LECO (TruSpec CN) - Quick Reference Guide

Similar documents
Netzsch TGA Training Notebook

Creating Instrument Calibrations (08/12/15 edits are highlighted)

Use of the Bartington Instruments MS3 magnetic susceptibility meter

INSTALLATION AND OPERATIONAL INSTRUCTIONS FOR THE OFITE MODEL 1000 VERSION 1.0

MULTIMETER: YSI-ProDSS rev. 9/23/15

How to use Excel Spreadsheets for Graphing

Getting low-resolution GC-MS data on the JEOL GCMate Introduction

Nanosight NS300 NBTC User Instructions

PANalytical X-Ray Diffractometer with PW3830 Generator. Standard Operating Procedure. Revised on 02/15/2007

Department of Instructional Technology & Media Services Blackboard Grade Book

SmartWorks MFP V4 Help File

HARMONi G3. Quick Start Guide for HARMONi G3. imac Processor/FireWire Upgrade

GX-2009 User Setup Program Operator s Manual

Prepulse Inhibition (PPI) Boxes

832 Series: Sulfur/Carbon Analysis by Combustion

CSV Roll Documentation

Statistics Winter Tutorial 1. The following computer labs have machines which support Minitab/win Release 16:

Pulsed Frequency TM. Joint PEMF Software Download, Installation and User Guide

CITO2 Installation & User Instructions

Rio Hondo Prep Computer Applications Class

ARA FTS Flow Calibrator. Operation Manual August 1, 2016

Digital Ink Pad+ User Manual

CRESCENDO /PB G3. Processor Upgrade Card for Macintosh PowerBook 1400 Series Computers. Quick Start Guide for Crescendo/PB G3

DI-5C. Prisma INSTRUMENT USER MANUAL

FG-7000L Digital Force Gauge Operation Manual

Introduction to Spreadsheets

Canto II Operation Protocol

Gas Cooling Accessory

Gallios TM Quick Reference

Checklist for the SMPS & CPC (5.400 & 5.500) PART 1 Setup of the external Plumbing (for the DMA and the CPC):

Appendix C. Vernier Tutorial

Identifying Updated Metadata and Images from a Content Provider

Title: Bruce Furnaces Semiconductor & Microsystems Fabrication Laboratory Revision: C Rev Date: 02/21/06

FROG-4000TM Quick Start Guide

Depending on the computer you find yourself in front of, here s what you ll need to do to open SPSS.

HP Pavilion dv7-6c90us Cooling fan Replacement

HOW TO USE THE EXPORT FEATURE IN LCL

How to Archive s in Outlook 2007

Appendix 1: DataStudio with ScienceWorkshop Sensors Tech Tips

Esys Braille Series. Tutorial 5 Tools ESYS FEATURES INTRODUCTION

StreetSync Basic Quick-Start

Bent Pins Inside the Sample Chamber

Equipment Standard Operating Procedure Kimberly Appel

MODEL ATV 4/6 BATTERY POWERED DIGITAL VACUUM INSTRUMENT

How to Remove Duplicate Rows in Excel

CMOS Passwords and Resources

High-Temperature Susceptometer (Kappabridge)

GREISINGER electronic GmbH. D Regenstauf, Hans-Sachs-Straße 26. T-Logg 100. T-Logg 100 E

Introduction to SPSS

Job Aid: Replacing the Field Replaceable Units (FRUs) for the Avaya G430 Media Gateway

PC Phone User s Manual

Manual Control Unit GFCD 16

Barchard Introduction to SPSS Marks

Chapter 2 Autodesk Asset Locator... 3

Operating Instructions EB 8388 EN. Series 373x Positioners. EXPERT + Valve Diagnostics

GREISINGER electronic GmbH D Regenstauf, Hans-Sachs-Straße 26

FASgov Asset Accounting Software Topic 4: Changing the FASgov Import File

Webview Help Guide. OnSite Telemetry System. Distributed by: Chart Inc th Street NW New Prague, MN USA (800)

BD Laser User Guide * IMPORTANT * PLEASE READ THE FOLLOWING INSTRUCTIONS CAREFULLY AS IMPROPER USE MAY DAMAGE THE MACHINE AND VOID THE WARRANTY.

