Three-Dimensional Laser Scanner. Field Evaluation Specifications

Similar documents
INVESTIGATION ON LASER SCANNERS *

SURVEY AND ALIGNMENT OVERVIEW: FERMILAB MAIN INJECTOR RING

COORDINATE MEASUREMENT MACHINES AS AN ALIGNMENT TOOL *

HIGH PRECISION SURVEY AND ALIGNMENT OF LARGE LINEAR COLLIDERS - HORIZONTAL ALIGNMENT -

12 Alignment. Technical Synopsis

Gregory Walsh, Ph.D. San Ramon, CA January 25, 2011

THE ALIGNMENT OF THE ADVANCED PHOTON SOURCE AT ARGONNE NATIONAL LABORATORY*

Exterior Orientation Parameters

Alignment with the help of the alignment service

Chapters 1 7: Overview

LCLS Undulator Quadrupole Fiducialization Plan

1.0 Introduction. 1.1 General

Displayed with permission The American Surveyor April/May Copyright 2009 Cheves Media

Boresight alignment method for mobile laser scanning systems

User s guide. November LSE S.r.l. All rights reserved

REPRESENTATION REQUIREMENTS OF AS-IS BUILDING INFORMATION MODELS GENERATED FROM LASER SCANNED POINT CLOUD DATA

Insertion Device Alignment for the Diamond Light Source

Terrestrial Laser Scanning assignment

STRAIGHT LINE REFERENCE SYSTEM STATUS REPORT ON POISSON SYSTEM CALIBRATION

Forward Time-of-Flight Geometry for CLAS12

FILTER PACKAGE USER S GUIDE

Ch 22 Inspection Technologies

Aligning Parallelism of Rolls Has Never Been Faster. L-732 Precision Dual-Scan Roll Alignment System

V-STARS Measurement Report

RAILWAY PROJECT Geodetic Reference System, Geodetic Control Network and Rail Construction Measurements

SUMMARY. Page INTERIOR SURVEY - SAN FRANCESCO CHURCH (TERNI) EXTERIOR SURVEY - UFFIZI & PIAZZA DUOMO TUNNEL SURVEY TERNI

Robust Automatic 3D Point Cloud Registration and Object Detection

AN EVALUATION OF NON-CONTACT LASER SCANNER PERFORMANCE USING CALIBRATED ARTIFACTS

EDUCATIONAL SPECTROPHOTOMETER ACCESSORY KIT AND EDUCATIONAL SPECTROPHOTOMETER SYSTEM

Status report on the Survey and Alignment of the Accelerators at CERN

2. INITIAL ALIGNMENT 2.1 Tolerances The FFTB is composed of four straight sections as shown in fig. 1. FFTB composed of 4 Straight Sections

TLS Parameters, Workflows and Field Methods

ASSETS DATA INVENTORY BASED ON BUILDING INFORMATION MODELLING

PARCC Geometry Practice Test Released April,

1. Safety instructions

High Precision Optical Instrumentation for Large Structure Position Monitoring: The BCAM System Applied to the CMS Magnet

Calibration of IRS-1C PAN-camera

Module 7 Defining Coordinate Systems

FARO Scanning Plugin

Multi-Hole Velocity Probes

Stand-alone manual for the RasCLiC set-up in TT1

cse 252c Fall 2004 Project Report: A Model of Perpendicular Texture for Determining Surface Geometry

ADS40 Calibration & Verification Process. Udo Tempelmann*, Ludger Hinsken**, Utz Recke*

Integrating the Generations, FIG Working Week 2008,Stockholm, Sweden June 2008

Assignment 1 Due September 6, 2011

INTRODUCTION TO MEDICAL IMAGING- 3D LOCALIZATION LAB MANUAL 1. Modifications for P551 Fall 2013 Medical Physics Laboratory

TERRESTRIAL LASER SCANNING AND APPLICATION IN GEODETIC ENGINEERING

Homework Set 3 Due Thursday, 07/14

Institute of Mechatronics and Information Systems

Leica Cyclone 5.4 Technical Specifications

AUTOMATIC TARGET IDENTIFICATION FOR LASER SCANNERS

CE 260 SURVEYING CHAPTER 5 THEODOLITES

INSTRUCTIONS RPD INFORMATION RPD PRODUCT INFORMATION. Address 5218 Barthel Industrial Drive Albertville, MN Website

A RADIAL WHITE LIGHT INTERFEROMETER FOR MEASUREMENT OF CYLINDRICAL GEOMETRIES

Advanced Vision Guided Robotics. David Bruce Engineering Manager FANUC America Corporation

