Operation Manual MODEL 3173/3173R. MICROCOMPUTER BASED Conductivity/Salinity/TDS &Temperature BENCH METER 3173/3173R

Similar documents
Operation Manual. Hand-held Conductivity/Salinity/TDS/ Temperature Meter 3010M

JENCO ELECTRONICS, LTD. MANUFACTURER OF PRECISION INSTRUMENTS

JENCO ELECTRONICS, LTD. MANUFACTURER OF PRECISION INSTRUMENTS

Operation Manual. MODEL ph/ec80. Microcomputer Based ph/condctivity/tds/ Sanility & Temperature Pocket Meter. ph/ec80

Operation Manual. MODEL ph6175. Microcomputer Based ph/orp/temperature Benchtop Meter

Operation Manual. Hand-held ph/orp/temperature Meter 6010M

Operation Manual. Hand-held ph/mv/temperature Meter. ph6810

Operation Manual. MODEL ph 630. Microcomputer Based ph&temperature Pocket Meter. ph630

Operation Manual CX10 1/8 DIN Microcomputer Based Conductivity/Resistivity Controller

Operation Manual CX20 1/8 DIN Microcomputer Based Conductivity/Resistivity Controller

Operation Manual MODEL 3621 JENCO ELECTRONICS, LTD. Microcomputer Based ph/orp Controller MANUFACTURER OF PRECISION INSTRUMENTS

Operation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter

Operation Manual TX10 1/8 DIN Microcomputer Based ph/orp Controller

JENCO ELECTRONICS, LTD. MANUFACTURER OF PRECISION INSTRUMENTS

JENCO ELECTRONICS,LTD. MANUFACTURER OF PRECISION INSTRUMENTS

Operation Manual MODEL 3631 JENCO ELECTRONICS, LTD. Microcomputer Based ph/orp Transmitter MANUFACTURER OF PRECISION INSTRUMENTS

User Manual DOCUMENT #601119REF. EcoSense ph1000a. ph, mv (ORP), and Temperature Benchtop Meter USER MANUAL. English

Datalogging Conductivity/ TDS Meter. Instruction Manual

OPERATION MANUAL MODEL TT-LP

Cole-Parmer Instrument Company

User Manual DOCUMENT #606042REF. EcoSense EC300A EcoSense EC300M Portable Conductivity, Salinity and Temperature Instruments. USER MANUAL English

JENCO INSTRUMENTS, INC. JENCO ELECTRONICS, LTD.

Cole-Parmer Instrument Company

OPERATION MANUAL. JENCO MODEL 6312 CTB MICROCOMPUTER BASED Conductivity/TDS & Temperature TRANSMITTER & CONTROLLER

DDS-12DW Benchtop Conductivity Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD

WalkLAB HC Professional Conductivity-TDS-Salinity Meter. Operations Manual 15,0,11,4,1,14,5,10,13,2,9,6,3,12,7,8

Operations Manual EcoSense ph100

Operations Manual EcoSense ph100a

USER MANUAL AD331 AD332 Waterproof Portable Meters for EC and TDS Measurements

TDS/SALT/TEMP TRACERTM POCKETESTER CODE 1749-KIT. Pool Professional s Meter

Professional Benchtop Conductivity meter BC3020

Oyster Meter ph-conductivity-tds-salinity-orp(mv)

Dissolved Oxygen Meter Kit DOK. Instruction Manual

DO Meter Kit. Instruction Manual

INSTRUCTION MANUAL CONDUCTIVITY-METER MODEL CC-01

Instruction Manual Temp JKT Temperature Meter Series

Water Quality Meter. Water Quality Meter Kit. Instruction Manual

Large Display Salinity & Conductivity Pens &

Global Water Instrumentation, Inc.

