DFI 9000 PORTABLE DIGITAL INDICATOR USER S GUIDE

Similar documents
DFI 7000 DIGITAL FORCE INDICATOR USER S GUIDE 1.0 INTRODUCTION 2.0 DESCRIPTION. 2.1 Configuration. 2.2 Specifications

TM0-1 AMPLIFIER / CONDITIONER MODULE OPERATORS MANUAL. Transducer Techniques R

SA SERIES Arlingate Lane, Columbus, Ohio (614)

MODEL 245 / 345 High Range

MODEL 245 / 345 DigiComp

DTW 400/DTW Digital Torque Tester/Screwdriver User s Guide

PIECAL 322 Automated Thermocouple Calibrator Operating Instructions. Product Description. Practical Instrument Electronics

PIECAL 510B/511B RTD Simulator Operating Instructions

Model SC500. Programmable Single-Channel Transducer Indicator/Conditioner DESCRIPTION FEATURES

LCA LOAD CELL AMPLIFIER SIGNAL CONDITIONER

HI ma Simulator

STATIC METERS Models 211 & 212. SURVEYORSTAT Models 211XL & 212XL

PCM-7140 Pulsed Current Source Operation Manual

PIM-Mini Pulsed Current Source Operation Manual

User's Guide. MiniTec TM Series Model MN25 MultiMeter

PIECAL 520B & 521B Thermocouple Source Operating Instructions

PRODUCT CONTENTS... 3 II. REQUIRED MATERIALS... 3 III. OVERVIEW... 3 IV. BENCH LAYOUT...

HI ma Simulator

AUTOMATION. Operator s Manual. PG5 Digital Pressure Gauge

PSM-2 POWER SUPPLY MODULE INSTRUCTIONS

V-9939B MICROPHONE ADAPTER

PSD Strain Gauge or Load Cell Hand Held Display TEDS enabled

PRO 360. Owner's Manual

AUTOMATION. Operator s Manual PG-5. Full Access. DOC Rev. A1 3/10

OPERATION MANUAL MODEL TT-LP

Operation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter

Operation Manual TX10 1/8 DIN Microcomputer Based ph/orp Controller

PRO Owner's Manual

Model 206 Operating Instructions. Setra Systems, Inc. 159 Swanson Road, Boxborough, MA

I/O SIGNAL CONDITIONER

Installation & Operation

DVC-350C CALIBRATORS, INC HANDHELD BATTERY-POWERED VOLTAGE CALIBRATOR FEATURES INTRODUCTION GENERAL DESCRIPTION

Speaker Selectors Models SSW-L4 EX and SSW-L6 EX. User Manual. SSW-L4 EX (bottom) and SSW-L6 EX (top)

CTI 2550 EIGHT CHANNEL ISOLATED ANALOG INPUT MODULE INSTALLATION AND OPERATION GUIDE. Version 2.0 CTI Part # IOG $25

7561-PSD Manual Portable Battery Powered Indicator

P027 Series Pressure Transducer Installation & Operation Manual

DFI 1550/1650 SIGNAL CONDITIONING, SELF-CALIBRATING INDICATOR USER S GUIDE

Model VT120 Fast Timing Preamplifier Operating and Service Manual

Model SRV945-BT 24 X 36 Platform Scale App-Enabled Operating and Service Manual Part No.: MAN945-BT_ Page 1 of 16.

LCI User Manual mantracourt.com

Two Channel 48 Volt Phantom Power Supply. Artcessories. User's Manual

LCI Load Cell Junction Box with Fault Monitor

Omnitron Systems Technology, Inc. 1. iconverter. 19-Module Managed Power Chassis User s Manual

User's Guide. Model High Precision Quad Output DC Power Supply

HASTINGS INSTRUCTION MANUAL HASTINGS POWER SUPPLY MODEL A-PAGE1

Pennsylvania Scale Company Model 400B Battery Powered Digital Indicator Operation & Calibration Manual

Global Water Instrumentation, Inc.

4-20mA Display. Installation and Operation Manual. Rev 4/3/2019 Part #

Cole-Parmer Instrument Company

B2800 FLOW MONITOR. - For Liquid Meters - PROGRAMMING & INSTALLATION MANUAL Simplified Version

PRO 360. Owner's Manual

OPERATION MANUAL BO BLASTER S OHMMETER

Manual No Revision MODEL GLI-9835 HAND-HELD/PORTABLE CONDUCTIVITY/TDS METER

DWYER INSTRUMENTS, INC. Series AVUL Air Velocity Transmitter. Specifications - Installation and Operating Instructions.

