FAB-2S FRIABILITY TESTER (with variable speed setting) Instruction Manual Rev : 0 Date : 5/26/2015

Similar documents
HDT-400L HARDNESS TESTER. Instruction Manual

PTF 3DR multiple drum Tablet Friability Test Instrument

PTF 600. Six Drum Tablet Friability Test Instrument

OPERATING MANUAL. Gilson Rice Shaker SGA-5R & SGA-5RT

FMFDUSPMBC! Automated Tap Density Tester INSTRUCTION MANUAL ETD 1020 !!FMFDUSPMBC!

Haze 800 DMX. User Guide. English

VBT-75 S2 VACUUM BOTTLE TESTER

INTRODUCTION CT87E FEATURES AND CONTROLS

H Manual Stretch Wrap Machine uline.com FILM ROLL CAPACITY TECHNICAL DATA ELECTRICAL SPECIFICATIONS TURNTABLE SPECIFICATIONS

GOLANDER PUMP BQ80S Microflow Variable-Speed Peristaltic Pump Operation Manual

High Precision Quad Output DC Power Supply

INSTRUCTION MANUAL. Instruction Manual. Analog Multi-Tube Vortexer Digital Multi-Tube Vortexer

Instruction Manual Standard Multi-Position Stirrers Advanced Multi-Position Stirrers

Replacement Instructions

ASTRO UW-1C and RW-1C LABEL PRINTER UNWINDER & WINDER

TEL: / FAX:

OPERATING MANUAL. Micro-Deval Apparatus MD-2000, MD-2000C & MD-2000F

Model P4470 Save A Watt Operation Manual

New Brunswick Products. Operating Manual M S Revision B+

Gardner-scrub Abrasion Tester

Service and Technical Support PLEASE CONTACT YOUR NEAREST DISTRIBUTOR If unknown then fax: 44 (0)

Digital Overhead Stirrer

Z-350 FAZER USER MANUAL

Operating Manual. AdirPro HV8RL Red Beam Horizontal/Vertical Laser level

Operational Manual Spectrophotometer Model: SP-830 PLUS Metertech Inc.

Power Supply, 17-inch

Serial No. OWNER S MANUAL. Installation & Operation

GoboLED 80 D AL2502. Manuale di istruzioni Instructions manual

Electric Actuators and Control Systems. Established Leaders in Valve Actuation AD-8140 & AD Instruction Manual. Publication P971E Issue 08/09

BROTHER HL 4040 TN-110 TN-115 CARTRIDGE REMANUFACTURING INSTRUCTIONS

ARGO LAB SKO-D XL Orbital and Linear digital shaker. User manual

TITLE: Field adjusting EECO 1000/2000/3000 Power Supply and Printer Mechanism Print Density

GOLANDER PUMP. Operation Manual. For Microflow Variable-Speed Peristaltic Pump BT50S/BT102S.

90 DEGREE PEEL TESTER

Vacuum Maintenance Manual (EXCERPT Tim Benedict)

TP3005P EN PROGRAMMABLE DC LAB POWER SUPPLY 0-30V/ USB 2.0

180 DEGREE PEEL TESTER

Deluxe In/Out Remote Thermometer with Trend Chart

Operation Manual. for Micromas ter I Micros copes. T his manual covers catalog numbers : S 11031, S 11033, and S 11035

VK-3iX WARRANTY REGISTRATION FORM

Members: Sponsor: Dalton Hamilton Preston-Eastin Levi Edens Kenneth Mui Brice Abbott

EASTERN LABS. Instructions Model RM /25/2007. Operation. Junction Box Assembly. Different Types of Switch Installations

ColorMaxLP Label Roll Rewinder

Installation Guide Philips MP20/30/40/50/60/70 IntelliVue M-Series Arm Rail Mount Kit

National Optical & Scientific Instruments Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210)

SL-35 MARK III INSTRUCTION MANUAL

TRC-190 User s Manual

Operation Manual. Concorde 600 Power Supply. *This instrument is intended for laboratory use only.

