T 1 Relaxation Measurement: The Inversion-Recovery Experiment (Using IconNMR) Revised

Similar documents
KJM Proton T1 Spectra on the AVI-600 and AVII-600. Version 1.0

Diffusion Ordered Spectroscopy (DOSY) Overview BRUKER last edit 5/14/12

Processing data with Bruker TopSpin

IconNMR Operating Procedures

VNMRJ 4.2 INSTRUCTIONS: QANUC 500 FOR CHEMISTS

DCIF NMR Training Guide 400 MHz Bruker AVANCE III HD b400 last edit 3/9/2015

DCIF NMR Training Guide 401 MHz Bruker AVANCE III HD b401 last edit 3/11/2015

UC DAVIS NMR FACILITY BRUKER SHORT MANUAL FOR TOPSPIN 3.X

Short instructions for the basic operation of the AVANCE II 400 MHz Bruker NMR Using ICON-NMR (PS751)

Bruker Topspin NMR Training Manual

Using OPUS to Process Evolved Gas Data (8/12/15 edits highlighted)

VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400

Bruker AVANCE III HD 400 MHz Spectrometer Operation

VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 AND VARIAN VNMRS 500

Vanderbilt Small Molecule NMR Center

Fitting NMR peaks for N,N DMA

NMR Spectroscopy with VnmrJ. University of Toronto, Department of Chemistry

NMR Spectrometer Crib-Sheet

TopSpin QuickStart. Default State TopSpin should be running at all times. If this is not the case, you may need to Log In and Start TopSpin.

NMR Training VNMRJ 4.2A/VNMRS 400

PROCESSING 2D SPECTRA USING VNMRJ JB Stothers NMR Facility Materials Science Addition 0216 Department of Chemistry Western University

Quick Guide to Topspin

PINMRF. Bruker AV-III / Avance DRX NMR Spectrometers running TopSpin Training Supplement for Advanced 1D NMR Spectroscopy

1D NMR Training Manual for Operating the Bruker 700 MHz Avance III HD NMR. Spectrometer without the Sample Changer and IconNMR

Routine 1 H and 13 C NMR Data Acquisition Basic Topspin Processing Guide For PSC and NSC Bruker 400 NMR Spectrometer

TOCSY 15. Goto. Introduction Pulse Sequence Diagram AVANCE User s Guide Bruker 167

2 Dimensional NMR User s Brief Guide

c. Saturday and Sunday: 10 minute time blocks and unlimited time.

DCIF NMR Training Guide 400 MHz Bruker DRX B401 Bruker DPX B400

Brief IconNMR and Topspin 3.5 User Guide for Bruker NMR Spectrometers Avance IIIHD 400MHz NMR with Autosampler in Chemistry room 93 DISCLAIMER

On Artemis and Cronus, make sure you are adjusting the db scale and not the watts

Select the Points You ll Use. Tech Assignment: Find a Quadratic Function for College Costs

1D and 2D Experiments Step-by-Step Tutorial

MestReC Cheat Sheet. by Monika Ivancic, July 1 st 2005

Walkup NMR. Varian NMR Spectrometer Systems With VNMR 6.1C Software. Pub. No , Rev. A0800

500 MHz Solution-state NMR Procedure. (Bruker AVANCE Machines running TopSpin under WINDOWS XP)

NMR Users Guide Organic Chemistry Laboratory

Spectrometer Visible Light Spectrometer V4.4

Single-sided Magnet Diffusion Measurements

Fluorimeter User s Booklet

Agilent VnmrJ 3.2. Quick Start Guide

TraceFinder Analysis Quick Reference Guide

General Information for Automated Inova 500 MHz NMR (Indy)

CH142 Spring Spectrophotometers with Vernier Data Acquisition Software

Galaxie Photodiode Array Software

NMR INSTRUMENT INSTRUCTIONS: Safety and Sample Preparation

Bruker NMR Topspin Manual

Appendix E: Software

Protein Deconvolution Quick Start Guide

Chancellor s Office Information Technology Services Corporate Information Management FIRMS Budget Review Hyperion Process

TOPSPINPLOT(XWINPLOT)

TopSpin. Multiplet Analysis Tutorial

KJM NMR processing everywhere with Windows Remote Desktop and TopSpin 3.5 at nmr-server.uio.no. Windows Server Version 1.

