Image Display DICOM Conformance Statement EasyViz 7.2

Similar documents
Image Display DICOM Conformance Statement

Digital Lightbox 1.0

POP-Net Essential DICOM Conformance Statement

MediPlus TM PACS&RIS Version 2.6

Version 7 DICOM Conformance Statement. Document Version 3.02, June 2009

Version 9 DICOM Conformance Statement. Version 3.05, September 2015

Technical Publication. DICOM Conformance Statement. DICOM Proxy 4.0. Document Revision 11. May 18, Copyright Brainlab AG

Technical Publication. DICOM Conformance Statement. iplan 3.0. Document Revision 1. November 17, Copyright BrainLAB AG

Technical Publication. DICOM Conformance Statement. Patient Browser 2.1. Document Revision 2. April 13, Copyright BrainLAB AG

VM PACS DICOM Conformance Statement

DICOM Conformance Statement Radiotherapy Document Revision 2 July 1, Copyright Brainlab AG

Visage. DICOM Conformance Statement

DICOM Conformance Statement Surgery Document Revision 4 August 26, Copyright Brainlab AG

DICOM Conformance Statement Product Name HCP DICOM Net Version

Caseaccess 1.0 DICOM Conformance Statement

Sep, th Edition 897N101668H

1 CONFORMANCE STATEMENT OVERVIEW

AGFA HEALTHCARE DICOM Conformance Statement

Copyright FUJIFILM Corporation, Japan. August, th Edition 897N100760F

Mirada XD DICOM Conformance Statement

Inobitec DICOM Server (PACS) Conformance Statement. Version: 1.4.x

JiveX Web-enabled solutions for medical imagecommunication

Mirada XD DICOM Conformance Statement

ClearCanvas Workstation DICOM Conformance Statement. Document Version: 1.5

SecurView RT/TechMate Workstation DiCom Conformance statement

Punctual Dicom Workstation

Technical Publication. DICOM Conformance Statement. Trauma 3.0. Document Revision 2. October 5 th, 2010

Parascript AccuDetect CAD Software System

efx Software DICONDE Conformance Statement

DICOM Conformance Statement RT Elements Document Revision 3 January 19, Copyright Brainlab AG

AGFA HEALTHCARE DICOM Conformance Statement

DICOM V3.0 Conformance Statement. SenoIris 1SP2

DICOM 3.0 Conformance Statement

DICOM Conformance Statement RT Elements Document Revision 5 December 14, Copyright Brainlab AG

AGFA HEALTHCARE DICOM Conformance Statement

Patterson DICOM Imaging. DICOM Conformance Statement

McKesson Radiology EXP1

DICOM Conformance Statement

DICOM Conformance Statement Application: Linkverse DICOM Provider 2.0

SecurView DX Workstation DICOM Conformance Statement

COPYRIGHT VITAL IMAGES, INC ALL RIGHTS RESERVED

DICOM CONFORMANCE STATEMENT

ClearCanvas Workstation DICOM Conformance Statement

ClearCanvas Workstation DICOM Conformance Statement

DICOM Conformance Statement. CharruaPACS

CoActiv, LLC CoActiv Medical Business Solutions EXAM-PACS Conformance Statement

Carl Zeiss Surgical GmbH

DICOM Conformance Statement * /02* /02 MADE IN GERMANY

Carl Zeiss Surgical GmbH

DBSWIN DICOM Conformance Statement. DBSWIN DD-DICOM Interface. DICOM Conformance Statement V2.1

Carl Zeiss Meditec AG

Conformance Statements for DICOM PaCentric Connect / Traveler

DICOM CONFORMANCE STATEMENT FOR DIAGNOSTIC ULTRASOUND SYSTEM MODEL SSA-640A V5.0

Health Engine DICOM Conformance Statement. Internal document number: Date:

Technical Publications

Software Version 6.0 DICOM Conformance Statement

DICOM Conformance Statement

NumaLink-3.0 DICOM 3.0 Conformance Statement

DICOM. Conformance Statement EPIQ 7 and 5 Release 1.8.x Affiniti 70, 50 and 30 Release 1.8.x Rev A 24 April 2017

DICOM Conformance Statement for Selenia Dimensions Acquisition Workstation Software Version 1.8 Part Number MAN Revision 002

AVIA (Dx MM) DICOM 3.0 Conformance Statement

Carl Zeiss Meditec AG

Documento n Revision 1.5 Dicom Conformance Statement del sistema PACS di AGFA

AGFA HEALTHCARE DICOM Conformance Statement

AGFA HEALTHCARE DICOM Conformance Statement

DICOM Conformance Statement. CR-VW 674 for DICOM Storage DICOM Print DICOM MWM DICOM MPPS DICOM Media Storage

DICOM Conformance Statement

DICOM CONFORMANCE STATEMENT FOR VANTAGE-GALAN TM VERSION V5.0

COPYRIGHT VITAL IMAGES, INC ALL RIGHTS RESERVED

PACSware Migration Toolkit (MDIG)

Technical Publication. DICOM Conformance Statement. BrainSUITE NET 1.5. Document Revision 1. March 5, Copyright BrainLAB AG

Technical Publications

DICOM Conformance Statement

Matrix Server and Client DICOM Conformance Statement

NumaStore 1.0 DICOM 3.0 Conformance Statement

DICOM Conformance Statement

Abstract DCV DICOM CONFORMANCE. XrayVision DCV DICOM Conformance Statement. Abstract

Technical Publications

Cover will be supplied by PMG MARCOM group.

Technical Publications

1 Introduction Scope and audience References Acronyms and abbreviations 3. 2 Implementation model... 4

DICOM Conformance Statement. Xcelera R4.1 L1SP1

C-DAC s Medical Informatics Software Development Kit (SDK) for DICOM PS Conformance Statement

DICOM Conformance Statement

DICOM CONFORMANCE STATEMENT. BrainLAB PatientBrowser 1.0

DICOM CONFORMANCE STATEMENT FOR PET/CT SCANNER. Celesteion TM V6.4 OR LATER (PCA-9000A)

DICOM Conformance Statement

SIEMENS. DICOM Conformance Statement

DICOM Conformance Statement iconnect Access v5.0

DICOM. Conformance Statement EPIQ 7, 5 and CVx Release 4.0.x Affiniti 70, 50 and 30 Release 4.0.x Rev A October 15, 2018

DICOM Conformance Statement. Version: 2.X

HCP DICOM Net DICOM Conformance Statement

RamSoft. PowerServer. DICOM Conformance Statement

Diagnostics Workstation 3-0-X DICOM Conformance Statement Part Number MAN-00418

Technical Publications

dicom PACS 5 Version 5.2

SYNAPSE 3D Conformance Statement FUJIFILM SYNAPSE 3D V3.0. Conformance Statement. Revision Z45N

Philips Medical Systems DICOM Conformance Statement. EasyWeb 2.0. Document Number April 1999

AGFA HEALTHCARE DICOM Conformance Statement

InstaPACS / InstaRISPACS / InstaViewer (v Release) DICOM Conformance Statement. Meddiff Technologies Pvt. Ltd.

