3M Static Sensor 718, 3M Air Ionizer Test Kit 718A. User s Guide

Similar documents
3M Workstation Monitor Checker CTE701A User s Guide

3M Wrist Strap and Ground Monitor 773. User s Guide

3M Constant Monitors WSMONITOR1, WSMONITOR 2, WSMONITOR3

3M Ground Master Equipment Ground Monitor Models CTC065-RT and CTC065-5 User s Guide

Electrostatic Fieldmeter Model 775

Lotus DX. sit-stand workstation. assembly and operation instructions. MODEL # s: LOTUS-DX-BLK LOTUS-DX-WHT

Super Stack. (Little Stack Wireless + Big Stack)

charge stream desk stand Made for Apple, Samsung, and other Qi-enabled smartphones USER MANUAL

Logitech Alert 700i/750i System Requirements & Support Guide

user s manual Battery case model #: IPB3500S Battery Case Charger for iphone 6

OPERATING INSTRUCTIONS POWERSMART 10 10,000 MAH PORTABLE POWER PLEASE READ BEFORE OPERATING THIS EQUIPMENT

Bluetooth Enabled Access Control MODEL BG-FE. Operating Instructions

MobilePro Users Guide Page 1 of 35. MobilePro HPC. User's Guide. Notices Proprietary Notice and Liability Disclaimer Table of Contents

Introduction. Package Contents. System Requirements

SySTIUM TECHNOLOGIES. Assembly Guide. Model 133i

MP3 Speaker USER GUIDE

HD40H(X) Performance Series Camera. User Guide

Nokia Wireless Charging Plate DT-900 User Guide

This simple chassis houses one media-converter module and features an external AC power supply and also a terminal block for DC power input.

SATA Hard Drive Enclosure

W IRELESS8000 T. Powerful Universal Portable Charger PLEASE READ BEFORE OPERATING THIS EQUIPMENT

Bluetooth Hands-Free Speaker Phone Custom Accessories 23685

STATIC METERS Models 211 & 212. SURVEYORSTAT Models 211XL & 212XL

*520886* IntelliTouch Pool & Spa Control System MobileTouch Wireless Controller. User s and Installation Guide. P/N Rev A

User Guide Wireless Charging Plate DT-903

Honor Whistle Smart Control Rechargeable Headset User Guide

NIMBUS a personal dashboard for your digital life

User Guide Nokia Wireless Charger DT-601

To force your device to restart, press and hold the power button until your device vibrates.

LITE TUNES #V

PrintPAD MC65. User Guide

User Guide Microsoft Portable Power (DC-21)

Owner s Manual. 5200mAh Mobile Power Bank

SMART SWITCH. User s Manual. This product is to be used with the MySmartBlinds automation kit.

Digital Camera USER GUIDE

ÿ ÿ HA09 - HANDY REMOTE CONTROLLER INSTRUCTIONS

SPOTTER the multipurpose sensor

COBALT C INSTALLATION GUIDE RFID CONTROLLER ESCORT MEMORY SYSTEMS. High Frequency Passive Radio Frequency Identification Controller

SHINE 3000 WITH FASHION CABLE

powerstation keychain Made for USB-C smartphones and wearables USER MANUAL

Dome Network Camera. Quick Start Guide

Nokia Bluetooth Headset BH-300 User Guide Issue 1

Version 1.0 December isappos 4 / isappos 5 Scanner Jacket

IRONKEY H80 USER GUIDE

N331 Wireless Mini Optical Mouse User s Guide

USB-A to Serial Cable

Owner s Manual. Industrial-Grade USB 3.0 Hub. Models: U IND (4-Port), U IND (7-Port), U IND (10-Port)

8-button wireless laser mouse USER GUIDE

Product Manual. USB Charging Hub 10 Port Hi-Power 2.4A USB Charger Hub. Coolgear, Inc. Version 1.1 September 2017 Model Number: U2CHGRHUB10

Charging Pad / Charging Stand

3.5 External IDE Drive Case

charge force battery Made for OtterBox universe cases USER MANUAL

3M I/O Connector Mounting Options and Hardware

PrintPAD MC65. User Guide

Quick Start Guide KB20. A3 10 Type Case CJB1G9KALAAB

Owner's Manual. For latest instructions please go to

USER GUIDE. AXIS T8120 Midspan 15 W 1-port ENGLISH

P OCKET POWER 9000 T. Powerful Universal Portable Charger PLEASE READ BEFORE OPERATING THIS EQUIPMENT

