Unique Features of the GE Senoclaire Tomosynthesis System. Tyler Fisher, M.S., DABR Therapy Physics, Inc.

Similar documents
Certificate Extension II 8/1/2016. Certification Extension Process for Digital Breast Tomosynthesis and Medical Physicists Role

RECENTLY APPROVED GE FFDM DBT TESTING PROCEDURES.

8/2/2016. Hands-on GE SenoClaire DBT: technical characteristics & quality control. Disclosures and Acknowledgements

Lorad FFDM QC Procedures for Medical Physicist. Tao Wu, Ph.D. Hologic, Inc.

ACMP 25th Annual Meeting

4/19/2016. Deborah Thames R.T. (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT

Z-MOTION. Universal Digital Radiographic System Z-MOTION. Control-X Medical CONTROL-X MEDICAL

10/9/2018. Deborah Thames BSRS RT (R)(M)(QM) Theory & Technology and advancement in 3D imaging DBT

Assessment of 3D performance metrics. X-ray based Volumetric imaging systems: Fourier-based imaging metrics. The MTF in CT

Acknowledgments and financial disclosure

Design Considerations in Optimizing a Breast Tomosynthesis System

Design and performance characteristics of a Cone Beam CT system for Leksell Gamma Knife Icon

MEDICAL EQUIPMENT: COMPUTED TOMOGRAPHY. Prof. Yasser Mostafa Kadah

Quality control phantoms and protocol for a tomography system

Fujifilm DR Solution. FDR AcSelerate. The new pinnacle in diagnostic imaging from Fujifilm ISS. CsI. Dynamic Visualization. Technology.

연구용유방초음파질관리 원광대학병원김혜원

ENGLISH. Digital Breast Tomosynthesis

Digital Breast Tomosynthesis: The Fundamentals

CT Protocol Review: Practical Tips for the Imaging Physicist Physicist

Optimisation of Toshiba Aquilion ONE Volume Imaging

DISC QA Mamchex Tool Use Instructions with CARESTREAM DIRECTVIEW CR Mammography Systems

SUMMARY OF DENTAL MEASUREMENT PROCEDURES (Abridged Protocol)

3/27/2012 WHY SPECT / CT? SPECT / CT Basic Principles. Advantages of SPECT. Advantages of CT. Dr John C. Dickson, Principal Physicist UCLH

AAPM Standard of Practice: CT Protocol Review Physicist

Fits you like no other

Chapter 1 General Information

AIDR 3D Iterative Reconstruction:

DX-D 300. Flexible Direct Radiography System

DX-D 100 with wireless detector

product data Raybow dr mobile digital radiographic system with wireless detector product data

Fits you like no other

Digital Scatter Removal in Mammography to enable Patient Dose Reduction

DX-D 300 FLEXIBLE DIRECT RADIOGRAPHY SYSTEM DX-D 300 DR SYSTEM

Planmed Sophie Classic Mobile. English. dedicated for mobile mammography

Background 8/2/2011. Development of Breast Models for Use in Simulation of Breast Tomosynthesis and CT Breast Imaging. Stephen J.

CT NOISE POWER SPECTRUM FOR FILTERED BACKPROJECTION AND ITERATIVE RECONSTRUCTION

DX-D 100 with WITH ITS EXCELLENT IMAGE QUALITY AND

O-ARM IMAGING SYSTEM TECHNICAL SPECIFICATION GUIDE

RaySafe Solo. Specifications

Digital Imaging and Communications in Medicine (DICOM) Supplement 116: 3D X-Ray Storage SOP Class

Design and development of a phantom for tomosynthesis with potential for automated analysis via the cloud

Synthesized 2D Mammographic Imaging

FLEXIBLE DIRECT RADIOGRAPHY SYSTEM DX-D 300

ImPACT. Information Leaflet No. 1: CT Scanner Acceptance Testing

A comparative study of limited-angle cone-beam reconstruction methods for breast tomosynthesis

Product data. Mammographic System

DX-D 300. Flexible Direct Radiography System

Spiral CT. Protocol Optimization & Quality Assurance. Ge Wang, Ph.D. Department of Radiology University of Iowa Iowa City, Iowa 52242, USA

Solid Capabilities Are Built Into the Supria Plus. Putting You On The Path of High Quality, Cost-Effective CT Scanning

Photon beam dose distributions in 2D

Monte Carlo simulation of scatter field for calculation of contrast of discs in synthetic CDMAM images

