Standard Operating Procedure for the Kaye Validator AVS System

Similar documents
Standard Operating Procedure For The Kaye Validator 2000 Thermal Mapping System

ON/OFF HUMIDIFIER CONTROL (ROOM 0R DUCT SENSOR) WITH OUTDOOR TEMPERATURE RESET OVERVIEW WALL CONTROLLER INSTALLATION (TCY-BH-T-U)

Kaye Validator AVS Version Z3007, Rev D June Advanced Validation System Console SW Enhancements

GE Sensing. Kaye RF ValProbe. Wireless Thermal Validation System. User s Guide

USER MANUAL MULTI COLOR TOUCH SCREEN PAPERLESS RECORDER MODEL : ARC2020

USER MANUAL MULTI COLOR TOUCH SCREEN PAPERLESS RECORDER TPLR-96 Series

Model HM-535 Power Supply Installation and Service Instructions

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER

AEROTRAK PORTABLE AIRBORNE PARTICLE COUNTER MODEL 9310/9350/9510/9550/9500 QUICK START GUIDE

Product Documentation

USB-5100 Series Multi-Channel Data Loggers

OPERATING INSTRUCTION

DM650TM SENSOR / WALL MOUNTED TEMPERATURE DISPLAY

S-14 S-14. Compact Digital Multimeter. Compact Digital Multimeter

BTU-4D/BTU-4D-I Digital Temperature Unit & Sensors. Installation and Operation Manual

HP R/T2200 UPS. Overview. Precautions. Installation Instructions. The HP UPS R/T2200 features power protection for loads up to 2200 VA/1600 W.

Oven Loggermaster BEVS User Manual

Platinum RTD Thermometer MODEL NO

L Series Humidity/-temperature sensors for use in heating, ventilation and air conditioning Technical data Humidity

AM3 Humidity. Temperature. Dew Point Indicator INSTRUCTION MANUAL

FIELD CALIBRATION PROCEDURE FOR ALTEK MODEL 222A

EMSwitch RF Switch Plug-In Card

Technical Manual Nova: Cabinet Security Management System (CSMS)

Software... pg3 File Maintenance Training

Autoranging True RMS Multimeter User Manual

E1135C PDU and Pod Upgrade Procedure

INSTALLATION INSTRUCTIONS

Computer Temperature Interface model CTI-2. Description. Installation. Control Module

JUMO ecotrans ph 03 Microprocessor transmitter/ switching device for ph/redox voltage and temperature

Accsense Wireless System. Temperature Monitor

SX90 Process Controller

Product Manual. Surge Protected Metal 7- Port USB 2.0 Hub DIN RAIL Mount Kit NEC Chip

Platinum RTD Thermometer MODEL NO

XL112, XL114 ポータブルデータロガ

Refrigerated incubators with forced convection

Product specification

HiTemp140-FP-TSK High Temperature Data Logger with Flexible RTD Probe and Thermal Shield

Line Interactive 1000VA/1400VA/2000VA Uninterruptible Power System

BTU 4 & Sensors. Installation and Operation Manual

Automotive Electronics Council Component Technical Committee

DSTHM-2 COMBINED T AND RH DUCT TRANSMITTER. Mounting and operating instructions

HygroClip M22 Series. 2-wire loop powered humidity temperature transmitters INSTRUCTION MANUAL

Enhanced insulation thickness. Environment friendly polyurethane foam, water based foaming.

Contents 1 Warnings, Cautions, and Notes Description Features... 1

Front-door station series for surface-mounting Profi-Economic PES Profi-Design PDS

DXTH DUCT SENSOR / SWITCH FOR TEMPERATURE AND HUMIDITY. Mounting and operating instructions

PI 500 Hand-held instrument for industry

HP UPS R/T3000 G2. Overview. Precautions. Kit contents. Installation Instructions

Model 7700 Multiplexer Module

Section 11. Electrical System. 42--CF200--MA--09 (formerly 42-65) Issued 5/96 P/N B. E 1996 Nordson Corporation All rights reserved

User Manual RecoMedia Presentation made easy

Main contacts Auxiliary contact Number of input channels Rated voltage Model Category 24 VAC/VDC 3PST-NO SPST-NC 1 channel or 2 channels possible

4-Port USB 2.0 Mini Metal High-Speed Hub. Product Manual. Coolgear, Inc. Version 1.1 September 2017 Model Number: USBG-4PUSB2-MH.

