Data Acquisition. Software and Hardware. Demonstration Guide. For: NetDAQ Logger for Windows Hydra Logger for Windows Trend Link for Fluke

Similar documents
Data Acquisition Software and Hardware Demonstration Guide

Data Acquisition Software and Hardware. Demonstration Guide. For: NetDAQ Logger, Hydra Logger, and Trend Link for Fluke V3.1

Fluke 8845A/8846A Digital Multimeters. Precision and versatility for bench or systems applications

FlukeView Forms. Documenting Software. Getting Started

Fluke 8845A/8846A Digital Multimeters. Precision and versatility for bench or systems applications

2680 Series Data Acquisition Systems

Fluke Metrology Software

HVR LINE INSTALL GUIDE

2680A/2686A. Data Acquisition System/Data Logging System. Users Manual

ChartView Out-of-the-Box Setup, Acquisition, & Real-Time Display Software

ScopeMeter. Accessories. PC Software. Protective Cases. Powering and Charging

TS MANAGER OPERATIONS MANUAL

Data Acquisition. Portable, Wireless, Networked. The complete picture in data acquisition

SW860 FlukeView Software

AEMLog Users Guide. Version 1.01

DATA ARCHIVING SOFTWARE V.

Cal-Bay Systems XY Plotter, Time-Base Recorder, Automated Tester. Users Guide. Rev 3.1

CORD-XL Dual-Channel Electronic Chart Recorder User s Manual

PowerView for the EnergyPro

Docking Stations DS-U1, DS-U2, DS-U4, DS-U4-WL and DS-U Adjustment and configuration procedure for the analog inputs

DX100 DAQSTANDARD Software

SWR Director. (Version 1.11)

41126 Cognento (MODENA) Italy Via Bottego 33/A Tel: +39-(0) Internet: Fax: +39-(0)

GL900 USER S MANUAL MANUAL NO. APS(GL900-4/8)-UM-151

Logger Pro 3. Quick Reference

Lite H.264 DVR Setup Guide

4-20 ma Current Data Logger

CVI SPECTRAL PRODUCTS. 111 Highland Drive Putnam CT, (860) SM32Pro

960H H.264 DVR Setup Guide

ScopeMeter. Accessories. PC Software. Protective Cases. Powering and Charging

ScopeMeter. Accessories. PC Software. Protective Cases. Powering and Charging

Quick Start Guide. Cole-Parmer USB-based Data Acquisition Software

Standard H.264 DVR Setup Guide

WELDSEQ Plus. Operation / Installation Manual. Weld Sequence PLC Terminal Program. Computer Weld Technology, Inc.

Data Acquisition. Where you need it When you need it How you need it

WebPakCS Software Version 1.0

FlukeView Forms. Users Manual. Documenting Software

Data Acquisition. where you need it when you need it how you need it

NUBIX H.264 DVR Setup Guide

Instructions for Installing FlashUpdate and Downloading Updates for Super Buddy Satellite Meter

DATA LOGGING SOFTWARE FOR 4-WAY BALL VALVE DIGITAL MANIFOLD

PQ ANALYZER PROCUREMENT SAMPLE SPECIFICATION

9120 Redundant Probe System (RPSC)

AEMLog users guide V User Guide - Advanced Engine Management 2205 West 126 th st Hawthorne CA,

Student Quick Reference Guide

Personal Daqs USB Data Acquisition Modules Models /55 & /56

WARNING: Do not use the thermometer/data logger before you read the users manual and the following instructions.

15. Check this point when you can t communicate

ScanVision Program for Setup and Monitoring of RJS SV Type Scanner/Verifier Products P/N Rev D

ProjectorNetTM Adapter Quick Start Guide

Comgraph32. Data Retrieval and Analysis for COMBILOG. Version Page 1. Comgraph32. data retrieval and analysis for Combilog version 3

Advanced Test Equipment Rentals ATEC (2832)

Developing Networked Data Acquisition Systems with NI-DAQ

Navigator Software User s Manual. User Manual. Navigator Software. Monarch Instrument Rev 0.98 May Page 1 of 17

