Operation Manual. Concorde 600 Power Supply. *This instrument is intended for laboratory use only.

Similar documents
EPS Power Supply

PowerPro Power Supplies

User's Guide. MiniTec TM Series Model MN25 MultiMeter

PowerPac HC Power Supply. Instruction Manual

Mini Digital Multimeter

250 Volt Electrophoresis Power Supply

BS 181 SINGLE CHANNEL POWER SUPPLY USER MANUAL

This 4200-RM Rack Mount Kit is for installation in 4200-CAB series cabinets only.

Model P4470 Save A Watt Operation Manual

Instruction Manual Standard Multi-Position Stirrers Advanced Multi-Position Stirrers

1-36V, 0-3A DC Power Supply

2200 Series. Quick Start Guide SAFETY CHECKLIST. Safety Made Simple

Mini Digital Multimeter

USER MANUAL. Mini Multimeter with Non-Contact Voltage Detector (NCV) Model EX310

Owner's Manual. True RMS Multimeter. Model No Safety Operation Maintenance Español

User Guide True RMS Multimeter Extech EX205T

AutoRanging Digital MultiMeter

Digital Overhead Stirrer

Model 7705 Control Module

P4472 Save A Watt HD Operation Manual

Model 2460-KIT. Screw Terminal Connector Kit. Description / September 2014 *P * 1

Cantata m100 Amplifier

User s Guide. 600A True RMS AC/DC Clamp Meter. Model 38389

PREMIUMAUDIOVIDEOLIGHTINGANDPOWERPRODUCTS

Mini Digital Multimeter

Model 2380 Rack-Mount Kit

User's Guide. Digital Multimeter. Model MN42

INSTRUCTION MANUAL. Instruction Manual. Analog Multi-Tube Vortexer Digital Multi-Tube Vortexer

Autoranging Multimeter plus IR Thermometer Extech 450 Patented

HV-CS kv Edge Mount Triaxial Jack

impact INSTRUCTIONS VS-LCD Series Flash lighting equipment and accessories

Mini Digital Multimeter Model MN15. User's Guide

MyGenie TM 96/384 Thermal Block

Operating Instructions

Operating Instructions

tracer Operating Manual Non-Contact AC Voltage Detector U.S. Patent No. 9,091,707

PORTABLE WIRELESS PA SYSTEM WITH LITHIUM -ION RECHARGEABLE BATTERY AWP6042 OWNER S MANUAL. Handheld Microphone Headset Transmitter

Model 8020-KHV. Kelvin Keithley Triaxial Connector Card. Description / October 2014 *P * 1

MyGenie TM 32 Thermal Block

Autoranging Mini Multimeter

700 Series 200 Amp Clamp Meters

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER

USERS GUIDE MCX-VTH. VGA to HDMI Converter. Manual Number:

MODEL 805 USER MANUAL

MultiTherm Shaker. Operations Manual Item: H5000-H and H5000-HC

CVU-200-KIT. 200 V Bias Tee Kit. Description. Parts list / October 2014 *P A* 1

TH2683 Insulation Resistance Meter. User s Mannual

OWNER S MANUAL CD-2 V 1.3

USERS GUIDE MCX-STH. 3G SDI to HDMI Converter. Manual Number:

Models 2601B, 2602B, and 2604B System SourceMeter Instruments Quick Start Guide

WEB SITE:

Economy Single Channel Output DC Power Supply

Automated Tuner System Power Distribution Hub

Model 8020-STC. Kelvin Standard Triaxial Connector Card. Description / October 2014 *P * 1

User s Guide. 600A AC Clamp Meter. Model 38387

Model A Mini AC/DC Clamp Meter. User's Guide

ZEPHYR Compact Haze Machine

CM-220 True RMS AC CLAMP METER INSTRUCTION MANUAL

User's Guide. Extech AM A AC Analog Clamp Meter

Model 2380 Rack-Mount Kit

User's Guide. True RMS Multimeter plus IR Thermometer. Extech 470 Patent Pending

BS 181 SINGLE CHANNEL POWER SUPPLY USER MANUAL

A120 A W/240W PA Mixer-Amplifier 70V/100V

Z-350 FAZER USER MANUAL

Model HM-535 Power Supply Installation and Service Instructions

PowerPac Universal Power Supply. Instruction Manual

BS 287 DUAL CHANNEL POWER SUPPLY. User Manual. January 2017 V1.0

TABLE OF CONTENTS 1. SAFETY PRECAUTIONS GENERAL DESCRIPTION FEATURES AND FUNCTIONS 4. CONNECTIONS 6. APPLICATIONS...

