Getting Data. This guide will take you step by step, button click by button click, through using S1PXRF and your Tracer

Similar documents
Using the Bruker Tracer III-SD Handheld X-Ray Fluorescence Spectrometer using PC Software for Data Collection

Technical Procedure for X-Ray Fluorescence Spectrometer (XRF)

This presentation takes you through Region of Interest (ROI) Analysis.

X-MET8000 Series Operator s Manual

TraceFinder Analysis Quick Reference Guide

SVG3 (Basic Equipment)

MODULE 3. FACTORS AFFECTING 3D LASER SCANNING

REFLEX XRF V2.2. Frequently Asked Questions. Advanced instrumentation technology for greater productivity and precision.

Rietveld refinements collection strategies!

Provide Accurate Results. Operational Manual. Visible Spectrophotometer. SP-880 Mate. PC Software. Metertech Inc. Version 1.2.3

Bruker. SDD X-ray Detector

PerkinElmer Series 200 HPLC: Operating Procedure

Perkin-Elmer Lambda 3B Spectrophotometer

CommCare for Android Smartphones

Agilent 6400 Series Triple Quadrupole LC/MS System

Cornell Spectrum Imager (CSI) Open Source Spectrum Analysis with ImageJ Tutorial

The first thing we ll need is some numbers. I m going to use the set of times and drug concentration levels in a patient s bloodstream given below.

X-Lab Pro 5 Operator Manual

Overview. Experiment Specifications. This tutorial will enable you to

Manual for Jasco-IR-spectrometer

1/4 Specifications Hidex 300 SL automatic TDCR liquid scintillation counter

Cygnos360 V2 Installation Manual

Activity concentration for exempt material (Bq/g)

NORAN System 7 Adding value to the Microprobe

Using Excel for Graphical Analysis of Data

UCS-20. Operating Manual Macintosh Edition. Spectrum Techniques, Inc. Oak Ridge, Tennessee USA. Universal Computer Spectrometer.

TEL-218DRV TEL-X-Driver

Probe for EPMA Quick Start Instructions

Evex NanoAnalysis EDS Acquisition

Interpretype Bluetooth Setup Procedure

Vibration Analyzer Version 1.0.2

Customer ( you ) must review these notes prior to installing or operating the Attune NxT Software version ( Attune Software v1.1.0 ).

MalaRom flashing Guide

Multivariate Calibration Quick Guide

X-Ray fluorescence and Raman spectroscopy

Comprehensive, Low-Cost Gas Analysis Software for Increased Productivity. FabGuard Explorer. Gas Analysis Software

FlowAccess TM GETTING STARTED. Windows software for Skalar SAN plus systems. Version and up

sc1000 Software Updates from SD Card

Tutorial 7: Automated Peak Picking in Skyline

Cygnos360 V2 Installation Manual

Operating Procedure for Horiba Raman Microscope

Nicolet is50 FTIR User s Booklet

Hidex 300 SL automatic TDCR liquid scintillation counter

EDX Spectra. Used to set the amount of time for which the spectrum will be acquired (dwell), and to perform the following functions:

XRF for Regulatory Programs RoHS and Consumer Safety Compliance

UV / VIS SPECTROPHOTOMETER

GENESIS SPECTRUM USER S MANUAL. COPYRIGHT EDAX INC. ALL RIGHTS RESERVED EDAX INC. 91 McKEE DRIVE MAHWAH, NJ USA

Loudness Logger 1/12. Software Tools. Features of release and higher

Probe for EPMA: Software for Electron Probe MicroAnalysis

NMR INSTRUMENT INSTRUCTIONS: Safety and Sample Preparation

Thermo Scientific Spectronic 200 Visible Spectrophotometer. The perfect tool. for routine measurements

Agilent G2727AA LC/MS Data Browser Quick Start

Touch Screen Selection Guide

Scientific Production Company Doza Configurator Software

67 Series Spectrophotometer PC Software

M2500 Engine Controller Operator Manual

MPE+ Frequently Asked Questions & Troubleshooting

Jasco FP-6500 Spectrofluorimeter Updated November 14, 2017

Precautions The XRD uses x-ray radiation. The instrument is designed to prevent radiation exposure.

