New approach to jitter reduction of an x-ray streak camera in accumulation mode

Similar documents
Text for the class, Pump and probe technique for picosecond time-resolved x-ray diffraction at the Cheiron School

Laser readiness for all optical EUV FEL

Section 3 DEVELOPMENT IN SUBPICOSECOND RESEARCH

Text for the class, Pump-Probe Technique for Picosecond Time-resolved X-ray Diffraction at Cheiron School

Time-Resolved measurements by FEL spontaneous emission: A proposal for sub-picosecond pumps & probe structural and spectrometric investigations

Detectors for Future Light Sources. Gerhard Grübel Deutsches Elektronen Synchrotron (DESY) Notke-Str. 85, Hamburg

Gating Image Intensifiers

ClEtVEb. Ultrafast X-ray Diffraction of Laser-Irradiated Crystals. DEC 0 8 p997 OSTI ERNEST ORLANDO LAWRENCE BERKELEYNATIONAL LABORATORY

Letter of Intent for provision of ultrafast streak camera-based detectors for LCLS experiments

Attosecond Optics. Zenghu Chang Kansas State University University of Central Florida

Development of automated ultraviolet laser beam profiling system using fluorometric technique

Company Pioneer in Ytterbium ultrafast lasers High quality manufacturing Intense and active R&D Located in Bordeaux and Paris US offices in Boston and

Physics 625 Femtosecond laser Project

Technical Specifications for High speed PIV and High speed PIV-PLIF system

Synchronization and pump-probe experiments

Section 2 PROGRESS IN LASER FUSION

L. Pina, A. Fojtik, R. Havlikova, A. Jancarek, S.Palinek, M. Vrbova

APPLICATIONS KEY FEATURES. High-speed intensified Camera Attachment

LIFA KEY FEATURES APPLICATIONS. Fluorescence Lifetime Attachment LIFA 15001A02 16/03/2015


Progress of the Thomson Scattering Experiment on HSX

High-Power Femtosecond Lasers

TRiCAM APPLICATIONS KEY FEATURES. Time Resolved intensified CAMera. TRiCAM 13001A01 31/10/2013

ADVANCED TEMPORAL CONTROL IN HIGH ENERGY ULTRAFAST LASERS DEDICATED TO MICROMACHINING AND BIOMEDICAL APPLICATIONS

Gated MCP framing camera system

LIFA. SPECIFICATIONs. Fluorescence Lifetime Attachment LIFA14001A02 25/02/2014

Development of Ultrafast CXRS system in Heliotron J. Graduate School of Energy Science Kyoto University LU XIANGXUN 03/15/2016

Image Diode APPLICATIONS KEY ATTRIBUTES

Detector systems for light microscopy

Real-world applications of intense light matter interaction beyond the scope of classical micromachining.

A CAMERA SHUTTER USED IN HIGH RESOLUTION BUBBLE CHAMBER PHOTOGRAPHY* J. D. Ferrie, R. C. Field and D. A. Talaska

Compatible with Windows 8/7/XP, and Linux; Universal programming interfaces for easy custom programming.

Recent Advances in Ultrafast Laser Subtractive and Additive Manufacturing

Active Light Time-of-Flight Imaging

Pump-probe. probe optical laser systems at FLASH. S. Düsterer

Spectroscopic equipment. Multispectral Imaging

PI-MAX 4: 1024 x 256 SPECTROSCOPY GROUP

Pump laser requirements for next generation lasers.

PI-MAX 4: 1024 x 256

PICOSECOND TRANSIENT THERMAL IMAGING USING A CCD BASED THERMOREFLECTANCE SYSTEM

Use of streak camera for time-resolved

UV-NIR LASER BEAM PROFILER

PI-MAX 4: 1024 EMB. Applications:

1. THEORETIC STUDY OF MCP GATED BY PICOSECOND PULSE

XUV pulse FLASH

P R I C E L I S T. Laser Diagnostic Tools

Development of planar microchannel plate photomultiplier at Argonne National Laboratory

THIRD GENERATION UED INSTRUMENTATION *

CinCam CCD - Technical Data -

Analysis of Cornell Electron-Positron Storage Ring Test Accelerator's Double Slit Visual Beam Size Monitor

GuideStar II Customer Presentation. February 2012

SYSTEM 8500 Power Converter Family

CHAPTER 4 DUAL LOOP SELF BIASED PLL

PI-MAX 4: 1024i-RF. Compatible with Windows 8/7/XP, and Linux; Universal programming interfaces for easy custom programming.

