Rev Carbon Dioxide (CO2) Gas Sensor. TG100 User Manual

Similar documents
Rev Carbon Dioxide (CO2) Gas Sensor. TG100 User Manual

Trueyes Inc. Carbon Dioxide (CO2) Module

General. Copyright 2013 EXSEN Company. All Rights Reserved. Rev /5/2013 Page 1 of 6

Rev 1.3, Air-Farm User Manual. CO2 / Temperature / Humidity Transmitter

MH-Z16 Intelligent Infrared Gas Module

Intelligent Infrared CO 2 Gas Sensor. (Model: MH-711A) Manual. Version: 3.3. Valid from: May 1st, 2014

HMC1022 Digital Compass

Manual iaq-engine Indoor Air Quality sensor

Manual iaq-core Indoor Air Quality sensor module

RoHS. CO 2, Temperature & Humidity Transmitter Applications

KCD-HP. KCD-HP200x, 300X. [Figures] Top : 1% Sensor probe Bottom left : 10% / 20% Sensor probe Bottom right : Holding bracket(optional) Measurement

Product Specification. CO 2 Engine K33-LP T. Sensor Module and OEM Platform. Document PSP0120. Rev 4. Page 1 (13)

CAN / RS485. Product Description. Technical Reference Note. Interface Adapter. Special Features

MH-Z16 Intelligent Infrared Gas Module Manual

JMY504M User's Manual

JMY505G User's Manual

Model 305e OEM CO 2 concentration sensor

2F. No.25, Industry E. 9 th Rd., Science-Based Industrial Park, Hsinchu, Taiwan

2.2 Main technical parameters. Detection range. 0~5%vol(0~100%vol optional) Warm-up time. SPAN <±500ppm Long-term drift

Page 1 MRK-D-0011, V2.0 Aeroqual SM50 User Guide

ZG07 CO2 Module. ZG07 series CO2 Module. User Manual

RADON SS SENSOR pci/l

Page 1 MRK-D-0011, V1.1 Aeroqual SM50 User Guide

Page 1 of 11 Version 1.2 Release Date: PenMount PM2204 PCI Controller Board Data Sheet. Table of contents

Gas and Air Sensors. Product Specification. CO 2 Engine K30. Sensor Module and OEM Platform. Document PSP 110. Rev 4. Page 1 (11)

Product Specification. CO 2 Engine ICB. Sensor module for bio applications

XS S ERIES TM PMB US TM O PTION C ARD

EXCELL PRECISION CO., LTD.

NHD-0220D3Z-FL-GBW-V3

PRELIMINARY! DIGITAL HUMIDITY SENSOR

Product Specification. CO 2 Engine K30 STA. Sensor Module and OEM Platform

Product Specification. Senseair LP8. Miniature CO 2 sensor module for battery-powered applications

Installation and operation manual ReciFlow Gas

T1K MODBUS Base Controller Specifications

EZO-CO2. Gaseous CO2. Reads. 0 10,000 ppm. Range. Factory calibrated. Calibration. 1 reading per second. Response time. 1 ppm.

MH-Z14 Intelligent Infrared Gas Module User s Manual

NHD-0216K3Z-NS(RGB)-FBW-V3

Page 1 of 13 Version 1.5 Release Date: PenMount PM1410 PCI Controller Board Data Sheet. Table of contents

NHD 0216K3Z FL GBW. Serial Liquid Crystal Display Module

Amphenol Advanced Sensors. SM-UART-04L Laser Dust Sensor. Application Note. Features. Applications

Product Specification. Sensor Module and OEM Platform

T67xx C0 2 Sensor Module Application Note

WBI Series Mass flow sensors for gases

NHD-0220D3Z-FL-GBW-V3

DS Wire Digital Thermometer and Thermostat

APPLICATION NOTE VACUUM FLUORESCENT DISPLAY MODULE

DS1845 Dual NV Potentiometer and Memory

an ISO9001/14001 company

Dual Interface LCD Display Controller

JP2380 Sensor Module. JP Sensor. Features: MCU Embedded. Specifications. Sensor with total solution

Controller Board for Projected Capacitive Touch Screen DUS2000 Product Specification

APPLICATION NOTE VACUUM FLUORESCENT DISPLAY MODULE

Analog signal to RS485 or RS232 Converter, A/D Converter

Signal Isolators & Conditioners

RS232-ADC16/24 Manual

User Manual for TeraRanger Evo with: USB and I2C/UART backboard

Datasheet: K-33 BLG Sensor

UWE-1707 SERIES CONTENTS OPERATION MANUAL ELECTRONIC WEIGHING INDICATOR 1. SPECIFICATIONS 2. INSTALLATION

OEM-ORP ORP. Reads mV mV. Range. 1 reading every 420ms. Response time. Any type & brand. Supported probes. Single point.

