Mounting of large lenses in Infrared instruments

Similar documents
INGRID. Lens Mount Assembly and Alignment Procedure. Author: Paul Jolley

Gabby Kroes Lars Venema, Ramón Navarro ICSO 2012 AJACCIO GABBY KROES PAPER 0013

A Cryogenic Heat Transport System for Space-Borne Gimbaled Instruments

Opto- Mechanics and Cryostat 3/29/13 1

Cyrostat Design -Camera Section Justin Holwell

Vincotech's Compliant Pin. Application Note. Advantages of Vincotech's Power Modules with Press-fit Technology

Orbital forming of SKF's hub bearing units

LHe/LN2 Cryogenic Probe Station

Design, Development and Verification of the HIFI Alignment Camera System

High Precision Lens Mounting

Simulation Method for Connector Packaging

Lens Mounting with Elastomeric Bonds Jon Blanchard University of Arizona, OPTI Dec08

Highly Efficient Assembly of Lenses with. OptiCentric and OptiCentric Cementing

Physics 625 Femtosecond laser Project

A U T O C O L L I M A T O R S

This paper presents the design of the flexure stage, the finite element analysis, and the measured results obtained in the laboratory.

Opto-mechanical review of a light weight compact visible zoom camera

Cylinders in Vs An optomechanical methodology Yuming Shen Tutorial for Opti521 November, 2006

Equipment Standard Operating Procedure Kimberly Appel

White Paper Daniel Nindl and Mirko Wiebking February Leica Geosystems Accuracy analysis of surveying tripods

Cryostation Cold science made simple

Reduction of squeal on laminated brake disc fastened with distributed contact pressure

These devices require adequately trained and experienced operating personnel.

Setup Information Panosaurus May 3, 2011

Error Budget as a Design Tool For Ultra-Precision Diamond Turning Machines Form Errors

ishell Design Review Cryostat & Optics Bench Dan Kokubun 9/11/2013

Assembly of thin gratings for soft x-ray telescopes

Prof. Alexander Slocum John Hart Pat Willoughby

COMPACT PRECISION LINEAR MOTORIZED ACTUATORS LSMA Series LSMA-173

EXPERIMENTAL VERIFICATION OF THE AXIAL AND LATERAL STIFFNESS OF LARGE W7-X RECTANGULAR BELLOWS. Abstract

Product Information. RCN 6000 Absolute Angle Encoder with Integral Bearing and Large Hollow Shaft

Introduction. Features and Highlights

GEMINI 8-M Telescopes Project

Specification. Radiamatic Detectors and Accessories

Modeling Strategies for Dynamic Finite Element Cask Analyses

Operators Guide. Vision 100 Pan & Tilt Head. Vinten Camera Control Solutions

WP1 NUMERICAL BENCHMARK INVESTIGATION

EDUCATIONAL SPECTROPHOTOMETER ACCESSORY KIT AND EDUCATIONAL SPECTROPHOTOMETER SYSTEM

Removal and Installation8

FLAMINGOS Camera Dewar Cooldown On Kitt Peak

Allen-Bradley Drives. Instructions. (For 6180 Industrial Computers)

Technical Description of the 2.3m Telescope System. Volume 5_4 Acquisition and Guiding System (AGU)

Influence of geometric imperfections on tapered roller bearings life and performance

Further Information can be found at

AMG Series. Motorized Position and Rate Gimbals. Continuous 360 rotation of azimuth and elevation including built-in slip ring

HIE ISOLDE ALIGNMENT AND MONITORING SYSTEM TECHNICAL DESIGN AND PROJECT STATUS

A Global Laser Brand. Versatile Module

LS-785 High Throughput Lens Spectrograph User Manual

ABM's High Resolution Mask Aligner Features:

Design and production of stereoscopic instruments. User manual for the 3-D macro lens model General. Exposure

CAMERA ASSEMBLY. Removal/Replacement of the Camera Box Assembly APR-CA. Install Camera Assembly. Remove Camera Assembly

TRIBOMETERS MICRO-MACRO TRIBOLOGY TESTING

Optimal Support Solution for a Meniscus Mirror Blank

ETC. Source Four Zoom Ellipsoidal Series. Source Four Zoom. Connector Designation

Source Four Zoom 15-30

Lyte-MV Analogue Userguide Red & IR Versions

Device for utilisation of solar energy. Technical field

Influence of mesh quality and density on numerical calculation of heat exchanger with undulation in herringbone pattern

