FACSLSRFortessa SORP QUICK REFERENCE GUIDE

Similar documents
FACSAria Fusion SORP QUICK REFERENCE GUIDE

UNC Flow Cytometry Core Facility LSR II Operation 2018

Canto II Operation Protocol

BD FACSDiva Software Quick Reference Guide for BD FACSCanto Systems with Loader Option

LSRFortessa Guide. Cell Evaluation and Therapy Core HO324

Flow Cytometry & FACS Core Facility

Unscrew and remove the cap from the waste tank

Log into the computer using your CU Denver username and password. Log into the DIVA software using your CU and Username and the password flow

FACSLyric TM General User instructions

Setup protocol for Firefly particles. On BD LSR II and LSRFortessa cytometers with single tube loader function

BD FACSMelody Cell Sorter Quick Reference Guide

Setup protocol for Firefly particles. On BD FACSCanto cytometers with single tube loader function

CyAn ADP Guide. Starting Up

FACCalibur Users Guide

BD CellQuest Pro Acquisition Tutorial

NovoExpress Software Guide. Software Version 1.2.4

Gallios TM Quick Reference

USER GUIDE TO BECKMAN COULTER FC500

BD Influx Standard Operation Protocol - Basic Operation

Release Notes Life Technologies Attune NxT Software v2.2

BD CellQuest Pro Software Acquisition Tutorial

Release Notes Thermo Fisher Scientific Attune NxT Software v2.4

Standard Operating Procedure For: Independent End User Training and Maintaining Optimal Instrument Performance:

Release Notes Life Technologies Attune NxT Cytometric Software v 2.0.0

Release Notes Life Technologies Attune NxT Software v2.3

Applied Biosystems Attune Cytometric Software version 1.2 Release Notes

Customer ( you ) must review these notes prior to installing or operating the Attune NxT Software version ( Attune Software v1.1.0 ).

Tutorial: FCAP Array Software with BD FACSArray Bioanalyzer

FACScan 4 and 5 Color Upgrade Rainbow Users Guide

JMJ-Group Microbiology BMB SDU

CELLQuest Acquisition Tutorial

Standard Operating Procedure For: Independent End User Training and Maintaining Optimal Instrument Performance:

Quick Start Guide: COLLECT

The Danish Stem Cell Center (DanStem) Danish Stem Cell Center Flow Cytometry Core Facility. LSR Fortessa Guide

Science is hard. Flow cytometry should be easy.

BD CellQuest Pro Analysis Tutorial

Hacking FlowJo VX. 42 Time-Saving FlowJo Shortcuts To Help You Get Your Data Published No Matter What Flow Cytometer It Came From

Attune NxT Acoustic Focusing Cytometer

FloMax Software for Cytometry. Operating Manual Instrument Control and Acquisition. FloMax Operating Manual Instrument Control and Acquisition

Technical Note. Display of BD Accuri C6 Software- Generated FCS 3.0 Files Using FlowJo for PC and Mac v7.6. Introduction

MACSQuantify Software 2.11

SPA III Quick Reference Guide for Custom Panels

Import Files from Other Applications

BD Lyoplate Human Screen Analysis Instructions For analysis using FCS Express or FlowJo and heatmap representation in Excel 2007

STANDARD OPERATING PROCEDURE FOR THE LSR II

Flow Cytometry Analysis Software. Developed by scientists, for scientists. User Manual. Version Introduction:

LEGENDplex Data Analysis Software Version 8 User Guide

FlowJo Software Lecture Outline:

Flowlogic. User Manual Version GraphLogic: Developed by scientists, for scientists. Graphing and Statistical Analysis.

Flow Cytometry/FACS Bio-Rad S3e Cell Sorter

BD Accuri C6 with single tube loader protocol. For multiplex assays

Offline Flow Cytometry Analysis Software: Importance of Scaling and Data Visualization

BD Multiwell AutoSampler Additional Features Tutorial

Gallios Flow Cytometer

BD High Throughput Sampler User s Guide

Basic Tutorial. Data Analysis Software for Flow Cytometry Windows/Mac user documentation

Spherotech, Inc Irma Lee Circle, Lake Forest, IL

Flowlogic. User Manual. Developed by scientists, for scientists. Version Flow Cytometry Analysis Software

Data acquisition and analysis of MACSPlex Exosome Kit using the MACSQuant Analyzer Express Mode in MACSQuantify Software version 2.

