Commissioning and quality assurance of Calypso four dimensional target localization system in linear accelerator facility

Similar documents
Commissioning and quality assurance of Calypso four dimensional target localization system in linear accelerator facility

Initial Clinical Experience with 3D Surface Image Guidance

A novel-integrated quality assurance phantom for radiographic and nonradiographic radiotherapy localization and positioning systems

Thank-You Members of TG147 TG 147: QA for nonradiographic

7/31/2011. Learning Objective. Video Positioning. 3D Surface Imaging by VisionRT

Optical Guidance. Sanford L. Meeks. July 22, 2010

Clinical implementation of photon beam flatness measurements to verify beam quality

ph fax Deming Way Middleton WI USA

Calibration of Video Cameras to the Coordinate System of a Radiation Therapy Treatment Machine

DAILY LINAC QA BEAM QA

Choosing and Commissioning a Video Based Motion Management System

Data. ModuLeaf Mini Multileaf Collimator Precision Beam Shaping for Advanced Radiotherapy

Mech. Engineering, Comp. Science, and Rad. Oncology Departments. Schools of Engineering and Medicine, Bio-X Program, Stanford University

Automated Quality Assurance for Image-Guided Radiation Therapy

Tomotherapy Physics. Machine Twinning and Quality Assurance. Emilie Soisson, MS

Feasibility study of performing IGRT system daily QA using a commercial QA device

8/3/2016. Image Guidance Technologies. Introduction. Outline

New Technology in Radiation Oncology. James E. Gaiser, Ph.D. DABR Physics and Computer Planning Charlotte, NC

Performance and characteristics of an IR localizing system for radiation therapy

Investigation of tilted dose kernels for portal dose prediction in a-si electronic portal imagers

Table of Contents. About AQUA

Simple quality assurance method of dynamic tumor tracking with the gimbaled linac system using a light field

A NEW LASER FOCUS ON PATIENT SAFETY

Measurement of inter and intra fraction organ motion in radiotherapy using cone-beam CT projection images

INSTRUCTIONS RPD INFORMATION RPD PRODUCT INFORMATION. Address 5218 Barthel Industrial Drive Albertville, MN Website

Computational modelling and analysis: IOP Medical Physics Group

Digital phantoms for the evaluation of a software used for an automatic analysis of the Winston-Lutz test in image guided radiation therapy

PyCMSXiO: an external interface to script treatment plans for the Elekta CMS XiO treatment planning system

1. Learn to incorporate QA for surface imaging

IAEA-TECDOC-1583 Commissioning of Radiotherapy Treatment Planning Systems: Testing for Typical External Beam Treatment Techniques

Using a research real-time control interface to go beyond dynamic MLC tracking

Calibration and quality assurance for rounded leaf-end MLC systems

THE WIRELESS PHANTOM PERFORM ACCURATE PATIENT QA IN LESS TIME THAN EVER!

Centralite CT Moving Laser Patient Positioning System (MRR-1)

IMSURE QA SOFTWARE FAST, PRECISE QA SOFTWARE

Monte Carlo methods in proton beam radiation therapy. Harald Paganetti

UNCOMPROMISING QUALITY

FIRST DEMONSTRATION OF COMBINED KV/MV IMAGE-GUIDED REAL-TIME DYNAMIC MULTILEAF-COLLIMATOR TARGET TRACKING

IMRT and VMAT Patient Specific QA Using 2D and 3D Detector Arrays

PET-CT in Radiation Treatment Planning

A NEW APPROACH IN THE DETERMINATION OF PATIENT S REFERENCE POINT IN CONFORMAL EXTERNAL RADIOTHERAPY

A simple method to test geometrical reliability of digital reconstructed radiograph (DRR)

Gregory Walsh, Ph.D. San Ramon, CA January 25, 2011

Disclosure. Outline. Acknowledgments. Ping Xia, Ph.D., Page 1. LINAC and MLC QA for IMRT. Received research support from Siemens Medical Solutions

State-of-the-Art IGRT

I. INTRODUCTION. Figure 1. Radiation room model at Dongnai General Hospital

Electromagnetic Detection and Real-Time DMLC Adaptation to Target Rotation During Radiotherapy

An approach for measuring the spatial orientations of a computed-tomography simulation system

