PLATINblue MSQ Plus. Connection Guide V6951A UHPLC/HPLC. Mass spectrometry

Similar documents
2-1. the ESI Probe Assembly. the PEEK Safety Sleeve and Sample Tube to the ESI Probe

CryoFocus-4 GC Cryogenic Trap

Agilent 6400 Series. System. Installation Guide. Agilent Technologies

Quick start guide for i5 520 ( or )

CBT LW 4-PUMP ADD-ON INTERNATIONAL INSTALLATION GUIDE

Pre-installation guide for analytical systems User manual HPLC

Removal and Installation8

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes. Agilent 6230B Series TOF LC-MS System Installation Checklist

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes. System Information. Agilent Technologies

Operating Manual. GC-APPI Inferface for:

Instruction Manual. M Pump Motor Controller. For file reference, please record the following data:

Dionex SRD-10 Suppressor Regenerant Detector Installation Instructions

Haze 800 DMX. User Guide. English

ACQUITY UPLC Systems with 2D Technology Capabilities Guide

2.2. Facilities Requirements

Agilent G1978B Multimode Source for 6410 Triple Quad LC/MS

Customers should provide all necessary operating supplies upon request of the engineer.

Unpacking and Installing the Flora 2512 UV Printer. Steps 1: Unscrew the 10mm bolts holding the top. Then remove the top and put in a safe place.

IBM Systems. Quick start guide for IBM System p5 505 ( )

Quick start guide for p5 520 ( )

Control Box Setup - PRSalpha

Rosemount 370XA Natural Gas Chromatograph

Installing the Cisco AS5400XM Universal Gateway

Discovery SDT650. Site Preparation Guide. Revision A Issued March Page 1

PerkinElmer Series 200 HPLC: Operating Procedure

DC100 Digital Dispenser Operating Manual

870 KF Titrino plus. Installation and tutorial EN

SI-150-GIST. DART Thermo Quantiva, Endura, and Fusion VAPUR Interface User Manual. IonSense Inc. 999 Broadway Suite 404 Saugus, MA 01906

ASTRO UW-1C and RW-1C LABEL PRINTER UNWINDER & WINDER

Bio Compatibility Kit

9 Maintenance. Agilent Technologies. This chapter describes the maintenance of the module. InfinityLab LC Series Diode Array Detectors User Manual

Indoor Protected Entrance Terminal 4588 Series

Model VDL 115 Vacuum drying chambers for flammable solvents

INSTRUCTION MANUAL. Instruction Manual. Analog Multi-Tube Vortexer Digital Multi-Tube Vortexer

6 x 6 Peltier-Thermostatted Multicell Holder Series II Accessory

Installation Guide. Retrofit Kit for USB Ready Intraoral Systems

Standard Operating Procedure

Dionex AS-AP Sample Conductivity and ph Accessory Setup and Operation Guide

Avanti HP Series Centrifuges Preinstallation

ASXPRESS PLUS Rapid Sample Introduction System Quick Installation Guide

Ioline CrystalPress. Quick Start Guide

Procedure to Upgrade from B&W Maintenance Panel to TIB Color Touchscreen Door in a Working Maxum or Maxum II Analyzer Equipped With a SYSCON2

Model VD 23 Vacuum drying chambers for non-flammable solvents

User s Guide. Pressure Controller for liquid delivery systems

MK-101 TILE SAW OWNER S MANUAL & OPERATING INSTRUCTIONS SERIAL NUMBER:

PIX 515/515E. PIX 515/515E Product Overview CHAPTER

Description: Detailed procedure on removing old bushing and installing new Brake Bushing Replacement Kit 10447

Software. LabVIEW Plug & Play Driver P 2.1S, P 4.1S V7062A

DART SVP. Hardware and Network Installation Manual For DART SVP Source and Controllers. Revision 6 Document #

IBM. Rack Installation Instructions

DNG-200 LOW FOG GENERATOR USER MANUAL DNG-200 低煙機使用說明書

Power Supply and Fan Module Installation

Rack Installation Instructions

Micro GC Fusion Quick Start Guide

Checklist for the SMPS & CPC (5.400 & 5.500) PART 1 Setup of the external Plumbing (for the DMA and the CPC):

Pretium Rack-mountable 1U Housing Hardware

Thermal Diffusivity Instruments DLF-2 with EM-1600

Configuring and optimizing an IC-MS system using a compact IC and a single quadrupole mass spectrometer

Thermo. EASY-nLC User Guide for LC Devices

MX Series II Operating Manual. MX Series II Modular Automation for Nano and Analytical Scale HPLC And Low Pressure Fluid Switching Applications

Upgrading the UltiMate 3000 RSLCnano System with ProFlow Technology

Print Mechanism Maintenance Kit

User Guide Automatic Transfer Switch (ATS)

Parts and Service Manual IMCO

Quick Start Guide and Health and Safety Information

Cisco CRS 3-Phase AC Power Distribution Unit Installation Guide 2. Cisco CRS 3-Phase AC Power Distribution Unit 2

AD Leak Tester INSTRUCTION MANUAL 1WMPD

Operation Manual. Imara S100 & S60 DMX. Imara S100 DMX Yoke Mount. Imara S100 DMX Pole-Op

Agilent ICP-OES SVS 2 Productivity Package Installation and Upgrade Instructions

VBAR 270. User Manual. Version 1.2

Dell Inspiron XPS and Inspiron 9100 Service Manual

Agilent InfinityLab LC Series 1290 Infinity II Preparative Column Compartment

BAGGER V2 PX3 PARTS CATALOG

Table of Contents Quick Install Guide page Introduction Safety Rack System Precautions ESD Precautions...