Docking Station DS-U4WEB with web server version 1 Instruction Manual

RS-232/422/485 to Copper or Fiber. Ethernet Converter. User s Manual

Model 620A Maintenance and Calibration

CyberComm Pro Data Acquisition Software Installation & User Guide

ENVI Classic Tutorial: A Quick Start to ENVI Classic

EQ7000. User Manual. Rev 1.00

ME 365 EXPERIMENT 3 INTRODUCTION TO LABVIEW

ST-HD-CVR4CH ST-HD-CVR8CH ST-HD-CVR16CH. Professional Digital Video Recorders QUICK START GUIDE

Lab IP Addresses and Network Communication

6220 Ethernet-Based Voltage Measurement Module

# Byrne Rd, Burnaby, BC, V5J 3J1, Canada Phone: or

HD32MTLogger Software Manual for the instrument HD32MT.1 Introduction

VisionTouch 5 Software

USER MANUAL WINATEQ300 SOFTWARE Full version 1.01

WebDAQ 316. Internet Enabled Thermocouple Logger. Remote Configuration and Monitoring. Virtually Unlimited Storage. Integrated HW and SW

DL160 Datalogger Software Guide

RC-SV Configuration Guide (Rev 4)

RKAT Audit Trail Module RK-LINK TM Software For the Radio Key 600 Series

LC1003A IP Security CCTV Connector Installation Guide

CHAPTER 6: EMAX CURRENT ANALYSIS

Lucent Technologies CAS for Windows Version 2.2 Software Installation Notes

NEC Versa 4200 Windows 98 Upgrade Information

Windows 95/98 Infoway Connection Instructions 1/4/2002; rev. 1/9/2002

HI 6500 series Quick Start Guide

OptiView Integrated Network Analyzer. Vision on every level.

SERVICE ADVANTAGE MODEL 575 MANUAL

OM-DLTH DATA LOGGER INSTRUCTIONS

RemoteWare OS/2 Client

OPERATION MANUAL. MV-410HS Layout Editor. Version higher. Command

DD-600. A member of the Data Dolphin Family of Data Logging Systems. Data Dolphin Installation and Usage Manual

Install the Software and Hardware

How to Get Started. Figure 3

TouchKit Touch Panel User manual for WindowsNT4 Version: 3.1.4

DI-154RS DI-195B. User's Manual. The way PC-based instrumentation should be. 4-Channel Data Acquisition Module

EIL3000/SVS. Supervisory Software Operating Manual. page 1

Windows Me Plug-and-Play

BCSE 1000 SOFTWARE MANUAL

HI 6600 series Quick Start Guide

Manual Supplement. Manual Title: 2680A/2686A Users Supplement Issue: 1 Print Date: June 2002 Issue Date: 9/06 Revision/Date: 1, 4/04 Page Count: 9

TCweb Data Logger. November 9, 2005 Rev D Tidal Engineering Corporation Page 1 of 30

Small rectangles (and sometimes squares like this

RK-LINK TM Version 2.x Instructions

Pre-Installation Information--S8700 IP Connect

C-322. Model. Instruction Manual. Datalogging Sound Level Meter. reedinstruments. www. com. Tripod sold separately (Model BS-6)

Tach Facts V3.0 Software

Ansur Index 2XL. Users Manual. Plug-In

Transcription:

Data Acquisition Software and Hardware Demonstration Guide For: NetDAQ Logger for Windows Hydra Logger for Windows Trend Link for Fluke

Introduction This demonstration guide contains exercises designed to quickly familiarize you with the features and functions of Fluke s software and hardware for data acquisition and analysis. NetDAQ Logger, Hydra Logger, and Trend Link for Fluke demonstrations are covered in this guide. NetDAQ and NetDAQ Logger for Windows NetDAQ Networked Data Acquisition Units transmit data over general or dedicated (isolated) networks. Each NetDAQ has 20 channels which may be programmed individually to perform virtually any signal conditioning function the user requires. Two models of NetDAQ offer the user a choice of higher speed scanning to 1000 Readings/sec., 0.01% basic Vdc accuracy, and up to 300 Volts channel to channel isolation. NetDAQ Logger for Windows is a configuration and data management program for the NetDAQ system. This program is available as a 32-bit application, supports up to 20 units (400 channels), utilizes DDE, and is available with or without Trend Link trending and data analysis capabilities. Hydra and Hydra Logger for Windows Hydra is Fluke s portable series of data acquisition products. Three models of Hydras offer the user 21 individually programmable channels of signal conditioning, a unique Universal Input Module for quick easy connections, PCMCIA memory storage, and precision rarely seen in portable data acquisition equipment. Hydra is easy to configure from the front panel while in the field, or from computer via simple RS232 connection. Hydra Logger for Windows is a configuration and data management program for the Hydra Series of products. This program is available as a 32-bit application, contains modem support for remote data communications, utilizes DDE, and is available with or without Trend Link for Fluke trending and data analysis capabilities. 2 Fluke Corporation DAQ Demonstration Guide