Single Photon Counting System

Comparison of Beam Shapes and Transmission Powers of Two Prism Ducts

The goal of this lab is to give you a chance to align and use a Pockel s Cell.

Optical properties and characterization

Condenser Optics for Dark Field X-Ray Microscopy

DESY Summer Student 2008

Model REEF-SS ASF-200

PARTICLE IMAGE VELOCIMETRY (PIV) AND VOLUMETRIC VELOCIMETRY (V3V) SYSTEMS

Applications of Piezo Actuators for Space Instrument Optical Alignment

Feasibility of Laser Induced Plasma Micro-machining (LIP-MM)

Minimizes reflection losses from UV - IR; Optional AR coatings & wedge windows are available.

MEASUREMENT OF WIGNER DISTRIBUTION FUNCTION FOR BEAM CHARACTERIZATION OF FELs*

Picosecond Fluorescence Lifetime Measurement System C11200

UV PULSE SHAPER: SIMULATIONS AND EXPERIMENTAL CHARACTERIZATION. C. Vicario, D. Filippetto (INFN/LNF) Abstract

FEMTO OpTicsTM in us! O j rld wo T as F ra T ul in an E liv

TABLE OF CONTENTS PRODUCT DESCRIPTION CINCAM CCD TECHNICAL DATA SENSOR RESPONSE DIMENSIONS CINCAM CCD LARGE FORMAT TECHNICAL DATA SENSOR RESPONSE

Outline. Nuclear Forensics & Laser-Induced Breakdown Spectroscopy. Interferometry & Shadowgraphy. Conclusion & Future Work. Experimental Setup

PGx01 series. High Peak Power. Available models

Optics for nonlinear microscopy

RAL Talk 8th. April 2003 Fast Diagnostics Optical and X-ray

Spherical Crystal X-ray Imaging for MTW, OMEGA, and OMEGA EP

Comissioning of the IR beamline at FLASH. DESY.DE

Diamond FLQ-Series. Q-Switched Fiber Lasers FEATURES & BENEFITS

Realization of Tiled-Grating Compressors for the OMEGA EP Petawatt-Class Laser System

DC - Devices and Controller

Video frame rates and higher to efficiently synchronize with high repetition rate lasers

A second-order autocorrelator for single-shot measurement of femtosecond laser pulse durations

Three Dimensional Measurements by Deflectometry and Double Hilbert Transform

HyperRapid NX SmartCleave

Engineered Diffusers Intensity vs Irradiance

Picosecond Fluorescence Lifetime Measurement System C11200

Hydrodynamic Instability and Particle Image Velocimetry

Diagnostic Tools for the Transverse Coherence of an X-FEL

WAVELENGTH MANAGEMENT

dicam pro intensified digital 12 bit CCD camera system

Uniformity scan of the 6cm tubes. Jingbo Wang Argonne National Laboratory, Lemont, IL

Development of a Simulation Method of Three-Dimensional Ultrafine Processing by Femtosecond Laser Shunsuke Nabetani 1, a, Hideki Aoyama 1, b * 1, 2, c

WinCamD-LCM 1" CMOS Beam Profiling Camera, SuperSpeed USB 3.0, * nm * model-dependent

Picosecond fluorescence lifetime measurement system C11200

HOLOGRAPHIC FEMTOSECOND LASER PROCESSING AND THREE-DIMENSIONAL RECORDING IN BIOLOGICAL TISSUES

CALYS 50. Field multifunction calibrator for basic use. Calibration measurement

Ink Volume Displacement In An Impulse Printhead With Bilaminar Transducer

REVIEW OF STREAK CAMERAS FOR ACCELERATORS : FEATURES, APPLICATIONS AND RESULTS

Be careful not to leave your fingerprints on the optical surfaces of lenses or Polaroid sheets.

