Safety Standards. Model Number:

Similar documents
Safety Standards. Model Number:

Safety Standards. Model Number:

Safety Standards. Model Number:

Safety Standards. Model Number:

Safety Standards. Model Number:

Safety Standards. Model Number:

060AAS19 B MDS. MDS Medical AC-DC Adapter 19Volt, 60Watt / MDS-060AAS19 B. Highlights & Features. Safety Standards. Model Number: General Description

Safety Standards. Model Number:

400AUS24 B MDS AUS. MDS Medical AC-DC Open Frame 24Volt, 400Watt / MDS-400AUS24 B. General Description. Model Information. Model Numbering ```

Safety Standards. Model Number:

040APS12 B MDS. MDS Medical AC-DC Open Frame 12Volt, 40Watt / MDS-040APS12 B ``` Highlights & Features. Safety Standards.

100BPS15 B. MDS Medical AC-DC Open Frame 15Volt, 100Watt / MDS-100BPS15 B ``` Highlights & Features. Safety Standards.

TECHNICAL DATASHEET MDS Medical AC-DC Open Frame 065APS24 B Highlights & Features Safety Standards Model Number: Unit Weight Dimensions (W x L x H):

400AUS24 B MDS AUS. MDS Medical AC-DC Open Frame 24Volt, 400Watt / MDS-400AUS24 B. General Description. Model Information. Model Numbering ```

Safety Standards. Model Number: Unit Weight:

Safety Standards. Model Number:

Safety Standards. Model Number:

030AAC Highlights & Features

350A Series MDS 350 A D 701 A A. Technical Datasheet Medical ATX Power Supply ATX 12V 1U Power Supply

Medical AC-DC Enclosed Power Supply 24V 200W 2 x4 High Density / MDS-200APB24 AA

Safety Standards. Model Number

MDS-300APB24 AA. Medical AC-DC Open Frame Power Supply 24V 300W with 5V/0.5A Standby / MDS-300APB24 AA MDS 300 A P B 24. Highlights & Features

Safety Standards. Model Number Input Voltage Output Voltage Convection Current Output Forced Air Current Output MDS 300 A D B 24

Chrome DIN Rail Power Supply 12V 9.96W 1 Phase (Class II & NEC Class 2) / DRC-12V10W1A

Chrome DIN Rail Power Supply 5V 7.5W 1 Phase (Class II & NEC Class 2) / DRC-5V10W1AZ

Chrome DIN Rail Power Supply 24V 10W 1 Phase (Class II & NEC Class 2) / DRC-24V10W1AZ

MDS-400ADB12 AA. Medical AC-DC Enclosed Power Supply 12V 400W 3.5 x5.5 High Density MDS-400ADB12 AA MDS 400 A D B 12 AA. Highlights & Features

Chrome DIN Rail Power Supply 24V 10W 1 Phase (Class II & NEC Class 2) / DRC-24V10W1A

Safety Standards. Model Number:

Chrome Class II DIN Rail Power Supply 24V 10W 1 Phase (NEC Class 2 & Household) / DRC-24V10W1HZ

MDS-250A Series. Medical AC-DC Open Frame Power Supply 24V 250W with 5V/0.5A Standby / MDS-250A series MDS 250 A B 24. Highlights & Features

Model Number: Unit Weight:

Medical / Industrial AC-DC Power Supply 600W Convection Cooling / MEU-600C series

MDS-400APB12. Medical AC-DC Open Frame Power Supply 12V/400W 3 x5 High Density MDS-400APB12 MDS 400 A P B 12. Highlights & Features.

