User's Guide. Model W 3-in-1 Switching DC Power Supply

Similar documents
NTP-5521/5531/5561 SWITCHING MODE POWER SUPPLY

Laboratory Switch Mode Power Supply Series. Instruction Manual

HCS-3600 / 3602 / 3604 Laboratory Grade & High RFI Immunity Switching Mode Power Supply with Rotary Encoder Control

CIRCUIT-TEST. Switching Mode Power Supply with USB Remote Control PSC VDC / 5A PSC VDC / 3A PSC VDC / 1.

NTP-6621 / 6631 / 6661 SWITCHING MODE POWER SUPPLY with USB REMOTE CONTROL

High Precision Quad Output DC Power Supply

User's Guide. Model High Precision Quad Output DC Power Supply

HCS-3200/3202/3204 Laboratory Grade & High RFI Immunity Switching Mode Power Supply with Rotary Encoder Control

SIM-9106/9303 High Current Switching Mode Power Supply with C.C., C.V., Remote Control & Sensing User Manual

Universal AC Power Source + AC Power Analyzer

Model DCP60 and Model DCP Switching Mode DC Power Supplies 60V / 10A Adjustable USER GUIDE

Model DCP42 160W Laboratory Grade Constant Power, Switching Mode Power Supply

HCS-3200/3202/3204 Laboratory Grade & High RFI Immunity Switching Mode Power Supply with Rotary Encoder Control

DC POWER SUPPLY. Model: PW-3063R / PW-4063R

CIRCUIT-TEST OPERATION MANUAL. Switching Mode Power Supply PSC-6916 PSC-6932 PSC A 30A 15A

CIRCUIT-TEST OPERATION MANUAL. Switching Mode Power Supply PSC-6118 PSC-6136 PSC A 5A 2.5A

ADC7520 SERIES. 1600W Battery Chargers and Power Supplies

Mini Digital Multimeter

User Manual. 400Amp AC Clamp Meter + NCV. Model MA430. Additional User Manual Translations available at

Model A Mini AC/DC Clamp Meter. User's Guide

User's Guide. Extech AM A AC Analog Clamp Meter

User's Guide. MiniTec TM Series Model MN25 MultiMeter

USER GUIDE. Dual Input True RMS AC Voltage/Current Datalogger. Model DL160

80W (16V/5A) Power Supply with USB Output. User Manual

99 Washington Street Melrose, MA Phone Toll Free Visit us at INSTRUCTION MANUAL

LABORATORY GRADE REMOTE PROGRAMMING SWITCHING MODE DC REGULATED POWER SUPPLIES SDP / 3636 / User Manual

Mini Digital Multimeter

User s Guide. 600A AC Clamp Meter. Model 38387

1-36V, 0-3A DC Power Supply

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODELS:

User's Guide. Digital Multimeter. Model MN42

User s Guide. 600A True RMS AC/DC Clamp Meter. Model 38389

User's Guide. Phase Sequence and Motor Rotation Tester Model

Mini Digital Multimeter Model MN15. User's Guide

USER MANUAL. RND 320-KA3000 Series

User's Guide. Programmable DC Power Supply 200 Watt (40 Volts / 5 Amps) Model Introduction

BENCHTOP INSTRUMENT. Programmable Switching DC Power Supply Operation Manual V1.0

Mini Digital Multimeter

Dual Range DC Power Supply

Multiple Output Linear DC Power Supply

User s Manual PROGRAMMABLE DC POWER SUPPLY

Earth Resistance and Resistivity Tester

Autoranging Mini Multimeter

Model HM-535 Power Supply Installation and Service Instructions

USER MANUAL. Mini Multimeter with Non-Contact Voltage Detector (NCV) Model EX310

Installation Manual INTELLIGENT SERVER. Model name: BMS-LSV6UL

INSTRUCTION MANUAL. Model True RMS AC/DC 30A Mini Clamp-on Meter. Introduction. True RMS AC Current and Voltage

DUAL LASER PACK & CONTROLLER USERS MANUAL

Features. Line Regulation. Load Regulation. Safety. Cooling. Input Connectors. Output Connectors. Overvoltage Protection. Over Temperature Protection

True RMS AC Voltage/Current Datalogger

USER S MANUAL DC REGULATED POWER SUPPLY

Variable Regulated DC Power Supplies PAD-LA Series

Features. Conducted Emission. Radiated Emission. Line Regulation. Load Regulation. Safety. Cooling. Input Connectors.