Trend Link for Fluke Trend Link is a comprehensive trend plotting and analysis package that plots data graphically in real time. Trend Link allows you to quickly scroll through real time and historical data, view data from all channels simultaneously, compare real time data to an historical template in batch mode, and superimpose selected channels for closer analysis. Any portion of a trend may be examined more closely by utilizing Trend Link s Zoom function. Within the confines of this demonstration guide we are able to give you only a brief introduction to some of Trend Link s basic functions. Upon finishing the structured sections in this guide we would invite you to spend some time exploring this unique software package. Demonstration Software Demonstration programs are fully functional and do not inhibit program features in any way. This allows you to explore a full-featured package to ensure it fully meets your application needs. These programs do however operate for 60 min. and then close down, requiring a restart of the program to continue. Obtaining Demonstration Software If you do not already have the software you wish to review or would like to have software sent to a colleague you may obtain it from the following sources: Contact your local Fluke office and ask that demonstration software be sent to you for evaluation. In the United States and Canada you can call 800-44-FLUKE for assistance in obtaining software. For Europe call (31 40) 2 675 200 and from other countries call 1-425-446-5500. Demonstration software and hardware may also be obtained from your local Fluke representative. World Wide Web: You may download the demonstration software of your choice by connecting to: http//www.fluke.com. Once connected, select the product information area and then select Data Acquisition Products. There you can review product specifications, application notes, and download demonstration software. Obtaining Demonstration Hardware Hydra or NetDAQ hardware may be made available through your local Fluke representative. Depending on your needs and current availability, our representatives can often accommodate you within a short time with the product you would like to review. Fluke Corporation DAQ Demonstration Guide 3

Overview Requirements: This guide covers the installation of NetDAQ Logger, Hydra Logger, and Trend Link software as well as hardware setup and configuration of NetDAQ and Hydra. To use this guide, follow the instructions for only the hardware and software you are working with, either NetDAQ or Hydra. This guide will step you through setting up connections on the Universal Input Module and show you how to verify communication settings of your NetDAQ or Hydra unit. (If you are evaluating software only you may skip the hardware section and start at the section for software installation for the package you are using.) Note that all demonstration software packages include Trend Link for Fluke so regardless of the package you are using you will need to follow the instructions for the installation of this software. The last section of this guide will step you through a brief introduction of Trend Link features and invite you to further explore Trend Link s extensive features. NetDAQ Requirements: The following equipment is required to perform the exercises in this guide when using the actual NetDAQ demonstration instrument (2640A or 2645A) supplied to you by your Fluke representative: NetDAQ Networked Data Acquisition Unit (2640A or 2645A) NetDAQ Logger for Windows demo disk set PCMCIA Ethernet adapter Ethernet cable supplied with NetDAQ (2) 50 Ohm Terminators (included with the NetDAQ unit) (2) BNC connectors ( Y or T style) Type T thermocouple (provided with the NetDAQ demo unit) You have an available PCMCIA slot on your computer (For hardware demo only) You have no network management software installed on the computer other than that supplied by your Windows operating system. 4 Fluke Corporation DAQ Demonstration Guide

Hydra Requirements: The following equipment is required to perform the exercises in this guide when using a Hydra Series instrument: Model 2620A, 2625A or 2635A Data Logger RS43 serial communication cable Type T thermocouple (provided with the Hydra demo unit) You are using COM2 (communication port 2) on your PC to communicate with the Hydra (If you are using COM1, you need to change all mentions of COM2 to COM1.) Computer Requirements: 486 PC with Windows 95, 98 or Windows NT 4 Mb RAM 5 Mb free disk space You have already installed and know how to use Microsoft Windows 95, 98 or Windows NT 4.0 A note to Windows 3.1 users. If you are using Windows 3.1 you may still review these programs. Each Fluke demonstration program automatically detects and installs the appropriate 16 or 32-bit version of software at the time of initial software installation. Please follow the special instructions for Windows 3.1 in the Software Installation sections. Software Only Operation To use this guide without hardware, begin on Step 4 (Hydra) or Step 5 (NetDAQ). Fluke Corporation DAQ Demonstration Guide 5

