OEM Digital NDIR CO sensor

Similar documents
OEM Digital NDIR CO 2

Model 305e OEM CO 2 concentration sensor

CDD Carbon Dioxide Transmitter

Agenda. Charging the QRAE II

2F. No.25, Industry E. 9 th Rd., Science-Based Industrial Park, Hsinchu, Taiwan

Page 1 MRK-D-0011, V1.1 Aeroqual SM50 User Guide

Product Specification. CO 2 Engine K30 STA. Sensor Module and OEM Platform

Product Specification. Sensor Module and OEM Platform

MDP200 Series. Differential Pressure Sensor FEATURES APPLICATIONS DESCRIPTION

Gas and Air Sensors. Product Specification. CO 2 Engine K30. Sensor Module and OEM Platform. Document PSP 110. Rev 4. Page 1 (11)

Datasheet: K-33 BLG Sensor

CDD4 Series Room CO2 Transmitter Installation Instructions

NSGV Series "90DM3" Carbon Dioxide Detector

ZBXYAF Oxygen sensor control circuit board

TOXALERT MODEL AIR 2000

INSTRUCTION MANUAL FOR GUARDIAN INFINITY SIM PROGRAM

CDD4 Duct Carbon Dioxide Transmitter

Product Specification. CO 2 Engine K33-LP T. Sensor Module and OEM Platform. Document PSP0120. Rev 4. Page 1 (13)

PRODUCT CONTENTS... 3 II. REQUIRED MATERIALS... 3 III. OVERVIEW... 3 IV. BENCH LAYOUT...

The unit is designed for connecting to Direct Digital Control (DDC). The linear output functions are pre-programmed as CO 2

Rev Carbon Dioxide (CO2) Gas Sensor. TG100 User Manual

DP DP2500 DP0250

Rev Carbon Dioxide (CO2) Gas Sensor. TG100 User Manual

MODEL 400. CO2 Concentration Sensor. user s Manual

Model 620A Maintenance and Calibration

Carbon Monoxide Sensor - ModBus

KCD-HP. KCD-HP200x, 300X. [Figures] Top : 1% Sensor probe Bottom left : 10% / 20% Sensor probe Bottom right : Holding bracket(optional) Measurement

Optimizer Collect User Manual. for MFC/MFM calibration with Integrator 110 and Drycal

Calibration & Connectivity

Page 1 MRK-D-0011, V2.0 Aeroqual SM50 User Guide

Instruction Manual CTC-1

UniPak UP448 Bridge Input Isolating Signal Conditioner

SERIES CMT CARBON MONOXIDE GAS TRANSMITTER

MH-Z16 Intelligent Infrared Gas Module

QC-3265 Checkweigher User s Manual

Installation Guide. Carbon Dioxide (CO 2. ) Detectors. General Information. Mounting WARNING CAUTION. SAE-1011/1012 Space Mount SAE-1011/1012/1062

Product Specification. tsense VAV. CO 2 -, Temperature- and RH- sensor in a housing with colour touch display

LAUREL. Laureate Digital Panel Meter for Load Cell & Microvolt Input ELECTRONICS, INC. Features. Description

Product Specification. CO 2 Engine ICB. Sensor module for bio applications

ioselect Z-NET Z-SG Bridge Input Isolating I/O Module

SNAP-IT-PM Installation Guide

QUICKSTART GUIDE. Infrared Gas Analyzer. Teledyne Analytical Instruments

ZG07 CO2 Module. ZG07 series CO2 Module. User Manual

PC104P-16AO20 20-Channel 16-Bit High-Speed Analog Output PC104-Plus Board With 440,000 Samples per Second per Channel, and Simultaneous Clocking

AQS51/61-KAM COMBINED CO 2 -/TEMPERATURE DUCT SENSOR/CONTROLLER

Operator Quick Guide ORBISPHERE 3650

TELEDYNE HASTINGS INSTRUMENTS

GMM220 CO 2. Transmitter Modules. Introduction. Mounting

PCI-16HSDI: 16-Bit, Six-Channel Sigma-Delta Analog Input PMC Board. With 1.1 MSPS Sample Rate per Channel, and Two Independent Clocks

HWS Series Wall Mount with Relay Setpoints, LCD, and Humidistat Control

A C C S for Road Tunnels. NOVICOM CO/NO/Visibility Monitor Contact us for Information about:

Gas Analyzers Components for Gas Analysis

INSTRUCTION MANUAL STATION CONTROLLER SC1000 MOTOR PROTECTION ELECTRONICS, INC.

