Copra Moisture Tester. Operating Manual

Similar documents
Instant Moisture Meter for Grains, Seeds, Coffee and other small products PM-650. Operating Manual

Single Kernel Grain Moisture Tester PQ-520. (type PQ ) Operating Instructions

Dual Type Coating Thickness Tester LZ-990. Operating Manual

I. PANEL DESCRIPTION... 1

Thermo Scientific Orion Star A222 and Star A322 Portable Conductivity Meter

Multifunction Calibrator

RS Stock No Instruction Manual RS Input Data Logging Thermometer

NIDEC-SHIMPO INSTRUMENTS

Pasta Moisture Tester PB Operating Manual

Thermo Scientific Orion Star A324 Portable ph/ise Meter. Instruction Sheet

Hygro-Thermometer / Datalogger

Thermo Scientific Orion Star A325 Portable ph/conductivity Meter. Instruction Sheet

MANUAL PCE-123

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER

INSTRUCTION MANUAL PORTABLE APPLIANCE TESTER KEW6205

373 Series Coating Thickness Testers LE-373/LH-373/LZ-373. Operating Manual

ph/orp Meter SD Card real-time Datalogger

Digital Torque Screwdriver Model STC2-G

Torque Meter. Instruction Manual

NIDEC-SHIMPO INSTRUMENTS

Series FGE-XY Force Gauges Operation Manual

Moisture Content Meter Plus Dew Point, Wet Bulb, and Thermocouple Temperature functions

Video Borescope Inspection Camera

User Guide. Moisture Meter. Model MO250

12-Channel Thermocouple Datalogger

RS Stock No Instruction Manual RS Input Data Logging Thermometer

MIYAWAKI ULTRASONIC TESTER PM11 OPERATOR S MANUAL. Be sure to read this manual for safe and proper operation. Store this manual carefully after use.

RH/Temperature SD Card Datalogger. Instruction Manual

Torque Meter. Instruction Manual

Thermo Scientific Orion Star A223 and Star A323 Portable RDO/DO Meter. Instruction Sheet

Visible Light SD Card Datalogger

USER MANUAL. TL400B Fingerprint Lock with Voice Guide Feature. Version: 1.1 Date: Jul. 2017

Anemometer SD Card Datalogger. Instruction Manual

Thermo Scientific Orion Star A329 Portable ph/ise/conductivity/rdo/do Meter. Instruction Sheet

OPERATING INSTRUCTION

User s Guide. 600A True RMS AC/DC Clamp Meter. Model 38389

BENCH-TOP DIGITAL MULTIMETER INSTRUCTION MANUAL

S-14 S-14. Compact Digital Multimeter. Compact Digital Multimeter

INSTRUCTION MANUAL. Portable ph and ph/ise Meters. Models 210Aplus 230Aplus 250Aplus 290Aplus

INSTRUCTION MANUAL. Model Dual Input RTD Thermometer. Measures two temperatures simultaneously. Dual RTD probe inputs

Biometric Lock Programming & Operating Instructions. info. code. Biometric Lock Information. Setting The Master Code. Enrolling A Fingerprint

HI ma Simulator

3 Channel Datalogging Thermometer

Thermo Scientific Orion Star A214 Benchtop ph/ise Meter. Instruction Sheet

Light Meter SD Card real-time datalogger

Exposure Monitoring. Model 8560 INSPECTAIR CO 2. Meter. Operation and Service Manual , Revision B June 2006

Mk II 3DR MULTI-LINE LASER INSTRUCTION MANUAL

QUICK START USER GUIDE. Data Logger Model L452

200J Series Coating Thickness Testers LE-200J/LH-200J/LZ-200J. Operating Manual

DPL4000 Portable Low Range Dew Point Analyzer

Platinum RTD Thermometer MODEL NO

HI ma Simulator

WBGT SD Card Datalogger. Instruction Manual

User's Guide. MiniTec TM Series Model MN25 MultiMeter

Operating Instructions

Thermo Scientific Orion Star A326 Portable ph/rdo/do Meter. Instruction Sheet

HI-POT TESTER. User s Manual

User's Guide. 3 Channel Datalogging Thermometer Model SD200

AutoRanging Digital MultiMeter

Vibration Meter SD Card real-time datalogger

Operating Instructions

Electronic Alliance PORTABLE SURGE PROTECTIVE DEVICE TESTER MODEL SPD SPECIFICATION

