Service Manual and Repair Guide

Similar documents
ADC7520 SERIES. 1600W Battery Chargers and Power Supplies

Abstract. GLV User Manual 1

EC-12RT Resting ECG Machine. Technical Specification

Switched Mode Power Supply Operating manual PAP800

Emerson Network Power provides customers with technical support. Users may contact the nearest Emerson local sales office or service center.

Part No. Z , IA Mar OPERATION MANUAL. High Voltage Digitalmeter A

EXPANSION PERIPHERAL CVM-R10 C. (Code ) INSTRUCTION MANUAL ( M / 00 A ) (c) CIRCUTOR S.A.

ECG MACHINES. Kizlon Medical. Website:

DRTS 33. The new generation of advanced test equipments for Relays, Energy meters, Transducers and Power quality meters

PUE C/31 measuring indicators PUE C/31H measuring indicators PUE C/31H/EX measuring indicators PUE C41H measuring indicator...

TOP - 1. Instruction Manual. Version 1.0 Produced in Jan. 2004

DRTS 64. The new generation of advanced test equipment for Relays, Energy meters, Transducers and Power quality meters.

DRTS 66. The new generation of advanced test equipments for Relays, Energy meters, Transducers and Power quality meters

LOGIC CONTROLLER INSTALLATION MANUAL NTR 756 B /E

DRTS 33. The new generation of advanced three phase relay test set

Manual Control Unit GFC 16

Part No. Z , IA Nov OPERATION MANUAL. High Voltage Digitalmeter A

4170 POS System Installation Guide

Druck DPI 620 advanced modular calibrator

Argentine Manufacturer of Hospital Equipment. Electrocardiograph Pulse Oximeters Vital Signs Monitors

DEUTSCH ENGLISH NEDERLANDS FRANÇAIS NORSK ITALANIO ČEŠTINA Hersteller DAB650SI

GRID-TIED SOLAR INVERTER 1.5KW ~ 6.0KW V.1.2

Line Interactive 1000VA/1400VA/2000VA Uninterruptible Power System

Digital Speed Controller User Manual

Plasma Panel Replacement Guide DU-42PX12X

PRO-1250D CT MID DIN rail three phase four wire energy meter.

Model RD-4 Remote Time Display

Owner's Manual. True RMS Multimeter. Model No Safety Operation Maintenance Español

Manual Control Unit GFC 32

DPP 4004 / DPP V POWER SUPPLY - RACK UNIT

PUE C/31 measuring indicators PUE C/31H measuring indicators PUE C/31H/EX measuring indicators PUE C/41H measuring indicator...

PCM SLK- Series Inverter Specification

MCX06C Programmable controller

DRTS 3 PLUS Advanced Protection Relay Test Set and Measurement System

Programmable controller Type MCX06C

LED Accu Penta 9x12W RGBWAU

MDM 011-Z1 Regen Resistor

Micro-Ohmmeter. Digital 200A. Model Measure low resistance with high accuracy

SDM120 SERIES. SDM120 series User Manual. Single phase two wire DIN rail energy meter (One module)

Always there to help you. Register your product and get support at DS3480. Question? Contact Philips.

METERSUK. SmartLink D Y N A M i T E. 3 MOD CT Series. User Manual KEY FEATURES INDEX 3. TECHNICAL DESCRIPTION.

QUALITY POWER SOLUTIONS BACKED UP BY REAL-WORLD ENGINEERING EXPERIENCE

Liebert MicroPOD 230V

isma-b-fcu FCU Hardware User Manual Global Control 5 Sp. z o.o. Warsaw, Poland

QIT600F1 USER'S GUIDE

RAD-DI8-IFS. I/O extension module with 8 digital inputs or 2 pulse inputs. INTERFACE Data sheet. 1 Description

3 analog linear/rotary transducers and 3 digital transducers inputs for travel/ speed analysis

Diamond Cabinet Slant Top Users Manual Revision 1.0

DS-1H05 Ethernet-over-Coax Extender. User Manual

USER MANUAL INSTALLATION MANUAL. Isolation transformers. ITR Isolation Transformer 7000 W 230V/32A

