GE Digital Solutions Druck Limited Fir Tree Lane Groby Leicester LE6 0FH England Tel: 0116 231 7100 SOFTWARE HISTORY Programmable Automated Calibration Equipment PACE1000 Pressure Indicator Page 1 of 15 PACE1000 Software History.doc SR0092 @ 10 November 2017
PACE1000 current release Rev E12 software component versions Instrument main code DK0406 V01.10.12 Instrument OS Build DK0366 V01.10.01 PACE1000 correct operation is only supported when component versions installed into the product agree with Changes implemented DK0406 Added compatibility to accept old sensors which have data format set incorrectly to 6. Software versions shown in the software installed status menu will now show the beta key word for a beta release of software or red dot if newer version is available. Closes underlying TCP sockets properly when connection is lost. Added Scpi command :SENSE:PRESSURE:RANGE:CAT? Which returns all real ranges with their serial numbers. Added Scpi command :SENSE:PRESSURE:RANGE[n]:SN? Where n goes from 1 to 3, either gets serial number of the range displayed in one of the three tabs or sets the tab with the range which serial number is provided. Corrected Scpi calibration commands procedure, where for gauge sensor 3 Scpi set point commands need to be given and for absolute sensor 2 Scpi set point commands need to be given. Corrected Scpi calibration abort command which now exists calibration mode as well. Made change which now displays correct limit for pseudo ranges when area of use is changed from Europe to USA. Added functionality to use any valid range displayed in any of three tabs or P1-P2 or P2-P1 as input for analogue card. Input is selected from analogue card options. External IDOS calibration message disappears after 5 seconds. Old sensor warning message disappears after 2 seconds. If # flag is disabled for all the communication ports, then it will also not be displayed on the screen or range menu. Made correction to display correct values of Max, Avg and Min. 10 th November 2017 Page 2 of 15 PACE1000 Software History.doc SR0092 @ 10 November 2017
PACE1000 release Rev E12 beta 2 software DK component versions Software release revisions are now alpha numeric, with old numeric release shown to provide historical trail. Instrument main code DK0406 V01.10.04 (beta) Instrument OS Build DK0366 V01.10.01 PACE1000 correct operation is only supported when component versions installed into the product agree with Changes implemented DK0406 Fixed an issue with the :CAL[x]:PRES:POIN? query not replying with the correct number of calibration points required to complete a calibration. Created new SCPI queries/command to allow the SCPI user to distinguish between sensors of the same range and also to move each sensor into the top display position for calibration, as the calibration routine only calibrates the sensor in the top display position, and for query of pressure readings using SENS:PRES? from the main top display area: :SENS:PRES:RANG:CAT? allows query of all sensor ranges with the serial number of each sensor. :SENS:PRES:RANGx:SN? allows query of the sensor serial number displayed in the display windows [x], where x is a numeric. :SENS:PRES:RANGx:SN yyyy allows a sensor with serial number yyyy to be moved into a new display (top=1, middle =2 or lower=3) position [x], where y and x are both numeric. 1 st December 2016 Software releases named beta : Each beta software is provided to customer hereunder on an as-is basis Without any representation, warranty or condition of any kind whatsoever. Customer acknowledges and agrees that neither ge nor any of its affiliates Have made any express warranties, representations or conditions relating to The beta software. Neither GE nor any of its affiliates make any Representations, warranties or conditions, express or implied, to customer or any other entity, including any warranties, representations or conditions regarding the merchantability, suitability, noninfringement of third party rights, originality, fitness for a particular purpose, or results to be derived or created from the use of the beta software provided hereunder. In no event Shall GE, any of its affiliates, or any of their respective directors, officers, employees, agents or subcontractors be liable under any theory of tort (including negligence), contract, strict liability or other legal theory for any lost profits, loss of data, exemplary, punitive, special, incidental, direct, Indirect or consequential damages, each of which is hereby excluded by agreement of the parties regardless of whether either party or any other such entity has been advised of the possibility of such damages. Customer acknowledges that neither GE nor any of its affiliates has any liability. Page 3 of 15 PACE1000 Software History.doc SR0092 @ 10 November 2017