Operation Manual. MODEL ph 618N. Microcomputer Based ph & Temperature Pocket Meter. ph618n

Light Meter with Bar Graph

USER MANUAL AD 3000 EC/TDS/Temperature Bench Meter

Portable Conductivity Meter M1410. User Manual 1.002

A150 Benchtop Conductivity Meter Instruction Manual

24-hour Continuous Field Monitors

ph/orp 300 (Cat. No ) MICROCOMPUTER BASED OPERATION MANUAL ph /ORP/ Temperature CONTROLLER Phone: , Fax:

EC/TDS/SAL TRACERTM POCKETESTER CODE Conductivity/TDS/Salinity/Temperature/Meter

Benchtop Multi-parameter Water Quality Meter

Instruction Manual HI 9033 HI Waterproof Multi-Range EC and TDS Meters for Field Applications.

ELECTROCHEMISTRY. BOECO ELECTRODES The electrodes are not included with the ph meters and have to be ordered separately:

CONDUCTIVITY METER OPERATION MANUAL. Model : CD-4322

WalkLAB ph mv/orptemperature

C100 Handheld Cond./TDS/Sal. Meter. Instruction Manual

ph, ORP, and TDS ORP Continuous Monitor OPERATING INSTRUCTIONS 8. Warranty 9. Return of Goods

Instruction Manual. αlpha TDS /8 DIN Total Dissolved Solids Controller with Temperature display and Transmitter. 68X rev 0 08/2002

AquaShock ph Meter K

The Economy Conductivity Benchtop Meter

HI HI N HI HI

Tracer TM Salt/EC/TDS PockeTester Code 1749

MULTI- PARAMETER TESTER

Colorado Series. CR-10 Portable ph / mv / Temperature Meter Operation Manual

Introduction: Features:

Manual No Revision MODEL GLI-9835 HAND-HELD/PORTABLE CONDUCTIVITY/TDS METER

User's Manual. Model Heavy Duty Datalogging Module with Windows Software

MODEL 470 PORTABLE CONDUCTIVITY/TDS METER OPERATING MANUAL CONTENTS

Thermo Scientific Orion Star A222 and Star A322 Portable Conductivity Meter

Research Grade Meters. ph, ISE, EC and DO. Laboratory bench meters with a color, graphic display and capacitive touch keypad

HI HI N HI HI

User Guide. ExStik TM Model RE300 Waterproof ORP Meter. Patent Pending RE300 ORP

HI 2300 Auto-ranging Microprocessor EC/TDS/NaCl/ºC Bench Meter

and Multiparameter Meters

Datalogging ph Meter

Tracer ph/tds/salt PockeTester TM Code 1766

Instruction Manual Cond 6+ Conductivity/Temp TDS 6+ TDS/Temp Salt 6+ Salinity/Temp

Operating Instructions

AquaController Apex Probe Module 2

Visible Light SD Card Datalogger

WK7200. Water Quality Test Kit USER S MANUAL

RH/Temperature SD Card Datalogger. Instruction Manual

INSTRUCTION MANUAL. Portable ph and ph/ise Meters. Models 210Aplus 230Aplus 250Aplus 290Aplus

Operating Instructions

Datalogging Hot Wire CFM Anemometer

ph, mv and Temperature Instrument

WEB SITE:

Instruction Manual CON 6/TDS 6. Hand-held Conductivity/TDS Meter 68X Technology Made Easy... Rev.2 01/04

OPERATION MANUAL. ph 500 (Cat. No ) MICROCOMPUTER BASED ph / Temperature TRANSMITTER & CONTROLLER

Instruction Manual CON 400 and CON 410 Waterproof Hand-held Conductivity/TDS Meter

POCKET DIGITAL REFRACTOMETERS ~

User's Guide. ExStik EC400. Conductivity / TDS / Salinity / Temperature Meter MODE HOLD

Waterproof Construction. Notepad Function. Automatic Datalogging. Automatic Calibration. Good Laboratory Practices. Serial Interface Port