Chrome DIN Rail Power Supply 24V 30W 1 Phase (Class II & NEC Class 2) / DRC-24V30W1AZ

Safety Standards. Model Number

Chrome DIN Rail Power Supply 24V 60W 1 Phase (Class II & NEC Class 2) / DRC-24V60W1A

PMT. Safety Standards. Model Number:

PMT. Model Number: Model Number Input Voltage Range Output Voltage Output Current PMT-5V350W1A (Selectable by Switch)

TECHNICAL DATASHEET Medical AC-DC Enclosed Power Supply MDS-400ADB24 Highlights & Features Safety Standards Model Number: Unit Weight:

Safety Standards. Model Number:

PJ Open Frame Power Supply 24V 50W 1 Phase / PJ-24V50W NA

Safety Standards. Model Number:

Model Number: Model Number Input Voltage Range Output Voltage Output Current PMC-24V100W1AX Vac ( Vdc) 24Vdc 4.17A

CliQ DIN Rail Power Supply 24V 60W 1 Phase / DRP024V060W1AZ

Chrome DIN Rail Power Supply 12V 25W 1 Phase (Class II & NEC Class 2) / DRC-12V30W1A

PMC. PMC Panel Mount Power Supply 24V 50W 1 Phase / PMC-24V050W1A. Highlights & Features. Safety Standards. Model Number: General Description

PMC. Model Number: Model Number Input Voltage Range Output Voltage Output Current PMC-12V060W1NA Vac ( Vdc) 12Vdc 5.

Sync DIN Rail Power Supply 5V 15W 1 Phase (NEC Class 2) / DRS-5V30W1NZ

Chrome DIN Rail Power Supply 12V 72W 1 Phase (Class II) / DRC-12V100W1AZ

PMT. Model Number: CC Code PM 24V 200W 1 A Panel Mount Product Type T Enclosed L L Frame*

PMT. PMT Panel Mount Power Supply 5V 50W 1 Phase / PMT-5V50W1A. Highlights & Features. Safety Standards. Model Number: General Description

Safety Standards. Model Number:

PMC. Safety Standards. Model Number:

PMC. Model Number: Model Number Input Voltage Range Output Voltage Output Current PMC-24V600W1BA Vac ( Vdc) 24Vdc 25.

PMC. Model Number: Model Number Input Voltage Range Output Voltage Output Current PMC-48V150W1B Vac ( Vdc) 48Vdc 3.

PMT. Model Number: CC Code PM 24V 150W 1 A Panel Mount Product Type T Enclosed L L Frame*

Safety Standards. Model Number:

PMT. Safety Standards. Model Number:

Model Number: Model Number Input Voltage Range Rated Output Voltage Rated Output Current PMT-24V100W2AA Vac 24Vdc 4.50A

Sync DIN Rail Power Supply 24V 30W 1 Phase (Class II & NEC Class 2) / DRS-24V30W1AZ

Sync DIN Rail Power Supply 24V 30W 1 Phase (NEC Class 2) / DRS-24V30W1AZ

PMT. Model Number: Model Number Input Voltage Range Output Voltage Output Current PMT-D2V100W1A. V1: 3.50A (Selectable by Switch) V2: 5Vdc

Model Number: Model Number Input Voltage Range Rated Output Voltage Rated Output Current PMT-12V100W2B Vac 12Vdc 8.5A PMT-24V100W2B 24Vdc 4.

CliQ DIN Rail Power Supply 24V 240W 1 Phase / DRP024V240W1AA

Medical/Industrial AC-DC Configurable Power Supply Up to 2100 Watt / MEG-A Series

PMT. Safety Standards. Model Number:

PJ Open Frame Power Supply 12V 100W 1 Phase / PJ-12V100W A

CliQ DIN Rail Power Supply 12V 30W 1 Phase / DRP012V030W1A

PMF. Safety Standards. Model Number:

LNE. LED Driver LNE-100W Series / LNE- V100W. Highlights & Features. Safety Standards. Model Number: General Description

PJT Highlights & Features

PMR. PMR Panel Mount Power Supply 4.2V 252W 1 Phase / PMR-4V320WC A. Highlights & Features. Safety Standards. Model Number: General Description

Chrome DIN Rail Power Supply 24V 91.2W 1 Phase (Class II & NEC Class 2) / DRC-24V100W1A

Watts. xppower.com. General. Input. Environmental. EMC & Safety. Output. IT & Medical Approvals. Energy Efficiency Level V, CEC2008 & EISA 2007

Safety Standards. Model Number:

Output Voltage in Constant Voltage Mode. Model Number

PMT. Safety Standards. Model Number:

PJ Open Frame Power Supply 24V 100W 1 Phase / PJ-24V100W A

PMT. PMT Panel Mount Power Supply 12V 35W 1 Phase / PMT-12V35W1A. Highlights & Features. Safety Standards. Model Number: General Description

PMT. Safety Standards. Model Number:

PMH. Safety Standards. Model Number:

Sync DIN Rail Power Supply 24V 50W 1 Phase (NEC Class 2) / DRS-24V50W1N

PMT. Model Number: Model Number Input Voltage Range Output Voltage Output Current PMT-12V35W1A Vac 12Vdc 2.92A

Sync DIN Rail Power Supply 5V 50W Series / DRS-5V50W1A / DRS-5V50W1N

AC-DC POWER SUPPLIES UNIVERSAL INPUT, 65 WATT MEDICAL APPLICATIONS MUI65 SERIES

CliQ DIN Rail Power Supply 24V 240W 3 Phase / DRP024V240W3AA

PMR. PMR Panel Mount Power Supply 5V 300W 1 Phase / PMR-5V320WD A. Highlights & Features. Safety Standards. Model Number: General Description

VER05 Series. xxx Series. 5 Watt. AC-DC Power Supplies. Models & Ratings. Input. Output

Safety Standards. Model Number:

MPM-U653-DCF-B M/N: MPM-U653-DCF-B. 650W Medical AC / DC SPECIFICATION FOR SWITCHING POWER SUPPLY. Revision History

LPS. ME20 20W Single Output External Power Medical Series. Features

Features. Regulated Converter. RAC35-G/OF 35 Watt 4 x 2 Open Frame RAC35- SG/OF. AC/DC Converter

CliQ DIN Rail Power Supply 12V 30W 1 Phase / DRP012V030W1A

LNE. LED Driver LNE-185W Series / LNE- V185W. Highlights & Features. Safety Standards. Model Number: General Description

PMT. Model Number: Model Number Input Voltage Range Rated Output Voltage Rated Output Current PMT-12V50W1A Vac 12Vdc 4.17A

12-15VDC (12V±5%) 24-28VDC (24V±5%)

PMT. Safety Standards. Model Number:

PMC. Safety Standards. Model Number:

ME to 30W Single Output External Power Medical Series. Features

Transcription:

060BAS12 A Highlights & Features Meet Efficiency Level VI Safety Approvals to IEC 60601-1 3rd ed. & IEC 60950-1 Compliant with IEC 60601-1-2 4th Ed. Requirements Low touch current (<0.1mA Normal &,0.3mA single fault) Over-Voltage/Load/Temperature & Short Circuit protections 1 Million Hours MTBF 2 x MOPP (means of patient protection) 3 years warranty Safety Standards General Description CB Certified for worldwide use Model Number: MDS-060BAS12 A Unit Weight: 340 grams (12.0 ounces) Dimensions (W x L x H): 62.0 x 135.0 x 34.1 mm 2.4 x 5.3 x 1.3 in The MDS series of external power supply come with universal AC input at 90Vac to 264Vac. Other features include low touch current, risk management report available and the electric shock protection comply with 2 x MOPP. The MDS series is certified for EMC standards according to EN 55011 for industrial, scientific and medical (ISM) radio-frequency equipment and EN 55022 for Information Technology Equipment (ITE) radio-frequency equipment. In addition, only recognized Japanese capacitors are used. The MDS series of external power supply come with both medical and ITE safety approvals including UL/cUL/CCC/CE and CB certification and are fully compliant with RoHS Directive 2011/65/EU for environmental protection. Model Information Medical AC-DC Adapter Model Number Input Voltage Range Output Voltage Output Current MDS-060BAS12 A 90-264Vac 12Vdc 5A Model Numbering MDS - 060 BAS 12 A Delta Medical power Supply Max wattage in the product series. Maybe lower at some voltage. 060 60W Family Code Output Voltage Single Output: 12 for 12V 1