CF3000 Dealer Diagnostic Tool Instruction Manual

RF Thermal Validation System. Kaye RF ValProbe PRESSURE HUMIDITY. High accuracy measurement of temperature and humidity

Marine User Guide. USBT202 Series High Power USB Charging Port

SER8300 Series Line Voltage Room Controller with Optional PIR Sensor Installation Guide

RXTP ROOM TEMPERATURE

12-Bit 0-5 Volt Input Adapter (Part # S-VIA-CM14)

Installation Instructions for: Channel Thermocouple Amplifier

CDD4 Series Room CO2 Transmitter Installation Instructions

Adafruit USB Power Gauge Mini-Kit

Druck DPI 620 advanced modular calibrator

MFD Thermal Process Controller Manual

KM-4800w. Installation Guide

H290 HUMIDITY TRANSMITTER INSTRUCTION MANUAL

360 Test Labs 4 Executive Plaza Yonkers, New York Toll-Free (855) LABS Page 1 of 8

MiG2 CONTROLLERS. 2 & 4 Stage General Purpose Controllers, with Air-conditioning Facilities

IntesisBox SMS - Mitsubishi Electric air conditioners

MultiTherm Shaker. Operations Manual Item: H5000-H and H5000-HC

HP Data Center Environmental Edge Hardware Installation Guide

EnCell Battery Cell Monitor

PRO REMOTE 25 LV CAPACITANCE PROBE

Whisper-Ride

UniStream CPU-for-Panel

Installation Manual. Mounting Instructions Mechanical Mounting. Luminato. Teleste Corporation

Z CAM K1 Pro. User Manual (Draft) v1.0

icp installation guide

Product and Function Description

V-3000 Series Miniature Basic Switches

Quick Start Guide. Mounting the Essential Plus and Elite

Gospeed.Racer Hub Assembly for the Asus EEEpc

Pure sine wave inverter. Product Manual

1523/1524 Reference Thermometers

TITAN S8. Transportable Data Acquisition Device. Product Brochure EN

ZTE MOBILE HOTSPOT QUICK START GUIDE

HOBO Temp/RH 3.5% Data Logger (UX ) Manual

Emerson Network Power provides customers with technical support. Users may contact the nearest Emerson local sales office or service center.

! Ranges from -45º to 650ºC. ! Rapid heating, cooling and settling. ! Switch test, ramp, step and preset functions

HWS Series Wall Mount with Relay Setpoints, LCD, and Humidistat Control

NIST RE CERTIFICATION GUIDE DRAFT!!!

Manual# Installation Manual. 200E Series. DCU 210E/208E Engine Panel RP 210E/220E Remote Panel

HOBO Temp/RH 2.5% Data Logger (UX ) Manual

AquaController Apex Probe Module 2

USP-070-B08 USP-104-B10, USP-104-M10 USP-156-B10

Harbortronics Time-Lapse Package Troubleshooting Guide This document is intended to help diagnose any problems with your Time-Lapse Package.

PT230 PONTOON THRUSTER INSTALLATION MANUAL (PATENT PENDING) VOLUME 1.2, JUNE 2015

Electrical Demand Specification (Reference SOP: )

Model 2460-KIT. Screw Terminal Connector Kit. Description / September 2014 *P * 1

PeopleNet TABLET Installation Guide. Version 1.1. A guide for new TABLET & new OBC installations.