Handheld Water Quality Meters 100 Series

Resolution selectable from 0.01 and ph. Range ph. Accuracy ± ph

USER GUIDE. Salinity Meter Pen Style Water Quality Meter. Model EC170

Torque Meter. Instruction Manual

Vacuum Meter Instruction Manual

Anemometer SD Card Datalogger. Instruction Manual

Instruction Manual. Laboratory EC Meter NeoMet Series. EC-400L (EC/TDS/ Salinity/Temp)

Instruction Manual COND 600/610 Waterproof Handheld Conductivity Meter

CODUCTIVITY METER OPERATION MANUAL. Model : CD-4303

Dissolved Oxygen Meter

Transcription:

Operation Manual MODEL 3173/3173R MICROCOMPUTER BASED Conductivity/Salinity/TDS &Temperature BENCH METER 3173/3173R

INITIAL INSPECTION..... 2 GENERAL INTRODUCTION.. 2 USING THE METER. 3 THE DISPLAY.... 3 THE KEYPAD.... 4 CONNECTORS.. 4 REPLACING THE BATTERIES..... 5 TURNING ON / OFF THE INSTRUMENT 5 MODES OF THE METER 6 CALIBRATION SET-UP... 7 BEFORE CALIBRATE. 7 CALIBRATION PROCESS.. 7 MEASUREMENTS 9 RS-232C INTERFACE OPERATION (For 3173R only).....10 INTRODUCTION....10 PREPARATION... 10 DB-9 CONNECTOR CABLE.. 11 SOFTWARE...... 11 PREPARING & STORAGE STANDARD SOLUTIONS... 11 ERROR DISPLAYS.. 13 SPECIFICATIONS...14 WARRANTY....15 RETURN OF ITEMS......16 3173/3173R 1

INITIAL INSPECTION Carefully unpack the instrument and accessories. Inspect for damages made in shipment. If any damage is found, notify your Jenco representative immediately. All packing materials should be saved until satisfactory operation is confirmed. GENERAL INTRODUCTION The Jenco Model 3173/3173R Bench System is a microprocessor based instrument designed for use in field, laboratories and process control applications. Using a two-wire electrode cell, it is user-friendly for the precise measurements of conductivity, salinity, TDS (Total Dissolved Solids) and temperature. The model 3173/3173R micro-processor allows the user to easily recalibrate the parameters for the probe. A few keystrokes will adjust all the parameters for compensated and uncompensated conductivity and will also give the user the option to select three types (0.1, 1.0, 10) of probe cell constant for a better selection of available probes and applications. And the user can input K value of the cell by keyboard directly. The system simultaneously displays temperature in along with either Compensated Conductivity, Uncompensated Conductivity, Salinity or TDS. The user can switch back and forth from all these displays by just pushing a single [MODE] key. This instrument is powered by six AAA-size alkaline batteries or with a UL approved AC adapter (OUTPUT: DC 9V). The instrument also displays a "LO BAT" message when the batteries are in need of replacement. Optional RS-232C can trace all displayed values and interface the meter with recorders, printers, or computers, etc. This is only for 3173R. Other features include large LCD display, touch keys with audio feedback, percentage of power supply display, auto-ranging conductivity, auto-saving calibration values and high 50/60Hz AC noise rejection. 2

USING THE METER THE DISPLAY FIGURE 1 1.BAT. Low battery indicator 2.CAL Calibration mode indicator 3.AUTO RANGING Auto ranging indicator 4.10.00 Main display 5. ms us Unit indicator 6. UNIT Conductivity units 7.Conductivity/Salinity/Temperature Parameters of this meter 8. Temperature unit 9.25.0 Secondary display 10.ppt TDS unit or salinity unit 11.ppm TDS unit 12.% Percentage unit 13.CELL K= K value 14.UNCOMPENSATED Uncompensated indicator 15.SALINITY ppt Salinity and it s unit 3

