Operation MANUAL. CARBOMETER - CO2 concentration measurement unit. LAMBDA CARBOMETER - CO 2 concentration measurement unit

Similar documents
OPERATION MANUAL. REDOX Measurement Unit. LAMBDA REDOX Red-Ox Potential measurement unit

USER MANUAL. FNet Fermentation Control Software

OPERATION MANUAL. MEGAFLOW Peristaltic Pump Tubing Pump. LAMBDA MEGAFLOW Peristaltic / Tubing pump

OPERATION MANUAL. Peristaltic Pump Tubing Pump. LAMBDA HIFLOW Peristaltic pump / Tubing pump

USER MANUAL. SIAM Industrial fermentation software. LAMBDA MINIFOR Fermenter-Bioreactor

Gas detection just got personal

DMA 500. Density/Specific Gravity/ Concentration Meter. ::: Unique Density & Concentration Meters

Micro-Ohmmeter. Digital 200A. Model Measure low resistance with high accuracy

Pressure transmitter for highest pressure applications Up to 217,500 psi (15,000 bar) Model HP-2

KCD-HP. KCD-HP200x, 300X. [Figures] Top : 1% Sensor probe Bottom left : 10% / 20% Sensor probe Bottom right : Holding bracket(optional) Measurement

Analog Input and Output Options for Syringe Pumps

The Art of Measuring. Compact Transmitter with Modbus Interface. MemoRail Modbus

DMA 500. Density/Specific Gravity/ Concentration Meter. ::: Unique Density & Concentration Meters

DCT 531. Industrial Pressure Transmitter with RS485 Modbus RTU. Stainless Steel Sensor

Micro-Ohmmeter. Digital 200A. Model Measure low resistance with high accuracy

HPS-M -2 DIFFERENTIAL PRESSURE TRANSMITTER. Mounting and operating instructions

RST ROOM TEMPERATURE TRANSMITTER. Mounting and operating instructions

Carbon Monoxide Sensor - ModBus

High-end pressure controller Model CPC8000

Manual DTM Digital Pressure Transmitter

User's Guide. Programmable DC Power Supply 200 Watt (40 Volts / 5 Amps) Model Introduction

EE80 Series. HVAC Room Transmitter and Switches for CO 2, Relative Humidity and Temperature EE80. Applications. Typical

2.2 Main technical parameters. Detection range. 0~5%vol(0~100%vol optional) Warm-up time. SPAN <±500ppm Long-term drift

Reference Guide METTLER TOLEDO WMH Weighing Platforms

The PM1000 series is a universal 4 digit LED plug-on display for transmitters with 4-20mA 2 wire output and fitted with DIN43650 connector.

Micro calibration bath Models CTB , CTB

TEC-BT temperature controller User manual

High-Precision Pneumatic Pressure Controller Series CPC8000-L/-X

SF10 SF10. Features: The SF10 laser altimeter is ideal for automated landings and precision hovering.

RADIOMETER PHOTOMETER L-420 INSTRUCTION MANUAL

Thermometer T12. Precision Multichannel Thermometer System. Typical applications: REFLECTING YOUR STANDARDS

Versatile Transmitter Series For Intelligent Measurement Solutions. INGOLD Leading Process Analytics

RS-FS-N01 Wind speed transmitter user's Guide (485type)

Model KB 400 Cooling incubators with compressor technology

Model VD 23 Vacuum drying chambers for non-flammable solvents

DSTHM-2 COMBINED T AND RH DUCT TRANSMITTER. Mounting and operating instructions

VAT series controller and transmitter. Manual

Features. Description. Applications

GMM220 CO 2. Transmitter Modules. Introduction. Mounting

Micro-Ohmmeter. Digital 200A. Model Measure low resistance with high accuracy. Technical Hotline: (800)

Digital Indicator Solutions

USB pressure transmitter model CPT2500 with model CPA2500 USB adapter and USBsoft2500 software

Assembly Instructions scancontrol 2700/2750 (500)