Carbon Monoxide (CO) Gas Detection and Control System

Bosch LSU4 Wide Band UEGO Controller

Data Cal 2000 Operator s Manual

and/or b. an optional internal Iridium satellite modem.

Model 815 User s Manual

CO Analyzer. The Value Leader TM

MTII4200 Level Transmitter Installation, Operation & Maintenance Instructions

Insertion Flowmeter for compressed air and gases DN50 - DN700 (2-28 )

2.2 Main technical parameters. Detection range. 0~5%vol(0~100%vol optional) Warm-up time. SPAN <±500ppm Long-term drift

TELEDYNE HASTINGS INSTRUMENTS

MS4525DO DESCRIPTION STANDARD RANGE (PSI)

Alma 10201/10401 ALMA 10201/ ALMA O2 UEGO Simulator Module 2 CH ALMA O2 UEGO Simulator Module 4 CH REV. 1.

WebDAQ 316. Internet Enabled Thermocouple Logger. Remote Configuration and Monitoring. Virtually Unlimited Storage. Integrated HW and SW

DGS-NO August 2017

HPS-M -2 DIFFERENTIAL PRESSURE TRANSMITTER. Mounting and operating instructions

CHAPTER IV Computerized Data-acquisition Systems

Instruction and Operation Manual

Sidewinder Pumps Inc. AC Timer/Controller

CPCI-16HSDI. 16-Bit, Six-Channel Sigma-Delta Analog Input Board. With 1.1 MSPS Sample Rate per Channel, and Two Independent Clocks.

sensor, which compensates for ageing effects, is highly insensitive to pollution and offers outstanding long term stability.

Vaisala Veriteq VL 2000 Temperature Humidity Loggers

Calibration Procedure

XMC-24DSI24WRC Wide-Range 24-Bit, 24-Channel, 200KSPS XMC Analog Input Module With 24 Wide-Range (High-Level, Low-Level) Delta-Sigma Input Channels

RPS-401A, RPS-426A Self-Contained Ultrasonic Sensor

WEA-Base. User manual for load cell transmitters. UK WEA-Base User manual for load cell transmitters Version 3.4 UK

Version MICROSTRAIN, INC. Microminiature Sensors. OEM Inertia-Link /3DM-GX2 Modules. Guidelines for Use of OEM Inertia-Link /3DM-GX2 Modules

asense m III in industrial wall mounted housing / CO sensor with built-in general purpose controller

Indoor Air Quality Meter/Datalogger

TRACKER 240 SERIES. Load Cell and Weighing Indicators. A Precision Measurement Instrument with Outstanding Features

The IQ240 panel mount load cell indicator is a precision digital indicator for load cell and strain gauge applications.

Technical Note Series

LAUREL. Laureate Digital Panel Meter for Process and Ratiometric Signals ELECTRONICS, INC. Features. Description

Sensor for car parks and tunnels

B. Rev. Release Date Description of Change ECR 0A 10/10/2014 Initial Release --- 0B 10/30/2014 Added firmware version to setup menu ---

Model OI-6900-LEL Sensor Assembly

DLA. DMX512 Analyzer. DLA Users Manual SV2_00 B.lwp copyright ELM Video Technology, Inc.

Installation and operation manual ReciFlow Gas

WMA-5. Operation Manual. Version PP Systems. All Rights Reserved. 3 rd November PP Systems. 110 Haverhill Road, Suite 301

K33-BLG/ELG Sensor module for environment logging Data Sheet and Manual

The CO 2 Engine Infrared Gas Sensing For OEM Applications

2001VT Temp & CO 2. - air quality sensor/controller NOTE!!! Manual for Installation. General. SenseAir AB

The IQ300 wall mount load cell indicator is a precision digital indicator for load cell and strain gauge applications.