SECURITEST USER S MANUAL TABLE OF CONTENTS SAFETY PRECAUTIONS SAFETY INFORMATION... 2 WARNINGS... 28

Model Instruction Manual. Heat Stress Meter. reedinstruments. www. com

IDEAL INDUSTRIES INC. TECHNICAL MANUAL MODEL:

Datalogging Conductivity/ TDS Meter. Instruction Manual

Amphenol PROTIMETER. Advanced Sensors. Digital Mini. Protimeter Moisture Meter. Instruction Manual

MANUAL. Handgelenksband WST 100. Handgelenksband- und Schuhtester. performance test for Electrostatic Wrist Straps and Conductive Shoes

2 in 1 LAN Tester and Multimeter Model:

USER MANUAL. Mini Multimeter with Non-Contact Voltage Detector (NCV) Model EX310

Optical Power Meter. TrendCommunications. User Guide Testing the World s Networks

Digital Quick-Response Thermometer

NOTES TABLE OF CONTENTS

Amphenol PROTIMETER. Advanced Sensors. Digital Mini. Protimeter Moisture Meter. Instruction Manual

EB300 Weighing Scale Operation Manual

User's Guide. Digital Multimeter. Model MN42

Mini Digital Multimeter

Pro II. Sound Level Meter V-9992

User's Guide. Video Borescope Inspection Camera. Model BR70

600 g x 0.02 g DIGITAL BALANCE

User Manual. Heavy Duty Differential Pressure Manometer. Model SDL720. Additional User Manual Translations available at

CM-220 True RMS AC CLAMP METER INSTRUCTION MANUAL

T0046 PROFESSIONAL LAN CABLE TESTER USER GUIDE

DIGITAL TEMPERATURE LOGGER

User s Guide. 600A AC Clamp Meter. Model 38387

Video Borescope Inspection Camera

Portable H/L Voltage Clamp Meter

200J Series Coating Thickness Testers LE-200J/LH-200J/LZ-200J. Operating Manual

PARTS CATALOGUE / TECHNICAL GUIDE Cal. S149A

Datalogging Hygro-Thermometer

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODELS:

SPRITE. Operations manual. VRM10 Digital Audio Variometer. Recorded alarm messages including undercarriage warning

Dual Moisture Meter Pin / Pinless Moisture Meter

Digital Recorder USER S MANUAL WVR-165 REC VOL.+ VOL.- Digital Recorder

4-Channel Thermometer / Datalogger

WBGT SD Card Datalogger. Instruction Manual

USER GUIDE Current Calibrator

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL:

DTT Series Digital Torque Testers User Manual

User Manual. 400Amp AC Clamp Meter + NCV. Model MA430. Additional User Manual Translations available at

Transcription:

Copra Moisture Tester HX-120 Operating Manual

Copra Moisture Tester Safety Precautions If the safety precautions for the copra moisture tester are not observed, injuries or damage to property may result. While the safety of the product has been given considerable attention, please read the precautions in the operating manual and use the instrument properly. Please observe the safety precautions. Please read the precautions noted in the operating manual. Do not use if broken. If the tester breaks or malfunctions, please consult our repair service. Meaning of warning symbols. In order to prevent damage resulting from erroneously operating the equipment, the following symbols are indicated in the operating manual and on the product. The following are their meanings. Caution This symbol for caution indicates something that should be strictly observed in order to use the product safely.

CONTENTS 1. Features... 4 2. Specifications... 5 3. Part Names... 6 4. Display... 7 5. Description of Keypad... 8 6. Directions... 9 6-1. Before Measurement... 9 6-2. Measurement Method... 10 6-3. Setting the Measurement Mode... 12 6-4. Displaying the Average Value... 13 6-5. How to make Bias Adjustment... 14 6-6. Setting the Alarm... 15 6-7. Printer Output... 16 7. Error Display... 17