Operators Guide. Vision 8 Pan & Tilt Head. Vinten Camera Control Solutions

SPIcam: an overview. Alan Diercks Institute for Systems Biology 23rd July 2002

DESIGN CRITERIA AND APPLICATIONS OF MOTION SIMULATORS USED IN RESEARCH AND TESTING OF INERTIAL SENSOR PACKAGES FOR SPACE APPLICATIONS ABSTRACT

Operation Manual. Autocollimator Test Wedge for quick testing of Electronic Autocollimators

Operators Guide. Vision 11 Pan & Tilt Head. Vinten Camera Control Solutions

Sense it! Connect it! Bus it! Solve it! ENCODERS

ABSTRACT 1. EMIR MECHANICAL CONFIGURATION

Top Layer Subframe and Node Analysis

Behaviour of cold bent glass plates during the shaping process


Mac Mini Mid 2010 SSD Installation

Elos Fixturlaser AB. Generator alignment

11.0 Measurement of Spindle Error Motion

2016 Product Catalog

CypLite Fiber Laser Cutting Head User Manual

Good Practice guide to measure roundness on roller machines and to estimate their uncertainty

XEROX PHASER 5500 TONER & DRUM CARTRIDGE REMANUFACTURING INSTRUCTIONS

3 TESTING OF OPTICAL COMPONENTS

Friction & Roughness Measurement of Elastomer

Laserlyte-Flex Userguide

Cryogenic Engineering for Omega2000: Design and Performance

3SAE Large Diameter Splicing (LDS) System

STEP-BY-STEP INSTRUCTIONS FOR BUILDING A MICHELSON INTERFEROMETER. TECHSPEC Optical Cage System

Innovations in touch-trigger probe sensor technology

Operating Instructions for FC Adjustable Fiber Optic Adapter

BPL SERIES INSTALLATION INSTRUCTIONS THIS SHEET CONTAINS IMPORTANT SAFETY INSTRUCTIONS. SAVE THESE INSTRUCTIONS.

User English Manual for Sputnik Stereo Camera

MOUNT PROBE ONTO PRECISION MANIPULATOR

Laser Diode Collimators Type 48TE, 48-0, and 44TE, universal modular system for self-assembly

Complete Barrel Measuring and Inspection System. PS Series. User s manual

Length Measurement. Measuring lengths up to 30 m

Chapter 3. Experimental Procedure

Principles of Kinematic Constraint

GH-50. Gimbal Head. You re on steady ground

Powerful & homogeneous projector

Sonicator -Standard Operating Procedure [Updated September 10, 2014] Location of Machine: Composites Lab - Nanoparticle Containment Room, RFM 1218

Status report. Eryk Kielar, Krzystof Oliwa, Wojciech Wierba, INP PAN, Cracow, Poland Wojciech Slominski, Jagiellonian University, Cracow, Poland

IBM. Rack Installation Instructions

Rack Installation Instructions

Introduction to the FTA4000

STEP-BY-STEP INSTRUCTIONS FOR BUILDING AN AUTOCOLLIMATOR. TECHSPEC Optical Cage System

User Manual. Binocular Zoom Stereo Microscope with Boom Stand. MicroscopeNet.com

Transcription:

Mounting of large lenses in Infrared instruments Jean Louis Lizon, Gotthard Huster European Southern Observatory, Karl-Schwarzschild-Strasse 2, 85748 Garching bei München, Germany ABSTRACT Infrared instruments are generally operated at cryogenic temperature which causes a serious problem for the mounting of optical components. This is specially the case for complex dioptric systems where the optical quality is strongly dependant from the relative centering of the lenses. In the frame of the pre-design of the VLT high resolution and large field infrared imager (HAWK-I) a design using spring system has been intensively tested. The paper describes the test carried out in order to assess the thermal and mechanical behavior of the mount. A special technique has been used to accurately verify the conservation of the lens position from room temperature to cryogenic temperature. Keywords: Infrared instrument, optic mounting.. INTRODUCTION The mounting of large lens in a cryogenic environment is a technical problem that has to be carefully studied. Actually there are already a few instruments successfully using lenses of similar size. It was nevertheless important at this stage to be sure to have a suitable design and technical solutions. The mounting should ensure a safe maintaining of the lens in a stable and well-known position, thus even in case of vibration or reasonable shocks. It should also ensure an efficient and safe cooling of the lens. ESO already has some experience with the use of lenses in cryogenic instruments. Both ISAAC and SOFI include a number of objectives among of them some still suffer from alignment problems. ISAAC and SOFI also include a large BaF2 lens. The ISAAC lens collimator with 6 mm is the largest one. The mount has been designed by ESO, but with very low centering and positioning requirement as it is a single lens. The idea of the test campaign is to establish the limitation of the ESO mount and especially to verify if it could be suitable for the mounting of the various individual lenses of a complex system with tight positioning requirements. For this operation a new mount has been realised. This mount has been used with one of the first ISAAC original lenses, which was damaged during cutting at an early stage of the project where the lens should have a square shape. This document presents the design of the mount and the various measurements, which have been carried out on the system in order to qualify it for the mounting of well-centred lenses in a cryogenic environment. 2. DESCRIPTION OF THE MOUNT Figure shows a cut through the assembly and a 3D exploded view of the ISAAC lens collimator assembly. On the two figures we can easily recognise the main components. The mount itself is manufactured in one massive part () which receives directly on the optical surface of the lens. This surface is the main axial reference, and this contact will also act as the main heat sink for the lens. An elastic ring (2), centred and screwed in the cell is used to ensure the radial positioning of the lens. This ring will accommodate for the different thermal expansion coefficients. A series of compression springs (4) is used to press an aluminium clamp (3) onto the second surface of the lens. This ensures the axial immobilisation of the lens accommodating also for the different thermal expansion coefficients.

4 3 2 4 2 F radial F axial LENS Figure : Lens mount Figure 2 shows the spring force measured on one of the spring blades of the ring. The lens weights 54g. The ring holds it with a radial force of 26 N. This is the force required to radially move the lens, measured at room temperature. At 8K the lens diameter changes from 59.72 mm to 59.35 mm (measured). The ring diameter changes from 59.6 mm to 59.4 mm. In that situation the radial force should increase to approximately: 54 N. Force (N) 4.5 4 3.5 3 2.5 2.5.5 At 8K At room temperature 2 3 4 5 Deformation (micron) Figure 2: Radial spring force 3 The radial fixation is reinforced by the friction of the lens on its support and on the axial clamp. The axial fixation force is provided by 2 springs where each of one provides 8 N. 3. DESCRIPTION OF THE TEST The main goal of the test is basically to verify that the lens does not move relative to the mechanical part during thermal cycling and that vibration or shock does also not affect the position. Figure 3 shows the set-up used to measure both the linear and angular position of the lens inside the cryostat. The autocollimator A coupled with a convex lens is used to measure the centre of curvature of two halfspherical mirrors. One half is fixed on a bridge which is attached on the mount while the second half is attached on a small part punctually glued on the middle of the lens. In combination with a well-suited lens the electronic autocollimator allows to measure motion in the order of. micron. The autocollimator A2 is used to directly measure the angular position of two small mirrors, one attached on a bridge link to the mount and one mounted on a small part glued on the centre of the lens. Figure 5 and 6 show the two sides of the lens equipped with the four mirrors.

Lens Mirror B Mirror 2B Autocollimator A Mirror A Mirror 2A Autocollimator A2 Figure 3: Measurement set-up Figure 4: Angular measurement Figure 5: Linear measurement Figure 6: Test setup The lens mount is bolted on the cold plate of a bath cryostat, which is fitted with two windows. Figure 6 shows the cryostat on the optical table with the peripheral equipments (vacuum pump, LN2 tank, temperature measurement system) and the two autocollimators. 4. RESULTS A total of 25 cool-downs have been carried out with different parameters. Every time between two cooling operations, the cryostat has been opened, the clamping flange removed to allow the lens to re-centre itself. This also explains why the test campaign took longer than foreseen. During the first cooling operations, the position of the lens has been recorded every hour. For every cooldown the final position of the lens at 8 K has been compared with the initial position at room temperature. Shock tests have been performed at the end of cooling 6, 7and. They have been repeated at the end of cooling 8 and 9 after increasing the radial supporting force. The 9 first cool-downs have been performed in 2, 3 hours, the cool-downs 2, 2 and 22 in 8 hours and the cool-downs 23,24 and 25 in 6 hours.

