Vibration Sensor Type 640

Similar documents
Vibration Control Typ 663

Safety Manual. Vibration Control Type 663. Standard Zone-1-21 Zone Edition: English

PHOENIX CONTACT - 01/2008

Operating instructions (Safety-related part ATEX and IECEx) Inductive NAMUR sensors NN504A NN505A NN506A NN507A /00 08/2015

Operating Instructions (Translation) 3. Safety Information. 1. Description. 2. Explosion Protection. Supply module Type 17-21BB-170x

Providing flexible power supply and signal connection there, where it is needed even in hazardous areas for the Zones 1and 2.

Documentation KM2042. Sixteen channel digital output module with D-Sub Connector. Version: Date:

Inductive sensor with analog output BI15-M30-LI-EXI

Signal processor MS96 for flow sensors MS96-11EX0-R/230VAC

Inductive dual sensor For rotary actuators NI4-DSU35TC-2Y1X2/S933

Operating Instructions

Inductive sensor NI3-EG08K-Y1-H1341

Inductive sensor Dual sensors with extended temperature range NI4-DSU35TC-2Y1X2/S97

Magnetic field sensor Magnetic-inductive Proximity Sensor BIM-M12E-Y1X

Inductive sensor For rotary actuators NI4-DSU35TC-2Y1X2/S933

MANUAL LML-P, LML-S, LML-Ex

Diffuse reflection light scanner with background suppression. Dimensioned drawing

PHOENIX CONTACT - 05/2007

SolConeX Wall-Mounted Socket, 16 A

Documentation EM2042. Sixteen Channel Digital Output Module with D-Sub Connector. Version: Date:

PHOENIX CONTACT Pin. SUBCON-PLUS M ARCNET, MULTI/MININET (B&R), SYSTEM 2003 (B&R), P-NET SUCONET K1, K2 (EATON/Moeller),

Inductive sensor BI1.5-EG08-Y1-H1341

Magnetic Field Sensor Magnetic-inductive Proximity Sensor BIM-M12E-Y1X-H1141

PTB 01 ATEX Example / Beispiel / Exemple. Type Operating Instructions

SolConeX extra-low voltage flange socket

Inductive sensor BI1.5-EG08-Y1-H1341

Inductive sensor For rotary actuators NI4-DSU35TC-2Y1X2

Inductive angle sensor with analog output Ri360P1-DSU35TC-ELi-Exi

Operating instructions Safety sensor EX-BNS40S. 1. About this document. Content

Technical Data. General specifications Switching element function PNP NO/NC Rated operating distance s n 5 mm

PTB 01 ATEX 2064 U, IECEx PTB U. Example / Beispiel / Exemple: Type Operating Instructions

Magnetic Field Sensor Magnetic-inductive Proximity Sensor BIM-EG08-Y1X-H1341

ADAMCZEWSKI. Ordering data AD-STVEX 710 GVD Option: Factory-software configuration as per customer specifications.

Intrinsically safe batch controller Batching Master 110i

Manual Mouse Module CHALLENGER Mi-PS2 / Mi-PS2-B

ExBin-D Transducers for ExPro-B... sensors (probes) ExPro-B... Thermostats / humidistats ( C, % rh)

Documentation ZB8610. Fan cartridge for EtherCAT and Bus Terminals. Version: Date:

Inductive sensor For rotary actuators NI4-DSU35-2Y1X2-H1140

Conveyor Belt Alignment Switches from IDEM

MANUAL Surge Protectors

Operating instructions Hinge safety switch TV.S 335 TV.S About this document. Content

Operating instructions Safe AS-i input module ASIM-C-M About this document. Content

Explosion-Protected Arc-Fault Protection ExAFCI Arc-Fault Circuit Interrupter

Operating instructions Safety sensor BNS About this document. Content. 6 Disassembly and disposal 6.1 Disassembly Disposal...

Explosion proof UL certified VoIP telephone. FernTel IP Z2 UL. Tabletop unit. Manual FHF BA /12

Operating instructions Position switches EX-Z/T G/D. 1. About this document. Content

STR-3 STANDARD 3-PHASE 230 VAC TRANSFORMER CONTROLLER. Mounting and operating instructions

Electronic Temperature Controller. Instruction Manual Version

Operating Manual. English. Digital-Analog Converter DACON8-UMB.