Chapter 2: Disassembly

PRO1 V21.xxx series MID DIN rail single phase two wire energy meter

MINI MCR-SL-UI-2I. Configurable signal duplicator. Data sheet. 1 Description

Voice Alarm System VX-3000 Series

MODEL: M5DY SEN TRONIC AG CURRENT LOOP SUPPLY. [2] POWER INPUT AC Power M: V AC (Operational voltage range V, Hz)

ECHO AIO TEST INTERFACE

Operating instructions. Switching amplifier DN0210 DN / / 2015

F1000 User's Manual. (Version: V1.01)

LXEM145 SERIES Lanx Australis single Phase KWh Meter 45Amp. User manual

Millenium II +: general characteristics

HuddlePod Air DUO Dual Wireless Audio Pods Installation and Operation Manual

3156 LEAK CURRENT HiTESTER

RAD-DO8-IFS. I/O extension module, eight digital transistor outputs. Data sheet. 1 Description

Electrocardiograph Electroencephalograph

RJ-2030/2050/2140/2150

HITACHI. EH-150 series PLC EH-RTD8 Resistance Temperature Detective input module Instruction manual. Safety precautions

MULTIPOWER MANUAL. Designed and Manufactured by ITEC Tontechnik und Industrieelektronik GesmbH 8200 Laßnitzthal 300 Austria / Europe

Programmable controller MCX06C

DRTS 66 can test all the following relays

DLA. DMX512 Analyzer. DLA Users Manual SV2_00 B.lwp copyright ELM Video Technology, Inc.

Residual overvoltage relay

Part No. Z , IA Jul OPERATION MANUAL. High Voltage Digitalmeter A

AVG-UDA2. Features. 2 simultaneous HDMI outputs. Supports Hot-plugging. Compliant with HDMI 1.4, able to transmit 1080p 3D& 4K signals

Safety Relay Unit (I/O Unit Type)

Operating instructions. Speed monitor D / / 2014

PRO1250D CT MID DIN rail three phase four wire energy meter. User manual Version 1.5

ekorrps MULTIFUNCTIONAL PROTECTION UNIT CONFIGURATION AND GENERAL CHARACTERISTICS LIB

Version /11/2015. User Manual. PowerLine AV HD Gigabit Passthrough Kit DHP-P601AV

DM-918 OPERATIONS MANUAL AUTORANGING MULTIMETER

USER MANUAL MULTI COLOR TOUCH SCREEN PAPERLESS RECORDER TPLR-96 Series

DOT MATRIX PRINTER SP6000 SERIES

Circuit Breaker Analyzer System

PI-EX-ME-2NAM/COC-120VAC

IDEAL INDUSTRIES INC. TECHNICAL MANUAL MODEL:

Model A Mini AC/DC Clamp Meter. User's Guide

FOR VISUAL CARD READER / WRITER. Installation Guide & Specifications

Digital Multimeter User's Manual

HI-POT TESTER User Manual

THE NEW GENERATION OF ADVANCED TEST EQUIPMENTS FOR RELAYS, ENERGY METERS, TRANSDUCERS AND POWER QUALITY METERS.

Manual Simatek Control Unit GFC 32

Voice Alarm System VX-3000 Series

PI-EX-ME-2NAM/COC-24VDC

Model HM-535 Power Supply Installation and Service Instructions

Electronic Balance Ionizer STABLO-AP Service Manual

Axiom Manufacturing. Users Manual. for PROJECT DEVELOPMENT BOARD AXM xiom anufacturing

1-36V, 0-3A DC Power Supply

Model Number Structure

Voice Alarm System VX-3000 Series

Inform IT Single Pen Circular Chart Recorder

POWER AVAILABILITY 2U POD USER MANUAL. Power Output Distribution 230 Volt

Transcription:

M - T R A C E E C G A P P L I A N C E Service Manual and Repair Guide E D I T I O N 1. 0 2 M4Medical Sp. z o. o. 10/7 Ogrodowa Street Lublin, PL 20-075 Phone/Fax +48 12 285 3755 Cell +48 501 407 094 E-mail: biuro@m4medical.com Revisions: Pre-release edition