DSCUSB Standard User Manual. mantracourt.com DSCUSB. Digital Sensor Card USB Strain Gauge Input

Cole-Parmer Instrument Company

User s Guide. 600A True RMS AC/DC Clamp Meter. Model 38389

Analog Interface Unit Operation Manual

CTI 2552-A 8-CHANNEL ISOLATED RTD INPUT MODULE INSTALLATION AND OPERATION GUIDE. Ver. 1.2 CTI Part #

SIAC-PS 3G SIGNAL ISOLATOR WITH POWER SUPPLY (PART NO. 8890) INSTALLATION INSTRUCTIONS

Model CR103J. DC Current and Voltage Calibrator/Standard ±0.01µA to ± mA ±0.1µV to ± V. Operating Manual

Model 500 Digital Weight Indicator

SY021 Portable Load/Force Meter User instructions

4-20/10-50 ma/v Loop Calibrator

Model CD-AW. Infrared Environmental CO2 Sensor. Operator's manual

DMS-20LCD-4/20 Series

CR87B TEMPERATURE RECORDER

Secured Series: Hub Plus Kit Single Door Controller Package Installation Manual

SERVICE MANUAL MODEL WPP-531-X-ADA MODEL SSP-511- X-ADA

LUDLUM MODEL ALPHA/BETA DETECTOR. August 2015

OPERATION MANUAL. for the BO BLASTER S OHMMETER

Hybrid AC Driver [GCNC-1110]

TSD-DA28 2x8 Balanced Line Distribution Amplifier

TECHNICAL MANUAL PRESSURE TRANSDUCER TYPE MLP

PC 150. Digital Bench Scale. Operation Manual

Global Water globalw.com. Global Water. Instrumentation, Inc.

TELEDYNE HASTINGS INSTRUMENTS

Proliphix EPA-60 Installation Guide

Model ROTARY TORQUE, W/ENCODER and AUTO ID. Installation and Operating Manual

MWR30. INSTALLATION/OWNER S MANUAL Wired Remote Control MARINE

MODEL ATV 4/6 BATTERY POWERED DIGITAL VACUUM INSTRUMENT

ANC Series RS-422 Serial Communications Adapter

V-9908 MESSAGE/PAGE PANEL

Remote Control for Inverter RC-300. Owner's Manual. Please read this manual BEFORE operating your RC-300 Remote Control

MANUAL INSTRUCTION MMP-10. Multiplexer

UWE-1707 SERIES CONTENTS OPERATION MANUAL ELECTRONIC WEIGHING INDICATOR 1. SPECIFICATIONS 2. INSTALLATION

Xxxxx Xxxxxx OM-CP-QUADVOLT. OM-CP-OCTVOLT 8-Channel Low Level DC Voltage Data Logger. 4-Channel Low Level DC Voltage Data Logger INSTRUCTION SHEET

VIP-804 QUAD ENHANCED NETWORK AUDIO PORT

RAILROAD SYSTEM OPERATION MANUAL

POWER SERIES Plus Watt-VAR / Watt-Power Factor Digital Switchboard Meter User s Manual IM2492WVPF-1

INSTRUCTION MANUAL. ph/mv CONTROLLER/RECORDER

V CHANNEL MODULAR PRE-AMP/MIXER V-9984 REMOTE INPUT MODULE V CHANNEL MODULAR PRE-AMP/MIXER

PSI DITHERLESS BIAS CONTROLLER USER GUIDE

Cantata m100 Amplifier

SERVICE MANUAL MODEL WPP-331-D WEATHERPROOF TELEPHONE MODEL HOB-331-D WEATHER RESISTANT TELEPHONE. and MODEL SSP-311-D STAINLESS STEEL PANEL TELEPHONE

NOTE: The F2 button and the 4 LED annunciators do not function on the LD-ACF-I. To use these features, ask about the LD-ACF-R4/R4A.