Matrix printer uni-bit. Application areas

OEM Digital NDIR CO 2

Data sheet * * Digital 2-channel transmitter for direct connection to bus-capable automation devices. Design and mode of operation

PHYSICS Digital Measuring Devices

Sensor for car parks and tunnels

Micro-Ohmmeter Model 6290

NIR Moisture Analyzer KB-30

OEM Digital NDIR CO sensor

FUTURA Egg-Shell-Tester V. 2.0 User Manual

Mass Flow Meters, Controllers and Manifolds

RXTP ROOM TEMPERATURE

(see also technical drawings) Width with inlay

ProLab 1. ProLab 1000: ph for the professional. ph Meter ProLab 1000: Superior technique in a classy metal body. Precise measurements

SVG3 (Basic Equipment)

Thermoelectric Cooler Controller TED1000

High Precision Quad Output DC Power Supply

ADC7520 SERIES. 1600W Battery Chargers and Power Supplies

Installation and operation manual ReciFlow Gas

MH-Z16 Intelligent Infrared Gas Module

DRTS 3 PLUS Advanced Protection Relay Test Set and Measurement System

Temperature dry-well calibrator Models CTD9100-COOL, CTD , CTD , CTD

Pressure sensor For highest pressure applications to 15,000 bar Model HP-2

Model VDL 115 Vacuum drying chambers for flammable solvents

Series Watt DC Power Supplies

DMT PERFORMING UNDER PRESSURE DMT. Applications. Special features. Description 1/9

CEM M-RS485 INSTRUCTION MANUAL (M014B A)

Part No. Z , IA Jul OPERATION MANUAL. High Voltage Digitalmeter A

Temperature dry-well calibrator Models CTD , CTD

Micro Motion Model 7950 and Model 7951 Signal Converters

Part No. Z , IA Nov OPERATION MANUAL. High Voltage Digitalmeter A

Peristaltic Pumps. Doser. Sysmatec. Products Multiflow Pump. Hiflow / Maxiflow. Laboratory instruments

RS-FX-N01 wind transmitter user's Guide (485type)

The sensor fulfills the specifications of the EMC requirements, if the instructions in the manual are followed.

How to build a Conductivity Measurement System

HIGH-VOLTAGE DC METER

Temperature dry-well calibrator Models CTD9100-COOL, CTD , CTD , CTD

D Issue A. Instruction Manual. Active Strain Gauge

AM3 Humidity. Temperature. Dew Point Indicator INSTRUCTION MANUAL

PRESSURE TRANSMITTER

Meter Centraliser Alarm Centraliser Unit LM50-TCP (Code M31521) INSTRUCTION MANUAL (M A)

Dräger Polytron 8200 Stationary gas detector (CatEx)

A C C S for Road Tunnels. NOVICOM CO/NO/Visibility Monitor Contact us for Information about:

Linear Position Transducer HLT 1000-R2

DXTH DUCT SENSOR / SWITCH FOR TEMPERATURE AND HUMIDITY. Mounting and operating instructions

Pneumatic High-Speed Pressure Controller Model CPC3000

Pneumatic Precision Pressure Controller Model CPC6000

TP32MTT.03 TP32MTT [ GB ] Probes for soil thermal profile measurement

GOLANDER PUMP. Operation Manual. For Microflow Variable-Speed Peristaltic Pump BT50S/BT102S.

Distance and Position Sensors

RXTH DUAL ROOM SENSOR / SWITCH

GA-12. madur Polska sp. z o.o.