Transcription:

Document #: Revision: A Kaye Validator AVS Standard Operating Version 1.0 This is a version controlled document that is released with Amphenol internal approvals. The Approvals section is provided for optional customer use to approve and accept this document. It may be amended, deleted or marked as N/A at the customer s discretion. Approvals (Name and Title) Technical Reviewer: Date: Validation: Date: Technical Services: Date: Quality Control: Date: Management: Date:

Document #: Revision: A PAGE: 2 of 16 TABLE OF CONTENTS 1. PURPOSE...3 2. SCOPE...3 3. SAFETY...3 4. EQUIPMENT...3 5. REFERENCE...4 6. SYSTEM DESCRIPTION...4 6.1. Kaye Validator AVS...4 6.2. Kaye Validator Software Version 1.28...4 6.3. Sensor Input Modules (SIMs)...4 6.4. Temperature Reference...5 6.5. IRTD...5 7. SYSTEM ACCESS...5 8. PROCEDURES...6 8.1. Wiring SIMs...6 8.2. Starting the Kaye Validator AVS Software...7 8.3. Creating/Modifying a Setup...7 8.4. Connecting the System...9 8.5. Checking Hardware and Communications Connections...9 8.6. Operation... 10 8.6.1. Sensor Calibration... 10 8.6.2. Qualification Study... 11 8.6.3. Calibration Verification... 12 Generating Reports... 13 8.6.4. Generating Setup Reports... 13 8.6.5. Generating Calibration Reports... 13 8.6.6. Generating Qualification Reports... 13 8.6.7. Generating Verification Reports... 15 8.7. Using the Kaye Validator Audit Trail... 15 8.7.1. Accessing the Audit Trail... 15 8.7.2. Filtering and Printing the Audit Trail... 15

Document #: Revision: A PAGE: 3 of 16 1. PURPOSE 2. SCOPE To provide an operational procedure that describes the setup and use of the Kaye Validator AVS system. This standard operating procedure provides safety information, a list of required equipment, a description of the system, and instructions for connecting inputs to Sensor Input Modules (SIMs), starting the Kaye Validator AVS software, creating/modifying a setup file, connecting the system, calibrating sensors, running a qualification study, verifying sensor calibration, generating reports, and accessing the Kaye Validator AVS audit trail. This procedure assumes that the Kaye Validator AVS software version 1.0 has been installed and the System Administrator has created user accounts. See Kaye Validator AVS User s Guide for more information. 3. SAFETY The following safety precautions shall be observed when working around steam, high temperature ovens, and autoclaves during qualification studies: 4. EQUIPMENT Ensure all electrical plugs are plugged in completely. Ensure all extension cords are correctly routed from wall receptacles. Do not route cords across active hallways or walkways without taping to floor or routing overhead. Know location of emergency shut-off switches, if applicable. Remove any heat-sensitive instruments from the system that could be damaged during steaming. Ensure all gaskets are in place before steaming. Ensure all doors, hatches, and so forth are closed before starting qualification studies. Steaming temperatures may exceed 121 C. Do not touch any exposed surfaces while steaming. Ensure equipment being tested is cool before attempting to remove thermocouples. A complete Kaye validation system includes: The Kaye Validator AVS instrument with Version 0.0.1.42 firmware. The Kaye Validator AVS software version 1.28 for Microsoft Windows Embedded 8.1 Professional (64-bit). A printer that supports both portrait and landscape page orientation. Up to four Kaye Sensor Input Modules (SIMs). A Kaye high temperature reference (HTR 400), low temperature reference (LTR-90, LTR 25/140 or LTR 40/140), or a cold temperature reference (CTR 40 or CTR 80) for sensor calibration and verification. A Kaye Intelligent RTD Probe (IRTD). Premium grade thermocouples. Additional temperature and non-temperature sensors for qualification studies (optional).