THE KEYPAD FIGURE 2 1. The [MODE] key. In normal operation this key will change the display to Conductivity, compensated Conductivity, Salinity and TDS displays. In calibration mode this key will exit the current parameter calibration and go to the next calibration parameter. 2. The [CAL] key. During normal operation this key will change the mode from normal to calibration mode. See section CALIBRATION SET-UP. 3. The [ ] key. This key is used during calibration to increment the value / choose parameters of the display. 4. The [ ] key. This key is used during calibration to decrement the value / choose parameters of the display. 5. The [ENTER] key. During calibration set-up, this key will save the current parameter. 6. The [ON/OFF] key. This key will turn on or turn off the instrument. The last display mode will be saved. CONNECTORS FIGURE3 1. RS-232C connector (for 3173R only) 2. AC ADAPTER connector 3. PROBE connector 4. No use 4

REPLACING THE BATTERIES Follow the procedures to replace the internal batteries. 1.Replace batteries when the BAT indicator appears on the LCD. The instrument can operate within specifications for approximately one hour after BAT starts to appear. 2.The battery compartment is located at the bottom side of the meter (Refer to Figure 4). Flip the meter over so that the bottom side is facing up. Lift the battery compartment cover to expose the battery compartment. 3.Replace all six type AA batteries. 4.Replace battery compartment cover. FIGURE 4 TURNING ON / OFF THE INSTRUMENT Once the batteries are installed correctly and / or an AC adapter is plugged in the unit, it automatically turns on. If not, press the [ON/OFF] key to turn it on. It will activate all segments of display for a few seconds then it will display the current cell constant. After a little while it will display the percentage of power supply. After the self-diagnostic is completed the temperature will be displayed on the LCD and the instrument is ready to make a measurement. Immerse the probe half-way in the solution. If possible do not allow the probe to touch any solid object in the solution. There should be no air bubbles around the probe either. Shaking or moving the probe vigorously before recording any measurement will dislodge any bubble formed in the probe. When the unit is not in use, the user should turn off the instrument by pressing the [ON/OFF] key to save battery life. By just unplugging the AC adapter will not turn off the instrument if batteries are present, it would automatically switch to battery power and will continue to operate. 5

MODES OF THE METER This instrument has four modes as follow: 1.Conductivity (us or ms): A measurement of the conductive material in the solution with no regard to temperature. 2. Compensated Conductivity (us or ms): Also known as temperature compensated conductivity which automatically adjusts to a calculated value which would have been read if the sample had been at 25 ( or some other reference temperature which the user chooses). See section CALIBRATION SETUP. 3. Salinity (ppt): A measurement of the salinity in the solution, the calculation is based on the Conductivity and Temperature. 4. TDS (ppm or ppt): A measurement of the total dissolved solids (TDS), the calculation is based on the Compensated Conductivity. Temperature is always displayed in all modes. 6

To change between modes, simply press and release the [MODE] key, the unit will cycle through these four modes. When change mode, please carefully observe the annunciators on the LCD. Note : Every time the unit is turned OFF the last mode will be saved so that when turn the instrument ON again it will return to this mode. CALIBRATION SET-UP BEFORE CALIBRATE System calibration is rarely required as Model 3173/3173R were all factory pre-calibrated. However, from time to time it is wise to check the system calibration and make adjustment when necessary. When the new probe / instrument is used, CALIBRATION should be performed, since the cell constant of conductivity probes may differ from each other. CALIBRATION PROCESS 1. After the power-on diagnostics, press the [CAL] key and the CAL annunciator will appear on the display. TEMPERATURE COEFFICIENT 2. The unit will display 1.91% or a value set previously using this procedure. Use the [ ] and [ ] keys to change the value to the desired new temperature coefficient. 3. Press the [ENTER] key to save the new value of temperature coefficient or press the [MODE] key to exit (and go to the next calibration parameter). TEMPERATURE REFERENCE 4. The unit will display 25.0C or a value set previously using this procedure. Use the [ ] and [ ] keys to change the value to the desired new temperature reference. 5. Press the [ENTER] key to save the new value of 7