User's Guide. MiniTec TM Series Model MN25 MultiMeter

Bad Elf Wombat User Manual

USER GUIDE. Laptop Charger NS-PWLC591/NS-PWLC591-C. Before using your new product, please read these instructions to prevent any damage.

juice pack Made for Samsung Galaxy S7 & Samsung Galaxy S7 edge USER MANUAL

Introduction. 1. RF Module

Wireless Stow-N-Go Rechargeable Notebook mouse USER GUIDE

SCSI PCI Card. 80 MBytes/sec Ultra2 SCSI PCI Controller Card PCISCSIU2W. Instruction Manual. Actual product may vary from photo

User s Manual. Bluetooth Slim Keyboard. Page

Voice Recording Presenter with Laser Pointer USER GUIDE

SATA Hard Drive Enclosure

BLUETOOTH SPEAKER. for ipod/iphone/ipad/android. User s Manual. Item #TA-11BTSP. PC Mac Ultrabook COMPATIBLE BLUETOOTH WIRELESS USB

Quick Installation Guide

P OCKET POWER 6000 T. Powerful Universal Portable Charger PLEASE READ BEFORE OPERATING THIS EQUIPMENT

DoorEye Camera MDOOREC101012E

P OWERBUDS. The perfect accessory for anyone who wants to be truly wireless PLEASE READ BEFORE OPERATING THIS EQUIPMENT

USER S GUIDE. LS355 Numeric Pager

INSTRUCTIONS FOR USE. RECHARGEABLE System for Hearing Aids

The Portable Air Pollution Monitor

powerstation AC Made for Smartphones, Tablets, Laptops, & Other USB & AC Devices USER MANUAL

Wa r n i n g-wa r n i n g-wa r n i n g

USB Transfer Cable. for Windows and Mac User s Guide

Gigabit Miniature Media Converters. Connect Gigabit Ethernet copper ports to fiber optic cable.

Z-Wave Ceiling-mount Smoke Detector HZS-200E/HZS-200A

CardBus PCI Host Controller Card

LIFX is color changing, Wi-Fi lighting that you control with your smartphone or tablet.

Instruction Manual. for. Audio Ape Remotes. 1

WCC100 IN-VEHICLE CHARGING CRADLE OWNER S MANUAL

Bluetooth USB Adapter TALUS. User Guide

Owner s Manual. DVI Single Link with Audio Splitter. Models: B A and B A

2U ATX12V 2.01 Power Supply

USER MANUAL. Elo Touch Solutions USB Expansion Module

User s Manual. Rechargeable Wireless Presenter Mouse. Page

Instruction Guide. Two Port ISA Serial Card ISA2S550. The Professionals Source For Hard-to-Find Computer Parts

powerstation & powerstation XXL Made for Apple and USB devices USER MANUAL

PCI Serial Card 2 Port RS-422/RS-485 PCI Serial Card with Terminal Block Connectors PCI2S422TB Instruction Manual Actual product may vary from photo

Register your product and get support at. SPP3038A. User manual

ipod /iphone Audio/Video Dock with Remote Control DX-IPDR3 USER GUIDE

1101 Universal Transmitter

Trademarks and registered trademarks are proprietary to their respective manufactures.

juice pack Made for Samsung Galaxy Note9 USER MANUAL

1. Product description

Owner s Manual HDMI Matrix Switch

Transcription:

3M Static Sensor 718, 3M Air Ionizer Test Kit 718A User s Guide

Table of Contents Section Page Safety Information... 2 1 Description... 3 2 Power Requirements and Battery Installation... 3 3 Operation of the 3M Static Sensor 718... 4 4 Operation and Use of the 3M Air Ionizer Test Kit 718A... 7 5 Service/Calibration... 8 6 Physical Characteristics (All values typical)... 10 7 Parts Included... 11 Regulatory Information... 12 Warranty... 13 1