Technical Data. General specifications. Output type Rated operating distance s n 8 mm

EX- Documentation VA 450 / VA 452. thermal mass flow sensor

Operating instructions Safety sensor BNS About this document. Content

Inductive Angle Sensor With Analog Output RI360P1-DSU35TC-ELI-EXI

PI-EX-ME-2NAM/COC-120VAC

CPCI-PS24 24V-Power Supply

PI-EX-ME-2NAM/COC-24VDC

BVS 17 ATEX E 117 X / IECEx BVS X Types 3320, 3321, 3323, 3360, 3361, 3363 AE3320, AE3321, AE3323, AE3360, AE3361, AE3363

MINI MCR-SL-RPS-I-I. Repeater power supply, active. Data sheet. 1 Description

Technical Data. General specifications Switching element function DC Dual NC Rated operating distance s n 3 mm. Short-circuit protection

Operating instructions Door handle switch BNS-B About this document. Content. 6 Disassembly and disposal 6.1 Disassembly Disposal...

Signal conditioning instruments for level switches. Conductive VEGATOR 256C VEGATOR 532 VEGATOR 631. Product Information

Inductive sensor BI5-EM18-Y1X-H1141

MINI MCR-SL-CVS NC

Operating instructions Safety sensor BNS About this document. Content

Operating Manual Extension for Vane Anemometer MiniAir6/60/64-Ex

PHOENIX CONTACT - 12/2007

MC 11 EB-2 Power supply cabinet with external bus, AC version

BVS 17 ATEX E 117 X / IECEx BVS X Types 3320, 3321, 3323, 3360, 3361, 3363 AE3320, AE3321, AE3323, AE3360, AE3361, AE3363

Pneumatic Control/Feedback

PROCESS AUTOMATION INSTRUCTION MANUAL FIELDBUS SURGE PROTECTOR F*-LBF-I1.32 F*-LBF-D1.32

Pneumatic Control/Feedback

Magnetic Field Sensor Magnetic-inductive Proximity Sensor BIM-EG08-Y1X

Operating instructions Safety-monitoring module AES 1235 AES About this document. Content

Operating Manual / IEC. Explosion-Proofed Junction Boxes, Polyester. Wallücker Bahndamm 7 DE Kirchlengern

USER MANUAL VIBRATION CONTROL RMA-POWER-BOX 107/230

Minimum cross sectional area of Conductor mm²

Documentation. KM2604 and KM2614. Four channel relay module. Version: Date:

Emergency Luminaire for Fluorescent Lamps Series EXLUX 6009

Thread length (overall length) material

Control Stations & Switches: Switches, Circuit Breakers and "Break Glass" Call Points SWE Series: 16 and 20 A Switches Increased Safety

MINI MCR-SL-PT100-UI. Active resistance thermometer measuring transducer. Data sheet. 1 Description

Operating Manual UMB ISO Converter ISOCON Order Number: 8160.UISO

Operating Instructions for Level Switch for Liquids. Model: RFS

Product and Applications Description. Application Programs. Note. instabus EIB Technical Product Information. June 2005

Operating Instruction

6173 AG GJB A0138

Metal Position Switch

DMP 333 DMP 333. Industrial Pressure Transmitter For High Pressure. Stainless Steel Sensor. Pressure Transmitter. Industrial

Operating Manual. icam100

Sensor boxes SRBC. Festo core product range Covers 80% of your automation tasks

Product manual plug, sockets and coupler 16 A.- Z0251

PROCESS AUTOMATION. MANUAL Z-System. Zener Barriers ISO9001

Redundancy Module QUINT-DIODE/40

Linear Luminaire with LED Series 6412/5

OPERATING INSTRUCTIONS

ILUMEN PID SOLUTION INDOOR OUTDOOR ILUMEN PIDBOX MINI INSTALLATION MANUAL. Ilumen PIDbox mini Version 1.5

Operating instructions Safety-monitoring module AES 1235 AES About this document. Content

PCT2000. User Manual. Precision current transducer 2000 A

Operating and mounting instructions

Application The DA03 serves as indicator of differential. Follow mounting instructions in accordance to application.