TRC-190 User s Manual

Operation Manual. 4Bank DMX

Installation and Start-Up Instructions

3M Indoor Protected Building Entrance Terminal 4588V-QCS Series

頤拓企業有限公司 TEL: TEKHNE TK-100 Dewpoint Transmitter. etorch Co., L. Users Manual. TEKHNE Corporation

Breeze 2 Installation and Configuration Guide

9 Maintenance and Repair

CLIMB2 DUAL MONITOR SIT/STAND WORKSTATION

Industrial pressure controller Model CPC4000

BS 287 DUAL CHANNEL POWER SUPPLY. User Manual. January 2017 V1.0

1. Carefully unpack the um260 s shipping carton and check the contents for damage.

TDM To MiniMech conversion ProceDure

EASY-nLC Series User Guide for the Xcalibur Data System (Xcalibur version 2.1 or later)

Agilent GC, MS, and ALS

Upgrading and Servicing Guide

Steps for LCMS-8040 Triple Quadrupole

R u b b e r d i s t r i b u t o r

Gobo Projector XP 80W

VOYAGER PUMP DOWN PROCEDURE (BENCH-TOP MODELS)

Installation Manual INTELLIGENT SERVER. Model name: BMS-LSV6UL

Precision hand-held pressure indicator Model CPH6400

LH 700. Spare parts list. Forward and reversible plate. Valid from serial number. LH 700 Hatz Electric start 810 mm Bio

Agilent G1978B Multimode Source for 6100 Series Single Quad LC/MS

Gas Cooling Accessory

Illustrated Parts List

TCC-100 Thermostatted Column Compartment Quick Installation Guide August 2003 Rev. 2 Page Dionex

Transcription:

PLATINblue MSQ Plus Connection Guide V6951A UHPLC/HPLC Mass spectrometry

3 Table of contents Note: For your own safety, be sure to read the manual and always observe the warnings and safety information on the device and in the manual! Introduction................................................ 4 Connecting the Forepump........................................ 4 Using the Oil Mist Filter Kit and the Drain Oil Return Kit.................. 6 Connecting the Oil Mist Filter to the Forepump...................... 7 Connecting the Drain Adapter to the Oil Mist Filter................... 8 Connecting the Hose Adapter to the Forepump...................... 8 Connecting the Drain of the Oil Mist Filter to the Gas Ballast Port......... 9 Connecting the Exhaust Lines..................................... 10 Incorrect Setup!............................................. 12 Connecting to the Nitrogen Source................................. 12 Connecting the Nitrogen Supply to the Mass Detector................ 13 Checking for Air Leaks........................................ 14 Connecting to Line Power........................................ 15 Connecting to the Data System Computer........................... 16 To connect the USB cable...................................... 17 Connecting the UHPLC system.................................. 18 Checking the Ethernet Card Settings................................ 18 To confirm the assignment of the 3Com 3C905B-TX Ethernet card...... 18 Connecting the mass spectrometer to UHPLC system................... 18 Connecting the autosampler to the mass spectrometer............... 20 Connecting high-pressure pump to mass spectrometer............... 20 Connecting the LC Outlet to the Mass Detector....................... 22 Installing the API Probe....................................... 23 Switching from ESI to APCI....................................... 23 Switching from APCI to ESI....................................... 25 Control of External Devices.................................... 27 MSQ Plus Mass Detector User I/O Terminals.......................... 27 External Devices Controlled by Xcalibur Data System................... 27 Wiring a System Controlled by Xcalibur for Contact Closure............ 28 Selecting the Appropriate Start Instrument......................... 28 External Devices Not Controlled by the Xcalibur Data System............. 30 Wiring Your System for Contact Closure........................... 30 Running a Sequence from the Xcalibur Data System.................. 31 Table of figures............................................. 33 Index.................................................... 34 V6951A

4 Introduction Introduction This chapter explains how to connect your MSQ Plus Mass Detector to your nitrogen source, line power, and data system computer. It also describes how to connect the Edwards forepump to your mass detector and how to connect the exhaust lines from the mass detector and the forepump to your laboratory venting system. Fig. 1 MSQ Plus Mass Detector and UHPLC To connect the Edwards forepump to the mass detector Connecting the Forepump Your MSQ Plus Mass Detector is shipped with an Edwards forepump. Before connecting the forepump, ensure that it is filled to the correct level with the factory-supplied oil. For more detailed information on these procedures, refer to the manual that comes with the Edwards forepump or to the MSQ Plus Mass Detector Hardware Manual. Note: At the time of installation, a KNAUER service engineer will connect the forepump to the MSQ Plus Mass Detector. 1. Use a KF25 O-ring (part number A66531) and a KF25 clamp (part number A66533) to connect the T-piece to the inlet port of the Edwards forepump. 2. To connect the vacuum hose that exits the rear of the MSQ Plus Mass Detector through the source manifold to one side of the T-piece (see connection 1 in Figure 1), do the following:

5 Introduction Fit a hose clamp over the free end of the vacuum hose and insert a KF25 adapter into the end of the hose. Tighten the clamp to secure the adapter. Use a KF25 O-ring and a KF25 clamp to connect the KF25 adapter on the end of the vacuum hose one side of the T- piece. Tighten the clamp to secure the connection. 3. To connect the vacuum hose that exits the rear of the MSQ Plus Mass Detector through the backing manifold to the other side of the T-piece (see connection 2 in Figure 1): Fit a hose clamp over the free end of the vacuum hose and insert a KF25 adapter into the end of the hose. Tighten the clamp to secure the adapter. Use a KF25 O-ring and a KF25 clamp to connect the KF25 adapter on the end of the vacuum hose to the other side of the T-piece. Then tighten the clamp to secure the connection. Note: The vacuum hoses are hard-wired into the MSQ Plus Mass Detector. Do not attempt to remove them. 4. Connect the power cord of the Edwards forepump to the PUMP OUT receptacle on the rear panel of the MSQ Plus Mass Detector. A MSQ Plus Mass Detector A B Connection 1: Source to forepump C Connection 2: Backing to forepump D Connection 3: Forepump power plug to PUMP OUT receptacle E T-Piece F Exhaust port G Power switch B C D E F G Fig. 2 Connecting the forepump to the MSQ Plus Mass Detector For information on connecting the forepump exhaust line, refer to Connecting the Exhaust Lines.

