Digital thermometer-hygrometer with external temperature probe Instruction manual

Similar documents
Digital thermo-hygro-barometer with external probe Manual for use

Temperature logger with display Instruction Manual

Dual channel temperature logger with display Instruction Manual

Dual channel temperature logger with two binary inputs and LCD display Instruction Manual

Dual channel temperature logger with two voltage inputs 0-5V Instruction Manual

Temperature logger with internal sensor Instruction manual

Programmable transmitter of temperature, relative humidity and other derived humidity values with 0 to 10 V outputs Instruction Manual

4-Channel Datalogging Thermometer

AM3 Humidity. Temperature. Dew Point Indicator INSTRUCTION MANUAL

Hygro-Thermometer / Datalogger

Temperature and humidity regulator H3020 H3021 H3023 H3027. Instruction Manual

DTM-321 REL HOLD T2 K-TYPE TEMP: 20 C 60 C 140 F C F HUMIDITY: %RH POWER-UP OPTIONS

4-Channel Thermometer

Picture of the instrument

RS Stock No Instruction Manual RS Input Data Logging Thermometer


1. Introduction 2. Specifications 3. Symbol Definition and Button Location 4. Operation Instructions 5. Temperature Measurement 6.

Version 12/2013 EN. User Manual Radon Scout/Radon Scout Plus Radon logging

HygroLog HL20 and HL21 Data Logger. User Guide

C F MAX MIN TEMP:-20 C~ 60 C -4 F~140 F HUMIDITY: 0 ~ 100RRH HOLD TIMER T2 T1 RS-232 DC9V. h:m:m:s HUMIDITY TEMPERATURE METER POWER - UP OPTIONS

4-Channel Thermometer / Datalogger

LEGO Energy Meter. How to Get Started

HygroLog HL20 and HL21 Data Logger. User Guide

Supplied with. Calibration certificate. HD 200 Hygrometer

Model. Instruction Manual. WBGT Heat Stress Meter. reedinstruments. www. com

HygroPalm 0. Portable Humidity Temperature Indicator. Instruction Manual

RS Stock No Instruction Manual RS Input Data Logging Thermometer

Hand-held thermometer Model CTH7000

HDT-318 Thermo-Hygrometer with Data Logger. (Air Humidity/Temperature) Instruction Manual

N9002 THERMOMETER OPERATING INSTRUCTIONS

Model Instruction Manual. Heat Stress Meter. reedinstruments. www. com

BESANTEK USB Data Logger Instructional Manual (BST-DL10, BST-DL13, BST-DL14)

is then retained absolutely without interruption.

Programmable transmitter of temperature, relative humidity and other derived humidity values with RS485 serial output Instruction Manual

High Performance Infrared Thermometer with High D:S Ratio, Adjustable Emissivity, and Built-in Laser Sighting

Programmable temperature transducer from Pt1000 sensor to current loop 4-20 ma Instruction Manual

12-Channel Thermocouple Datalogger

Moisture Content Meter Plus Dew Point, Wet Bulb, and Thermocouple Temperature functions

GB/UK. Operating manual. Data logger EBI USB

GREISINGER electronic GmbH D Regenstauf, Hans-Sachs-Straße 26

RMS Mini Wireless Data Logger

User Guide. Indoor Air Quality Meter/Datalogger. Model EA80

Type K/J Thermometers

RMS Mini Wireless Data Logger

Battery operated recorders. Built-in battery operated GSM modem Text message alerts Data collecting to Comet database software and Comet cloud

NHT-310. Broadband EM Field Meter: DC 40 GHz. HAVSCo LTD. Key Features:

Light Meter SD Card real-time datalogger

CT-510. Multifunctional CCTV Tester. User Manual

GREISINGER electronic GmbH D Regenstauf, Hans-Sachs-Straße 26

24 V DC typ. 0,65 W. Liquid level with conductive probes (type SK1, SK5) Resistance measurement E1-E0, E2-E0, E3-E0, E4-E0

Operating Instructions

INSTRUCTION MANUAL. Model Dual Input RTD Thermometer. Measures two temperatures simultaneously. Dual RTD probe inputs

TEMPERATURE AND HUMIDITY TRANSDUCER P18 TYPE USER S MANUAL

ph/orp Meter SD Card real-time Datalogger

AVM-8880 USB Logging Anemometer

PV Remote Unit A 1378 Instruction manual Version 2.1, Code no

Indoor Air Quality Meter/Datalogger

1450 First Ave. Page 1 of 8 Chippewa Falls, WI (800) or (715) Fax: (715)

DP C10.UART CONTENT Content Normal indication area Introduction Button definition Appearances and dimensions Normal operation Specifications

