Agilent 85130C NMD 3.5 mm to Type-N Adapter Kit

Similar documents
Keysight 85130D NMD 3.5 mm to 3.5 mm Adapter Kit

Keysight 85130F NMD 2.4 mm to 3.5 mm Adapter Kit

Keysight 11878A 50 Ohm 3.5 mm Adapter Kit

Operating and Service Manual

Operations and Installation Manual. Keysight 85132E/F 3.5 mm to 7 mm Test Port Return Cables

Keysight 11636B APC-3.5 mm Power Divider DC to 26.5 GHz

Agilent CSV Export Utility

Agilent N2739A 1000 Series Oscilloscope Rack Mount Kit

BenchCel Workstations Software

Agilent OpenLAB. Data Store. Backup and Restore Guide

Agilent OpenLAB. Data Store. Maintenance Guide

Installation Note. Source Attenuators and Bias Tees Upgrade Kit. For E8362B/C PNA Series Microwave Network Analyzers. Network Analyzer Model Number

Agilent Series Logic Analysis System

Agilent 1260 Infinity Purification Solution

Agilent OpenLAB Chromatography Data System (CDS)

Agilent N2916B 6000 and 5000 Series Oscilloscope Rack Mount Kit

Agilent OpenLAB Data Analysis Upload Download Tool. User s Guide

Agilent Technologies E5385A 100-Pin Probe

Agilent InfiniiMax II 1168A/1169A Probes

Agilent OBSAI Protocol Tester

BIT-3000 Dynamic Sequencing Generator and Analyzer. Datasheet 1.11

Keysight Add Source and Receiver Attenuators Upgrade Kit

Firmware Guide. Keysight PXIe Chassis Family

Agilent Technologies E5346A 38-Pin Probe and E5351A 38-Pin Adapter Cable

Agilent OpenLAB Chromatography Data System

Keysight Second Source, Combiner, and Mechanical Switches Upgrade Kit

Agilent W2100 Antenna Modeling Design System

Microscan Barcode Reader

Installation Note. Removable Hard Disk Drive Upgrade Kit. For All PNA Series RF Network Analyzers. Network Analyzer Model Number

Automated Plate Labeling

Agilent EZChrom SI. Startup Guide

Reference Guide. Agilent Technologies Electronic Calibration Modules. RF Two-Port 85091C 85092C 85093C 85096C 85098C 85099C. RF Four-Port N4431B

Agilent OpenLAB Chromatography Data System (CDS)

Agilent Lab Advisor. IT Administrator's Guide. Agilent Technologies

Agilent OpenLAB Data Store Administration. Guide for Administrators

Agilent E2943A/E2944A ATCA Probes for Advanced Switching Interconnect

Keysight 54006A 6 GHz Passive Divider Probe Kit. User s Guide

Agilent OpenLAB Chromatography Data System (CDS)

Agilent ChemStation Plus

Setting Up Probes. Online Help

Extended Frequency Range Upgrade Kit (50 GHz to 67 GHz)

Agilent OpenLAB Data Store. Maintenance Guide

Keysight Second Source, Combiner, and Mechanical Switches Upgrade Kit

Installation Note. IF Access Upgrade Kit. For PNA Series Microwave Network Analyzers. Network Analyzer Model Number. Upgrade Kit Part Number

Agilent N5410A Fibre Channel Automated Test Application

Keysight N7015A/16A Type-C Test Kit. User Guide

Keysight M9393A PXIe Vector Signal Analyzer

Nimsoft Monitor. proxy Guide. v3.1 series

Agilent OpenLAB Chromatography Data System (CDS)

Agilent OpenLAB Chromatography Data System

Agilent Dissolution Tester Driver

86100 Series Infiniium DCA Oscilloscope

Keysight E5864A Removable Hard Drive for Series Logic Analyzers. Installation Guide

Agilent OpenLAB Chromatography Data System (CDS)

Agilent CytoGenomics 2.5

* * Agilent Power Distribution Unit (PDU) Installation Guide

Keysight N108xA-Series IEEE Ethernet KR/CR Compliance and Debug Application

G5584A Labware MiniHub

Agilent ChemStation Plus

Agilent SureCall 3.5. Installation Guide. Agilent Technologies. For Research Use Only. Not for use in diagnostic procedures.