Simulation of Mammograms & Tomosynthesis imaging with Cone Beam Breast CT images

Digital breast tomosynthesis: comparison of different methods to calculate patient doses

DX-D 100 WITH WIRELESS DETECTOR

Background. Outline. Radiographic Tomosynthesis: Image Quality and Artifacts Reduction 1 / GE /

Dose Distributions. Purpose. Isodose distributions. To familiarize the resident with dose distributions and the factors that affect them

Micro-CT Methodology Hasan Alsaid, PhD

The digital EVOlution in cassette size format

Optimization of CT Simulation Imaging. Ingrid Reiser Dept. of Radiology The University of Chicago

Philips MicroDose L30

Power Spectrum Analysis of an Anthropomorphic Breast Phantom Compared to Patient Data in 2D Digital Mammography and Breast Tomosynthesis

DX-D WITH DX-D 40/45 DETECTOR FAMILY MOBILE DR SOLUTION

DX-D 300 FLEXIBLE DIRECT RADIOGRAPHY SYSTEM DX-D 300

Limits of Dose Reduction in CT: How low is too low? Acknowledgement. Disclosures 8/2/2012

CLASS HOURS: 4 CREDIT HOURS: 4 LABORATORY HOURS: 0

Detection of microcalcification clusters by 2D-mammography and narrow and wide angle digital breast tomosynthesis

xorantech.com Suite of DR Products

Scatter Correction for Dual source Cone beam CT Using the Pre patient Grid. Yingxuan Chen. Graduate Program in Medical Physics Duke University

C-View 1.0. DICOM Conformance Statement. MAN Rev 001

Conflicts of Interest Nuclear Medicine and PET physics reviewer for the ACR Accreditation program

Slide 1. Technical Aspects of Quality Control in Magnetic Resonance Imaging. Slide 2. Annual Compliance Testing. of MRI Systems.

PRODUCT DATA. Advanced 128

Preparing for Part 2 of the ABR Diagnostic Physics Exam

(restyle 2011) Anatomic programs

S. Guru Prasad, Ph.D., DABR

8/3/2011. Implementation Status. Yes (EI and DI) On all new equipment Legacy systems can be upgraded (digitizer (plate reader) upgrade not required).

INTERNATIONAL STANDARD

LAMBDA DATA SHEET. ANALOGIC mammography unit. ANALOGIC mammography unit. Rev. n.05 of 15/01/13

Radiographic Systems

James R Halama, PhD Loyola University Medical Center

Nicholas Marshall Department of Radiology, University Hospitals Leuven, Herestraat 49, 3000 Leuven, Belgium

8/2/2016. Measures the degradation/distortion of the acquired image (relative to an ideal image) using a quantitative figure-of-merit

Philips SPECT/CT Systems

VIVIX Software.

Introduction to Biomedical Imaging

Developments in Dimensional Metrology in X-ray Computed Tomography at NPL

COMPREHENSIVE QUALITY CONTROL OF NMR TOMOGRAPHY USING 3D PRINTED PHANTOM

Automated Image Analysis Software for Quality Assurance of a Radiotherapy CT Simulator

Evaluation of AutoQA Lite TM Image Quality Measurement Software

5020S CR READER. Software Installation Guide & Calibration Process. Ver.: Document Part Number: CR /170208

Lucy Phantom MR Grid Evaluation

RITtrend allows you to effectively manage all your physics QA data in one powerful and customizable package.

Bucky to DR in one smart step. Philips ProGrade DR solution specifications

Future Topics. Projection Imaging Dose Reporting, XA 3D Volume Objects. for DICOM WG-02. Presented by Heinz Blendinger, Siemens Medical Solutions

INSTRUCTIONS RPD INFORMATION RPD PRODUCT INFORMATION. Address 5218 Barthel Industrial Drive Albertville, MN Website

Shadow casting. What is the problem? Cone Beam Computed Tomography THE OBJECTIVES OF DIAGNOSTIC IMAGING IDEAL DIAGNOSTIC IMAGING STUDY LIMITATIONS

Photon counting spectral CT versus conventional CT: comparative evaluation for breast imaging application

DICOM Conformance Statement for Selenia Dimensions Acquisition Workstation Software Version 1.8 Part Number MAN Revision 002

C a t p h a n / T h e P h a n t o m L a b o r a t o r y

GPU implementation for rapid iterative image reconstruction algorithm

Transcription:

Unique Features of the GE Senoclaire Tomosynthesis System Tyler Fisher, M.S., DABR Therapy Physics, Inc.