Dual Range DC Power Supply

Features. Conducted. Emission. Line Regulation. Load Regulation. Safety. Cooling. Input Connectors. Overvoltage Protection. Phase Loss.

AutoRanging Digital MultiMeter

User s Manual. Models Digital Force Gauge. Additional User Manual Translations available at

USER S MANUAL Linear Programmable DC Power Supply APS

Series 2200 Multichannel Programmable DC Power Supplies User Manual

HERCULES-5000 INSTRUCTION MANUAL

Keysight U8030 Series Triple-Output DC Power Supplies

Installation and Operation Back-UPS BR1000G-IN / BR1500G-IN

Autoranging Multimeter plus IR Thermometer Extech 450 Patented

USER MANUAL. True RMS Multimeter. Extech EX430A

TRC-190 User s Manual

LABORATORY GRADE. REMOTE PROGRAMMING SWITCHING MODE DC regulated Power Supplies

Heat Stress WBGT Meter

User's Guide. 800 Amp Clamp Meters. EX710 AC Clamp meter EX720 True RMS AC Clamp meter EX730 AC/DC True RMS Clamp meter

Part No. Z , IA Jul OPERATION MANUAL. High Voltage Digitalmeter A

DC Power Supply Models AX501, AX502 & AX503

Operating Instructions. LSP-KONSTANTER Series 32K Laboratory Power Supply 8/10.13

General Warranty. For more details, please refer to the user manual on the supplied CD, it can also be downloaded at

OVEN INDUSTRIES, INC.

PS-3005P Programmable DC Power Supply

An ISO 9001 Company. KLP POWER SUPPLY

Series Watt DC Power Supplies

INSTALLATION DKM-409 NETWORK ANALYSER WITH HARMONIC MEASUREMENT AND SCOPEMETER. Before installation:

Wise HP33 THREE PHASE HIGH PRECISION AVR SURVO-MOTOR AUTOMATIC VOLTAGE STABILIZER

MCCB-500 MOLDED-CASE CIRCUIT BREAKER TESTER

AP SERIES SYSTEMS POWER SUPPLIES. Autoranging power supplies for bench or systems use FEATURES

PLC-24V10AL(-PT/-TH) Quick Start Manual (Rev.1.10)

VBAR 270. User Manual. Version 1.2

INSTALLATION MANUAL SLI 50 INVERTER

OPERATING PROCEDURES for the UPS-S2 Universal Power Supply Series 2

BENCHTOP INSTRUMENT. Programmable DC Power Supply Operation Manual V1.0

INTELLIGENT SERVER. Model: BMS-LSV6E

Autoranging Mini Multimeter Model MN16A

ETHOS Auto Ranging Digital Multimeter

DIGITAL INPUT/OUTPUT RELAY INTERFACE BMS-IFDD01E

Safety. Features 1 ID: Make all the connections before you plug in the main power. DMX512 Controllable and RDM Configurable

Blue Weapon Laser. User manual UK. Version 1.0

MAINTENANCE MANUAL. EDACS REDUNDANT POWER SUPPLY SYSTEM 350A1441P1 and P2 POWER MODULE CHASSIS 350A1441P3, P4, AND P5 POWER MODULES TABLE OF CONTENTS

Installation Job Aid for Ethernet Routing Switch 3600 Series

R5050. Model. Instruction Manual. TRMS AC/DC Clamp Meter. reedinstruments. www. com

DLA. DMX512 Analyzer. DLA Users Manual SV2_00 B.lwp copyright ELM Video Technology, Inc.