Transcription:

Image Display DICOM Conformance Statement EasyViz 7.2 April 11, 2016 Document ID: 2016.04.035 Prepared By: Karos Health

Table Of Contents 1. Document History... 5 2. Conformance Statement Overview... 6 3. Introduction... 8 3.1 Audience... 8 3.2 Remarks... 9 3.3 Terms and Abbreviations... 9 3.4 References... 10 4. Networking... 11 4.1 Implementation Model... 11 4.1.1 Application Data Flow... 11 4.1.2 Functional Definitions of AE s... 12 4.1.2.1 Functional Definition: Query/Retrieve... 12 4.1.2.2 Functional Definition:... 13 4.1.2.3 Functional Definition:... 13 4.1.2.4 Functional Definition: Print Management... 13 4.1.2.5 Functional Definition: General Purpose Worklist... 13 4.1.3 Sequencing of Real World Activities... 14 4.1.4 AE Specification: Query/Retrieve... 17 4.1.4.1 SOP Classes... 17 4.1.4.2 Association Policies... 17 4.1.4.3 Association Initiation Policy... 18 4.1.4.4 Association Acceptance Policy... 19 4.1.5 AE Specification:... 19 4.1.5.1 4.1.5.2 SOP Classes... 19 Association Policies... 21 4.1.5.3 4.1.5.4 Association Initiation Policy... 22 Association Acceptance Policy... 25 4.1.6 AE Specification:... 25 4.1.6.1 4.1.6.2 SOP Classes... 25 Association Policies... 27 4.1.6.3 4.1.6.4 Association Initiation Policy... 28 Association Acceptance Policy... 28 4.1.7 AE Specification: Print Management... 39 4.1.7.1 4.1.7.2 SOP Classes... 39 Association Policies... 39 4.1.7.3 4.1.7.4 Association Initiation Policy... 40 Association Acceptance Policy... 40 4.1.8 AE Specification: General Purpose Worklist... 41 4.1.8.1 SOP Classes... 41 4.1.8.2 Association Policies... 41 4.1.8.3 Association Initiation Policy... 41 4.1.8.4 Association Acceptance Policy... 42 4.2 Network Interfaces... 42 Image Display DICOM Conformance Statement Page 2 of 60

4.2.1 Physical Network Interface... 42 4.2.2 Additional Protocols... 42 4.3 Configuration... 42 4.3.1 AE Title/Presentation Address Mapping... 42 4.3.1.1 Local AE Titles... 42 4.3.1.2 Remote AE Title/Presentation Address Mapping... 43 4.3.2 Parameters... 43 5. Media... 44 5.1 Implementation Model... 44 5.1.1 Application Data Flow Diagram... 44 5.1.2 Functional Definitions of AE s... 45 5.1.2.1 Functional Definition: Media FSR... 45 5.1.2.2 Functional Definition: Media FSC... 45 5.1.3 Sequencing of Real World Activities... 45 5.1.4 File Meta Information for Implementation Class and Version... 46 5.2 AE Specifications... 46 5.2.1 AE Specification: Media FSR... 46 5.2.1.1 File Meta Information: Media FSR... 47 5.2.1.2 Real-World Activities... 47 5.2.2 AE Specification: Media FSC... 49 5.2.2.1 File Meta Information: Media FSC... 49 5.2.2.2 Real-World Activities... 49 5.3 Augmented and Private Application Profiles... 50 5.4 Media Configuration... 50 6. Support of Character Sets... 50 6.1 Overview... 50 6.2 Character Sets... 50 6.3 Character Set Configuration... 51 7. Security... 51 7.1 Security Profiles... 51 7.2 Association Level Security... 51 7.3 Application Level Security... 51 8. Annexes... 52 8.1 IOD Contents... 52 8.1.1 Created SOP Instance(s)... 52 8.1.1.1 Key Object Selection Document IOD... 52 8.1.1.2 Grayscale Softcopy Presentation State IOD... 53 Image Display DICOM Conformance Statement Page 3 of 60

8.1.1.3 Common Modules... 53 8.1.1.4 Key Object Selection Document Modules... 54 8.1.1.5 Grayscale Softcopy Presentation State Modules... 56 8.1.2 Usage of Attributes of Received IOD s... 60 8.1.3 Attribute Mapping... 60 8.1.4 Coerced/Modified Fields... 60 8.2 Data Dictionary of Private Attributes... 60 8.3 Coded Terminology and Templates... 60 8.4 Grayscale Image Consistency... 60 8.5 Standard Extended/Specialized/Private SOP Classes... 60 8.5.1 MI 3D Softcopy Presentation State Private SOP Class... 60 8.6 Private Transfer Syntaxes... 60 Image Display DICOM Conformance Statement Page 4 of 60

1. Document History Revision Date Author Changes 2010-07-15 NHK Initial draft for EasyViz 3.0.4 Added support for: Multi-frame Grayscale Byte Secondary Image Multi-frame Grayscale Word Secondary Capture Multi-Frame True Color Secondary Capture 2011-01-28 TS Updated for 3.2 Added VL (retired) for. Added JPEG 2000 Lossless Only and JPEG 2000 Lossless and transfer syntaxes to accepted transfer syntaxes. Updated Implementation Version Name and Implementation Class UID. 2011-04-19 JBB Updated for 3.2. Added Description of GPWL support 2011-11-29 TS Updated for 4.0. No actual changes except document ID, software version and references to other 4.0 documents. 2013-04-25 MWN Updated for 5.0. No actual changes except document ID, software version and references to other 5.0 documents. 2013-05-07 MWN Converted document into new template 2014-08-05 MWN Updated for 6.0. No actual changes except software version and references to other 6.0 documents. 2015-09-14 MWN Updated for 7.0. No actual changes except software version and references to other 7.0 documents. 2016-04-11 Martin Wallengren Nilsson Changed Document Template Image Display DICOM Conformance Statement Page 5 of 60

2. Conformance Statement Overview The EasyViz Image Display implements the necessary DICOM services to query, retrieve and display CR, CT, DX, MG, MR, PT, RF, SC, US, and XA images. In addition, EasyViz Image Display may be used to: print DICOM instances to a network attached hardcopy device via the General 2D CAM, import DICOM instances from a CD-R/DVD via the EasyViz PACS Workflow Manager, or export DICOM instances to a CD-R/DVD via the EasyViz PACS Workflow Manager. The DICOM Networking Services supported by the EasyViz Image Display are listed in Table 1. Networking SOP Classes Transfer User of Service () Provider of Service () Hardcopy Grayscale SOP Class (Retired) Yes Yes Hardcopy Color SOP Class (Retired) Yes Yes Computed Radiography Yes Yes Digital X-Ray For Presentation Yes Yes Digital X-Ray For Processing Yes No Digital Mammography X-Ray For Presentation Yes Yes Digital Mammography X-Ray For Processing Yes No Digital Intra-oral X-Ray For Presentation Yes Yes Digital Intra-oral X-Ray For Processing Yes No CT Yes Yes Enhanced CT Yes Yes Ultrasound Multi-frame (Retired) Yes Yes Ultrasound Multi-frame Yes Yes MR Yes Yes Enhanced MR Yes Yes Nuclear Medicine (Retired) Yes Yes Ultrasound (Retired) Yes Yes Ultrasound Yes Yes Secondary Capture Yes Yes Multi-frame Grayscale Byte Secondary Capture Yes Yes Multi-frame Grayscale Word Secondary Capture Yes Yes Image Display DICOM Conformance Statement Page 6 of 60

Networking SOP Classes User of Service () Provider of Service () Multi-frame True Color Secondary Capture Yes Yes Grayscale Softcopy Presentation State SOP Class Yes Yes Color Softcopy Presentation State SOP Class Yes Yes Pseudo-Color Softcopy Presentation State SOP Class Yes Yes X-Ray Angiographic Yes Yes Enhanced XA Yes Yes X-Ray Radioflouroscopic Yes Yes Enhanced XRF Yes Yes Nuclear Medicine Yes Yes VL Endoscopic Yes Yes Video Endoscopic Yes Yes VL Microscopic Yes Yes Video Microscopic Yes Yes VL Slide-Coordinates Microscopic Yes Yes VL Photographic Yes Yes Video Photographic Yes Yes Ophthalmic Photographic 8 Bit Yes Yes Ophthalmic Photographic 16 Bit Yes Yes Basic Text SR Yes Yes Enhanced SR Yes Yes Comprehensive SR Yes Yes Mammography CAD SR Yes Yes Key Object Selection Document Yes Yes Positron Emission Tomography Yes Yes MI 3D Softcopy Presentation State Yes Yes Query/Retrieve Study Root Q/R Information Model FIND Yes No Image Display DICOM Conformance Statement Page 7 of 60