Dual Moisture Meter Pin / Pinless Moisture Meter

TEMPERATURE / HUMIDITY /DEW POINT METER

Model Instruction Manual. Digital Pocket Psychrometer. reedinstruments. www. com

Humidity Temperature Indicator. Instruction Manual

SD Model. Instruction Manual. Humidity/Temperature Datalogger. reedinstruments

DIGITAL POCKET HEAT INDEX MONITORS USER S MANUAL

TC41-Thermocouple Thermometers User Manual

Hand-Held Pressure Indicator

CATALOG OF MEASURING INSTRUMENTS

USER GUIDE. Differential Thermometer Datalogger. Model HD200

USER GUIDE. AC/DC Magnetic Meter. Model MF100

HI HI HI HI Portable Microprocessor Thermistor Thermometers

DMT340 Series Dewpoint and Temperature Transmitters for Very Dry Conditions

Manual. Meter for BlueSolar DUO 12/24V-20A For RVs, Caravans and boats

TRANSDUCER T4311, T4411

For Auto Locksmith Association only!!! Emergency Service Manual.

Supplied with. Calibration certificate. TM 200 Thermometer

User manual. paper moisture meter humimeter RH5 with sword sensor

AWA6223S Sound Level Calibrator. Instruction Manual HANGZHOU AIHUA INSTRUMENTS CO., LTD

LEGO Energy Meter. How to Get Started

RE48ATM12MW on-delay timing relay s..300 h V AC - 2 OC

DPL4000 Portable Low Range Dew Point Analyzer

Handheld Odor Meter OMX-ADM. Operation Manual. Ver

OWNER S MANUAL 9908-TE. HIGH PRECISION AUTO-RANGING DC/True RMS AC BENCH-TOP DIGITAL MULTIMETER

MID approval no. DE-08-MI004-PTB004 Date of release

Hand-held thermometer Model CTH6200

1450 First Ave. Page 1 of 11 Chippewa Falls, WI (800) or (715) Fax: (715)

Laboratory Accreditation Programmes

User Manual ELECTRONIC SCALE

Trend-1 Vibration Data Logger. User Manual

Humidity + Temp. + Dew Point Pocket size HUMIDITY METER. Model : HT-315

4-Channel Vibration Datalogger

Operation Manual for Multifunctional Data Logger. HUATO Electronic Co., LTD.

RH/Temperature SD Card Datalogger. Instruction Manual

Datalogging IAQ Meter Instruction Manual

Anemometer SD Card Datalogger. Instruction Manual

HI ma Simulator

PHOENIX CONTACT - 01/2007

Industrial Power Supplies

EasyLog 21CFR Compatible Temperature, Humidity & Dew Point Data Logger with LCD Screen

Really Performs! Here is a Hand-Held Data Logger that Royalton Rd. Cleveland, OH USA Tel: (440) Fax: (440)

Transcription:

COMMETER C3631 Digital thermometer-hygrometer with external temperature probe Instruction manual