Hardware Setup Universal Input Module (complete this section for both NetDAQ and Hydra) NetDAQ Hardware Set-Up 1 2 Configure (Hook-up) Universal Input Module 1. Connect the type T thermocouple (provided with your Demo unit ) to Channel 1 and 2 of the Universal Input Module in parallel fashion as shown in Figure 1. 2. Close the Universal Input Module and insert it into the rear of your demo unit. This concludes the setup of the Universal Input Module. Please proceed to the Hardware set-up section for the demo unit you are using, either NetDAQ or Hydra, in the following sections. Install the PCMCIA Ethernet Adapter 1. Close all applications and turn the host computer off. 2. Plug the PCMCIA adapter into the host computer s PCMCIA slot. 3. Connect the network cable adapter to the PCMCIA adapter as shown in Figure 2. Figure 2. 4. Turn on the host computer and verify that Windows recognizes the new Ethernet adapter by viewing PC card in Windows control panel. 5. Using the supplied cable, connectors and 50 Ohm terminators, connect the Ethernet adapter to the NetDAQ as shown in Figure 3. Ethernet Coaxial Cable (50-ohm) "T" Connector BNC Connector for 10BASE2 THIN ETHERNET 50-ohm Termination Figure 1. 6 Fluke Corporation DAQ Demonstration Guide

2640ANetDAQ COMM D/O MON ENTER 50-ohm Terminator Terminator Ground Wire Hydra Hardware Setup Host Computer 1 Figure 3. Connect to Ethernet Interface Port 10BASE2 Thin Ethernet (Typical) Instrument 1 3 NetDAQ NETWORKED DATA ACQUISITION UNIT Figure 4. Reset NetDAQ to Default Communication Settings 7. Turn on power to the NetDAQ while pressing the C button on the front panel until the display reads reset (about 4 seconds) (Figure 4). REVIEW LAST MAX REM MIN AUTO SCAN MON SET FUNC Mx+B ALARM C F RO mv AC DC x1mk Ω Hz F LIMIT 1 2 This concludes the Hardware setup for the NetDAQ demo. Please proceed to the Software installation for NetDAQ Logger. HI LO OFF CAL PRN EXT CH TR COMM DIO MON ENTER CAL ENABLE Set Hydra s Communication Parameters Set the Hydra s communication parameters to: 9600 baud, no parity, echo off. From the Hydra s front panel perform the following actions: 1. Press (shift) then C to view Hydra s communication parameters. 2. Press U to select 9600 baud and press E. 3. Press U to select no parity and press E. (Note: if you are using a 2635A Hydra Data Bucket, set CTS to off.) 4. Press U to select echo off and press E. 5. Connect the RS43 cable to the serial port on the back of Hydra and to COM2 of your computer. This concludes the Hardware setup for the Hydra demo. Please proceed to the Software installation for Hydra Logger. Fluke Corporation DAQ Demonstration Guide 7

Software Installation Hydra Logger Installation NetDAQ Logger Installation 4 Hydra Logger software Installation 1. Insert disk #1. From the Start menu in Windows (or Program Manager in Windows 3.1) select Run then enter A:\Setup.exe (Figure 5). Click OK. 2. Step through the installation prompts by clicking on Next, changing the language if desired. This completes the installation of Hydra Logger. Proceed to the Trend link installation section. 5 Ethernet adapter and network address settings (Note: go directly to step 10 if you are using Windows 3.1) 1. From Windows desktop, select Start, then Settings, then Control Panel. 2. Select Network and open it. 3. If the Ethernet adapter is not listed in the Network Components window, click Add... and then select Adapter. Click Add. Figure 5. Figure 6. 8 Fluke Corporation DAQ Demonstration Guide