Migrating Label Formats from Legi for Windows 2.X to Legitronic Labeling Software 3.5.X

Using OPUS to Process Evolved Gas Data (8/12/15 edits highlighted)

AbraScan Test Strip Reader Digital Reading of Test Strips for the Determination of Cyanotoxins in Contaminated Water Samples

IDS 101 Introduction to Spreadsheets

Hardware and Installation

Model FP-6500 Spectrofluorometer Instruction Manual. FP-6500 for Windows

Cardiac Dual Energy CT: Technique

ML reconstruction for CT

HP AP5000 Installation Guide (XP)

Cygnos360 V2 Installation Manual

Appendix D: Equipment

QUANTAX EDS SYSTEM SOP

Enhanced material contrast by dual-energy microct imaging

The latest trend of hybrid instrumentation

Onwards and Upwards, Your near space guide. Figure 1. CheapBot Line Follower

Tutorial: FCAP Array Software with BD FACSArray Bioanalyzer

8 Setting Preferences

FRONTPAGE STEP BY STEP GUIDE

New Generation Explorer Handheld XRF. EXPLORER 7000 Mineral Ore Analyzer

the digital darkroom essential skills Chris Neylon

Haldia Institute of Technology. Touch Screen Monitors

Navigator Software User s Manual. User Manual. Navigator Software. Monarch Instrument Rev 0.98 May Page 1 of 17

Operating the COULTER COUNTER

Grazing Angle 2 Theta Phase Analysis

How to Analyze Materials

User Operation Manual

Generating a Custom Bill of Materials

SmartSoft AES Operator s Guide

ISB USER GUIDE HP EliteBook Folio 9470m. Laptop User Guide

XIV. XV. XVI. XVII. Manufacturer reserves the right to change the contents of this manual without notice.

Compatibility with graphing calculators 32 Deleting files 34 Backing up device files 35 Working with device screens 36 Capturing device screens 36

Copyright 2008 Casa Software Ltd.

MAESTRO 7. Multichannel Analyzer (MCA) Application Software

OCEAN OPTICS Chem 2000 Spectrophotometer

UV-VIS Spectrophotometer

Chapter 14 Speeding Up Windows Explorer 3

Fox Thermal Instruments, Inc.

Novo-Haze TX Transmission Hazemeter

Flo-Ware. Discovering. Model 260/268 Flo-Tote System Manual. Version 3.1. Instrument & Data Management System for Windows 95,98 & NT

Computer Basics: Step-by-Step Guide (Session 2)

Work with Shapes. Concepts CHAPTER. Concepts, page 3-1 Procedures, page 3-5

Transcription:

Getting Data You can easily obtain spectra from your Tracer using the S1PXRF software. This enables you to see your spectrum as it is gathered, and do both qualitative and quantitative analysis. This guide will take you step by step, button click by button click, through using S1PXRF and your Tracer

Getting Data Step 1 Setting up for Analysis

This is the home screen after launching S1PXRF. At this point, you are not connected to the instrument

The first step - and this is particularly important when opening the program for the first time, is to load a preexisting file. This way, you add in settings specific to your instrument. To do this, click File and then PDZ Preview...

Select a file run with your model of instrument, which is generally a Tracer III V+ or Tracer III SD. To see which instrument you have, look at your serial number.

The name of the file you have selected will appear in the window above A spectrum will appear in red You can select other files in the PDZ Preview menu

You are not yet connected to your instrument, to do that you will need to tell the software where to find it The options for this menu can be found under the DownLoad menu

First, select Baud Rate Make sure you select a baud rate appropriate to your instrument. For Tracer III V+ and Tracer III/IV SD models, 115200 is appropriate. For older models, smaller baud rates may be appropriate

To connect to your instrument, you will need to select the communications port (e.g. USB port) it is plugged into.

