Vanderbilt Small Molecule NMR Center

Similar documents
Bruker NMR Topspin Manual

DCIF NMR Training Guide 400 MHz Bruker AVANCE III HD b400 last edit 3/9/2015

DCIF NMR Training Guide 401 MHz Bruker AVANCE III HD b401 last edit 3/11/2015

Quick Guide to Topspin

Routine 1 H and 13 C NMR Data Acquisition Basic Topspin Processing Guide For PSC and NSC Bruker 400 NMR Spectrometer

TopSpin QuickStart. Default State TopSpin should be running at all times. If this is not the case, you may need to Log In and Start TopSpin.

NMR Spectrometer Crib-Sheet

Short instructions for the basic operation of the AVANCE II 400 MHz Bruker NMR Using ICON-NMR (PS751)

DCIF NMR Training Guide 400 MHz Bruker DRX B401 Bruker DPX B400

NMR Data Acquisition and Processing Procedure

VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400

Basic Procedure 1D NMR Data Collection

VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 AND VARIAN VNMRS 500

1D NMR Training Manual for Operating the Bruker 700 MHz Avance III HD NMR. Spectrometer without the Sample Changer and IconNMR

NMR FACILITY DEPARTMENT OF CHEMISTRY University of Washington XWINNMR 3.5 TRAINING NOTES. 08/8/2005 rajan paranji

TopSpin 1.3 patch level 10. DPX 200, DPX 300 and DRX 500. Windows XP

ONE DIMENSIONAL NMR SPECTROSCOPY. DPX 200 and DPX 300

Brief IconNMR and Topspin 3.5 User Guide for Bruker NMR Spectrometers Avance IIIHD 400MHz NMR with Autosampler in Chemistry room 93 DISCLAIMER

Welcome to the NMR Center

NMR Spectroscopy with VnmrJ. University of Toronto, Department of Chemistry

NMR INSTRUMENT INSTRUCTIONS: Safety and Sample Preparation

North Carolina State University Department of Chemistry Varian NMR Training Manual

PINMRF. Bruker AV-III / Avance DRX NMR Spectrometers running TopSpin Training Supplement for Advanced 1D NMR Spectroscopy

500 MHz Solution-state NMR Procedure. (Bruker AVANCE Machines running TopSpin under WINDOWS XP)

Bruker Topspin NMR Training Manual

VNMRJ 4.2 INSTRUCTIONS: QANUC 500 FOR CHEMISTS

Diffusion Ordered Spectroscopy (DOSY) Overview BRUKER last edit 5/14/12

NMR Training VNMRJ 4.2A/VNMRS 400

UC DAVIS NMR FACILITY BRUKER SHORT MANUAL FOR TOPSPIN 3.X

PINMRF. Bruker Avance / AV NMR Spectrometers running TopSpin Training Supplement for Advanced 1D NMR Spectroscopy

Bruker Avance 400 MHz Instructions

c. Saturday and Sunday: 10 minute time blocks and unlimited time.

On Artemis and Cronus, make sure you are adjusting the db scale and not the watts

Varian Solution NMR Procedure

500 MHz Solution-state NMR Procedure. (Bruker AVANCE Machines running TopSpin under WINDOWS XP)

Processing data with Bruker TopSpin

NMR FACILITY DEPARTMENT OF CHEMISTRY University of Washington TOPSPIN 2.1 TRAINING NOTES

Running Multiple Experiments on the Same Sample Automatically. (multizg, spooler/queue, iconnmr)

PINMRF. Bruker AV-III / Avance DRX NMR Spectrometers running TopSpin Training Guide for Basic 1D NMR Spectroscopy

Varian Solution NMR Automation Procedure (VNMRS Machines running VNMRJ 4.2 under Red Hat Enterprise 5.1)

Bruker AVANCE III HD 400 MHz Spectrometer Operation

Walkup NMR. Varian NMR Spectrometer Systems With VNMR 6.1C Software. Pub. No , Rev. A0800

CONTENTS 1D EXPERIMENTS D EXPERIMENTS APPENDIXES... 83

Agilent VnmrJ 3.2. Quick Start Guide

Delta V5 operation. Widetron Technologies Corp. Wu, Chung Ying. Progress through Synergy

MNova Version Installation Instructions.