Z-350 FAZER USER MANUAL

BASIC PA AMPLIFIER A-1031 A-1061 A-1121 OPERATING INSTRUCTIONS TABLE OF CONTENTS

OWNER S MANUAL 9908-TE. HIGH PRECISION AUTO-RANGING DC/True RMS AC BENCH-TOP DIGITAL MULTIMETER

Mini Multimeter with Non-Contact Voltage Detector (NCV)

Spectrum Laser. User manual UK. Version 1.0

Transcription:

User's Guide Model 382260 80W 3-in-1 Switching DC Power Supply

Introduction Congratulations on your purchase of the Extech 80W 3-in-1 Switching DC Power Supply. The Model 382260 has three output ranges that perform like 3 power supplies in one. It can be used for many applications including bench testing, field service, hobby and telecommunication equipment use. This power supply is shipped fully tested and calibrated and, with proper use, will provide years of reliable service. Safety This manual contains important safety and operation instructions for correct use of the power supply. Read through the manual and pay special attention to the markings and labels of this unit and equipment to be connected. Pay special attention to these two types of notices used in this manual WARNING : Failure to observe this warning may cause injury to persons and damage to power supply or connected equipment. CAUTION : Failure to observe this warning may result in damage to equipment and Improper functioning of the power supply. WARNING : 1. Do not use this power supply near water. 2. Do not operate or touch this power supply with wet hands. 3. Do not open the casing of the power supply when it is connected to ac mains. 4. Refer all servicing to qualified service personnel only. 5. Before replacing the AC fuse find out and clear up the cause first. 6. Replace the AC fuse with the same type and rating as the original fuse. CAUTION : 1. Use a grounded 3 pin AC source. 2. This unit is for indoor use only. 3. Do not operate or place this unit in a humid, dusty, in direct sunlight location or near any heat source. 4. Before plugging into local AC mains, check with the rating label at the back of the unit. 5. Do not block any ventilation openings of the unit. 6. This unit must be used within the specified rating; regular excessive continuous loading may cause damage to the power supply. 7. The gauge size of input power cable must be at least 0.75mmsq and the total length of power cable must not exceed 3m. 8. Input Fuse Recommended: T3.15AL250V Time-delay / GDC (5x20mm) 2

Power Supply Description 1. Power On/Off switch 2. Output terminals 3. UVL adjustment screw 4. Current adjustment 5. Voltage adjustment 6. Upper Voltage Limit button 7. Review button 8. Range selector 9. Output On/Off button 10. UVL indicator 11. Constant Voltage indicator 12. Constant Current indicator 13. Overvoltage or Temperature alarm 14. Slave indicator 15. LED Voltage Display 16. LED Current Display 17. Master/Slave switch 18. Master/Slave control terminals 19. Remote Sensing terminals 20. Rear Output terminals 21. AC power socket 3

Stand Alone Operation 1. Set the Master/Slave switch located on the back of the unit to the Master position. 2. Without any load attached, use the POWER button to switch on the power supply and the LED indicator will light. There should not be any reading on the voltmeter and ammeter. 3. Press the Output On/Off button and its green LED will light up, the voltmeter will display the previously set output voltage. Use the voltage control knob to adjust the voltage to the desired level. 4. Use the Output On/Off button to turn off the output. Note: The output will automatically shut off when one of the other voltage/current range selection buttons is pushed. This is to prevent damage to a connected load if the voltage/current values are set too high. 5. Press and Hold the Review button to display the previously set voltage and current limiting values. With this button held down, you can set the current limiting value without connecting to a load or shorting the output terminals. Note: The Output button must be in the OFF position. Note: Front and rear output terminals are connected together internally. Setting the Upper Voltage Limit (UVL) Value The Upper Voltage Limit Value is an added protection for voltage sensitive loads. When the output voltage exceeds the set UVL, the output terminal will automatically shut off and the ALM LED will light. Note: Only one UVL value can be set for all three ranges. 1. Press the UVL/Voltage and insert a small screwdriver into the UVL adjustment screw. 2. Turn clockwise to increase the voltage limit and counter-clockwise to decrease the UVL value. Remote Sensing Operation When the output current is high or the load connection is long, there is a voltage drop across the connecting cables. This results in a difference between the voltage at the output terminals and load point. By connecting the Remote Sense Terminals to the load point, the reading at the output display and load point will be the same. Caution: Observe correct polarity and never short the Remote Sensing terminals together. 1. Connect power supply Output terminals to the Load. (Use the front Output terminals to source higher current) 2. Connect the Remote Sensing Terminals to the Load. (Push in on the Orange tabs to open the wire clamp inside the connector) Connect RS+ to + Positive connection at the Load. Connect RS- to -Negative connection at the Load. Note: Always disconnect the Remote Sense terminal connections first before the Load connections. 4