1. Features Measure moisture content by simply piercing with electrodes You can measure moisture content simply by piercing the sample with the electrodes. Handheld device with large built-in display The portable handheld device features a large 76mm x 52mm display that's very easy to read. Automatic temperature compensation Compensation for the temperature of the copra is automatically performed by means of a built-in temperature sensor. Bias adjustment The moisture value can be corrected corresponding to each calibration in the range from 9.9% to +9.9%. Auto power off When inactive, the unit will automatically turn itself off after approximately 5 minutes to avoid draining the battery. Alarm setting Set any moisture value, and an alarm will sound if the measured moisture value is higher than the set moisture value. 4

2. Specifications Measurement Principle : Electrical resistance Applications : Copra Standard method : ISO 665 Measurement Range : 3 to 20% (dry base) Measurement Precision : Standard error 0.4% (moisture values of 10% or lower) Display : Digital (LCD) Resolution : 0.1% Operating Temperature Range : 0 to 40 C Functions : Moisture value bias adjustment ( 9.9 to +9.9%), average value, Upper limit alarm, Automatic temperature compensation, Auto power off (automatically turns off after approximately 5 minutes) Power Supply : 1.5 V batteries (AA alkaline) (6) Power Consumption : approximately 0.45 W Dimensions : 110 mm (W) x 210 mm (D) x 50 mm (H) Weight : 0.5 kg Accessories : Two needle proble, wrench, shoulder strap, 1.5 V battery (AA alkaline) (6), carrying case, operating manual Options : Printer (VZ-330), printer connecting cable (VZC-26) 5

<Main body> 3. Part Names <Accessories> Connector Display Two needle probe (Cap) Shoulder strap Control Panel 1.5 V battery (AA) (6) Wrench Battery cover (back) 6 Carrying case Operating manual

4. Display 3 1 2 4 5 Description of display Item No. Name of display area Description 1 Battery indicator Comes on when battery is depleted. 2 Number of measurment times Displays the number of times measured. 3 Alarm set-point value Displays the set-point value for the alarm for the upper moisture limit that was set. 4 Moisture value Displays the measured moisture value (%). 5 Bar graph Displays the measured moisture value by means of a bar graph. Displays up to 50% in 2% increments. 7

5. Description of Keypad Of the numeric keys of 0 to 9, some combine numerical input and other functions. Key Function <Control panel> Power On and Off Bias correction Not used for this device. Average value Entering a minus input Setting of upper alarm limit Setting printer output Use for measuring. You can use either the left or right side key. Continuous measurement 8

6. Directions 6-1.Before Measurement 1.The unit is powered by six 1.5 V batteries (AA, alkaline). Remove the rear battery cover, place the batteries into the compartment taking care to correctly orient the positive ( ) and negative ( ) terminals. Then attach the battery cover. is displayed when the batteries are depleted. Replace all six with new batteries. 2.Plug the probe connector into the socket on the main unit, and fasten and secure the lock ring. 2 Tighten the lock ring 1 Securely plug the connector in 3.Prepare the copra sample to be measured. Allow the sample to equilibrate to the same temperature as the main unit. Errors may occur when there is a large difference in the temperature of the copra and the temperature of the unit. To make more accurate measurements, allow the temperature of the copra to equilibrate with the temperature of the unit. 9

6-2.Measurement Method 1.Press the key to turn on the power. All elements of the LCD will be displayed for approximately three seconds. Subsequently, the TIMES, and % will be displayed. If the display shows something other than that described above, there may be a problem with the unit. Refer to p. 17 7. Error Display. 2.Hold the two needle probe firmly, and insert the sensor needles completely into the copra. Please make sure to stick the probe in from outside the shell. 10 The symbol denotes a blinking display.

3.Press the key. The decimal point will blink. After about 3 seconds a beep will sound and the Number of measurement times, Moisture value, and Bar graph will be displayed. The key on the left and the key on the right both have the same function so it doesn't matter which one you press. Please use the one that's easier to press. The bar graph displays a 50% full scale in 2% increments. If the moisture value is outside of the measuring range, HI will be displayed if the value higher than the measuring range, and LO will be displayed if it is lower. 4.Remove the two needle probe from the copra. At this time, the moisture value will remain on display. To continue to make further measurements, start the procedure from Step 2. Hold the two needle probe firmly.... When finished making measurements, press the key to turn off the power. With its auto power off function, the unit will automatically turn itself off when inactive after approximately 5 minutes. The symbol denotes a blinking display. 11