Automated Plate Centrifugation

87421/22A Power Supply. Operating and Service Manual

Agilent U2941A Parametric Test Fixture

Agilent Genomic Workbench 6.0

Agilent ChemStation for UV-visible Spectroscopy

PlateLoc ActiveX v3.1

Agilent E2094M IO Libraries

Keysight E7515A UXM Firmware Update Tool

Agilent 89600B VSA. Software Installation Guide

U85026A Detector 40 to 60 GHz

Agilent 34826A BenchLink Data Logger for 34980A. Getting Started Guide. Agilent Technologies

Cary 50. Pre-Installation Manual

Agilent Gb/s Serial BERT

Agilent G2721AA/G2733AA

Agilent OpenLAB ECM Intelligent Reporter

Getting Started with the Agilent Serial BERT N4906B. You only need a few minutes to get started with the Serial BERT.

Keysight Technologies 85131C/D/G NMD-3.5 mm -fto 3.5 mm Semi-Rigid Test Port Return Cables

Agilent Protocol Analyzer and Jammer for USB. Quick Start Guide

User s and Service Guide

Agilent ChemStation. ECM Interface Guide. Agilent Technologies

Advanced Test Equipment Rentals ATEC (2832)

DDR/LPDDR Custom Configuration Creator Tool. User s Guide

Keysight 85052D 3.5 mm Economy Calibration Kit

Agilent Technologies 11667A DC to 18 GHz Power Splitter

Keysight Technologies 85052B 3.5 mm Calibration Kit

N2099A PXIT Synthesizer Module. Version User s Guide. Agilent Technologies

Agilent Technologies E5339A 38-Pin Low-Voltage Probe

Keysight Combiner and Mechanical Switches Upgrade Kit

Programming Note. Agilent Technologies Quick Reference Guide For the 8757D/E Scalar Network Analyzer

Agilent N4219B Packet Analysis Probe for Serial ATA. User s Guide

User s and Service Guide. Agilent Technologies 85052C 3.5 mm Precision Calibration Kit

User s and Service Guide

User s and Service Guide. Agilent Technologies 85033E 3.5 mm Calibration Kit

Agilent E4418/19A Power Meter Hardware Upgrade Kit

Agilent N5394A DVI Electrical Performance Validation and Compliance Software

Keysight N5191A/93A UXG X-Series Agile Signal Generator

EZChrom Elite Compact. What s New in EZChrom Elite Compact Service Pack 2

User s and Service Guide

Errata. Agilent References in this manual. About this manual. Support for your product

Automated Serial Dilution

Transcription:

Agilent 85130C NMD 3.5 mm to Type-N Adapter Kit Operating and Service Manual Agilent Technologies

Notices Agilent Technologies, Inc. 1987, 2004, 2008 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws. Manual Part Number 85130-90010 Edition November 15, 2013 Supersedes: November 2008 Agilent Technologies, Inc. 5301 Stevens Creek Blvd Santa Clara, CA 95051 Warranty The material contained in this document is provided as is, and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control. Technology Licenses The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license. Restricted Rights Legend If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as Commercial computer software as defined in DFAR 252.227-7014 (June 1995), or as a commercial item as defined in FAR 2.101(a) or as Restricted computer software as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies standard commercial license terms, and non-dod Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data. Safety Notices CAUTION A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met. WARNING A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.

85130C Adapter Kit General Information To obtain optimum performance from this adapter kit, observe these simple precautions: Make connections carefully to avoid misalignment and connector damage, which will result in inaccurate measurements. Keep the connectors free of dirt and any particles. When you clean the connectors, try using compressed air first. Do not use abrasives. With a clean foam swab, apply only isopropyl alcohol. For more information, refer to the Connector Care for RF and Microwave Coaxial Connectors document. It can be viewed online by searching for part number 08510-90064 at www.agilent.com. Description The 85130C adapters are used when type-n devices must be attached to a 3.5 mm test set in a network analyzer system. The test set end of the adapters has a NMD-3.5 mm (f) connector while the Device Under Test (DUT) end has a precision slotless type-n connector (PSC-N). Contents The 85130C kit contains the following: Test port adapter, NMD-3.5 mm (f) to PSC-N (m) (Agilent part number 85054-60030) Test port adapter, NMD-3.5 mm (f) to PSC-N (f) (Agilent part number 85054-60029) Storage box, foam-lined Operating and Service Manual Spanner wrench 85130C Operating and Service Manual 1

