BetaProbe TI/TI+ High Accuracy Digital Thermometer. User Reference Manual

Similar documents
TC100 Precision Thermocouple Calibrator

MC1000 Multi-Function Calibrator

Reference Manual Industrial Pressure Indicator AMETEK JOFRA IPI. Copyright 2005 AMETEK Denmark A/S

LC-110/H. Reference Manual

User's Guide. Phase Sequence and Motor Rotation Tester Model

CA846 THERMO-HYGROMETER ENGLISH. User Manual. 99 Washington Street Melrose, MA Phone Toll Free

Micro-Ohmmeter Model 6292

IAQ-CALC INDOOR AIR QUALITY METER MODEL 7525

User Manual. Thermocouple Datalogging Thermometer. Model EA15

CL600 SIMPLE LOGGER RMS CLAMP-ON MODULE E N G L I S H. User Manual

RH/Temperature SD Card Datalogger. Instruction Manual

USER INSTRUCTIONS - TP7C & TP9

INSTRUCTION MANUAL. Universal AC Input Switching Power Supply 24 Vdc Output DIN-Rail Models PSD1000, PSD1000F PSD PSD1000F

99 Washington Street Melrose, MA Fax TestEquipmentDepot.com DPI 820. Dual Input Thermometer. User manual.

IAQ-CALC TM Indoor Air Quality Meter

IAQ-CALC Indoor Air Quality Meter

IAQ-CALC TM Indoor Air Quality Meter

Platinum RTD Thermometer MODEL NO

L610 L620 L630 SIMPLE LOGGER THERMOCOUPLE MODULE E N G L I S H. User Manual

Platinum RTD Thermometer MODEL NO

HI HI N HI HI

Anemometer SD Card Datalogger. Instruction Manual

RS Stock No Instruction Manual RS Input Data Logging Thermometer

200A MICRO-OHMMETER ENGLISH. User Manual

CO2 Controller Operating Instructions Models: RAD-0501, RAD-0501A, RAD-0501E 1. Product Description

QUICK START USER GUIDE. Data Logger Model L452

374 FC/375 FC/376 FC. Clamp Meter. Safety Information

USER GUIDE. Differential Thermometer Datalogger. Model HD200

1. Introduction 2. Specifications 3. Symbol Definition and Button Location 4. Operation Instructions 5. Temperature Measurement 6.

Visible Light SD Card Datalogger

Installation and Maintenance Instructions for GC55 Differential Pressure Transducer Version 1.0 5/11

4-Channel Datalogging Thermometer

INSTRUCTION MANUAL. Model Dual Input RTD Thermometer. Measures two temperatures simultaneously. Dual RTD probe inputs

Sound Level Meter SLM700 AUDAC PROFESSIONAL AUDIO EQUIPMENT. Sound Level Meter SLM700. User Manual

4-Channel Thermometer

Indoor Air Quality. Thermohygrometer. Alnor Model TH720 AIRFLOW TM Model RH720. Operation and Service Manual


Indoor Air Quality Meter/Datalogger

Temperature & Humidity Datalogger

Instruction Manual. T3110 T3110L T3110Ex

SuperHeat Psychrometer Models RH350 and RH355 (kit)

Datalogging Conductivity/ TDS Meter. Instruction Manual

Heavy Duty Psychrometer + IR Thermometer Model HD500

User's Guide. Cup Thermo-Anemometer. Model AN400

IVC-222HPII Precision Voltage/Current Calibrator

User's Guide. Vibration Meter and Laser Combination Tachometer Model

AC/DC Current Probe Model SL201

-auto Power Off- T1/T2/T1-T2

User Guide. Indoor Air Quality Meter/Datalogger. Model EA80

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER

True RMS AC Voltage/Current Datalogger

Sound Level Pen Instruction Manual

ICAM501/502. .com. Installation & Maintenance Instructions & Operating Manual. ATEX/IECEx & Class1 Division1 Safety Instructions.

OX 5022-CK OX 5042-CK

VELOCICALC Air Velocity Meter

DOSTMANN electronic. Precision Measuring Instrument P600-EX/P605-EX/P650-EX/P655-EX. Operation manual

EPS Power Supply

Dissolved Oxygen Meter Kit DOK. Instruction Manual

Operation Manual. Concorde 600 Power Supply. *This instrument is intended for laboratory use only.