PHYSICS. Chapter 34 Lecture FOR SCIENTISTS AND ENGINEERS A STRATEGIC APPROACH 4/E RANDALL D. KNIGHT

Dedicated Software Algorithms for 3D Clouds of Points

USING A LASER SCANNER FOR THE CONTROL OF ACCELERATOR INFRASTRUCTURE DURING THE MACHINE INTEGRATION

PCMM System Specifications Leica Absolute Tracker and Leica T-Products

Unwrapping of Urban Surface Models

PCMM System Specifications Leica Absolute Tracker and Leica T-Products

Perceptron Plugin. Activating the Perceptron Plugin

NEW MONITORING TECHNIQUES ON THE DETERMINATION OF STRUCTURE DEFORMATIONS

On the Kinematics of Undulator Girder Motion

UNIT IV - Laser and advances in Metrology 2 MARKS

TERRESTRIAL 3D-LASER SCANNER ZLS07 DEVELOPED AT ETH ZURICH: AN OVERVIEW OF ITS CONFIGURATION, PERFORMANCE AND APPLICATION

AP Physics: Curved Mirrors and Lenses

Laser Scanning. 3D Model is not existing and is required for: studies revamping maintenance HSE integration in another 3D model archiving

Straightness measurement

Operation Manual. Autocollimator Test Wedge for quick testing of Electronic Autocollimators

EOSC 454 Lab #3. 3D Magnetics. Date: Feb 3, Due: Feb 10, 2009.

METHOD FOR DETERMINING WAFER FLATNESS USING THE MOVING AVERAGE QUALIFICATION METRIC BASED ON SCANNING LITHOGRAPHY

Level 6 Graduate Diploma in Engineering Engineering surveying

Rigorous Scan Data Adjustment for kinematic LIDAR systems

Digital Image Correlation of Stereoscopic Images for Radial Metrology

NEDO X-LINER SERIES. Dot and line lasers for interior finishing X-LINER SERIES

High Resolution Tree Models: Modeling of a Forest Stand Based on Terrestrial Laser Scanning and Triangulating Scanner Data

BEST-FIT ADJUSTMENTS OF THE REFLECTORS IN A COMPACT RANGE 1

ZED Tunnel Guidance Ltd

Developing a Method Using Field Surveying Techniques and Instruments to Create 3D Terrain Models. Faheem Muhammad

With FARO PointSense for Revit. Jörg Braunes Platform Owner Software Chris Palmer Sales Engineer 1

Teledyne PDS. Multibeam Calibration. Version January Teledyne RESON B.V. Stuttgartstraat AS Rotterdam The Netherlands

HIE ISOLDE ALIGNMENT AND MONITORING SYSTEM TECHNICAL DESIGN AND PROJECT STATUS

FARO Scanning Plugin

Automated front wall feature extraction and material assessment using fused LIDAR and through-wall radar imagery

Light and the Properties of Reflection & Refraction

Strategy. Using Strategy 1

Selective Space Structures Manual

Module 4A: Creating the 3D Model of Right and Oblique Pyramids

SEMI Draft Document 4537 Revision to SEMI M PRACTICE FOR DETERMINING WAFER-NEAR-EDGE GEOMETRY USING PARTIAL WAFER SITE FLATNESS

IGTF 2016 Fort Worth, TX, April 11-15, 2016 Submission 149

5 member consortium o University of Strathclyde o Wideblue o National Nuclear Laboratory o Sellafield Ltd o Inspectahire

MultiBeam-Laser MBL 5

Impact of 3D Laser Data Resolution and Accuracy on Pipeline Dents Strain Analysis

ARCHITECTURE & GAMES. A is for Architect Simple Mass Modeling FORM & SPACE. Industry Careers Framework. Applied. Getting Started.

ATLANTIC SHUTTER HARDWARE Storm Rated Installation Instructions

Measurements using three-dimensional product imaging

SIMPLE ROOM SHAPE MODELING WITH SPARSE 3D POINT INFORMATION USING PHOTOGRAMMETRY AND APPLICATION SOFTWARE

GEOMETRICAL SURVEY OF COMPACT ANTENNA TEST RANGES USING LASER TRACKER TECHNOLOGY

The process components and related data characteristics addressed in this document are:

Transcription:

Stanford University June 27, 2004 Stanford Linear Accelerator Center P.O. Box 20450 Stanford, California 94309, USA Three-Dimensional Laser Scanner Field Evaluation Specifications Metrology Department Alignment Engineering Group