Manual for use of thermometer-hygrometer COMMETER C3631 with external temperature probe Instrument is designed for measurement of temperature and relative humidity (attached sensors) and temperature from an external probe. It enables to display the dew point temperature and the difference between external temperature and dew point temperature. Measured values are displayed on a dual line LCD display. Temperature is measured by RTD sensor Ni1000/6180ppm. Instrument compares measured values of temperature, humidity, dew point and temperature difference with two adjustable levels for each measured quantity. Breaking the level is indicated by blinking the proper value on display and by audio indication (switchable). Instrument is equipped with minimum and maximum memory and Hold function. Minimum and maximum values and Hold value are possible to display on the LCD anytime. Technical parameters: Parameters of measurement: Ambient temperature: Range of measurement: -10 to +60 C Resolution: 0.1 C Accuracy: ±0.4 C External probe temperature: Range of measurement: -50 to +250 C Resolution: 0.1 C Accuracy including probe: ±0.4 C from -50 to +100 C ±0.5% from reading from 100 to 200 C Humidity: Range measurement: 5 to 95 %RH Resolution: 0.1 %RH Accuracy: ±2.5 % RH at the range of 5 to 95 %RH Dew point temperature (calculated from temperature and humidity): Range: -40 to +60 C Resolution: 0.1 C Accuracy: ± 1.5 C at ambient temperature T < 25 C and RV > 30 %, for more details see Appendix A Measuring interval and display reading refresh: approximately 0.7 s in FAST mode 0.7 to 5 s in dynamic mode Power: battery 9V or ac/dc adapter 12V with NiCd accumulator 9V Average current consumption: 0.2 to 0.8 ma (depending on operation mode) Typical life of Zinc-Chloride battery: 4 months Typical life of Alkali-Mangan battery: 6 months Operation conditions: Ambient temperature range: -10 to +60 C Ambient relative humidity range: 5 to 95 %RH Outer characteristics in accordance with EN 33-2000-3: normal environment with characteristics AD1, AE1, AF1, AG1, AH1, AK1, AL1, AN1, AP1, AQ1, AR1, AS1, BA1, BE1 Not allowed manipulation: it is not allowed to touch sensors under the cover to avoid sensors damaging or to effect calibration. The sensors (under the cover) should not be exposed to direct contact with water or other liquids. Storing conditions: temperature -10 to +60 C relative humidity 5 to 95 %RH Dimensions: 141 x 71 x 27 mm Weight including battery: approximately 150 g 2 IE-COM-C3631-10

Battery life depends on selected display refresh mode (see below). In FAST mode display is refreshed in shortest possible interval with highest current consumption. In dynamic mode display is refreshed in interval up to 5 s in case measured values remain stable. Refresh interval is shortened to approximately 0.7 s only if measured values change. Current consumption in this mode in usual operation is lower, battery life is up to 4 times longer. The FAST mode is recommend to use only in cases, when slower display response is not acceptable. Battery voltage drop below 7 V is indicated with blinking of "BAT" in default display mode (displaying of actual values) and FAST mode is automatically canceled to save the battery. At the same time audio indication of alarms is automatically switched OFF. If instrument is powered from external ac/dc adapter, internal 9V battery is replaced with rechargeable NiCd accumulator. In usual operation from adapter accumulator is charged only with small current. If accumulator is totally discharged, its full charging in instrument takes approximately 100 hours. Instrument with accumulator is not recommended for permanent operation without ac/dc adapter plugged. Accumulator works only as a standby source in case of power mains failure. Switching ON and OFF the instrument If you wish to use the external temperature probe, connect it before instrument switching ON. Switch ON the instrument by pressing ON/OFF key. After switching ON the instrument all symbols on the LCD are displayed. If the ON/OFF key is being held pressed, all LCD symbols are displayed till the key is released. In usual operation instrument then starts the measurement mode and actual measured values are displayed. It is possible to switch OFF the instrument anytime and all setting will be saved. Setting will not be saved if battery is disconnected for long than 1 minute. Displaying of actual measured values In this mode is instrument anytime after switching ON. It is possible to enter this mode from other modes by pressing or by repeating pressing of MENU key. Ambient temperature in C is displayed on the upper LCD line (symbol INT) and relative humidity (%RH) is displayed on the lower LCD line. Press DISPLAY key to display other readings - external probe temperature in C (symbol EXT) and dew point temperature on lower line ( C DP). If external probe is not connected properly or temperature is out of measuring range, symbol -- is displayed. Press DISPLAY key to display other readings - external probe temperature in C (symbol EXT) and difference between external temperature and dew point temperature (symbol DIF). The difference is defined as t dif = t ext - t dp. IE-COM-C3631-10 3

Function HOLD (storing of actual measured values) and minimum a maximum memory Press HOLD key in the default mode (displaying of actual measured value) to store actual measured values to internal memory (indicated by short beep). Anytime it is possible to display stored values from MENU (see below). Each pressing of the HOLD key in the default mode causes values stored in HOLD memory are replaced with actual ones. Switched ON instrument permanently updates minimum and maximum memory of each measured values. Press MIN key (resp. MAX key) in the default mode to display minimum (resp. maximum) reading. These minimum and maximum readings are indicated by MIN (MAX) symbols on the LCD. Press DISPLAY key to display minimum (resp. maximum) value of other values. Pressing MIN (MAX) or MENU key again to return to default mode. Minimum and maximum memory is cleared from menu after confirmation selection CLR (see below). Values in HOLD, MIN and MAX memories remain stored even after instrument is switched OFF. Functions and settings available from menu Press MENU key to enter mode of viewing menu items one by one. Press arrow keys up and down to list all menu items. Press MENU key again to return to default mode (displaying of actual measured values). Pressing the ENTER key enables to enter the mode of setting alarm limits for all quantities (see below). This item indicates if audio signaling of alarm indication is switched on (On) or switched off (OFF). Press ENTER key to change actual setting. Notice: if the battery voltage is low, audio indication is out of operation to reduce current consumption independently on this selection. Clearing of minimum and maximum memory of all values. Memory is cleared after pressing ENTER key. Clearing is confirmed by reading YES on the LCD lower display. Press ENTER key to display values stored in the HOLD memory. Press DISPLAY key to display other stored values (dew point temperature). Press MENU key to leave this mode. Battery voltage of partially loaded battery is displayed. This value illustrates battery condition. 4 IE-COM-C3631-10