Transcription:

Vibration Sensor Type 640 Standard Zone-1-21 Zone-2-22 Instruction Manual English

Instruction Manual Vibration Sensor Type 640 Standard Zone-1-21 Zone-2-22 Edition: 12.04.10 Attention! Before Start-Up Procedure the Instruction Manual must be read and understood! Should any question arise, please contact: HAUBER-Elektronik GmbH Fabrikstraße 6 D-72622 Nürtingen Germany Phone: +49 (0) 7022 / 62393 Fax: +49 (0) 7022 / 64143 info@hauber-elektronik.de www.hauber-elektronik.de

Content 1 Safety Informations...4 2 Instruction Manual Scope...5 3 The Vibration Sensor Type 640...5 Standard, Zone-1-21, Zone-2-22 4 Intended Use...5 5 Documents and Certificates...5 6 Responsibility for the Safe Operation / Disclaimer...5 7 Application Fields...6 8 Delivery Contents...6 9 Electrical Data...7 10 Mechanical Data...8 11 Connections...9 12 Mounting and Demounting...10 12.1 Fastening at the Mounting Surface...10 12.2 Zone-2-22 - Fastening Safety Clip / Protective Cover...11 13 Installation and Start-Up...12 14 Maintenance and Repairs...12 3

1 Safety Instructions In General The safety instructions serve the protection of persons and things from damage and danger that arise from not intended use and further misuse of products especially in explosion endangered areas. Therefore read the instruction manual carefully, before working with or starting-up the product. To the operating personnel the instruction manual has to be accessable anytime. Before the starting-up or miscellaneous works with the product please check, wether all the documents are available completely. If not all the documents are committed completely or futher copies are required, they can be obtained in different languages. Our product is designed to the latest state of the art. Nevertheless there are a number of residual risks. This means that each person in the operators firm, concerned with mounting and dismounting, installation, start-up, operating or maintanance of the product, has to have read and understood the instruction manual. This means furthermore that each person in the operators firm, concerned with mounting and dismounting, installation, start-up, operating or maintanance of the product, has to be an authorized expert, familiar with the safety instructions for handling electrical components. For handling ATEX-certified products within explosion endangered surroundings the expert in addition has to be familiar with the safety instructions relevant there. Used Symbols This symbol indicates an explosion hazard. This symbol indicates a risk from electrical current. This symbol indicates a (non-safety relevant) information. 4

2 Instruction Manual Scope The present instruction manual of the Vibration Sensor Type 640 is applicable for the following variants: Standard, Zone-1-21 and Zone-2-22. The functionality of the variants is identical. In addition the variants have certifications and labellings, that allow operation in explosion endangered areas. (see chap. 7, Operation Areas) 3 The Vibration Sensor Type 640 The Vibration Sensor Type 640 ist applied for measurement of machines absolute bearing vibration, referring to DIN ISO 10816. It works to Two-Wire-System. The Measurement parameter is the root mean square (rms) of the vibration velocity, the unit is mm/s. The Sensor contains an electronic device that delivers an interference-free 4...20 ma Output-Signal, proportional to the Sensors measuring-range. 4 Intended Use The Type 640 exclusively serves for the measurement of mechanical vibrations of machines and mechanical facilities. The operation is valid exclusively within the specifications mentioned in this manual. Main areas of application: Industrial fans, ventilators, blowers, electric motors, pumps, centrifuges, seperators, generators, turbines, and similar, oscillatory mechanical equipment. 5 Documents and Certificates Following Type 640 Documents und Certificates can be consulted on www.hauber-elektronik.de: EC-Conformity-Declaration EC-Type-Examination-Certificate ATEX-Zone 1 und 21, no.: PTB 06 ATEX 1072 Statement-of-Conformity ATEX-Zone 2 und 22, no.: LU 06 ATEX 0018X 6 Responsibility for the Safe Operation / Disclaimer The correct layout of the electrical plant under conditions of explosion protection, as well as the correct switch on procedure, is the sole responsibility of the user of the plant. The current valid explosion protection rules and security regulations must be adhered to and must be under given circumstances checked by a competent person. Should the plant on the order of the user be erected by a subcontractor, the plant must only be switched on after the subcontractor has submitted an installation certificate as prove of the correct nature of the installation, according to the relevant valid regulations. The primary switch on of explosion protected plants or part of plants, as well as the subsequent switch on after major adjustments or maintenance work, must be reported to the relevant authorities by the owner. 5