CARS 2008 Computer Assisted Radiology and Surgery

Verification of dynamic and segmental IMRT delivery by dynamic log file analysis

3DVH : SUN NUCLEAR On The Accuracy Of The corporation Planned Dose Perturbation Algorithm Your Most Valuable QA and Dosimetry Tools *Patent Pending

INTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1. Modifications for P551 Fall 2013 Medical Physics Laboratory

TomoTherapy Related Projects. An image guidance alternative on Tomo Low dose MVCT reconstruction Patient Quality Assurance using Sinogram

Determination of rotations in three dimensions using two-dimensional portal image registration

URGENT IMPORTANT FIELD SAFETY NOTIFICATION

Application Notes for Team Hydrostatic Pad Bearings

FAST, precise. qa software

LASER SOLUTIONS. Ensure Accurate Setup from Simulation to Treatment. MOVEABLE: CT SIM+ // FIXED: MICRO and MICRO+

Version 5.6. Quick Start Guide

Acknowledgments. Ping Xia, Ph.D., UCSF. Pam Akazawa, CMD, UCSF. Cynthia Chuang, Ph.D., UCSF

Laser Systems for Patient Alignment

Determination of maximum leaf velocity and acceleration of a dynamic multileaf collimator: Implications for 4D radiotherapy

Image Guidance and Beam Level Imaging in Digital Linacs

ICARO Vienna April Implementing 3D conformal radiotherapy and IMRT in clinical practice: Recommendations of IAEA- TECDOC-1588

An Automated Image-based Method for Multi-Leaf Collimator Positioning Verification in Intensity Modulated Radiation Therapy

8/4/2016. Emerging Linac based SRS/SBRT Technologies with Modulated Arc Delivery. Disclosure. Introduction: Treatment delivery techniques

ADVANCING CANCER TREATMENT

The team. Disclosures. Ultrasound Guidance During Radiation Delivery: Confronting the Treatment Interference Challenge.

ADVANCING CANCER TREATMENT

MRI-LINAC Dynamic Phantom

Improvement and Evaluation of a Time-of-Flight-based Patient Positioning System

Image Co-Registration II: TG132 Quality Assurance for Image Registration. Image Co-Registration II: TG132 Quality Assurance for Image Registration

Automatic measurement of air gap for proton therapy using orthogonal x ray imaging with radiopaque wires

Data Fusion Virtual Surgery Medical Virtual Reality Team. Endo-Robot. Database Functional. Database

Position accuracy analysis of the stereotactic reference defined by the CBCT on Leksell Gamma Knife Icon

Multi-Purpose Body Phantom

THE SIMULATION OF THE 4 MV VARIAN LINAC WITH EXPERIMENTAL VALIDATION

Verification of Gated Radiation Therapy: Dosimetric Impact of Residual Motion

Chapter 9 Field Shaping: Scanning Beam

INNOVATIVE LASER POSITIONING SOLUTIONS IN RADIATION ONCOLOGY.

Facility Questionnaire PART I (General Information for 3DCRT and IMRT)

Photon beam dose distributions in 2D

IHE Radiation Oncology Technical Framework Supplement. Treatment Delivery Plan Content (TDPC) Rev. 1.1 Trial Implementation

Use of TrueBeam developer mode for imaging QA

An investigation of temporal resolution parameters in cine-mode four-dimensional computed tomography acquisition

Volumetric Modulated Arc Therapy - Clinical Implementation. Outline. Acknowledgement. History of VMAT. IMAT Basics of IMAT

A novel method to correct for pitch and yaw patient setup errors in helical tomotherapy

A novel phantom and procedure providing submillimeter accuracy in daily QA tests of accelerators used for stereotactic radiosurgery *

SUMMARY OF DENTAL MEASUREMENT PROCEDURES (Abridged Protocol)

Design approach for a Highly Accurate Patient Positioning System for NPTC

Evaluating the Performance of a Vehicle Pose Measurement System

Cemar Electro Inc. MAXX-1100 Laser Positioning System. Installation and Set-up Manual

Room scatter effects in Total Skin Electron Irradiation: Monte Carlo simulation study

Evaluation of AutoQA Lite TM Image Quality Measurement Software

Contents. - i - Ver.:2

IMRT site-specific procedure: Prostate (CHHiP)