Acquiring Data in VnmrJ 4.2A

PROCESSING 2D SPECTRA USING VNMRJ JB Stothers NMR Facility Materials Science Addition 0216 Department of Chemistry Western University

1D and 2D Experiments Step-by-Step Tutorial

NMR Users Guide Organic Chemistry Laboratory

See XwinNMR Acquisition Manual, Sections 1.4, 1.5 (see list of acquistion parameter descriptions starting , pg. A-39), 1.6

Varian Solution NMR Automation Procedure

2 Dimensional NMR User s Brief Guide

TopSpin is the successor to a program called XWinNMR, which appeared in the mid- 90s (I believe) and was used to run Bruker spectrometers until it

General Information for Automated Inova 500 MHz NMR (Indy)

T 1 Relaxation Measurement: The Inversion-Recovery Experiment (Using IconNMR) Revised

MestReC Cheat Sheet. by Monika Ivancic, July 1 st 2005

Fitting NMR peaks for N,N DMA

VNMRJ 2.1B Quick overview and BioPack features

Using OPUS to Process Evolved Gas Data (8/12/15 edits highlighted)

TOCSY 15. Goto. Introduction Pulse Sequence Diagram AVANCE User s Guide Bruker 167

FOCUS. Image Based Automatic Shimming Using B 0 Gradients. Installation and Users Guide Version 0.9

Introduction. Prerequisites. Ballbar - QC20-W - Analysis LAST UPDATED: 12/01/2018

IconNMR Operating Procedures

Ballbar QC20-W - Analysis

15. HOMONUCLEAR COSY

Bruker OminiFlex MALDI-TOF Mass Spectrometer Operation Quick Start

PINMRF. Varian 300 MHz NMR Spectrometers Training Guide for Basic 1D NMR Spectroscopy

KJM NMR processing everywhere with Windows Remote Desktop and TopSpin 3.5 at nmr-server.uio.no. Windows Server Version 1.

Determination of Drag Coefficient

Gemini-300, Manual Operation (version 1.0, 4Feb95) 1

Operating Procedure for Horiba Raman Microscope

CH142 Spring Spectrophotometers with Vernier Data Acquisition Software

Troubleshooting Acquisition Related Problems

VnmrJ Walkup. Varian NMR Spectrometer Systems With VnmrJ 1.1D Pub. No , Rev. 0604

EQNMR Facility Rules of Use

Content. Getting data organized Good data processing Plotting

KJM Proton T1 Spectra on the AVI-600 and AVII-600. Version 1.0

CHEMISTRY DEPARTMENT Category Work Instruction

67 Series Spectrophotometer PC Software

Frequently Asked Questions (FAQ)

Open Access Manual for Superusers

Instructions for 1 H-, 13 C-, 19 F-, and 31 P-Spectra on the Varian Mercury-Vx-300

VnmrJ Walkup. Varian, Inc. Inova and MercuryPlus NMR Systems With VnmrJ 2.2D Software Pub. No , Rev. A 0808

Standard Operating Procedure for the Horiba FluroMax-4

ME 365 EXPERIMENT 3 INTRODUCTION TO LABVIEW

Manual. User Reference Guide. Analysis Application (EMG) Electromyography Analysis

User Manual. Thermocouple Thermometer with Dual Type K/J Inputs and Data Logging. Model with NIST-Traceable Calibration

Varian NMR Instructions - 1D. Department of Chemistry NMR Facility University of Minnesota

TopSpin. Multiplet Analysis Tutorial

There are two ways to launch Graphical User Interface (GUI). You can either

Data Acquisition with CP-2002/2003 Micro-GC Control

Bruker AVANCE-360 User s Guide for the UWChemMRF

Getting low-resolution GC-MS data on the JEOL GCMate Introduction

Spectrometer Visible Light Spectrometer V4.4

Microsoft Windows XP. Operating System. Starting Windows XP. You might be asked to enter your username and password

Seized Drugs Operational Guidelines for the Shimadzu UV/VIS Comparative and Analytical Division

Condensed AFM operating instructions:

Quick Start Guide to nmrpipe Processing of 15 N-HSQC data

TopSolids Where Expertise meets Convenience

Transcription:

Vanderbilt Small Molecule NMR Center Instructions for Using TOPSPIN version 2.0 DRX 400: Setting up 1H, 13C (1H decoupled), and 31P NMR Experiments Getting started: Sample Preparation: Make sure your sample has a minimum volume of 0.5ml. Samples with less volume will be more difficult to shim and may require large changes in Z3 and Z4 in order to see a good lineshape. Please remember to use a NMR tube holder for transporting tubes to the NMR Vault. They are available in the Chemistry Stockroom. At the spectrometer: Log into the PC using the username and password that was assigned to you by the NMRStaff Once you re logged in, double click on the TOPSPIN icon to launch the program: Find your data: All data is located in /opt/topspin/data. Simply locate your username on the left hand side and click on the icon to view the sub directories. To select a falie, click on the file and drag into the active window on the right. If you have multiple experiments in that directory the following window should appear: 1

A. To Create a New File: There are 2 Ways To Open the Window That Creates New Files 1) type edc 2) Go under File than New 2