Remote Control for Inverter RC-200. Manual. Please read this manual before operating your RC-200 Remote Control

KXR20 Voltage Regulator

User Manual DOCUMENT #601119REF. EcoSense ph1000a. ph, mv (ORP), and Temperature Benchtop Meter USER MANUAL. English

V-9941A ONE ZONE HANDSFREE TALKBACK CONTROL UNIT

Transcription:

DFI 9000 PORTABLE DIGITAL INDICATOR USER S GUIDE www.cooperinstruments.com PH: 540-349-4746 FAX: 540-347-4755

CONTENTS 1.0 INTRODUCTION...1 2.0 SETUP AND USE...1 2.1 Your DFI 9000...1 2.2 Additional Tools...1 2.3 Setup Overview...1 2.4 Connect the Transducer...2 2.4.1 Cable Type...2 2.4.2 Connection...2 2.5 Turn On the Unit...2 2.6 Set the Decimal Point Location...3 2.7 Verify Calibration...3 2.8 Fine Tune...4 2.9 Using the DFI 9000...4 3.0 CALIBRATION...4 3.1 When to Calibrate...4 3.2 Methods of Calibration...4 3.3 Calibration Overview...4 3.4 Center the Fine Tuning...4 3.5 Disable Scaling...4 3.6 Adjust for No Load...5 3.7 Adjust for a Known Load...5 3.7.1 Standard Calibration...5 3.7.2 Shunt Calibration...5 4.0 OPTIONS...6 4.1 Peak Detector...7 4.2 Analog Voltage Output...7 4.3 High Input...7 5.0 MAINTENANCE AND REPAIR...7 5.1 General Care...7 5.1.1 Cleaning...7 5.1.2 Storing...7 5.2 Troubleshooting...7 5.2.1 Low Battery...7 5.2.2 Overrange indications...7 5.2.3 Erratic Readings...8 6.0 SPECIFICATIONS...8 6.1 Physical Characteristics...8 6.2 Environmental...8 6.3 Power Supply...8 6.4 Transducer Interface...8 6.5 Amplifier...8 6.6 Digital Display...9 6.7 Options...9 7.0 WARRANTY REPAIR POLICY...9 CF12 DFI 9000 ii Rev. C May 2004

1.0 INTRODUCTION The DFI 9000 Portable Indicator is a battery-powered signal conditioner and readout device. The DFI provides precise electrical drive to a transducer and converts the transducer output into a digital of engineering units. The signal conditioner supports transducer outputs between 1 and 4 millivolts/volt (mv/v); a high input option supports outputs up to 40 mv/v. The readout device is a 4-1/2 digit, liquid-crystal display. Enclosed in a rugged, weatherproof aluminum case, the DFI 9000 is specially designed for remote field operations. Quick field setup allows the DFI 9000 to be used for a variety of applications, from checking the load on an oil well drilling bit to determining natural gas line pressure. Coarse zero and Course span controls allow broad adjustment for a wide variety of transducer outputs. To prevent accidental misadjustment, these controls are located on the edge of the circuit board inside the unit, easily accessible when the panel is removed from the case. Fine zero and fine span controls on the front panel allow precise adjustment for offsets and gain variations up to +/-15 percent of full scale. An optional peak detector toggles between peak mode and track mode via a front-panel switch. In both modes, the unit monitors the input signal. In peak mode, the DFI 9000 stores, digitizes, and displays the highest signal level. 2.0 SETUP AND USE 2.1 Your DFI 9000 The DFI 9000 can be purchased in one of two ways: As a stand-alone unit with a mating connector for the input connector As a set with a transducer and a mating connector for the input connector Some transducers are equipped with internal cables that are used to connect the transducer to the DFI 9000. Cooper offers separate cables for transducers without integral cables. If you purchased Assembly /Calibration with the DFI 9000 and a Cooper transducer, as a system, they are calibrated together as a system. If Assembly/Calibration was not purchased, follow the setup procedures in this section before using the unit. If you purchased the DFI 9000 as a stand-alone unit or if you replace the original transducer, follow the setup and calibration procedures before using the unit for the first time. 2.2 Additional Tools Use the following tools to set up and calibrate your DFI 9000: Two inches of #20 AWG wire (if needed) Needle-nose pliers Small, standard screwdriver Calculator Solder and soldering iron (if needed) 2.3 Setup Overview 1. Connect the transducer to the unit 2. Turn on the unit 3. Set the decimal point location 4. Verify the calibration 5. Fine tune CF12 DFI 9000 1 Rev. C May 2004