Conflict of Interest Disclosure I have no conflicts to disclose.

Learning Objectives Overview of GE SenoClaire System Medical Physicist QC Testing Requirements Common Pit Falls, Errors, and Issues Technologist QC Testing Requirements Review the FDA DBT Certificate Extension Program Note: This presentation assumes the physicist is proficient with 2D GE Physics Testing Procedures

SenoClaire System Summary Uses same Target/Filter combinations as 2D system. Similar Dose level in 3D as 2D FDA Approved for Screening as 2D CC and 3D MLO views. An exam consisting of one DBT MLO view plus one 2D CC view is clinically noninferior to a standard two view digital mammography examination as measured by the area under the receiver operating characteristic (ROC) curve 1 Available as an add on for all Senographe Essential systems currently installed. V Preview reconstruction option synthesized 2D image from 3D acquisition 1. SenoClaire GE Breast Tomosynthesis Clinical and Non Clinical Information, 5415896 3 1EN, Rev. 1, Page 7

SenoClaire Attachment 3 Available Paddles 24 x 31, Elevated 24 x 31, Sliding 19 x 23 9 images acquired over a 25 0 sweep. Step and Shoot Acquisition 100 micron pixel size no binning of pixels 5:1 Anti Scatter grid MTD attachment weighs over 12 kg and is not counterbalanced use caution if manually moving. Not available for Magnification, CESM, or negative angle exposures.

Installing the MTD

Caution! When installing and removing the MTD, proper care must be taken not to scratch, dent, or damage BOTH the Digital Detector and the MTD grid. The Cart that holds the MTD when not in use is helpful for initially placing the weight of the MTD on the Digital Detector, but should be removed before final alignment. Especially when removing the MTD, make sure that the MTD is clear of the guiderails before lowering the gantry out of the way. It is possible to shear the guiderails if the MTD is attached to the cart and the gantry is lowered prematurely.

Performing a 3D Acquisition Initialize 3D mode by pressing motion selection button on MTD until 3D light is illuminated. Move Tube Head into start position using the Gantry Arm Movement Control Buttons Select Right or Left Laterality at Control Console From patient s perspective, Gantry starting to the left must be a Right Laterality acquisition. Gantry starting on the right must be a Left Laterality acquisition. Press Foot Pedal, Prep, and Expose Buttons. Following completion of 3D acquisition, gantry is not ready for a second. Must manually move gantry into position again.

Image Processing Images are processed at Dr s Review Workstation using GE s Adaptive Statistical Iterative Reconstruction (ASiR) system. Images should be Automatically Pushed to Review Workstation for Reconstruction. At acceptance, remind service that this must be functional for you to complete the testing. Both slabs (10 mm) and planes (0.5 mm or 1 mm) are reconstructed. Technologist Control Station only sees 9 acquired images to review positioning. For submission to FDA, must submit a slice of the phantom showing best score (usually at 37 mm or 38 mm) from the Review Workstation. Image must have identifying data on it (kvp, ma, Unit Name, Type of Acquisition) Important to remind service that you will need a way to print or burn the image from the Review Workstation.

Medical Physicist QC Tests QAP (Quality Assurance Procedures) CNR and MTF Measurement with MTD Flat Field 3D Test Grid Texture Test AOP 2D and SNR Check with MTD AOP 3D Check Non QAP Phantom IQ Test with MTD Phantom IQ 3D Test Compression Paddle Border to Chest Wall Alignment with MTD Breast Entrance Exposure and Average Glandular Dose with MTD Breast Entrance Exposure and Average Glandular Dose in 3D Mode Artifact Evaluation and Flat Field Uniformity with MTD Volume Coverage

Many of the QAP tests are nearly identical to 2D tests, but with MTD system in place. Browser QAP Button MTD Flat Field 3D Test CNR and MTF Measurement with MTD 2D AOP and SNR Check with MTD

Grid Texture Test Objective: To measure the amount of grid texture in 2D Images. Necessary due to differences in grid texture of MTD device compared to gain calibration of traditional system. Uses 25 mm thick flat field phantom. Acquire 10 images. Record Result. Action Limit: Texture Level must not exceed 0.002. If this fails, try re installing the MTD and performing the test again.

