TPG 261 TPG 262. Communication Protocol. Single-Channel Measurement and Control Unit for Compact Gauges

Similar documents
Communication protocol

SingleGauge Single-Channel Measurement and Control Unit for Compact Gauges

Contents. Product Identification 6 Validity 6 Intended Use 7 Scope of Delivery 7

MaxiGauge TPG 256 A. Operating manual incl. Declaration of conformity. Vacuum measurement and control unit for Compact Gauges

vacuum technologies FRG Control Unit INSTRUCTION MANUAL

Total Pressure Gauge Controller

Profibus-DP Interface Board for Total Pressure Gauge Controller TPG300

Profibus DP/V1 Interface for Bayard-Alpert Pirani Capacitance Diaphragm Gauge

Contents General Data Data Exchange Mode Block Model

Profibus DP/V1 Interface for Bayard-Alpert / Pirani Gauge and High Pressure / Pirani Gauge

VGC402 VGC403. Operating Manual Incl. EC Declaration of Conformity. Two-Channel Measurement and Control Unit

QualyTest QualyTest Select QualyTest Dry QualyTest Dry+ Helium leak detector

DeviceNet Interface for Bayard-Alpert Pirani Gauge

EtherCAT for Pirani Capacitance Diaphragm and Pirani Standard Gauges

Communication Protocol DeviceNet Interface for Bayard-Alpert / Pirani Gauge and High Pressure / Pirani Gauge BPG400-SD HPG400-SD tira03e1-b (0310) 1

EtherCAT for Bayard-Alpert Pirani Gauge

Addmaster Corporation

PRECISION DIGITAL CORPORATION

A GUIDE TO RS-232 COMMUNICATION WITH FX PLCS

Series 370 Stabil-Ion Gauge Controller with RS-485 Interface Option

S-Series Sensor ASCII Protocol v8.1.0

Operating instructions

Elotech Standard Protocol. for Single R8200 SC

DeviceNet for Capacitance Diaphragm Gauge

RS232C Serial Interface for Pirani Diaphragm and Pirani Standard Gauges. Caution. binary 8 data bits 1 stop bit no parity bit no handshake

RS232C / RS485C Serial Interface for Cold Cathode and Cold Cathode Pirani Gauges

EtherCAT. CDG025D2 CDG045D CDG045D2 CDG045Dhs CDG100D CDG100D2 CDG100Dhs CDG160D CDG160Dhs CDG200D CDG200Dhs. Communication Protocol

DeviceNet for Pirani Capacitance Diaphragm and Pirani Standard Gauges

EXPERIMENT 8: Introduction to Universal Serial Asynchronous Receive Transmit (USART)

1.1. INTRODUCTION 1.2. NUMBER SYSTEMS

JetWeb JX6-INT1 Function Description

RS232C / RS485C Serial Interface for Pirani Capacitance Diaphragm and Pirani Standard Gauges

Profibus DP/V1 Interface for Capacitance Diaphragm Gauge

Modbus Remote Communication Protocol for REM 54_. Technical Description

High Pressure / Pirani Gauge

Single-Axis Counter BDD 611

OPERATING MANUAL AND TECHNICAL REFERENCE

Bayard-Alpert Pirani Gauge

Manual Electronic Identification Systems BIS Processor BIS C-6_0 RS 232 C-6_ _836905_E_0704.p65 Deutsch bitte wenden!

Connect Digital GagesandSerial Devices intoaserialport

THT-500-A/R (C5) 1. System Configuration COMMUNICATION INSTRUCTION MANUAL. No. THT5CE

MK D Imager Barcode Scanner Configuration Guide

ECR-Interface ZVT-Protocol. Transport-Protocol Application-Protocol

ACS 2000 Single-channel Controller

RS485 INSTRUCTION MANUAL. DM4500 Series. Page 1 of 6

Data Representation and Binary Arithmetic. Lecture 2

ADDMASTER. Addmaster Corporation. IJ-3080 Journal/Validation Printer. Specification. IJ-3080 Specification