User Manual for. TeraRanger Evo with: USB and I2C/UART backboard

Before powering on your driver, read this manual thoroughly. If you have any doubt or suggestion, please do not hesitate to contact us!

INTRODUCTION Laser Particle Sensor Module PM2005

Controller Board for Projected Capacitive Touch Screen DUS1000 Product Specification

ILI2511. ILI2511 Single Chip Capacitive Touch Sensor Controller. Specification ILI TECHNOLOGY CORP. Version: V1.4. Date: 2018/7/5

CO 2 Sensor Evaluation Board EM7160

K33-BLG/ELG Sensor module for environment logging Data Sheet and Manual

JSD81 series products are based on SCM's intelligent monitoring and control system, users set the calibration value,

Compact Card Dispenser

WM1030 Rev Introduction. Ultra low power DASH7 Modem. Applications. Description. 868 / 915 MHz. Features. WIZZILAB Technical datasheet 1/10

Altimeter / Barometer Module SMD500 ultra low power, low voltage

User Manual for TeraRanger Evo single point distance sensors and backboards

WIZ220IO / WIZ220IO-EVB User s Manual (Ver. 1.0)

THERMAL PRINTER SPEC BAS Basic Features. 1.) Type : PANEL Mounting or DESK top type. 2.) Printing Type : THERMAL PRINT

DS WIRE INTERFACE 11 DECOUPLING CAP GND

PAT9125EL: Optical Tracking Miniature Chip

AQ_G24 GSM Terminal Card Motorola Cellular GSM Engine

PCD DIGITAL Differential pressure probe

User Manual for TeraRanger Evo single point distance sensors and backboards

ZG01 CO2 Module. ZG01 CO2 Module. User Manual

Revision History Rev. Date By Summary Remark /07/16 Ken Hsu New release

DS1624 Digital Thermometer and Memory

MF1-RW-TTL-PCB Mhz Mifare ISO14443A reader/writer module with TTL interface CONTENT 1. MAIN FEATURES...,, PINNING INFORMATION...

TABLE OF CONTENTS. Communication Functions

Signal Isolators & Conditioners

RN-174. WiSnap M2 Super Module. Features. Description. Applications. ~ page 1 ~ rn-174-ds v1.1 6/1/2011

MODEL : TRP-100-II Receipt Printer User s Manual

FXL6408 Fully Configurable 8-Bit I 2 C-Controlled GPIO Expander

Chapter. Modbus. In This Chapter...

EZmoto V4.1 Product description Rev. 2 30/07/2015

Compact Card Dispenser

OEM Digital NDIR CO 2

User-configurable Resolution. 9 to 12 bits (0.5 C to C)

ORDERING INFORMATION. OPERATION Measuring Temperature A block diagram of the DS1621 is shown in Figure 1. DESCRIPTION ORDERING PACKAGE

CW2013. Low-Cost 1s Fuel Gauge IC with Low-SOC Alert. General Description. Features. Applications. Order Information

MDP200 Series. Differential Pressure Sensor FEATURES APPLICATIONS DESCRIPTION

Product Specification

OEM Digital NDIR CO sensor

Modbus on tsense. Table of contents:

! User Manual Rev. 1

DCC-8 DIGITAL TO EIGHT CURRENT LOOP CONVERTER OPERATING MANUAL

Transcription:

Rev. 2.93 TG100 User Manual

The TG100 measuring carbon dioxide (chemical formula CO2) is a NDIR (Non-Dispersive Infrared) gas sensor. As it is contactless, it has high accuracy and longer life than sensors having a different way. This sensor has been developed with standard gas and is efficiently calibrated using our state of the art gas sensor calibration technology. Furthermore, this sensor comes with Serial Communication (UART), I 2 C interface and Analog Output. So, CO2 density can be checked using these inter faces easily. Features NDIR Gas Sensor High Precision Expected Life Span, 10 years Compact Size, 37 x 28 mm RoHS Compliance Model : TG100 Series Interfaces UART, 38400 Baud Rate I 2 C Interface Analog Output Status LED GPIOs for calibration. 1

Specifications Table 1. Electrical Specifications Parameter Symbol Min Typ. Max Unit Notes Regulated DC Power Supply is used, Test Condition V DD = 5 V Power Supply V DD 4.75 5 5.25 V Current Consumption I DD 12-21 ma ready state I DD 100-210 ma while gas measurement * The noise on power supply may cause incorrect measurement. Table 2. Sensor Specifications Parameter Symbol Min Typ. Max Unit Notes Measuring Range 0-10k ppm Accuracy ±5% ±30ppm +degree C ±7% ±50ppm -degree C Detection Limit - 10 - ppm ( Resolution @ 0ppm region) Response Time t 1/e - - 18 sec t 90 - - 30 sec Operating Temperature T O * T OL - * T OH Operating Humidity H 0-95 %,RH Non Condensing * T OL and T OH are mentioned at ordering guide on Page 17. Refer the ordering guide page. Storage Temperature T S -30-70 Storage Humidity H 0-95 %,RH Non Condensing Warming-Up Time t w 30 120 - sec for accuracy Data Sampling Time t s - 2 - sec Pulse Operation 2

Table 3. UART Interface Specifications Parameter Symbol Min Typ. Max Unit Notes Recommended Request Time T UR 0.5 1 - sec Transmitted Time T UT - 2 - sec Acceptable TTL Signal High Level V HU 2.31 3.3 5.25 V TXD, RXD Acceptable TTL Signal Low Level V LU - - 0.99 V TXD, RXD Table 4. I 2 C Interface Specifications Parameter Symbol Min Typ. Max Unit Notes Operating Frequency f I2C - - 200 khz Request Time T I2C 1 - - sec Acceptable TTL Signal High Level V HI 2.3 3.3 5.25 V SCL, SDA Acceptable TTL Signal Low Level V LI - - 0.99 V SCL, SDA Table 5. Analog Output Specifications Parameter Symbol Min Typ. Max Unit Notes Analog Output Range V A 10-4000 mv 10mV 25 ppm Conversion Time T V - 2 - sec * The range of analog output depend on maximum value of measured ppm. 3

UART Interface This sensor comes with a serial interface for user to read ppm values. Also, several commands are supported. About commands, refer to Table 9. Also, there are examples on next page. To use UART interface, refer to serial communication configuration in Table 6. Table 6. Serial Communication Configuration Baud Rate Stop Bits Parity Check Data Length TTL Level 38,400 bps 1 bit None 8 bits 3.3V or 5V If the communication configuration is set correctly, user will get the protocol message as mentioned below. Table 7. Protocol Format Byte 1 1 1 1 1 1 1 1 1 1 1 1 Value D6 D5 D4 D3 D2 D1 <SP> p p m <CR> <LF> Table 8. Protocol Format Explanation Value D6 ~ D1 <SP> ppm <CR> <LF> Gas density is indicated as string, Maximum length is 6 bytes. Space : 0x20 ppm : 0x70 0x70 0x6D Carriage Return : 0x0D Line Feed : 0x0A The number of <SP> between D6 and D1 will be different for different gas densities. So, user will get ppm values in ASCII format, shown below. To write commands, Follow the important points below to keep in mind while sending serial commands and refer to the examples given on pages 5 to 7. 1 This protocol format conforms to ASCII format. 2 Before send command, Enter command mode. 3 To enter in command mode, Send ATTN<CR><LF> and to exit, Send RUN<CR><LF>. 4 All commands should be delimited by<cr><lf>. Table 9. UART Command Table Code ACAL MCAL Period of automatic calibration can be changed. Manual Calibration is requested. 4