Master/Slave Operation Two or more Model 382260 Power Supplies can be connected in parallel to increase output current to the sum of the connected supplies. In this mode of operation, the Master power supply will control all of the voltage and current settings of the Slave units. 1. Set all of the power supplies to the same upper voltage limit (UVL) and voltage/current range. 2. Set the voltage and current limit of all of the slaves to maximum values. 3. Switch off all of the power supplies. 4. Set the Master/Slave switch to the Slave position on all of the Slave units. 5. Connect the power supplies together using the Master/Slave control circuit on the back of the power supplies. Note: For proper performance all connecting cables and output cables should be the same gauge and length. 6. Using either the output terminals on the front or back of the units, connect the output cables in parallel to the load. 7. Switch on the Master unit first and set to the desired voltage. 8. Switch on the slave units and the Slave Indicator LED will light to confirm correct connection. 9. All output voltages and currents are now controlled by the Master unit. 5

Specifications Display Dual 4-digit LED Output voltage accuracy +/- (1.0% (reading) + 15 digits) Output current accuracy +/- (1.0% (reading) + 15 digits) Display Accuracy +/- (1.0% (reading) + 15 digits) Voltage Output, DC 0.0 to 16.40V, 0.10 to 27.60V, 0.10 to 36.80 Current Output, DC 0.100 to 5.100A, 0.100 to 3.100A, 0.1 to 2.00A Ripple and Noise <30mVp-p Line Regulation < 4mV/<10mA Load Regulation < 20mV/<10mA Efficiency >75% Power Factor >0.9 Operating %RH 10 to 80% RH Altitude 2000m Installation Category CAT 2 Pollution degree 2 Power 90 to 250VAC (47 to 63Hz) Fuse 3.15A/250V (T3.15AL250V) Time-delay/ GDC (5x20mm) Supply voltage fluctuation ± 10% of stated operating voltage Dimensions 13 x 5 x 2" (330 x 127 x 53.5mm) (DxHxW) Weight 4.2lbs. (1.9 kg) 6

Trouble Shooting PROBLEM INDICATIONS POSSIBLE CAUSES SUGGESTED SOLUTIONS Panel display, LED Indicators not on Power supply not working No DC output power Voltage meter reading grossly inaccurate Ampere meter reading grossly inaccurate The activated (auto- cross over ) current limiting value is different from the preset cc value Voltage meter zero indication. ALM (13) Alarm led on CC (12) led on Actual output voltage is grossly different from the meter read out In CV mode, the actual current measured is grossly different from the Amp meter reading All indicators and display are normal, only CC mode has the problem A. AC power input not connected B. AC input fuse blown A. Output on off button not on. B. UVL (10 ) protection triggered C. OTP protection triggered D. Output short circuit A. Volt meter shows not the output voltage. B. Possible misalignment in voltmeter calibration Possible misalignment in calibration. For stand alone unit, Master & Slave Switch (17) in wrong position (Slave). Wrong setting or connection in Master & Slave or s e t u p procedure. A. Check AC power connection, B. Contact factory A. Check output LED (9) is on or not, push (9) to on. B. Check UVL set voltage by pushing (6), re-set UVL to applicable limit. C. Check vent holes at top and bottom is clear, ambient temperature too high D. Check and undo short circuit of output connection. A. Check the LED indicator of UVL on the top right hand side of the Volt meter, if it lights up then the volt meter only shows the set UVL value. Push (6) to go back to output voltage. B. Contact factory Contact factory Check the Master & Slave switch (17) correct position. Use stand alone mode to check power supply separately without any connection to slave units. Follow the procedure carefully and make sure there is only one Master unit. Copyright 2012-2015 FLIR Systems, Inc. All rights reserved including the right of reproduction in whole or in part in any form. www.extech.com 7