Bruker Avance 400 MHz Instructions

Barchard Introduction to SPSS Marks

Intro to 2D NMR: Homonuclear correlation COSY, lr-cosy, and DQ-COSY experiments

Introduction to Windows

CSImage Tutorial v August 2000

CRL MASS SPECTROMETRY FACILITY USER MANUAL. A Guide to Oligonucleotide Analysis by MALDI TOF/TOF

Agilent ChemStation Plus

Metrics Tutorial. Table of Contents. 1. Introduction

Frequently Asked Questions (FAQ)

Processing With Topspin

500 MHz Solution-state NMR Procedure. (Bruker AVANCE Machines running TopSpin under WINDOWS XP)

Years after US Student to Teacher Ratio

TraceFinder Analysis Quick Reference Guide

Podium Data Analysis Software. User Manual. SWIS10 Version

Agilent MassHunter Qualitative Data Analysis

Manual. User Reference Guide. Analysis Application (EMG) Electromyography Analysis

Put the Graphs for Each Health Plan on the Same Graph

NMR Assignments using NMRView I: Introduction to NMRView

Bruker OminiFlex MALDI-TOF Mass Spectrometer Operation Quick Start

Process Eye Professional. Recall

FIFI-LS: Basic Cube Analysis using SOSPEX

Topics in Linguistic Theory: Laboratory Phonology Spring 2007

Jasco FP-6500 Spectrofluorimeter Updated November 14, 2017

Barchard Introduction to SPSS Marks

Chapter 3 Set Redundancy in Magnetic Resonance Brain Images

GET TO KNOW FLEXPRO IN ONLY 15 MINUTES

Lesson 1: Routine ITC Data Analysis and Fitting. Routine ITC Data Analysis. Lesson 1:Routine ITC Data Analysis and Fitting

ChromQuest 5.0 Quick Reference Guide

TopSpin is the successor to a program called XWinNMR, which appeared in the mid- 90s (I believe) and was used to run Bruker spectrometers until it

Nicolet is50 FTIR User s Booklet

Models for Nurses: Quadratic Model ( ) Linear Model Dx ( ) x Models for Doctors:

Agilent 6400 Series Triple Quadrupole LC/MS System

Version 1.1. COPYRIGHT 1999 Tufts University and Vernier Software. ISBN (Windows) ISBN (Macintosh)

Agilent EZChrom Elite. PDA Analysis

INTViewer Tutorial Cube Tutorial

HOW TO USE THE EXPORT FEATURE IN LCL

(Coins & Distributions & Bias s)

Double Click in in this space to open the the Calculator/Formula

Running Multiple Experiments on the Same Sample Automatically. (multizg, spooler/queue, iconnmr)

All textures produced with Texture Maker. Not Applicable. Beginner.

Basic 1D Processing. Opening Saved Data. Start the VnmrJ software. Click File > Open.

Plane wave in free space Exercise no. 1

Quick Guide to the Star Bar

Content. Getting data organized Good data processing Plotting

Guide to geometric morphometrics

MALDI Software Users Guide (simplified version)

Model FP-6500 Spectrofluorometer Instruction Manual. FP-6500 for Windows

Transcription:

T 1 Relaxation Measurement: The Inversion-Recovery Experiment (Using IconNMR) Revised 4-21-2015 Relaxation times are divided into two types: longitudinal, which concerns change in magnetization along the z-axis (T 1, an enthalpy effect), and transverse, which concerns change in magnetization in the x-y plane (T 2, an entropy effect). The inversion-recovery experiment is used to determine the time constant for the T 1 process. One of the assumptions made in the T 1 calculations is that the equilibrium is approached exponentially, and therefore, the magnetization along the z-axis is represented by equation 1. Here, M 0 is the magnetization at thermal equilibrium, t is the time elapsed, and T 1 is the time constant that is obtained by comparing different values of M z at different times. M z = M 0 (1 - e -t/t1 ) (1) The pulse sequence used in the inversion-recovery experiment is shown below: d1 - p1(180 ) - d2 p2(90 ) FID In this experiment, the nuclei are first allowed to relax to equilibrium (d1, obviously, in the first run through the pulse sequence d1 is superfluous). A 180 pulse (p1) is then applied to invert the signals. The signals are then allowed to relax for a length of time (d2) that is varied for each experiment. After the variable d2, a 90 pulse (p2) is applied and the FID is recorded. The FID records the spectrum intensity as a function of the variable delay d2, and the longer d2 is, the more the signal will have relaxed. The peak intensity will reflect the extent to which each signal has relaxed during the d2 period. Procedure using IconNMR 1. Working first in TopSpin: It is best to use a fairly minimal acquisition window (SW) when running the inversion-recovery experiment. If you already have a regular proton spectrum of your sample, you can use that to determine the proper acquisition window, otherwise, acquire a normal spectrum. 2. Open the spectrum file and process as usual. Expand around the peaks of interest, so that in the display, you leave about 0.5 ppm of baseline on either side of the spectrum. You may exclude the solvent peak, if desired, unless you want to know its T 1 (it is likely to be quite long). After expanding the spectrum, click on the button Set SW, SFO1 from current region, shown to the right. 3. A dialog box will appear and will give you the values of SW and O1 that correspond to the displayed spectral region. Write down the values for SW and O1 (both in ppm). You ll need these to set up the T1 experiment. Close the dialog box. 4. Move to IconNMR and click on the appropriate entry line that corresponds to your sample in the sample changer. Fill in the filename, experiment number, solvent and select the experiment: N ProtonT1.

The Inversion-Recovery Experiment page 2 If&you&already&have&a& folder&with&data&for&this& sample,&you&can&use&that& filename&and&increment& the&experiment&number.& Scroll&(way)&down&in&this&window&to&find&and&select&the&N"ProtonT1"experiment& 5. Various parameters need to be set for this experiment. From the menu items along the top of the Icon window, select Parameters > Pertinent Acquisition Parameter Editor, for a set of the relevant acquisition parameters for the ProtonT1 experiment. Type%your%desired%SW%(in%ppm)%here.% Enter%a%value%for%the%relaxaHon%delay,%D1% % at%least%15%s.% Click%the%edit%(E)%buCon%for%the%VDLIST% file% %the%file%with%the%variable%delays%for% the%t1%experiment.% Type%the%appropriate%O1%value%here.% If%you%measured%the%90 %pulse,%you%can%enter%your%value% here,%or%just%accept%the%current%value%(from%the%prosol% file).% As indicated in the figure above, enter the values you noted earlier for SW and O1. You also need to enter the relaxation delay, D1. The default is 10 s, but ideally you want this to be 5 times your longest T1. The longer you make this, the longer the experiment will take. At minimum set to 15 s. If you determine that a proton has a very long T1 value, you may want to re-run the experiment with a longer D1. If you measured the 90 pulse, you can enter that value, too (or just accept the default value taken from the Prosol file). Next create the list of variable delays to be used in the experiment. Click on the E (edit) button for the VDLIST (Variable Delay list), which will open the following dialog box shown to the right. Enter&10&values,&with& more&values&closely& spaced&at&short&8mes,& where&the&exponen8al& curve&will&be&changing& fastest.& The VDLIST file (named t1delay) will have delay values from the last person who saved this file. The T1 experiment assumes When&done,&save&the&file&(File%>% Save),&then&close&the&VDLIST& window&(file%>%close)& you will enter 10 values. You can enter more (or less), but if you do you will have to set one more parameter to tell the instrument how many values are in your list (see below). The values shown in the figure are fine, but I like to go out a little longer if I m not in a hurry, especially if you expect some long T1s. You want the values to be closely spaced at the short delays, where the exponential curve changes most rapidly. Example suggested values: 0.01, 0.06, 0.12, 0.3, 0.7, 1.3, 3, 7, 15, 25.