Networking SOP Classes User of Service () Provider of Service () Study Root Q/R Information Model MOVE Yes No Print Management Basic Grayscale Print Management Meta SOP Class Yes No Presentation LUT SOP Class Yes No Basic Annotation Box SOP Class Yes No General Purpose Worklist General Purpose Worklist Information Model - FIND Yes No General Purpose Scheduled Procedure Step SOP Class Yes No General Purpose Performed Procedure Step SOP Class Yes No Table 1: Network Services supported by EasyViz Image Display. The DICOM Media Services supported by the EasyViz Image Display are listed in Table 2. Media Application Profile Compact Disk Recordable Write Files (FSC/FSU) Read Files (FSR) General Purpose CD-R Yes Yes DVD General Purpose on DVD-RAM Media Yes Yes Table 2: Media Services supported by EasyViz Image Display. 3. Introduction 3.1 Audience This document is intended for hospital staff, health system integrators, software designers or implementers. It is assumed that the reader has a working understanding of DICOM. Image Display DICOM Conformance Statement Page 8 of 60

3.2 Remarks DICOM, by itself, does not guarantee interoperability. However, the Conformance Statement facilitates a first-level validation for interoperability between different applications supporting the same DICOM functionality. This Conformance Statement is not intended to replace validation with other DICOM equipment to ensure proper exchange of information intended. The scope of this Conformance Statement is to facilitate communication with Medical Insight A/S and other vendors medical equipment. The Conformance Statement should be read and understood in conjunction with the DICOM Standard [DICOM]. However, by itself it is not guaranteed to ensure the desired interoperability and a successful interconnectivity. The user should be aware of the following important issues: The comparison of different conformance statements is the first step towards assessing interconnectivity between Medical Insight A/S and non-medical Insight A/S equipment. Test procedures should be defined to validate the desired level of connectivity. The DICOM Standard will evolve to meet the users future requirements. Medical Insight A/S is actively involved in developing the standard further and therefore reserves the right to make changes to its products or to discontinue its delivery. 3.3 Terms and Abbreviations Definitions, terms and abbreviations used in this document are defined within the different parts of the DICOM standard. A list of abbreviations and terms can be seen in Table 3. Term AE AET BSPS CAD CD-R CR CT DICOM DX FSC FSU FSR Description DICOM Application Entity. Application Entity Title. Blending Softcopy Presentation State. Computer Aided Diagnostics. CD Recordable. Computed Radiography. Computed Tomography. Digital Imaging and Communications in Medicine. Digital X-Ray. File Set Creator. File Set Updater. File Set Reader. Image Display DICOM Conformance Statement Page 9 of 60

Term GPWL GSDF GSPS MG MR PT RF SC SOP US XA Description General Purpose Work List Grayscale Standard Display Function. Grayscale Presentation State. Mammography. Magnetic Resonance. Positron Emission Tomography. X-Ray Radioflouroscopic. Secondary Capture. DICOM Service Class Provider (DICOM Server). DICOM Service Class User (DICOM Client). DICOM Service-Object Pair. Ultrasound. X-Ray Angiographic. Table 3: Abbreviations and Terms. 3.4 References Reference ID Description Digital Imaging and Communications in Medicine (DICOM), NEMA PS 3.1-3.20, 2015c. National Electrical Manufacturers Association (NEMA), 1300N 17th Street, Rosslyn, Virginia 22209, USA. Image Display DICOM Conformance Statement Page 10 of 60

4. Networking 4.1 Implementation Model The EasyViz Image Display DICOM Networking Services are implemented in a number of processes launched and terminated by the user. The EasyViz PACS Workflow Manager, launched at user login, provides DICOM Services to query and transfer DICOM Instances via the Query/Retrieve and the Application Entities. The EasyViz CAMs, which are launched from the EasyViz PACS Workflow Manager, provides DICOM Services to retrieve and present DICOM Instances via the Query/Retrieve and Application Entities. Moreover, the EasyViz General 2D CAM provides DICOM Services to store Grayscale Softcopy Presentation States, Key Objects and Secondary Captures via the Application Entity and DICOM services related to hardcopy printing via the Print Management Application Entity. The EasyViz General 3D CAM provides DICOM Service to store private SOP-class 3D softcopy presentation states via the Application Entity. 4.1.1 Application Data Flow Application data flow diagram for the EasyViz Image Display can be seen on Figure 1. Image Display DICOM Conformance Statement Page 11 of 60

Local User Requests Transfer Image Display Requested Instances Sent to Remote AE Local User Queries Local Database Local User Requests Retrieval Image Display Query/Retrieve Remote AE Receives Query or Retrieve Command Image Display Remote AE Sends Requested Instances Local User Requests Print Image Display Print Management Requested Formatted Instances Sent to Remote AE DICOM Standard Interface Figure 1: Application Data Flow Diagram. 4.1.2 Functional Definitions of AE s 4.1.2.1 Functional Definition: Query/Retrieve The Query/Retrieve Application Entity provides DICOM Services to query and retrieve DICOM Instances from remote DICOM Application Entities. The Query/Retrieve Application Entity does not always query a remote DICOM Application Entities about images before retrieving the images. The Query/Retrieve may also use the information in the internal database to obtain information on available studies. The internal database is updated by EasyViz VPS. Image Display DICOM Conformance Statement Page 12 of 60

4.1.2.1.1 The Query/Retrieve Application Entity Provides DICOM Services to: - Query and retrieve DICOM Instances from remote DICOM Application Entities using the DICOM Query/Retrieve SOP (Acting as ). 4.1.2.2 Functional Definition: The Application Entity provides DICOM Services to transfer DICOM Instances to a remote DICOM Application Entity. This Application Entity is intended to be used to transfer DICOM Instances to the local image display, i.e. transfer images from a local CD-ROM/DVD drive to the local archive. Note: the EasyViz Image Display supports transfer of DICOM Instances, which are not supported for displayed, i.e. the is capable of transferring Nuclear Medicine images, which currently cannot be displayed. 4.1.2.2.1 The Application Entity Provides DICOM Services to: Transfer DICOM Instances to a remote DICOM Application Entity using a number of DICOM Image SOPs (Acting as ). 4.1.2.3 Functional Definition: The Application Entity provides DICOM Services to receive images and other DICOM Instances from remote DICOM Application Entities. 4.1.2.3.1 The Application Entity Provides DICOM Services to: Receive DICOM Instances from remote DICOM Application Entities using a number of DICOM Image SOPs (Acting as ). 4.1.2.4 Functional Definition: Print Management The Print Management Application Entity provides DICOM Services to print grayscale images including Presentation LUTs and Annotations on a remote DICOM Application Entity. 4.1.2.4.1 The Print Management Application Entity Provides DICOM Services to: Print Grayscale Images incl. an optional Presentation LUT to remote DICOM Application Entities using the DICOM Basic Grayscale Print Management Meta SOP Class (Acting as ) and the DICOM Presentation LUT SOP Class (Acting as ). Print annotations to DICOM peer Application Entities using the DICOM Basic Annotation Box SOP (Acting as ). 4.1.2.5 Functional Definition: General Purpose Worklist The GPWL Application Entity provides DICOM Services to work with General Purpose Scheduled Procedure Steps and related entities 4.1.2.5.1 The GPWL Application Entity Provides DICOM Services to: Perform a general purpose worklist C-FIND against a worklist. Claim/unclaim Scheduled-Procedure-Step. Perform an N-set to change state of an existing Scheduled-Procedure-Step. Create and update Performed-Procedure-Step related to a claimed SPS using N-SET and N-GET. Image Display DICOM Conformance Statement Page 13 of 60