6 Introduction Note: You must place the MSQ Plus Mass Detector on a benchtop within 2.5 m of a connection to your laboratory venting system. Using the Oil Mist Filter Kit and the Drain Oil Return Kit The oil mist filter captures oil mist from the exhaust port of the forepump that would otherwise be ejected into the atmosphere. The oil drain return kit allows you to periodically return oil trapped in the oil mist filter back to the forepump. The oil mist filter kit contains both the EMF20 oil mist filter kit and the drain oil return kit. Instructions for installing the oil mist filter and the drain oil return are included in this kit, as well as in this section. The EMF20 oil mist filter kit contains the parts listed in Table 1. The drain oil return kit contains the parts listed in Table 2. Table 1: Parts included in the oil mist filter kit Part Quantity Oil mist filter 1 Adapter 1 O-ring seal 1 Centering ring and O-ring 1 Clamp 1 Table 2: Parts included in the drain oil return kit Part Quantity Drain adapter 1 Gas ballast adapter 1 O-ring 1 Bonded seals 2 Banjo bolt 1 Flexible return tubing 1 Flow restrictor 1 Hose clips 2 To connect the oil mist filter and the drain oil return to your forepump, follow these procedures: 1. Connecting the Oil Mist Filter to the Forepump 2. Connecting the Drain Adapter to the Oil Mist Filter

7 Introduction 3. Connecting the Hose Adapter to the Forepump 4. Connecting the Drain of the Oil Mist Filter to the Gas Ballast Port To connect the oil mist filter to the exhaust port of the forepump A Oil mist filter B Bonded seal C Drain plug D Clam E Centering-ring F Adaptor G O-ring H Gas ballast knob I Exhaust port Connecting the Oil Mist Filter to the Forepump Follow these steps to connect the oil mist filter to the forepump. 1. Turn the power switch on the Edwards forepump to the Off position. If its power cable is attached to Pump Out receptacle on the rear panel of the MSQ Plus Mass Detector, detach it. See Figure 1. 2. Use a 1.25 in. wrench to remove the brass pipe fitting (see Figure 1) that extends from the exhaust port of the Edwards forepump. 3. Clean the area around the exhaust port of the forepump. 4. Fit the O-ring seal to the adapter. 5. Screw the adapter into the exhaust port of the forepump and hand-tighten. See Figure 2. Do not overtighten the adapter or you will damage the O-ring seal. 6. Slip the centering ring onto the adapter. 7. Position the flange of the oil mist filter on the centering ring. 8. Fit the clamp over the flange and hand tighten. A B C D E F G H I Fig. 3 Connecting the oil mist filter to the Edwards forepump

8 Introduction To connect the drain adapter to the oil mist filter To connect the hose adapter to the gas ballast port of the forepump Connecting the Drain Adapter to the Oil Mist Filter Follow this procedure to connect the drain adapter to the oil mist filter. 1. Remove the drain plug and the bonded seal from the oil mist filter (Figure 2). 2. Place the bonded seal onto the drain adapter and screw the drain adapter into the opening that you created by removing the drain plug. Connecting the Hose Adapter to the Forepump Follow these steps to connect the hose adaptor to the forepump. 1. Using a right-hand circlip removal tool, remove the circlip from the gas ballast port on the forepump. 2. Using a screwdriver, pry the wire mesh and filters out of the gas ballast port. 3. Fit the O-ring to the gas ballast adapter, and screw the gas ballast adapter into the gas ballast port on the forepump. 4. Using the banjo bolt and the two bonded seals, attach the hose adapter to the gas ballast adaptor. A Circlip B Wire mesh C Filters D Gas ballast port E Gas ballast knob A B C D E Fig. 4 Removing the circlip, wire mesh, and filters

9 Introduction A Banjo bolt B Bonded seal C Host adapter D Bonded seal E Gas ballast adaptor F O-ring G Gas ballast port H Gas ballast knob A B C D E F G H Fig. 5 Installing the hose adaptor To connect the drain adapter on the oil mist filter to the hose adapter on the gas ballast port Connecting the Drain of the Oil Mist Filter to the Gas Ballast Port This section explains how to connect the oil mist filter drain to the gas ballast port. 1. Cut a suitable length from the flexible oil return tubing. Ensure that the ends of the tubing are free of burrs and that they are squarely cut. 2. Insert the brass restrictor into one end of the tubing. 3. Attach one end of the tubing to the drain adaptor on the oil mist filter and the other end of the tubing to the hose adaptor. 4. Using the two hose clips, secure the ends of the tubing. Ensure that the tubing is neither taut nor kinked and that the flow path from the drain adaptor on the oil mist filter to the hose adaptor on the forepump is downhill all the way.

10 Introduction A Host adaptor B Hose clip C Flexible tubing D Hose clip E Drain adaptor A B C D E Fig. 6 Connecting the flexible oil return tubing You must connect the oil mist filter to a vent. A blue exhaust hose with a 1 in. ID is included in the Edwards forepump installation kit for this purpose. Exhaust hose clamps are included in this kit; however, clamping the exhaust hose is not required, because the press fit makes an adequate seal. For instructions on connecting the oil mist filter to a vent, see the next section, Connecting the Exhaust Lines. Connecting the Exhaust Lines The forepump exhaust must be vented to atmosphere external to the laboratory. In addition, the liquid waste that exits the system through the MSQ Plus Mass Detector exhaust port must be captured in a solvent trap. This trap must be vented to atmosphere external to the laboratory. Both of these exhausts must have a separate vent, and the venting must be carried out by way of separate user-supplied fume hoods or industrial vents. Consult local regulations for the proper method of exhausting gases from your system. Caution! The tubing that extends from the Exhaust port must be connected to a solvent trap. Vapor phase solvents produced in the source chamber drain through this tubing. Therefore, connecting this tubing to the forepump exhaust can cause serious instrument damage. Note: Place the forepump within 2.5 m of the laboratory connection to an external vent.