Real-time self-calibration of a tracked augmented reality display

Cemar Electro Inc. MAXX-1100Laser Positioning System. Installation and Set-up Manual

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

Evaluation of 3D Gamma index calculation implemented in two commercial dosimetry systems

Transcription:

Technical Note Commissioning and quality assurance of Calypso four dimensional target localization system in linear accelerator facility K. R. Muralidhar, Krishna Komanduri, Birendra Kumar Rout, K. K. D. Ramesh Department of Radiation Oncology, American Oncology Institute, Hyderabad, Andhra Pradesh, India Received on: 14.02.13 Review completed on: 19.05.13 Accepted on: 20.05.13 ABSTRACT Four dimensional (4D) target localization system (Calypso System) was installed at our hospital, which is equipped with Beacon Transponders, Console, Electromagnetic Array, Optical System, Tracking Station, Treatment table overlay, and Calypso kvue Couch top. The objective of this presentation is to describe the results of commissioning measurements carried out on the Calypso System to verify the manufacturer specifications and also to evolve a quality assurance (QA) procedure which can be used to test its performance routinely. The QA program consists of a series of tests (QA for checking the calibration or system accuracy, Camera Calibration with L frame fixture, Camera Calibration with T frame fixture, System calibration Fixture targets test, Localization, and Tracking). These tests were found to be useful to assess the performance of the Calypso System. Key words: Calypso system, electromagnetic array, target localization, tracking Introduction The Calypso System is designed to provide accurate, objective, and continuous localization of a treatment target for patient alignment and target position monitoring during radiation therapy. Use of the Calypso System for target localization is based on the system s detection of electromagnetic signals generated by markers called Beacon transponders. Studies of this electromagnetic tracking system proved sub millimeter position measurement accuracy during real time tracking. [1 4] Address for correspondence: Dr. K. R. Muralidhar, Department of Radiation Oncology, American Oncology Institute, Nallagandla, Lingampally, Hyderabad 500 019, Andhra Pradesh, India. E mail: krmuralidhar@rediffmail.com Quick Response Code: Access this article online Website: www.jmp.org.in DOI: 10.4103/0971-6203.116376 Implanted Beacon transponders are indicated for permanent implantation in the prostate [5,6] and the periprostatic tissue (i.e., prostatic bed) to align and monitor the treatment isocenter in real time during radiation therapy. Surface Beacon transponders are indicated for temporary external placement on the skin, to monitor respiratory motion, and other patient motion in real time during radiation therapy. Tracking accuracy, setup accuracy, and precision at any point in a treatment or setup is computed as the magnitude (length) of the average error vector over 300 valid target positions for a three beacon plan or 200 valid target positions for a two beacon plan. There are three independent rotations (Rx (pitch), Ry (roll), and Rz (yaw)) around the Optical Localization Subsystem frame axes (Xol, Yol, and Zol), respectively. Calypso System is high precision equipment and requires not only proper operation but also stringent quality assurance checks. [7] It is essentially to carry out routine QA checks to keep calypso operating safely, effectively, and reliably. The QA program consists of several tests, covering QA for system accuracy, Camera Calibration with L frame fixture, Camera Calibration with T frame fixture, System calibration test, Localization and Tracking. [8] In this paper, we have described the manufacturer specified procedures of the tests included in the QA program and presented the results of the measurements. Journal of Medical Physics, Vol. 38, No. 3, 143-7, 2013