temperature reference or press the [MODE] key to exit (and go to the next calibration parameter). PROBE BASIC CELL CONSTANT 6. The unit will display the actual (calibrated previously or default) cell constant while the lower part of the LCD will display the current selected cell number. Using the [ ] and [ ] keys, you can now change the probe basic cell constant to that you use from the 3 available cell constants (0.1, 1 and 10). 7. Press the [ENTER] key to save the new value of probe basic cell constant or press the [MODE] key to exit (and go to the next calibration parameter). CELL CONSTANT CALBRATION 8. Immerse the probe to a known standard conductivity solution (See section PREPARING & STORAGE STANDARD SOLUTIONS) that is chosen to calibrate your probe. The unit will display the conductivity value of the standard solution. During cell constant calibration, the following parameters are over-ridden: temperature reference (fixed to 25.0 ) and temperature coefficient (fixed to 1.91%). 9. Wait for the values of temperature and conductivity to stabilize for a few seconds. If you want input K value directly, please press the [MODE] key to go to the Step 11. Using the [ ] and [ ] keys, you can now adjust the reading of the display until it matches the value of the known standard conductivity solution at 25. 10.Press the [ENTER] key to calculate and save the new value of cell constant or press the [MODE] key to exit (and go to the next calibration parameter). 8

K VALUE INPUT 11.The instrument will display the conductivity value of the standard solution with the CELL K= flashing. If the CELL CONSTANT has been calibrated by using Step 9 & 10, please press the [MODE] key to exit calibration and return to normal operation. When you press the [ ] or [ ] keys, the unit will display K value with the CELL K= staying on. Using the [ ] and [ ] keys, you can now input the K value (from 70% to 130% of the probe basic cell constant). After release the [ ] or [ ] keys for 2 seconds, the unit will display the conductivity value with the CELL K= flashing again. 12.Adjust the K value until the conductivity value displayed on the LCD matches the value of the known standard conductivity solution at 25. 13.Press the [ENTER] key to save the new K value of the cell or press the [MODE] key to exit calibration and return to normal operation. MEASUREMENTS The Model 3173/3173R is an autoranging instrument. This means that, regardless of the conductivity of the solution (within the specifications of the instrument), all you need to do to get the most accurate reading is to put the probe into the sample. After immersing the probe into the solution, the unit will automatically find the right range for the solution. During the search, the unit will display rang on the LCD. The AUTORANGING will take a few seconds after which the measurement will be displayed. Press and release the [MODE] key to show values in other parameters. 9

RS-232C INTERFACE OPERATION (For 3173R only) INTRODUCTION The RS-232C interface feature can track all displayed values and interface your Jenco meter with recorders, printers, computers, etc. For example, the user can log all data simultaneously to an interfaced IBM PC/AT compatible computer. This section assumes you are familiar with the basics of data communication, the RS-232C interface, programming with computer languages, like Delphi, VB, VC, etc. A simple program must be written in order to send command and receive data from the unit. If you do not have that knowledge, you must have a rudimentary knowledge and a copy of the more popular Windows 98 operating system or a later version of Microsoft s Windows operating system capable of using COM port. In addition you must also have an interface cable of the DB-9 type (sometimes also called a Serial port connector) and that the wiring of that connector is wired to be compatible to the RS-232C standard for RxD, Receive Data and TxD, Transmit Data and Circuit Common. To use the RS-232C interface, you must choose either to download on line from the Jenco website at http://www.jencoi.com/ a general Windows compatible program that allows you to receive data from the Jenco meter as pre-programmed by that Windows compatible program or you must custom write your own simple program to send the commands you want to the Jenco meter and to receive back the specific data sampled at the time intervals you specify in your custom written program. PREPARATION The Jenco unit communicates with a PC computer (IBM PC/AT compatibles) through a DB-9 interface connector. The DB-9 must be configured to be compatible with the RS-232C standard for TxD, RxD and Circuit Common. (Please see accompanying diagram.) After you have connected the cable, turn on both the Jenco unit and the computer, you are now ready for the software part of the operation. 10