Autoranging Mini Multimeter Model MN16A

Instruction Manual. Rocking Platform Shaker. 3-D Rotator Waver

1 Channel Strobe Controller ORDERCODE 40226

SafeBlue Electrophoresis System. Instruction manual

User Manual. 400Amp AC Clamp Meter + NCV. Model MA430. Additional User Manual Translations available at

USER MANUAL. Uninterruptible Power Supply Line-interactive VCL Series UPS VA. GE Critical Power

FT-20 Fire Training Fog Machine User Manual English

Autoranging True RMS Multimeter User Manual

OPERATING INSTRUCTION

OPR Power Series AC to DC POWER SUPPLY SERIES WITH REMOTE MANAGEMENT AND ALARM SYSTEMS. Model Nos. OPR065-48S / OPR065-48R. Manual

Haze 800 DMX. User Guide. English

Installation and Operation Back-UPS BR1000G-IN / BR1500G-IN

S-14 S-14. Compact Digital Multimeter. Compact Digital Multimeter

Model R5010. TRMS Digital Multimeter. Instruction Manual

Ares. Modular Audiophile System. by Thrax Audio. Operating Manual. Manual issued 15/03/2018 CAUTION

CVU-3K-KIT. 3 kv Bias Tee Kit. Description. Parts list / October 2014 *P * 1

INSTRUCTION MANUAL. Model True RMS AC/DC 30A Mini Clamp-on Meter. Introduction. True RMS AC Current and Voltage

USERS GUIDE ASP-18H-4K. HDMI 4K2K 1x8 Splitter with EDID Management. Manual Number:

Model R5010. Instruction Manual. TRMS Digital Multimeter. reedinstruments www.

Platform Rocker PMR-100

MS-SL12-RD. 17-Outlet LCD Display Power Strip 20Amp INSTRUCTION MANUAL

SHUTTERBOSS VERSION II USER MANUAL

PS 289 DUAL CHANNEL POWER SUPPLY USER MANUAL

PREMIUMAUDIOVIDEOANDPOWERPRODUCTS V-RVC-PRO. Owners Manual

Antenna Distributor ORDERCODE D1444

OPERATING INSTRUCTIONS 7 SERIES STATIC GENERATORS

Powerware 3105 UPS User s manual

Powerware 3105 UPS User s manual

Model 4001A Modular System Bin Operating and Service Manual

USER'S MANUAL DCL-650

MX-T6-PRO/S Tube Rollers

Toll Free: Tel: Fax:

Transcription:

Concorde 600 Power Supply Operation Manual Cat.no. R10-1001011 *This instrument is intended for laboratory use only http://www.recenttec.com E-mail : support@recenttec.com Version 1.1

Packing List x 1 main body x 1 power cable x 1 manual x 1 warranty card 7F, 72 Song-Te Road, Taipei 110, Taiwan, R.O.C. Tel: +886-2-2723-0886 Fax: +886-2-2723-0876 http://www.recenttec.com E-mail: support@recenttec.com

Index A. Important Notice ----------------------------------------------------------------------------------------- 1 1. Warranty 2. Technical and Service Contact 3. Safety Notice B. Introduction ----------------------------------------------------------------------------------------------- 4 1. Specifications 2. Product Descriptions C. Installations ----------------------------------------------------------------------------------------------- 8 1. Package List 2. Setup and Operation of the Concorde 600 Power Supply 2-1. Unpacking 2-2. Setup and operation D. Troubleshooting --------------------------------------------------------------------------------------- 8 1. Basic troubleshooting 2. Error codes 3. Replacing the fuses E. Care and Maintenance ----------------------------------------------------------------------------- 10 F. Order Information -------------------------------------------------------------------------------------- 11