High-end pressure controller Model CPC8000

Octostrip ORDERCODE 42220

703 Laboratory Conductivity Meter

POSICHRON PCQA Mounting Sets

ATS-16 HV USER MANUAL. Automatic Transfer Switch 16A / 230Vac V090318

Low Pressure Sensors 0.1 H2O to 3 psi. 5 Minute Product & Pricing Guide

Transcription:

CARBOMETER - CO2 concentration measurement unit LAMBDA CARBOMETER - CO 2 concentration measurement unit

LAMBDA CO 2 -meter: CO 2 concentration measurement unit LAMBDA CARBOMETER (CO 2 -METER) allows the measurement of the concentration of CO 2 gas (0 100%) in the outgas / exit gas of the MINIFOR fermentor and bioreactor and the digital transfer of the data to the PC achieved through its RS-485 interface. The measured data can be visualized and recorded for example by the industrial fermentation software SIAM. It does not require any additional connector, cable or side neck. Connected to the glass outgas condenser (or Peltier outgas condenser) of the MINIFOR laboratory fermentor and bioreactor, the CO 2 -METER measures the concentration of the CO 2 gas produced. With SIAM industrial fermentation software, the RQ (respiratory quotient) from the CPR and OUR (obtained from LAMBDA O 2 -METER) during the fermentation process can be calculated. LAMBDA Laboratory Instruments LAMBDA Laboratory Instruments develops innovative, high quality lab-scale instruments with an excellent price to performance ratio for biotechnology, microbiology, food and agricultural, chemical and pharmaceutical industries, research and development as well as for general laboratory and research applications. LAMBDA MINIFOR highly innovative and compact fermenter/bioreactor system for laboratory scale fermentation and cell cultures LAMBDA OMNICOLL fraction collector-sampler for unlimited number of fractions LAMBDA PRECIFLOW, MULTIFLOW, HIFLOW, MAXIFLOW AND MEGAFLOW peristaltic pumps reliable, precise and extremely compact LAMBDA SAFETY POWDER DOSER AND HI-DOSER allows automatic feeding of powders without spoon. Safe operation with hazardous material (GLP) LAMBDA VIT-FIT (HP) polyvalent syringe pump with extremely robust mechanics programmable infusion and filling from micro syringes to large volume syringes of 150 ml without adapter LAMBDA MASSFLOW precise gas flow measurement and control with data acquisition option LAMBDA PUMP-FLOW INTEGRATOR with LAMBDA pumps and doser allows the visualization and recording of the pumped volume www.lambda-instruments.com 1

TABLE OF CONTENTS 1 SETTING UP THE CO 2 -METER... 3 1.1 Overview of CO 2 -METER connections... 3 1.2 Power supply... 4 1.3 Measurement of CO 2 concentration in the outgas / exit gas... 4 2 PC CONTROL & SOFTWARE... 4 2.1 Setting up the address of CO 2 -METER... 4 2.2 PC connection... 5 2.3 SIAM industrial fermentation software... 5 3 CALIBRATION PROCEDURE FOR THE CO 2 -METER... 5 3.1 Zero setting... 5 3.2 Sensitivity setting... 5 4 TECHNICAL SPECIFICATIONS... 6 4.1 General specification... 6 4.2 Remote control (Inputs/outputs)... 6 5 ACCESSORIES... 7 6 GUARANTEE... 7 7 APPENDIX... 8 7.1 RS communication protocol of LAMBDA CO 2 -METER AND O 2 -METER... 8 7.2 RS-connection scheme... 9 www.lambda-instruments.com 2

1 SETTING UP THE CO 2 -METER 1.1 Overview of CO 2 -METER connections Table 1 Connection overview of CO 2 -METER: Stand-alone instrument, with MINIFOR fermentorbioreactor or PC software CO 2 -METER (stand-alone) CO 2 -METER with MINIFOR CO 2 -METER with PC software Remote control - Pump remote control (analog and digital) cable (8 poles) RS-interface - RS-485 interface (incorporated with CO 2 -METER) PC connection - Please refer MINIFOR operation manual MINIFOR connection 1. REMOTE (rear of CO 2 -METER) 2. POWER (rear of CO 2 -METER) 3. IN nozzle (rear of CO 2 -METER) 4. OUT nozzle (rear of CO 2 -METER) - PUMP - socket at the rear of MINIFOR control unit - o Connect one end of 8-pole remote control cable to the PUMP - socket at the rear of the MINIFOR control unit o Other end of the 8- pole remote control cable to REMOTEsocket at the rear of CO 2 -METER o Plug the connector of power supply cable into the POWER - socket (12 V) at the rear of CO 2 -METER o Universal plug-in power supply (100-240 V AC/50-60 Hz, 12 VDC, 12 W) to mains RS-485 interface (incorporated with CO 2 -METER) USB or serial port o Connect the RS- 485 connection kit to the PC with the help of the USB connector. o Plug-in the other end of the connection kit (RS- 485 connection cable) to the REMOTE socket of CO 2 -METER - o Plug the connector of power supply cable into the POWER - socket (12 V) at the rear of CO 2 -METER o Universal plug-in power supply (100-240 V AC/50-60 Hz, 12 VDC, 12 W) to mains Connect the gas tubing from the outgas / exit gas condenser to the IN nozzle of CO 2 -METER and secure it with clamps Connect tubing to the OUT nozzle of CO 2 -METER to lead the measured gas to O 2 -METER / CH 4 -METER / to measure total gas flow or others - - www.lambda-instruments.com 3