nual k Excella E-2 Classic Platform Shaker N) manual New Brunswick Excella E-2 Open Air Shaker Operating manual

RPX Series USER S MANUAL

NightStar. Installation, Operation, & Maintenance Instructions M /04/2016

Analog Multi-Tube Vortexer Digital Multi-Tube Vortexer

Installation, Operation and Maintenance Manual

Model N47KB Energy Monitor with Built-in Timer

Operation Manual WARNING. Be sure to read this Operation Manual before use. Universal Space Amusement Equipment Ltd.

Triax A/T QUICK START GUIDE

CALOG CALOG - LC DISPLAY. Loadcell Display. English Users Manual. Rev. 1.10

Power Supply (48Vdc, 2.5A)

M20 Metallurgical Microscope User s Manual. Fein.

BAUM K20B KNIFE FOLDER UNIT INSTRUCTION MANUAL

NTP-5521/5531/5561 SWITCHING MODE POWER SUPPLY

DS5/DS10/DS20 Fluid Head Operator s Manual. #180 DS5 Fluid Head #182 DS10 Fluid Head #184 DS20 Fluid Head

Energy Logger. User's Guide. Model EM Washington Street Melrose, MA Phone Toll Free

Installation Guide. Universal Camera Mount Bracket

User Manual Charter Street Los Angeles CA

User Manual. Binocular Stereo Microscope. MicroscopeNet.com

Memory, 20-inch. Replacement Instructions. Tools Required. Part Location

Linear Shaker & Orbital Shaker

Chinon 35 FS-A

Operating Instructions

PD Way Pan/Tilt Head. You re on steady ground

ABB Drives. User s Manual Pulse Encoder Interface Module MTAC-01

skyline 70 Fluid Head OPERATOR S MANUAL #1050 Skyline 70 Fluid Head

VBAR 270. User Manual. Version 1.2

Rocking Shaker & 3D Shaker

Cel-Gro Tissue Culture Rotators

EL7060 Series SERVICE MANUAL

MX-T6-PRO/S Tube Rollers

MT400 MOTION CONTROLLER

BE Series. Dual Action Ballhead. You re on steady ground

DT-900 Techstrobe Operation Manual

P4472 Save A Watt HD Operation Manual

Polarization Microscope. Omano OM349P Series. User Guide

ASSEMBLY FEATURES. Please check to make sure the following items are included with the main unit in the carton:

TRC-190 User s Manual

User's Guide. Model High Precision Quad Output DC Power Supply

Model DT-315P DIGITAL STROBOSCOPE INSTRUCTION MANUAL

INSTALLATION INSTRUCTIONS FOR YOUR MULTI-6

* IMPORTANT * REGISTERING YOUR MACHINE

MCCB-500 MOLDED-CASE CIRCUIT BREAKER TESTER

VERTICAL CONTROL LASER OWNERS MANUAL

STATIC SHEAR INTEGRATED OVEN

2 in 1 LAN Tester and Multimeter Model:

Ostar Eye K18. Beam and Kaleido Effects. User manual. 6. Fixture Cleaning. Please read the instructions carefully before use

Alpha-3R. Direct Thermal Portable Printer SERVICE MANUAL

DC100 Digital Dispenser Operating Manual

Treadmill Embedded Touch Screen Won t Power Up

HDR TM K HOIST DRUM ROTATION INDICATOR KIT

SW-250 Wireless Control High Output Snow Machine User Manual

Part No. Z , IA Jul OPERATION MANUAL. High Voltage Digitalmeter A

Transcription:

(with variable speed setting) Instruction Manual Rev : 0 Date : 5/26/2015 LOGAN INSTRUMENTS CORP. 19 C School House Rd., Somerset, NJ 08873 Tel: 732-302-9888 Fax: 732-302-9898 E-Mail: info@loganinstruments.com www.loganinstruments.com