6-1 LECTURE #6: OPTICAL PROPERTIES OF SOLIDS. Basic question: How do solids interact with light? The answers are linked to:

Two slit interference - Prelab questions

Transcription:

New approach to jitter reduction of an x-ray streak camera in accumulation mode J. Liu *a, A. G. MacPhee b, C. Liu a, B. Shan c, Z. Chang c and J. Wang a a Advanced Photon Source, Argonne National Laboratory, Argonne, IL 60439 b Lawrence Berkeley National Laboratory, Berkeley, CA 94720 c J. R. Macdonald Laboratory, Kansas State University, Manhattan, KS 66506 ABSTRCT An x-ray streak camera operating in accumulation mode was developed for studying ultrafast dynamics at synchrotron facilities. A laser-triggered photoconductive switch was used as a sweeping unit to obtain low timing jitter. The fast rise time of the ramp pulse generated by the switch (90 ps) combined with the fast response of the traveling wave deflection plates (150 ps) significantly reduced the jitter caused by the shot-to-shot laser fluctuation. At ~1% rms (root mean square) laser energy fluctuation, the resolution of the camera is 1.1 ps when over 5000 laser shots were accumulated. This is two times better than that of the previous design with slower response (300 ps) deflection plates. Keywords: fast detector, streak camera, photoconductive switch, ultrashort laser, trigger jitter, synchronization 1. INTRODUCTION X-rays from synchrotron radiation have recently been used as structural probes of solids in the ultrashort time domain. [1-3] The pulse width of the x-ray from a third-generation light source, typically 50 to 100 ps, is limited by the electron bunch length in the storage ring. By combining such a long x-ray pulse with ultrafast x-ray detectors, it is possible to carry out ultrafast dynamic studies with a few picosecond or even subpicosecond time resolution. The time-resolved experiments can use a femtosecond laser as a pumping pulse, relatively long x-ray pulse from synchrotron radiation source as a probing pulse, and an x-ray streak camera as a fast detector. Accumulation of signals over thousands of laser and x-ray shots is required due to low signal level. In this operating mode, the resolution of the streak camera is limited by the timing jitter from shot to shot. The timing jitter of the commonly used avalanche transistor circuits for driving the deflection plates was on the order of tens of picosecond (rms). Therefore operating the streak camera in accumulating mode to achieve a few picoseconds time resolution is difficult. When a high-voltage photoconductive switch was first used on a streak camera as a sweeping circuit, [4] it reduces the shot-to-shot jitter to a few picosecond. [5,6] Compared with the avalanche transistor circuit, the photo-conductive switches exhibit much lower jitter and are commonly referred to as jitter-free switches. The timing jitter for photoconductive switches is caused by the shot-to-shot laser pulse energy fluctuation. The jitter has been reduced to subpicosecond levels by improving the laser stability. [7] To reduce the jitter even further, an extra GaAs switch can be used to compensate the sweep pulse amplitude resulting from laser intensity fluctuation. [8] Realizing that the jitter is inversely proportional to the rise time of sweeping ramp voltage, we demonstrated that the timing jitter is significantly reduced by using a laser-triggered, fast GaAs sweeping unit and fast-response meander-type deflection plates. The measured time resolution of the streak camera is 1.1 ps in accumulation mode with the laser RMS fluctuation being ~1%,

which is significantly better than a similar camera with slower switch and deflection plates. [6] 2. METHODS OF REDUCING TIMING JITTER The streak tube that was used in this work is similar to our previous design [9,10] except that we redesigned the deflection plates. In the streak tube, the x-ray-produced electrons from the photocathode are accelerated by the high field (10 kv/mm) between the photocathode and the anode. The electrons then enter the region between a pair of deflection plates. The deflection plates are driven by a ramp voltage produced by a photoconductive switch that will be described later. Finally, the electrons are focused on the microchannel plate /phosphor screen by a magnetic lens. It is well known that the temporal resolution of streak camera are limited mainly by the transit time dispersion of the photoelectrons as they travel from the photocathode to the deflection plates. It is also limited by the spatial resolution of the electron optics in the tube and by the sweeping speed. It was demonstrated before that the contribution from these two factors is about 0.5 ps for a previously developed camera. [10] When the camera operates in the accumulation mode, the dominating factor that limits the time resolution is the timing jitter. While the response time of the photoconductive switch to the triggering laser may be instantaneous, the timing jitter is not zero unless the laser pulse is absolutely stable. At the present, the lasers used in synchrotron facilities for time-resolved x-ray studies are Ti:sapphire lasers operated at kilohertz repetition rate. The same laser is used to pump the sample under investigation and to trigger the photoconductive switch. The pulse energy fluctuation is on the order of 1 % rms. This energy fluctuation causes the output amplitude of the photoconductor to change from shot to shot, which in turn produces timing jitter. The timing jitter caused by laser fluctuation is illustrated in Figure 1, where the time dependence of the output amplitude of the switch from two laser shots is shown. We assume that the first pulse (V 1 ) has higher amplitude than the (V 2 ). If properly calibrated, it will take a longer time for the second pulse to reach a certain level of driving voltage (for example, at half of the averaged pulse high, V a ), which corresponds to the center of the phosphor screen. The time difference (t 2 -t 1 ) represents the time jitter due to the switch response to the fluctuation of laser intensity. The jitter of the switch can be simply expressed as t jitter = 0.5T rise V V a. (1) Where T rise is the rise time of sweeping pulse, V/V a is the rms fluctuation in the shot-to-shot switch output voltage, which is related to the laser intensity fluctuation nonlinearly (if the switch is operated in a saturated mode to suppress the effect of the laser intensity fluctuation). Therefore, the jitter can be reduced by decreasing the rise time.