1.2 Power supply The LAMBDA CO 2 -METER is powered by MINIFOR Fermentor-Bioreactor. The CO 2 -METER is connected with the corresponding 8-pole remote control cable (art. no. 4810) to the PUMP - socket at the rear of the MINIFOR laboratory fermenter-bioreactor. The other end of the 8-pole remote control cable (art. no. 4810) is plugged into the REMOTE - socket at the rear of the CO 2 -METER. When used independently of the MINIFOR laboratory bioreactor-fermentor, a universal plug-in power supply (100-240 V AC/50-60 Hz, 12 VDC, 12 W) is used (art. no. 4820). Connected to the power supply, all LEDs and the display of the CO 2 -METER light shortly. This allows a function control of all signal elements. 1.3 Measurement of CO 2 concentration in the outgas / exit gas The gas tubing from the outgas / exit gas condenser (glass or Peltier outgas condenser of MINIFOR fermentor-bioreactor) is connected to the IN nozzle at the rear of CO 2 -METER to measure the concentration of CO 2 (0-100%). The concentration of CO 2 in the outgas / exit gas can be measured in terms of 0-9.99% and 10.0-99.9%. The gas tubing from the OUT nozzle of CO 2 -METER can be connected to the other outgas concentration measurement instruments like O 2 -METER, CH 4 -METER (IN nozzle) or measure total gas flow with a MASSFLOW gas flow meter. 2 PC CONTROL & SOFTWARE The measured CO 2 outgas concentration data can be visualized and recorded, for example by the industrial fermentation software SIAM. 2.1 Setting up the address of CO 2 -METER To look up/modify the instrument address: Disconnect the 8-pole remote control cable from CO 2 -METER (when used together with the MINIFOR fermentor-bioreactor) or the power supply (stand-alone or PC) Press the CO2 [%] button continuously and at the same time connect the 8-pole remote control cable to CO 2 -METER (when used together with the MINIFOR fermentor-bioreactor) or the power supply (stand-alone or PC) The message A and two numbers will appear on the display. This number from 00 to 99 is the current address of the CO 2 -METER To change the address, press the buttons Λ Λ Λ under the display until the desired number is obtained. To confirm and save the address, press the OK button. www.lambda-instruments.com 4