Trend Link 4. Select 3Com then Ethernet III LAN PC Card Adapter. Click OK. 5. When the adapter installation is complete, click Add and select Protocol then press Add. Choose Microsoft TCP/IP and press OK. 6. When the protocol installation is complete, highlight TCP/IP protocol and select Properties. 7. Select the IP Address tab. Then select Specify an IP address. 8. Enter an IP address of 198.178.246.101 and a Subnet Mask of 255.255.255.0, then press OK. 9. Press OK, then restart your computer. Install NetDAQ Logger 10. Insert disk #1. From the Start menu in Windows (or Program Manager in Windows 3.1) select Run then enter A:\Setup.exe (Figure 5). Click OK. 11.Click Next to accept the default of Isolated Network (Figure 6). 12. Step through the installation prompts by clicking on Next, changing the language if desired. 13.Accept the 3C589 Adapter if using Windows 3.1. 14. When prompted, insert disk 2 (or enter the path for the location of disk 2 files if you downloaded this program from another source) and click OK. (Windows 3.1 users should click OK when prompted for a driver disk). This completes the NetDAQ Logger installation. Proceed to the Trend Link Installation section. 6 Trend Link Installation 1. Insert disk #1. From the Start menu in Windows (or Program Manager in Windows 3.1) select Run then enter A:\Setup.exe (Figure 5). Click OK. 2. Step through the installation prompts by clicking on Next, changing language if desired. 3. Insert program disks as prompted. This completes the Trend Link installation. At this point, if you are reviewing software only, skip the following section on communication setup (sections 7 and 8) and proceed to the configure unit section (section 9). If you are using hardware proceed to the Communication Setup section for the Hydra or NetDAQ. Fluke Corporation DAQ Demonstration Guide 9

Hydra Logger NetDAQ Logger 7 Communication Setup 1. Start Hydra Logger. Click on Edit Hydra Configuration. Select Communications. 2. Click in the Baud Rate list box. Select 9600 Baud from the list. 3. Click in the Parity list box. Select No Parity from the list. 4. Click Verify, you should see the message, Connection Successful. 5. Click OK if you have a properly connected unit. 6. Click on OK several times to return to the main screen. This completes the communication setup for Hydra Logger. Proceed to the Configure Unit section. 8 Communication Setup 1. Start NetDAQ Logger. 2. Select Communication Configuration from the Setup menu. The communications configuration file dialog as shown in Figure 7 will appear. 3. If you are using a NetDAQ 2645A, go directly to #6 of this section. If you are using a NetDAQ 2640A continue. (The model number of the NetDAQ you are using is marked just above the Universal Input Connector on the back panel of the unit.) 4. Click on the Modify button in the Communications Configuration File dialog box. 5. Select 2640A NetDAQ and click OK. 6. Click Verify, you should see the message, Connection Successful. 7. Click OK if you have a properly connected unit. 8. Click on Close to return to the main screen. This completes the communication setup for the NetDAQ Logger. Proceed to Configure Unit section. 10 Fluke Corporation DAQ Demonstration Guide Figure 7.

Notice: The following steps are for both HYDRA and NetDAQ Logger software. The figures in the following sections reflect the NetDAQ Logger screen display. If you are using Hydra Logger the screen examples will look slightly different from those shown on your computer s screen, however the function selections for Hydra are the same and you should have no difficulty making the correct selection. 9 Configure Unit, Hydra and NetDAQ Configure Unit for Scanning 1. Select the Instrument Configuration command from the Setup menu. 2. Click the Channels button on the Instrument Configuration dialog box as shown in Figure 8. 3. Select channel 0101 in the Analog Channel box (ch #1 for Hydra). This represents Instrument # 01; analog channel #01. (Figure 9) 4. Click the Function button. 5. Click on the down arrow of the Function list and select Thermocouple. (Figure 10) 6. Click on the down arrow of the Range list and select T. 7. Click OK. Repeat steps 3-7 for channel 0102 (Hydra Channel #02) to set it to a type T thermocouple. 8. Click OK to accept your configurations and back up to the Instrument Configuration screen. (Figure 8) Figure 8. Figure 9. Figure 10. Fluke Corporation DAQ Demonstration Guide 11