Document #: Revision: A PAGE: 4 of 16 5. REFERENCE Kaye Validator AVS User s Guide, Document M5100 6. SYSTEM DESCRIPTION This section describes the Kaye Validator AVS system and associated equipment. 6.1. Kaye Validator AVS Self-contained thermal validation device with high accuracy sensor measurement hardware and secure data recording capability. Standalone functionality when calibrating sensors, running qualification studies, and verifying sensor calibration. Accepts up to 48 inputs in any combination of thermocouple, voltage, and current inputs. System accuracy after calibration, taking into account all sources of error including the Kaye Validator and calibration and reference standards: 0.28 C at 121 C 0.38 C at 250 C 4 GB of internal memory to maintain qualification data during a study. Qualification data resides in internal memory until system is powered down or a new setup is loaded. Lithium ion backup battery to provide short-term power in case of AC power loss. A fully charged battery will provide approximately 120 minutes to perform an orderly shutdown of the system. 6.2. Kaye Validator Software Version 1.28 Create user accounts and assign user IDs, passwords and permission levels. Create validation study setups that can be run standalone or Console-controlled. When calibrating sensors and running qualification studies from the Console, the Kaye Validator AVS software features expanded tools for viewing and displaying data. Generate reports from secure data files that contain original calibration and qualification data recorded during the study. Access the Kaye Validator AVS audit trail, which provides a complete listing of events that affect the integrity of the Kaye Validator AVS software and the Kaye Validator AVS instrument. Fly out help guide in software operation. 6.3. Sensor Input Modules (SIMs) Sensor input module to which thermocouples or other measuring devices are connected. Each SIM has 12 sensor connections. Each SIM has a memory chip to store calibration offsets, SIM serial number, slot location, and the serial number of the Kaye Validator AVS measurement board on which calibration is performed. SIMs can be wired by the user or purchased pre-wired.

Document #: Revision: A PAGE: 5 of 16 6.4. Temperature Reference Temperature references provide the stable temperature required for sensor calibration. Amphenol Advanced Sensors has five temperature reference models available. They are designed to provide different temperature ranges and are compatible with the Kaye Validator. 6.5. IRTD LTR 90 (setpoint range 95 C to 140 C at 23 C ambient) Recommended for calibration of sensors used in freeze dryers, freezers, cryogenic units, cold rooms, incubators, and steam autoclaves. The LTR 90 accepts up to 12 thermocouples 7 stranded. LTR 25/140 (setpoint range 25 C to 140 C at 25 C ambient) Recommended for calibration of sensors used in freezers, cold rooms, incubators, and steam autoclaves. The LTR 25/140 accepts up to 18 thermocouples. LTR 40/140 (setpoint range 40 C to 140 C at 25 C ambient) Recommended for calibration of sensors used in freezers, cold rooms, incubators, and steam autoclaves. The LTR 40/140 accepts up to 18 thermocouples. HTR 400 (setpoint range 25 C above ambient to 400 C) Recommended for calibration of sensors used in steam autoclaves, dry heat ovens, and tunnel sterilizers. The HTR 400 accepts up to 24 thermocouples. For stability chambers and steam sterilizers, PTFE-insulated Type T thermocouples are used. For dry heat and tunnel sterilizers, Kapton-insulated Type T thermocouples are used. CTR 80 (setpoint range 80 C to 100 C) Recommended for calibration of sensors used in freeze dryers, freezers, and cryogenic units. The CTR 80 accepts up to 36 thermocouples. CTR 40 (setpoint range 40 C to 150 C) Recommended for calibration of sensors used in freezers, cold rooms, incubators, and steam autoclaves. The CTR 40 accepts up to 36 thermocouples. The Kaye IRTD 400 (M2801) temperature measurement standard is required for fully automatic sensor calibration and calibration verification. The IRTD 400 has a measurement accuracy of ±0.025 C within a temperature range of 195 C to 420 C. 7. SYSTEM ACCESS The Kaye Validator software is designed for three levels of users. Each level has specific permissions that define system access. System Administrator Responsible for the security of the program. The System Administrator creates and maintains user accounts, sets site options, backs up and restores user information, and views, prints, and maintains the audit trail. Unless administrator permissions are modified, an administrator cannot perform qualifications or actually operate the equipment. Supervisor Creates and modifies setups, calibrates sensors, runs qualification studies, and generates reports. Operator Uses prepared setups to calibrate sensors, run qualification studies, and generate reports.