Operating instructions

FD-011WU. 2D Barcode Reader User Guide V1.6CC

ModBUS Protocol for SMART 3

PCG550, PCG552, PCG554

NC-1200 BARCODE SCANNER. Configuration Guide - 1 -

PSG550, PSG552, PSG554

Profibus DP/V1 Interface for Pirani Capacitance Diaphragm and Pirani Standard Gauges

ECO History: Date Revision ECO# Description of Change

BD-6500BT Bluetooth 2D Barcode Scanner Configuration Guide

MODEL 715AN QUADRATURE DEGREES COUNTER

BARCODE SCANNER. Configuration Guide - 1 -

RS-422 Code-Operated Switches

Precision Digital Modbus Register Tables Serial Communication

C30xx Digital communication

Manual DTM Digital Pressure Transmitter

BPG400 ATM to Ultra-High Vacuum Gauge

Operating instructions

Mounting Dimensions / Viewing 2 Mounting Options 3. Wiring Configuration 4. Quick Set up Procedure 5. Changing Intensity 6.

PSG55x ATM to Medium Vacuum Gauge

North American Service Manual

Diagnostic Port Serial Interface for Capacitance Diaphragm Gauges

JCL-33A (C5) Warning. 1. System Configuration COMMUNICATION INSTRUCTION MANUAL

OPERATION MANUAL Card Reader Communication

Serial Communication Interface PCP-3016

Chapter 3. Information Representation

EXPERIMENT 7: Introduction to Universal Serial Asynchronous Receive Transmit (USART)

Configuration Manual PULSAR C CCD SCANNER. Table of Contents

BPG402-Sx ATM to Ultra-High Vacuum Gauge

Pins 2, 3 - Baud rate select - Selects the baud rate for the serial transmission.

2D BARCODE SCANNER CA-SC-20200B

-DIN & DIN RaPID TEMPERATURE CONTROLLERS. Product Manual (June 1999)

PM130 Powermeters Reference Guide ASCII Communications Protocol

I/O232-A User s Guide

VeriColor. Solo Identification System. Command Users Manual Document ver Rev. 7/20/05

Line Mode: Digitizer outputs coordinate data points when a button is down; plus one when the button is released.

Operate the Hydrolab SDI-12 / Modbus / RS232 TTY Communications Module (HL Series Sonde)

INTELLIS. Modbus Direct Network Monitor

General Specifications

RITZ / Turbine controls DISPLAY

Connecting UniOP to Datalogic Barcode Readers

OPERATING PANELS. PROGRAMMING Manual EBELT OPERATING PANELS

INSTRUCTION MANUAL SERIES

Sartorius Comparator. Interface Description for the CC Model Series

HMC1022 Digital Compass

AST1-01X055(1/15) For form ZX smart sensors Interface unit type ZX-SF11 Compoway/F Message communication command specifications

PCG55x ATM to Medium Vacuum Gauge

Nokeval. No Manual. Model for serial inputs

List of Contents 1. INTRODUCTION DEFINITIONS AND ABBREVIATIONS REFERENCES TECHNICAL DATA GENERAL MODBUS RTU...

SERIAL INTERFACE MANUAL. Precision Integrating Sound Level Meter. with 1/3 octave band real-time analyzer NA-27

Table of Contents Sleep Settings How to Configure the Scanner. 7 Chapter 2 System Setup

PD1100 STAND-ALONE PROGRAMMING & USER S GUIDE. use the freedom

Xi2000-BT Series Configuration Guide

User s Manual. Xi3000 Scanner. Table of Contents

Transcription:

Communication Protocol Single-Channel Measurement and Control Unit for Compact Gauges TPG 261 VACUUM Dual-Channel Measurement and Control Unit for Compact Gauges TPG 262 VACUUM BG 805 198 BE (0109) 1

