VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400

Similar documents
VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 AND VARIAN VNMRS 500

VNMRJ 4.2 INSTRUCTIONS: QANUC 500 FOR CHEMISTS

NMR Spectroscopy with VnmrJ. University of Toronto, Department of Chemistry

NMR Training VNMRJ 4.2A/VNMRS 400

Agilent VnmrJ 3.2. Quick Start Guide

Brief IconNMR and Topspin 3.5 User Guide for Bruker NMR Spectrometers Avance IIIHD 400MHz NMR with Autosampler in Chemistry room 93 DISCLAIMER

Vanderbilt Small Molecule NMR Center

General Information for Automated Inova 500 MHz NMR (Indy)

NMR Spectrometer Crib-Sheet

c. Saturday and Sunday: 10 minute time blocks and unlimited time.

Bruker Avance 400 MHz Instructions

North Carolina State University Department of Chemistry Varian NMR Training Manual

Varian Solution NMR Procedure

Short instructions for the basic operation of the AVANCE II 400 MHz Bruker NMR Using ICON-NMR (PS751)

TopSpin QuickStart. Default State TopSpin should be running at all times. If this is not the case, you may need to Log In and Start TopSpin.

NMR INSTRUMENT INSTRUCTIONS: Safety and Sample Preparation

Acquiring Data in VnmrJ 4.2A

1D NMR Training Manual for Operating the Bruker 700 MHz Avance III HD NMR. Spectrometer without the Sample Changer and IconNMR

DCIF NMR Training Guide 400 MHz Bruker DRX B401 Bruker DPX B400

DCIF NMR Training Guide 401 MHz Bruker AVANCE III HD b401 last edit 3/11/2015

Routine 1 H and 13 C NMR Data Acquisition Basic Topspin Processing Guide For PSC and NSC Bruker 400 NMR Spectrometer

Processing data with Bruker TopSpin

DCIF NMR Training Guide 400 MHz Bruker AVANCE III HD b400 last edit 3/9/2015

Bruker NMR Topspin Manual

Varian Solution NMR Automation Procedure (VNMRS Machines running VNMRJ 4.2 under Red Hat Enterprise 5.1)

Varian Solution NMR Automation Procedure

PROCESSING 2D SPECTRA USING VNMRJ JB Stothers NMR Facility Materials Science Addition 0216 Department of Chemistry Western University

VnmrJ Walkup. Varian NMR Spectrometer Systems With VnmrJ 1.1D Pub. No , Rev. 0604

Quick Guide to Topspin

Bruker AVANCE III HD 400 MHz Spectrometer Operation

VnmrJ Walkup. Varian, Inc. Inova and MercuryPlus NMR Systems With VnmrJ 2.2D Software Pub. No , Rev. A 0808

Basic 1D Processing. Opening Saved Data. Start the VnmrJ software. Click File > Open.

Delta V5 operation. Widetron Technologies Corp. Wu, Chung Ying. Progress through Synergy

Bruker Topspin NMR Training Manual

Diffusion Ordered Spectroscopy (DOSY) Overview BRUKER last edit 5/14/12

NMR Users Guide Organic Chemistry Laboratory

MNova Version Installation Instructions.

UC DAVIS NMR FACILITY VNMRJ SHORT MANUAL FOR VARIAN/AGILENT NMR SPECTROMETERS

EQNMR Facility Rules of Use

IMSERC NMR MANUAL 02: Basic Processing of Varian 1D NMR Data

PINMRF. Bruker AV-III / Avance DRX NMR Spectrometers running TopSpin Training Supplement for Advanced 1D NMR Spectroscopy

PINMRF. Varian 300 MHz NMR Spectrometers Training Guide for Basic 1D NMR Spectroscopy

Installing TopSpin 4

Welcome to the NMR Center

Instructions for 1 H-, 13 C-, 19 F-, and 31 P-Spectra on the Varian Mercury-Vx-300

NMR Data Acquisition and Processing Procedure

Gemini-300, Manual Operation (version 1.0, 4Feb95) 1

T 1 Relaxation Measurement: The Inversion-Recovery Experiment (Using IconNMR) Revised

