Alpy 600 Mixed Mode Calibration

Similar documents
Quick Value Statistics in Presentation

Galaxie Photodiode Array Software

SPECTRAL IMAGING VIEWER SOFTWARE MANUAL

Model FP-6500 Spectrofluorometer Instruction Manual. FP-6500 for Windows

Using the Bruker Tracer III-SD Handheld X-Ray Fluorescence Spectrometer using PC Software for Data Collection

AnalytPro. Analysis software. User Manual

New software features of the 7010 Particle Size. Analyzer Danielle Chamberlin May May 5, Page 1

Using the Link to Petrel

DICKEY-JOHN GAC 2500-UGMA PRODUCT CALIBRATION DOWNLOAD

TIDY LABELS. User Guide

WiFi Advisor Wireless LAN Analyzer Quick Start Guide. October 2015 Revision OCT-2015 Technical Assistance: GO VIAVI /

Quick Start Guide. ARIS Architect. Version 9.8 Service Release 2

StellarNet Single-Channel SpectraWiz for LabVIEW

Jasco FP-6500 Spectrofluorimeter Updated November 14, 2017

MAN0314/2.00. Software update summary

Launch old style dialogue boxes from the dialogue box launchers at the bottom of the ribbon.

Acquiring Data (Through UV WinLab Software) Procedure for Lambda 35 UV2

Simulation and Auxiliary Data Management

Operating Procedure PerkinElmer LAMBDA 750

67 Series Spectrophotometer PC Software

Quick Start Guide. ARIS Architect. Version 9.7

Quick Start Guide: Welcome to OceanView

Opening Screen When you first enter Slick!, a blank opening screen will appear as shown below.

Multivariate Calibration Quick Guide

Product Instruction. Drew3 Control Assay Data File Internet Download and Installation Instructions

Agilent Dissolution Tester Driver

This presentation takes you through Region of Interest (ROI) Analysis.

NHSC HIFI DP workshop Caltech, September A Tour of HIFI Data. - page 1

FIFI-LS: Basic Cube Analysis using SOSPEX

BEAWebLogic Server. Using the WebLogic Diagnostic Framework Console Extension

Delta V5 operation. Widetron Technologies Corp. Wu, Chung Ying. Progress through Synergy

(Fiber-optic Reosc Echelle Spectrograph of Catania Observatory)

Creating and Viewing Curriculum

Orifice Flow Meter

Introduction to Remote Sensing Wednesday, September 27, 2017

Spectrometer Visible Light Spectrometer V4.4

V-600 Operation. JASCO V-600 for Windows Software Manual

Protein Deconvolution Quick Start Guide

ARIS Architect QUICK START GUIDE. Version Service Release 7

Welcome to noiselab Express 1.0

Allocation Tool. User Manual on the use and functionality of the Allocation Tool

MockupScreens - User Guide

Metrics Tutorial. Table of Contents. 1. Introduction

CH142 Spring Spectrophotometers with Vernier Data Acquisition Software

SPICAM Archive Tutorial. Principal Investigator : J.L. Bertaux Archive Manager : A. Reberac

USER MANUAL. igui Revision: 2. Date: page 1 of 12

MonoVista CRS+ Raman Microscopes

How to use character and paragraph styles

TOSHIBA GA Printing from Mac OS

Optics Vac Work MT 2008

Fluorescence. Requirements. Introduction. Models: FluorescenceExampleBegin.oml. Properties: FluorescenceExampleProperties.txt

Author A.Kishore/Sachin WinSCP

Contour Interface Manual

Scientific Production Company Doza Configurator Software

PANalytical Epsilon 5 XRF spectral acquisition, processing and reporting

Spectroscopy techniques II. Danny Steeghs

Interactive comment on Quantification and mitigation of the impact of scene inhomogeneity on Sentinel-4 UVN UV-VIS retrievals by S. Noël et al.

Interactive comment on Quantification and mitigation of the impact of scene inhomogeneity on Sentinel-4 UVN UV-VIS retrievals by S. Noël et al.

Zahner 12/2011. Light Spectra Display and Analysis

MIMO OTA Environment Builder. User Manual

Tutorial. COPYRIGHT 2014 IGE+XAO. All rights reserved TUTORIAL. Your first steps with SEE Electrical Expert. The specialist of electrical software

BCC Linear Wipe. When the Invert Wipe Checkbox is selected, the alpha channel created by the wipe inverts.