11 Introduction To connect the exhaust lines for your system A MSQ Plus Mass Detector B Connection 4: Exhaust to solvent trap C Connection 5: Solvent trap to vent D Connection 6: Forepump exhaust to vent E Oil return tubing F Oil mist filter G Gas ballast knob H Forepump I Solvent trap 1. Cut the tubing that extends from the exhaust port on the rear panel of the MSQ Plus Mass Detector to an appropriate length for your system. Then connect the free end of the tubing to the solvent trap. See connection 5 in Figure 7. 2. Use the extra length of exhaust tubing that you just cut to connect the solvent trap to the user-supplied vent for the exhaust fumes from the solvent trap. See connection 4 in Figure 7. 3. Use an adapter, a centering ring, and a clamp to attach the blue hose to the top of the oil mist filter. 4. Direct the other end of the blue hose to a vent that you have supplied for the forepump exhaust. See connection 6 in Figure 7. A B C D E F G H I Fig. 7 Caution! Connecting exhaust lines from the source and forepump without oil mist filter For the solvent trap to work effectively, you must place it on a lower level than the instrument so that solvent exiting the exhaust port can flow into it. Do not raise the tubing from the exhaust port above the level of the port, because this will cause solvent to flow back into the instrument. Serious instrument damage can occur as a result. Cut the tubing to a suitable length so that it does not loop down below the top of the solvent trap. If it does, solvent will collect in the tubing and might cause a blockage. See Figure 2.

12 Introduction Incorrect Setup! If the tubing is above the exhaust outlet, the solvent condensing from the exhaust gases can flow back into the instrument. Fig. 8 Caution! Incorrectly connecting exhaust lines from the source and forepump Do not allow the oil mist filter to become overfilled with oil. To empty the oil mist filter, open the gas ballast knob at the end of a sequence run or at the end of the day before you place the system in the Off mode. After the oil in the filter returns to the forepump, close the gas ballast knob. Connecting to the Nitrogen Source The MSQ Plus Mass Detector requires a nitrogen gas supply for the source. The nitrogen must be of high purity (99%). The required gas pressure is 520 kpa (5.2 bar, 75 psi). Nitrogen gas consumption is 12 L per min in ESI mode and 8 L per min in APCI mode. Therefore, KNAUER recommends you use one of the following sources for your nitrogen supply:

13 Introduction A large, sealed, thermally insulated cylinder containing liquid nitrogen, from which the nitrogen is boiled off. The 230 psi model is recommended. A typical cylinder of size 240 L yields 143850 L (5080 ft 3 ) of gas. Replacement frequency is approximately every 9 to 12 days. Note: Liquid nitrogen conversion factors: 1.0 lb of liquid nitrogen = 0.5612 L 1.0 kg of liquid nitrogen = 1.237 L A nitrogen generator, such as the Peak Scientific Model N118LA (part number A66501), with minimum capacity of 720 L per hour at 99% purity with 100 psi at the side panel. Maximum consumption of nitrogen gas is 12 L per minute. Replacement frequency is continuous source with no replacement required. Routine maintenance is required for commercial nitrogen generators, usually on an annual basis. Connecting the Nitrogen Supply to the Mass Detector Nitrogen enters the MSQ Plus Mass Detector by way of an inlet labeled GAS IN on its rear panel. The MSQ Plus Mass Detector is supplied with the following gas line connections: 6 mm OD PTFE tubing Built-in pressure regulator 6 mm 1/4 in. BSP half union (part number A66532) Swagelok stainless steel fitting for 1/4 in. OD tubing (part number A66530) You must supply suitable adapters to connect your gas line to the 6 mm 1/4 in. BSP half union. The Swagelok fitting is easily adaptable to typical (U.S.) 1/4 in. laboratory gas supply lines. See Figure 4. A 6 mm 1/4 in. BSP half union B Stainless steel fitting for 1/4 in. OD tubing A B Fig. 9 Fitting for 6 mm OD tubing and stainless steel fitting for 1/4 in. OD tubing To connect the MSQ Plus Mass Detector to the nitrogen supply 1. Cut a suitable length of the supplied 6 mm OD PTFE tubing. 2. Screw the 6 mm 1/4 in. BSP half union into the stainless steel adapter for 1/4 in. OD tubing. Connect the free end of the stainless steel adapter to your nitrogen supply source. 3. Push one end of the 6 mm PTFE tubing into the 6 mm - 1/ 4 in. BSP half union.

14 Introduction 4. Connect the other end of the 6 mm OD PTFE tubing to the inlet marked GAS IN inlet on the rear of the MSQ Plus Mass Detector. See connection 7 in Figure 10. A connection 7: Nitrogen supply to the GAS IN port A Fig. 10 Connecting the nitrogen supply to the GAS IN port To check the gas supply system for leaks Checking for Air Leaks Follow this procedure to check the gas supply system for air leaks. 1. Prepare a test solution of 50:50 isopropanol/water (v/v). 2. Set the pressure to 10 psi: Pull out the control dial on the auxiliary pressure regulator, and turn it clockwise. See Figure 11. When the pressure gauge reads approximately 10 psi, push the adjuster back to its locked position. 3. Apply the test solution to all the joints between the regulator and the MSQ Plus Mass Detector. Ensure that no test solution enters the MSQ Plus Mass Detector through the gas inlets. The test solution indicates a leak by foaming. 4. If you observe any foaming of the test solution, tighten the connection until the foaming stops, or replace the connection. 5. Increase the pressure to 50 psi. Then check for leaks. If the system is airtight, go to step 6 If you find a leak, return to step 4