PACE1000 release Rev E12 beta 1 software DK component versions Software release revisions are now alpha numeric, with old numeric release shown to provide historical trail. Description DK Number Version Instrument main code DK0406 V01.10.00 (beta) Instrument OS Build DK0366 V01.10.01 PACE1000 correct operation is only supported when component versions installed into the product agree with Changes implemented DK0406 Changes to TCP socket which now closes all underlying sockets once connection is lost. 3 rd November 2016 Software releases named beta : Each beta software is provided to customer hereunder on an as-is basis Without any representation, warranty or condition of any kind whatsoever. Customer acknowledges and agrees that neither ge nor any of its affiliates Have made any express warranties, representations or conditions relating to The beta software. Neither GE nor any of its affiliates make any Representations, warranties or conditions, express or implied, to customer or any other entity, including any warranties, representations or conditions regarding the merchantability, suitability, noninfringement of third party rights, originality, fitness for a particular purpose, or results to be derived or created from the use of the beta software provided hereunder. In no event Shall GE, any of its affiliates, or any of their respective directors, officers, employees, agents or subcontractors be liable under any theory of tort (including negligence), contract, strict liability or other legal theory for any lost profits, loss of data, exemplary, punitive, special, incidental, direct, Indirect or consequential damages, each of which is hereby excluded by agreement of the parties regardless of whether either party or any other such entity has been advised of the possibility of such damages. Customer acknowledges that neither GE nor any of its affiliates has any liability. Page 4 of 15 PACE1000 Software History.doc SR0092 @ 10 November 2017
PACE1000 current release Rev D11 software component versions Released via SAP, software release version changes numeric to Alpha from this point onwards, with old numeric release shown. Instrument main code DK0406 V01.09.02 Instrument OS Build DK0366 V01.10.01 PACE1000 correct operation is only supported when component versions installed into the product agree with Changes implemented DK0406 Fixed issue with ROC not being updated in Aeronautical mode for configurable TERPS sensors. Added user control enable/disable hash (#) be part of the scpi pseudo range commands for all communication ports. Added delays when analogue card is attach to overcome hanging boot up. 5 th February 2016 Page 5 of 15 PACE1000 Software History.doc SR0092 @ 10 November 2017
PACE1000 SR0092 Rev C10 current software version Released via SAP, software release version changes numeric to Alpha from this point onwards, with old numeric release shown. Instrument main code DK0406 V01.08.05 Instrument OS Build DK0366 V01.10.01 PACE1000 correct operation is only supported when component versions installed into the product agree with Changes implemented DK0406 Added Qc Zero when two IRS sensors are fitted. Disabled the debug print to file in SD card to prevent intermittent corruption of SD Card. Added change which now displays Barometer as Barometer (0) not Barometer0 Added Sensor Type Configuration which now deletes sensor config file and allows to configure the sensors again. Pseudo gauge Range minimum is 7bar. Now it checks the OS files at OS upgrade not just OS folder. Increased the limit of absolute sensors which can be zeroed from 20bar abs to 210bar abs. 28 th September 2015 Page 6 of 15 PACE1000 Software History.doc SR0092 @ 10 November 2017
PACE1000 SR0092 Rev B9 component versions Released via SAP, software release version changes numeric to Alpha from this point onwards, with old numeric release shown. Instrument main code DK0406 V01.07.00 Instrument OS Build DK0366 V01.10.01 Changes implemented Enhancements: None General Bug fixed: DK0406 - Aeronautical mode airspeed leak test issue corrected where over-range is displayed with 2off RPTs fitted in Ps/Pt. 18 th August 2015 Page 7 of 15 PACE1000 Software History.doc SR0092 @ 10 November 2017