Cytomics FC 500 Quick Reference

North America/International Japan Europe

EditGrid Excel Plus Links

FX500 Exchangeable Fluidics Cell Sorter

Offline Flow Cytometry Analysis Software: Importance of Scaling and Data Visualization

Data acquisition and analysis of MACSPlex Cytokine Kits using the MACSQuant Analyzer Express Modes in MACSQuantify Software version 2.

APS (Automatic Population Separator)

A Bright Cytometer Easy to use

LSM510 Confocal Microscope Standard Operation Protocol Basic Operation

Comparing technical specification documents

User Manual. For Windows operating systems Excelsior Blvd. Suite Pécs St. Louis Park, MN Kedves street 20

NovoSampler Pro Operator s Guide

Hands-On Lab. Lab: Developing BI Applications. Lab version: Last updated: 2/23/2011

1. INTRODUCTION to Object Storage

Cytometry Data Analysis in FlowJo V10. Timothy Quinn Crawford, PhD Application Scientist FlowJo, LLC

Operating the COULTER COUNTER

Tyler Dashboard. User Guide Version 5.8. For more information, visit

xponent Version 4.2 RELEASE NOTES

Contents. Common Site Operations. Home actions. Using SharePoint

DIAplex. DIAplex Pro 1.0 Software. Instructions for use. Fast Track Your Research. For research use only

Agilent Feature Extraction Software (v10.5)

Instruction Manual XLComm Data Acquisition Software For XL Series Meters

Roche LightCycler 480

Tutorial: Create an Excel Dashboard. Overview. **** Download the Example Dashboard ****

BD Spigot to BD FACS Sortware Conversion Guide

BioRobot 9604 Operating Procedure

MIS 0855 Data Science (Section 006) Fall 2017 In-Class Exercise (Day 15) Creating Interactive Dashboards

Instructions for Use. COULTER EPICS XL Flow Cytometer COULTER EPICS XL-MCL Flow Cytometer SYSTEM II Software. PN CA (September 2010)

0Acknowledgement. This application has been developed with the cooperation of the Inter-agency Group for Child Mortality Estimation.

Agilent MassHunter Workstation Software Report Designer Add-in

The walkthrough is available at /

DataPro Quick Start Guide

Standard Operating Procedure for the HPLC-ICP-MS System By Celina Dozier Fall 2017

Designing a Web Page Using MS SharePoint Designer Reference Manual

Table of Contents RURO, Inc. All Rights Reserved

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Luminex Prima

Data acquisition and analysis of MACSPlex Cytokine Kits using the MACSQuant Analyzer Express Modes in MACSQuantify Software version 2.

Table of Contents RURO, Inc. All Rights Reserved

UMHS Financial Systems Workspace & Smart View Templates

Hands-On Lab. Developing BI Applications. Lab version: Last updated: 2/23/2011

CARE USER MANUAL REVISION MAY 2017

Transcription:

FACSLSRFortessa SORP QUICK REFERENCE GUIDE INSTRUMENT: 1. The computer is left on at all times. Note: If not Username: Administrator Password: BDIS 2. Unlock the screen with your PPMS account (UTSW username and password) 3. Turn ON the "Cytek" Fluidic Manager 4. Check sheath and waste levels. If emptying waste, add 0.5L household bleach + ~500µL of Anti Foam (Sigma) to the new waste tank. 5. Launch Tera Term, Click on Serial and make sure the Port is on COM1 and click OK First user of the day 6. Turn on cytometer (Green button-right side). 7. Then Launch BD coherent 3 (make sure all lasers windows are there and lasers have their power ON) 8. Allow ~20 minutes for lasers to warm-up. Continue while waiting. 9. Launch FACSDiva Software and log in at the prompt with your ID and password. 10. Move SIP arm to side. Remove H2O tube. Prime x2. 11. Clean the fluidics: Tubes: half tube with arm on the side + 5min arm under the tube for each of these solutions: 12. Standby. CST: Run CS&T if they haven t been run on the same day 1. Cytometer > CS&T. 2. Verify lot number. Run on Low + 5 turns of the fine adjustment knob. First user of the day Note 1: Failed CS&T is NOT acceptable. Note 2: 1 drop CS&T beads in 300uL 1