CONTENTS INTRODUCTION... 3 ACCESSING AND MODIFYING FLEX SAMPLES... 4 CREATE AND POPULATE A NEW FLEX DOCUMENT... 6

LED MINISPECTROMETER LMS-R

INSTRUCTIONS FOR UPDATING SOFTWARE ON THE i-stat PORTABLE CLINICAL ANALYZER

UV WinLab Software The Lambda Series

PT-6315-C virtual cash drawer

Quick Start. Getting Started

Introduction to Raman spectroscopy measurement data processing using Igor Pro

GIRO Payment & notification

About This Version ParkSEIS 2.0

PM1703MO-1BT/PM1401GNB USER SOFTWARE GUIDE

WaveAR User s Manual. Blake J. Landry and Matthew J. Hancock. WaveAR User s Manual page 1 of 19

UNIVERSITY OF NORTH CAROLINA AT CHARLOTTE

Excel window. This will open the Tools menu. Select. from this list, Figure 3. This will launch a window that

Atlas Edge Alignment Calibration Page 1

DIFFUSION TENSOR IMAGING ANALYSIS. Using Analyze

Performing whole genome SNP analysis with mapping performed locally

<Partner Name> RSA ARCHER GRC Platform Implementation Guide. RiskLens <Partner Product>

Instytut Fizyki Doświadczalnej Wydział Matematyki, Fizyki i Informatyki UNIWERSYTET GDAŃSKI

OOIBase32 Spectrometer Operating Software Operating Instructions

OIW-EX 1000 Oil in Water Monitors

Skyline irt Retention Time Prediction

4) Study the section of a worksheet in the image below. What is the cell address of the cell containing the word "Qtr3"?

1) Merge the cells that contain the title and center the title

X-ray Spectra Part I: Loading, Visualizing, & Grouping Data

HESP PIPELINE v. 1.0

A database implementation of data analysis and quality control for the Brewer

Position Table Creation

Copyright 2008 Casa Software Ltd.

QuiC 1.0 (Owens) User Manual

X-Lab Pro 5 Operator Manual

MultiMap. Name. Operating Manual

SoftWare. M.Wave Professional PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION

CVI SPECTRAL PRODUCTS. 111 Highland Drive Putnam CT, (860) SM32Pro

v SMS 11.1 Tutorial BOUSS2D Prerequisites Overview Tutorial Time minutes

Formatting a Table of Contents Using Word 2013

COGNOS User-Defined Prompts and Maps User Guide

Little Math User Manual

RSpec Update Version

Transcription:

Alpy 600 Mixed Mode Calibration Paul Luckas October, 2016 1 Introduction Alpy 600 Calibration using the mixed mode involves the following basic steps. 1. Use a sequence of spectra from a type A or B star as the source. 2. Set up the project in ISIS using the conventional file naming and structure. 3. Use the Calibration assistant. 2 General settings Ensure general settings as per Figure 1. Mixed mode uses the pre-defined spectral calibration mode. Wavelength registration should be deactivated due to [currently] unexplained wavelength shifts in processed spectral profiles that appear when this feature is enabled. 1

Figure 1: Settings on the General tab. 3 The Calibration assistant Proceed to the Calibration tab and select Calibration assistant. The calibration assistant uses a mixed mode of Balmer lines present in the spectra of type A or B stars, and calibration lines produced by the Alpy calibration module. Detection of these features is automatic, but involves a number of steps. Figure 2 shows the default screen with Root name and Object name automatically populated. Other fields may or may not be populated depending on how the assistant was last used. Important: Pressing Go adjacent to the Object name dialogue on the calibration assistant default window will produce the unexpected result of renaming the generic, calibration and response file names back on the General tab. Accordingly, this step should be ignored if file names have already been correctly configured on the General tab. 3.1 Producing a mean spectral image The calibration assistant will produce a mean image from a sequence of spectral images. Together with a calibration image, this improved image will be used in determining the calibration. 2

Figure 2: The calibration assistant default screen. 1. Enter the Generic name using the dialogue to navigate to the appropriate file. 2. Select the Number of images auto-complete button. 3. Type a meaningful name into the Result (mean image) dialogue. 4. Press Go. The mean image will be created and saved to the working directory as shown in Figure 3. It should also automatically appear in the top right Calibration selection dialogue. 3.2 Determining the calibration The mean image together with a neon calibration spectrum is used as part of the mixed method. The inputs for these, and the process for determining the calibration are oulined as follows: 1. Ensure the mean image created in the last step appears in the top left Calibration star spectrum dialogue. It can also be manually selected by navigating to the file. 2. Enter the star s radial velocity (if known). 3. Ensure the Calibration neon spectrum image file name is correct. If not, it is selected by placing a cursor in the CALIBRATION file name box located back on the GENERAL tab and pressing ENTER. There is no need to close the calibration assistant during this step. 4. The horizontal and vertical coordinates of the H-alpha line are entered by selecting the feature on the open spectral image. Ensure the correct absorption feature is selected! 3

Figure 3: Computing the mean of an image sequence. 5. Experiment with UV calibration to increase the data range used in the computation for a potentially improved result. The result is shown in Figure 4. Figure 4: A successfully completed spectral calibration calculation. The dispersion coefficients are transferred to the spectral dispersion dialogue on the Profile tab (Figure 5). 4

Figure 5: The updated dispersion dialogue. 4 Checking the result The newly created calibration can be used to calibrate the open spectral images. The resultant profile can then be compared to a Miles standard to confirm alignment of the major absorption features of the type A or B star used to create the calibration. Creating and applying an instrument response profile, and scaling the comparison spectrum to the same spectral range allows a very precise visual comparison. 5

Figure 6: Comparison profiles. 6