User s Guide TG Series C. Senva Sensors 9290 SW Nimbus Ave Beaverton, OR 97008

cpci6u-24dsi32r 32-Channel 24-Bit Delta-Sigma Analog Input Board

Analog input module. Data sheet. for MULTICAL 603

MS4515DO SPECIFICATIONS FEATURES APPLICATIONS

SENSIT HCN Hydrogen Cyanide Analyzer

Transcription:

OEM Digital NDIR CO sensor with flow-through gas cell Model 2015SPI-CO-20 The VALTRONICS Model 2015SPI-CO-20 is an OEM NDIR CO sensor with digital signal processing and temperature compensation. The firmware VERSION depends upon the specific customer interface requirements. The SPI (Serial Peripheral Interface) is described on pages 2 and 3. Each serial numbered sensor is individually gas calibrated and temperature compensated at the factory. An RS-232 Test Board (which can be connected to a USB port, via an adapter) is needed for field gas calibration & diagnostics (See Application Note A66). It has on-board & remote switches for calibration. RANGE setting from 3% up to 20% defines 0-1 volt output (see page 4). Model 2015SPI-CO-20 Specifications: Method: Gas:.. NDIR with Digital Signal Processing and Temperature Compensation Carbon Monoxide (CO) Full-scale and RANGE:... 0-20% CO RANGE user-selectable from 3% to 20%, see page 4 CAUTION:.. Lower Explosive Limit (LEL) is 12.5% CO by volume in air Input Power:. +12 VDC (@ 0.250 amp max, 0.135 amp typical, 16.0 volts max, 8.0 volts min) Accuracy:. ±0.25% CO from 0 to 5.90% CO, and 5% of reading from 6% to 20% CO... 16-bit analog to digital converter: Delta-Sigma Conversion Method Resolution / Repeatability:... ±0.05% CO (challenge with same sample gas multiple times & assure zero) Stability:... Less than 0.05% CO in any 20-second period of time Warm-up Time: Less than 3 minutes for use and 5 minutes for accuracy Output Signal:.. Digital SPI (Serial Peripheral Interface), Application notes A59 and A61... Linear 0 to 1 volt output signal, 12-bit resolution. See pages 3 & 4 RS232 Test Board:... For troubleshooting & gas calibration, see Application note A66 LED Indicators: IR Source ON/OFF Indicator, Power ON Indicator, Cal Switch Indicators Input Signal:. Digital SPI input for calibration & diagnostic modes. See Application Note A59 Calibration Switches:... Operating Temperature Range: SW1 (Zero), SW2 (Span Target (STAR)), SW3 (Span), SW4 (Range adj.), remote via J3 0 to 50 C (32 to 122 F), see Application Note A12 Ambient Relative Humidity: 0 to 95% RH non-condensing: see Application note A30 CO 2 Interference:. Less than 1% CO response to 100% CO 2 Storage Temperature Range: -40 to +70 C (-40 to 158 F) Weight:. Less than.25 pound (<.011 kilogram) External Dimensions (PCB): 4.9 x 2.9 x 3.5. See page 2 for mounting Model 2015SPI-CO 20% CO Page 1 of 5 Revised: 7/18/2017 1994-2017 VALTRONICS, INC.

See Application Note A67 for gas conditioning requirements and information about gas sampling pumps and filters. Use a Hydrophobic Filter immediately in front of the gas inlet hose barb. Push gas through the gas cell at a rate of between 0.3 and 1.0 liter per minute. See Application Note A79 for more information about the hydrophobic filter. Model 2015SPI-CO 20% CO Page 2 of 5 Revised: 7/18/2017 1994-2017 VALTRONICS, INC.

* See Application Note A75 for J1, J2 & J3 mating connector part numbers. Keep J1 interface cable shorter than 18. See Application Note A61 for 16-bit serial *Model 2015SPI-3 shown on this page for board reference Model 2015SPI-CO 20% CO Page 3 of 5 Revised: 7/18/2017 1994-2017 VALTRONICS, INC.