DB-9 CONNECTOR CABLE Below is the wiring diagram showing the DB-9 connector wired to be compatible to the RS-232C standard for TxD, RxD and Circuit Common. Note: The Standard RS-232C cable is a 25-Pin connector. The essential Circuit Common and TxD and RxD are shown here adapted to the DB-9 connector. DTE DCE Data Terminal Equipment Data Communications Equipment ( Your computer or recorder, etc.) ( Your Jenco Meter.) Transmit Data! Pin 2 Receive Data? Pin 3 Circuit Common Pin 7 Pin 2 Receive Data? Pin 3 Transmit Data! Pin 7 Circuit Common SOFTWARE A demo program and the protocol are included in the accompanying CD. Using the demo, you can easily track or recall data from the Jenco meter. Please read the protocol to understand more about the procedure in detail. You may also choose to download from our website a pre-programmed Windows compatible program but you will be limited to tracking and recalling data as preset in those programs. PREPARING & STORAGE STANDARD SOLUTIONS Suitable conductivity standards are available commercially or the user can prepare them using research grade reagents. Here are some standard solutions the user can prepare to calibrate the probe of the Model 3173/3173R. 1.Standard solution of 1413 µs at 25 C : Accurately weight out 0.746 grams of research grade dried Potassium Chloride (KCl). Dissolve it in 1000ml of deionized water 2.Standard solution of 12.90 ms at 25 C : Accurately weight out 7.4365 grams of research grade dried Potassium Chloride (KCl). Dissolve it in 1000ml of deionized water 11

3.Standard solution of 111.0 ms at 25 C : Accurately weight out 74.2640 grams of research grade dried Potassium Chloride (KCl). Dissolve it in 1000ml of deionized water You can store the remaining solution in a plastic container for one week but the air space between the cap and the solution must be kept to an absolute minimum. The storage life can be prolong by storing the solution below 4 C. If you have any doubt of the accuracy of the stored solution, a fresh batch should be prepared. 12

ERROR DISPLAYS Main Display OuEr OuEr OuEr OuEr LErr OuEr OuEr OuEr OuEr LErr OuEr Undr Secondary Display -10.0 120.0 OuE/Und -10.0 120.0 OuE/Und -10.0 120.0-10.0 120.0 OuE/Und -10.0 120.0 OuE/Und -10.0 120.0 OuE/Und Mode/State Uncompensated conductivity measurement Compensated conductivity measurement Salinity measurement TDS measurement Cell constant calibration Temperature measurement Reasons Uncompensated conductivity is out of the range. Temperature is out of the -10.0 120 range. Compensated conductivity is out of the range. Temperature is out of the -10.0 120 range. Temperature exceeds the limit (compute using the current temperature coefficient and temperature reference). Salinity value is out of the range. Temperature is out of the -10.0 120 range. TDS value is out of the range. Temperature is out of the -10.0 120 range. Temperature exceeds the limit (Compute using the current temperature coefficient and temperature reference). Cell constant calibration is greater than ±30% of the selected basic cell constant. Temperature is out of the -10.0 120 range. 13