Specifications Input Ratings / Characteristics Nominal Input Voltage Input Voltage Range Nominal Input Frequency Input Frequency Range Input Current (max) Average Efficiency (min) Standby Power (max) Inrush Current (typ.) 100-240Vac 90-264Vac 50-60Hz 47-63Hz 1.5A @ 115Vac, 0.75A @ 230Vac 88% @ 115Vac and 230Vac 0.21W @ 115Vac and 230Vac 50A @ 115Vac, 100A @ 230Vac Touch Current (max) 0.1mA @ 264Vac NC 1), 0.3mA @ 264Vac SFC 2) 1) NC: normal condition 2) SFC: single fault condition Output Ratings / Characteristics Fig.1 Efficiency versus output load Nominal Output Voltage 12Vdc Output Voltage Tolerance ± 5.5% Output Current 5A Output Power 60W Line Regulation (max) ±0.5% Load Regulation (max) ±5% Ripple & Noise (typ.) 1% pk-pk @ Full load (20MHz BW) Start-up Time (max) 3000ms @ 115Vac Hold-up Time (min) 10ms @ 115Vac Dynamic Response (Overshoot & Undershoot O/P Voltage) ± 10% @ 50-100% load *Periodic and Random Deviation 2

Ripple & Noise measurement circuit PSU 5cm 0.47uF Ceramic 10uF Electrolytic Load Scope Test Fixture Mechanical Case Chassis PC Case Cover PC Dimensions(W x L x H) 62.0 x 135.0 x 34.1 mm (2.4 x 5.3 x 1.3 in) Unit Weight 340 grams (12.0 ounces) Indicator NA Cooling System NA Terminal Input IEC320 C8 socket Output 4pin Din / Barrel type / Tuning fork type Environment Surrounding Air Temperature Operating 0 C to +40 C Linearly derate, at rate of 1.5 watts per degree C, from 60W at 40 C to 45W at 50 C Operating Humidity Operating Altitude Shock Test (Non-Operating) Vibration (Operating) Storage -40 C to +85 C 5-95% RH (Non-Condensing) 5,000 meters (16,400 feet) 50G, 11ms, 3 shocks for each direction 5-500Hz, 2.09Grms, 20 minute for each three axis 3

Protections Overvoltage (max) Over load / Over current (max) Over Temperature Short Circuit Degree of Protection Protection Against Shock 150%, Latch Mode 180% of rated load current, Hiccup Mode, (Non-Latching, Auto-Recovery) Hiccup Mode, (Non-Latching, Auto-Recovery) Hiccup Mode, (Non-Latching, Auto-Recovery) IP22 Class II Reliability Data MTBF (typ.) 1 million Hours based on Telecordia SR-332 Safety Standards / Directives Medical Safety IEC60601-1 3 rd and 3.1 rd edition CB report IEC60601-1 edition 3.1rd (2012), EN60601-1 (2006) + A11 + A1 + A12, CAN/CSA-C22.2 NO. 60601-1:14, ANSI/AAMI ES60601-1:2005/(R)2012 ITE Safety CE Material and Parts Galvanic Isolation Input to Output 4000 Vac IEC60950-1 (Ed.2,2005), GB4943.1-2011, GB9254-2008, GB17625.1-2003 MDD Directive 93/42/EEC RoHS Directive 2011/65/EU Compliant 4

EMC (Compliant with IEC 60601-1-2 4th Ed. Requirements) EMC / Emissions EN55011/EN55022,FCC Title 47:Class B Harmonic Current Emissions IEC61000-3-2 Meet Class A limit Immunity to Voltage Flicker IEC61000-3-3 Electrostatic Discharge IEC61000-4-2 Level 4 Criteria A 1) Radiated Field IEC61000-4-3 Criteria A 1) Air Discharge: 15kV Contact Discharge: 8kV Electrical Fast Transient / Burst IEC61000-4-4 Level 3 Criteria A 1) : 2kV Surge IEC61000-4-5 Level 3 Criteria A 1) Differential Mode 3) : 1kV Conducted IEC61000-4-6 Level 2 Criteria A 1) Power Frequency Magnetic Fields IEC61000-4-8 Criteria A 1) Voltage Dips IEC61000-4-11 Criteria A 1) 1) Criteria A: Normal performance within the specification limits 2) Criteria B: Output out of regulation, or shuts down during test. Automatically restore to normal operation after test. 3) Symmetrical: Differential mode (Line to line) 80MHz-2700MHz, 10V/m AM modulation 385MHz-5785MHz, 28V/m Pulse mode and other modulation 150kHz-80MHz, 3Vrms, 6Vrms at ISM bands and Amateur radio bands Magnetic field strength 30A/m 0% UT, 0.5 cycle (10ms), 0 o /45 o /90 o /135 o /180 o /225 o /270 o /315 o /360 o Criteria B 2) 0% UT, 1 cycle (20ms), 0 o Criteria A 1) 70% UT, 25 cycle (500ms), 0 o Criteria B 2) 0% UT, 250 cycle (5000ms), 0 o 5