10 Select a Data File 1. Click on the Data File button on the Instrument Configuration window. This will move you to the Data File Configuration screen as shown in figure 11. 2. Enable data file recording with the Trend Link file format. 3. Click on Select Data File. 4. Enter sim.set in the File Name window. 5. Click OK on each window until you have returned to the main wndow. 11 Save a Setup File 1. Select Save Setup As from the Setup menu. 2. Enter setup06.stp in the File Name Box as shown in figure 12. 3. Click Save. (TO SIMULATE HARDWARE: Select Logging from the main menu then select Simulate Logging on start ) 12 Start Instrument Logging Session and Open Trend Link 1.Click on the Start Instrument button on the tool bar. A Logging Status window appears showing the number of scans received. 2. Select Show Trend Link from the Plot/Trend menu (or click on the Chart button). 3. Maximize the Trend Link window. Figure 12. Figure 11. 12 Fluke Corporation DAQ Demonstration Guide

13 Review Trend Plot Values 1. Allow a few moments for the traces to move into view and then Click and hold the mouse button with the mouse pointer over the trace in the window (a vertical cursor line will appear in the window). 2. Move the mouse right or left while holding the mouse button to view the value where the cursor intersects the trace. The time each measurement was recorded is displayed at the top of the cursor. (Figure 13) 14 Change Time Frame of Plot Window 1. Double-click on the small centered rectangular window toward the bottom of the window that has the time format in it displayed as: [hh:mm:ss]. 2. Insert 2 min. (00:02:00) in the Change Time dialog box (Figure 14) then close the box by clicking OK. The total time frame you are looking at from the left to right edges of the chart is about 2 minutes. Figure 13. Figure 14. Fluke Corporation DAQ Demonstration Guide 13

15 Review Channel Trend Statistics 1. Double-click on one of the small chart Icons to the right of the chart. A dialog box similar to that shown in Figure 15 will appear. 2. Click on statistics to review information regarding your selected trace (Figure 16). 3. Click OK to return to the curve information screen. 4. Click OK to return to the Trend Link main window. Figure 15. 16 Superimpose Two Traces 1. Click and hold a numeric scale ( on the left side of the chart ) and drag this scale over another numeric scale on the chart. 2. Release the mouse button. 3. Note the two traces are now overlaid on the chart and matched in time as shown in Figure 17. Figure 16. 14 Fluke Corporation DAQ Demonstration Guide Figure 17.

17 Zoom in on the Data 1. Double-click the mouse to the left of a selected part of the data on your chart. A dashed line will appear. 2. Move your mouse to the right and double-click the mouse again. You will zoom in on the data between the two places you double clicked. You are now in historical mode (Figure 18). 3. Notice that the time in the time span box has changed and reflects a new chart edge-to-edge time frame. 4. Click on the undo icon. It s the green arrow located on the tool bar. 5. Notice that your time span changes back to your original time frame. 18 Exit Trend Link and Hydra Logger or NetDAQ Logger Programs 1. Select Close from the File menu on Trend Link s main window. 2. Select NO in the Save chart before proceeding? dialog box. You will now switch to the main window of either Hydra Logger or NetDAQ Logger. 3. Stop the instrument from the Logging menu. 4. Select Exit Application from the Setup menu. 5. Select NO in the Save Changes dialog box. Figure 18. Fluke Corporation DAQ Demonstration Guide 15

Summary This Demonstration Guide gave you a brief overview of NetDAQ Logger or Hydra Logger and Trend Link for Fluke as well as some hands on experience with hardware. Now that you have finished the step-by-step portion of this demonstration guide you may wish to explore some of the many features within these programs. Feel free to roam through the program using the intuitive menu system and tool bar buttons available to you. There is extensive on line Help available from the main menu bar. Thank you for your time and interest in our products. Fluke. Keeping your world up and running. Fluke Corporation PO Box 9090, Everett, WA USA 98206 Fluke Europe B.V. PO Box 1186, 5602 BD Eindhoven, The Netherlands For more information call: U.S.A. (800) 443-5853 or Fax (425) 446-5116 Europe/M-East/Africa (31 40) 2 675 200 or Fax (31 40) 2 675 222 Canada (800) 36-FLUKE or Fax (905) 890-6866 Other countries (425) 446-5500 or Fax (425) 446-5116 Web access: http://www.fluke.com 2000 Fluke Corporation. All rights reserved. Printed in U.S.A. 9/2000 1263520 D-ENG-N Rev B