Introduction The specifications document 1 describes three tests which will be used to evaluate the proposed laser scanners. This document provides details for carrying out these tests and for the evaluation of the data yielded from these tests. 1.0 Test 1: Tunnel Test The specifications document on page 5 defines Test 1 as follows: The field testing and evaluation of the laser scanner will include a scan of an area at the Stanford Linear Accelerator Center such as a tunnel in which magnets and other typical components are present. The scan will include measurements of control points that have a common datum with existing Alignment Engineering Group control monuments. This will ensure that the laser scans can be integrated with existing and new surveying control data for a positional tie to an existing network datum. The three-dimensional scans will have the ability to be overlapped and connected due to the unique geometry of the tunnel having a limited width but substantial length. More specifically, the location of the test will be most likely at a section of the beamline in the South Final Focus of the Stanford Linear Collider. It will be around 10 meters long and comprise at least three unique magnets plus a BPM (Beam Position Monitor) and a collimator. Scanner coverage will also include the existing pipes, wires and hoses as well as the walls in the region. The classical tooling used at the Stanford Linear Accelerator Center consists of a series of 1½ diameter steel spheres. They are placed on a nest (also known as a Hubbs Cup) that itself fits into a ¼ hole located on a component. With this setting, the center of the sphere is straight over the axis of the ¼ hole at a height offset of exactly 1. For use with the laser scanner, the steel spheres will be painted or preferably plated for optimal scanning reflectivity. The Stanford Linear Accelerator Center will supply these for the test. The steel spheres will be mounted on all accessible ¼ holes for each magnet as well as on the tunnel walls. The tunnel is approximately 3 meters in width and height and the walls are made of shotcrete. The beamline is located closer to one wall and is on a girder approximately 1 meter wide and 1.5 meters high. Testing in this area will ensure that this typical tunnel geometry including the minimum range requirements describe in Section 2 can be adequately covered with the candidate laser scanner. 1.1 Area to be scanned The tunnel test will be carried out in the Final Focus section of the now defunct Stanford Linear Collider (SLC). Scanner coverage shall include the girder components as marked in Picture 1 and include any ceiling components, existing pipes, wires, hoses and the walls. Picture 1: Horizontal demarcation of scanning area and area of special interest This task should be performed in several scans. SLAC will deliver 3D-coordinates of fiducial points (see table 1), which can be signaled with 1.5 spheres (as described in the specifications). 1 Three Dimensional Laser Scanner Technical Specifications & Field Evaluation written by Brian Fuss and Catherine LeCocq (March, 31, 2004, SLAC). Field Test Specifications.doc 2 of 8

Table 1 Fiducial Point Coordinates and Lay-out The vendor shall carry out the registration of the individual scans and the analysis of the data. As part of the test, the vendor shall demonstrate the Ease-of-use and functionality of the proposed software for data acquisition, data handling (simultaneous loading of multiple scans into memory for registration) and data modeling. Additionally, the vendor shall deliver the whole model (all scans brought to the global system, as defined by the control points) in ASCII-format for independent analysis. The evaluation of the measurement results will focus on the Completeness, the Level-of-Detail (LOD) and the Geometrical Trueness. To evaluate the laser scanner data against these criteria, but at the same time keep the data analysis manageable, the measurements shall emphasize on the areas of specific interest as described below. Field Test Specifications.doc 3 of 8

1.2 Areas of specific interest From the point cloud, three areas of specific interest will be selected and treated according to the above mentioned criteria. The areas selected are two vertical profiles (P1 and P2), perpendicular to the tunnel axis, and the area A indicated by the green boundary line in Picture 1. The profiles will be marked by spray paint to facilitate the easy identification of the profiles in the point cloud. Pictures 2 and 3 show the Profile 1 area in more detail. Picture 3 Collimator between two magnets Picture 2: 270º View of Profile 1 Area P1 is approx. 30-50 cm wide. Inside the shown area the LOD and the Completeness will be investigated. In the rest of the complete Profile P1, the geometrical deviations from the true geometry of the tunnel and ceiling mounted components (there are two magnets suspended from the ceiling) will be analyzed. In order to acquire all the relevant details (Pic. 3) the scanner has to be positioned in the vertical profile. If the tested scanner is restricted in the vertical measurement range, the vendor should propose solutions for digitizing the space between the magnets and overhead. Field Test Specifications.doc 4 of 8