7 Application Fields Variante Application Fields Labelling Standard None explosion endangered none areas Zone-1-21 Explosion endangered areas Zone 1 and 21 II 2G Ex d IIC T4 II 2D Ex td A21 IP65 T120 C Zone-2-22 Explosion endangered areas Zone 2 and 22 II 3G Ex nc II T4 II 3D Ex td A 22 IP55 T125 C 8 Delivery Contents Variant Delivery Contents Standard Vibration Sensor Type 640 Instruction Manual Zone-1-21 Vibration Sensor Type 640 Instruction Manual Zone-2-22 Vibration Sensor Type 640 Instruction Manual Safety Clip Protective Cover for M12-Plug Available Supplies Evaluation Equipment Type 656 Evaluation Equipment Type 652 Various Adapters, e.g. M8 -> M10 Allocable Mating Connector, M12, 8-pole Connection Cable, M12-Socket, 4-pole, 0,34 mm 2, L= 2 m, 5 m oder 10 m or on request Magnet Foot 6

9 Electrical Data Before Starting-Up the Sensor, the mains must be secured with a microfuse ( time delay, 32 ma, breaking capacity C)! Measuring Range: 0... 8 mm/s 0... 16 mm/s 0... 32 mm/s 0... 64 mm/s 0... 128 mm/s 0... 256 mm/s 0... 512 mm/s 0... 1000 mm/s Measuring accuracy: 5% Each Sensor has one of the listed measuring ranges. Further ones on request. Frequency range: 5 Hz...1000 Hz Output signals: 4...20 ma (Proportional to the Measuring Range) Voltage supply: 24V DC 10% Power input (max.): 25 ma Shock (max.): 1000 g Burden/Load: 500 Fusing: Microfuse (Time delay, 32 ma, breaking capacity C) typical amplitude response of vibration velocity db 10 5 0-5 -10-15 -20 1 10 100 1000 10000 Hz Fig. 1: Frequency Range 5 Hz...1000 Hz Valid Operating Temperature Ranges of all Variants Standard Zone-1-21 Zone-2-22 Ambient Temperature -20 C +60 C -20 C +60 C -20 C +60 C Measuring Head- Temperatur (at the Fastening) -40 C +85 C optional: -40 C +125 C -20 C +100 C -40 C +100 C 7

10 Mechanical Data Housing Material: Stainless Steel V2A; material no: 1.4305 M12-connector / cable gland: CuZn (brass), nickel plated Fastening: Wrench Size: 24 (hexagon), M8 x 8 Securing: Weight: ca. 150 g Protection Style: IP 67 The Sensor must be earthed via the M8 fastening (see chapt.12). Housing Dimensions and Direction of Measurement Standard, Zone-2-22 Zone-1-21 Measures in mm Connection Cable Standard: L= 2 m 69 8 9.3 59 17.5 7 50 10 8 M8 Ø28 Ø16.5 Ø5 M8 Ø28 2 M12-Plug Cable Gland Direction of Measurement Fig. 2: Housing Dimensions and Direction of Measurement: Standard, Zone-2-22. Direction of Measurement Fig. 3: Housing Dimensions and Direction of Measurement: Zone1-21 Note: Direction of Measurement = Direction of Fastening 8

11 Connections Standard, Zone-2-22 M12-Plug Zone-1-21 Connection Cable 4 3 1 2 Fig. 4: M12-Plug, 4-pole. Pin allocation see Connection Plan. Fig. 5: PUR-sheathed-cable, 4-pole, 0,25 mm², Standard: L= 2 m. Pin allocation see Connection Plan. Connection Cable Socket (Supplies) Fig. 6: Connection cable socket M12, 4-pole, 0,34 mm². Pin allocation see Connection Plan. Connection Plan for all 3 Variants Type 640 bu Evaluation- Device To avoid capacitive Coupling Interferences, the Pins 2 and 4 have to stay open resp. free! 9