4.1.3 Sequencing of Real World Activities The sequencing constraints for the EasyViz Image Display Application Entities can be seen on Figure 2, Figure 3 and Figure 4. The Query/Retrieve may either query the internal DB or a Peer AE for information on SOP Instances, so either the transaction marked by *) is used or the transaction marked by **), see Figure 2. The Print Management will not update an instance of the Basic Grayscale Image Box SOP Class before it has been received by the Image Display. If collation is disabled, the Print Management will send the two transactions marked by *), otherwise the two transactions marked by **) are sent see Figure 3. Peer Query/Retrieve Peer Image Display Image Display Query/Retrieve *) Query Internal DB for information related to SOP Instances **) Query Peer AE for information related to SOP Instances **) Peer AE returns information related to SOP Instances Request Retrieval of SOP Instances Notification of Instances to be sent to C-MOVE Destination AE in Response Requested Instances sent to C-MOVE Destination AE Figure 2: and Query/Retrieve : Sequencing Constraints. Image Display DICOM Conformance Statement Page 14 of 60

Peer Peer Print Management Image Display Print Management Image Display Transfer of SOP Instances Printer N-GET Film Session N-CREATE Presentation LUT N-CREATE Film Box N-CREATE Image Box N-SET Annotation Box N-SET *) Film Box N-ACTION *) Film Box N-DELETE **) Film Session N-ACTION **) Film Session N-DELETE Printer N-EVENT-REPORT Figure 3: and Print Management : Sequencing Constraints. Image Display DICOM Conformance Statement Page 15 of 60

Figure 4: General Purpose Worklist : Sequencing constraints. Image Display DICOM Conformance Statement Page 16 of 60

4.1.4 AE Specification: Query/Retrieve AE Specifications 4.1.4.1 SOP Classes The Query/Retrieve provides Standard Conformance to the SOP Classes listed in Table 4. SOP Class Name SOP Class UID Query/Retrieve Study Root Q/R Information Model - FIND Study Root Q/R Information Model MOVE 1.2.840.10008.5.1.4.1.2.2.1 Yes No 1.2.840.10008.5.1.4.1.2.2.2 Yes No 4.1.4.2 Association Policies Table 4: Query/Retrieve : Supported SOP Classes. 4.1.4.2.1 General The Query/Retrieve will propose Association Requests for DICOM Query/Retrieve Service. The DICOM standard application context name for DICOM 3.0 is always proposed, see Table 5. Application Context Name 1.2.840.10008.3.1.1.1 Table 5: Query/Retrieve : DICOM Application Context. 4.1.4.2.2 Number of Associations The Query/Retrieve does not support multiple simultaneous associations, see Table 6. Maximum number of simultaneous associations 1 Configurable. Table 6: Query/Retrieve : Number of simultaneous associations. 4.1.4.2.3 Asynchronous Nature The Query/Retrieve does not support asynchronous communication. Multiple outstanding transactions are not supported, see Table 7. Maximum number of outstanding asynchronous transactions 1 Configurable. Table 7: Query/Retrieve : Asynchronous nature. 4.1.4.2.4 Implementation Identifying Information The identifying information for the Query/Retrieve can be seen in Table 8. Implementation Class UID 1.3.6.1.4.1.16978.0.3.5.4.28 Implementation Version Name EV-3.5.4-28 Table 8: Query/Retrieve : DICOM Implementation Class and Version. Image Display DICOM Conformance Statement Page 17 of 60

4.1.4.3 Association Initiation Policy 4.1.4.3.1 Activity: Query Instances 4.1.4.3.1.1 Description and Sequencing of Activities The Query/Retrieve sends an Association Request to a remote DICOM Application Entity acting as Query/Retrieve when the user queries a PACS from the EasyViz Workflow Manager. Note that by default, worklist in the EasyViz Workflow Manager are populated using the internal database. Sequencing constraints for the Query Instances activity can be seen in Figure 2. 4.1.4.3.1.2 Proposed Presentation Contexts The Query/Retrieve will propose the Presentation Context listen in Table 9: Query/Retrieve : Proposed Presentation Contexts for the Query Instances activity. for the Query Instances activity. Abstract Syntax Transfer Syntax Role Ext. Neg Name UID Name List UID List Study Root Q/R Information Model FIND 1.2.840.10008.5.1.4.1.2.2.1 Implicit VR Little.2 Table 9: Query/Retrieve : Proposed Presentation Contexts for the Query Instances activity. Relational-retrieval 4.1.4.3.1.3 Extended Negotiation The Query/Retrieve will try to negotiate Relational-Retrieval with the Peer Application Entity acting as Query/Retrieve. If the Query/Retrive fails to negotiate Relational-Retrieval, then only baseline behavior is performed. 4.1.4.3.2 Activity: Retrieve Instances 4.1.4.3.2.1 Description and Sequencing of Activities The Query/Retrieve sends an Association Request to a remote DICOM Application Entity acting as Query/Retrieve when the user launches a CAM from the EasyViz Workflow Manager. Sequencing constraints for the Retrieve Instances activity can be seen in Figure 2. 4.1.4.3.2.2 Proposed Presentation Contexts The Query/Retrieve will propose the Presentation Contexts listed in Table 10 for the Retrieve Instances activity. Abstract Syntax Transfer Syntax Role Ext. Neg Name UID Name List UID List Study Root Q/R Information Model MOVE 1.2.840.10008.5.1.4.1.2.2.2 Implicit VR Little.2 Table 10: Query/Retrieve : Proposed Presentation Contexts for the Retrieve Instances activity. Relational-retrieval 4.1.4.3.2.3 Extended Negotiation The Query/Retrieve will try to negotiate Relational-Retrieval with the Peer Application Entity acting as Query/Retrieve. If the Query/Retrive fails to negotiate Relational-Retrieval with the Query/Retrieve, then only baseline behavior is performed. Image Display DICOM Conformance Statement Page 18 of 60

4.1.4.4 Association Acceptance Policy The Query/Retrieve does not accept associations. 4.1.5 AE Specification: 4.1.5.1 SOP Classes The provides Standard Conformance to the SOP Classes listed in Table 11. SOP Class Name SOP Class UID Transfer Hardcopy Grayscale SOP Class Hardcopy Color SOP Class 1.2.840.10008.5.1.1.29 Yes No 1.2.840.10008.5.1.1.30 Yes No Computed Radiography 1.2.840.10008.5.1.4.1.1.1 Yes No Digital X-Ray For Presentation Digital X-Ray For Processing Digital Mammography X-Ray Image For Presentation Digital Mammography X-Ray Image For Processing Breast Tomosynthesis For Presentation Digital Intra-oral X-Ray For Presentation Digital Intra-oral X-Ray For Processing 1.2.840.10008.5.1.4.1.1.1.1 Yes No 1.2.840.10008.5.1.4.1.1.1.1.1 Yes No 1.2.840.10008.5.1.4.1.1.1.2 Yes No 1.2.840.10008.5.1.4.1.1.1.2.1 Yes No 1.2.840.10008.5.1.4.1.1.13.1.3 Yes No 1.2.840.10008.5.1.4.1.1.1.3 Yes No 1.2.840.10008.5.1.4.1.1.1.3.1 Yes No CT 1.2.840.10008.5.1.4.1.1.2 Yes No Enhanced CT 1.2.840.10008.5.1.4.1.1.2.1 Yes No Ultrasound Multi-frame (Retired) 1.2.840.10008.5.1.4.1.1.3 Yes No Ultrasound Multi-frame 1.2.840.10008.5.1.4.1.1.3.1 Yes No MR 1.2.840.10008.5.1.4.1.1.4 Yes No Enhanced MR 1.2.840.10008.5.1.4.1.1.4.1 Yes No Image Display DICOM Conformance Statement Page 19 of 60