User's Guide. Extech AM A AC Analog Clamp Meter

SP6R Level Controller Operation Manual

INSTRUCTION MANUAL. Models: H100 Light Meter

AC/DC Current Probe Model SL206 USER MANUAL

INSTRUCTION MANUAL. Model True RMS AC/DC 30A Mini Clamp-on Meter. Introduction. True RMS AC Current and Voltage

Clamp-on Ground Tester Model 6417

Auto-timer Datalogging Temperature / Humidity Meter

TRUE TEMPERATURE INDICATOR MODEL TTI-10 User Manual/Handbook

Advanced Datalogging Thermometer. Instruction Manual

Operation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter

Copyright [W.L. Walker Co., Inc.]. All Rights Reserved

User Guide. Moisture Meter. Model MO250

Heavy Duty Datalogging Light Meter

User's Guide. EasyView K-Type Thermometer. Model EA11A

VELOCICALC AIR VELOCITY METER MODEL 5725

Operating Instructions LX-Series Digital Tension Meter

L605 SIMPLE LOGGER TEMPERATURE MODULE E N G L I S H. User Manual

SHUTTERBOSS VERSION II USER MANUAL

BETAGAUGE 301 Reference Manual

Precision Measuring Instrument P650-EX/P655-EX

Operating Instructions STX Series Digital Strap Tension Meter

LINE VOLTAGE TESTER CT101 USER S MANUAL. Please read this manual carefully and thoroughly before using this product.

Datalogging IAQ Meter Instruction Manual

DPL4000 Portable Low Range Dew Point Analyzer

Operation Manual MODEL 3173/3173R. MICROCOMPUTER BASED Conductivity/Salinity/TDS &Temperature BENCH METER 3173/3173R

TC2100 HOLD MAX AVG MIN SETUP THERMOMETER. K T1-T2 F C mv TYPE KJTERSN USB MAX MIN AVG MEN. K INTERVAL hour:min POWER T1 T2.

DELUXE FAST RESPONSE WATERPROOF DIGITAL STEM THERMOMETER USER S MANUAL

GE Infrastructure Sensing. Druck DPI 832. Electrical loop calibrator User manual - K393

VELOCICALC Air Velocity Meter

Energy Logger. User's Guide. Model EM Washington Street Melrose, MA Phone Toll Free

UÊ, Ê* - 1 Ê Ê Ê " /",Ê

9040/9040UK. Users Manual. Phase Rotation Indicator

RS Stock No Instruction Manual RS Input Data Logging Thermometer

Temperature Humidity Meter

ENGLISH. User manual


Water Quality Meter. Water Quality Meter Kit. Instruction Manual

User's Guide. MiniTec TM Series Model MN25 MultiMeter

POCKET MULTIMETER 5115 ENGLISH User Manual

Operation Manual. MODEL ph6175. Microcomputer Based ph/orp/temperature Benchtop Meter

UWE-1707 SERIES CONTENTS OPERATION MANUAL ELECTRONIC WEIGHING INDICATOR 1. SPECIFICATIONS 2. INSTALLATION

Transcription:

BetaProbe TI/TI+ High Accuracy Digital Thermometer User Reference Manual

1. Introduction The BetaProbe TI/TI+ is designed to be a high precision digital thermometer that can be used in place of mercury liquid in glass (LIG) thermometers. It can also be used as a reference standard to check the accuracy of other types of digital or analog thermometers. Because of its intrinsically safe certification it can be used in many of the same hazardous locations as analog thermometers. The probe houses a fast responding thin film (TI) or wire wound (TI+) sensor that allows very accurate readings to be taken quickly and with minimal insertion depth. A rotatable probe mount allows for both horizontal and vertical orientation of the display module. A unique feature of the BetaProbe TI/TI+ is the trend indication arrows which the user can configure to show when the readings are considered to be stable enough to record the reading. Additional features such as min/max storage, selectable sample rate, auto-shutoff, backlight and damping allow the user to tailor the TI/TI+ for their specific application. Battery life is up to 300 hours. With the auto-shutoff feature the typical user may only have to change the batteries a few times per year. 1.1 Customer Service Corporate Office: www.martelcalibrators.com e-mail: sales@martelcorp.com Tel: (603) 434-1433 800-821-0023 Fax: (603) 434-1653 1.2 Standard Equipment Martel Electronics 3 Corporate Park Drive Derry, NH 03038 Check that your BetaProbe TI/TI+ has arrived intact. Batteries are factory installed. Save the packing materials at least until you have verified that there is no concealed damage. 1.3 Safety information A Warning identifies conditions and actions that pose hazard(s) to the user; a Caution identifies conditions and actions that may damage the Calibrator or the equipment under test. 1