If you are not connected to the right port, then you will receive a message like this

You will need to open up Devices and Printers which can be found either in the Start menu or in Control Panel The Tracer will be under the Unspecified category, and will be listed as Prolific USB-to- Serial Com Port. The number underneath, in this case Com3, is the port your Tracer is located. Remember this number

To change the Com Port number, click on the current setting

This will open up a small window where you can enter in a new com port number

Enter the correct number as identified in the Devices and Printer menu. In this example, that number was 3

A Start button will appear as well Once you are connected, the red dot will turn green

Once connected, navigate to Setup and then Instrument Setup

A window will open up next Back Scatter, when selected, provides a safety feature - your instrument will shut off automatically if X-rays are not all accounted for. However, this can also prevent measurements for small objects. You can temporarily turn it off for some measurements Settings on the left are set when you open an existing PDZ file

In this case, Back Scatter is turned off After this, you can click Done Selecting PC Trigger will give your computer control over the instrument - this is the best setting for gathering data

The next step for setting up your Tracer is navigating to Tube, KTI Tube, and Read. This will open up a menu that allows you to change energy, current, and filter parameters on your Tracer

Here, you have a range of energies in kev. These determine what elements you are able to see The Anode Current determines how many photos you send into your sample The Filter (for Tracer IV systems) determines what filter is being used in your measurement 1 - Yellow Filter (1mil Al, 1 mil Ti) 2 - No Filter 3 - Red Filter (1 mil Al, 1 mil Ti, 1 mil Cu) 4 - Green Filter (1 mil Al, 1 mil TI, 6 mil Cu) 5 - Blue Filter (1 mil Ti)

Select PC Trigger to turn on the beam in your Tracer

With PC Trigger on, you can see the voltage (kev) and current (μa) update to confirm your settings Here, we have 40 kev and 14.8 μa selected

You can see your voltage and current change after making a new selection Select a new setting by clicking one of the buttons

Your chosen settings should be visible here Now, you are ready to begin taking measurements

Getting Data Step 2 Collecting Data

Place an object on the nose of the Tracer and click Start to begin a measurement. The Start button will instantly turn to a Stop button A spectra will appear a fraction of a second later (you will see a red light come on your Tracer first).

You can click Stop to end a measurement at any time As the Tracer measures, you can see a red counter in the bottom left of the screen increase - this is how may seconds you have been measuring As you measure, the spectra will become more and more smooth as error is averaged out

After clicking Stop, the Start button will return After stopping your measurement, you the counter will turn black and keep the total amount of time you were measuring visible Here, more time results in a cleaner measurement

Keep an eye on your count rates. Too many, and your Tracer will not analyze the data correctly. For Tracer II/IV S systems, make sure the count rate is below 100,000. For Tracer II/IV V+ systems, make sure it is below 10,000. In this case, we have a Tracer IV SD, so we are fine with a count rate of around 13,000.

Click the <> button to expand your window to focus on specific elements. You can also press the Up and Down buttons on your keyboard. You can also click (or on tablets touch) and spread your spectra as well

Click the ID menu item to open up a window that gives multiple options for analyzing spectra

Click Elem to open up a periodic table The periodic table is the way in which we navigate the spectra

K-alpha for Zirconium If we click on Zr, or Zirconium, a blue line will mark where that element lies K-beta for Zirconium

You can label the k-alpha peak by clicking the Add button

The element currently selected is displayed in the top left. In this case, it is Strontium You can click the Z- button to move down the spectrum K-alpha for Strontium K-beta for Strontium

You can label the k-alpha peak by clicking the Add button for multiple elements

You can click the Home button to return to see the whole spectrum To take a controlled measurement, select Timed then Timed Assays...