1.0 SPECIFICATIONS 2.0 DESCRIPTION OF 3.0 DESCRIPTION OF CONTROLS 4.0 INSTALLATION AND SET UP 5.0 OPERATING PROCECURES 6.0 MAINTENANCE 7.0 TROUBLE SHOTTING 8.0 SPARE PARTS AND SERVICE INFORMATION LOGAN FRIABILITY TESTING SYSTEM TABLE OF CONTENTS Page 2 of 12

1.0 SPECIFICATIONS Drum Position: 2 LOGAN FRIABILITY TESTING SYSTEM Drum Speed: Speed Accuracy Per Minute: Maximum Testing Time: Maximum Count of Revolutions: Variable Speed 25-70 RPM ±1 RPM 9 HR 59 MIN. 59 SEC. 99999 Revolutions Size: 12 x 16 x 14 Weight: Voltage: 20LBS. 110V/50-60Hz 220V/50-60Hz Page 3 of 12

2.0 DESCRIPTION OF Logan FAB-2S Friability Tester is a double-drum Tablet Friability Test instruments, manufactured in compliance to the USP, Europe, and other Pharmacopoeias. One of the testing criteria of mechanical strength of tablets and cores is Friability and Abrasion testing. During the process of coating, transportation, packing etc. the tablet will lose some mass. To measure the mass loss the samples are counted and weighed. Thereafter the Friability test performed following the individual monographs of the Pharmacopoeia. When finished the samples have to be dusted and weighed again. Actually the difference between the weight before and after the test is determined as Friability, which usually should not exceed 1%. The standard Friability drum causes the samples to roll and fall during the test while the abrasion drum continuously stresses the sample by abrasion of lamellas. FAB-2S Friability Tester can be used 1 or 2 drum. A noiseless gear motor rotates the drums with a variable speed of 25-70 RPM that programmed by users. The actual drum is made from Plexiglas and separated into two parts, the drum body and a removable cover, opened to fill, discharge and clean the drum inside when required. The instrument itself is made from high quality metal which meets in full the actual GLP requirements. includes the following: Main Drive Unit 2 Drums (L & R) 2 Drum Nuts (L & R) Page 4 of 12

2.1 FAB-2S AUTOMATED DISSOLUTION TESTER FRONT VIEW Page 5 of 12

3.0 CONTROL LOCATIONS ON FRONT PANEL OF FAB-2S Power Switch Location: Left side control panel Purpose: Turns the system on/off Display Purpose: To display timer or counter Timer Indicator Light Purpose: To indicate Friability Tester is in timer mode. Counter Indicator Light Purpose: To indicate Friability Tester is in counter mode. Start/Stop Button Purpose: To start or stop the timing or counting Counter/Timer Button Purpose: To select counter mode or timer mode Reset Button Purpose: To reset Friability Tester to re-start Enter Button Purpose: To enter the digits Set Button Purpose: To set the digits Page 6 of 12

4.0 INSTALLATION AND SET UP 4.1 Remove the unit from packing box. 4.2 Remove both drums from packing box. 4.3 Remove drum nut from Friability Tester 4.4 Install drums inf. Shafts. Both drums are marked. One marked L goes to left side. One marked R goes to right side. Make sure the drum drive slot engage with drive bar and tighten with drum nut. 4.5 Connect the power cord to the Friability Tester 5.0 Operating Procedure 5.1 Turn power switch on. 5.2 Select timer mode or counter mode by pressing Timer/Counter button. 5.3 Timer Mode: With Timer indicator light on, display will show a default time of 0:04:00. This means when Start/Stop button has been pressed drum will rotate for 4 minutes and then stop. To Set Test Time: (Make sure timer mode indicator light is on before proceeding.) Set Seconds: A. Press set button, the last digit will start flashing. B. Press Enter to increase the flashing digit. (Skip step B if no change is required.) C. Press Set button, the second to last digit will start flashing. D. Press Enter to increase the flashing digit. (Skip step D if no change is required.) Page 7 of 12