2.4 Connect the Transducer 2.4.1 Cable Type Determine the distance between the transducer s installed location and the DFI 9000. If the distance is less than 30 feet, use a four-wire hookup. If it is greater than 30 feet, use a six-wire hookup and the sense feature. Transducers with integral cables generally have four wires, designated: + Excitation - Excitation + Signal (or + Output) - Signal (or Output) For long six-wire cables, the additional two wires are used for sense feature and are designated: + Sense - Sense The sense feature reduces the effect of signal voltage loss over a long cable, increasing display accuracy. The sense feature is only needed for cables longer than 30 feet. 2.4.2 Connection Use the transducer s integral cable if available. If the transducer is not equipped with an integral cable, wire your own cable as follows: Four wire hookup (Figure 1)-Disable the sense feature by soldering two small wires inside the mating connector to jumper: + Excitation (pin A) to + Sense (pin B) - Excitation (pin C) to Sense (pin D) DFI 9000 MATING CONNECTOR Six-wire hookup (Figure 2)- Jumpers are not needed in this configuration. DFI 9000 MATING CONNECTOR 2.5 Turn On the Unit Flip the on/off toggle switch on the front panel to on and allow the DFI 9000 to stabilize for a few seconds. 1. Make sure that no load is applied to the transducer, then press and hold the shunt cal button. 2. If the number on the display drifts, the unit is not yet stable. 3. Release the shunt cal button and try again in a few seconds. CF12 DFI 9000 2 Rev. C May 2004

2.6 Set the Decimal Point Location The DFI 9000 is factory set to the correct decimal point location for its original Cooper transducer. As a standalone product, it is factory set to xxx.xx. You can easily change the decimal point location to match your load and desired display. Each decimal point has two terminals that must be connected. For example, to set the decimal point to xxxx.x for a 1000 psig load, jumper the first and second terminals (from the front of the board). For smaller loads (such as 200 psig), setting the decimal point to xxx.xx may provide higher resolution, but may also result in an unstable last digit. For stable last digit, set the decimal point to xxxx.x. To set the decimal point location: 1. Remove the unit from the case by loosening all four captive fasteners on the front panel and lifting the panel out of the case. 2. Find the decimal point jumper locations 3. Use needle-nose pliers to remove the existing jumper and place it in the desired terminals. TOP OF FRONT PANEL FIGURE 3: DECIMAL POINT JUMPER LOCATIONS 2.7 Verify Calibration For best results, verify the calibration daily. Note: When not using the original Cooper transducer, calibrate the DFI 9000 before use. 1. Obtain the following values from the transducer calibration data sheet: - Shunt-cal output (or shunt-cal factor ) in mv/v - Calibration factor (or full-scale output ) in mv/v - Full-scale engineering units (or capacity ) in psig, lbs, etc. 2. Perform this calculation and record the value. Shunt-cal output X full-scale engineering units = shunt-cal engineering units Calibration factor Note: Place a decimal point in the full-scale engineering units value in the equation in the same place as set in the display. CF12 DFI 9000 3 Rev. C May 2004

3. With no load on the transducer, press and hold the shunt cal button. Compare the displayed value with the result of step 2. If the values are not equal, adjust the fine span potentiometer according to the procedure below. 2.8 Fine Tune Fine tune the DFI 9000 (the controls for both zero and span are on the front panel): 1. Fine zero- With no load on the transducer, use a small screwdriver to turn the zero potentiometer until the display reads 0000. 2. Fine span- With the known load on the transducer, use a small screwdriver to adjust the span potentiometer until the display matches your known load. OR 3. With no load on the transducer, press and hold the shunt cal button. Use a small screwdriver to adjust the span potentiometer until the display matches the value calculated in Section 2.7. Note: Turn the potentiometer clockwise to increase the displayed value. Turn it counterclockwise to decrease the displayed value. 2.9 Using the DFI 9000 Verify the calibration daily for best results. To use the DFI 9000, simply apply the load on the device to be measured and read the display. Recalibrate the DFI 9000 each time you use it with a different transducer. For details about using the peak detector and other options, see Section 4. 3.0 CALIBRATION 3.1 When to Calibrate Calibrate the DFI 9000 if you are not using the original Cooper transducer and each time you change transducers. 3.2 Methods of Calibration Two methods of calibration are available: standard and shunt. Use standard calibration when you can place a known load on the transducer. Use shunt calibration to stimulate the load when you cannot apply a known load or foe quick field setup. 3.3 Calibration Overview Remove the unit from the case by loosening all four captive fasteners on the front panel and lifting the panel out of the case, then do the following: 1. Center the fine tuning 2. Turn off the scaling 3. Adjust for no load 4. Adjust for a known or calibrated load 5. Adjust the scaling 3.4 Center the Fine Tuning Center the zero and span fine-tuning potentiometers to allow precise adjustment in either direction. Use a small screwdriver to turn each potentiometer up to 15 turns clockwise until it clicks. Then, turn it back halfway, 7-1/2 turns counterclockwise (Figure 4). 3.5 Disable Scaling Scaling is the ratio of the output voltage to the displayed reading. Disable scaling so it does not interfere with calibration. Use a small screwdriver to turn the scaling potentiometer up to 15 turns clockwise until it clicks (Figure 4). CF12 DFI 9000 4 Rev. C May 2004