Getting Started with UART Commands User can communicate with sensor using PC software / terminal program or another device. Which supports serial communication on same baud rate as mentioned in serial interface specification. If user has connected their serial device correctly, they will receive string message from sensor. The message will be transmitted at intervals of about two seconds. Example RX> <SP>643<SP>ppm<CR><LF> Example 1. ASCII Code Message Example RX> 20 36 34 33 20 70 70 6D 0D 0A Example 2. HEX Code Message Entry Command Mode To use any command, sensor should be entered in command mode. To enter in command mode, user should send ATTN<CR><LF>. Otherwise, the commands will not recognized by the sensor. Example TX> ATTN<CR><LF> RX> ATTN<CR><LF> RX> <ACK> Example 3. ATTN Message Exit Command Mode To exit from the command mode, user should send RUN<CR><LF>. When user is not using serial interface, to continue with normal running mode. Example TX> RUN<CR><LF> RX> RUN<CR><LF> RX> <ACK> Example 4. RUN Message 5

ACAL Command Mode About automatic calibration, The information is on Page 13. Read the page before use this command The period of automatic calibration can be changed using this command. ACAL07<CR><LR> If the user wants to run ACAL for 7 days, user should send ACAL07<CR><LF>. Refer to Example 5 Example TX> ACAL07<CR><LF> RX> ACAL07<CR><LF> RX> <ACK> RX> ACAL<SP>Period<SP>is<SP>7<SP>Days<CR><LF> Example 5. ACAL 7 days Message ACAL30<CR><LR> If the user wants to run ACAL for 30 days, user should send ACAL30<CR><LF>. Refer to Example 6 Example TX> ACAL30<CR><LF> RX> ACAL30<CR><LF> RX> <ACK> RX> ACAL<SP>Period<SP>is<SP>30<SP>Days<CR><LF> Example 6. ACAL 30 days Message 6

MCAL Command Mode At Page 14, There are instructions about manual calibration. Read the instructions before use this command. MCAL=<Value><CR><LR> When ppm value is put that user wants, Sensor will response like Example 7 message mentioned below. It means that manual calibration is started. And then when manual calibration is finished, sensor will response like Example 8 message. Example TX> MCAL=500<CR><LF> RX> MCAL=500<CR><LF> RX> <ACK> RX> MCAL<SP>ON<SP>:<SP>500<SP><CR><LF> Example 7. Starting MCAL Message Example RX> MCAL<SP>is<SP>activated<CR><LF> Example 8. Finishing MCAL Message 7

I 2 C Interface This sensor communicates with the host controller, over a digital I 2 C interface. The 7-Bit base slave address is 0x31. Table 10. I 2 C Slave Address Byte Bit 7 6 5 4 3 2 1 0 Data 0 1 1 0 0 0 1 R/W * R/W : Read = 1, Write = 0 Master I 2 C device communicates with our sensor, using a command structure. The commands are listed in, Table 11. Commands, other than those documented below are undefined and should not be sent to the device. Table 11. I 2 C Command Table COMMAND DATA n 0x52 7 Byte Read PPM and state of configuration 0x53 2 Byte Request a manual calibration In the I 2 C sequence diagrams in the following sections, bits produced by the master and slave are color coded as shown : Master Slave Table 12. I 2 C Bit s Value Symbol START S SDA goes low while SCL high STOP P SDA goes high while SCL high Repeated START Sr SDA goes low while SCL high. It is allowable to generate a STOP bef ore the repeated start. SDA can transition to high before or after SCL goes high in preparation for generating the START READ R Read bit = 1 WRITE W Write bit = 0 ACK A Acknowledge NACK NA No acknowledge Inside sensor, there are pullup resistors on each I 2 C line. Refer to the circuit below. 8