Product Identification SingleGauge TPG 261 DualGauge TPG 262 BG 805 195 BE BG 805 196 BE Validity This document applies to products with part number PTG28030 (SingleGauge ) PTG28280 (DualGauge ). The part number (No.) can be taken from the product nameplate. This document is based on firmware number 302-510--. If your unit does not work as described in this document, please check that it is equipped with the above firmware version ( 13). We reserve the right to make technical changes without prior notice. Intended Use The RS232C interface is used for operating the TPG 261 / TPG 262 via a computer or a terminal. Trademarks DualGauge FullRange Inficon AG Inficon AG Contents Product Identification 2 Validity 2 Intended Use 2 Trademarks 2 1 RS232C Interface 3 1.1 Installation 3 1.2 Data Transmission 3 1.2.1 Definitions 3 1.2.2 Flow Control 3 1.2.3 Communication Protocol 4 2 Mnemonics 5 2.1 Measurement Mode 6 2.2 Parameter Mode 9 2.2.1 Switching Function Parameters 9 2.2.2 Gauge Parameters 10 2.2.3 Gauge Control 12 2.2.4 General Parameters 12 2.2.5 Test Parameters 13 2.3 Example 17 XY) is used, for cross- For cross-references within this document, the symbol ( references to other documents, the symbol ( [Z]). 2 BG 805 198 BE (0109) TPG26x.cp

1 RS232C Interface The serial interface is used for communication between the TPG 261 / TPG 262 and a computer. A terminal can be connected for test purposes. When the TPG 261 / TPG 262 is put into operation, it starts transmitting measured values in intervals of 1 s. As soon as the first character is transferred to the TPG 261 / TPG 262, the automatic transmission of measured values stops. After the necessary inquiries or parameter modifications have been made, the transmission of measured values can be started again with the COM command ( 7). 1.1 Installation SingleGauge TPG 261 DualGauge TPG 262 BG 805 195 BE BG 805 196 BE 1.2 Data Transmission The data transmission is bi-directional, i.e. data and control commands can be transmitted in either direction. Configuration of the interface SingleGauge TPG 261 DualGauge TPG 262 BG 805 195 BE BG 805 196 BE Data format 1 start bit, 8 data bits, no parity bit, 1 stop bit, no hardware handshake 1.2.1 Definitions The following abbreviations and symbols are used: Symbol Meaning HOST Computer or terminal [...] Optional elements ASCII American Standard Code for Information Interchange Dec Hex <ETX> END OF TEXT (CTRL C) Reset the interface 3 03 <CR> <LF> <ACK> <NAK> CARRIAGE RETURN Go to beginning of the line LINE FEED Advance by one line ENQUIRY Request for data transmission ACKNOWLEDGE Positive report signal NEGATIVE ACKNOWLEDGE Negative report signal 13 0D 10 0A 5 05 6 06 21 15 "Transmit": Data transfer from HOST to TPG 261 / TPG 262 "Receive": Data transfer from TPG 261 / TPG 262 to HOST 1.2.2 Flow Control After each ASCII string, the HOST must wait for a report signal ( or <NAK> <CR><LF>). The input buffer of the HOST must have a capacity of at least 32 bytes. BG 805 198 BE (0109) TPG26x.cp 3

1.2.3 Communication Protocol Transmission format Messages are transmitted to the TPG 261 / TPG 262 as ASCII strings in the form of mnemonic operating codes and parameters. All mnemonics comprise three ASCII characters. Spaces are ignored. <ETX> (CTRL C) clears the input buffer in the TPG 261 / TPG 262. Transmission protocol HOST TPG 261 / TPG 262 Explanation Mnemonics [and parameters] > <CR>[<LF>] > Receives message with "end of message" < Positive acknowledgment of a received message Reception format When requested with a mnemonic instruction, the TPG 261 / TPG 262 transmits the measurement data or parameters as ASCII strings to the HOST. must be transmitted to request the transmission of an ASCII string. Additional strings, according to the last selected mnemonic, are read out by repetitive transmission of. If is received without a valid request, the ERROR word is transmitted. Reception protocol HOST TPG 261 / TPG 262 Explanation Mnemonics [and parameters] > <CR>[<LF>] > Receives message with "end of message" < Positive acknowledgment of a received message > Requests to transmit data < Measurement values or parameters < <CR><LF> Transmits data with "end of message" : : > Requests to transmit data < Measured values or parameters < <CR><LF> Transmits data with "end of message" Error processing The strings received are verified in the TPG 261 / TPG 262. If an error is detected, a negative acknowledgment <NAK> is output. Error recognition protocol HOST TPG 261 / TPG 262 Explanation Mnemonics [and parameters] > <CR>[<LF>] > Receives message with "end of message" ***** Transmission or programming error ***** < <NAK><CR><LF> Negative acknowledgment of a received message Mnemonics [and parameters] > <CR>[<LF>] > Receives message with "end of message" < Positive acknowledgment of a received message 4 BG 805 198 BE (0109) TPG26x.cp