Symbol Symbols Used The following table lists the International Electrical Symbols. Some or all of these symbols may be used on the instrument or in this manual. Description Power OFF Power ON This unit was designed to meet the above requirements using: EN 60079-0:2006 ELECTRICAL APPARATUS FOR EXPLOSIVE GAS ATMOSPHERES -- PART 0: GENERAL REQUIREMENTS (IEC 60079-0:2004 (MOD)) EN 60079-11:2007 EXPLOSIVE ATMOSPHERES -- PART 11: EQUIPMENT PROTECTION BY INTRINSIC SAFETY "I" (IEC 60079-11:2006 (EQV) + CORRIGENDUM DEC. 2006 (EQV)) Earth ground Risk of Danger. Important information. Refer to manual. Battery Hazardous Voltage Conforms to ATEX requirements Conforms to relevant European Union directives. Wheeled bin, conforms to EC directive 2002/96/EC Hazard Location Information/Approvals Ex Hazardous Areas An Ex-hazardous area as used in this manual refers to an area made hazardous by the potential presence of flammable or explosive vapors. These areas are also referred to as hazardous locations. Ex ib IIB T4 Gb ( 10 C Ta +50 C) ITS10ATEX27114X Ex ib IIB T4 Gb II 2 G IECEx ITS10.0049X 0344 Manufactured by Martel Electronics, Inc., 3 Corporate Park Dr. Derry, NH, USA Special Conditions for Safe Use: Misuse Should the BetaProbe TI/TI+ be overheated or exposed to sudden physical shock (i.e. being dropped) it should be examined for any damage that may cause a safety concern. If in doubt please return the unit for evaluation to Martel Electronics. Please refer to the Customer Service Section for contact information. Check for proper operation of the instrument before taking it into a hazardous area. Should the instrument not operate as expected do not take the instrument into a hazardous area and return it to the manufacturer for inspection. The BetaProbe TI/TI+ is intended for installation only in locations providing adequate protection against the entry of solid foreign objects or water capable of impairing safety. Caution To avoid possible damage to calibrator or to equipment under test: If the message changes to OL the range limit is exceeded and the BetaProbe TI/TI+ must immediately be removed from the heat source to prevent damage. 2 3

2. BetaProbe TI/TI+ Display and Controls 3. Operation Insertion Depth and Usage The BetaProbe TI/TI+ should be inserted to a minimum depth of 2.5" to obtain specified accuracy of measurement. The final 2 inches of the probe should always be exposed to prevent high temperatures from damaging the instrument's electronics or housing. See illustration below. Item Description 1 Power Key Used to turn the unit on and off 2 Min/Max Key Used to toggle the main display value between min, max, and trend modes 3 Setup Key Used to change the various set-up parameters of the unit 4 Enter Key Used to enter data in set-up modes 5 Primary Display Used to display the temperature values in C or F. Also used in set-up mode. 6 Icon Area Used in regular and setup modes 7 Stability Display Used to show the trend and stability of data. 8 Arrow Keys Secondary functions of the MAX/MIN and SETUP keys. Used for data entry in the setup modes. 9 Backlight Key Secondary function of the key, turns backlight on or off. 10 Low Battery Display Flashes when the battery needs to be changed. For best results, allow ample time for the probe to stabilize before recording measurements. The stability display is helpful for this. Power Push the power button momentarily to turn the unit on or off. Normal Operation Controlling the Backlight Press the enter button to turn the backlight on or off. Controlling the Primary Data Display Use the MIN/MAX key to toggle the numeric display between the current reading, minimum, maximum and 1 minute trend values in either C or F as calculated since power up or since the last reset of the values. Press and hold the Min/Max key for 2 seconds until "CLR" is displayed to reset the MIN/MAX/TREND values. 4 5