SOP Class Name SOP Class UID Nuclear Medicine (Retired) 1.2.840.10008.5.1.4.1.1.5 Yes No Ultrasound (Retired) 1.2.840.10008.5.1.4.1.1.6 Yes No Ultrasound 1.2.840.10008.5.1.4.1.1.6.1 Yes No Secondary Capture 1.2.840.10008.5.1.4.1.1.7 Yes No Multi-frame Grayscale Byte Secondary Capture Multi-frame Grayscale Word Secondary Capture Image Store Multi-frame True Color Secondary Capture Grayscale Softcopy Presentation State SOP Class Color Softcopy Presentation State SOP Class Pseudo-Color Softcopy Presentation State SOP Class 1.2.840.10008.5.1.4.1.1.7.2 Yes No 1.2.840.10008.5.1.4.1.1.7.3 Yes No 1.2.840.10008.5.1.4.1.1.7.4 Yes No 1.2.840.10008.5.1.4.1.1.11.1 Yes No 1.2.840.10008.5.1.4.1.1.11.2 Yes No 1.2.840.10008.5.1.4.1.1.11.3 Yes No X-Ray Angiographic 1.2.840.10008.5.1.4.1.1.12.1 Yes No Enhanced X-Ray Angiographic Image 1.2.840.10008.5.1.4.1.1.12.1.1 Yes No X-Ray Radioflouroscopic 1.2.840.10008.5.1.4.1.1.12.2 Yes No Enhanced X-Ray Radioflouroscopic 1.2.840.10008.5.1.4.1.1.12.2.1 Yes No Nuclear Medicine 1.2.840.10008.5.1.4.1.1.20 Yes No VL Endoscopic 1.2.840.10008.5.1.4.1.1.77.1.1 Yes No Video Endoscopic 1.2.840.10008.5.1.4.1.1.77.1.1.1 Yes No VL Microscopic 1.2.840.10008.5.1.4.1.1.77.1.2 Yes No Video Microscopic 1.2.840.10008.5.1.4.1.1.77.1.2.1 Yes No VL Slide-Coordinates Microscopic Image 1.2.840.10008.5.1.4.1.1.77.1.3 Yes No VL Photographic 1.2.840.10008.5.1.4.1.1.77.1.4 Yes No Video Photographic 1.2.840.10008.5.1.4.1.1.77.1.4.1 Yes No Ophthalmic Photographic 8 Bit Image 1.2.840.10008.5.1.4.1.1.77.1.5.1 Yes No Image Display DICOM Conformance Statement Page 20 of 60

SOP Class Name SOP Class UID Ophthalmic Photographic 16 Bit Image 1.2.840.10008.5.1.4.1.1.77.1.5.2 Yes No Basic Text SR 1.2.840.10008.5.1.4.1.1.88.11 Yes No Enhanced SR 1.2.840.10008.5.1.4.1.1.88.22 Yes No Comprehensive SR 1.2.840.10008.5.1.4.1.1.88.33 Yes No Mammography CAD SR 1.2.840.10008.5.1.4.1.1.88.50 Yes No Key Object Selection Document 1.2.840.10008.5.1.4.1.1.88.59 Yes No Positron Emission Tomography Image 1.2.840.10008.5.1.4.1.1.128 Yes No MI 3D Softcopy Presentation State 1.3.6.1.4.1.16978.3.1 Yes No Table 11: : Supported SOP Classes. 4.1.5.2 Association Policies 4.1.5.2.1 General The will propose Association Requests for DICOM Services. The DICOM standard application context name for DICOM 3.0 is always proposed, see Table 12. Application Context Name 1.2.840.10008.3.1.1.1 Table 12: : DICOM Application Context. 4.1.5.2.2 Number of Associations The does not support multiple simultaneous associations, see Table 13. Maximum number of simultaneous associations 1 Configurable. Table 13: : Number of simultaneous associations. 4.1.5.2.3 Asynchronous Nature The does not support asynchronous communication. Multiple outstanding transactions are not supported, see Table 14. Maximum number of outstanding asynchronous transactions 1 Configurable. Table 14: : Asynchronous nature. 4.1.5.2.4 Implementation Identifying Information The identifying information for the can be seen in Table 15. Implementation Class UID 1.3.6.1.4.1.16978.0.3.5.4.28 Implementation Version Name EV-3.5.4-28 Table 15: : DICOM Implementation Class and Version. Image Display DICOM Conformance Statement Page 21 of 60

4.1.5.3 Association Initiation Policy 4.1.5.3.1 Activity: Transfer Instances 4.1.5.3.1.1 Description and Sequencing of Activities The sends an Association Request to a remote DICOM Application Entity acting as when the selects import in the EasyViz Workflow Manager. This is intended to be used to import selected DICOM Instances from CD-ROM/DVD to the local archive. 4.1.5.3.1.2 Proposed Presentation Contexts The will propose the Presentation Contexts listed in Table 16 for the Transfer Instances activity. Abstract Syntax Transfer Syntax Role Ext. Neg Name UID Name List UID List Hardcopy Grayscale SOP Class Hardcopy Color Image SOP Class Computed Radiography Digital X-Ray Image For Presentation Digital X-Ray Image For Processing Digital Mammography X-Ray For Presentation Digital Mammography X-Ray For Processing Breast Tomosynthesis For Presentation Digital Intra-oral X-Ray For Presentation Digital Intra-oral X-Ray For Processing 1.2.840.10008.5.1.1.29 Implicit VR Little 1.2.840.10008.5.1.1.30 Implicit VR Little 1.2.840.10008.5.1.4.1.1.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.1.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.1.1.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.1.2 Implicit VR Little 1.2.840.10008.5.1.4.1.1.1.2.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.13.1.3 Implicit VR Little 1.2.840.10008.5.1.4.1.1.1.3 Implicit VR Little 1.2.840.10008.5.1.4.1.1.1.3.1 Implicit VR Little CT 1.2.840.10008.5.1.4.1.1.2 Implicit VR Little.2.2.2.2.2.2.2.2.2.2.2 Image Display DICOM Conformance Statement Page 22 of 60

Abstract Syntax Transfer Syntax Role Ext. Neg Name UID Name List UID List Enhanced CT Image 1.2.840.10008.5.1.4.1.1.2.1 Implicit VR Little Ultrasound Multi-frame 1.2.840.10008.5.1.4.1.1.3 Implicit VR Little (Retired) Ultrasound Multi-frame 1.2.840.10008.5.1.4.1.1.3.1 Implicit VR Little MR 1.2.840.10008.5.1.4.1.1.4 Implicit VR Little Enhanced MR Image Nuclear Medicine (Retired) Ultrasound Image (Retired) Ultrasound Image Secondary Capture Multi-frame Grayscale Byte Secondary Capture Multi-frame Grayscale Word Secondary Capture Image Store Multi-frame True Color Secondary Capture Grayscale Softcopy Presentation State SOP Class Color Softcopy Presentation State SOP Class Pseudo-Color Softcopy Presentation State 1.2.840.10008.5.1.4.1.1.4.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.5 Implicit VR Little 1.2.840.10008.5.1.4.1.1.6 Implicit VR Little 1.2.840.10008.5.1.4.1.1.6.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.7 Implicit VR Little 1.2.840.10008.5.1.4.1.1.7.2 Implicit VR Little 1.2.840.10008.5.1.4.1.1.7.3 Implicit VR Little 1.2.840.10008.5.1.4.1.1.7.4 Implicit VR Little 1.2.840.10008.5.1.4.1.1.11.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.11.2 Implicit VR Little 1.2.840.10008.5.1.4.1.1.11.3 Implicit VR Little.2.2.2.2.2.2.2.2.2.2.2.2.2.2 Image Display DICOM Conformance Statement Page 23 of 60