15 Introduction 6. Increase the pressure to 100 psi. Then check for leaks. If the system is airtight, go to step 7. If you find a leak, return to step 4. 7. After you have determined that the system is airtight, reduce the pressure of the nitrogen gas supply line to 75 psi. The recommended operating pressure required for the gas inlet on the rear of the MSQ Plus Mass Detector is 75 to 80 psi. A Pressure gauge B Control dial A B Fig. 11 Pressure regulator on MSQ Plus Mass Detector To connect the MSQ Plus Mass Detector to a power supply Connecting to Line Power Refer to the MSQ Plus Mass Detector Preinstallation Guide for information on power supply requirements. Ensure that the power cable is fitted with the correct cable termination for your electrical power supply outlets. 1. Remove the 2 Phillips-head screws that secure the MAINS IN panel on the rear of the MSQ Plus Mass Detector. See Figure 12. Retain the screws. 2. Plug the 3-pin IEC socket into the port marked MAINS IN on the rear of the MSQ Plus Mass Detector. See connection 8 in Figure 12. 3. Slide the 2 Phillips-head screws through the brackets in the 3- pin IEC socket, and tighten the screws. 4. Plug the other end into a suitable electrical wall outlet.

16 Introduction Caution! Do not switch on the power until the installation and all connections are complete. A Connection 8: Phillips head screws A Fig. 12 Note: Connecting the MSQ Plus Mass Detector to the line power If your local area is subject to power fluctuations or power interruptions, a power-conditioning device or an uninterruptible power supply (UPS) should be installed in your laboratory. (Refer to the Power Conditioning Devices topic in the MSQ Plus Mass Detector Preinstallation Guide.) Connecting to the Data System Computer The data system for the MSQ Plus Mass Detector consists of a computer, a monitor, and a user-supplied printer. The MSQ Plus Mass Detector communicates with the data system computer (PC) through a USB cable.

17 Introduction A Series A socket B Series B socket C Series A plug D Series B plug To connect the USB cable 1. Locate the USB cable, consisting of series A plug linked to a series B plug. 2. Connect the series B plug to the port marked USB on the rear of the MSQ Plus Mass Detector. 3. Connect the series A plug to the rear of the computer, in the USB slot. See Connection 11 in Figure 14. A B C D Fig. 13 USB cable connection plugs At a minimum, do not use a USB hub adjacent to the MSQ series A plug. Note: Leave the other USB slots on the rear of the PC vacant. Use the USB slots located on the front of the data system computer to connect other USB devices, such as the keyboard and the mouse. A Data system computer B MSQ Plus Mass Detector A B Fig. 14 Connecting the MSQ Plus Mass Detector to the data system computer

18 Connecting the UHPLC system Connecting the UHPLC system This chapter describes aspects of connecting the MSQ Plus Mass Detector to the KNAUER UHPLC system. For detailed information on connecting the MSQ Plus Mass Detector to UHPLC, refer to the Getting Connected Guide. An UHPLC system consists of UHPLC pumps, an UHPLC autosampler, a termostat (oven), and an optional UHPLC PDA Detector. Checking the Ethernet Card Settings Before you connect the system interconnect cable, check the settings of the Ethernet card. To confirm the assignment of the 3Com 3C905B-TX Ethernet card 1. From the Windows XP taskbar, choose Start > Control Panel. 2. Double-click the Network Connections icon to open the Network Connections system folder. 3. Right-click the Local Area Connection 3 icon, and then choose Properties from the shortcut menu. The Local Area Connection Properties dialog box opens. 4. Select Internet Protocol (TCP/IP) from the This Connection Uses The Following Items list box. Click [Properties] to open the Internet Protocol (TCP/IP) Properties dialog box. 5. Confirm that the IP address for the 3Com 3C905B-TX Ethernet card is 172.16.0.101. 6. Click [OK] to close the Internet Protocol (TCP/IP) Properties dialog box, and click [OK] to close the Network dialog box. Connecting the mass spectrometer to UHPLC system Installing a local area network Network connection Connecting the LC-MS modules The analysis system, workstation and router are interconnected to form a LAN. Once the router has assigned an IP address to all modules, the Xcalibur software can begin configuring the modules (remote control). The network connections are located at the rear of the modules. The mass spectrometer is connected to the autosampler using a contact cable. The contact cable transfers a start signal issued by the autosampler to the mass spectrometer. If no autosampler is used, the high-pressure pump can be connected to the mass spectrometer. As soon as the high-pressure

19 Connecting the UHPLC system Connecting the mass spectrometer to the autosampler pump starts up, it sends the start signal to the mass spectrometer. 1 2 3 Fig. 15 Schematic display of connections for mass spectrometer and autosampler; legend: 1 On/Off switch 2 Fuse 3 Power connection Connecting the mass spectrometer to the high-pressure pump Fig. 16 Schematic display of connections for mass spectrometer and high-pressure pump Accessories Multicore contact cable, A1467 WAGO terminal strip, A1421 WAGO microstrip, A1420 Network cable, A5255 Note: The contact cable is fitted with a DB15 connector at one end and three sheathed wires in different colors at the other.

20 Connecting the UHPLC system Connecting the autosampler to the mass spectrometer 1. Plug the black and red wires of the contact cable into the WAGO terminal strip (1) 2. On the rear side of the mass spectrometer, insert the WAGO terminal strip into the female connector (2) so that the black wire is connected to START IN + and the red wire is connected to START IN - 1 2 Fig. 17 Connection Mass spectrometer 3. On the rear side of the autosampler, insert the DB15 connector (4) into the IN/OUT female connector (3) 3 4 Note: The two ends of the multicore contact cable are connected by a WAGO terminal strip. Connecting high-pressure pump to mass spectrometer 1. Plug the black and red wires of the contact cable into the WAGO terminal strip (1) 2. On the rear side of the mass spectrometer, insert the WAGO terminal strip into the female connector (2) so that the black wire is connected to START IN + and the red wire is connected to START IN - 1 2 Fig. 18 Connection Mass spectrometer

21 Connecting the UHPLC system 3. On the rear side of the pump, insert the terminal strip into the female connector for remote control so that the blackencased wire is connected to OUT (START) and the redencased wire is connected to GND (START) Fig. 19 Note: Connection High-pressure pump Xcalibur software is used to configure the modules and program controller methods for the system.