General information Precautions, Technical Specification, Tools and Parts To keep the maintenance and repair process correct and faultless it is essential to meet the information enclosed in this manual and follow them during everything you do with the ECG appliance, before any action or activity is taken. Precautions and Safety Familiarizing with the safety basics is the most vital to avoid any disturbing and unexpected cases during service maintenance process of the ECG appliance. It also has a purpose to keep the technical personnel away from any danger and unnecessary complications. To make it easy, read and follow the information below: First of all, one is to ensure that work stand is clean, well ventilated, and lamp-lit properly. All the operations should be performed by personnel, skilled at electronics and electricity basics, equipped with necessary tools. Before performing any maintenance, servicing or repairing actions personnel should be trained in by manufacturer according to the servicing duties regarding the M-TRACE appliance. Hygiene precautions are vital. Handling the product, keep your hands and working surfaces clean of any dirt, dust, debris and hand oil. Before removing the external cover, disassembling or making any mechanical maintenance disconnect the appliance from external power source by turning off the main switch on the back of it located just by the power socket and unplugging main power cord from it. If any specific operation requires the appliance to be connected to mains please DO FOLLOW STRICTLY THE INSTRUCTIONS INCLUDED IN THE RELAYING PARAGRAPH OF THIS MANUAL. To get any necessary, and not included in this publication user's maintenance instructions, please refer to most-up-to-date user's manual. To get the latest information contact the manufacturer as well. 2

Tools M-TRACE is characterized by simple, modular construction, however upon maintaining any service process some certain tools might be found helpful. Here s the list with tools which might be required to perform any maintenance described in this manual: PH1 (Phillips type) magnetic-tip screwdriver Square-tip screwdriver Pliers Tweezers Pen-drive Multimeter 3

Technical specification Appliance description M-TRACE electrocardiograph is a sophisticated and modern electronic device. It is dedicated to record ECG impulses in full range of 12 standard channels. Printout is made on thermal paper. The device is equipped with high resolution linear thermal printing unit and color LCD. The built-in battery allows to use it quickly wherever it is necessary. Aesthetic plastic cover together with membrane keyboard makes the device easy to keep clean. Specification Dimensions (W x H x D) Weight Power supply Built-in battery Power consumption ECG leads 60 W x 52 H x 220 D mm <1,5 kg AC 90-240 V @50-60 Hz Li-ion 7,2 V 2,2 Ah Can be replaced only by qualified servicing personnel <35 VA Sensitivity 2,5/5/10/20 mm/mv 5% Recording speed 5/10/25/50 mm/s 5% Common Mode Rejection Ratio Frequency band Input impedance Control range 12 Standard ECG leads: Einthoven s limb leads I, II, III Goldberger s limb leads avr, avl, avf Wilson s precordial leads V1, V2, V3, V4, V5, V6 >100 db 0,05-150 Hz >10 MΩ >300 mvpp 10 mvpp Resolution 2,5 µv Sampling frequency Digital filters LCD screen Safety Class / Group Operating environment Conditions 1000 Hz 50Hz, 60Hz, 35Hz, 25Hz, anti-drift Color graphic display, 320x240 Protection Type CF (EN60601-1) Class I Protector Class II (EN60601-1) Class A / Group 1 (CISPR-11) Temperature: +10 to +40 C (+50 to +104 F) Relative Humidity: 25 to 95% (non-condensing) 4

M-TRACE SERVICE MANUAL (1.0) M-TRACE unit top view 5

M-TRACE unit's structure Both external (power grid) and internal (built-in battery) sources are used to power the appliance. Battery charger is integrated with the mainboard. It is vital to notice the right internal battery type during its replacing. Using any but the right one can cause bad damage done by explosion being natural behaviour for Li-ion type cells that are wrongly matched or not properly installed. Li-ion type batteries are highly fragile to overcharging so be sure the required parameters of the mounted one are strictly met. The built-in power supply unit delivers DC 9 V to its output connector. It is wired to mainboard where the proper elements produce all the other voltage levels required by the M-TRACE to be fully operational. There are four following major voltage converters responsible for the above: Voltage level Powered sections DC 3,3 V DC 5 V DV 1,5 V DC 10,4 V Directly from power supply unit output or from internal battery Most of the digital elements Printing thermal head USB controller FPGA on-board programmable logic LCD backlit Thermal printer's motor drive Analog amplifier's board ECG signals, collected from patient, after being amplified by amplifiers and AD converted on analog board, are passed using SPI opto-isolated (which is eligible and essential to separate patient from the device's circuits) transmission to mainboard. Mainboard's duties are to keep all the internal modules communication running properly, analyse and show all the computed data and interact with the user and patient. None of the components requires any adjustments or tuning. All of them are ready instantly to serve as properly assembled and powered. Parts list No. Part Symbol 1. Power supply unit MBU30-104 2. Analog amplifiers module X300-20-xx 3. Mainboard X300-10-xx 4. Thermal printer LTPV-445 5. Battery CGR18650CF-2S1P-PCB 7,2 V 2,2 Ah 6. LCD screen AM-320240N1TMQW-30H 7. Keyboard 8. Power socket 9. Casing 6