Document #: Revision: A PAGE: 6 of 16 8. PROCEDURES This section describes the steps necessary to: Wire SIMs 8.1. Wiring SIMs Start the Kaye Validator software Create/modify a setup Connect the system Calibrate sensors Run the qualification study Verify sensor calibration Create reports from study data Access the audit trail There are two methods for wiring SIMs. You can assign sensor locations in the setup file and then wire the sensors according to the Setup Report, or you can buy pre-wired SIMs or wire the SIMs yourself and then create a setup that reflects the sensor locations. This section shows you how to wire SIMs yourself. 1. Loosen the two screws in the top of the SIM and open the cover. 2. Using appropriate thermocouple wire, cut lengthwise through the outer insulation to separate the red and blue wires, remove the resulting excess exterior sheathing, and then strip back each wire approximately ½ inch in order to make the connection with the SIM connectors. Connect the positive (+) lead to the positive (+) connector and the negative (-) lead to the negative (-) connector. The negative thermocouple lead is normally red, in conformance with ANSI standards. For Type T thermocouples, the positive lead is blue. 3. When validating moist heat processes, thermocouples with a sealed PTFE tip should be used with a drip cut through the outer insulation close to the SIM module to reduce the possibility of drawing moisture into the SIM. To add a drip cut, remove 4 inches (10 cm) of the outer insulation from each thermocouple at a point where natural drainage can take place to prevent water reaching the terminal screws. 4. AMPHENOL ADVANCED SENSORS also recommends shaving open about 1/4 inch from each of the red and blue inner wire jackets at opposite ends of the outer insulation drip cut to allow condensate drainage from within the individual wires. If moisture does collect in the SIM, remove the SIM from the instrument, open, and allow to air dry before storage. 5. Connect dry contact inputs directly to the SIM connectors. 6. Connect voltage inputs (up to 10 VDC) to the SIM connectors. Connect the positive (+) lead to the positive (+) connector and the negative (-) lead to the negative (-) connector. 7. For current applications, connect a 250Ω precision shunt resistor to the SIM connector in parallel with the input leads. The 250Ω resistor converts a 4-20 ma signal to 1-5 volts. Connect the positive (+) lead from the power supply to the positive (+) SIM connector, the positive (+) lead from the device to the negative (-) SIM connector, and the negative (-) lead from the device to the negative (-) power supply connector. 8. Once all sensors are connected, route the wire harness around the terminal connectors and out the drain hole. Secure the wire harness with tie wraps. 9. Close the SIM cover and tighten the screws.

Document #: Revision: A PAGE: 7 of 16 8.2. Starting the Kaye Validator AVS Software On the Console home page, click the Kaye Validator AVS tile. The Kaye Validator AVS user login screen appears, followed by the Hub page. The Hub page provides access to the major functions of the Kaye Validator software. 8.3. Creating/Modifying a Setup The setup defines everything required to calibrate sensors, run qualification studies, and verify sensor calibration. The setup is created/modified using the Kaye Validator software. The Console does not have to be connected to the Kaye Validator AVS to create/modify setups. 1. On the Hub page, click Assets tile, then on Asset Details page click Setups tab. 2. Existing setups are listed by the date and time the files were saved. To create a new setup, click New, enter up to 35 characters to describe the setup, and click OK. To modify an existing setup, select the setup from the file list and click on edit icon. The Sensors screen appears. 3. Add, modify or delete sensor configurations as necessary. Select sensors by clicking a sensor tiles in the SIM. To add or modify a sensor definition, enter/modify the sensor label and sensor description, and select a sensor type. Specify scale factors as required. To automatically number multiple sensor definitions, select both Automatic Numbering check boxes to append 2 digits in sequence to the sensor label and description label. Click OK. 4. To remove sensor location(s), click the sensor tile again to deselect it. 5. Identify/modify groups. Click the Group Sensors tab. Change group titles or add groups as necessary (up to 25 groups, 35 characters) in the group list box. 6. Assign sensors to groups. Select a group from the list box. Assign sensor locations as necessary. Repeat for all groups as necessary. Note: Sensors of different types should not be mixed in a group. 6. Specify calculations and events. Click the Calculations tab. Specify calculations as necessary. Define the event as necessary. Repeat for all sensors as necessary. 7. Specify calibration parameters. Click the Calibration tab. Set the Temperature Setpoints as necessary. Set the Stability Criteria as necessary. Set the Deviation Criteria as necessary. 8. Specify qualification study conditions.