With the previous camera, [6] the rise time was limited to 300 ps by the bandwidth of the deflection plates. The plates were redesigned so that the response time improved to 150 ps. The measured response time is shown in Figure 2. We expected the two-fold reduction in the response time would result in a two-fold reduction of the timing jitter. amplitude 500 V a 0.5V a V 1 V 2 Amplitude (V) 400 300 200 100 0 150 ps t 1 t 2 time 47.4 47.6 47.8 48.0 48.2 48.4 48.6 Time (ns) Figure 1. Illustration of timing jitter caused by amplitude fluctuation of sweeping pulse. V a is an averaging amplitude of sweeping pulses. t = t 2 t 1 is the timing jitter. Figure 2. Response time (150 ps) of the newly designed deflection plates. It was measured by sending a pulse with 90 ps rise time to the deflection plate. A 20 GHz Tektronics sampling oscilloscope was used for the measurement. 3. PHOTOCONDUCTIVE SWITCH The overall rise time of the field in the deflection plates is determined by the response time of the deflection plates and the photoconductive switch. It is desirable that the response time of the switch is much smaller than that of the deflection plates. The design of our switch is eased by the fact that the deflection sensitivity of our streak tube is high at 20 cm/kv. Only ±50 volts are needed to sweep electrons across a 20-mm-diameter phosphor screen. The photoconductive switch was made from semi-insulate GaAs with a resistivity of 10 7 Ω-cm. The active area of GaAs is about 0.5 cm 2. To reduce the contact resistance, two copper electrodes were deposited on the switch with a gap of 2 mm between them. In order to get a fast response time and reduce the pulse reflection, the switch is mounted on a circuit board with a 50 Ω transmission line. Figure 3 shows a schematic diagram of the circuit. This design is much simpler and much easier to use than that of the ramp voltage generation with the two GaAs switch configuration.

600 90ps 500 400 (+) amplitude(v) 300 200 100 (-) 0 45.6 45.8 46.0 46.2 46.4 46.6 46.8 47.0 time(ns) Figure 3. Schematic diagram the of photoconductive switch circuit. (DP1 and DP2 : deflection plates. C 1, C 2 : DC blocks; R 3, R 4 : matching resistors). Figure 4. Symmetrical ramp pulse wave forms from the photoconductive switch with rise time 90 ps respectively, which used to drive the two arms of deflection plates. With our design, the two pulses are generated by the same switch at the same time. The effect of laser energy fluctuation on the timing jitter is reduced when a photoconductive switch is operated in saturated mode at high laser illumination flux. Only 20 µj laser energy is needed to saturate our photoconductive switch. As the laser flux increases, the peak of the output signal becomes more stable. However, the laser pulse flux can not be too high, since the pre-pulses and /or amplified spontaneous emission may start to distort the output of the switch or even damage the switch. Figure 4 shows the symmetrical ramp pulses produced by the switch measured by a TEK11801A Sampling Oscilloscope. The sweeping pulse rise time is 90 ps with 500 V amplitude. 4. MEASUREMENT OF TIMING JITTER The measurements were done using the newly established high-intensity laser facility at Kansas State University. A 25 nm gold layer coated on fused silica was used for the streak camera photocathode. The experimental setup is shown in Figure 5. Since the streak camera is also sensitive to uv light, we used ultrashort third-harmonic pulses to test the camera. The third-harmonic pulses of a Ti: sapphire laser are generated by focusing the 800 nm, 20 fs, 1 mj, 1 khz laser pulses in air with an f = 100 mm lens. A prism was used to separate the uv beam from the infrared beam. Shot-to-shot laser stability is about 1% evaluated by oscilloscope. A Michelson interferometer is used as a timing scale. One of the optical path lengths was varied to produce two pulses with known delay. The reflectivity of another arm s mirror is 95%, which provides about 20 µj laser energy to trigger the photoconductive switch. The GaAs photoconductor is triggered at 1 khz repetition rate. The delay time between the sweep pulses and the photoelectrons emitted by the photocathode can be adjusted continuously by moving the photoconductor switch on a rail. The streak images were integrated by a Liquid-Cooled Photometrics CCD camera (1024 1024 pixels). The static spatial resolution is about 120 µm full width half maximum. The streaked images have been accumulated on the CCD camera with a CCD integration time of 5 seconds. Figure 6 shows the lineout of the streak image obtained with the 5000 shots. The FWHM of the plot is 1.1 ps while the interval time between the two pulses is 3.3 ps. The sweep speed used in the calibration was 0.35 µm/fs on the CCD camera, comparable to the speed of light (0.3 µm/fs). The trigger jitter of the streak camera can be written as