A second vertical Profile P2 will be selected. P2 is small (compared to P1) and 3 to 4 m away from P1. The analysis of P2 will be done with the point cloud acquired from the position in P1. In this configuration a profile from an off-the-profile-scan-position will be evaluated which is the standard case in long-tunnel objects like accelerator tunnels. Cables magnet Cables magnet Concrete table Concrete table 1.2 Data Evaluation 1.2.1 General Data Analysis Geometrical trueness is evaluated by comparing the local (global) scanner coordinates with their nominal values. A set of well distributed known points (see table 1) is to be used as identical points for the local-to-global transformation of the point cloud(s). The coordinates of these points are given and based on these coordinates a transformation into the beam-line-coordinate-system is to be performed. The unknown fiducial points (as listed in Table 1, test points) signaled with 1.5 spheres should be scanned and determined. In order to investigate the quality of the registration, some of the test points will be signaled only for the first scan and others only for the subsequent ones. The following information should be extracted from the point cloud: Coordinates of fiducial Diameter of spheres signaling the fiducial points 1.2.2 Area A In Area A several objects are selected and calculated to check for repeatability, i.e. consistency within the point cloud and resulting geometrical trueness. These tests will be performed using Magnet M, which is the red object on the very right in Picture 3. The top, side and end plane should be intersected to determine the coordinates of the visible corners. Additionally, the diameter and the length of the collimator (the stainless steel cylinder to the left of Magnet M) should be determined. 1.2.3 Profile P1 In P1 the Level of Detail (LOD) will be investigated. The criteria are: - completeness - resolution (density of the point cloud) - perturbations in the point cloud by total reflections and other optical deviations - geometry (position, size, orientation) of some objects. Field Test Specifications.doc 5 of 8

The comparison will be done by visual inspection and by different extractions from the point cloud (section, cross sections, etc ) 1.2.4 Profile P2 P2 is a small profile (compared to P1). The main criteria here are the deviation of the point cloud from the real tunnel (wall or inventory), i.e. geometrical trueness and the completeness of the entire profile. The evaluation will be done by comparing to the known profile, by visual inspection, and by different extractions from the point cloud (section, cross sections, etc ) 2.0 Test 2: Building Test This test is designed to test and evaluate the range of the scanner and the flexibility of the whole measurement equipment. Test Location For part one of this test, the left two sections of the north facade of Bldg 26 (see Pic. 4) shall be scanned with the roll-up door leading into the alignment laboratory opened. Subsequently, the interior of the alignment lab (see Pic. 5) is to be scanned such that the exterior and interior scans can be registered. The interior scan should yield a floor plan of the laboratory. No control points are provided. Picture 4 Building 26 Facade Picture 5 Interior Room Scan Area 3.0 Test 3: Accuracy Test 3.1 Test Location All tests will be carried out in the Sector 10 Calibration Laboratory. 3.2 Tests to be performed Eleven sphere mounts with 1.5 spheres are attached to the East facing wall of the laboratory at about 1.8 m height. This test field will be scanned from different positions. Additionally, a resolution target is to be scanned at three different distances. Field Test Specifications.doc 6 of 8

3.2.1 Horizontal relative distance test The scanner will be placed in the prolongation of the line of spheres (Pos. 3.2.1). From this position all targets are to be measured in one scan. Spheres 3.2.2 Horizontal absolute distance test The scanner will be placed in the middle of the test field, very close to the line of spheres (Pos. 3.2.2). From this position all targets are to be measured in one scan. 3.2.3 Horizontal Angle test The scanner will be placed in the middle of the test field, as far away as possible from the line of spheres (Pos. 3.2.3). From this position all targets are to be measured in one scan. 3.2.4 Vertical Angle test Seven sphere mounts have been mounted to the wall about equally spaced along a vertical line. The scanner shall be set-up with an instrument height of about 1.5 m close to the opposite wall. From this position, the instrument should scan all seven spheres. 3.2.3 3.2.2 3.2.1 Picture 6 Horizontal Distance Test Picture 7 Vert. Distance Test 3.3 Resolution The resolution test uses a target as proposed by Dr. Boehler from the i3mainz group. The front panel has slots which are about 30 mm wide at the outside becoming smaller towards the center. If a scanner has a high resolution (small angular increments and a small laser spot) there should be reflections not only from the front panel but also from the bottom of the box which is about 60 mm behind the front panel. If the resolution is very good, these reflections from the bottom should not Field Test Specifications.doc 7 of 8

only be present in the outer regions but also near the center. This target shall be scanned to detect resolution information from three different distances, 3 m, 12 m, and 25 m. 3.4 Data evaluation From each point cloud the position of each sphere should be determined in the local scanner coordinate system. Each determination has to be calculated with nominal and unknown radius. In the latter case the radius is also part of the result. Field Test Specifications.doc 8 of 8