I 2 C Sequence to perform a ppm measurement and read manual calibration state The master device should request to slave device. The slave device means this sensor. S Slave Address W A COMMAND A Bit Slave Address 0b0110001 Byte Command 0x52 * If Slaver Address = 0b0110001, They can be represented in hexadecimal as Slave Address = 0x31. Sr Slave Address R A DATA 0 A DATA 1 A DATA 2 A Configuration PPM data Byte 0x08 (fixed) Byte 0x0000 ~ 0x2710 * DATA 1 = High Byte, DATA 2 = Low Byte If DATA1 = 0b00000011, DATA2 = 0b11101000 (Binary), They can be represented in hexadecimal as DATA1 = 0x03, DATA2 = 0xE8. And they can be combined as 0x03E8, It also can be represented in decimal as 1000. So DATA1 and DATA2 are represented 1000 as ppm density. DATA3 A DATA 4 A Byte ACAL. Period 0x07 (7days) 0x1E (30days) Byte MCAL. State 0x00 (Ready) / 0x01 (Running) To know a finished manual calibration, Check a change that 0x01 0x00. The default value is 0x1E. It is unrelated to ACAL State. DATA 5 NA P Byte ACAL. State 0x00 (Disable) 0x01 (Enable) The ACAL State can be changed using Hardware Pin, refer to Page 13. Figure 2. I 2 C Packet while reading data 9

I 2 C Sequence to write a value in a manual calibration. At Page 14, There are instructions about manual calibration. Read the instructions before use this function. The master device should request to slave device. After communication using this sequence, User should check the MCAL. State (Refer Figure 2). If the MCAL. State is still 0x00, It means entered data was rejected. S Slave Address W A COMMAND A DATA 0 A DATA 1 A P Command Entering PPM data Byte 0x53 Byte 0x0190 ~ 0x2710 Figure 3. I 2 C manual calibration packet 10

Analog Output, Voltage (V) Analog Output This sensor has analog output to read ppm value. To read ppm value, the analog voltage value should be convert to ppm value. Refer to Function 1. ppm = V Analog Output 0.0004 Function 1. Voltage to ppm 4 Voltage vs. ppm value 3 2 1 0 0 2500 5000 7500 10000 CO2 Concentration (ppm) Graph 1. I 2 C Write Packet In case of analog output, the accuracy of ppm may be different when compared with other interface. 11

Check LED This sensor has one led on the bottom. User can check the state of sensor easily. The led will be light up in mentioned situations below. Refer to the conditions. - Conditions when the green led is light up. 1 When sensor responses after receive any data trough UART, the led will be twinkled. 2 When manual calibration is running, the led will be light up until finish the calibration. As see the Dimension 1, user can know the location of green led. Green LED Dimension 1. Green LED Position 12

ABC (Automatic Calibration) This sensor was calibrated in a factory. But, it will be recalibrate monthly to get the accuracy of ppm. It mean if the ABC is enabled, sensor will have these operation. If user wants to disable the ABC, use a hardware pin. As use a hardware pin, the ABC can be disabled easily. Refer to instruction at Table 13 and Figure 4. If UART or I 2 C is used, the period can be changed. In the initial state, the period is one month. It will be selected as one week or one month. How to Enable and Disable the ABC function Table 13. Hardware State vs ABC State Hardware State Open Short ABC is Enabled ABC is Disabled Figure 4. Hardware Pin Location To change the period of ABC, use a UART or I 2 C at Page 6 (UART), Page 10 (I 2 C), there are instructions that how to use. Table 14. ABC Period Period 7 days 30 days (Default) 13

Manual Calibration If user wants recalibrated data manually, this sensor can be calibrated as use a manual calibration function. To use manual calibration, Please, Read all comments and cautions very carefully. CAUTIONS Please, Make sure that air conditions should be stabilized. It means that the CO2 density and temperature should be stabled in vicinity of sensor. If manual calibration is run with unstable air conditions, calibration run time will be increased. Furthermore, the sensor data after finishing such manual calibration may be incorrect. Manual Calibration Sequence 1 Request a manual calibration command using Hardware Pin (Page 14) or UART (Page 7) or I 2 C (Page 10) interface. 2 When sensor receives requested command, the manual calibration will be started. User can check it by watching Green LED glowing up during calibration process. Also, it can be checked through UART or I 2 C communication. CHECK LED ON Figure 5. Green LED glowing up while manual calibration. How to use Hardware Pin At the moment, when the MCAL pin is connected to GND (at least 2 seconds), this sensor will start the manual calibration with set default value (default value = 400 ppm) and it will be ended automatically after calibration time. Which is depen dent on surrounding conditions. So, in Figure 6, it is shown when the switch (SW) is pressed, the manual calibration will be started. Sensor Figure 6. Manual Calibration Application Circuit, in case of using external hardware 14