Display refresh mode is indicated. In the FAST mode refreshment is fastest with regular interval approximately 0.7 s. In the dynamic refresh mode (DYN.) each 15 s refresh interval of display is doubled to maximum 5 s if measured values are stable. If measured values change, refresh interval decreases to approximately 0.7 s. This dynamic mode prolongs battery life significantly. Select the desired mode by ENTER key. Notice: if battery voltage is low, the FAST mode is out of operation to reduce current consumption independently on this selection. Each pressing of ENTER key causes displaying of service information on software version (upper LCD line) together with instrument configuration on the LCD lower line. Alarm indication and setting It is possible to set lower and upper limit for each measured quantity. Breaking of the limit is indicated by blinking of the appropriate value on the display. If a new alarm was indicated (i.e. it was not active in the previous measurement), display starts to display the value out of limits. If at least one alarm is active, audio indication can be activated, if menu AUDI "On" is selected (see setting described above). Alarm activation of each value can be disabled by setting lower alarm limit of the desired value up to its maximum. This is indicated by OFF reading at the position of numeric value. Value of upper limit of the same alarm is indifferent. To set alarms press MENU key, select ALAR from menu items and confirm by pressing ENTER key. Reading ALLO indicates adjusted lower alarm limit (here air temperature). Set the desired value by means of the arrow keys. Press and hold the arrow key UP to make value increase fast. Press and hold the arrow key DOWN to make value decrease fast. Release the arrow key and press ENTER to confirm new limit. Reading ALHI indicates adjusted upper limit of the same alarm (here air temperature). Set the desired value in the same way as in above lower limit. If needed it is possible to get back to lower limit setting of the same alarm by pressing DISPLAY key. Press ENTER key to confirm new upper limit. Then you are offered to set alarm of other input value (here relative humidity). The procedure is the same as the above temperature limit setting. Alarm activation of each value can be disabled by setting lower alarm limit of the desired value up to its maximum. This is indicated by OFF reading at the position of numeric value. Value of upper limit of the same alarm is indifferent. IE-COM-C3631-10 5

Instruments further offers alarm setting of other measured quantities. It is possible to leave the alarm setting mode by pressing MENU key. New adjusted limits up to pressing MENU key are stored in memory. After pressing ENTER key it is possible to set alarm limits for dew point temperature. Battery replacement Low battery voltage is indicated on the display with blinking reading "BAT". It is necessary to replace it with new one as soon as possible. Battery is located under small cover on the instrument lower side. It is absolutely necessary to replace battery with instrument switched OFF, otherwise setting of d.ref. and AUDI (from menu selections) and data in memory HOLD, MIN and MAX will be lost. For the same reason do not disconnect the battery for longer than 1 minute even if instrument is switched OFF. If it happens (or if battery is totally discharged), it is necessary to set again in appropriate menu selection LCD refreshment mode (d.ref.), alarm audio indication (AUDI) and clear the minimum and maximum memory (CLR). Commeter instruments passed the following electromagnetic compatibility (EMC) tests: Device conforms in accordance with EN 61326-1 these norms: radiation: EN 55022 class B immunity: EN 61000-4-2 (levels 4/8 kv, class A) EN 61000-4-3 (intensity of electromagnetic field 3 V/m, class B) EN 61000-4-4 (levels 1/0,5 kv, class A) EN 61000-4-6 (intensity of electromagnetic field 3 V/m, class B) EN 61000-4-11 (class A) EN 61000-4-5 (class A) Appendix A - Accuracy of dew point measurement 6 IE-COM-C3631-10