22 Connecting the UHPLC system Connecting the LC Outlet to the Mass Detector To connect the LC System to the MSQ Plus Mass Detector, use the fittings for 1/16 in. OD tubing and the red 0.005 in. ID PEEK tubing that are included in the MS accessory kit to connect the outlet of the LC column or the outlet of the LightPipe flowcell to the inlet of the MSQ Plus Mass Detector. For high-speed chromatography, minimize bandspread by connecting the outlet of the LC directly to the grounded union at point A being careful not to crimp the LC tubing when you close the door. If you disconnect the original calibration line from point A, you must re-plumb the system to perform subsequent autotune procedures. A Connection from LC System B Point A A B Fig. 20 Connecting to the MSQ Plus Mass Detector

23 Installing the API Probe Installing the API Probe This chapter describes how to install the ESI probe or the APCI probe in the MSQ Plus Mass Detector. The MSQ Plus Mass Detector is shipped with an ESI probe and an APCI probe. The API probe that is not being used is stored in the front door of the mass detector. A APCI probe, installed B ESI probe A B Fig. 21 MSQ Plus Mass Detector with opened front door Switching from ESI to APCI Follow these steps to install the APCI probe in the MSQ Plus Mass Detector. Caution! Allow the probe heater to cool before you remove the ESI probe. To switch from ESI mode to APCI mode 1. Turn off the LC pump flow. If you are using the cone wash pump, turn it off. 2. Turn off the nitrogen gas, ion optics, and probe heater by doing one of the following:

24 Installing the API Probe From the Status page in the Information view in the Xcalibur data system, right-click the MSQ Plus listing to display a shortcut menu, and choose Turn Device Off. Fig. 22 Status page in the Xcalibur data system or Open the Per Method Parameters table in the Tune window. Take the system out of Operate mode by clicking the Operate On/Off toggle button. Then turn off the nitrogen gas by clicking the Nitrogen Gas On/Off toggle button. A Operate On/Off B Nitrogen Gas On/Off A B Fig. 23 Operate and Nitrogen Gas toggle buttons 3. Allow the probe heater to cool. 4. Unscrew and remove the PEEK fingertight fitting from the ESI probe. See Figure 18. 5. Turn the locking plate on the ESI probe clockwise to the open position. Pull the ESI probe out of the probe heater. See Figure 18. 6. Remove the APCI probe from the holder located in the door of the MSQ Plus Mass Detector, and replace it with the ESI probe.

25 Installing the API Probe A PEEK fitting B Locking plate in locked position C Probe heater D Probe mount E Corona pin knob F Source housing 7. Turn the locking plate on the APCI probe clockwise to the open position. Insert the APCI probe into the probe heater, as shown in Figure 18. Turn the locking plate counterclockwise to the closed position. 8. Turn the corona pin knob 90 degrees to its full horizontal position. 9. Insert the PEEK fingertight fitting into the APCI probe and screw in. A B C D E F Fig. 24 Note: MSQ Plus Mass Detector setup for APCI mode You might obtain a better signal using pin positions up to 1-2 centimeters below a full right angle Switching from APCI to ESI Follow these steps to install the ESI probe in the MSQ Plus Mass Detector. Caution! Allow the probe heater to cool before you remove the APCI probe. To switch from APCI mode to ESI mode 1. Turn off the LC pump flow. If you are using the cone wash pump, turn it off. 2. Turn off the nitrogen gas, ion optics, and probe heater by doing one of the following: From the Status page of the Information view in the Xcalibur data system, right-click the MSQ Plus listing to display

26 Installing the API Probe A PEEK fitting B Locking plate in locked position C ESI probe D Probe mount E Probe heater F Corona pin knob G Source housing a shortcut menu, and choose Turn Device Off. See Figure 19. or Open the Per Method Parameters table in the Tune window. Take the system out of the Operate mode by clicking the Operate On/Off toggle button. Then, turn off the nitrogen gas by clicking the Nitrogen Gas On/Off toggle button. See Figure 20. 3. Allow the probe heater to cool. 4. Unscrew and remove the PEEK fingertight fitting from the APCI probe. See Figure 18. 5. Turn the corona pin knob 90 degrees to its full vertical position. 6. Turn the locking plate of the APCI probe clockwise to the open position and remove the APCI probe from the mass detector. 7. Remove the ESI probe from the holder located in the door of the MSQ Plus Mass Detector and replace it with the APCI probe. 8. Turn the locking plate on the ESI probe clockwise to the open position. Insert the ESI probe into the probe heater, as shown in Figure 18. Turn the locking plate counterclockwise to the closed position. 9. Insert the PEEK fingertight fitting into the ESI probe and screw in. A B C D E F G Fig. 25 MSQ Plus Mass Detector setup for the ESI mode

27 Control of External Devices Control of External Devices This chapter describes how to connect the MSQ Plus Mass Detector to an external device (typically an autosampler) used to trigger data acquisition. The MSQ Plus Mass Detector can start data acquisition upon receiving a contact closure (closed contact or open contact) signal from an external device. The external device can be controlled from the Xcalibur data system or it can be controlled from outside the Xcalibur data system, for example, from a front panel keypad. MSQ Plus Mass Detector User I/O Terminals Table 3 provides functional descriptions and typical uses for the terminals (pins 1 and 2) located on the MSQ Plus Mass Detector user I/O panel. Table 3: MSQ Plus Mass Detector user I/O panel Pins Functional description Typical use 1 and 2 Start In The Start In internal circuit is a digital latch that provides a 5 V High TTL signal. The external device must be capable of pulling the signal Low to less than 2.5 V. Note: The external circuit must have a true ground. The Start In circuit allows an external device to signal a start to the MSQ Plus Mass Detector. External Devices Controlled by Xcalibur Data System The Xcalibur data system can respond to control from external devices (for example, autosamplers, pumps, and detectors) from KNAUER and several other manufacturers. You can find additional information about using these devices to trigger data acquisition from the mass detector in the following documents: For third-party LC devices controlled by the Xcalibur data system, refer to the manuals on the LC devices CD. To trigger data acquisition with an external device controlled by the Xcalibur data system, you must connect the external device to the MSQ Plus Mass Detector for contact closure. In addition, you must ensure that the appropriate device (typically an autosampler) is listed as the start instrument in the Xcalibur Run Sequence dialog box.