144 All safety tests are in compliance with the IEC60601 1 2 recommendations. [9] Materials and Methods Calypso System components The Calypso System has five main components: [10] Beacon transponders, the console, the electromagnetic array, the optical system, and the tracking station. Beacon transponders Beacon transponders [Figure 1], three in number, are used as permanent or surface transponders in or near the treatment target as fiducial markers during radiation therapy. Each Beacon transponder consists of a sealed glass capsule containing a miniature electrical circuit measuring 1.85 mm in diameter and 8.0-8.7 mm in length. They do not have internal energy source and are passive. Figure 1: Beacon transponder components Transponders emit signals when excited by the non ionizing electromagnetic field generated by the system s array. Each transponder emits a response signal at a unique frequency specific to that transponder which can be detected by sensors within the array. The system then interprets the shape of the transponder signal measured across the array to determine the position of each transponder. Console The Calypso System console [Figure 2] is a movable unit contains the system components that generate and detect the electromagnetic signals used for patient alignment and continuous monitoring of target position. The console contains electronic power supply, system control components, computer with Calypso System software, and a touch screen display. Calypso System software provides graphics and numerical data for performing quality assurance and calibration procedures for positioning the array and the patient. Electro Magnetic array The array is a flat panel [Figure 2], which contains the electronic components that enable the Calypso System to excite the Beacon transponders and estimate their position from the transponders signals. The array is positioned over the patient during patient setup and remains in place during radiation treatment to track the motion of the treatment target relative to the machine isocenter. Optical targets in the array can communicate with the optical system to determine the array position with respect to the machine isocenter. The array is designed to be left in position during treatment. Optical system The optical system [Figure 2] is equipped with infrared cameras permanently mounted on the ceiling of the treatment room, a power supply and a hub unit. This system is connected to the tracking station in the control room Figure 2: Calypso system console optical handles with optical targets and optical system via Ethernet. The optical system continuously monitors the optical targets embedded in the top of the array to determine the position of the array relative to the machine isocenter. As long as five of the nine optical targets are visible to two of the three cameras, the system has adequate data to calculate the array s position continuously. The optical system is calibrated to the isocenter of the linear accelerator. Tracking station Tracking station [7] is located in the control room connected to optical system and console in the treatment room through Ethernet. Data continuously stream between the console and the tracking station to track the target position relative to the machine isocenter. Target position information is presented to the radiation therapist using simple and objective on screen graphics, numerical data, and auditory indicators when the target moves outside of predefined limits for target motion. Quality assurance for checking the calibration or system accuracy Before system is used a QA procedure was performed daily to ensure proper operation of the Calypso System. QA fixture phantom [Figure 3] and the Calypso fixture stabilizer tools were used. The QA fixture is a block

145 containing three Beacon transponders rests on three thumbwheels that allows adjusting the fixture s height. It has Scribe lines on 3 vertical sides and the top to permit alignment with the lasers in the treatment room. This procedure is used to verify the system performance using Beacon transponders and to identify the changes in calibration or system accuracy. Carbon fiber table top was removed by Calypso kvue Couch top. QA was selected in the console computer. QA fixture was placed on the treatment table at isocenter. QA fixture level was adjusted with the help of thumbwheel feet and also table height was adjusted to align the lasers with the target lines on the QA fixture. Array is positioned within the acceptable limits from QA fixture. The screen continuously updates numerical and graphical position indicators that display the direction and distance the array should be moved or rotated, relative to the fixed position of the machine isocenter. Figure 3: QA fixture phantom With the help of software we can begin the test which takes less than 1 minute which shows the isocenter offset result as either pass or fail. If the test result is fail, reposition the fixture. This test evaluates the detected position of the fixture. The isocenter off set was calculated with the help of the following formula: D= X + Y + Z 2 2 2 QA QA QA D: Distance between QA fixture isocenter to Calypso System isocenter. X QA, Y QA, and Z QA are the distances of the QA fixture s isocenter from the calibrated Calypso System isocenter in the X, Y, and Z axes, respectively. Figure 4: L Frame fixture Camera calibration with L frame fixture Camera calibration was performed only if system calibration fails. The tools that were used were two fixtures, the L Frame fixture [Figure 4] and the T Frame fixture [Figure 5]. The L Frame fixture is shaped like an L and has four spherical optical reflectors. This fixture is also used to calibrate the system cameras. L frame calibration fixture was placed on top surface of the table such that fixtures corner reflector was as close as possible to isocenter and other reflectors on their respective laser planes. The four reflectors were correctly detected by the optical targets on the L Frame fixture by the infrared cameras. When less than four light sources were detected reflective objects from the camera view were uncovered if it exists and L Frame position was adjusted. In case of more light sources, unnecessary reflective objects were blocked. With the help of software, test was done which showed the camera calibration result as either pass or fail. Figure 5: T Frame fixture Camera calibration with T frame fixture Table was positioned such that it is out of view of the cameras (in the lowest vertical position and away from the gantry in the longitudinal direction). Gantry rotated to 180. Optical field of view was defined by moving the T Frame fixture [Figure 5] throughout the field of view using a small circular motion.

