' key moves into gr A-1 1. gra - 1 FUNCTION : Gravity Compensation KEY LCD SCREEN DESCRIPTION : Shift one digit to the left : Shift one digit to the right : Increase one value Gravity of calibration place Set the value Use to complete the input Go to Next gra - 2 1. gra - 2 FUNCTION : Gravity Compensation KEY LCD SCREEN DESCRIPTION : Shift one digit to the left : Shift one digit to the right : Increase one value Use to complete the input Go to CAL Mode Move to normal mode by pressed ENTER Key.

Amsterdam 9.813 m/s² Havana 9.788 m/s² Paris 9.809 m/s² Athens 9.800 m/s² Helsinki 9.819 m/s² Rio de Janiero 9.788 m/s² Auckland NZ 9.799 m/s² Kuwait 9.793 m/s² Rome 9.803 m/s² Bangkok 9.783 m/s² Lisbon 9.801 m/s² San Francisco 9.800 m/s² Birmingham 9.813 m/s² London 9.812 m/s² Singapore 9.781 m/s² Brussels 9.811 m/s² Los Angeles 9.796 m/s² Stockholm 9.818 m/s² Buenos Aires 9.797 m/s² Madrid 9.800 m/s² Sydney 9.797 m/s² Calcutta 9.788 m/s² Manila 9.784 m/s² Taiwan 9.788 m/s² Chicago 9.803 m/s² Melbourne 9.800 m/s² Taipei 9.790 m/s² Copenhagen 9.815 m/s² Mexico City 9.779 m/s² Tokyo 9.798 m/s² Cyprus 9.797 m/s² Milan 9.806 m/s² Vancouver BC 9.809 m/s² Djakarta 9.781 m/s² New York 9.802 m/s² Washington DC 9.801 m/s² Frankfurt 9.810 m/s² Oslo 9.819 m/s² Wellington 9.803 m/s² Glasgow 9.816 m/s² Ottawa 9.806 m/s² Zurich 9.807 m/s²

SEALING & OPTIONAL DRAWING Perform the sealing as shown in the below figure after the calibration is completed. Open the top cover. Cover CAL switch in the left top of cover with sealing plate and screw2 sealing bolt. Connect the sealing wire as follows and compress the sealing lead. OPTIONAL DRAWING

ERROR MESSAGE DESCRIPTION & MANAGEMENT Errors in Weighing Mode Err 01 Cause The Indicator can't be initialized due to platform shaking. Management Put the platform on a flat place without shaking and power on. Err 02 Cause The load cell is connected wrongly or A/D conversion part has a fault. Management Check that the load cell is well connected with main unit. Err 03 Cause The calibration isn't executed or the data of internal memory is lost due to a electrical shock. Management Perform the required specification again in setting mode. Err 09 (Zero Key) Cause The current weight exceeds zero range. Err 10 Cause The specified TARE weight exceeds the maximum capacity. Management Set the TARE weight lower than maximum capacity. Or set the maximum capacity higher than the desired TARE weight in calibration menu and perform the calibration again.

Err 13 (Zero Range with power on) Cause The initial zero range exceeds above 10% of maximum capacity. Management Check the load cell state. Over Cause Too heavy object is put on platform so that it may exceed the maximum capacity. Management Never put the weight exceeding the maximum capacity. If the load cell is damaged, replace the load cell. Errors in Weighing Mode Err 22 Cause The span weight is over maximum capacity. Management Set the span weight less than maximum capacity. Err 23 Cause Setting weight is set under 10 % of the Maximum Capacity. Management Set span setting weight equal to or over 10 % of the Maximum Capacity in CAL 3 of CAL Menu. Err 24 Cause The resolution is set too high in comparison with load cell output. Management Set the minimum division in CAL 2 high.

Err 25 Cause The resolution is more than 1/10,000. Management Set the minimum division in CAL 2 high Errors in Count Mode Err 61 Cause The count unit weight and sample weight are set too high. Management Set the count unit weight low or the count sample quantity high. Err 62 Cause The count unit weight and sample weight are set too low. Management Set the count unit weight high or the count sample quantity low.