KUHLMAN INSTRUMENT COMPANY 5000K WARRANTY

GX-6000 Data Logger Management Program Operator s Manual

Datalogging IAQ Meter Instruction Manual

CDD4 Duct Carbon Dioxide Transmitter

Procedures of the simaa 6000 (perkin elmer verison 2.5 aa winlab) Simultaneous multi-element graphite furnace Atomic absorption spectrophotometer(aa)

STEP 2: VITALSBRIDGE TO PATIENT MONITOR CONNECTION GUIDE VITALSBRIDGE AND 100

Computer Interface for Multi-Sensor Instruments

DwyerLog Logger Interface Software

User Guide. ExStik FL700. Fluoride Meter

Computer. Test Fixture SPA. Key board

Equipment Overview: Safety:

COPYRIGHTED MATERIAL. Making Excel More Efficient

CypLite Fiber Laser Cutting Head User Manual

SAMSUNG ML TONER CARTRIDGE REMANUFACTURING INSTRUCTIONS

Objective 1: Familiarize yourself with basic database terms and definitions. Objective 2: Familiarize yourself with the Access environment.

SCANNING IN COPA. To scan in the COPA system, you will need the computer attached to the scanner. Swipe Card screen

Central Processing Unit (CPU) Replacement

* IMPORTANT * REGISTERING YOUR MACHINE

11 Analytical Sequences

Removing and Replacing Parts

.txt - Exporting and Importing. Table of Contents

TopPoint User's Guide

Pulsed Frequency TM. MR7 Software Download, Installation and User Guide

CAT-100 ATMOSPHERIC CARBON POTENTIAL ANALYZER OPERATIONS MANUAL

Installing the upgrade memory kit in the G450 and G430 Media Gateway

Ritsumeikan University Researcher s Database Batch Registration Manual

KINETICS CALCS AND GRAPHS INSTRUCTIONS

A dialog box will appear informing you that you will need to restart Enterprise to see the Calendar changes. Click Ok and then restart Enterprise.

YGL 3.0 // The Basics

Version 1 1. Ultratrend DMS 5.3 Data Management System INSTRUCTION MANUAL

Pulsed Frequency TM. EM272/EM272A Software Download, Installation and User Guide

NOTIFY PLUS - MY SERVICES MANAGEMENT

Pivot Tables, Lookup Tables and Scenarios

Equipment (Tools) Needed

FG-7000 Digital Force Gauge Operation Manual

CHAPTER 1 - INTRODUCTION...2 WHAT IS ISUPERSUITE?...2 WEB-BASED APPLICATION...3 CHAPTER 2 - INSTALLATION CONFIGURATIONS...4 STANDALONE...

Welcome to FertiCom JEPOL Help

Drives. Document Part Number: This guide explains the computer hard drive and optical disc drive. October 2005

OPENING A LEADS.TXT FILE IN EXCEL 2010

User Manual. Administrator s guide for mass managing VirtueMart products. using. VM Mass Update 1.0

Statistics with a Hemacytometer

Transcription:

LECO (TruSpec CN) - Quick Reference Guide Logging On (if necessary) Coming soon Items to check prior to beginning analysis Gases Compressed Air should read 40psi. Helium and Oxygen should read 35psi on the left gauge. If the pressure on the Oxygen or Helium tanks is less than 300-400 psi (gauge closest to the tank), check with Dan or Colin. If you run out of oxygen, the LECO will simply stop running samples. If you run out of helium, you can cause expensive damage to the TCD element. Counters To access the counter window, select Configuration from the menu bar, then click on Counters... The Warn column indicates the maximum number of samples that should be run before that item needs replacement or maintenance. The Count column will increase by 1 each time you run a blank and/or standard and sample. Make sure that the Count column will not exceed the Warn value during your run. (Note: This is especially important for the crucible, as overfilling the crucible can result in expensive damage to the machine and considerable downtime.) For plant material the Warn value should read 200, for soil material the Warn value should read 65, (change the Warn values if necessary; Do not change the values in the Count column). Please check with Dan or Colin, if you are unsure whether or not to replace the crucible before running your samples. Resetting the Counters (if necessary) After changing the crucible or performing other maintenance please reset the Count to 0. To access this window, select Maintenance, then click Login Select the item that maintenance has been performed on, click OK. Click Yes if maintenance was performed; click No if maintenance was not performed. Return to the Counter window to ensure that the proper item was reset. Furnace Temp - should be close to 950 C (bottom of spreadsheet screen) Check to see if a crucible is in the furnace (same procedure used to replace crucible) (Note: failure to ensure a crucible is in place can result in costly repairs that may be charged to your account) 1. Carefully lift off sampler carousel and gently set aside 2. Loosen the 3 main screws on the loading head very carefully remove the loading head and set on it s side on top of the machine