Gas calibration may be initiated via a command from the SPI input on J1 (see Application Note A59) or from the on-board or remote switches via J3 (Logic "0" to initiate). For the locations of the on-board switches, test points and LEDs 3-6, refer to the illustration on the previous page. ZERO (SW1 or remote J3-3): With nitrogen flowing in gas calibration tube press and hold SW1 for 2 seconds. LED 3 through 6 will flash on & off together. If they flash on/off sequentially the sensor has detected an error and the RESET button must be pressed. Wait 1 minute and continue where you left off. The 0 to 1 volt output should snap to 0.0 ±0.01 volt measured with a DVM "+" lead connected to TP7 and "-" lead connected to GND test point. LED 3 will be ON to indicate a ZERO calibration. RANGE (SW4 or remote J3-2): To set the full scale or range press and hold SW4. From the chart on page 5 find the voltage value that corresponds to the full scale that you want from 3 to 20%. LED 6 will be ON. Use SW1 as an UP and SW3 as a DOWN switch to adjust this value (examples: 20% = 1.00 v, 10% = 0.50 v, 3% = 0.15 v). STAR (SW2 or remote J3-5): To set the Span Target (calibration gas value) press and hold SW2. LED 4 will be ON. Use SWI as an UP and SW3 as a DOWN switch to adjust this value read on the DVM. See chart on page 5. SPAN (SW3 or remote J3-4): To SPAN calibrate while flowing certified span gas (like 5.0 ±0.1% CO) in gas calibration tube for at least 30 seconds at about 300 ml/min. Press & hold SW3 for 2 seconds. LED 5 will be ON. The DVM voltage should snap to the STAR value entered above & LED 3 through 6 will flash on & off together. On the next page is a table that shows the 0 to 1 volt output equivalent to any Full Scale RANGE or Span TARget (STAR) anywhere from 3 to 20% CO. Version 20.34 or later allows a STAR to be as low as 1.00% CO. The ideal Span TARget gas for best accuracy is near mid-scale, like 10.0 ±0.2% certified CO, or near your typical measurement point. A typical application would be a full-scale RANGE of 20.0 and a STAR of 10.00. If the serial digital output is used (see App Note A61) the RANGE value does not affect it. The serial digital output has a 16-bit resolution over the whole 0 to 20% scale. The Full-Scale RANGE will set what % CO will give an output of 1.00 volt. A Range of 5.0 will give a 0-1V output of 0.600 volt for a gas reading of 3.0% CO. A Range of 10.0 will give a 0-1V output of 0.500 volt for a gas reading of 5.0% CO. A Range of 20.0 will give a 0-1V output of 0.250 volt for a gas reading of 5.0% CO. In other words, the 0-1V output is a decimal representation of the percent of full scale: for a Range of 20.0, 5.0% gas is ¼ or 25% of the full-scale (5 divided by 20), which is a decimal value of.25, which would be.250 volt. 10% gas is ½ or 50% of the full-scale, which is a decimal value of.5, which would be.500 volt. Please remember that calibration using the RS-232 Test Board will give the user much better visibility as to what RANGE and what STAR value is selected 1. Version 20.34 or later allows a STAR to be as low as 1.00% CO. 1 Most computers nowadays do not have an RS-232 port, so a RS-232 to USB adapter would be needed. Model 2015SPI-CO 20% CO Page 4 of 5 Revised: 7/18/2017 1994-2017 VALTRONICS, INC.