28 Control of External Devices Wiring a System Controlled by Xcalibur for Contact Closure For information on controlling your LC devices from the Xcalibur data system, refer to the Help available from the Xcalibur Instrument Setup window. For front-panel (keypad) operation of the LC devices and maintenance procedures for the LC devices, refer to the documentation provided with the LC. Note: The external device providing the start signal must have a good ground. A MSQ Plus Mass Detector A B C B Two-wire trigger cable C Triggering Fig. 26 Contact closures for the MSQ Plus Mass Detector and an external device To select the appropriate start instrument Selecting the Appropriate Start Instrument Ensure that the appropriate device is listed as the start instrument. 1. On the Xcalibur Roadmap view, click the Sequence Setup icon,, to open the Sequence Setup window. 2. Open the sequence that you want to run: Choose File > Open to open the Open dialog box. Browse to the appropriate folder, and then select a sequence file. Click [Open] to open the sequence, and exit the Open dialog box. Sequence files are identified by their.sld file extension.

29 Control of External Devices 3. Choose Actions > Run Sequence or Actions > Run This Sample to open the Run Sequence dialog box. Fig. 27 Run Sequence dialog box with the UHPLC selected as the start instrument 4. Verify that the appropriate device is listed as the start instrument in the Acquisition Options box. 5. If the appropriate device is not listed as the start instrument, change the starting device: Click [Change Instruments] to open the Change Instruments In Use dialog box. Click the blank field in the Start Instrument column to the right of the appropriate triggering device (typically an autosampler). The word Yes moves to this field. Click [OK] to save the setting, and close the Change Instruments In Use dialog box.

30 Control of External Devices 6. Complete the remaining selections in the Run Sequence dialog box. Click [OK] to save the settings, close the dialog box, and start the sequence or queue it. Fig. 28 Change Instruments In Use dialog box with the UHPLC selected as the start instrument External Devices Not Controlled by the Xcalibur Data System To trigger data acquisition by using an external device that is not controlled by the Xcalibur data system, you must connect the external device to the MSQ Plus Mass Detector for contact closure. In addition, you must ensure that the mass detector is not listed as the start instrument in the Xcalibur Run Sequence dialog box. Note: The output (start) signal from the external device must be normally High (+5 V) and go to Low momentarily to start data acquisition on the MSQ Plus Mass Detector. If you cannot configure the external device to go from normally High to Low momentarily, you cannot use it with the MSQ Plus Mass Detector. Wiring Your System for Contact Closure To hard-wire an external device that is not controlled by the Xcalibur data system to the MSQ Plus Mass Detector, connect the two-wire trigger cable (part number A66527 and provided in the MSQ Plus Mass Detector installation kit) shown in Figure 26 from the user I/O of the mass detector to the contact closure terminal of the external device. Follow the wiring scheme shown in Table 5. Fig. 29 MSQ Plus Mass Detector remote start cable

31 Control of External Devices Table 5: Wiring the MSQ Plus Mass Detector and an external device not controlled by the Xcalibur data system for contact closure MSQ Plus Mass Detector user I/O Start In + Start In - External device contact closure terminal Output (start) terminal Ground terminal Running a Sequence from the Xcalibur Data System When you are ready to inject a set of samples, ensure that the MSQ Plus Mass Detector is not listed as the start instrument in the Xcalibur data system. 1. On the Xcalibur Roadmap view, click the Sequence Setup icon,, to open the Sequence Setup window. To start the sequence run 2. Open the sequence that you want to run: Choose File > Open to open the Open dialog box. Browse to the appropriate folder, and then select a sequence file. Click [Open] to open the sequence, and exit the Open dialog box. Sequence files are identified by their.sld file extension. 3. Choose Actions > Run Sequence or Actions > Run This Sample to open the Run Sequence dialog box. The MSQ Plus Mass Detector is not the start instrument. Fig. 30 Run Sequence dialog box

32 Control of External Devices 4. Click [Change Instruments] to open the Change Instruments In Use dialog box. MSQ Plus Mass Detector is not selected as a start instrument. Fig. 31 Change Instruments In Use dialog box The MSQ Plus Mass Detector should not be in the Start Instrument: Yes mode. Practical tip 5. Check the status of the MSQ Plus Mass Detector in the Acquisition Options group box. If the MSQ Plus Mass Detector is in the Start Instrument: Yes mode, go to step 6. If the MSQ Plus Mass Detector is not in the Start Instrument: Yes mode, click [OK] to close the dialog box and go on to step 7. 6. In the MSQ Plus Mass Detector row of the Start Instrument column, click [Yes] to change the mode to Off (field is blank), and click [OK] to save the setting and close the dialog box. 7. In the Acquisition Options group box, select the Start When Ready check box, and click [OK] to save the settings, close the dialog box, and start the sequence or queue it. The instrument method is downloaded to the MSQ Plus Mass Detector, and the Status page displays Waiting - Contact Closure. If the Information view is not displayed, click the Information view toggle button,, to display it. Then click the Status tab to display the Status page. 8. Push the [Start] button on the external device to start the external device. Data acquisition from the MSQ Plus Mass Detector begins after the external device sends the contact closure signal to the MSQ Plus Mass Detector.