Abstract Syntax Transfer Syntax Role Ext. Neg Name UID Name List UID List SOP Class X-Ray Angiographic Enhanced X-Ray Angiographic Image X-Ray Radioflouroscopic Enhanced X-Ray Radioflouroscopic Nuclear Medicine VL Endoscopic Image Video Endoscopic VL Microscopic Image Video Microscopic VL Slide-Coordinates Microscopic Image VL Photographic Image Video Photographic Ophthalmic Photographic 8 Bit Ophthalmic Photographic 16 Bit 1.2.840.10008.5.1.4.1.1.12.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.12.1.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.12.2 Implicit VR Little 1.2.840.10008.5.1.4.1.1.12.2.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.20 Implicit VR Little 1.2.840.10008.5.1.4.1.1.77.1.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.77.1.1.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.77.1.2 Implicit VR Little 1.2.840.10008.5.1.4.1.1.77.1.2.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.77.1.3 Implicit VR Little 1.2.840.10008.5.1.4.1.1.77.1.4 Implicit VR Little 1.2.840.10008.5.1.4.1.1.77.1.4.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.77.1.5.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.77.1.5.2 Implicit VR Little.2.2.2.2.2.2.2.2.2.2.2.2.2.2.2 Basic Text SR 1.2.840.10008.5.1.4.1.1.88.11 Implicit VR Little Image Display DICOM Conformance Statement Page 24 of 60

Abstract Syntax Transfer Syntax Role Ext. Neg Name UID Name List UID List Enhanced SR 1.2.840.10008.5.1.4.1.1.88.22 Implicit VR Little Comprehensive SR 1.2.840.10008.5.1.4.1.1.88.33 Implicit VR Little Mammography CAD SR 1.2.840.10008.5.1.4.1.1.88.50 Implicit VR Little Key Object Selection Document Positron Emission Tomography Image MI 3D Softcopy Presentation State 1.2.840.10008.5.1.4.1.1.88.59 Implicit VR Little 1.2.840.10008.5.1.4.1.1.128 Implicit VR Little 1.3.6.1.4.1.16978.3.1 Implicit VR Little.2.2.2.2.2.2.2 Table 16: : Proposed Presentation Contexts for the Transfer Instances activity. 4.1.5.3.1.3 Extended Negotiation No extended negotiation is performed. 4.1.5.4 Association Acceptance Policy The does not accept associations. 4.1.6 AE Specification: 4.1.6.1 SOP Classes The provides Standard Conformance to the SOP Classes listed in Table 17. SOP Class Name SOP Class UID Transfer Hardcopy Grayscale SOP Class (Retired) Hardcopy Color SOP Class (Retired) 1.2.840.10008.5.1.1.29 No Yes 1.2.840.10008.5.1.1.30 No Yes Computed Radiography 1.2.840.10008.5.1.4.1.1.1 No Yes Digital X-Ray For Presentation 1.2.840.10008.5.1.4.1.1.1.1 No Yes Image Display DICOM Conformance Statement Page 25 of 60

SOP Class Name SOP Class UID Digital Mammography X-Ray Image For Presentation Breast Tomosynthesis For Presentation Digital Intra-oral X-Ray For Presentation 1.2.840.10008.5.1.4.1.1.1.2 No Yes 1.2.840.10008.5.1.4.1.1.13.1.3 No Yes 1.2.840.10008.5.1.4.1.1.1.3 No Yes CT 1.2.840.10008.5.1.4.1.1.2 No Yes Enhanced CT 1.2.840.10008.5.1.4.1.1.2.1 No Yes Ultrasound Multi-frame (Retired) 1.2.840.10008.5.1.4.1.1.3 No Yes Ultrasound Multi-frame 1.2.840.10008.5.1.4.1.1.3.1 No Yes MR 1.2.840.10008.5.1.4.1.1.4 No Yes Enhanced MR 1.2.840.10008.5.1.4.1.1.4.1 No Yes Nuclear Medicine (Retired) 1.2.840.10008.5.1.4.1.1.5 No Yes Ultrasound (Retired) 1.2.840.10008.5.1.4.1.1.6 No Yes Ultrasound 1.2.840.10008.5.1.4.1.1.6.1 No Yes Secondary Capture 1.2.840.10008.5.1.4.1.1.7 No Yes Multi-frame Grayscale Byte Secondary Capture Multi-frame Grayscale Word Secondary Capture Image Store Multi-frame True Color Secondary Capture Grayscale Softcopy Presentation State SOP Class Color Softcopy Presentation State SOP Class Pseudo-Color Softcopy Presentation State SOP Class 1.2.840.10008.5.1.4.1.1.7.2 No Yes 1.2.840.10008.5.1.4.1.1.7.3 No Yes 1.2.840.10008.5.1.4.1.1.7.4 No Yes 1.2.840.10008.5.1.4.1.1.11.1 No Yes 1.2.840.10008.5.1.4.1.1.11.2 No Yes 1.2.840.10008.5.1.4.1.1.11.3 No Yes X-Ray Angiographic 1.2.840.10008.5.1.4.1.1.12.1 No Yes Enhanced X-Ray Angiographic Image 1.2.840.10008.5.1.4.1.1.12.1.1 No Yes X-Ray Radioflouroscopic 1.2.840.10008.5.1.4.1.1.12.2 No Yes Image Display DICOM Conformance Statement Page 26 of 60

SOP Class Name SOP Class UID Enhanced X-Ray Radioflouroscopic 1.2.840.10008.5.1.4.1.1.12.2.1 No Yes Nuclear Medicine 1.2.840.10008.5.1.4.1.1.20 No Yes VL (retired) 1.2.840.10008.5.1.4.1.1.77.1 No Yes VL Endoscopic 1.2.840.10008.5.1.4.1.1.77.1.1 No Yes Video Endoscopic 1.2.840.10008.5.1.4.1.1.77.1.1.1 No Yes VL Microscopic 1.2.840.10008.5.1.4.1.1.77.1.2 No Yes Video Microscopic 1.2.840.10008.5.1.4.1.1.77.1.2.1 No Yes VL Slide-Coordinates Microscopic Image 1.2.840.10008.5.1.4.1.1.77.1.3 No Yes VL Photographic 1.2.840.10008.5.1.4.1.1.77.1.4 No Yes Video Photographic 1.2.840.10008.5.1.4.1.1.77.1.4.1 No Yes Ophthalmic Photographic 8 Bit Image Ophthalmic Photographic 16 Bit Image 1.2.840.10008.5.1.4.1.1.77.1.5.1 No Yes 1.2.840.10008.5.1.4.1.1.77.1.5.2 No Yes Basic Text SR 1.2.840.10008.5.1.4.1.1.88.11 No Yes Enhanced SR 1.2.840.10008.5.1.4.1.1.88.22 No Yes Comprehensive SR 1.2.840.10008.5.1.4.1.1.88.33 No Yes Mammography CAD SR 1.2.840.10008.5.1.4.1.1.88.50 No Yes Key Object Selection Document 1.2.840.10008.5.1.4.1.1.88.59 No Yes Positron Emission Tomography Image 1.2.840.10008.5.1.4.1.1.128 No Yes MI 3D Softcopy Presentation State 1.3.6.1.4.1.16978.3.1 No Yes Table 17: : Supported SOP Classes. 4.1.6.2 Association Policies 4.1.6.2.1 General The will accept Association Requests for DICOM Services. The DICOM standard application context name for DICOM 3.0 is always accepted, see Table 18. Application Context Name 1.2.840.10008.3.1.1.1 Table 18: : DICOM Application Context. Image Display DICOM Conformance Statement Page 27 of 60