Package Dimension Dimension 2. Product Dimension Caution : Do not attempt to reassemble or give physical stress while handling. 15

Pin - TOP Table 15. Upper Pin No. Pin Assigned 1 Analog Output 2 GND 3 UART TXD 4 UART RXD 5 I 2 C SDA 6 I 2 C SCL 7 GND 8 V DD Table 16. Lower Pin No. Pin Assigned Figure 7. Assigned Pin on Top 9 GND 10 GND 11 GND 16

Pin - BOTTOM Table 17. Upper Connector-1 Table 18. Upper Connector-2 No. Pin Assigned No. Pin Assigned 1 I 2 C SDA 2 I 2 C SCL 3 GND 4 V DD 5 UART RXD 6 UART TXD 7 GND 8 Analog Output Figure 8. 4x2 Pin Header, 1.27mm SMD Type Figure 9. Assigned Pin on Bottom Table 19. Lower Connector-1 Table 20. Lower Connector-2 No. Pin Assigned No. Pin Assigned 9 GND 10 *MCAL Pin 11 Reserved 12 NC *MCAL : Manual Calibration 13 Reserved 14 *ABC Pin 15 Reserved 16 NC *ABC : Manual Calibration Figure 10. 3x2 Pin Header, 1.27mm SMD Type Caution : Reserved pins should not be connected to anything. 17

Ordering Guide Table 21. Ordering Guide Model Supply Voltage ( V DD ) Operating Range ( ) T OL T OH TG100 NDIR Type CO2 Gas Sensor Module 5 V 0 50 TG100-R NDIR Type CO2 Gas Sensor Module 5 V -20 50 TG100-R2 NDIR Type CO2 Gas Sensor Module 5 V -25 50 TG100-R3 NDIR Type CO2 Gas Sensor Module 5 V -35 50 * The R denotes what product can operate in temperatures below zero like a refrigeration system. 18

Revision History Table 22. Revision History Revision Date Author 2.3 Last Version - - 2.4 2.5 The previous version is destroyed because modified template and major parameters. Major changed contents : - Electrical Spec. was changed. - Details about interface, dimensions were added The pitch value at the dimension section was modified. - Changed : 2.5±0.2 2.54 PITCH Specific define for resolution ( @0ppm) 5 Jan 2018 JH Yi 8 Jan 2018 JH Yi 2.6 I2C section was modified with details. 15 Jan 2018 JH Yi 2.7 2.8 2.9 The page 2, on specifications - The range of power supply was modified. - Thus, current consumption were modified. - Location of Operating Humidity just is moved. The table 4, I2C frequency was changed according to mater I2C. On I2C sequence, the slave address was added. On Analog Output page, The equation was modified. On Page 13, Release time was added. read the contents. Dimension was changed to old version. - Add sensor gross weight at dimension drawing page - Correct operating temp -20~50-25 to 50 - Remove slave frequency description on I2C spec - Add mechanical tolerance on dimension - Dimension was changed to new version On page 15, Pin was added. On Page 17, Ordering Guide was added. 29 Jan 2018 JH Yi 31 Jan 2018 JH Choi 3 July 2018 DH Jeong 2.91 - On Page 3, Correct Request Time unit ( ms sec ). 12 July 2018 DH Jeong 2.92 - On Page 6, ACAL Mode example was modified. - The Table 11, Set a period of automatic calibration command is removed. - On page 9, Correct ACAL Period unit ( 0x30 (30days) 0x1E (30days) ). - On page 10, I 2 C automatic calibration packet is removed. 1 August 2018 DH Jeong 2.93 - On Page 15, Bottom dimension was added. 3 August 2018 DH Jeong 19

This document contains information of products and processes in development. All information provided here is subject to change without notice. Information in this document supersedes and replaces information previously supplied in any prior versions of this document. We do not guarantee the performance of this device in case of disassembling, operating without complying with instructions in this document. Contact your TRUEYES representative to obtain the latest forecast, schedule and specifications. Or It may be obtained by calling +82-2-2083-2377~8 or by visiting http://www.trueyes.co.kr Copyright 2018 TRUEYES Incorporated. All rights reserved.