2 Mnemonics ADC A/D converter test 15 BAU Baud rate (transmission rate) 13 COM Continuous mode 7 CAL Calibration factor 10 DCD Display control digits (display resolution) 13 DGS Degas 12 DIC Display control (display changeover) 13 DIS Display test 15 EEP EEPROM test 15 EPR EPROM test 14 ERR Error status 8 FIL Filter time constant (measurement value filter) 10 FSR Full scale range (measurement range of linear gauges) 10 IOT I/O test 16 LOC Keylock 14 OFC Offset correction (linear gauges) 11 OFD Offset display (linear gauges) 11 PNR Program number (firmware version) 13 PR1 Pressure measurement (measurement data) gauge 1 6 PR2 Pressure measurement (measurement data) gauge 2 6 PRX Pressure measurement (measurement data) gauge 1 and 2 6 PUC Penning underrange control (underrange control) 11 RAM RAM test 14 RES Reset 8 RST RS232 test 16 SAV Save parameters to EEPROM 13 SC1 Sensor control 1 (gauge control 1) 12 SC2 Sensor control 2 (gauge control2) 12 SCT Sensor channel change (measurement channel change) 8 SEN Sensors on/off 7 SP1 Setpoint 1 (switching function 1) 9 SP2 Setpoint 2 (switching function 2) 9 SP3 Setpoint 3 (switching function 3) 9 SP4 Setpoint 4 (switching function 4) 9 SPS Setpoint status (switching function status) 9 TID Transmitter identification (gauge identification) 7 TKB Keyboard test (operator key test) 16 TLC Torr lock 14 UNI Pressure unit 12 WDT Watchdog control 14 BG 805 198 BE (0109) TPG26x.cp 5

2.1 Measurement Mode Measurement data gauge 1 or 2 PRx <CR>[<LF>] Pressure measurement Measurement value x = 1 > Gauge 1 2 > Gauge 2 x,sx.xxxxesxx <CR><LF> (always exponential format) Measurement value *) [in current pressure unit] Status x = 0 > Measurement data okay 1 > Underrange 2 > Overrange 3 > Sensor error 4 > Sensor off (IKR, PKR, IMR, PBR) 5 > No sensor (output: 5,2.0000E-2 [mbar]) 6 > Identification error *) For logarithmic gauges, the 3 rd and 4 th decimal are always 0. Measurement data gauges 1 and 2 PRX <CR>[<LF>] Pressure measurement x,sx.xxxxesxx,y,sy.yyyyesyy <CR><LF> (always exponential format) Measurement value gauge 1 *) [in current pressure unit] Status gauge 2 Measurement value gauge 1 *) [in current pressure unit] Status gauge 1 x = 0 > Measurement data okay 1 > Underrange 2 > Overrange 3 > Sensor error 4 > Sensor off (IKR, PKR, IMR, PBR) 5 > No sensor (output: 5,2.0000E-2 [mbar]) 6 > Identification error *) For logarithmic gauges, the 3 rd and 4 th decimal are always 0. 6 BG 805 198 BE (0109) TPG26x.cp