PACE1000 SR0092 Rev A8 component versions First release via SAP, software release version changes numeric to Alpha from this point onwards, with old numeric release shown. Instrument main code DK0406 V01.06.07 Instrument OS Build DK0366 V01.10.01 Changes implemented Enhancements: DK0406 - Sensor gain resolution increased for pressure calculation from PACE RPT, TERPS and Piezo resistive sensors. Support for 2bara and 3.5bara TERPS sensors added. DK0366 - Ethernet port firewall added. General Bug fixed: ZERO history file length limited to prevent DPI.EXE error. 5 th June 2015 Page 8 of 15 PACE1000 Software History.doc SR0092 @ 10 November 2017
PACE1000 SR0092 release 7 component versions Instrument main code DK0406 V01.05.00 Instrument OS Build DK0366 V01.09.00 Changes implemented in Instrument main code - DK0406:V01.05.00 General Bug fixed: Sensor status screen, incorrectly displayed IPSn rather than IRSn status for: R202S-055B-nn-4695 (750-1150mbar Grade nn) RPT Barometric sensors. Unidentified or Un-Programmed sensors Flagged up during Power-Up. 4 th July 2014 Page 9 of 15 PACE1000 Software History.doc SR0092 @ 10 November 2017
PACE1000 SR0092 release 6 component versions Instrument main code DK0406 V01.04.00 Instrument OS Build DK0366 V01.09.00 Changes implemented in Instrument main code - DK0406:V01.04.00 Enhancements: Ability to allow Alias names to each sensor range. Added 9 more languages. Support IDOS sensors as plug & play.( Hot-pluggable) Ability to configure TERPS sensors as Barometers sensors. Able to copy instrument settings to USB stick. Able to zero sensors via SCPI commands. Display additional pseudo ranges as #a or #g as applicable. General Bugs fixed: Error fixed during transfer of large Datalog files to USB stick. Fixed ability to re-set TERPS (Barometer) back to Terps. abs sensor. High pressure sensors are now correctly recognised. Sensors correctly recognised during power-up. Able to set Ethernet Subnet Mask of 255.255.255.0 18 th June 2014 Page 10 of 15 PACE1000 Software History.doc SR0092 @ 10 November 2017
PACE1000 SR0092 release 5 component versions Instrument main code DK0406 V01.03.01 Instrument OS Build DK0366 V01.09.00 CHANGES Instrument main code DK0406 V01.03.01 Support Aeronautical unit s feet and metres in Airfield task. Adjustable Airfield Trend period. APPLICABILITY OF NEW ISSUE 3 rd June 2013 Page 11 of 15 PACE1000 Software History.doc SR0092 @ 10 November 2017
PACE1000 SR0092 RELEASE 4.0 COMPONENT VERSIONS Instrument main code DK0406 V01.02.02 Instrument OS Build DK0366 V01.09.00 CHANGES Instrument main code - V01.02.02 Corrected Entry for 1000 bar range in the Range Table. This is because the instrument would display 10,000 bar with a 1000 bar sensor on the range selection pages when the area of use is set to Europe. APPLICABILITY OF NEW ISSUE 15 th May 2012 Page 12 of 15 PACE1000 Software History.doc SR0092 @ 10 November 2017
PACE1000 SR0092 RELEASE 3.0 COMPONENT VERSIONS Instrument main code DK0406 V01.01.32 Instrument OS Build DK0366 V01.09.00 CHANGES DK0365 Change to flash memory block unlock command because of Numonyx flash memory fault Disable boot diagnostic output on RS232 interface to prevent problems with connected equipment DK0366 Change to flash memory block/unlock command for Numonyx flash memory fault APPLICABILITY OF NEW ISSUE 13 th Feb 2012 Page 13 of 15 PACE1000 Software History.doc SR0092 @ 10 November 2017
PACE1000 SR0092 RELEASE 2.0 COMPONENT VERSIONS Instrument main code DK0406 V01.01.32 Instrument OS Build DK0366 V01.08.01 Instrument Boot ROM DK0365 V01.02.00 CHANGES Instrument main code DK0406 V01.01.32 Includes Aeronautical Task Includes Range Compare Function Includes Ethernet Communications as standard Support for Terps Sensors General Bug fixes. APPLICABILITY OF NEW ISSUE 8 th Dec 2011 Page 14 of 15 PACE1000 Software History.doc SR0092 @ 10 November 2017
PACE1000 SR0092 RELEASE 1.0 COMPONENT VERSIONS Instrument main code DK0406 V01.00.00 Instrument OS Build DK0366 V01.08.01 Instrument Boot ROM DK0365 V01.02.00 CHANGES N/A APPLICABILITY OF NEW ISSUE 15 th Feb 2011 Page 15 of 15 PACE1000 Software History.doc SR0092 @ 10 November 2017