Document #: Revision: A PAGE: 8 of 16 Click the Qualification tab. Click the Modify button. The Start/Stop Conditions screen appears. Set the Start/Stop Conditions as necessary and click OK. Set the Data Storage File Rate and Kaye Validator Printer Rate as necessary. Set the Output Relay to actuate based on event if necessary. 9. Save the setup file. Click Save. The Save Setup File screen appears. Enter User Id and password. Click OK. The Setup tile screen appears. 10. Click home to return to the Home page.

Document #: Revision: A PAGE: 9 of 16 8.4. Connecting the System All system connections are on the back of the Kaye Validator AVS. Each connection port is labeled with an icon representing its function. Connection ports are as follows: 2 IRTD ports (4 pin DIN connector) 1 temperature reference port (4 pin DIN connector) 1 USB port labeled with a USB icon, is available for connecting an ICAL device 2 USB/Ethernet communications ports and one built in WiFi port for connecting the I/O 2 Relay NC/NO port 1. Make sure the Kaye Validator AVS and the temperature reference are powered off before connecting the system. 2. Insert the IRTD into the temperature reference. 3. Connect the IRTD to one of the appropriate (4 pin DIN connector) sockets, using the cable supplied with the IRTD. 4. Connect the temperature reference to its appropriate (4 pin DIN connector) socket, using the cable supplied with the temperature reference. 5. Connect the Console to AVS Dock. 6. Connect the temperature reference power cable to a grounded power source. 7. Connect the Kaye Validator AVS power cable to a grounded power source. 8. Insert the wired SIMs into the SIM slots. Label each SIM with the associated slot number. 9. Turn the temperature reference power switch to the ON position. 10. Turn the Kaye Validator AVS power switch to the ON position. 8.5. Checking Hardware and Communications Connections The Kaye Validator AVS software graphically displays the validation system hardware and communications connections. It is good practice to check communications before starting the validation process. 1. Click the Hardware tab. Each hardware item is shown with a communication link. A Device Not Found symbol (Grey color line) indicates a communication problem.

Document #: Revision: A PAGE: 10 of 16 8.6. Operation This section includes procedures for operating the Kaye Validator AVS. s include sensor calibration, running the qualification study, and verifying sensor calibration. 8.6.1. Sensor Calibration Calibrate sensors before running the qualification study. Standalone sensor calibration requires a Kaye temperature reference and IRTD. 1. Place the Kaye Validator AVS (with temperature reference and IRTD connected) in a location with stable and even temperature, not exposed to any local heating or cooling sources (i.e., close to a sterilizer, an open door causing a draft, etc.). 2. Power up the Kaye Validator AVS and let it run for approximately 30 minutes in the operating environment in order to acclimate to the ambient temperature. Note: The fan of the temperature reference should not blow toward the Kaye Validator AVS. 3. Insert the thermocouple sensors into the temperature reference. 4. Load the setup from Console into the Kaye Validator AVS. On Console, go to Set Up tab of Asset details screen, select Set Up file and click on initiate calibration button. Select Kaye Validator AVS. 5. Select sensors to calibrate. Select Sensors on sensor selection screen appears. To select individual sensors by clicking on it or use select all button Click Initiate Calibration button 6. Calibrate sensors. Press Start. The Kaye Validator AVS automatically performs sensor calibration at the specified setpoints. When calibration is complete Press Save to save the file.