Continuous output of measurement values (RS232) COM [,x] <CR>[<LF>] Continuous mode Mode x = 0 > 100 ms 1 > 1 s (default) 2 > 1 min. <ACK> is immediately followed by the continuous output of the measurement value in the desired interval. x,sx.xxxxesxx,y,sy.yyyyesyy <CR><LF> (always exponential format) Measurement value gauge 1 *) [in current pressure unit] Status gauge 2 Measurement value gauge 1 *) [in current pressure unit] Status gauge 1 x = 0 > Measurement data okay 1 > Underrange 2 > Overrange 3 > Sensor error 4 > Sensor off (IKR, PKR, IMR, PBR) 5 > No sensor (output: 5,2.0000E-2 [mbar]) 6 > Identification error *) For logarithmic gauges, the 3 rd and 4 th decimal are always 0. Turning a gauge on/off SEN [,x,x] <CR>[<LF>] Sensors on/off Gauge 2 x = 0 > No status change 1 > Turn gauge off 2 > Turn gauge on Gauge 1 x,x <CR><LF> Status gauge 2 x = 0 > Gauge cannot be turned on/off 1 > Gauge turned off 2 > Gauge turned on Status gauge 1 Gauge identification TID <CR>[<LF>] Gauge identification x,x <CR><LF> Identification gauge 2 x = TPR (Pirani Gauge) IKR9 (Cold Cathode Gauge 10-9 ) IKR11 (Cold Cathode Gauge 10-11 ) PKR (FullRange CC Gauge) PBR (FullRange BA Gauge) IMR (Pirani / High Pressure Gauge) CMR (Linear gauge) nosen (no SEnsor) noid (no identifier) Identification gauge 1 BG 805 198 BE (0109) TPG26x.cp 7

Measurement channel change SCT [,x] <CR>[<LF>] Sensor channel change Display channel x = 0 > Gauge 1 1 > Gauge 2 x <CR><LF> Display channel Error status ERR <CR>[<LF>] Error status xxxx <CR><LF> xxxx = 0000 > No error 1000 > Controller error (See display on front panel) 0100 > NO HWR No hardware 0010 > PAR Inadmissible parameter 0001 > SYN Syntax error The ERROR word is cancelled when read out. If the error persists, it is set again. Reset RES [,x] <CR>[<LF>] Reset x = 1 > Cancels currently active error and returns to measurement mode [x]x,[x]x,... <CR><LF> List of all present error messages xx = 0 > No error 1 > Watchdog has responded 2 > Task fail error 3 > EPROM error 4 > RAM error 5 > EEPROM error 6 > DISPLAY error 7 > A/D converter error 9 > Gauge 1 error (e.g. filament rupture, no supply) 10 > Gauge 1 identification error 11 > Gauge 2 error (e.g. filament rupture, no supply) 12 > Gauge 2 identification error 8 BG 805 198 BE (0109) TPG26x.cp

2.2 Parameter Mode 2.2.1 Switching Function Parameters Threshold value setting, allocation SPx [y,x.xxxxesxx,x.xxxxesxx] <CR>[<LF>] Setpoint Upper threshold *) [in current pressure unit] (default = depending on gauge) Lower threshold *) [in current pressure unit] (default = depending on gauge) Switching function assignment y = 0 > Meas. channel 1 1 > Meas. channel 2 1 > Switching function 1 2 > Switching function 2 3 > Switching function 3 4 > Switching function 4 *) Values can be entered in any format. They are internally converted into the floating point format. y,x.xxxxesxx,x.xxxxesxx <CR><LF> Upper threshold [in current pressure unit] Lower threshold [in current pressure unit] Switching function assignment Switching function status SPS <CR>[<LF>] Setpoint status x,x,x,x <CR><LF> Status switching function 4 Status switching function 3 Status switching function 2 Status switching function 1 x = 0 > off 1 > on BG 805 198 BE (0109) TPG26x.cp 9

2.2.2 Gauge Parameters Measurement value filter FIL [,x,x] <CR>[<LF>] Filter time constant Gauge 2 x = 0 > fast 1 > medium (default) 2 > slow Gauge 1 x,x <CR><LF> Filter time constant gauge 2 Filter time constant gauge 1 Calibration factor CAL [,x.xxx,x.xxx] <CR>[<LF>] Calibration factor Gauge 2 log. 0.100... 9.990 (default = 1.000) lin. 0.500... 2.000 (default = 1.000) Gauge 1 x.xxx,x.xxx <CR><LF> Calibration factor gauge 2 Calibration factor gauge 1 Measurement range (F.S.) of linear gauges The full scale value of the measurement range (Full Scale) of linear gauges has to be defined by the user; the full scale value of logarithmic gauges is automatically recognized. FSR [,x,x] <CR>[<LF>] Full scale range Gauge 2 x = 0 > 0.01 mbar 1 > 0.1 mbar 2 > 1 mbar 3 > 10 mbar 4 > 100 mbar 5 > 1000 mbar (default) 6 > 2 bar 7 > 5 bar 8 > 10 bar 9 > 50 bar Gauge 1 x,x <CR><LF> Measurement range gauge 2 Measurement range gauge 1 10 BG 805 198 BE (0109) TPG26x.cp