Creating a New File A NEW file can be created by either changing the Name OR by keeping the same Name and incrementing the EXPNO #. If the name is changed use up to a maximum of 13 characters (alphabetical or numeric). Do not use any special characters such as (!@#$% &?). Also, DO NOT include any spaces since this will create problems when the data is transferred. B. Insert Sample and Locking it Next, open up the Lock Window by either typing in lockdisp or selecting the Lock icon found on the top bar or by double clicking on the lock icon on the bottom of the window Once TOPSPIN is opened, the sample can be placed inside the magnet. 1) You might want to check the temperature (type edte) 2) Make sure the Lock, Autoshim and Spin are OFF before ejecting the sample from the magnet!! Anything on the BSMS keypad that is 3

illuminated green is on. To turn off the lock or spin, push the appropriate keypad button 3) Remove the bore cap 4) Eject sample (press lift on/off button, top left hand corner of keypad) 5) Clean Sample and Place in depth gauge (make sure gauge is set for 18 mm) 6) Lower Sample (by pressing the lift on/off button) (type) lock or click on the LOCK button When the solvent list displays, select the proper solvent by clicking on it. Computer will automatically adjust Field & auto Lock Since data files are overwritten once a new experiment is started it is very important to create a new file before a new acquisition is started!! The type of experiment can be selected by clicking on the icon across from Experiment 4

For 1H experiments select AA_PROTON.MV For 13C experiments select A_C13CPD.MV For Coupled 31P experiments select A_P31.MV For Decoupled 31P experiments select A_P31CPD.MV *Finally, Type gpro to read in the current power levels* OR the correct power levels can be recalled by clicking on the appropriate Icon: Under the AcquPars window click on the following icon: 5

C. Probe Tuning To tune the probe either type in the command atma or click on the TUNE button This will automatically start the tuning program. The computer will notify you once the program has finished and the proper tune/match values have been achieved Spin Sample (optional) Since the main probes on the high field instruments are very sensitive it is recommended not to spin the sample due to Q modulations that can be detected and often distort the baseline. D. Shimming Once the sample is spinning and you have locked on to the appropriate solvent, the next step is to shim the sample. The recommended method for sample shimming uses pulse field gradient shim routine which is fully automated. Gradient Shimming using TOPSHIM Before you start the gradient shimming program it is VERY important to return the shims to their default value. This is done by typing the following: 6

Type rsh chloshim.600 Once the default shims values are recalled, the gradient shimming program is started by typing topshim gui or click on the SHIM button Before starting the gradient shim make sure that you select Z X Y XZ YZ Z under the After section in TUNE box. This will shim the off axis shims once the on axis shims are complete. To begin shimming simply select the Start button located on the bottom. When the automatic gradient shimming is completed the Start button will again be available for selection. To determine if the shim was successful, simply note the lock level before you started shimming and see if it increases upon completion of the shimming routine. If the lock level becomes too high on the display and goes out of view, decrease lock gain until the signal is observable on the lock display. An additional shim may be required if the lock increased dramatically. It is recommended to repeat the shimming procedure two times to ensure optimal lineshape. 7

E. Acquiring a Spectrum After maximizing the lock level, take one scan and check the lineshape. Before you take a scan you want to set the receiver gain. The command to set the receiver gain is: (Type) rga or simply click on the icon Wait for the instrument to message that is has completed setting the receiver gain. Once it is done setting the gain you are ready to take a scan. The command to start an experiment is: (Type) zg or click on the icon located on the top right of the TOPSPIN window Type ns to set the number of scans or set the number of scans under the AcquPars window. The recommended number of scans is 16. 8

After setting the number of scans start the acquisition as described above. You might want ot check the shimming and parameter settings by acquiring a single scan (NS=1) experiment followed by a fourier transformation and phase correction as described below E. Processing Spectrum Fourier Transformation (Type) efp or click on the EFP button Keep in mind, that this uses the LB parameter (LB for 1H ~ 0.3 Hz, 13C ~ 2Hz). 9

use these buttons to increase the scale so the baseline becomes easier to view use the up and down buttons to bring the spectrum to the center of the screen and to the bottom use these buttons to expand or contract your spectrum use this button to go back to the full display for your spectrum 10

PHASE CORRECTION: Click on the Phase icon This should open the following window: 11

To phase the spectrum, click and hold in on the 0 button for zero order phasing and the 1 button for 1st order phasing. The R button resets the phase back to its original value while the 90, 90 and 180 buttons apply a 90, 90, and 180 degree phase automatically. Once the appropriate phase is applied click on the icon in the red box to save. Check the solvent peak and other narrow peaks to make sure the lineshape is narrow and symmetric. This requires expanding the spectrum display. EXPAND SPECTRUM: Click mouse 1 on either side of the spectrum where you would like to expand and hold down while dragging the mouse. A second cursor should appear. The expansion will finish once you release mouse 1. 12