2.2 PC connection CO 2 -METER with PC software: CO 2 -METER with the RS-485 connection kit (includes: RS- 485 connection cable, RS-232/485 converter, Power supply for RS-232/485 converter and USB to RS-232 converter) is connected to the USB port of the PC (when used independently of MINIFOR fermentor-bioreactor) for the control by software. CO 2 -METER with MINIFOR: CO 2 -METER is connected to the PUMP socket at the rear of MINIFOR fermentor-bioreactor. MINIFOR fermentor-bioreactor is connected to the SIAM industrial fermentation software with the help of PC connection kit. 2.3 SIAM industrial fermentation software SIAM industrial fermentation software is intended for the automation of fermentation and cell culture processes at laboratories. When used together with MINIFOR fermentor-bioreactor, the SIAM industrial fermentation software allows to calculate the RQ (respiratory quotient or respiratory coefficient) from the CPR (Carbon dioxide Production Rate) and OUR (Oxygen Uptake Rate, obtained from LAMBDA O 2 -METER) during the fermentation process. 3 CALIBRATION PROCEDURE FOR THE CO 2 -METER The CO 2 concentration can be calibrated on the LAMBDA CARBOMETER (CO2-Meter). The calibration is a two-point calibration consisting of zero-setting and sensitivity setting: 3.1 Zero setting On the rear side of the LAMBDA CARBOMETER (CO2-METER), there are two buttons labeled N 2 and Air. These buttons are used for the zero setting of the instrument. Connect a source of nitrogen (N 2 ) gas to the CARBOMETER input barb. The gas flow rate should be around 200 ml/min. The display shows a value, which is gradually stabilizing and approaching zero. After the value on the display stabilizes, press the button in the hole on the rear side labeled N 2. The display shows 0.00. A simpler but less accurate way is to use an air flow for the zero setting (air flow rate should be around 200 ml/min). In this case, after the value on the display stabilizes, press the button in the hole on the rear side labeled Air. The display shows 0.04, which corresponds to the expected concentration of CO 2 in air (400 ppm). 3.2 Sensitivity setting Connect a source of known concentration of CO 2 to the CARBOMETER input barb. The gas flow rate should be around 200 ml/min. The display will stabilize with the actual value of CO 2 concentration. After the actual value of CO 2 concentration has stabilized the correction factor can be calculated as the ratio between the real CO 2 concentration and the actual measured CO 2 concentration. Unplug the power supply and continuously press the button located below the highest order digit of the display and reconnect the power supply. The display shows SLP and www.lambda-instruments.com 5

then the value of the correction factor. Use the buttons under the display to set the desired value of the correction factor and press button. The correction factor is saved, the display shows the actual CO 2 concentration and the CARBOMETER is switched to operation mode. 4 TECHNICAL SPECIFICATIONS 4.1 General specification Type: LAMBDA CARBOMETER CO 2 concentration measurement (0-100%) Accuracy: ± 3% reading or 1 digit Measurement range: [CO 2 ] 0 9.99 % and 10.0 99.9% Interface: RS-485 Power supply: 90 240 V/50 60 Hz AC plug-in power supply with DC 12V/12W output Dimensions: 10.5 (H) x 8 (W) x 17 (D) cm Weight: 0.6 kg Safety: CE, meets IEC 1010/1 norm for laboratory instruments Operation temperature: 0-40 C Operation humidity: 0-90% RH, not condensing For safety reasons the voltage of the external signal must not exceed 48 V to earth! 4.2 Remote control (Inputs/outputs) No. Colour Description 1 yellow (+) input remote speed control 0-10V *) 2 grey not used 3 green earth, 0 V 4 brown + 12 V 5 white (+) input remote ON/OFF; 0V = ON, 3 12 V = OFF (this logic can be inversed on demand) 6 pink earth, ground (GND) 7 red RS 485 B (-) 8 blue RS 485 A (+) *) (zero line connected to the contact no. 3) Figure 3.2-1 8 pole connector www.lambda-instruments.com 6

5 ACCESSORIES ART. NO. ACCESSORIES 4817-kit PC connection RS-485 connection kit (for connection to a serial port or USB port) Control 4810 Pump remote control (analog and digital) cable, 8 poles connector 800202 Quadruple plug box (power and RS-connection for up to 4 LAMBDA laboratory instruments) Spare parts 4820 Plug-in power supply (12 V / 12 W) [Plug-type: AU, CH, EU, UK, US] 4815 Silicone tubing 3/5 mm (10 m) 800202 Quadruple plug box (power and RS-connection for up to 4 LAMBDA laboratory instruments) 800083 Electronic Peltier air output condenser (no water connection required) Outgas concentration measurement 8080 OXYMETER (O 2 -METER) O 2 concentration measurement (0-25%) 8081 CARBOMETER (CO 2 -METER) CO 2 concentration measurement (0-100%) 8082 METHAMETER (CH 4 -METER) CH 4 concentration measurement (0-100%) 6 GUARANTEE LAMBDA provides a two-year guarantee on material and manufacturing defects, if the instrument was used according to the operation manual. Conditions of guarantee: The instrument must be returned with a complete description of the defect or problem. In order to send back the equipment for repair, you will need a returns authorization number from LAMBDA. The customer will send the instrument to our service office. Damage or loss of items during transport will not be compensated for by LAMBDA. Failure to fulfil these requirements will disqualify the customer from compensation. Serial Number: Guarantee from: www.lambda-instruments.com 7