The Inversion-Recovery Experiment page 3 Save the VDLIST file (File > Save) and close the VDLIST window (File > Close). [Note: Typically, all users are using the same VDLIST file (named t1delay, as indicated in the figure above). If this Icon experiment will not be running for some time and someone comes along and changes the VDLIST, they may overwrite the values you entered. You can give your VDLIST file a unique file name, if desired (e.g. t1delay-dave), and then there is little risk that someone would overwrite the list before your experiment runs.] 6. Close the Parameter window and return to IconNMR by clicking on the Return to IconNMR button. 7. If you entered more or fewer than 10 delays in your VDLIST, you need to change TD (number of time domain points) in the F1 dimension to match the number of entries in your list. To do so, along the top menu select Parameters > Edit all Acquisition Parameters. As before, you ll be bumped into TopSpin and a more extensive list of parameter will appear. Here you can enter the correct TD value in the F1 dimension. Then, return to IconNMR. Click%here%to%return%to%IconNMR,% aser%you%entered%the%f1% dimension%td%value% Enter%the%number%of%entries%in% your%vdlist%file.% 8. Submit your sample to be run, and start icon, if its not already running. Processing the T 1 data with TopSpin 1. In TopSpin, open your T1 experiment file. 2. Type rser 10 (or the number that corresponds to your last VDLIST entry, if you entered more or fewer than 10 delays). Type ef to process your spectra. Spectrum 10 (or your last spectrum) will be displayed, and this spectrum, with the longest variable delay, needs to be phased positively. 3. Phase this spectrum manually: click on the Process tab in the TopSpin Menu. Click the Adjust Phase button. Your spectrum will probably be 180 out of phase, unlike the one below.

The Inversion-Recovery Experiment page 4 Zero:order% phase% correc;on% First:order% phase% correc;on% Apply%and%save% the%phase%to% all%spectra% Move%cursor%on%peak% on%one%side%of%the% spectrum%or%the%other,% right%click,%and%select% Set$Pivot.% The%Pivot% Set the red pivot to a peak at one side of the spectrum or the other by moving the cursor to that peak, right click, and select Set Pivot. The red pivot should now be on that peak. Focusing on the peak with the pivot, put the cursor on the zero-order phase correction button and, while holding down the left mouse button, slide up or down to phase this peak positively (similar to what you do in MNova, to manually phase). Now, without moving the pivot, focus on the peaks at the other end of the spectrum and apply the first-order phase correction by placing the cursor over the First-order phase correction button and, while holding down the left mouse button, slide the cursor up or down to phase these peaks. Once phase properly, click on the button shown to the right, to save the 2-D phase values for all of the spectra in your file. The spectrum will go back to being unphased, but don t worry about it, the phases are saved for the 2D spectrum processing to come. 4. Click on the button shown to the right to exit the phase routine. 5. Click on the button to go to the 2D spectrum display. 6. Click on the arrow in the button and select the Process Only F2 Axis (xf2) option. You ll see a 2D spectrum something like the one shown below

The Inversion-Recovery Experiment page 5 The T 1 Calculation 1. Click on the Analyse button on the TopSpin menu bar, then click on the TopSpin T1/T2 Module button. Note: the flow of buttons change, as shown below. Note: A series of dialog boxes will appear as the following steps are executed. Read the messages and follow the instructions. 2. Click on (Extract Slice) and a dialog box appears: 3. Click on Spectrum, then another dialog box appears:

The Inversion-Recovery Experiment page 6 4. Input the slice number of your last spectrum (10, unless you had more or fewer entries in your VDLIST. Click OK. Spectrum 10 appears. 5. Click on (Define Peaks/Ranges). A dialog box appears: 6. Click Manual Integration. Another dialog box appears:

The Inversion-Recovery Experiment page 7 7. Click OK, then a spectrum appears with the Integral routine activated. 8. Use the cursor to define the integral regions in the same way you would in MNova. Then click on the button: Another dialog box appears: Select Export Regions to Relaxation Module and.ret. 9. Click on (Relaxation Window), and a dialog box appears: Click OK. A plot will appear.

The Inversion-Recovery Experiment page 8 10. Click on (Fitting Functions) to curve fit all the datasets. Another dialog box appears. Click Close. 11. Click on (Start Calculating). Another dialog box appears. Click Close, and a plot appears. 12. Select Area for Fitting Type (see next page).

The Inversion-Recovery Experiment page 9 Select& Area &for& Fi.ng&Type.& 13. In the upper left corner of the data display, above, click on the button, to calculate all the regions selected. The T1s will be displayed in the Brief Report. 13. Click on (Display Report) for a detailed report. This report can be printed from the File menu. Note: saved VDLISTS are stored in: C:\Bruker\TopSpin3.2\exp\stan\nmr\lists\vd