Specifications Agilent Technologies guarantees that your adapters will equal or exceed the following specifications in the +20 to +26 C (+68 to +79 F) temperature range. Frequency DC 8 GHz Return Loss 34 db 8 18 GHz 28 db The allowable recession of the center conductor of the NMD-3.5 mm (f) connector is 0.0006 to 0.0022 inches ( 0.015 to 0.056 mm) below the mating surface of the outer conductor. The allowable pin position for type-n connectors is: male = 0.2075 to 0.2079 inch ( 5.2705 to 5.28066 mm) female = +0.2061 to +0.2065 inch (+5.23494 to +5.2451 mm) In a mated pair of type-n connectors, the center conductor s mating plane is offset from the outer conductor s mating plane by 0.207 inch (5.2578 mm) in the direction of the male connector. See Figure 1 on page 3. Zero the gages before each use by attaching the appropriate gage master to the end of the gage, torquing the connection to 12 in-lb with a 3/4 inch torque wrench, and adjusting the dial on the face of the gage so the gage reads zero. Both type-n gages, male and female, read zero when the center conductor to outer conductor offset is nominally 0.2070 inch (5.2578 mm). When gaging a male connector, the actual recession of the center conductor is 0.2070 inch plus the reading on the gage. When gaging a female connector, the actual recession of the center conductor is +0.2070 inch plus the reading on the gage. Example: for a gage reading of 0.0001 inch, the following is true: Gage Reading Actual Pin Position (versus outer mating plane) -- Male Female 0.0001 inch 0.2070 + ( 0.0001) = 0.2071 inch +0.2070 + ( 0.0001) = +0.2069 inch 2 85130C Operating and Service Manual

Figure 1 Type-N Connectors Proper Use Attach the adapters to the test ports and tighten them finger tight. Use the spanner wrench to hold the test set end of the adapter and torque the test set connector with a 20 mm torque wrench set to 96 N-cm (8 in-lb). Performance Tests Use a network analyzer to perform the following return loss test on your adapters as soon as you receive them. Periodically repeat the test to determine if the performance meets the electrical specifications on the previous page, or if they need to be replaced. An initial period of one year between performance tests is recommended. Required Equipment Network analyzer 18 GHz (or higher) measurement capability with time domain option 010 Type-N loads (included in the 85054B calibration kit) Type-N 50 ohm airline, 12.5 cm (included in the 85055A verification kit) Agilent Model or Part Number PNA with Option 010 (See PNA Family Microwave Network Analyzers Configuration Guide, part number 5990-7745EN) 00909-60011 (male) 00909-60012 (female) 85055-60001 85130C Operating and Service Manual 3

Figure 2 Return Loss Setup Return loss is measured by connecting a 50 ohm fixed load termination through a 12.5 cm airline to the adapter, then attaching the adapter to port one of the test set (see Figure 2). The effects of an imperfect load may be gated out using the network analyzer time domain option. NOTE Refer to your network analyzer s Help system for specific instructions on using the functions mentioned in the return loss test below. 1. Preset the analyzer. 2. Set a stimulus start frequency of the analyzer s lowest frequency. 3. Set a stimulus stop frequency of 18 GHz. 4. Set an IF bandwidth of 100 Hz. 5. Perform, and then save, a 3.5 mm 1-port S 11 calibration. 6. With correction turned on, select the time domain mode. 7. Set a stimulus start time for the sweep to 0.05 nano-seconds. 8. Set a stimulus stop time for the sweep to 1.9 nano-seconds. 9. Select the gating function and gate-out everything but the adapter. See Figure 3. 10. Select the analyzer s frequency domain mode. 11. Use the markers to read the return loss value. 4 85130C Operating and Service Manual

Figure 3 Analyzer Trace Showing Location of Gates and Airline Performance Test Record ELECTRICAL SPECIFICATIONS Tested by: Date: Adapter Frequency Range Return Loss Measured NMD-3.5 mm (f) to PSC-N (m) and NMD-3.5 mm (f) to PSC-N (f) DC to 8 GHz 34 db 8 GHz to 18 GHz 28 db Replaceable Parts There are no replaceable components for the adapters. A worn or damaged adapter must be replaced in whole. 85130C Operating and Service Manual 5

Equipment and Supplies The following equipment and supplies are required for the maintenance and use of, but are not supplied with, your 85130C adapter. kit. Item Part Number 3.5 mm gage kit (part of the 85052B calibration kits) 85052-80010 Type-N gage kit (part of the 85054B calibration kit) 85054-80011 Torque wrench, 3/4, 136 N-cm (12 in-lb) (part of the 85054B calibration kit) 8710-1766 Torque wrench, 20 mm, 96 N-cm (8 in-lb) 8710-1764 Document: Connector Care for RF and Microwave Coaxial Connectors 08510-90064 Contacting Agilent Assistance with test and measurement needs and information on finding a local Agilent office are available on the Web at: www.agilent.com/find/assist NOTE In any correspondence or telephone conversation, refer to the Agilent product by its model number and full serial number. With this information, the Agilent representative can determine whether your product is still within its warranty period. 6 85130C Operating and Service Manual