Power supply unit MBU30-104 P1 connector: AC 90-240 V input 1 L Power line L (live) 2 N Power line N (neutral) P2 connector: DC 9 V output 1 +9 V +U power supply 2 +9 V +U power supply 3 +9 V +U power supply 4 GND 5 GND 6 GND 7

Analog amplifiers module - X300-20-xx J1 DA15 type connector patient's ECG signal cable 1 V2 Precordial V2 lead 2 V3 Precordial V3 lead 3 V4 Precordial V4 lead 4 V5 Precordial V5 lead 5 V6 Precordial V6 lead 6 GND Ground 7 8 9 R Right upper limb (arm) lead 10 L Left upper limb (arm) lead 11 F Left lower limb (leg) lead 12 V1 Precordial V1 lead 13 14 N Right lower limb (leg) lead 15 8

J2 Molex type connector processor programming lines J3 Molex type connector wiring with X300-10-xx mainboard 1 SCLK Clock output 2 GND Ground 3 MOSI Data output 4 GND Ground 5 MISO Data input 6 ENABLE Data transmission enabling signal 7 POWEROFF Analog module on/off switching line 8 3V3 Opto-isolators power supply 9 GND Ground 10 +U power supply Voltage converter power supply 9

Mainboard X300-10-xx J1 connector J2 connector J3 connector J4 connector J5 connector J6 connector J7 connector J8 connector J9 connector J10 connector J11 connector J12 connector Thermal printer's unit motor drive control line Thermal printer's unit heating head's control line Not connected for future use USB host External PCL5/6 printer output FPGA circuit programming lines USB slave M-TRACE-PC data transmission Keyboard LCD data signal wiring with analog amplifiers module X300-20-xx (J3) USB host programming lines Internal battery socket 1 TERM Thermistor protector 2 GND Ground 3 +U battery Battery power supply +7,2 V Power supply unit socket 1 L Power line L (live) 2 N Power line N (neutral) J13 connector ARM CPU programming lines J14 connector LCD backlit power supply line 10

Electrocardiograph appliance meets all the required following standards: EN 60601-1, EN 60601-2-25, EN 60061-2-51. DO NOT test an insulation resistance with B-d (EN 60601-1) procedures it can cause appliance damage. To perform insulation resistance, tests should be checked as follows: External power supply SIP/SOP section with AC 4000 V charge 60 seconds duration, Application circuits SIP/SOP section with AC 2500 V charge 60 seconds duration. Analog amplifiers frequency band check should be performed with isoline filters turned off (refer to user's manual). ------------------------------------------------------------------------------------------------------------------------ TROUBLESHOOTING For usage Lack of paper Paper cover LED diode on the keyboard doesn t light ECG paper is out. Put new roll of paper. Follow the instruction Paper loading in this manual. The paper cover is not closed properly, Open the cover and close it again properly. The accumulator is not charging. 1. Check switch on/off button (1/0) at the back side of the unit. It should be switchd on in the position 1. 2. Check if there is voltage in the power sypply socket. For service Lack of ECG recording LCD disturbances The unit doesn t work as described in the manual The unit doesn t charge The unit doesn t print on the printer printer breaks the printouts Check the patient cable Check PC board X300-20-xx Check PC board X300-10-xx Check PC board X300-10-xx Check PC board MBU-104 Check ECG paper Check PC board X300-10-xx Check printer LTPV-445 11