Linear position Figure 7 shows the motion of the lens during the first cool-downs. The amplitude is rather important and the lens did not recover its position after warm-up. 9 Y 8 7 6 X Z Y motion (micron) 5 4 3 2 -.2 -...2 X motion (micrron) Figure 7: Motion of the lens during the three first cool-downs The third test carried out with the cryostat horizontal shows that it was not related to the gravity but only to the direction of the heat sink. The logic conclusion was that it was related to the cooling and more precisely with the cooling homogeneity. Then some heavy copper straps have been added in order to coll also the part of the mount which is far from the cold plate. Figure 8 shows a photograph of the new mounting. Figure 8: Lens mount with additional cooling straps Figure 9 shows the motion of the lens measured during various cool-downs after addition of the cooling straps.

A homogeneous cooling of the lens considerably improves the stability. Cooling n: 8 with the lens mounted with a higher radial force shows also the best result. This is also confirmed by the measurement of the final position after cooling 8, 9, and 2 to 25 shown in figure..4 Y motion (micron).3.2. -.5 -. -.5.5..5.2 -. -.2 -.3 Y Position (micron )..5 -.5 -. -.5 -.5.5..5.2.25 -. -.5 -.2 -.25 X Position (micron ) -.4 X motion (micron) Figure 9: Motion of the lens during cool-down with homogeneous cooling σ Cooling 4, σ Cooling 2, ν Cooling 8 Figure : End position of the lens After cooling 4 to 7, σ after cooling 8 and 9, after cooling 2 to 25 A few measurements have been done in order to establish the influence of the gravity on the position of the lens relative to the mount. No detectable motion has been recorded over 36 degrees rotation around Z. As already mentioned, the influence of shocks has been investigated. For this test the cryostat has been placed on the plate off a vertically sliding support. Then the support was left falling free from different heights in order to create various accelerations. The acceleration is given at the level of the lens mount with some % uncertainty. A calibration has been done prealably on the open cryostat with an accelerometer on the lens mount.. Y motion of the lens (micron ) -. -.2 Cool-down 6 -.3 Cool-down 7 -.4 Cool-down -.5 Cool-down 8 Cool-down 9 -.6.5.5 2 2.5 3 3.5 4 Acceleration (g) Figure : Effect of shocks along Y Figure 3 gives the results of the measurement for shocks along Y. The record has shown very low motion along X, therefore this information is not given graphically. In order to improve stability, the radial force has been slightly increased after the cooling 7. For this the two halves of the ring have been moved toward the centre by 5 microns. This increases the radial stiffness only along Y, which is the direction of the shock.

The cryostat is not suited to do some shock tests in other direction. Nevertheless it has been verified that shock of an amplitude of g along z did not cause any motion of the lens. Angular position All along these operations the angular motion of the lens has been measured with the second autocollimator. During the 3 first cool-down the lens rotates slightly around X. the amplitude of the motion was around 2-arc sec. The motion was less reproductive and less systematic than for the linear motion. Angular position Rot Y (Arc sec.).5 -.5 - -.5 -.5.5.5 2 - -.5-2 -2.5 Angular position Rot Y (Arc sec.) Figure 2: Angular position of the lens After cooling 4 to 7, σ after cooling 8 and 9, after cooling 2 to 25 Figure 4 shows the final angular position of the lens compared to the mount measured at 8 K. There is a clear correlation between the angular and linear positions. It also seems to help to have a uniform cooling here. Thermal aspect, cooling limitation The thermal contact conductance between the lens and the lens centering barrel has been measured to be:. W/K. Temperature (K) 32 27 22 7 2 7 Cooling to 9 cooling 2 to 22 Cooling 23 to 25 Cooling 26 Cooling 27 cooling 28 Cooling 29 2 3 4 5 6 7 8 9 2 Time (hours) Figure 3: Temperature evolutions

In addition to the 25 cool-down operations mentioned before, a few more cool-downs were done in order to assess the risk of braking the lens while cooling too fast. These cool-downs were not done in the cryostat which was limited to 6 hours cooling time. For these fast coolings the lens mount has been slowly dived into an open bath of liquid nitrogen. The lens has been successfully cooled in 3 hours. The lens also survives two more coolings, one in hour and one in a little bit less than one hour. Finally the attempts to cool the lens in some half an hour have been fatal. The lens brakes at K under the internal tension. The brakeage does not seem to be related to the mount. Figure 4 shows the lens after the last cooling. Figure 4: The lens brakes during a cooling in 4 mn