REFERENCING Now you can set the reference. The reference is usually set to a peak who s position you are certain of such as a solvent peak or TMS. The command sref does a rough calibration of the spectrum. To set the reference click on the Calibrate button: Selecting the calibrate button will open the following window With the red cursor in the center of the peak you would like to use as a reference click with mouse 1 and the following window should appear Simply type the correct value into the box and select OK Now that you have set all the parameters, it s time to collect a spectrum with multiple scans. If you click on the 1D1H_ACQ Icon it will 13

automatically acquire, process and integrate a 1H NMR spectrum with 16 scans. BASELINE CORRECTION: Select the baseline correction icon and the following window will open: Selects the proper function ( polynomial, sine, exponential). The default is polynomial Use these buttons to compensate of the offset (A) and to make sure the fit is running through the center of the spectral baseline (B,C,D,E) Once corrections are made select this button to SAVE and RETURN 14

INTEGRATION: Click on the Integrate icon The following window should appear: Before you integrate, expand around the area of interest following the directions for expansion listed above. Click on this icon to activate the integral mode and the following window will appear (once the icon is active it will change color to green) 15

To integrate, click on one side of the peak of interest using the left mouse and hold down. A second cursor will appear. Release the mouse when the cursor area covers the peak you would like to integrate. 16

Once an integral is set, a value will appear. To calibrate the integral to a reference number, click on the right mouse between the integral limits and select Calibrate in the window that appears. A new box will appear where you can modify the value. You can also delete the current integral using the same window Use these keys to adjust the basis and slope of the integral Use these keys to define and delete integral regions Use these keys to manipulate vertical scale of integrals USE THIS KEY TO SAVE INTEGRALS Once integrals are saved, they may disappear from your active window. To make your integrals appear in your active window, right click on the mouse and select DISPLAY PROPERTIES. The following window will appear: 17

Simply click on the box for the parameter you would like to appear on your active window. PEAK PICKING: Click on the Peak Pick icon The following window should appear: 18

To pick peaks, click on the left mouse and drag. A green box should appear as in the following: Once you release the mouse the peaks should appear. Please remember, the lower the box is towards the baseline, the MORE peaks will be selected. To delete peaks, right click and select Delete All Regions 19

Use this icon to delete all peaks Use this icon to manually pick peaks USE THIS ICON TO SAVE PEAKS To make your peaks appear in your active window, right click on the mouse and select DISPLAY PROPERTIES and make sure you select the Peak Labels box. TITLE: Before printing you can set the title for the spectrum by selecting the Title icon. The following window should appear: 20

To set the title, just click in the pink box and type the desired title. There are no character or space limitations. PLOTTING: In version 1.3 of TOPSPIN, plotting is done using the Plot editor. To open up the plot editor click on the PLOT button or select File on the top left side of the window and than choose Print Once you select Print, the following box will appear: 21

Make sure Print with layout start Plot Editor [plot] is selected and a Layout is selected as well. To select a layout, click on the arrow button. Once the proper layout is chosen the Plot editor should open. Finishing Up 1) Lock = Off Spin = Off Eject Sample (make sure bore cap is removed before ejecting) Autoshim = off (if used) 2) Put standard sample back inside the magnet 3) Put the bore cap back in place!!!! 4) exit Topspin program and log out of instrument (right click on desktop) If Problems Occur Please Contact Don or Markus 22

Copying Data onto a Flash Drive 1) Open up data window by clicking on your home icon located on the Desktop. 2) All data is located in the following path:/opt/topspin/data/your_username/nmr 3) Open up data stick window by clicking on icon 4) To transfer data: Simply drag a file from your data directory to the USB Stick directory 5) Once you drag the file over the following message will appear: 23

Select Copy Here and file should be saved on USB Stick 6) Remove USB Stick properly: a. Right click on USB drive icon b. Select the unmount option c. Remove USB stick 24

Acquiring 13 C/ 31P/ 11B NMR Spectrum 1) Create a new file as described above: Under Experiment click on the icon and select the following: FOR 13C: A_C13CDP.MV A_C13DEPT135.MV A_C13DEPT90.MV 25

FOR 31P: A_P31.MV (no 1H decoupling) A_P31CPD.MV (with 1H decoupling) FOR 11B: A_B11.MV (no 1H decoupling) A_B11CPD.MV (with 1H decoupling) Remember, the Probe MUST BE tuned before you start any X nucleus NMR experiment. Type atma or click on the TUNE button Once tuning is complete, you can begin the acquisition: Set ns 128 scans = 6 mins 512= 24 min ZG or tr 2000= 1.5 hrs 20000 = 15 hrs. starts the experiment (lets you look at the data before it s finished) wait for the message checklockshift: finished then type efp (phase as needed) halt to stop acquisition (do not use stop, this will not save the FID) efp to Fourier Transform and display the final spectrum 26

27