SAFETY INFORMATION Intended Use: The 3M Static Sensor 718 is a portable, handheld instrument designed for measuring voltages associated with electrostatic charge. Its intended use is for measuring the amount of voltage, in a range from 19,900 kilovolts, associated with an electrostatic charge buildup on a surface. Any deviation from this intended use could impair the instrument s effectiveness and possibly lead to an unsafe operating condition. The 3M Air Ionizer Test Kit 718A is a set of accessories for use in conjunction with the Static Sensor 718. Their intended use is for verifying the operation of air ionizers by measuring the neutralization time for a static charge placed on a stationary metal plate. Any deviation from this intended use could impair the instrument s effectiveness and possibly lead to an unsafe operating condition. CAUTION The Static Sensor 718 and 3M Charger 718A use 9VDC power supplied by a 9V alkaline battery. Usage of any other power source may cause damage to the instruments. The Static Sensor 718 and Air Ionizer Test Kit 718A have no user-serviceable parts. Do not disassemble the products for any reason. UNAUTHORIZED SERVICE WILL VOID THE WARRANTY. The Static Sensor 718 and Charger 718A are NOT designed for usage in hazardous environments where the possibility of explosion or fire exists. Read and understand all safety information before installing and operating this equipment. 2

1.0 Description The 3M Static Sensor 718 is a portable handheld instrument used for locating and measuring electrostatic charges. It can beused to locate ESD trouble-areas, and is a valuable tool for the ESD-control engineer. Used in conjunction with the 3M Air Ionizer Test Kit 718A (available separately), it can be used for verification and auditing of air ionizers.the Static Sensor 718 is battery-powered and has several measurement features: Range: measurements can be taken in a 0-1,990 kv or 19,900 kv range. Automatic Zero: push button feature allows easy adjustment to zero. No screws or dials to turn. HOLD function: allows the User to freeze a displayed measurement, for later evaluation. Automatic shutoff: conserves battery power by shutting off the instrument after 20 minutes of inactivity. 2.0 Power Requirements and Battery Installation 2.1 Both the Static Sensor 718 and 3M Charger 718A use a 9V alkaline battery. 2.2 To install the battery on either unit: 2.2.1 Remove the battery cover, located in the lower back of the unit. To do this, press down on the cover and slide it downward. 2.2.2 Pull the battery connector out of the housing, and align the male/female ends of the connector with the proper terminals on the battery. 2.2.3 Connect the two and place the connected battery into the housing by inserting the connector end first, then following up with the other end of the battery. 2.2.4 Replace the cover. 3

2.3 The 3M Static Sensor 718 has a Low Battery indicator. Once the battery is depleted to approximately 6.5 volts, the instrument will show BAT in the display. At this time, the Static Sensor 718 will not produce accurate results and the battery should be replaced. 2.4 The 3M Charger 718A also has a low battery indicator. This is an LED located at the lefthand side of the unit. When the battery voltage drops below operating level, the LED will light up. At this time, the user shouldreplace the battery. Usage of the Charger 718A under low battery conditions would lead to insufficient voltage levels being generated. 3.0 Operation of the Static Sensor 718 Supplemental notes: For accurate measurements during usage, it is recommended that the outside housing of the Static Sensor 718 be connected to an electrical ground. This can be accomplished by having the user holding the instrument connected to ground through either a static control wrist strap, or while wearing static control footwear. The enclosure of the Static Sensor 718 is made of conductive plastic and is, therefore, electrically connected to whomever is holding the instrument. In addition, ground for the instrument can be provided through the Voltage Monitor Output on the front of the case. 4