6-3.Setting the Measurement Mode There are two measurement modes: Normal measurement mode and Continuous measurement mode. Immediately after pressing the key and turning on the power, the unit will be set to Normal measurement mode. If the unit is set to Continuous measurement mode, measurements can be made without pressing the key for each measurement. 1.Continuous measurement mode setting In step 6-2. Measurement Method 3. Press the..., press the key. The decimal point will blink, and the moisture value and bar graph will be displayed. If the measured value is outside of the measuring range, either HI or LO will be displayed, and the decimal point will blink. 2.Canceling the continuous measurement mode Press the continuously for at least 2 seconds until a beep sounds. The display will momentarily go blank. When you take your finger off the key, the unit will return to normal measurement mode. Pressing the key turns off the power and automatically cancels this mode. Note that the battery life is shorter in continuous measurement mode. 12 The symbol denotes a blinking display.

6-4.Displaying the Average Value When the number of measurements is from 2 to 9, and you press, the AVE, average value, and number of measurement times will be displayed. At this time, the AVE, average value, and number of measurement times remain displayed, but if you proceed to make a measurement, the number of measurement times resets to 1. If the number of measurement times exceeds 9, the measured value up to that point resets and measurement starts from measurement number 1. The average value cannot be determined when making measurements in continuous measurement mode. 13

6-5.How to make Bias Adjustment The calibration (scale) for this unit was prepared by determining the relationship between the moisture of copra resulting from a drying method and the electrical resistance of the copra, and then statistically processing these results. However, depending on various conditions, there are cases where the standard measurement method and the moisture value do not agree. In these cases, a moisture value bias adjustment can be made ( 9.9% to 9.9%) by the following method. 1.Press the key. The BIAS indicator will blink, and the previously entered bias value will be displayed. The bias is initially set to 0.0%. 2.Enter the bias value. Enter a 2-digit number. To enter 1.2%, press the key and the key in succession. For a negative value, press the key before the numbers. 3.Press the key to make measurements. If a bias adjustment value has been entered, BIAS will be displayed during measurements. To cancel the bias adjustment value, enter 0.0%. 14 The symbol denotes a blinking display.

6-6.Setting the Alarm An upper limit moisture alarm can be set. When copra is measured having moisture higher than the set moisture value, the alarm sounds with a beeping tone. 1.Press the key. The numbers to the right of ALARM will blink. 2.Enter the alarm value. Enter a 2-digit number. To enter 8%, press the key and the in succession. 3.The set value for the alarm is displayed. Then, press the key to make measurements. To cancel the set value for the alarm, enter 0.0%. 15 The symbol denotes a blinking display.

6-7.Printer Output Measured values can be printed out by using the optional printer (VZ-330). The output content is number of measurement times, moisture value, and average moisture value. 1.How to set up Press the in normal measurement mode. P will be displayed to the right of TIMES. Set up is complete. The timing of the output is just after the measurement or just after the display of the average value. However, there is no output if the measured value is outside the measuring range. After setting up the printer (VZ-330, sold separately), connect the printer cable (VZC-26, sold separately). Refer to the operating manual for the printer for how to set up the printer. [Printout example] HX-120 TIMES [%] 1 6.8 2 6.9 3 6.7 4 7.0 5 6.9 6 6.8 AVERAGE 6.9 2.How to cancel the settings Press the key in normal measurement mode. OFF will be displayed to the right of TIMES. The settings have been cancelled. The settings in both 1 and 2 will be retained in memory when the power is turned off. 16

7. Error Display When there is a problem with the unit or with the measurement conditions, the following errors will be displayed for 4 seconds, and then the power will turn off. 1.There is a problem with the temperature sensor. Servicing is required. 2.When either of the errors on the right are displayed, there is a problem with the electronic circuit used for moisture measurement. Servicing is required. 17

3.The measurement cannot be made because the temperature of the unit is below 5 C. Bring the temperature of the unit up to the operating temperature range (0 to 40 C), and make the measurement again. 4.The measurement cannot be made because the temperature of the unit is above 50 C. Bring the temperature of the unit down to the operating temperature range (0 to 40 C), and make the measurement again. Indo-digital.com 18