Table used for Calibration for setting the Full-Scale & the Span TARget value, STAR (certified tank %). Range/STAR 0 to 1V Range/STAR 0 to 1V Range/STAR 0 to 1V Range/STAR 0 to 1V % gas Ouput % gas Ouput % gas Ouput % gas Ouput 20.0 1.000 15.0 0.750 10.0 0.500 5.0 0.250 19.9 0.995 14.9 0.745 9.9 0.495 4.9 0.245 19.8 0.990 14.8 0.740 9.8 0.490 4.8 0.240 19.7 0.985 14.7 0.735 9.7 0.485 4.7 0.235 19.6 0.980 14.6 0.730 9.6 0.480 4.6 0.230 19.5 0.975 14.5 0.725 9.5 0.475 4.5 0.225 19.4 0.970 14.4 0.720 9.4 0.470 4.4 0.220 19.3 0.965 14.3 0.715 9.3 0.465 4.3 0.215 19.2 0.960 14.2 0.710 9.2 0.460 4.2 0.210 19.1 0.955 14.1 0.705 9.1 0.455 4.1 0.205 19.0 0.950 14.0 0.700 9.0 0.450 4.0 0.200 18.9 0.945 13.9 0.695 8.9 0.445 3.9 0.195 18.8 0.940 13.8 0.690 8.8 0.440 3.8 0.190 18.7 0.935 13.7 0.685 8.7 0.435 3.7 0.185 18.6 0.930 13.6 0.680 8.6 0.430 3.6 0.180 18.5 0.925 13.5 0.675 8.5 0.425 3.5 0.175 18.4 0.920 13.4 0.670 8.4 0.420 3.4 0.170 18.3 0.915 13.3 0.665 8.3 0.415 3.3 0.165 18.2 0.910 13.2 0.660 8.2 0.410 3.2 0.160 18.1 0.905 13.1 0.655 8.1 0.405 3.1 0.155 18.0 0.900 13.0 0.650 8.0 0.400 3.0 0.150 17.9 0.895 12.9 0.645 7.9 0.395 2.9 0.145 17.8 0.890 12.8 0.640 7.8 0.390 2.8 0.140 17.7 0.885 12.7 0.635 7.7 0.385 2.7 0.135 17.6 0.880 12.6 0.630 7.6 0.380 2.6 0.130 17.5 0.875 12.5 0.625 7.5 0.375 2.5 0.125 17.4 0.870 12.4 0.620 7.4 0.370 2.4 0.120 17.3 0.865 12.3 0.615 7.3 0.365 2.3 0.115 17.2 0.860 12.2 0.610 7.2 0.360 2.2 0.110 17.1 0.855 12.1 0.605 7.1 0.355 2.1 0.105 17.0 0.850 12.0 0.600 7.0 0.350 2.0 0.100 16.9 0.845 11.9 0.595 6.9 0.345 1.9 0.095 16.8 0.840 11.8 0.590 6.8 0.340 1.8 0.090 16.7 0.835 11.7 0.585 6.7 0.335 1.7 0.085 16.6 0.830 11.6 0.580 6.6 0.330 1.6 0.080 16.5 0.825 11.5 0.575 6.5 0.325 1.5 0.075 16.4 0.820 11.4 0.570 6.4 0.320 1.4 0.070 16.3 0.815 11.3 0.565 6.3 0.315 1.3 0.065 16.2 0.810 11.2 0.560 6.2 0.310 1.2 0.060 16.1 0.805 11.1 0.555 6.1 0.305 1.1 0.055 16.0 0.800 11.0 0.550 6.0 0.300 1.0 0.050 15.9 0.795 10.9 0.545 5.9 0.295 0.9 0.045 15.8 0.790 10.8 0.540 5.8 0.290 0.8 0.040 15.7 0.785 10.7 0.535 5.7 0.285 0.7 0.035 15.6 0.780 10.6 0.530 5.6 0.280 0.6 0.030 15.5 0.775 10.5 0.525 5.5 0.275 0.5 0.025 15.4 0.770 10.4 0.520 5.4 0.270 0.4 0.020 15.3 0.765 10.3 0.515 5.3 0.265 0.3 0.015 15.2 0.760 10.2 0.510 5.2 0.260 0.2 0.010 15.1 0.755 10.1 0.505 5.1 0.255 0.1 0.005 To check what RANGE (Full-Scale for 0-1 volt output) is selected, press and hold SW4 (see page 4) and measure the voltage out at TP7. As an example, TP7 will read 0.50 volt for a full-scale RANGE of 10.0% CO. The STAR will be relative to the new full scale RANGE. As an example, a STAR of 5.0% CO with a full-scale RANGE of 10% will give a 0 to 1 volt output of 0.5 volt when the STAR switch SW2 is pressed. A STAR of 5.0% CO with a full-scale RANGE of 20% will give a 0 to 1 volt output of.25 volt when the STAR switch SW2 is pressed. CAUTION: It is easy to get confused when you are pressing these different switches. Please make sure to double-check where you have set the RANGE and STAR. If you attempt to gas calibrate and the STAR does not match the certified gas % when you press the SPAN SW3, the sensor will obviously give you incorrect measurements. Version 20.33 and later disables the digital and the analog 0 to 1 volt outputs during the first 40 seconds after a power-on RESET to prevent and out-of-range signal from being transmitted. Model 2015SPI-CO 20% CO Page 5 of 5 Revised: 7/18/2017 1994-2017 VALTRONICS, INC.