A. Important Notice Before setting up and operating the instruments, please carefully read these instructions to get familiarized with the installation and operation process. Instructions should be read by experienced individuals before operating the instruments. Any improper usage of the instruments may cause damage. Please refer to the safety notice included with this equipment. The instruments shall not be modified or altered in any way. Any modification or alteration will void the warranty, void the regulatory certifications and create potential safety hazard. Recenttec is not responsible for any injury or damage caused by using the instruments for any non-intended purpose or injury as a result of modification of the instruments by any person who is not authorized by Recenttec Inc. 1. Warranty CONCORDE 600 is warranted to be free from defects in materials or workmanship for a period of one year from the original invoice date, under normal usage. Any defects occurring during warranty period, Recenttec Inc. will repair or replace defective products or parts without charge unless the defects arise from conditions outlined below. The defects described below are specially excluded from Recenttec warranty policy. a. Improper operation of the instrument.

b. Repair or modification by any person who is not authorized by Recenttec Inc. c. Damage caused by any (in)-direct accident, neglect or misuse. d. Damage caused by disaster. e. Damage caused by any improper solvents or samples 2. Technical and Service Contact Most of the operation details are described in this instruction manual to assist and guide operator for an appropriate solution. For any other technical/ service questions, please contact your local representative or contact Recenttec international technical/ service specialist by E-mail: support@recenttec.com. 3. Safety Notice Concorde 600 power supply use high voltage output. To minimize the risk of electrical shock to the 2

user the following guidelines should be observed and followed when using Concorde 600 power supply. Always read the product manual before operating the unit. Concorde 600 power supply shall be operated at temperature between 0 and 40, with relative humidity between 10% and 90% non-condensing. Do not operate the Concorde 600 power supply in extreme humidity or where condensation can short the internal electrical circuits. Be sure that there is at least 6 cm clearance around the Concorde 600 power supply to ensure adequate cooling. Always connect the power supply using the AC power cord provided with the power supply or a similar 3-prong, grounded AC outlet. Use other than Recenttec s electrical cable with plugs or banana jacks is done at the user s own risk when connecting the external instrument to the power supply s high voltage output jacks. While doing so, always grasp the plug by the molded support at the rear of the plug. Do not grasp the individual prong ends. When taking the power supply into a cold room, the unit can be operated immediately. However, when removing the power supply from the cold room, let the unit equilibrate to room temperature for a minimum of two hours before using it. Never connect a high voltage output lead to earth ground. This defeats the floating electrical isolation of the power supply and exposes the user to potentially lethal high voltages. 3

B. Introduction CONCORDE 600 power supply is compact, space saving and light weight in design. It can be used with many electrophoresis applications, such as running standard PAGE gels, multiple PAGE mini gels, horizontal/vertical agarose gels, semidry transfer and mini-gel tank transfer. The LED display reports power supply status. Parameters are set by pressing UP/DOWN arrow buttons in combination with a MODE button. Four pairs of output jacks accept 4 mm plugs, which operate in parallel. The output jacks are color coded, recessed for safety, and accept both 4 mm unshielded and shielded (banana) plugs. The user interface consists of a three digit, LED readout, and five buttons on a membrane keypad. Diagrams of the front and rear panel are shown on the following pages. The CONCORDE 600 operates in constant voltage or constant current modes. Power output is controlled by setting a maximum value for voltage (up to 600 V), or current (up to 500 ma). The maximum output power at any time is 90 Watts. The Run duration can be continuous or programmed to a maximum duration of 999 minutes. The CONCORDE 600 automatically crosses over or switches the limiting parameter according to the programmed limits as the resistance of the electrical load changes during a run. 4

1. Specifications Type CONCORDE 600 Input 100 ~ 240 V AC, 50-60 Hz Output voltage (V) 5 600, 1 V steps Output current (ma) 5 500, 1 ma steps Maximum output power (W) 90 Output jacks 4 pairs Output connector 4-mm plugs Timer (Minutes) 1 999, 1 min increment User interface 3 digit, red LED membrane keypad Dimension (L x W x H) 320 x 240 x 73 mm Weight Net ~2.4 kg (5.3 lb) Gross ~3.4 kg (7.5 lb) Safety features No load detection, Sudden load change protection, Over voltage protection, Automatic power failure recovery Indoor use: 0 40 Operating conditions Relative humidity: 0% to 90% Non-condensing Altitude: < 2000 m Installation category: II 5

2. Product Description: CONCORDE 600 Power Supply Hardware Overview 6

Button Description 1 ON/OFF ON/OFF Power Supply 2 CONST Select voltage or current to be constant 3 MODE Select the volts, milli amps, or time to be displayed 4 Changes the displayed value of the selected parameter 5 RUN Starts and Stops the output from the Power Supply The front and rear panel buttons as shown above are used to set up and run the CONCORDE 600 power supply. The audible alarm design allows a beep sound for each and every button pressed on the front membrane keypad. A long beep alarm allows user to identify the run is complete or if an error occurs. It will continue every ten seconds until any button is pressed or the main power is turned off. 7