Grid Texture Test Measured 0.0164 8 times the allowable 0.002 Obvious gridlines in flat field image Corrected by service with new Gain Calibration In this case, I did not allow facility to perform 3D imaging until issue was corrected.

AOP 3D Check Verifies that Correct Parameters are Selected in AOP 3D Mode Uses 25 mm, 50 mm, and 60 mm acrylic plates. No SNR Measurement as with 2D. Acrylic Thickness (mm) Exposure Parameters For AOP 3D Mode Only Track/Filter mas kv 25 Mo/Mo or Mo/Rh 20 70 26 50 Rh/Rh 40 90 29 60 Rh/Rh 50 120 30 or 31

Phantom IQ Test with MTD Identical to 2D Phantom IQ Test but with MTD system in place. Rh/Rh, 29 kvp, 56 mas Action Limit: 4 fibers, 3 speck groups, 3 masses.

Phantom IQ 3D Test Set system to acquire 3D acquisition Manual Technique: Rh/Rh, 29 kvp, 56 mas Review Reconstructed Image on Dr. s Review Workstation View both slabs and planes. Select plane image with best score for submission to FDA Action Limit: 4 fibers, 3 speck groups, 3 masses.

Compression Paddle Border to Chest Wall Alignment Test with MTD Objective: To assure the paddle chest wall side border aligns with the chest wall side of the detector. Both Mo and Rh targets must be tested. Must perform test with all 3 paddles (if used clinically) 24 x 31 paddle Elevated 24 x 31 paddle Sliding 19 x 23 paddle FDA requires that you follow the test procedure as specified in the QC manual or show that your procedure is not significantly different. Show your work

Compression Paddle Border to Chest Wall Alignment Test with MTD Coin must be as close as possible, but inside of the tangent to the inner vertical surface of the compression paddle. Recommend taping coin on top of compression paddle as close to chest wall edge as possible. Without tape, coin will slide away from chest wall edge.

Compression Paddle Border to Chest Wall Alignment Test with MTD Set the paddle to approximately 4.2 cm above the breast support. There are 2 possible geometric cases for the coin:

Compression Paddle Border to Chest Wall Alignment Test with MTD Use segment tool to measure Full Diameter (Wd) & Partial Diameter (Pd) Note the segment tool measures in a plane 2 cm above breast support Calculate Alignment Deviation Z d = (Wd Pd) Calculate Zd = Z d * 1.063 Action Limit: Zd must be < 6.6 mm. Enter Results in Table

Breast Entrance Exposure and Average Glandular Dose with MTD Measure in AOP mode (STD, CNT, & Dose) Rotate Phantom 180 0 from Normal Orientation. **Verify CNT Dose Setting with MTD**

3D Breast Entrance Exposure and Average Glandular Dose Perform a 3D Stationary Acquisition AOP Mode or Manual Mode @ Rh/Rh, 29 kvp, 45 mas Rotate Phantom 180 0 from Normal Orientation.

Artifact Evaluation and Flat Field Uniformity with MTD Objective: To assess the degree and source of artifacts and to assure that the Flat Field image is Uniform. 2D Acquisition in Mo/Mo, Mo/Rh, and Rh/Rh modes. Review Raw Images from Browser There is a separate Gain Calibration for the MTD to create uniform images.

Volume Coverage Objective: To ensure that the entire imaged object is reconstructed on the Z Axis. 3D Acquisition with 25 mm & 60 mm phantoms with aluminum sheets above and below. Manual Exposure Technique: 25 mm: Mo/Mo, 26 kvp, 45 mas 60 mm: Rh/Rh, 29, kvp, 110 mas Visually Verify that the top and bottom Al sheets are visible in the reconstructed volume.

Volume Coverage

Volume Coverage

Summary of Physics Tests Test Pass/Fail Criteria Timeframe to Correct Phantom IQ Test with MTD 4 Fibers, 3 Speck Groups, 3 Masses Immediate CNR and MTF Measurement with MTD MTF at 2 lp/mm > 49% MTF at 4 lp/mm > 18% Change in CNR < 0.20 Immediate Flat Field 3D Test Brightness Non Uniformity and SNR Non Uniformity Must Pass Immediate Phantom IQ 3D Test 4 Fibers, 3 Speck Groups, 3 Masses Immediate Grid Texture Test Texture Level < 0.002 30 Days (Use Discretion) AOP 2D and SNR Check with MTD Exposure Parameters within Limits SNR > 50 Immediate AOP 3D Check Exposure Parameters within Limits Immediate