Stabilty Display The stability display shows the general stability level relative to a user configurable stability limit. There are 4 preset settings for stability represented in C or F (0.01, 0.1, 1.0, or 10.0). When the limit is exceeded, it also shows the general direction in which the readings are presently moving. The stability level is calculated from a moving sample window of six seconds duration, with the value extrapolated to a 1 minute time base. Segments are turned on to indicate the stability level as follows: Center only (stable) when half of the 1 minute trend value is less than or equal to the stability limit, i.e. maximum deviation limit to +limit. Center plus one arrow up or down (depending on the trend) when half of the 1 minute trend value is greater than the stability limit and is less than or equal to twice the stability limit. Center plus two arrows up or down (depending on the trend) when half of the 1 minute trend value is greater than twice the stability limit and is less than or equal to three times the stability limit. Center plus three arrows up or down (depending on the trend) when half of the 1 minute trend value is greater than three times the stability limit. Set-up and Configuration Use the set-up key to enter the setup modes. Inside the setup modes the keys work as described below. MAX/MIN Exit to the main display. SETUP Step to the next menu item, exiting to the main display after the last item. Enter the data edit mode for the present menu item. During the data edit mode, use the and keys to scroll through the values. Use to exit back to the menu item. The following paragraphs describe the set-up items in the order in which they appear. Temperature Unit Selection The Temperature Unit Selection menu allows the user to chose the unit used when displaying the primary variable. Change the unit C or F. Change the unit C or F. Stability Limit The stability limit setting is used to determine when the secondary display will indicate that the reading is stable. There are 4 preset settings for stability (0.01, 0.1, 1.0, or 10.0) expressed in the presently selected units. Change the unit stability limit setting to the next greater setting. Change the unit stability limit setting to the next lesser setting. Auto Off Selection The Auto Off parameters control when the unit automatically turns off after inactivity on the keypad. Battery Life Increase the auto off setting and the primary data field, advancing from "OFF" to 1 minute and stopping at 30 minutes. Decrease the auto off setting and the primary data field, advancing from 1 minute to "OFF" and stopping at "OFF". Initially the primary data field shows the percentage of battery life that is remaining. This display can be toggled to show the battery voltage. This display is updated continually to reflect the present battery capacity. Press the key to toggle between the two data displays. Operating Temperature This selection shows the operating temperature of the unit head in C or F. Change the unit C or F. Change the unit C or F. Note: Should the head temperature rise above 50 C (122 F) an Overtemp warning will be displayed. 6 7

Damping The Damping function is a running average of readings used filter out "noisy" temperature sources. The choices are OFF, 2, 5, or 10 sample average. Change the damp setting to the next item in the list, cycling from the last to the first. Change the damp the previous item in the list, cycling from the first to the last. Start or Stop Data Logging The logging status shows "FULL" if the data logging memory is full, "OFF" if not presently logging data, or "ON" if presently logging data. Pressing allows you to change the logging status. or If not presently logging data, choose between "START" and "OFF". If presently logging data, choose between "STOP" and "ON". The present data logging mode is not changed until the key is pressed. Stop or Start data logging Sampling Rate The sampling rate determines how often the Beta Probe samples data. The settings are (0.5, 1.0, or 2.0), expressed in samples per second. Change the rate setting to the next item in the list, cycling from the last to the first. Change the rate setting to the previous item in the list, cycling from the first to the last.. Display Resolution Display resolution is the number of digits to the right of the decimal point. The settings are (0.1, 0.01, or 0.001). Change the resolution setting to the next item in the list, cycling from the last to the first. Change the resolution setting to the previous item in the list, cycling from the first to the last.. RS232 Communication The baud rate can be set to 2400 or 9600. Change the RS232 communication setting in the list, cycling from the last to the first. Change the RS232 communication setting to the previous item in the list, cycling from the first to the last.. Note: The following Data Logging functions can only be enabled with the optional BetaLOG-TI program which is available from Martel or your Martel distributor. Detailed information on the Logging functions is included with the BetaLOG software. Display Free Log Memory Initially the primary data field shows the percentage of log memory. This display can be toggled to show the number of free records. This display is updated continually during data logging to reflect the present memory capacity. Press the key to toggle between the two data displays. Logging Interval The key is ignored if presently logging data. Otherwise, when the key is pressed to enter the data edit mode. Change the logging interval setting to the next item in the list, cycling from the last to the first. Change the logging interval setting to the previous item in the list, cycling from the first to the last. return to the corresponding menu item. Send Logged Data The key is ignored if presently logging data. Otherwise, when the key is pressed, the user is allowed to: or Confirm or cancel the choice to send data. Send or cancel data. Erase Logged Data The key is ignored if the presently logging data. Otherwise, when the key is pressed, the user is allowed to: or Confirm or cancel the choice to erase logged data. Erase or cancel data erase. 8 9