FRONT PANEL CIRCUIT BOARD DECIMAL PT. TERMINAL STRIP COARSE SPAN COARSE ZERO SCALING SHUNT CAL RESISTOR 3.6 Adjust for No Load FIGURE 4: POTENTIOMETER LOCATIONS 1. Remove any load from the transducer. 2. Coarse zero-this control zeroes the display when the fine zero is out of range or the transducer has a preload, offset, or tare. 3. Fine zero-use a small screwdriver to turn the fine zero potentiometer on the front panel until the display reads exactly 0000. Note: Turn the potentiometer clockwise to increase the displayed value. Turn it counterclockwise to decrease the displayed value. 3.7 Adjust for a Known Load If you can apply a known load to the transducer, use the standard calibration method. If you cannot apply a known load, use the shunt calibration method. 3.7.1 Standard Calibration 1. Coarse span-this control adjusts the span for each different transducer. (a) Apply a known load (near full scale) to the transducer. (b) Use a small screwdriver to turn the coarse span potentiometer counterclockwise until the display reads approximately 19999. (c) If the known load is not full scale, scale down this value proportionately. For example, if you apply an 80-lb load to a 100-lb transducer, scale down 19999 as shown: 80 x 19999 = 15999 100 2. Fine span-with the known load applied, use a small screwdriver to turn the fine span potentiometer until the display reads exactly 19999 (or the scaled-down value). 3. Scaling-This control scales the display to the desired units. With the known load applied, use a small screwdriver to turn the scaling potentiometer until the display reads the desired full-scale engineering units (found on the transducer s calibration sheet or the result of step 1c). After scaling, return the panel to the case and tighten the four captive fasteners. If the display does not read 0000, fine tune the unit again according to the procedure in Section 3.4. 3.7.2 Shunt Calibration Use shunt calibration when you cannot apply a known load to the transducer. When the shunt cal button is pressed (with no load on the transducer), a precise resistor of known value is shunted across one leg of the transducer. This resistor simulates a known load or pressure. The internal shunt cal resistor is installed on two pin jacks on the bottom of the circuit board. Note: The internal shunt cal resistor is normally 59 kohm when used with Cooper transducers. To install a different calibration resistor, remove the present resistor from the jack and plug in a new one. See the information sheet supplied with the transducer for the recommended value for the shunt cal resistor and other relevant information. CF12 DFI 9000 5 Rev. C May 2004