3. Screw the lance extractor tool into the lance tube, carefully remove the tube and place in the tray next to the machine 4. Replace crucible with tongs if necessary don t forget to reset counter if you replace the crucible 5. Replace the lance tube, use your finger wipe any grit from the O-ring 6. Replace the loading head by very carefully lining up the connecting pins and stems NOTE: if seated properly you will hear the compressed gas, this is normal and will stop once the loading head is tightened 7. Carefully push down on the loading head; tighten the bottom right-hand screw first, by turning until you feel resistance, move onto the other 2 screws and stop when you feel resistance then turn each screw roughly ¼ turn, alternating among the screws until they are all tight Balance Turn on if necessary; use the 42g/205g button to switch display to 0.0000 g if necessary. Leak Checks Before running any samples it's in your best interest to check for leaks. Select Diagnostics from the main menu, then click Leak Check. Select Whole O2, click Start, wait until Passed or Failed appears in the Results column. Select the Whole He button, click Start, wait until Passed or Failed appears in the Results column. Please note: the software considers any leak smaller than 5 mm Hg to have passed. This is not correct: any leak greater than 2 mm Hg is failing. If either of the leak checks fails, contact Dan or Colin. Calibration Running the initial Blanks First, make sure the next available (i.e. empty) cell on the spreadsheet has been selected. Select Samples from the main menu; click Login. From the Sample window use the drop down arrow next to Sample Name to select Blank. The Mass must be 1.0000 (which is the default), select 6 repetitions. Using the drop down arrow at the end of the Method line select the appropriate method:

Use Plant 10ml Loop for plant samples or samples with a large amount of carbon and/or nitrogen. Use Soil 10ml Loop for soil samples or samples with low carbon and/or nitrogen (Note on Methods check the C/N values on the blackboard and/or check with Dan or Colin if you re not sure which method to use. Do not use any other method unless told differently by Dan or Colin. Enter supervisor s name or project name on the comment line and enter your name on the operator s line. The other Attributes should be left blank in most cases. Click OK to add the Blanks to the sample list. Click Cancel if necessary to close the Sample window. Four blanks should have been added to the sample list. To help keep track of your samples, change the numbers in the Location column to match the numbers on the sample carousel. Running the initial Standards Select Samples from the main menu; click Login Drift Samples. Select the appropriate method using the drop down arrow; the associated calibration standard will appear in the Drift Standards box, along with the weight range (your standards must be within this range). Select 4 repetitions. The standards should appear after the blanks on the spreadsheet. Note: The weights in the Mass column for the standards will appear with a ~ symbol, this is just to remind you of the approximate weight you should use, the actual weights will replace the reminder weights as you enter them using the Print button on the balance. To start the analysis, click on the Analyze button or press F5 to start analyzing the blanks and standards. To halt the machine after the blanks (in order to run the blank calibration), first left-click on the row containing the first standard, then right-click; select Pause from the Samples menu, choose Manually from the drop down list. A red stop sign should appear next to the first standard. This acts as a Halt on the old machine and will stop the analysis temporarily after the last blank has run. Blank Calibration If the carbon values vary less than 0.04 and the nitrogen values vary less than 0.02, you may proceed with the calibration. If not, continue running blanks until the values stabilize. After the blanks have finished running (and the machine is in pause mode), choose the blanks you wish to use for the blank calibration by highlighting the appropriate group of blanks (left-click and drag, or left-click and use control to select individual blanks). Select Configuration from the main menu, click Blank. From the General Blank window, Include both Nitrogen and Carbon if necessary, using the Include/Exclude button. Click OK. Note, unlike the old machine we have noticed that the blank values (especially for Carbon) may not appear to change (and may even be higher that 0.04); this is fine, as the blank values are automatically subtracted from the standards and samples. After the blank calibration restart the analyses by clicking the Analyze button or pressing F5.

Standard calibration Select the appropriate standards (analyze additional standards if necessary in order to achieve precise values). Select Configuration from the main menu, click Drift, from the next menu click Drift again. Include both Nitrogen and Carbon if necessary, using the Include/Exclude button. Click OK.

Running Samples * Run one blank and one standard immediately after calibration to ensure the values are correct; if they are out of range, run a few more blanks and/or standards, checking to see if they stabilize in the correct range; if the values are still in question, see Colin or Dan. Insert a Pause at the beginning and end of the calibration check. The quickest way to enter a sample is to simply type the sample name in the Name column, you can hit Enter to proceed to the next line, or wait until you have weighed out the sample and pressed the Print button on the balance. * Run a blank and/or a standard approx. every 10 samples to insure the machine is running properly. * At the end of your run, do not shut off the machine, do not close the software - change the crucible if necessary, select and download your data, and fill out the log book. Exporting Data Select the samples you want to download by clicking and dragging in the row column. Goto Samples on the main menu bar and click on Text Export Data. Enter a filename and the appropriate drive, click Save. You can use either a 3.5 floppy or jump/flash USB memory stick (the USB port is in the front of the machine under the gray Dell panel). Your data is in comma delimited format, which seems open nicely in Excel. Printing data Select the rows containing the samples you want to print. Select Samples on the main menu bar, click Print, click OK. Updated 9-26-2011 (Colin Pinney)