Summary of Physics Tests Test Pass/Fail Criteria Timeframe to Correct Compression Paddle Border to Chest Wall Alignment with MTD Chest Wall edge of Paddle Not in Image Chest Wall edge < 6.6 mm beyond detector 30 Days Breast Entrance Exposure and Average Glandular Dose with MTD 3 mgy (300 mrad) Immediate Breast Entrance Exposure and Average Glandular Dose in 3D Mode 3 mgy (300 mrad) Immediate Artifact Evaluation and Flat Field Uniformity with MTD No Artifact or Non Uniformity that is expected to mimic or obscure clinical information 30 Days (Use Discretion) Volume Coverage Focal Plane of Top and Bottom Al sheets visible within the reconstructed volume Immediate

Technologist QC Tests Minimum Frequency Weekly Monthly Semi Annually Procedure Phantom IQ Test with MTD CNR and MTF Measurement with MTD Flat Field 3D Test Phantom IQ 3D Test Grid Texture Test AOP 2D and SNR Check with MTD AOP 3D Check Visual Checklist Compression Force Test All 2D QC TESTS MUST STILL BE PERFORMED! These tests are in addition to the routine GE QC

FDA DBT Certificate Extension Program Under MQSA each Manufacturer s DBT system is considered a new modality, and the personnel training requirements apply. The certificate extension program only applies to the DBT portion of the system. The 2D portion of the system must be accredited by one of the approved accreditation bodies. The facility must have the approval of the FDA prior to DBT clinical use.

FDA DBT Application Process Available at: http://www.fda.gov/radiation EmittingProducts/MammographyQuality StandardsActandProgram/FacilityCertificationandInspection/ucm413117.htm Includes Facility Information, Unit ID, Image Receptor ID, Monitor ID, Personnel Qualifications. 3D Phantom Image (Hardcopy or Softcopy) Include the Complete, Detailed Mammography Equipment Evaluation Include an Evaluation of the DBT Manufacturer s QC Program

FDA DBT Certificate Extension Submission Send All Application Materials to: FFDM and DBT Certification Extension Program Division of Mammography Quality Standards FDA/CDRH/OIR 10903 New Hampshire Ave., WO66 4528 Silver Spring, MD 20993 0002 Phone: 301 796 5919 Fax: 301 847 8502

In order to acquire the DBT image set, the GE system uses views over sweep angle of degrees? 20% 20% 20% 20% 20% 1. 15 views, 25 degrees 2. 9 views, 25 degrees 3. 15 views, 15 degrees 4. 12 views, 15 degrees 5. 9 views, 15 degrees 10

In order to acquire the DBT image set, the GE system uses views over sweep angle of degrees? 1. 15 views, 25 degrees 2. 9 views, 25 degrees 3. 15 views, 15 degrees 4. 12 views, 15 degrees 5. 9 views, 15 degrees Reference: GE Breast Tomosynthesis Operator Manual, 5415894 4 1EN, Revision 1. Page #24.

What is the minimum acceptable phantom score for the 3D image? 20% 20% 20% 20% 20% 1. 5 Fibers, 4 Speck Groups, 4 Masses 2. 4 Fibers, 4 Speck Groups, 4 Masses 3. 4 Fibers, 4 Speck Groups, 3 Masses 4. 4 Fibers, 3 Speck Groups, 3 Masses 5. 3 Fibers, 3 Speck Groups, 3 Masses 10

What is minimum acceptable phantom score for the 3D image? 1. 5 Fibers, 4 Speck Groups, 4 Masses 2. 4 Fibers, 4 Speck Groups, 4 Masses 3. 4 Fibers, 4 Speck Groups, 3 Masses 4. 4 Fibers, 3 Speck Groups, 3 Masses 5. 3 Fibers, 3 Speck Groups, 3 Masses Reference: GE Breast Tomosynthesis Quality Control Manual, 5415892 3 1EN, Revision 1. Page #21.

What is the Grid Ratio for the Anti Scatter grid? 20% 20% 20% 20% 20% 1. 2:1 2. 4:1 3. 5:1 4. 8:1 5. 12:1 10

What is the Grid Ratio for the Anti Scatter grid? 1. 2:1 2. 4:1 3. 5:1 4. 8:1 5. 12:1 Reference: GE Breast Tomosynthesis Operator Manual, 5415894 4 1EN, Revision 1. Page #198.