4.1.6.2.2 Number of Associations The does not support multiple simultaneous associations, see Table 19. Maximum number of simultaneous associations 1 Configurable. Table 19: : Number of simultaneous associations. 4.1.6.2.3 Asynchronous Nature The does not support asynchronous communication. Multiple outstanding transactions are not supported, see Table 20. Maximum number of outstanding asynchronous transactions 1 Configurable. Table 20: : Asynchronous nature. 4.1.6.2.4 Implementation Identifying Information The identifying information for the can be seen in Table 21. Implementation Class UID 1.3.6.1.4.1.16978.0.3.5.4.28 Implementation Version Name EV-3.5.4-28 Table 21: : DICOM Implementation Class and Version. 4.1.6.3 Association Initiation Policy The does not initiate associations. 4.1.6.4 Association Acceptance Policy 4.1.6.4.1 Activity: Receive Instances 4.1.6.4.1.1 Description and Sequencing of Activities A remote DICOM Application Entity acting as a may establish an association with the. The will accept these associations for the purpose of receiving supported SOP Class Instances. When the user launches a CAM from the EasyViz Workflow Manager, the CAM will initiate the Retrieve Instances activity as described in section 3.1.4.3. The C-MOVE request in the Retrieve Instances activity contains the as destination Application Entity. 4.1.6.4.1.2 Accepted Presentation Contexts The accepts the Application Presentation Contexts listed in Table 22. Abstract Syntax Transfer Syntax Role Ext. Neg Name UID Name List UID List Hardcopy Grayscale Image SOP Class (Retired) 1.2.840.10008.5.1.1.29 Implicit VR Little.2.4.50.4.51.4.57.4.70 Image Display DICOM Conformance Statement Page 28 of 60

Abstract Syntax Transfer Syntax Role Ext. Neg Name UID Name List UID List Hardcopy Color Image SOP Class (Retired) Computed Radiography Digital X-Ray Image For Presentation Digital Mammography X-Ray 1.2.840.10008.5.1.1.30 Implicit VR Little 1.2.840.10008.5.1.4.1.1.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.1.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.1.2 Implicit VR Little.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91 Image Display DICOM Conformance Statement Page 29 of 60

Abstract Syntax Transfer Syntax Role Ext. Neg Name UID Name List UID List For Presentation Breast Tomosynthesis For Presentation Digital Intra-oral X-Ray For Presentation 1.2.840.10008.5.1.4.1.1.13.1.3 Implicit VR Little 1.2.840.10008.5.1.4.1.1.1.3 Implicit VR Little CT 1.2.840.10008.5.1.4.1.1.2 Implicit VR Little.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80 Image Display DICOM Conformance Statement Page 30 of 60

Abstract Syntax Transfer Syntax Role Ext. Neg Name UID Name List UID List Enhanced CT Image Ultrasound Multi-frame (Retired) Ultrasound Multi-frame 1.2.840.10008.5.1.4.1.1.2.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.3 Implicit VR Little 1.2.840.10008.5.1.4.1.1.3.1 Implicit VR Little MR 1.2.840.10008.5.1.4.1.1.4 Implicit VR Little.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.4.50.4.51 Image Display DICOM Conformance Statement Page 31 of 60

Abstract Syntax Transfer Syntax Role Ext. Neg Name UID Name List UID List Enhanced MR Image Nuclear Medicine (Retired) Ultrasound Image (Retired) 1.2.840.10008.5.1.4.1.1.4.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.5 Implicit VR Little 1.2.840.10008.5.1.4.1.1.6 Implicit VR Little.4.57.4.70.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80.4.81 Image Display DICOM Conformance Statement Page 32 of 60

Abstract Syntax Transfer Syntax Role Ext. Neg Name UID Name List UID List Ultrasound Image Secondary Capture Multi-frame Grayscale Byte Secondary Capture Multi-frame Grayscale Word Secondary Capture Image Store 1.2.840.10008.5.1.4.1.1.6.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.7 Implicit VR Little 1.2.840.10008.5.1.4.1.1.7.2 Implicit VR Little 1.2.840.10008.5.1.4.1.1.7.3 Implicit VR Little.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57 Image Display DICOM Conformance Statement Page 33 of 60

Abstract Syntax Transfer Syntax Role Ext. Neg Name UID Name List UID List Multi-frame True Color Secondary Capture Grayscale Softcopy Presentation State SOP Class Color Softcopy Presentation State SOP Class Pseudo-Color Softcopy Presentation State SOP Class X-Ray Angiographic Enhanced X-Ray Angiographic Image 1.2.840.10008.5.1.4.1.1.7.4 Implicit VR Little 1.2.840.10008.5.1.4.1.1.11.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.11.2 Implicit VR Little 1.2.840.10008.5.1.4.1.1.11.3 Implicit VR Little 1.2.840.10008.5.1.4.1.1.12.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.12.1.1 Implicit VR Little.4.70.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.2.2.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70 Image Display DICOM Conformance Statement Page 34 of 60

Abstract Syntax Transfer Syntax Role Ext. Neg Name UID Name List UID List X-Ray Radioflouroscopic Enhanced X-Ray Radioflouroscopic Nuclear Medicine 1.2.840.10008.5.1.4.1.1.12.2 Implicit VR Little 1.2.840.10008.5.1.4.1.1.12.2.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.20 Implicit VR Little.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91 VL 1.2.840.10008.5.1.4.1.1.77.1 Implicit VR Little Image Display DICOM Conformance Statement Page 35 of 60

Abstract Syntax Transfer Syntax Role Ext. Neg Name UID Name List UID List (retired) VL Endoscopic Image Video Endoscopic VL Microscopic Image 1.2.840.10008.5.1.4.1.1.77.1.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.77.1.1.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.77.1.2 Implicit VR Little.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80 Image Display DICOM Conformance Statement Page 36 of 60

Abstract Syntax Transfer Syntax Role Ext. Neg Name UID Name List UID List Video Microscopic VL Slide-Coordinates Microscopic Image VL Photographic Image Video Photographic 1.2.840.10008.5.1.4.1.1.77.1.2.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.77.1.3 Implicit VR Little 1.2.840.10008.5.1.4.1.1.77.1.4 Implicit VR Little 1.2.840.10008.5.1.4.1.1.77.1.4.1 Implicit VR Little.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.4.50 Image Display DICOM Conformance Statement Page 37 of 60

Abstract Syntax Transfer Syntax Role Ext. Neg Name UID Name List UID List Ophthalmic Photographic 8 Bit Ophthalmic Photographic 16 Bit 1.2.840.10008.5.1.4.1.1.77.1.5.1 Implicit VR Little 1.2.840.10008.5.1.4.1.1.77.1.5.2 Implicit VR Little Basic Text SR 1.2.840.10008.5.1.4.1.1.88.11 Implicit VR Little Enhanced SR 1.2.840.10008.5.1.4.1.1.88.22 Implicit VR Little Comprehensive SR 1.2.840.10008.5.1.4.1.1.88.33 Implicit VR Little Mammography CAD SR 1.2.840.10008.5.1.4.1.1.88.50 Implicit VR Little Key Object Selection Document 1.2.840.10008.5.1.4.1.1.88.59 Implicit VR Little.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2.2.2.2.2 Image Display DICOM Conformance Statement Page 38 of 60

Abstract Syntax Transfer Syntax Role Ext. Neg Name UID Name List UID List Positron Emission Tomography Image MI 3D Softcopy Presentation State 1.2.840.10008.5.1.4.1.1.128 Implicit VR Little 1.3.6.1.4.1.16978.3.1 Implicit VR Little.2.4.50.4.51.4.57.4.70.4.80.4.81.5.4.90.4.91.2 Table 22: : Accepted Presentation Contexts for the Receive Instances activity. 4.1.6.4.1.3 Extended Negotiation No extended negotiation is performed. 4.1.7 AE Specification: Print Management 4.1.7.1 SOP Classes The Print Management provides Standard Conformance to SOP Classes listed in Table 23. SOP Class Name SOP Class UID Print Management Basic Grayscale Print Management Meta SOP Class 1.2.840.10008.5.1.1.9 Yes No Basic Annotation Box SOP Class 1.2.840.10008.5.1.1.15 Yes No Presentation LUT SOP Class 1.2.840.10008.5.1.1.23 Yes No Table 23: Print Management : Supported SOP Classes. 4.1.7.2 Association Policies 4.1.7.2.1 General The Print Management will propose Association Requests for DICOM Print Management Services. The DICOM standard application context name for DICOM 3.0 is always proposed, see Table 24. Application Context Name 1.2.840.10008.3.1.1.1 Table 24: Print Management : DICOM Application Context. Image Display DICOM Conformance Statement Page 39 of 60