The 3M Static Sensor 718 is a precision electronic instrument. Improper use or rough treatment can damage the unit, and render it incapable of providing accurate measurements.2.2.4 Replace the cover. 2.3 The Static Sensor 718 has a Low Battery indicator. Once the battery is depleted to approximately 6.5 volts, the instrument will show BAT in the display. At this time, the Static Sensor 718 will not produce accurate results and the battery should be replaced. 3.1 Turning the unit ON and OFF: To turn on the Static Sensor 718, momentarily press down on the membrane switch labeled POWER. Determining that the power is on can be verified by seeing that the liquid crystal display (LCD) is on, and that the red light-emitting diodes in the front of the sensor are lit. To shut off the Static Sensor 718, momentarily press down on the membrane switch labeled POWER. Determining that the power is off can be verified by seeing that the liquid-crystal display (LCD) is off. 3.2 Making Electrostatic Voltage Measurements: hold the instrument 1-inch (2.54 cm) away from the object being measured. The display will update with the voltage measurement in kilovolts. If the measured voltage is greater than the measurement range of the instrument, a -1. will be displayed. At this time, switch to a greater range. If over-ranging occurs even with the high range activated, the static charge on the object cannot be measured with the Static Sensor 718. 3.3 Measurement Range: all measurements are in kilovolts (kv) as stated on the front label of the unit. The Static Sensor 718 has two measurement ranges: 0-1,990 kv and 0-19,900 kv. The unit s current measurement range mode can be verified by checking the display. Three digits following the decimal point indicate that the unit is in 0-1,990 kv range. Two digits following the decimal point indicate that the sensor is in 0-19,900 kv range. To change between measurement ranges, press the RANGE/HOLD button once, momentarily. 3.4 HOLD Function: in the event that the user wishes to freeze the current measurement, the HOLD function of the Static Sensor 718 may be used. Simply press the HOLD switch momentarily and the currently displayed voltage will be frozen. A HOLD notice will also be displayed to alert the user that the instrument is currently in HOLD status. To unfreeze the display 5

and return to floating measurement, momentarily press the HOLD switch once again. Please note that during HOLD condition the distance indicating LED s are turned off. 3.5 Zero adjustment: the 3M Static Sensor 718 has a zero adjustment function, which sets a zero reference point for all subsequent measurements. This zero reference can be set by pointing the instrument at a known zero-voltage surface, and holding down the RANGE/ZERO button for longer than 3 seconds. After 3 seconds, the display will flash and adjust to zero. Repeat this step for both the low and the high ranges. The zero adjustment should be performed every time the unit is turned on. 3.6 Measurement Accuracy: 3.6.1 Distance Indicator: the Static Sensor 718 is factory calibrated to give accurate measurements when it is placed one inch (2.54 cm) away from the object to be measured. To assist the user in gauging this distance, two light-emitting diodes (LED s) are present on the front face of the instrument. These LED s emit two red, bullseye targets on the surface of the object being measured. As the instrument gets closer to the one inch measurement distance, the bullseyes begin to converge. When they converge and become one, the instrument is approximately one inch away, and the measurement can be made. For more accurate measurements, it is recommended that the user manually measure the distance between the front housing of the instrument and the object being measured. 3.6.2 Accuracy and Size of Object to be Measured: the minimum surface area on an electrostatically charged object which can be accurately measured is a 5 square inch (32.3 cm 2 ) area. 3.6.3 Measurements from Greater than One Inch (2.54 cm) Away: in the event that a one inch separation between object-to-be measured and the Static Sensor 718 cannot be achieved, it is possible to get approximate readings. Please contact your 3M representative for additional information. 6

3.7 Continuous Output: an output jack is provided on the front of the 3M Static Sensor 718. This output can be used to feed a continuous signal into a data storage device for continuous monitoring of measured voltages. Please use a 3/32 inch (2.5 mm) mono-phone plug to connect into the output jack. The output signal is dependent on the measurement range currently selected. For the low range, the output signal is 1/1000 of the measured electrostatic voltage. For the high range, the output signal is 1/10,000 of the measured voltage. 3.8 Automatic Shut-Off: the Static Sensor 718 will automatically shut-off 20 minutes after the last switch activity. This is done in order to conserve battery power. In the event that the user needs to have the unit stay ON continuously, when turning the unit on press the POWER AND RANGE switches simultaneously. This deactivates the Automatic Shut-Off feature. The BAT indicator will then flash three times to indicate that the automatic shut-off features has been disabled. The Automatic Shut-Off feature will reset itself the next time the instrument is turned on. 4.0 Operation and Use of the 3M Air Ionizer Test Kit 718A It is recommended that the user be familiar with ionizer test standards ANSI/ ESD S3.1 and draft standard ANSI/ESD SP3.3 if the Air Ionizer Test Kit 718A is used to perform verification testing on ionizer performance. 4.1 Assembly: slide the charge plate over the Static Sensor 718 until it stops. The charge plate slides onto the lower groove, on the sides of the Static Sensor 718. 4.2 Charging the plate: holding the Static Sensor 718 (with charge plate attached) in one hand, use the other hand to touch the probe of the 3M Air Ionizer Test Kit 718A to the charge plate. Press either the + button (for a positive voltage) or the button (for a negative voltage), then remove the probe from the charge plate. Be sure to keep the button pressed while removing the probe from the charge plate. The display on the Static Sensor 718 will 7