Walkup NMR. Varian NMR Spectrometer Systems With VNMR 6.1C Software. Pub. No , Rev. A0800

Open Access Manual for Superusers

Using OPUS to Process Evolved Gas Data (8/12/15 edits highlighted)

Basic Procedure 1D NMR Data Collection

Troubleshooting Acquisition Related Problems

ONE DIMENSIONAL NMR SPECTROSCOPY. DPX 200 and DPX 300

Frequently Asked Questions (FAQ)

Running Multiple Experiments on the Same Sample Automatically. (multizg, spooler/queue, iconnmr)

UC DAVIS NMR FACILITY BRUKER SHORT MANUAL FOR TOPSPIN 3.X

VNMRJ 2.1B Quick overview and BioPack features

Fitting NMR peaks for N,N DMA

Vnmrj 3.1 New and improved features

1. Log in at the RESLOG terminal by entering your code and select the instrument you want to use.

DCIF NMR Training Guide Varian Mercury 300 MHz Last Update

Training Manual For Chemistry NMR Facility Spectrometers

User Manual V

67 Series Spectrophotometer PC Software

CHEMISTRY DEPARTMENT Category Work Instruction

Agilent VnmrJ 3.2 Automation

Varian NMR Instructions - 1D. Department of Chemistry NMR Facility University of Minnesota

TOPSPINPLOT(XWINPLOT)

MestReC Cheat Sheet. by Monika Ivancic, July 1 st 2005

IconNMR Operating Procedures

500 MHz Solution-state NMR Procedure. (Bruker AVANCE Machines running TopSpin under WINDOWS XP)

Quick Start Guide to nmrpipe Processing of 15 N-HSQC data

500 MHz Solution-state NMR Procedure. (Bruker AVANCE Machines running TopSpin under WINDOWS XP)

User Manual.

DL150 Software Instructions

TopSpin is the successor to a program called XWinNMR, which appeared in the mid- 90s (I believe) and was used to run Bruker spectrometers until it

TopSpin 1.3 patch level 10. DPX 200, DPX 300 and DRX 500. Windows XP

DEPT on the U400. 2A. Enter the Macro In the experiment that contains the 13 C spectrum, enter: macro to set up a DEPT experiment in exp2

CORPORATE ACTIVITY REPORT (CAR) MANUAL. Rev 3/13

TopSpin. Multiplet Analysis Tutorial

XPS1 Automated Multi-Sample Run Procedure

NMR Spectroscopy with VNMRJ & EZ NMR

Manual for Jasco-IR-spectrometer

VnmrJ 2.2B Installation and Administration. Varian NMR Systems With VnmrJ 2.2B Software Pub. No , Rev. B 0107

NMR Experiment Procedure

WebPrint Quick Start User Guide

Updating Your Local Program Webpage

ChromQuest 5.0 Quick Reference Guide

Before you get started, make sure you have your section code since you ll need it to enroll. You can get it from your instructor.

Forms/Distribution Acrobat X Professional. Using the Forms Wizard

NMR Experiment Procedure

PASS Sample Size Software

KJM Proton T1 Spectra on the AVI-600 and AVII-600. Version 1.0

3D Surface Plots with Groups

Operating instructions - Cary 100 Bio UV-visible Spectrophotometer

RemoteMRC. Introduction Connecting to a Workstation Connecting to a Workstation Terminal Troubleshooting... 9

Mnova Training Basics

NMR FACILITY DEPARTMENT OF CHEMISTRY University of Washington XWINNMR 3.5 TRAINING NOTES. 08/8/2005 rajan paranji

Netzsch TGA Training Notebook

Transcription:

VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 January 10, 2019 1. Sample preparation a. Tubes of any length can be used b. Make your samples 4 cm deep. They will not shim as well if they are shorter; you will waste deuterated solvent if they are longer c. Make sure that the sample is clear: filter it if necessary d. Use a pen to label the cap or the top of your NMR tube if needed (do not use labels and do not write below the top few cm of the NMR tube this is to avoid ink getting into the inside of the spinner and from there getting into the NMR instrument) 2. The Varian Mercury 400 instrument can only run 1H, 19F, and 13 C. The spectrometer is reasonable for all three nuclei. 1D and 2D experiments are possible. This instrument is the best at McGill for 13C with 19F decoupling and for 13C-19F HSQC these do require some setup, so read the experiment guide. This instrument is fast: shimming is quick and there is no tuning. Better for 1H: Bruker 400, Bruker 500, Varian 500, QANUC 500 Better for 19F: Bruker 400, Bruker 500 Better for 2D: Bruker 400, Bruker 500, QANUC 500 Better for 13C: Bruker 400, Bruker 500 3. Experiment preparation a. Log in to the walkup NMR software

Figure 1: VnmrJ walkup interface login screen. Enter your username and password to log in. b. Inserting sample i. Place sample into spinner (if no spinner is available, then eject the current sample, step v., before continuing with step i) ii. Check that the depth gauge is adjusted to 67 mm iii. Adjust sample position with depth gauge (the sample should touch the bottom of depth gauge, unless the sample is very short, in which case it should be raised so that it is centered on the box marked at the back of the sample changer) iv. Wipe NMR tube with Kimwipe without affecting its position in the spinner v. Click the button marked eject or type e vi. Listen for the jet of air. When you hear it, remove the dummy or previous sample and place your sample on the jet of air Figure 2: Sample adjusted properly in the depth gauge. vii. Click the button marked insert or type i 4. Running experiments a. In the middle of the screen, enter a sample name (usually of the form <initials>_<lab notebook number>; for example, RSS_0216). Only use letters, numbers, hyphens, and underscores as your sample name. Do not use spaces or any special characters such as %, #,!, or anything else. Most special characters will prevent email from working and many will prevent the file from being saved properly.

b. If you want, enter a description of your sample in the Comments box. Anything entered here will appear in Mnova or other NMR processing program. c. The experiment PROTON is already in the queue. If you would like to run any additional experiment, choose one or more experiments from the list on the upper left of the screen, under the McGill tab. See the end of this handout for a description of the experiments available. Figure 3: Submitting experiments to the sample queue. At a minimum, choose an experiment, enter the sample name, choose the solvent, choose a sample position, and push submit. d. To remove an experiment from the list (for example, if you change your mind about running it), click on it and drag it to the trash can below the list e. To modify an experiment (for example, to add more scans), double-click on it in the list. Then change one or more of the options listed on the right. If you do not see the option you want listed, click on Acquisition (under the Acquire tab, just to the left of the options shown), or on other panels under Acquire. After customizing your experiment, click on Save (or click on Quit-No save to cancel any changes you have made). Use the Show time button to see how long an experiment will take.

Figure 4: Customizing an experiment. If the parameter you want to change is not displayed, click on "Acquisition" on the left menubar, under Acquire. f. Click on Submit to submit experiments to the queue, or click on Cancel (above the study queue list of experiments to submit) to cancel submitting experiments. Figure 5: Start panel showing location of sample name and email address fields. 5. Processing data at the spectrometer a. If your data is listed in the Study Queue on the left of the screen, you can doubleclick on it to open and auto process it.

Figure 6: Experiments that have run already (green) and can be viewed by double-clicking on them. Experiments in blue (currently running) can be viewed if the number of completed scans (ct) is greater than the number of scans in the block scan (bs). b. You can also go to the File menu and choose Open. Your data will be in /home/walkup/vnmrsys/data/<username>. c. The basic processing command for 1D data is wft; for 2D data, it is wft2da. (All typed commands are entered in the command bar at the top of the screen.) Or, you can go to the Process tab and click on the Auto process button. If you try to process a 2D dataset and you get an error message about strflt, go to More 2D (under Process) and deselect F1 under Linear Prediction. d. For phasing, type aph or click on Autophase in the Process tab. For manual phasing, click on the Phase mode icon to the right of the spectrum. Left click on the right side of the spectrum, then click and drag with the left mouse button (coarse control) or the right mouse button (fine control) for 0 order phasing (which affects all peaks across the spectrum). Make sure that the peaks at the right side of the spectrum are phased correctly. Then, left click on the right side of the spectrum. Click and drag with the left mouse button (coarse control) or the right mouse button (fine control) for 1 st order phasing (which affects the peaks on the left more than on the right). Make sure that the peaks on the left side of the spectrum are phased correctly. Repeat 0 and the 1 st order phasing as many times as needed. Then click on the Redraw icon to quit the phasing mode. e. To adjust the vertical scale, type vsadj or use the mouse wheel. If integrals are displayed, then this command will adjust the integrals, so remove the integrals by clicking