33 Table of figures Fig. 1: MSQ Plus Mass Detector and UHPLC........................ 4 Fig. 2: Connecting the forepump to the MSQ Plus Mass Detector........ 5 Fig. 3: Connecting the oil mist filter to the Edwards forepump.......... 7 Fig. 4: Removing the circlip, wire mesh, and filters................... 8 Fig. 5: Installing the hose adaptor................................ 9 Fig. 6: Connecting the flexible oil return tubing..................... 10 Fig. 7: Connecting exhaust lines from the source and forepump without oil mist filter.......................................... 11 Fig. 8: Incorrectly connecting exhaust lines from the source and forepump............................................ 12 Fig. 9: Fitting for 6 mm OD tubing and stainless steel fitting for 1/4 in. OD tubing........................................... 13 Fig. 10: Connecting the nitrogen supply to the GAS IN port............ 14 Fig. 11: Pressure regulator on MSQ Plus Mass Detector................ 15 Fig. 12: Connecting the MSQ Plus Mass Detector to the line power...... 16 Fig. 13: USB cable connection plugs.............................. 17 Fig. 14: Connecting the MSQ Plus Mass Detector to the data system computer............................................ 17 Fig. 15: Schematic display of connections for mass spectrometer and autosampler.......................................... 19 Fig. 16: Schematic display of connections for mass spectrometer and high-pressure pump.................................... 19 Fig. 17: Connection Mass spectrometer............................ 20 Fig. 18: Connection Mass spectrometer............................ 20 Fig. 19: Connection High-pressure pump.......................... 21 Fig. 20: Connecting to the MSQ Plus Mass Detector.................. 22 Fig. 21: MSQ Plus Mass Detector with opened front door.............. 23 Fig. 22: Status page in the Xcalibur data system..................... 24 Fig. 23: Operate and Nitrogen Gas toggle buttons................... 24 Fig. 24: MSQ Plus Mass Detector setup for APCI mode................ 25 Fig. 25: MSQ Plus Mass Detector setup for the ESI mode............... 26 Fig. 26: Contact closures for the MSQ Plus Mass Detector and an external device........................................ 28 Fig. 27: Run Sequence dialog box with the UHPLC selected as the start instrument.................................... 29 Fig. 28: Change Instruments In Use dialog box with the UHPLC selected as the start instrument.................................... 30 Fig. 29: MSQ Plus Mass Detector remote start cable.................. 30 Fig. 30: Run Sequence dialog box................................ 31 Fig. 31: Change Instruments In Use dialog box...................... 32

34 Index A air leaks 14 APCI probe 23 B backing manifold 5 banjo bolt 8 BSP half union 13 C Change Instruments In Use dialog box 29, 32 circlip 8 cone wash pump switching from APCI to ESI mode 25 switching from ESI to APCI mode 23 Connecting high-pressure pump to mass spectrometer 20 Connecting the autosampler to the mass spectrometer 20 Connecting the LC Outlet to the Mass Detector 22 contact closure signal 27 D drain oil return 8 drain oil return kit 6 E EMF20 oil mist filter kit 6 ESI probe 23, 25 Ethernet card 18 exhaust hose 10 exhaust hose clamps 10 exhaust lines 10, 12 external devices 27 F forepumps connecting exhaust lines 10 connecting to drain oil return 8 connecting to hose adapter 8 connecting to mass detector 4 connecting to oil mist filter 7 Edwards 4 G gas ballast port 8 H hose adapter 8 I ID PEEK tubing 22 Installing the API Probe 23 Instrument Setup view 28 Internet Protocol (TCP/IP) Properties dialog box 18 Introduction 4 ion optics 23 K KF25 clamp 4 KF25 O-ring 4 L LC pump switching from APCI to ESI mode 25 switching from ESI to APCI mode 23 LC-MS 18 LightPipe flowcell 22 Local Area Connection Properties dialog box 18 M mass spectrometer 18 MSQ Plus Mass Detector connecting to data system computer 16 connecting to external device controlled by Xcalibur data system 27 connecting to external device for data acquisition 27 connecting to external device not controlled by Xcalibur data system 30 connecting to nitrogen supply 13 connecting to UHPLC system 18 installing APCI probe 23 installing ESI probe 23, 25 terminals on user I/O panel 27 N nitrogen gas 12, 23 nitrogen generator 13 nitrogen supply 13 O OD PTFE tubing 13 OD tubing 22 oil mist filter connecting drain to gas ballast port 9 connecting to forepump 7 purpose 6

35 oil mist filter kit 6 Open dialog box 28, 31 P PEEK fingertight fitting 24, 26 Per Method Parameters table 24 power cables 7 power supply 15 probe heater 23, 24, 26 PUMP OUT receptacle 5 R Roadmap view 28 Run Sequence dialog box 29, 31 S Sequence Setup icon 28, 31 series A plug 17.sld files 28, 31 solvent trap 10 source manifold 4 Swagelok fitting 13 T T-piece 4 two-wire trigger cable 30 U UHPLC autosampler 18 UHPLC PDA detector 18 UHPLC pump 18 UHPLC system components of 18 connecting to MSQ Plus Mass Detector 18 USB cable 16 USB port 17 USB slots 17 V vacuum hose 4, 5 X Xcalibur data system controlling LC devices from 28 Instrument Setup view 28 Roadmap view 28 Xcalibur Run Sequence dialog box 27, 30

Wissenschaftliche Gerätebau Dr. Ing. Herbert Knauer GmbH All rights reserved. The information in this document is subject to change without prior notice. Original version of the manual. 2011-01-31 Printed in Germany. PLATINblue, ChromGate and BlueOrchid are registered trademarks of Wissenschaftliche Gerätebau Dr. Ing. Herbert Knauer GmbH Visit our website for more information www.platinblue.com www.knauer.net HPLC SMB Osmometry Wissenschaftliche Gerätebau Dr. Ing. Herbert Knauer GmbH Hegauer Weg 38 14163 Berlin, Germany Phone: +49-(0)30-809727-0 Fax: +49-(0)30-8015010 email: info@knauer.net Internet: www.knauer.net