FOR A NEW ANALYSIS EXPERIMENT: USE COMP BEADS and/or SINGLE STAINED CELLS: For a new experiment without application setting ever being done 1. Open a new experiment "Experiment>New Experiment. Select the experiment titled Blank Experiment. 2. Click cytometer settings. Delete all parameters then insert experiment specific fluorophores in the order you will run them for compensation purposes. 3. Create a new specimen by clicking the New Specimen button on the Browser toolbar 4. Expand the new specimen. 5. Place the Acquisition Pointer on Tube_001. Adjusting Area Scaling 6. Select the H checkbox in the "Parameters" tab to select height for each parameter. 7. In the Browser, right click Global Sheet1 and select Apply Analysis Template 8. In the Template dialog, select the Area Scaling worksheet and click OK 9. Adjust histograms and statistics according to fluorophores used (Use one fluorophores per laser). 10. Install an unstained sample onto the loading port and click Load in the acquisition dashboard. Note: sample will be sucked even as soon as you push run on the instrument 11. Adjust the FSC and SSC voltages to place the particles on scale. Note: Keep record of FSC and SSC voltages, if using compensation beads. 12. Adjust the P1 gate around the population of interest in the FSC vs SSC plot 13. Click the Laser tab in the Cytometer window and adjust the FSC Area Scaling factor until the FSC-A signal matches the FSC-H signal (A in the +10% of H value). Increase the area scaling factor if the FSC-A signal is lower than the FSC-H Decrease the area scaling factor if the FSC-A signal is higher than the FSC-H 14. Unload the unstained tube. 15. Load the compensation tubes Note: Adjust FSC PMT voltage in the parameter window if you are using compensation beads. 16. In the Laser tab adjust area scaling for each laser (being used) with appropriate fluorophores until the A signal matches the H signal (A in the +10% of H value). 17. In the Inspector, clear the checkbox for height for all parameters. (Uncheck the H!!!) 2

Optimizing PMT Voltages 18. Right-click Cytometer Settings > Application Settings > Create Worksheet. A sheet titled Global Sheet2 opens. 19. Load the unstained control sample onto the cytometer. Note: If compensation beads have been used, enter FSC and SSC voltage recorded earlier for cells under the parameters tab 20. Adjust the FSC threshold if needed. 21. Adjust the appropriate fluorescence voltage to center the negative population in the grey zone and around the + sign. Do not record. Note: However, lowering PMT voltage from the setting established by CS&T might affect the resolution of dim populations. 22. Load the fully stained sample to verify that all fluorescent parameters are on scale. If a positive population is off scale, lower the PMT voltage for that parameter until the positive population can be seen entirely on scale 23. Unload the fully stained sample from the cytometer. 24. If you are going to run the same type of experiments later on same cells, same staining-; Right-click Cytometer Settings > Application Settings > Save. a) Name the experiment including your initials and the date. b) Click Yes to apply the optimized threshold, if applicable. c) The newly saved Application Settings can be found under Cytometer > Catalogs to view or delete when no longer in use. d) The settings are now optimized for this new experiment and are relative to current CST. Note 1: Application settings are relevant until a new baseline is performed (~every 6 months). Note 2: Even though the saved experiment template contains Application Settings, these Application Settings are not automatically synchronized with current CS&T every time a new experiment is created from the template Note 3: When you return to perform a similar experiment and you want the setting to be relevant to previous runs. For a new experiment with recalling an existing application setting 1. Import the experiment template or create a blank experiment 2. Right-click Cytometer Settings > Application Settings > Apply 3. Select your application setting and Ok -if necessary click on Overwrite and Yes- If the confirm dialog appears. Cytometer Settings is now called Application Settings in the Browser window 3