4. Changing the Batteries Explosion hazard Batteries must only be changed in an area known to be non-hazardous. The RS-232 interface must not be used in hazardous areas. 5. Cleaning To clean the BetaProbe TI/TI+ use a cloth with a mild cleaning solution. 6. Calibration The BetaProbe TI/TI+ can be calibrated in the field by one of three methods depending on the extent of the calibration needed. 1) Single point Ro calibration (typically 0C) 2) Two-Point calibration Ro and gain (typically 100C). 3) A full 4 point calibration. For complete calibration instructions please contact Martel Electronics Corp. www.martelcalibrators.com Remove the battery door to expose the three AAA batteries. Replace only with AAA batteries approved in this document. Replace the battery door and reinstall the hex screw. Be sure to use only the approved list of batteries. Any substitution will void the safety ratings of this product. Battery Manufacturer Type Alkaline AAA Duracell LR03/MN2400 Alkaline AAA Rayovac LR03/824 Alkaline AAA Energizer LR03/E92 Alkaline AAA Panasonic LR03X Battery life Battery life is about 300 hours (12.5 days) of continuous use with the backlight off. There is a low battery icon in the lower right of the display. It will appear when battery level is low. Replace batteries per recommendations found in the specifications section of this manual. RS-232 Interface An RS-232 interface is standard on the BetaProbe TI/TI+. Serial communication can be used for configuration, calibration, and to transfer measurement data from the probe. When the data logging software is purchased, a RS-232 and USB cable is included. For customers wishing to do calibration only the RS-232 cable is available from Martel. Order P/N 1919069. 10 11

7. Specifications (23 ± 5 C unless noted) Measurement Range: BetaProbe TI: -50 C to 160 C (-58 F to 320 F) BetaProbe TI+: -50 C to 300 C (-58 F to 572 F) Accuracy (1 year): ± 0.06 C (0.1 F) Resolution: Selectable (0.1, 0.01, 0.001) factory default is 0.001 Sample Rate: User selectable 0.5/sec, 1/sec or 2/sec factory default is 1/sec Operating Temperature Range of Readout: -10 C to 50 C EMC Compliance: EN61326:2006 Annex C CISPR II, Edition 5.0-2009 Class B Humidity Range: 0 to 95% R.H. Non-condensing Storage Temperature Range: -20 to 60 C Enclosure Rating: IP50 Readout Temperature Coefficient: Add ± 10ppm/ C of full scale temperature from -10 C to 18 C and 28 C to 50 C Battery Save (auto-off) Range: Selectable from 1 to 30 minutes or can be disabled Size (readout only): Approximately 4"x2"x1" Probe Size: BetaProbe TI (1919876) 3/16 Diameter x 12 Length BetaProbe TI (1919918) 3/16 Diameter x 18 Length BetaProbe TI+ (1919877) 1/4 Diameter x 18 Length Weight: 6.9oz. 8. Warranty Martel Electronics Corporation warrants all products against material defects and workmanship for a period of twelve (12) months after the date of shipment. Problems or defects that arise from misuse or abuse of the instrument are not covered. This includes but is not limited to bent probes, broken cases, and blackened probe sheaths. If any product is to be returned, a Return Material Author ization form can be obtained from our website www.martelcorp.com under customer service. You can also call 1-800-821-0023 to have a form faxed. Martel will not be responsible for damage as a result of poor return packaging. Out of warranty repairs and recalibration will be subject to specific charges. Under no circumstances will Martel Electronics be liable for any device or circumstance beyond the value of the product. Substitution of components will impair suitability for hazardous locations. Power: 3 AAA alkaline batteries (Must use only approved batteries see section 4) Battery Life: Approximately 300 hours without back light 12 13

www.martelcalibrators.com e-mail: sales@martelcorp.com Tel: (603) 434-1433 800-821-0023 Fax: (603) 434-1653 Martel Electronics 3 Corporate Park Drive Derry, NH 03038 0219876 Rev. D 11/11