7 APPENDIX 7.1 RS communication protocol of LAMBDA CO 2 -METER AND O 2 - METER 7.1.1 Communication settings: Speed: Data format: Transmit-Receive switching: Transmit-Receive interval: 2400 Baud 8 bit, odd parity, 1 stop bit DTR (usually automatic switching) 10 ms 7.1.2 Data format for transmission from the computer (master): # ss mm z qs c send data where: ss is the address of the slave receiver mm is the address of the transmitter (master) z command qs Checksum c carriage return CR (ASCII 0D) 7.1.3 Commands for CO 2 -METER #ssmmkqsc send concentration of CO2 [%] #ssmmhqsc send humidity value [%] #ssmmtqsc send temperature [º C] #ssmmgqsc send the measure value of the concentration of CO2 [%] #ssmmvqsc send the measure value of the concentration of CO2 [%] * Commands G and V allow MASSFLOW simulation in SIAM 7.1.4 Commands for O 2 -METER #ssmmkqsc to send the value of the concentration of O 2 [%] #ssmmoqsc to send the value of O 2 partial pressure [mbar] #ssmmpqsc to send the value of the total pressure [mbar] #ssmmtqsc to send value of temperature [º C] #ssmmgqsc to send the measure value of the concentration of O 2 [%] #ssmmvqsc to send the measure value of the concentration of O 2 [%] * Commands G and V allow MASSFLOW simulation in SIAM www.lambda-instruments.com 8

7.1.5 Data format for transmission to the computer (master): < mm ss a xxxx qs c where: a indicator variables (1 character ASCII) xxxx data (4 ASCII number 0...9 - transmission from high to low) qs checksum in hexadecimal format (2 ASCII symbols 0...9 A B C D E F) c carriage return CR (ASCII 0D) <mmsskxxxxqsc value of O 2 or CO 2 concentration (xx.xx) [%] <mmssoxxxxqsc value of the partial pressure of O 2 (xxx.x) [ C] <mmsspxxxxqsc value of the total pressure (xxxx) [mbar] <mmsstxxxxqsc value of temperature (xxx.x) [ C] <mmsshxxxxqsc value of humidity (xx.xx) Simulation of MASSFLOW in SIAM (commands G and V): <mmssrxxxqsc value of the concentration of O 2 or CO 2 (xx.x) [%] 7.2 RS-connection scheme The 8-pole DIN connector REMOTE is used for the remote control and RS-485 connection. No. Colour Description 1 yellow (+) input remote speed control 0-10V *) 2 grey not used 3 green earth, 0 V 4 brown + 12 V 5 white (+) input remote ON/OFF; 0V = ON, 3 12 V = OFF (this logic can be inversed on demand) 6 pink earth, ground (GND) 7 red RS 485 B (-) 8 blue RS 485 A (+) *) (zero line connected to the contact no. 3) Figure 7.2-1 8 pole connector LAMBDA Laboratory Instruments Sihlbruggstrasse 105 CH-6340 Baar SWITZERLAND EUROPE Tel.: +41 444 50 20 71 Fax: +41 444 50 20 72 E-mail: support@lambda-instruments.com Web: www.lambda-instruments.com LAMBDA CZ, s.r.o. Lozíbky 1 CZ-61400 Brno CZECH REPUBLIC EUROPE Hotline: +420 603 274 677 www.fermentor.net www.bioreactors.eu www.lambda-instruments.com 9