CALCULATING COMPENSATION 25. Create compensation tubes including a tube for unstained cells/beads under Experiment > Compensation Setup > Create. 26. A specimen compensation controls has been created. Expand it and select Unstained tube; Normal worksheets have been created for each compensation tubes. 27. If compensation beads are used, change FSC & SSC voltages to the beads value. 28. Acquire and record unstained and single stained compensation samples. 29. Check the gates for all tubes, excluding outlier events as needed. 30. Calculate compensation under Experiment > Compensation Setup > Calculate Compensation. Choose Link & Save. 31. Turn FSC, SSC voltage down to the noted value for cells when switching from compensation beads to cells. ACQUISITION: 1. Place the Acquisition Pointer on Tube_001. 2. Click to go back on Global Worksheet 3. In the compensation Tab, make sure enable compensation is selected 4. Create a new Global Worksheet (#3) by click on the button in the browser window Develop the Global Worksheet with: a. Create your plots (do plots, contour plots and/or histograms), stats (Median Fluorescence Intensity, CV, etc ) and Population Hierarchy for your analysis. b. Create gates (Polygonal gates, squares, quadrants or intervals) to analyses correctly your data IF USING SINGLE TUBES: 5. Add additional tubes to the experiment. 6. Experiment > Experiment Layout. a. In Labels Tab Change tubes names and the axis labels. b. In Keyword Tab Add Patient ID & Sample ID. c. In Acquisition Tab. Click Global Worksheet column header then select Global Sheet 3 on the Global Worksheet pull-down tab. Select the number of events to acquire as well as the stopping gate by selecting the appropriate column header first. 7. Set up the global worksheet and turn on the bi-exponential display for the fluorescent log plots. 8. Run a fully stained sample and setup gates. 9. Make sure the "tube probe" is installed on the machine (Long probe) 10. Run tubes. 11. Record data. IF USING HTS (Plate Loader): 5. Experiment > New plate (V, U or flat bottom). 6. Select appropriate wells and create specimen(s) 7. Enter the appropriate setting (Speed, Volume, number of mixes and Wash volume) 8. Make sure the "HTS probe" on and the HTS connector installed on the probe (Short probe + screwing Brown HTS connector) 9. Experiment > Experiment Layout. a. In Labels Tab Change tubes names and the axis labels. b. In Keyword Tab Add Patient ID & Sample ID. 4

c. In Acquisition Tab. Click Global Worksheet column header then select Global Sheet 3 on the Global Worksheet pull-down tab. Select the number of events to acquire as well as the stopping gate by selecting the appropriate column header first. 10. Set up the global worksheet and turn on the bi-exponential display for the fluorescent log plots. 11. Run a fully stained sample and setup gates. 12. Click on Run Plate. Note 1: Export the Experiment Template to the hard drive for future use: File > Export > Experiment Template. Name, etc Note 2: Even though the saved experiment template contains Application Settings, these Application Settings are not automatically synchronized with current CS&T every time a new document is created from the template. When you return to perform a similar experiment and you want the settings to be relevant to previous runs, import the experiment template, right-click Cytometer Settings > Application Settings > Apply. If the Confirm dialog appears and you have a new set of compensation tubes, click Set all compensation to zero and proceed with step 9 above. Alternatively, you can keep compensation values from a previous run or Duplicate Experiment Without Data, but in the latter two examples the settings won t be relative to the current CST EXPORTING DATA: 1. For single tubes: Export > experiment to your computer folder. For HTS: Export > FCS > 3.0 to your computer folder 2. Connect to your online tools (= Cytobank.org/cytobank with your personnel username and password) to upload the folder on this online server. CLEANING BETWEEN USERS: 1. Clean the fluidics: Tubes: 1min with arm on the side + 5min arm under the tube for each of these solutions: Plates: Run the cleaning plate (4 wells of contrad, 2 wells of beach and 2 wells of dih2o) 2. Wipe down workspace. 3. Check sheath and waste levels. 4. Logout FACSDiva software 5. Logout of PPMS DAILY SHUTDOWN: 1. Clean the fluidics: Tubes: half tube with arm on the side + 5min arm under the tube for each of these solutions: Plates: Run the cleaning plate (4 wells of contrad, 2 wells of beach and 2 wells of dih2o) 2. Wipe down workspace. 3. Check sheath and waste levels. 4. Quit FACSDiva software, Laser software (BD Coherent 3) and Tera Term 5. Turn off cytometer main power 6. Turn off "Cytek" Fluidic Manager 7. Logout of PPMS 5