Block Diagram 6

Dimensions W x L x H: 62.0 x 135 x 34.1 mm Item Device Description 1 Cover 2 Chassis 3 Power Cord (UL1185, #18AWG) 4 Socket (IEC320 C8) 7

Functions Start-up Time The time required for the output voltage (Vo) to reach 90% of its set value, after the input AC voltage is applied. Rise Time The time required for the output voltage (Vo) to change from 10% to 90% of its steady state value. Hold-up Time Hold up time is the time when the AC input collapses and output voltage retains regulation for a certain period of time. The time required for the output to reach 95% of its set value, after the input voltage is removed. Graph illustrating the Start-up Time, Rise Time, and Hold-up Time 8

Dynamic Response The power supply output voltage will remain within ±3% of its steady state value, when subjected to a dynamic load change from 50 to 100% of its rated current. Short Circuit Protection The power supply s output OLP/OCP function also provides protection against short circuits. When a short circuit is applied, the output current will operate in Hiccup mode, as shown in the illustration in the OLP/OCP section on this page. The power supply will return to normal operation after the short circuit is removed. Overload & Over current Protections The power supply s Overload (OLP) and Over current (OCP) Protections will be activated when output current is between 110% and 130% of I O (Max load). Upon such an occurrence, V O will start to drop. Once the power supply has reached its maximum power limit, the protection will be activated. and the power supply will go into Hiccup mode (Auto-Recovery). The power supply will recover once the fault condition causing the OLP and OCP is removed and I O is back within the specified limit. Inrush Current Inrush current is the input current that occurs when the input voltage is first applied. For AC input voltages, the maximum peak value of inrush current will occur during the first half cycle of the applied AC voltage. This peak value decreases exponentially during subsequent cycles of AC voltage. Additionally, if the I O is <130% but >110% for a prolong period of time (depending on the load), the Over Temperature Protection (OTP) will be activated due to high temperature on critical components. The power supply will then go into hiccup mode until the fault is removed; and, the input voltage is removed, then reapplied. Overvoltage Protection The power supply s overvoltage circuit will be activated when its internal feedback circuit fails. The output voltage shall not exceed its specifications defined on Page 3 under Protections. Power supply will latch off, and require removal/re-application of input AC voltage in order to restart. Over Temperature Protection As mentioned above, the power supply also has Over Temperature Protection (OTP). This is activated when the overload condition persists for an extended duration and the output current is below the overload trigger point but >100% load. In the event of a higher operating condition at 100% load, the power supply will run into OTP when the surrounding air temperature is higher than the operating temperature. When activated, the output voltage will go into hiccup mode until the input voltage is removed; then, reapplied, and the surrounding air temperature drops to its normal operating temperature. 9

Certificate All Delta Medical Power products conform to the European directive 2011/65/EU. RoHS is the abbreviation for Restriction of the use of certain hazardous substances Delta has been certified as meeting the requirement of ISO 13485: 2003 and EN ISO 13485:2012 for the design and manufacture of switching power supply and adaptor for medical device. Delta is approved for the UL Total Certification Program (TCP) approved client laboratory for IEC62368-1. Delta also has participated UL Client Test Data Program (CTDP) for IEC 60601 Meet Level VI Efficiency Requirement 10