A window will then open up You can select the amount of time you wan to assay. Typically 2-3 minutes is used, but sometimes 5-10 minutes can be used for important measurements

Always select to save as a PDZ file - this is the file format for other Bruker software. TXT and CSV can be used in Excel and other programs - but this is the raw data, it may only be useful to advanced users Make sure Autosave is selected After you have selected settings, press the OK button

You will next be prompted to save the file. Pick a descriptive name, and make sure to keep data organized - you will gather a lot of it Click Save when done

Your measurement will begin. You will see a countdown from your selected time here

When you are finished, the total measurement time will be displayed in black

If we want to compare two measurements, we can go to Setup and Spectrum Overlay...

This will open a new window to the right. Click Move A >> B to move the current spectra to the background

After moving A to B, the original spectra will turn blue.

Once again, you can go to Timed Assay and save a file. Make sure all conditions are the same (energy, current, filter, time, and atmosphere)

You will see the new spectra (red) grow over the old spectra (blue)

Getting Data Step 3 Navigating and Analyzing Spectra

The spectra may not match exactly, as there are many things that can affect the measurement, including differences in composition and density

In this case, we are going to normalize to the Compton Peak - in Tracers this peak can be found at 18.5 kev. The Compton peak is determined by the elastic scatter of Rhodium, and it can be a proxy for the density of light elements (like Carbon, Oxygen, and Nitrogen) To account for differences, you can use the Normalize button in the Spectrum Overlay window.

After normalization, the area you selected will be matched, and the rest of the spectra adjusted accordingly When you click the Normalize button, it will change to NO Normalize

Click the Add button once you have selected an element Once you have selected and labeled the element, it s peak will be identifiable To begin analyzing your data, select an element from the periodic table. You can begin with a light element, such as Silica, and work your way up

You can move up the periodic table by clicking the Z+ button

Here, we can see a difference between the two elements - the presence of Potassium in the iphone (blue spectrum) and its lower abundance in the hotel glass (red spectrum). Potassium can make glass stronger and more resistant to scratches

The hotel glass, in contrast, has more Calcium

The next element at first glance seems to be Titanium - but the peaks do not match

In case of a mismatch, check the L shell for a heavier element It is always good to check the L shell. In this case, the element is actually Barium

Both the iphone glass (in blue) and the hotel glass (in red) have comparable amounts of iron

They also have comparable amounts of copper

Click on the home button to return to the entire spectrum

Here, it can be seen that the hotel (drinking) glass, in red, has Arsenic present. It is not uncommon to find unexpected elements

One prominent spectral peak will never match an element - that is the Compton peak located at 18.5 kev.

Rhodium will always be in your spectrum, as it is the x-ray source in the Tracer.

Palladium will be present as well, it is the collimator around the detector

Tin is present in the iphone (in blue), and absent from the hotel glass (in red)

Earlier, the L-shell of Barium was visible in the hotel glass (red spectrum). It is also visible in the Ks-shell. Barium is the beginning of the heavy elements for which you switch to the L-shell for best detection

Barium L-shell Barium K-shell

To remove the blue background spectrum, click Clear B in the Spectrum Overlay window

Getting Data Step 4 Quantifying Data

In this metal, we can see clearly defined peaks

These include Chrome, Iron, Nickel, and Molybdenum

We can quantify these results using Select Coef (SRZ, CFZ, or PMZ file) in the Setup menu

This will open a window to find a calibration file. Navigate to your instrument s serial number (in this case, T3S2430) The calibration files for metals are located in the Empirical Cals folder

Since Iron is the dominant element in this spectra, we will use the FE1.CFZ file

We can click the Conc button, now available in the menu bar, to obtain our quantified results

In the first column, we have the elements included in the calibration

The second column includes mathematical parameters used to estimate elemental weight percents

The third column includes the weight percents. In this case, we have ~23% Crome, 65% Iron, ~4% Molybdenum, ~2% Nickel and ~2% Manganese, with other smaller concentrations of other elements

We can select the cells in this window and copy them into Excel, Word, or any other software program