t jitter = [ ( t ) 2 + t meas ( ) 2 ] 1 2 ss. (2) where t meas is the measured time resolution under accumulation mode, which is 1.1 ps. t ss is the single-shot time resolution, 0.5 ps, that was measured before. [10] Therefore the timing jitter is estimated to be 1 ps, which is two times smaller than what was measured before with a lower response deflection system. [6] 800nm 1mJ 25fs Streak L Prism C M1 M2 CCD M3 GaAs Figure 5. Experimental setup. 3.3ps 1200 1000 count 800 1.1ps 1.1ps 600 400 15 20 25 30 35 40 45 50 tim e (ps) Figure 6. Measured time resolution of the streak camera under accumulation mode. 5. CONCLUSION We have developed an x-ray streak camera driven by a photoconductive switch. We found that the timing jitter of the camera can be reduced significantly by reducing the rise time of the sweeping field. More specifically, the new

deflection plates with 150 ps response time resulted in a 1 ps jitter in the streak camera at 1% laser fluctuation. Work is in progress to reduce the jitter to the femtosecond level by further improving the photoconductive switch and the deflection plates. 6. ACKNOWLEDGMENTS This work is supported by the U.S. Department of Energy under contract no. W-31-109-ENG-38 and DE-FG02-00ER15082. References: 1. J. Larsson, P. A. Heimann, A. M. Lindenberg, P. J. Schuck, P. H. Bucksbaum, R. W. Lee, H. A. Padmore, J. S. Wark and R. W. Falcone, Appl. Phys. A66, pp. 587-591, 1998. 2. F. Raksi et al., J. Chem. Phys. 104, pp. 6066-6069, 1996. 3. A. M. Lindenberg, I. Kang, S. L. Johnson, T. Misalla, P. A. Heimann, Z. Chang, J. Larsson, P. H. Bucksbaum, H. C. Kapteyn, H. A. Padmore, R. W. Lee, J. S. Wark and R. W. Falcone, Phys.Rev.Lett. 84, pp.111-114, 2000. 4. W. Knox and G. Mourou, Opt. Commun. 37, pp. 203, 1981. 5. A. Maksimchuk, M. Kim, J. Workman, G. Korn, J. Squier, D. Du, M. Bouvier, D. Umstadter and G. Mourou, Rev. Sci. Instrum., 67, pp. 697-699, 1996. 6. J. Larsson et al., Optics Letters, 22, pp. 1012-1014, 1997. 7. G. A. Naylor, K. Scheidt, J. Larsson, M. Wulff and J. M. Filhol, Meas. Sci. Technol. 12, pp. 1858-1864, 2001. 8. C. Belzile, C. Y. Cote, T. Oksenhendler, D. Kaplan and J. C. Kieffer, Rev. Sci. Instrum. 73, pp. 1617-1620, 2002. 9. Z. Chang, A. Rundquist, J.Zhou, M. M. Murnane and H. C. Kapteyn, X. Liu, B.Shan, J. Liu, L.Niu, M. Gong and X. Zhang, Appl. Phys. Lett. 69, pp. 133-135, 1996. 10. Z. Chang et al., High Speed Photography and Photonics, SPIE 2869, pp. 971-975, 1996. * contact: ljinyuan@aps.anl.gov; phone 1 630 252-5348; fax 1 630 2529303