Set Minutes: A. Press Set button, the third to last digit will start flashing. B. Press Enter to increase the flashing digit. (Skip step B if no change is required.) C. Press Set button, the fourth to last digit will start flashing. D. Press Enter to increase the flashing digit. (Skip step D if no change is required.) Set Hour: A. Press Set button, the first digit will start flashing. B. Press Enter button to increase the flashing digit. (Skip step B if no change is required.) C. Press Set button to end the setting process. 5.4 Counter Mode With counter indicator light on, display shows a default revolution count of 0:01:00. This means when Start/Stop button has been pressed drum will go 100 rotations and then stop. To Set Timer Mode: (Make sure the counter indicator light is on before proceeding.) Set Revolution Count: A. Press Set button, last digit will start flashing. B. Press Enter button to increase the flashing digit. (Skip step B if no change is required.) Page 8 of 12

C. Press Set button, the second to last digit will start flashing. D. Press Enter button to increase the flashing digit. (Skip step C if no change is required.) E. To change the remaining digits repeat the above procedure. 5.5 To Reset the Friability Tester Press reset button. FAB-2S will go back to default starting mode. Please note that default mode for FAB-2S is Timer Mode. To select Counter Mode you must press Timer/Counter button. 5.6 To Start Test Once you have finished setting your time or count, press Start/Stop button. FAB-2S will start rotation for timer or counter mode. 5.7 To Stop Test FAB-2S will automatically stop when it reaches the preset time or count. To stop test before it reaches its preset time or count, press Start/Stop button. Warning: Below information for the lab manager only, do not show technician how to change the rotation speed to avoid future problems. F. To change the rotation speed: 5.8 Turn on the POWER switch while pressing and holding the Start/Stop button, the display shows the last rotation speed. 5.9 Press Set and Enter button to input the new rotation speed. Proceed as to set the Revolution Count. The speed setting is from 20-70 RPM. 5.10 Press Reset button into test mode for the test. Page 9 of 12

To set speed (from 25 RPM to 70 RMP): LOGAN FRIABILITY TESTING SYSTEM A. Press and hold Start/Stop button and turn on the main power switch at the same time; this will put the FAB-2S into speed program mode. In speed program mode, only the last 2 digits can be programmed (25 RMP 70 RMP). B. Press the Set button, the last digit will start flashing. C. Press Enter to increase the flashing digit. (from 0 to 9) D. Press Set button, and the fourth digit will start flashing. E. Press Enter to increase the flashing digit. (from 2 to 7.) F. Press Reset to enter the new speed setting into the FAB-2S. G. Turn off the FAB-2S and power on again to check the new speed setting on the display to start the test at the new speed. Page 10 of 12

6.0 Maintenance Keep FAB-2S Friability Tester in a dry and flat surface 6.1 Drum After test please clean by using a dry cloth. Do not use any solvent. Lay on clean flat surface. Do not store at vertical position. 6.2 Drive Unit Keep on a flat and clean surface. Clean the surface by using a dry cloth. Do not use any solvent. 6.3 Drive Shaft Make sure both shafts are dry at all times. Clean shaft with dry cloth. Do not use any solvent. Fuse Replacement Instruction: To replace the fuse, turn off the power switch. Gently press the fuse holder knob in and turn counter clockwise. Take the old fuse out and replace it with a fuse (Part#3AG-5AMP at 250VAC) Ground Symbol Do not remove any screws with ground symbol next to it. Page 11 of 12

7.0 Trouble Shooting: Symptoms No Power When power is on constantly long beep No rotation Possible Cause Check Line Cord Check Fuse Check Buttons if join with case Check Motor and Belt If further assistance is needed please contact your local distributor or Logan Instruments: Address: 19 C School House Rd. Somerset, NJ 08873 Tel: 732-302-9888 Fax: 732-302-9898 Email: info@loganinstruments.com Page 12 of 12