1. From the transducer calibration data sheet, obtain the following values: - Shunt-cal output (or shunt-cal factor ) in mv/v - Calibration factor (or full-scale output ) in mv/v - Full-scale engineering units (or capacity ) in psig, lbs, etc. 2. Coarse span-this control alters the gain of the amplifier or readjusts the span when calibrating to a different transducer. Determine the unscaled shunt-cal display value using the following formula: Shunt-cal output x 19999 = unscaled shunt-cal display Calibration factor With no load on the transducer, press and hold the shunt cal button. Use a small screwdriver to adjust the fine span potentiometer until the display reads the exact unscaled shuntcal display value from step 2. 3. Fine span-with no load on the transducer. Press and hold the shunt cal button. Use a small screwdriver to adjust the fine span potentiometer on the front panel until the display reads the exact unscaled shunt-cal display from step 2. 4. Scaling-This control scales the display to the desired units. Determine the scaled shunt-cal value using the following formula: Shunt-cal output x full-scale units = scaled shunt-cal engineering units Calibration factor With no load on the transducer, press and hold the shunt cal button. Use a small screwdriver to adjust the scaling potentiometer until the display reads the scaled shunt-cal engineering units. After scaling, return the panel to the case and tighten the four captive fasteners. If the display does not read 0000, fine tune the unit again according to the procedure on page 5. Shunt Calibration Example: The following values are taken from a typical transducer calibration data sheet: shunt-cal output = 1.4683 mv/v calibration factor = 2.9999 mv/v full-scale units = 25 psig To calculate the unscaled shunt-cal display value: unscaled shunt-cal display= shunt-cal output x 19999 = 1.4683 mv/v x 19999 = 9788 calibration factor 2.9999 mv/v To calculate the scaled shunt-cal value: Scaled shunt-cal value = shunt-cal output x full-scale units = 1.4683 mv/v x 25 psig = 12.236 calibration factor 2.9999 mv/v 4.0 OPTIONS The DFI 9000 has three options: Peak detector Analog voltage output High input See below for brief descriptions of the DFI 9000 options. For more information about the options, contacts Cooper s Sales Department at 1-800-344-3921. CF12 DFI 9000 6 Rev. C May 2004

4.1 Peak Detector The peak detector operates in two modes: peak and track. In both modes, the unit monitors the input signal. Use the toggle switch on the front panel to select the desired mode. The peak detector always operates in either peak or track mode. In peak mode, the DFI 9000 detects and displays the highest positive peak signal level. After two conversions of the analog-to-digital converter (at 1.5 conversions per second), the unit stores the highest value. If the DFI 9000 identifies a higher peak, the peak detector reactivates to record and display the new value. To cancel a displayed peak value, flip the peak/track switch to track. In track mode, the peak detector continuously follows the input signal but does not store the highest value. 4.2 Analog Voltage Output Cooper offers an optional 0-1 volt analog output (for example, to connect to a chart recorder). For this option, a double banana-plug jack (positive red terminal) is mounted on the front panel. Loading current must not exceed 2 milliamps. 4.3 High Input Cooper makes a low-gain version of the DFI 9000 to support transducer outputs between 10 and 40 mv/v. The DFI 9000 can also be field-converted for high input. For more information, contact Cooper Instruments at 1-800- 344-3921. 5.0 MAINTENANCE AND REPAIR 5.1 General Care 5.1.1 Cleaning Clean the exterior and front panel of the DFI 9000 with a damp (not wet) soft cloth. 5.1.2 Storing Store the DFI 9000 at temperature of -20 F to 150 F. 5.2 Troubleshooting 5.2.1 Low Battery The red low battery light on the front panel indicated battery voltage below 4.6 volts or an incorrectly installed battery. Although the unit can be operated with weak batteries, it may produce inaccurate readings Replace the batteries according to the following procedure: 1. Flip the on/off toggle switch on the front panel to off. 2. Loosen all four captive fasteners and remove the panel from the case. 3. Replace all four D batteries, matching polarities to the indications on the battery holder. 4. Replace the front panel and tighten the fasteners. Note: We recommend alkaline batteries. To safeguard the circuitry, remove the batteries when not using the unit for a long period of time. 5.2.2 Overrange indications Note: The DFI 9000 indicates an ovverange condition by a flashing left arrow on the display. An overrange indication often results from faulty sense pin wiring (such as a missing or broken wire). Determine the source of the problem using the following procedure: 1. Check all electrical connections for continuity. 2. Check the wiring code for pin layout. 3. Verify that the unit was set up according to the instructions in this manual. CF12 DFI 9000 7 Rev. C May 2004