Offset correction (linear gauges) OFC [,x,x] <CR>[<LF>] Offset correction Gauge 2 x = 0 > off (default) 1 > on 2 > auto (offset measurement) Gauge 1 x,x <CR><LF> Gauge 2 Gauge 1 Offset display (linear gauges) OFD [,sx.xxxxesxx,sx.xxxxesxx] <CR>[<LF>] Offset display Gauge 2 Offset *) [in current pressure unit] (default = 0.0000) Gauge 1 sx.xxxxesxx,sx.xxxxesxx <CR><LF> Gauge 2 Gauge 1 *) Values can be entered in any format. They are internally converted into the floating point format. Underrange control PUC [,x,x] <CR>[<LF>] Penning underrange control Gauge 2 x = 0 > off (default) 1 > on Gauge 1 x,x <CR><LF> Gauge 2 Gauge 1 BG 805 198 BE (0109) TPG26x.cp 11

Degas DGS [,x,x] <CR>[<LF>] Degas Gauge 2 x = 0 > Degas off (default) 1 > Degas on (3 min.) Gauge 1 x,x <CR><LF> Degas status gauge 2 Degas status gauge 1 2.2.3 Gauge Control Gauge control SCx [,x,y,x.xxesxx,y.yyesyy] <CR>[<LF>] Sensor control OFF threshold ON threshold Controlling source for gauge deactivation x = 0 > no control 1 > automatic deactivation 2 > manual deactivation (default) 3 > external deactivation 4 > self control Controlling source for gauge activation x = 0 > no control 1 > automatic activation 2 > manual activation (default) 3 > external activation 4 > hot start Controlled gauge x = 1 > Gauge 1 2 > Gauge 2 x,y,x.xxesxx,y.yyesyy <CR><LF> OFF threshold ON threshold Controlling source for deactivating the gauge Controlling source for activating the gauge 2.2.4 General Parameters Pressure unit UNI [,x] <CR>[<LF>] Pressure unit Pressure unit x = 0 > mbar/bar (default) 1 > Torr 2 > Pascal x <CR><LF> Pressure unit 12 BG 805 198 BE (0109) TPG26x.cp

Transmission rate BAU [,x] <CR>[<LF>] Baud rate Transmission rate x = 0 > 9600 baud (default) 1 > 19200 baud 2 > 38400 baud x <CR><LF> Transmission rate As soon as the new baud rate has been entered, the report signal is transmitted at the new transmission rate. Display resolution DCD [,x] <CR>[<LF>] Display control digits Resolution x = x <CR><LF> Resolution 2 > Display x.x (2 digits) (default) 3 > Display x.xx (3 digits) Save parameters to EEPROM SAV [,x] <CR>[<LF>] Save parameters to EEPROM x = 0 > Save default parameters 1 > Save user parameters Display changeover DIC [,x] <CR>[<LF>] Display control Measurement display behavior when a Pirani gauge is combined with a linear gauge with 1000 mbar F.S. x = 0 >manual (default) 1 >automatic x <CR><LF> Measurement display behavior 2.2.5 Test Parameters (For service specialists) Firmware version PNR <CR>[<LF>] Program number 302-510-x <CR><LF> -x = Modification index (-- = original version) Firmware number BG 805 198 BE (0109) TPG26x.cp 13

Watchdog control WDT [,x] <CR>[<LF>] Watchdog control x = 0 > Manual error acknowledgement 1 > Automatic error acknowledgement *) (default) x <CR><LF> Watchdog control *) If the watchdog has responded, the error is automatically acknowledged and cancelled after 2 s. Torr lock TLC [,x] <CR>[<LF>] Torr lock x = 0 > off (default) 1 > on x <CR><LF> Torr lock status Keylock LOC [,x] <CR>[<LF>] Keylock x = 0 > off (default) 1 > on x <CR><LF> Keylock status RAM test RAM <CR>[<LF>] RAM test Starts the test (duration <1 s) xxxx <CR><LF> ERROR word EPROM test EPR <CR>[<LF>] EPROM test Starts the test (duration 5 s) xxxx,yyyy <CR><LF> Check sum (hex) ERROR word 14 BG 805 198 BE (0109) TPG26x.cp