C. Installation of CONCORDE 600 Power Supply 1. Package List Item Quantity CONCORDE 600 power supply 1 Power cable 1 Instruction package 1 2. Setup and Operation of the CONCORDE 600 Power Supply 2-1. Unpacking Unwrap all packages carefully and compare contents with the packing list, ensure the arrival of all the items. If any part is missing, contact your local Recenttec representative. Inspect all the components for damage that may have occurred while the unit was in transit. If any part appears damaged, contact the carrier immediately. Be sure to keep all packing material for damage claims or to use should it become necessary to return the unit. 2-2. Setup and Operation Caution: Use only one hand when making or breaking a connection to avoid making a complete circuit across your chest. Also while connecting or disconnecting the leads, do not allow any part of 8

your body to contact a grounded surface. Ensure your other hand is not touching anything that grounds you. Caution: If you are in the middle of a run and wish to connect or disconnect the leads, always turn the power off (RUN button) and wait for the display to read OFF. a. Connect the main power cord: Connect the main power cord to the power cord receptacle on the rear panel and to a suitable grounded AC power outlet. b. Connect the HV power leads: Connect the apparatus to the power supply by plugging the lead connectors into one pair of the recessed output jacks. The power leads are non-reversible as per to the color coordinated to the output jacks. c. Turn on the main power switch: Turn on the main power switch on the rear panel. The front panel and the main power LED indicator should be illuminated. d. Select constant mode: Press the CONST button to toggle between constant voltage and constant current mode. The selection is indicated by the green LED. e. Adjust limit: Press up and down buttons to set the constant variable to the desired value. 9

f. Check other run variable: Press the MODE button to check and verify the set value of the other (non-constant) run variable. This variable is set by default to the maximum output of the power supply. This is the normal condition or setting for simple electrophoretic separations. Optional: If there is another upper limit desired for this non-constant variable, use the up and down buttons to set it to the desired value. g. Set the timer (Optional): Press the MODE button to move the green indicator LED light to under clock symbol. Continuous run: Leave the timer at 0 to start a continuous run. The power supply will continue to operate indefinitely, until stopped manually. The timer will start counting up, indicating the length of the run in minutes. After reaching the maximum time of 999 minutes the timer will automatically reset and recount again from 0. Therefore for overnight operation the resulting operation time is the addition of the current timer value with the previous maximum timer 999 minutes. Timed run: Press the up and down buttons to set the timer to the desired value (minutes). The timer counts down during the run, indicating the time remaining. At the end of the timed run, a ~3 10

second beep sounds and the display changes to OFF. To reset the power supply press any button except the RUN button. PFD: The Power Failure Detection feature will save the time, voltage and current settings in the event of a power outage and automatically continue the run when the power is restored. To alert the user that a power failure occurred during the run, the error message E6 flashes at the end of the run instead of the usual OFF message. The E6 error message means that despite the power outage, the power supply successfully supplied the programmed voltage and current for the entire length of time that was programmed. The time shown at the end of the run is the time the gel ran since power was restored, not the total time of the run. The default settings for Power Failure Detection are: Run Time Setting PFD Code Timed Run 1 999 minutes ON PF1 Continuous run 0 OFF PF0 The Power Failure Detection feature can be deactivated during a timed run. Press the MODE button to move the indicator light under clock symbol. Press the UP and DOWN buttons to enter in the run time in minutes. 11

Press both the UP and DOWN buttons at the same time. The display will briefly show PF0, indicating that the PFD mode is disabled. Press both the UP and DOWN buttons at the same time again. The display will show PF1, indicating that the PFD mode is enabled. h. Start the run: Press the RUN button. The green LED light above the RUN button indicates that voltage and current are being applied to the output jacks. The green LED light above the CONST button indicates which parameter (V or ma) is limiting. The LED display shows the numerical value of the actual V or ma output, or the time (Press the MODE button to toggle between the values). The timer will count up in the case of a continuous run and count down if the timer was set for a specific timed duration. i. Stop the run: A timed run will automatically stop with a warning beep as described above. In the case of a continuous run, press RUN button to terminate the run (The green LED light above the RUN button will turn off indicating that the voltage and current which was being applied to the output jacks were stopped.). The display reads OFF. To reset the power supply press any button except the RUN button. 12