indicate a positive or negative charging voltage (1.1kV minimum). If a voltage of less than ±1.1kV is displayed, check to see if the low battery indicator on the 3M Charger 718A is illuminated. If illuminated, replace the battery in the charger. If the unit continues to supply an incorrect voltage to the charge plate, please contact 3M for additional instructions. 4.3 Testing ionizer discharge time: after charging the plate, hold the 3M Static Sensor 718 approximately one foot (30.5 cm) away from the ionizer. Monitor the display to see how quickly the 1.1 kv charge is dissipated to 0.1 kv. The speed at which this occurs (the discharge time) indicates how well the ionizer is operating. Please refer to the specific ionizer s operating manual or consult with the ionizer manufacturer to determine what this discharge time should be. Repeat this procedure for both a positively and a negatively charged plate. 4.4 Testing ionizer offset balance: zero the charge plate by touching it with a grounded object. This can either be the finger of a grounded person or some other item which is connected to electrical ground. In either case, zeroing the charge plate should make the display on the Static Sensor 718 read zero. Hold the Static Sensor 718 approximately one foot (30.5 cm) in front of the ionizer. Monitor the display. The value displayed is the offset balance of the ionizer, which is the difference between the number of positive and negative ions being emitted. Please refer to the specific ionizer s operating manual or consult with the ionizer manufacturer to determine what this offset balance should be. 5.0 Service/Calibration 5.1 Service and Repair: in the event that you believe the Static Sensor 718 or the 3M Air Ionizer Test Kit 718A is in need of repair, please contact your local 3M representative for troubleshooting help, and, as needed, repair information. There are no user-serviceable parts on either product. 8

5.2 Calibration: the 3M Static Sensor 718 and 3M Air Ionizer Test Kit 718A products are supplied by the factory pre-calibrated. 3M does not specify a minimum calibration cycle for the Static Sensor 718 or Air Ionizer Test Kit 718A products. The user, usually according to internal Quality procedures, determines calibration cycles. In the event that the user wishes to perform a self-calibration, the following steps should be followed for the Static Sensor 718 (user-calibration not possible on the Air Ionizer Test Kit 718A). 5.3 Equipment Needed: Test Fixture High-Voltage Power Supply, capable of supplying voltages up to 10,000V Voltmeter, with > 50 kω input impedance, capable of measuring voltages down to the μv range. Cable with a 3/32 inch (2.5 mm) mono plug and secondary connector to interface with voltmeter. 5.4 Test fixture: Metal plate of at least 5 square inches area (38.7 cm 2 ) area. Metal stand capable of supporting Static Sensor 718, and holding it one inch (2.54 cm) away from the metal plate, centered. Connectors on the plate with which it can interface with the high voltage power supply. Connections on the metal stand with which it can be connected to electrical ground. 5.5 Procedure: 5.5.1 Place the Static Sensor Static Sensor 718 on the metal stand. Verify that it is exactly one inch (2.54 cm) away from the metal plate, and that its position is centered relative to the plate. 5.5.2 Connect the stand to ground. 5.5.3 Turn on the Static Sensor 718 and set it to the low range. 5.5.4 Ground the plate. Zero the display. 5.5.5 Remove the ground from the plate, and connect it to the high voltage power supply. Apply a 1 kv charge to the plate. 9

5.5.6 The instrument should now be reading 1.000. If it is not, remove the battery door and use a screwdriver to turn the small screw located inside the battery compartment. Use a small screwdriver to turn the small screw, located on the front right side of the instrument. This screw should adjust the reading on the display. Once the display has been adjusted to read 1.000, the low range of the 3M Static Sensor 718 is now calibrated. 5.5.7 Repeat procedures 5.5.2-5.5.6 for the high of the meter, using a test voltage of 5,000 volts. 6.0 Physical Characteristics (All values typical) (at 1 inch (2.54 cm) distance from sensor to target) 3M Static Sensor 718 Description Dimensions Weight Typical Value 0.85" H x 2.4" W x 4.2" L 2.2 cm H x 6.1 cm W x 10.6 cm L 4.4 oz. (125 g) with battery Operating Conditions 41º F - 95º F (5º C - 35º C) Up to 80% RH, non-condensing Indoor use only For use at altitudes below 42, 300 ft. (2,000 m) Pollution Degree II Class III Enclosure Power Requirements Measurement Ranges Voltage Display Display Resolution Conductive Housing One 9-volt alkaline battery 0 1,990 kv 0 19,900 kv Liquid Crystal Display 1V/inch (0.39V/cm) @ low range 10V/inch (3.9V/cm) @ high range ± 10 counts 10