EEPROM test EEP <CR>[<LF>] EEPROM test Starts the test *) (duration <1 s) *) Do not keep repeating the test (EEPROM life). xxxx <CR><LF> ERROR word Display test DIS [,x] <CR>[<LF>] Display test x = 0 > Stops the test display according to current operating mode (default) 1 > Starts the test all LEDs on x <CR><LF> Display test status ADC test ADC <CR>[<LF>] ADC test [x]x.xxxx,[x]x.xxxx,x.xxxx,x.xxxx <CR><LF> ADC channel 4 Gauge 2 identification [0.0000... 5.0000 V] ADC channel 3 Gauge 1 identification [0.0000... 5.0000 V] ADC channel 2 Measurement signal gauge 2 [0.0000... 11.0000 V] ADC channel 1 Measurement signal gauge 1 [0.0000... 11.0000 V] BG 805 198 BE (0109) TPG26x.cp 15

I/O test Caution Caution: The relays switch irrespective of the pressure. Starting a test program may cause unwanted effects in connected control systems. Disconnect all sensor cables and control system lines to ensure that no control commands or messages are triggered by mistake. IOT [,x,yy] <CR>[<LF>] I/O test yy = Relay status (in hex format) 00 > All relays deactivated 01 > Switching function relay 1 activated 02 > Switching function relay 2 activated 04 > Switching function relay 3 activated 08 > Switching function relay 4 activated 10 > Gauge relay CH1 activated 20 > Gauge relay CH2 activated 40 > Error relay activated x = x,yy <CR><LF> Relay status I/O test status 7F > All relays activated 0 > Test stopped 1 > Test runs Operator key test TKB <CR>[<LF>] Keyboard test xxxx <CR><LF> Key 4 x = 0 > Not pushed 1 > Pushed Key 3 Key 2 Key 1 RS232 test RST <CR>[<LF>] RS232 test Starts the test (repeats each character, test is interrupted with <CTRL> C) 16 BG 805 198 BE (0109) TPG26x.cp

2.3 Example "Transmit (T)" and "Receive (R)" are related to Host. S: TID <CR> [<LF>] E: <ACK> <CR> <LF> S: E: TPR,CMR <CR> <LF> S: SEN <CR> [<LF>] E: <ACK> <CR> <LF> S: E: 0,0 <CR> <LF> S: SP1 <CR> [<LF>] E: <ACK> <CR> <LF> S: E: 0,1.0000E-09,9.0000E-07 <CR> <LF> S: SP1,1,6.80E-3,9.80E-3 <CR> [<LF>] E: <ACK> <CR> <LF> S: FOL,1,2 <CR> [<LF>] E: <NAK> <CR> <LF> S: E: 0001 <CR> <LF> S: FIL,1,2 <CR> [<LF>] E: <ACK> <CR> <LF> S: E: 1,2 <CR> <LF> Request for gauge identification Positive acknowledgement Request for data transmission Gauge identifications Request for gauge statuses Positive acknowledgement Request for data transmission Gauge statuses Request for parameters of switching function 1 (setpoint 1) Positive acknowledgement Request for data transmission Thresholds Modification of parameters of switching function 1 (setpoint 1) Positive acknowledgement Modification of filter time constant (syntax error) Negative acknowledgement Request for data transmission ERROR word Modification of filter time constant Positive acknowledgement Request for data transmission Filter time constants BG 805 198 BE (0109) TPG26x.cp 17

Notes 18 BG 805 198 BE (0109) TPG26x.cp

Notes BG 805 198 BE (0109) TPG26x.cp 19

Original: German BG 805 198 BD (0109) Emmeliusstrasse 33 D 35614 Asslar Deutschland Tel +49 (0) 6441 802-0 Fax +49 (0) 6441 802-202 info@pfeiffer-vacuum.de bg805198be www.pfeiffer-vacuum.de