146 The T Frame fixture is shaped like a T and has three spherical optical reflectors on its cross bar and a cushioned handle. The field of view is a cube with it center 14 cm above isocenter. This process was done in three positions spending equal amount of time in each position. In the first position, target bar was horizontal and the handle is parallel to the table top. In the second position, target bar is vertical and the handle is tilted approximately 30 in clockwise. In the third position, target bar is vertical and the handle is tilted approximately 60 in anti clockwise direction. The reflective sphere was facing the wall opposite the Linear accelerator (LINAC). The software system evaluated the data that was collected. During calibration process, the three camera images were observed that each camera is detecting the T Frame fixture optical targets uniformly throughout the optical field of view. It is visible as an increasing volume of dots. When each camera image displays uniform coverage, test result is shown as pass. System calibration target test In this Calypso System, coordinate reference frame is calibrated to the machine isocenter using the isocenter fixture [Figure 6]. It contains three Beacon transponders and has scribe lines on three vertical sides and the top for alignment with the lasers in the treatment room. In addition, the fixture has six optical reflectors, and three thumbwheels for adjusting the fixture s height. The fixture also contains an internal tungsten ball positioned at isocenter that may be used for image verification of the fixture position. Isocenter fixture was placed on the treatment table at isocenter. Fixture level was adjusted with the help of thumbwheels and also table height was adjusted to align the lasers with the target lines on the isocenter fixture. Anterior and lateral orthogonal films were taken with calibration fixture. Position of the isocenter was evaluated with respected to LINAC isocenter. Fixture was moved towards isocenter when the offset position of the fixture is outside tolerance. Numbers of optical targets in each camera were counted as three, six, and three in camera 1, camera 2, and camera 3, respectively. With the help of software testing was started which took less than 1 minute and showed that the isocenter offset result was either pass or fail. If the test result is fail, fixture was repositioned. This test defines the coordinate reference frame of the optical system by aligning the Calypso System coordinate reference frame with the machine isocenter. The isocenter off set was calculated with the help of the following formula: 2 2 2 D= X + Y + Z C C C D is the Isocenter Offset which is the distance (offset) of the measured position of the isocenter, fixture isocenter from the machine isocenter position. It must be less than or equal to 0.1 cm for the test to pass. X C, YC, and Z C are the distances of the fixture s isocenter from the calibrated Calypso System isocenter in the X, Y, and Z axes, respectively. The calibrated Calypso System isocenter was defined at 0, 0, 0. Localization and tracking Calypso System is used for both localization (alignment of the treatment target with the machine isocenter before radiation treatment delivery) and tracking (continuous monitoring of this alignment during radiation treatment delivery) of target. In localize, only technique Calypso System is used only for localization. These patients have treatment targets that are too deep to permit tracking during treatment because there is inadequate clearance between the array and the linear accelerator gantry. However, they still benefit from accurate and objective initial alignment. These tests represent maximal height array locations relative to the machine isocenter for both Localize only (very large patients) and Localize and Track Patients (array up to 19 cm from the isocenter). For localize only test, 30 seconds of localization data was collected, and for localize and track test, 60 seconds of tracking data was collected and the results were tabulated. Tracking accuracy can be performed on a monthly basis by tracking motion of the QA fixture within the system limits. [11] Results Figure 6: Isocenter calibration fixture Quality assurance for checking the calibration or system accuracy In system accuracy QA, the detected positions of the fixture with respect to calibrated Calypso System isocenter in lateral, longitudinal, and in vertical directions