SPECIFICATIONS DISPLAY RANGE RESOLUTION ACCURACY RANGE of TDS (0.5) Conductivity 0 20.00 us/cm 0.01 us/cm ±0.5% FS 0 10.00 ppm (PROBE CELL CONSTANT = 0.1) 20.0 200.0 us/cm 0.1 us/cm ±0.5% FS 10.0 100.0 pm 200 2000 us/cm 1 us/cm ±0.5% FS 100 1000 ppm Conductivity 0 200.0 us/cm 0.1 us/cm ±0.5% FS 0 100.0 ppm (PROBE CELL CONSTANT = 1) 200 2000 us/cm 1 us/cm ±0.5% FS 100 1000 ppm Conductivity (PROBE CELL CONSTANT = 10) 2.00 20.00 ms/cm 0.01 ms/cm ±0.5% FS 1.00 10.00 ppt 0 2000 us/cm 1 us/cm ±0.5% FS 0 1000 ppm 2.00 20.00 ms/cm 0.01 ms/cm ±0.5% FS 1.00 10.00 ppt 20.0 200.0 ms/cm 0.1 ms/cm ±0.5% FS 10.0 100.0 ppt DISPLAY RANGE RESOLUTION ACCURACY Salinity 0.0 80.0 ppt 0.1 ppt ±1% FS Temperature -10.0 120.0 0.1 ±0.5 14

Temperature sensor Thermistor, 10k, at 25 Cell constant 0.1, 1.0, 10 ; user selectable Reference temperature From 15.0 to 25.0, default 25.0 ; user adjustable Temperature coefficient From 0.00 to 4.00%, default 1.91%; user adjustable K value From 70% to 130% of the probe basic cell constant TDS constant 0.5 (Fixed) LCD character size 21mm / 16mm high Calibration Back-up Yes Automatic temperature compensation Yes Audio Feedback All Touch Keys Communication RS-232C output (for 3173R only) Power Supply 6 AA batteries / AC adaptor (Output: DC 9V) Ambient temperature From 0 to 50 Relative humidity Up to 95% Weight (Batteries not included) About 650 grams Dimensions 230mm x 165 mm x 80 mm WARRANTY Jenco Instruments, Ltd. warrants this product to be free from significant deviations in material and workmanship for a period of 3 years from date of purchase. If repair or adjustment is necessary and has not been the result of abuse or misuse, within the year period, please return-freight-prepaid and the correction of the defect will be made without charge. If you purchased the item from our Jenco distributors and it is under warranty, please contact them to notify us of the situation. Jenco Service Department alone will determine if the product problem is due to deviations or customer misuse. Out-of warranty products will be repaired on a charge basis. 15

RETURN OF ITEMS Authorization must be obtained from one of our representatives before returning items for any reason. When applying for authorization, please have the model and serial number handy, including data regarding the reason for return. For your protection, items must be carefully packed to prevent damage in shipment and insured against possible damage or loss. Jenco will not be responsible for damage resulting from careless or insufficient packing. A fee will be charged on all unauthorized returns. NOTE: Jenco Instruments, Inc reserves the right to make improvements in design, construction, and appearance of our products without notice. JENCO INSTRUMENTS, INC. 7968 Arjons Drive, Suite C San Diego, CA 92126 USA TEL: 858-578-2828 FAX: 858-578-2886 E-MAIL: jencoi@ix.netcom.com; sales@jencoi.com Homepage: http://www.jencoi.com/ JENCO ELECTRONICS, LTD. PO. BOX LINKOU 117 TAIPEI, TAIWAN TEL: 02 601-6191 FAX: 02 601-7206 E-MAIL: jencoe@ms2.hinet.net Homepage: http://www.jenco.com.cn/big5/ SHANGHAI JENCO ELECTRONICS, LTD. 18 Wang Dong Zhong Road Sijing Town,Songjiang SHANGHAI,CHINA TEL: (86-021)5761-9599 FAX: (86-021)5761-9598 E-MAIL: jencos@jenco.com.cn Homepage:http://www.jenco.com.cn/ Version 03 090629 16

3173 2004/09/06 K3173-02 2009-6-29 03 04-11-05:Beacause changing the input K value, the operation is not as same as the primary. So the manual should be rewrote. Version 2. 09-6-29: EMAIL 17