4. Determine if the problem is with the transducer or the DFI 9000: a. Remove the transducer from the DFI 9000 b. Jumper output (pin E) and +output (pin F). c. If the display reads close to 0000, the transducer may be faulty. Try using another transducer with the DFI 9000, calibrating first. 5. If an ovverange indication is still present, the problem may be with the DFI 9000. Return the DFI 9000 to Cooper, following our Warranty and Repair policy. 5.2.3 Erratic Readings Erratic readings could indicate that the batteries are weak. 1. Check the batteries for low voltage or mismatched polarities. Replace weak batteries. 2. If the batteries are strong and installed properly but erratic readings are still present, return the DFI 9000 to Cooper following our Warranty and Repair Policy. Caution: Any attempt to repair the unit by anyone other than the Cooper Customer Service Department voids the product warranty. 6.0 SPECIFICATIONS 6.1 Physical Characteristics Size 7in. wide x 5 in. deep x 5 in. high Number of channels 1 Material Deep-drawn, weatherproof aluminum case Weight 3.5 lbs. 6.2 Environmental Temperature, storage -20 F to 150 F Temperature, operating 32 F to 130 F 6.3 Power Supply Battery type Battery life 4 D cells (alkaline recommended) 250 hours (for alkaline) Note: To safeguard the circuitry, remove the batteries when not using the unit for a long period of time. 6.4 Transducer Interface 6.5 Amplifier Transducer excitation 4 Vdc (constant voltage) Transducer output range 1 to 4 mv/v 10 to 40 mv/v (high input option) Transducer bridge range 250 ohms minimum Electrical connection type PT02A-10-6S (mating connector) Zero balance +/-15% full scale (minimum) Noise and Ripple 30 µv Full-scale output (optional) 0-1V (output impendence < 1 ohm) Non-linearity (% F.S.) (max) +/-0.02% Drift (max. zero and span) +/-5mV Zero stability (% F.S./yr.) 0.1% Span stability (% F.S./yr.) 0.1% Fine span adj. (% range) +/-15% Coarse span adj. (% range) 100% Fine zero adj. (% range) +/-15% Coarse zero adj. (% range) 100% CF12 DFI 9000 8 Rev. C May 2004

6.6 Digital Display 6.7 Options Type Liquid-crystal display (LCD) Number of Display characters 4-1/2 digits Character size 0.4 in. Conversion rate 1.5/second Scaling 0.19,999 Scaling method Potentiometer Maximum display count 19,999 Polarity indications? Yes Programmable decimal points? Yes Overrange indicator Flashing left arrow Resolution 1/20,000 Maximum sensitivity 0.1µV/count High Input Analog voltage output Peak/hold detector 0-40 mv/v 0-1Vdc (2mA maximum) Bleed rate of 0.01% second or less 7.0 WARRANTY REPAIR POLICY Limited Warranty On Products Any Cooper Instruments product which, under normal operating conditions, proves defective in material or in workmanship within one year of the date of shipment by Cooper will be repaired or replaced free of charge provided that a return material authorization is obtained from Cooper and the defective product is sent, transportation charges prepaid, with notice of the defect, and it is established that the product has been properly installed, maintained, and operated within the limits of rated and normal usage. Replacement or repaired product will be shipped F.O.B. from our plant. The terms of this warranty do not extend to any product or part thereof which, under normal usage, has an inherently shorter useful life than one year. The replacement warranty detailed here is the buyer s exclusive remedy, and will satisfy all obligations of Cooper whether based on contract, negligence, or otherwise. Cooper is not responsible for any incidental or consequential loss or damage which might result from a failure of any and all other warranties, express or implied, including implied warranty of merchantability or fitness for particular purpose. Any unauthorized disassembly or attempt to repair voids this warranty. Obtaining Service Under Warranty Advance authorization is required prior to the return to Cooper Instruments. Before returning the item, contact the Repair Department c/o Cooper Instruments at (540) 349-4746 for a Return Material Authorization number. Shipment to Cooper shall be at buyer s expense and repaired or replacement items will be shipped F.O.B. from our plant in Warrenton, Virginia. Non-verified problems or defects may be subject to a $100 evaluation charge. Please return the original calibration data with the unit. Repair Warranty All repairs of Cooper products are warranted for a period of 90 days from date of shipment. This warranty applies only to those items that were found defective and repaired; it does not apply to products in which no defect was found and returned as is or merely recalibrated. It may be possible for out-of-warranty products to be returned to the exact original specifications or dimensions. * Technical description of the defect: In order to properly repair a product, it is absolutely necessary for Cooper to receive information specifying the reason the product is being returned. Specific test data, written observations on the failure and the specific corrective action you require are needed. CF12 DFI 9000 9 Rev. C May 2004