3M Static Sensor 718 - continued Description Typical Value Voltage Output Jack Voltage Output Automatic Shut-off Distance Indicator Measurement Accuracy Measurement Stability Certifications 3/32 in. (2.5 mm) monophone Tip: signal Sleeve: ground 1/1000 of measured voltage @ low range 1/10,000 of measured voltage @ high range 20 minutes after last switch activity LED targets. Aligned targets indicate 1 in. (2.54 cm) measurement distance Within 5% of measured voltage ± 10 counts culus, CE 3M Air Ionizer Test Kit 718A Description Typical Value Charge Plate Assembly Charge Plate Area Per ESD Association Draft Standard ANSI/ESD SP3.3 Aluminum bracket Bare stainless steel plate Teflon spacers isolate plate from bracket 3.25" W x 1.25" L 8.3 cm W x 3.2 cm L Charge Plate Assembly Weight 2.5 oz (70 g) Charger Dimensions Charger Weight Charger Power Requirements Charger Output (using Static Sensor 718 with charge plate) Certifications 0.87" H x 2.4" W x 4.9" L 2.2 cm H x 6.1 cm W x 12.4 cm L 5 oz. (140 g) with battery One 9 volt alkaline battery 1.1kV minimum for positive or negative voltage culus, CE 7.0 Parts Included Static Sensor 718 1 ea. Static Sensor 718 1 ea. 718/718A User s Guide Air Ionizer Test Kit 718A 1 ea. Charge Plate Assembly 718A 1 ea. Charger 718A 1 ea. 718/718A User s Guide 11

Regulatory Information WEEE Statement The following information is only for EU-members States: The mark shown to the right is in compliance with Waste Electrical and Electronic Equipment Directive 2002/96/EC (WEEE). The mark indicates the requirement NOT to dispose the equipment as unsorted municipal waste, but use the return and collection systems according to local law. Caution To reduce the risks associated with environmental contamination: Dispose of the monitor and power adapter in accordance with local, state, and federal regulations. culus Statement Meets UL Safety Requirements. FCC This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: Reorient or relocate the receiving antenna. Increase the separation between the equipment and receiver. Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. Consult the dealer or an experienced radio/tv technician for help. Restriction of the use of certain hazardous substances (RoHS): Meets 2011-65-EU Directive. 12

UL Listed to U.S. and Canadian Safety Standards. Mark of Conformity to European Directives (Conformité Européene). 3M is a trademark of 3M Company. Important Notice All statements, technical information, and recommendations related to 3M s products are based on information believed to be reliable, but the accuracy or completeness is not guaranteed. Before using this product, you must evaluate it and determine if it is suitable for your intended application. You assume all risks and liability associated with such use. Any statements related to the product which are not contained in 3M s current publications, or any contrary statements contained on your purchase order shall have no force or effect unless expressly agreed upon, in writing, by an authorized officer of 3M. Warranty; Limited Remedy; Limited Liability. This product will be free from defects in material and manufacture for one year from the time of purchase. 3M MAKES NO OTHER WARRANTIES INCLUDING, BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. If this product is defective within the warranty period stated above, your exclusive remedy shall be, at 3M s option, to replace or repair the 3M product or refund the purchase price of the 3M product. Except where prohibited by law, 3M will not be liable for any indirect, special, incidental or consequential loss or damage arising from this 3M product, regardless of the legal theory asserted. 3 Electronics Materials Solutions Division Static Control Products 926 JR Industrial Drive Sanford, NC 27332-9733 Toll-Free: 866-722-3736 International: 919-718-0000 Email: 3mstaticinfo@mmm.com www.3mstatic.com Please recycle. Printed in USA. 3M 2014. All rights reserved. 78-9102-1938-1 JHA