4.1.7.2.2 Number of Associations The Print Management does not support multiple simultaneous associations, see Table 25. Maximum number of simultaneous associations 1 Configurable. Table 25: Print Management : Number of simultaneous associations. 4.1.7.2.3 Asynchronous Nature The Print Management does not support asynchronous communication. Multiple outstanding transactions are not supported. Maximum number of outstanding asynchronous transactions 1 Configurable. Table 26: Print Management : Asynchronous nature. 4.1.7.2.4 Implementation Identifying Information The identifying information for the Print Management can be seen in Table 27. Implementation Class UID 1.3.6.1.4.1.16978.0.3.5.4.28 Implementation Version Name EV-3.5.4-28 Table 27: Print Management : DICOM Implementation Class and Version. 4.1.7.3 Association Initiation Policy 4.1.7.3.1 Activity: Print Images 4.1.7.3.1.1 Description and Sequencing of Activities The Print Management sends an Association Request to a remote DICOM Application Entity acting as Basic Grayscale Print Management Meta SOP Class when the user prints DICOM Instances from the EasyViz General 2D CAM. 4.1.7.3.1.2 Proposed Presentation Contexts The Print Management will propose Presentation Contexts as listed in Table 28 for the Print Images activity. Abstract Syntax Transfer Syntax Role Ext. Neg Name UID Name List UID List Basic Grayscale Print Management Meta SOP Class 1.2.840.10008.5.1.1.9 Implicit VR Little.2 Basic Annotation Box SOP Class 1.2.840.10008.5.1.1.15 Implicit VR Little.2 Presentation LUT SOP Class 1.2.840.10008.5.1.1.23 Implicit VR Little.2 Table 28: Print Management : Proposed Presentation Contexts for the Print Images activity. 4.1.7.4 Association Acceptance Policy The Print Management does not accept Association Requests. Image Display DICOM Conformance Statement Page 40 of 60

4.1.8 AE Specification: General Purpose Worklist 4.1.8.1 SOP Classes The General Purpose Worklist provides Standard Conformance to SOP Classes listed in Table 29. SOP Class Name SOP Class UID Work List General Purpose Scheduled Procedure Step SOP Class General Purpose Performed Procedure Step SOP Class General Purpose Worklist Information Model - FIND 1.2.840.10008.5.1.4.32.2 Yes No 1.2.840.10008.5.1.4.32.3 Yes No 1.2.840.10008.5.1.4.32.1 Yes No 4.1.8.2 Association Policies Table 29: General Purpose Worklist : Supported SOP Classes. 4.1.8.2.1 General The General Purpose Worklist will propose Association Requests for General Purpose Worklist Services. The DICOM standard application context name for DICOM 3.0 is always proposed. Application Context Name 1.2.840.10008.3.1.1.1 Table 30: General Purpose Worklist : DICOM Application context. 4.1.8.2.2 Number of Associations The general purpose worklist does not support multiple simultaneous associations. 4.1.8.2.3 Asynchronous Nature The general purpose worklist does not support asynchronous communication. Multiple outstanding transactions are not supported. 4.1.8.3 Association Initiation Policy 4.1.8.3.1 Activity: Query Worklist 4.1.8.3.1.1 Description and Sequencing of Activities The General Purpose worklist sends an Association Request to a remote DICOM Application Entity acting as a General Purpose Worklist Information Model FIND SOP Class when the user activates an EasyViz worklist profile. 4.1.8.3.1.2 Proposed Presentation Contexts The General Purpose worklist will propose Presentation Contexts as listed in Table 31 for the worklist query. Image Display DICOM Conformance Statement Page 41 of 60

Abstract Syntax Transfer Syntax Role Ext. Neg Name UID Name List UID List General Purpose Worklist Information Model - FIND 1.2.840.10008.5.1.4.32.1 Implicit VR Little Table 31: GPWL : Proposed Presentation Contexts for the worklist query activity 4.1.8.3.2 Activity: Claim/Update Work Item 4.1.8.3.2.1 Description and Sequencing of Activities The General Purpose worklist sends an Association Request to a remote DICOM Application Entity acting as a General Purpose Worklist Scheduled Procedure Step SOP Class or as a General Purpose Worklist Performed Procedure Step SOP Class when the application needs to update a Scheduled Procedure Step work Item, or needs to create/fetch/update a Performed Procedure Step Work Item 4.1.8.3.2.2 Proposed Presentation Contexts The General Purpose worklist will propose Presentation Contexts as listed in Table 32 for the worklist interactions. Abstract Syntax Transfer Syntax Role Ext. Neg Name UID Name List UID List General Purpose Scheduled Procedure Step SOP Class 1.2.840.10008.5.1.4.32.2 Implicit VR Little General Purpose Performed Procedure Step SOP Class 1.2.840.10008.5.1.4.32.3 Implicit VR Little 4.1.8.4 Association Acceptance Policy The General Purpose Worklist does not accept Association Requests. 4.2 Network Interfaces 4.2.1 Physical Network Interface The EasyViz Image Display Application Entities are indifferent to the physical medium over which TCP/IP is executed. This is entirely dependent on the underlying operating system and hardware. 4.2.2 Additional Protocols When hostnames rather than IP addresses are used to specify presentation addresses for remote Application Entities, the EasyViz Image Display Application Entities depends on the name resolution mechanism of the underlying operating system for proper operation. 4.3 Configuration 4.3.1 AE Title/Presentation Address Mapping 4.3.1.1 Local AE Titles The default Application Entity title and port number of the EasyViz Image Display Application Entities are listed in Table 29. Image Display DICOM Conformance Statement Page 42 of 60

The default AE Title for the is created from the hostname and port number of the host where the is installed. Application Entity Default AE Title Default Port Number Query/Retrieve ID_QR_ N/A ID_STORE_ N/A <hostname>-<port nr> 7000 Print Management ID_PM_ N/A Table 29: Local AE Titles. 4.3.1.2 Remote AE Title/Presentation Address Mapping 4.3.1.2.1 Remote Query/Retrieve and The remote Query/Retrieve and configuration is stored in the environment variable EV_PACS. The EasyViz Image Display assumes that the Query/Retrieve and are located on the same host. The EV_PACS environment variable has the following structure: hostname=<hostname>,port=<port number>,qr_aet=<aet>,store_aet=<aet> 4.3.1.2.2 Remote Print Management The Print Management configuration is stored in a printer profile in the internal database. For more information on how to setup a printer, see Error! Reference source not found.. 4.3.2 Parameters The configurable parameters of EasyViz Image Display are listed in Table 30. Parameter Configurable Default Value General Parameters PDU Size No 32Kb Time-out waiting for acceptance or rejection Response of an Association Open Request (Application level timeout) No General DIMSE level time-out values No Time-out waiting for response to TCP/IP connect request (Low-level timeout) Time-out waiting for acceptance of a TCP/IP message over the network (Low-level timeout) Time-out waiting for data between TCP/IP packets (Lowlevel timeout) Yes Yes Yes 180s [OS Specific] 180s [OS Specific] Adaptive 0.2-120s [OS Specific] Image Display DICOM Conformance Statement Page 43 of 60