twice on the Integrals icon to the right of the spectrum. To move the baseline of the spectrum, try typing vp=10 or click and drag the far left side of the spectrum up by treating the mouse wheel as a button. f. To peak pick, go to the Process / Default panel and click on Find Peaks. Then click on Peak Threshold to show the threshold. Use the left mouse button to position the threshold appropriately (click or click and drag). g. To integrate, first decide whether a baseline correction is necessary. If so, type the command dc. Then, click on the Process / Integration panel and click on the Partial radio button. Click on Clear Integrals and then on Interactive Resets. Left click to start and to stop every integral (ie left click on the left and left click on the right of each peak) and then click on Scale display to fit. To normalize the integrals, place the cursor on an integral with known proton number, enter that number in Integral Area, and click on Set integral value (below it). Click on Show integral values to see the numbers. 6. Plotting and generating PDF of data at the spectrometer a. Zoom in on the region you are interested in and adjust the vertical scale. b. Go to the Process / Plot panel. Select the items you want to plot (eg Integral lines (Partial), Integral values (Horiz), Peak labels (ppm, top in the drop-down menu), Axis (ppm), etc. c. To generate a PDF for saving or emailing, click on Preview, then Cancel Acrobat Reader will open in the back. d. To print, click on Plot (which will print). 7. Processing data offline Your data will be emailed to you as a ZIP file. Once you unzip it, you can drag and drop it into Mnova. Or, you can use the vconv command in TopSpin to open it there. See the TopSpin notes for more information about reading Varian data in TopSpin (1D data is straightforward, but some 2D data is problematic.) 8. Running experiments manually a. Locking Click on Auto lock (Start panel) to start the Find z0 routine to look for the largest 2H signal and lock on it b. Shimming

Click on Gradient shim (Start panel) to start the gradient shimming routine Figure 7: Auto lock and gradient shim buttons on the Start panel. Note: if shims are not good enough, go to the Shim page of the Start panel and click on Read default shims, then rerun Gradient shim. At this point, you can also adjust the X, Y, XZ, and YZ shims by right- or left-clicking on them so as to maximize the lock level. c. Choosing experiment Click on the experiment in the Experiment selector on the McGill tab (or on another tab, as long as the experiment uses only 1H and/or 31P) that you would like to run. Modify any parameters you like under the Acquire tab. You might like to set bs (block scans), the number of scans after which data is written to disk and you can start processing, to a small number (do this by typing, for example, bs=2 in the command entry line at the top of the screen. You can also set up kinetics experiments at this point. See section 11 for more information about all experiments, including kinetics experiments. d. Run the experiment Use the command go (to acquire and process) or ga (to acquire without processing) in the command bar or the Go button under the Acquire tab to start acquisition. (go is particularly recommended for 2D experiments, because it is easier to process them during acquisition than when ga is used.) If you need to stop the acquisition, type aa or click Stop. 9. Saving and emailing data manually If these steps are followed exactly, your data will be saved in the correct location (ie one where you can find it!) and you will receive an email of your data. a. Enter a sample name in the Sample Information field (Start panel) b. Choose File / Autosave (this will ensure that your data is saved to /home/walkup/vnmrsys/data/<username> and not to some other directory) c. Enter your email address in the Email field d. Type the command efid at the top of the screen to have the data emailed to you

10. Logging out of walkup interface to VnmrJ Click on Logout (right-middle of the screen, with Start selected) to log out of VnmrJ. Note: do not close the VnmrJ program or log out of Linux. After you log out, you should see the login screen (Figure 2) displayed. Figure 8: Log out of VnmrJ by selecting the "Logout" button. Do not quit VnmrJ and do not log out of Linux.