12 Mounting and Dismounting Mounting and Dismounting works at and with the Sensor may only be executed by an authorized expert, familiar with the safety instructions for handling electrical components. For handling ATEX-certified Sensors within explosion endangered surroundings the expert in addition has to be familiar with the safety instructions relevant there. Before mounting and dismounting works the Sensor has to be seperated from the mains! Seperated plug and socket devices always have to be disconnected from the mains! Otherwise danger of explosion because of sparkling, when operating ATEX-certified Sensors in explosion endangered areas! The Sonsor housing must be earthed via ist fastening - i. e. via machine earth or via a seperate earth wire (PE)! 12.1 Fastening at the Mounting Surface Preconditions Mounting surface clean and flat, i.e. free from paint, rust, etc. Threaded hole at the Mounting surface: Depth: 10 mm Thread: M8 Tools and Materials Allen wrench, SW24 Working Steps Tighten Sensor friction-locked into the threated hole of the mounting surface. To obtain exact measuring values, the Sensor has to be tighten friction-locked at the mounting surface! Avoid Auxiliary Constructions! If unavoidable, implement it as stiff as possible! 10

12.2 Zone-2-22 - Fastening Safety Clip / Protective Cover The operation of variant Zone-2-22 is not permitted without the Safety Clip, to avoid accidently disconnecting the plug-in connection! Otherwise danger of explosion because of sparkling, when operating in explosion endangered areas! Fastening Safety Clip 1. Plug in the connection cable socket into the M12-plug completely. (Pay attention to the code cam!). 2. Tighten firmly the lock-nut of the connection cable socket by hand. 3. Fasten the safety clip against accidental disconnection of the plug connection: 1. Put both shell halves of the safety clip around the plug connection. 2. Press together by hand both shell halves of the safety clip until the catch lock snaps in. 3. Put the arrow connected to one shell halve around the cable, then stick it through the eye on the other end, so that the notice sign is readable alongside the cable. Half Shell Arrow and Eye Sign Fig. 7: Safety Clip Fig. 8: Fastened Safety Clip Fastening Protective Cover After disconnecting the plug connection the protective cover has to be mounted! Disassemble the safety clip and mount the protective cover: 1. Disconnect mains / electric circuit. 2. Separate both shell halves of the fuse clip with a screw driver. 3. Fasten protective cover and skrew it tightly onto the sensor plug. Fig. 9: Protective Cover Fig. 10: Fastened Protective Cover 11

13 Installation and Start-Up Installing and starting-up the Sensor may only be executed by an authorized expert, familiar with the safety instructions for handling electrical components. For handling ATEX-certified Sensors within explosion endangered surroundings the expert in addition has to be familiar with the safety instructions relevant there. Prior to starting-up the Sensor, the mains must be secured with a microfuse ( time delay, 32 ma, breaking capacity C)! The connection cable and possible extension cables must be protected against electrical influenzes and mechanical damages. Here local regulations and commissions absolutely have to be considered. 14 Maintenance and Repair Repairing the Sensor may only be executed by an authorized expert, familiar with the safety instructions for handling electrical components. For handling ATEX-certified Sensors within explosion endangered surroundings the expert in addition has to be familiar with the safety instructions relevant there. Prior to repair and cleaning works the Sensor has to be seperated from the mains! Seperated plug and socket devices always have to stay disconnected from the mains! Otherwise danger of explosion because of sparkling, when operating ATEX-certified Sensors in explosion endangered areas! Defective connection cables immediately have to be replaced! Otherwise danger of explosion because of sparkling, when operating ATEX-certified Sensors in explosion endangered areas! A defective control has to be changed completely! Note: The Type 640 and ist variants are maintenance free! Errortable Error Cause Activity No Measured Value (4-20 ma) No Power Supply Check Power Supply and/or Connection Cable Connection Cable interrupted Replace Connection Cable Fuse defective Replace Fuse Connection wrong Polarity Provide correct Polarity Sensor defektive Replace Sensor Measured Value wrong Sensor mounting not friction-locked Mount Sensor friction-locked Sensor mounting at wrong position Mount Sensor at correct postion 12