Document #: Revision: A PAGE: 11 of 16 8.6.2. Qualification Study The Kaye Validator AVS can perform standalone qualification studies manually, automatically, or in combination, depending on the start/stop conditions defined in the setup. Before starting this procedure, be sure you have conducted a sensor calibration and have loaded the appropriate setup into the Kaye Validator AVS. 1. If not already on, power up the Kaye Validator AVS and let it run for approximately 30 minutes in the operating environment in order to acclimate to the ambient temperature. 2. Position sensors in vessel or chamber being tested. 3. Insert a new (or freshly formatted) memory card into the USB card port. 4. Start the qualification study. If the start condition is defined as Manual: On Console go to Set Up tab of Asset details screen, select Set Up file and click on initiate qualification button. Select Kaye Validator AVS Real Time screen appears. Press Start button. 5. View qualification study data. List view displays all the real time data of sensors in groups. Graph view displays real time data on line graph for group that is selected. 6. Start the exposure cycle. If the start condition is defined as Manual: Press Start Exposure on real time screen on Console 7. Stop the exposure cycle. If the stop condition is defined as Manual: Press Stop Exposure on real time screen on Console 8. Stop the qualification study. If the stop condition is defined as Manual: Press Stop Qualification. on real time screen on Console

Document #: Revision: A PAGE: 12 of 16 8.6.3. Calibration Verification Verify sensor calibration after running the qualification study. Standalone calibration verification requires a Kaye temperature reference and IRTD. 1. If not already on, power up the Kaye Validator AVS, temperature reference and IRTD and let them run for approximately 30 minutes in the operating environment in order to acclimate to the ambient temperature. Note: Again, be sure the Validator is in a location with stable and even temperature, not exposed to any local heating or cooling sources (i.e., close to a sterilizer, an open door causing a draft, etc.) The fan of the temperature reference should not blow toward the Kaye Validator AVS. 2. Insert the thermocouple sensors into the temperature reference. 3. Load the setup from Console into the Kaye Validator AVS. On Console, go to Set Up tab of Asset details screen, select Set Up file and click on initiate verification button. Select Kaye Validator AVS. 4. Select sensors to verify. Select Sensors on sensor selection screen appears. To select individual sensors by clicking on it or use select all button Click Initiate Verification button 5. Verify sensors. Press Start. The Kaye Validator AVS automatically performs sensor calibration at the specified setpoints. When Verification is complete Press Save to save the file.

Document #: Revision: A PAGE: 13 of 16 Generating Reports Reports are generated from secure data files that can only be read by the Kaye Validator software. You create reports each time from the secure data files. You can generate Setup, Calibration, Qualification, and Calibration Verification Reports to document the specifics of your validation study. You can generate PDF reports too. 8.6.4. Generating Setup Reports Setup Reports can be generated from setup file. This section documents the procedure for generating an individual Setup Report. : 1. Select the report file. On the Console, either go to Set Up tab of Asset details screen of Kaye Validator AVS application. Select Setup and click on edit icon. Go to review pane and generate Report in PDF. Or On the Console, go to report file selection screen after clicking on Asset tile of AVS Reporting Tool application. Select Setup file and click on Report icon Generate report in PDF. 8.6.5. Generating Calibration Reports This section documents the procedure for generating Calibration Reports from data stored during a sensor calibration. : 1. Select the report file. On the Console, either go to Calibration tab of Asset details screen of Kaye Validator AVS application. Select Calibration file and click on Generate Report button. Go to review pane and generate Report in PDF. Or On the Console, go to report file selection screen after clicking on Asset tile of AVS Reporting Tool application. Select Calibration file and click on report icon Generate report in PDF. 8.6.6. Generating Qualification Reports You can print qualification data exactly as it was recorded, or you can customize your report by adding statistical calculations, events, and event timers based on the stored data. The complete Qualification Report set includes the Setup, Qualification Detailed, and Qualification Summary Reports. : 1. Select the report file. On the Console, either go to Qualification tab of Asset details screen of Kaye Validator AVS application. Select Qualification file and click on Generate Report button.