147 Table 1: Localization and tracking values Test name Setup accuracy Tracking accuracy Precision cm cm cm cm cm cm 3 transponder phantom at isocenter in localize only mode 0.05 0.2 0.01 0.15 RMS 3 transponder phantom at isocenter 0.04 0.2 0.05 0.2 0.07 0.15 2 transponder (disable low frequency) phantom at isocenter 0.06 0.2 0.07 0.2 0.03 0.15 2 transponder (disable medium frequency) phantom at isocenter 0.07 0.2 0.09 0.2 0.05 0.15 3 transponder phantom at (4, 4, 0) cm from isocenter 0.04 0.2 0.05 0.2 0.09 0.15 were 0.12 cm, 0.15 cm, and 0.00 cm, respectively. From the above data, the derived distance between QA fixture isocenter to Calypso System isocenter (isocenter offset) was extremely small with sub millimeter precision of 0.02 cm. Camera calibration with L frame and T frame fixture The test result was passed in all the cases (block of reflective objects, exposing optical targets without any uncover) by proper positioning of L Frame and exposing optical targets when four light sources were detected. During the camera calibration, T Frame fixture optical targets were detected by each camera throughout the optical field of view which shows all tests were passed. System calibration Against the tolerance of 0.1 cm, the offsets that were observed in x, y, z directions were 0.02 cm, 0.02 cm, 0.01 cm and the derived isocenter offset was 0.03 cm observed in this test. This proved that the Calypso System coordinate reference frame was calibrated to the machine isocenter using the isocenter fixture with high precision accuracy of <0.02 cm. Localization and tracking The setup accuracy, tracking accuracy and precision were measured in localization and tracking procedure and all the values that were obtained were within the limits with sub millimeter accuracy [Table 1]. Conclusions Patient positioning and localization by non radiographic techniques become key components in the radiation therapy process. Quality assurance of these systems is needed, as only proper operation and use ensures that the planned dose distribution is actually delivered. These tests were found to be useful in detecting performance of the Calypso System that needs recalibration. Use of these tests provide accurate, objective, and continuous localization of a treatment target for patient alignment and target position monitoring during radiation therapy to an accuracy of 0.01 cm. References 1. Parikh P, Hubenschmidt J, Dimmer S, Vertatschitsch E, Eidens R, Low D. 4D verification of real time accuracy of the Calypso System with lung cancer patient trajectory data. Int J Radiat Oncol Biol Phys 2005;63:S26 7. 2. Burch D, Willoughby T, Meeks S, Kupelian P, Litzenberg D, Sandler H, et al. Real time prostate translation, rotation, deformation evaluated with Calypso Beacon transponders. Int J Radiat Oncol Biol Phys 2005;63:S195. 3. Kupelian P, Willoughby T, Litzenberg D, Sandler H, Roach M, Levine L, et al. Clinical experience with the Calypso 4D localization system in prostate cancer patients: Implantation, tolerance, migration, localization and real time tracking. Int J Radiat Oncol Biol Phys 2005;63:S197. 4. Willoughby T, Kupelian P, Pouliot J, Shinohara K, Aubin M, Roach III M, et al. Target localization and real time tracking using the calypso 4D localization system in patients with localized prostate cancer. Int J Radiat Oncol Biol Phys 2006;65:528 34. 5. Kupelian P, Willoughby T, Mahadevan A, Djemil T, Weinstein G, Jani S, et al. Multi institutional clinical experience with the Calypso System in localization and continuous, real time monitoring of the prostate gland during external radiotherapy. Int J Radiat Oncol Biol Phys 2007;67:1088 98. 6. Scott T. Efficiency and clinical workflow of delivering IMRT to the prostate within 2 mm tolerances. Int J Radiat Oncol Biol Phys 2009;75 (3 Suppl):S580. Abstract no. 2889. 7. Beyer D, Liu D, Flores N. Efficiency of a non ionzing target localization system for radiation therapy. Int J Radiat Oncol Biol Phys 2006;66:S372. 8. Balter JM, Wright JN, Newell LJ, Friemel B, Dimmer S, Cheng Y, et al. Accuracy of a wireless localization system for radiotherapy. Int J Radiat Oncol Biol Phys 2005;61:933 7. 9. International standard IEC 60601 1 8. Medical electrical equipment Part 1 2: General requirements for basic safety and essential performance. Collateral standard: Electromagnetic compatibility Requirements and tests. Geneva, Switzerland: International Electrotechnical Commission Publications;2003. 10. Calypso System Version 2.0 User s Manual. Palo Alto, CA 94304, USA: Varian Medical Systems; 2012. 11. Santanama L, Noel C, Willoughby TR, Esthappan J, Mutic S, Klein EE, et al. Quality assurance for clinical implementation of an electromagnetic tracking system. Med Phys 2009;36:3477 86. How to cite this article: Muralidhar KR, Komanduri K, Rout BK, Ramesh K. Commissioning and quality assurance of Calypso fourdimensional target localization system in linear accelerator facility. J Med Phys 2013;38:143-7. Source of Support: Nil, Conflict of Interest: None declared.