D. Troubleshooting 1. Basic troubleshooting Caution: Wait 10 seconds while disconnecting the main AC input power cord to allow voltage discharge within the power supply when restarting. Symptom Fan is not spinning LED & Display do not light up Front panel power on indicator off Fan is spinning LED & Display do not light up Front panel power on indicator off Fan is spinning LED & Display do not light up Front panel power on indicator on Action and possible causes 1. Check main power source 2. Check the fuse, replace a new same configuration fuse if necessary and try again 3. Check if power switch malfunction 4. Contact your local Recenttec representative 1. Check main power source 2. Check if any errors on the front panel LED display 3. Contact your local Recenttec representative 1. Contact your local Recenttec representative. 13

Fan is spinning Intermittent LED & Display Front panel power on indicator on Fuse blown 1. Contact your local Recenttec representative. 1. Replace a new same configuration fuse and try again 2. Contact your local Recenttec representative 2. Error codes The CONCORDE 600 power supply has built-in error messages to help troubleshoot problems with the electrical circuit. The error codes and possible corrective actions are listed below: Error Code E1 Explanation No load detected The voltage leads to the electrophoresis cell are not plugged. The current load for the separation is below 4 ma the lower limit of the power supply. Possible Solution 1. Check all the connections 2. Verify that buffer volumes are sufficient to contact all electrode wires and that buffer is contacting both surfaces of the gel. 3. Press the any button* to clear the error code. 14

E3 E4 E5 Change in Load Resistance Open circuit was detected at some point after a successful start. Buffer is leaking from the cells. There is a loose electrical connection in the circuit. Power Failure during a continuous run (no timer used) Power Failure during a timed run with PFD option disabled. Run was not complete. 1. Check and correct all potential resistance problems. 2. Check buffer volumes. 3. Check all voltage leads. 4. Check the power supply adapters fitting (If used) 5. Press any button* to clear the error code. 1. Restart the run, if desired. Press any button* to clear the error code. 1. Restart the run, if desired. 2. Press any button* to clear the error code. See page 11 for information on the PFD option. E6 Power Failure during a timed run with PFD option enabled. Despite the power outage the run was completed successfully. 15 1. Press any button* to clear the error code. See page 11 for information on the PFD option.

1. Check for any short circuit or evidence of Short Circuit electrical arching. E7 Load current exceeded 500 ma 2. Buffer concentration is too high. 3. Press any button* to clear the error code. *Press any button, except the RUN button, to clear the error message. 3. Replacing the fuses a. Unplug the power cord from the power supply. b. Locate the fuse compartment on the rear panel. c. Use a screwdriver to pry out the fuse compartment. There are two fuses in the compartment. Check and replace any blown fuses with a fuse of the same size and ratings. d. FUSES (2); 250 V, 3.15 A, 5 x 20 mm fast blow. 16

E. Care and Maintenance Turn the main power switch off and unplug the power cord before cleaning. Use a soft cloth dampened with water or a mild cleaning solution to clean the cabinet and display. If spilled liquids contact the circuit boards, unplug the power supply and allow to dry completely. Please contact your local representative or Recenttec international technical/ service specialist by E-mail: support@recenttec.com Recenttec Inc. offers complete technical support for all of our products. If you have any questions about how to use this product, or would like to arrange to repair it, please contact your local Recenttec Inc. representative or international technical / service specialist by email: support@recenttec.com. 17

F. Order Information Cat. No. Description R10-1001001 CONCORDE 200 Power Supply, 120 V, 50~60 Hz R10-1001002 CONCORDE 200 Power Supply, 230 V, 50~60 Hz R10-1001005 CONCORDE 300 Plus Power Supply, 120 V, 50~60 Hz R10-1001006 CONCORDE 300 Plus Power Supply, 230 V, 50~60 Hz R10-1001011 CONCORDE 600 Power Supply, 100~240V, 50~60 Hz R10-1001007 Mini CONCORDE Power Supply, 90~264 V, 50~60 Hz 18

Printed in Taiwan Recenttec Inc. 7F, 72 Song-Te Road, Taipei 110, Taiwan Tel: +886-2-2723-0886 Fax: +886-2-2723-0876 http://www.recenttec.com E-mail: support@recenttec.com