EX7000 SCPI COMMANDS

Similar documents
87421/22A Power Supply. Operating and Service Manual

Agilent Technologies E5339A 38-Pin Low-Voltage Probe

Firmware Updates Model 24 / 44 / 42 Cryogenic Temperature Controllers

VTEXSYSTEM DRIVER PROGRAMMER S MANUAL. P/N: Released July 30, VTI Instruments Corp.

Quick Start/Installation Guide

Contents. HP E1586A Rack Mount Terminal Panel User s Manual

Interface Modules for HP Series Instruments. I/O Function Guide. Copyright Hewlett-Packard Company All rights reserved

PCMCIA Flash Card User Guide

DAS 9200 Mainframe System Upgrade

Relay Board. User Manual. 1.0, Oct 2013

Release Information. Revision History. Version: build 018 Release Date: 23 rd November 2011

Agilent Technologies LogicWave

Instruction Manual. CT-4 High-Current Transformer

Redirector User Guide

Trimble S6 and SPS700 Total Station Firmware

STAINLESS STEEL CAST FITTINGS 150lb 304 & 316 Stainless Steel Cast Fittings Specifications

DCS-E 1kW Series, DLM-E 3kW & 4kW Power Supplies

Hardware Documentation

Quick Installation Guide TK-407K

FOR LT-DPLS401. Document No.: A-A Creation: Sep 28, 2010 Revision: 01, Sep 28, 2010

iconverter 2-Module Power Chassis

MiT. MOVING image TECHNOLOGIES INSTRUCTIONS FOR INSTALLATION, OPERATION, AND MAINTENANCE

OPERATIONS MANUAL PCM-DOC

EPSON Preferred. Priority Technical Support. Toll-Free Phone Number. On-Site Repair. Security and Peace of Mind

USB Server User Manual

PZ 97E. LVPZT OEM Amplifier. for Multilayer Bending Actuators. Technical Documentation

Programmable DC Power Supply IT6800 series Programming Guide

Automated Tuner System Power Distribution Hub

SM7100 MICROWAVE MATRIX USER S MANUAL. P/N: Released February 13, VXI Technology, Inc.

Quick Installation Guide TU2-700

Pasternack. PE11S390X SERIES USB Stick Synthesizer. Generation II.

OL-KT. Online Connection Kit Installation Manual

EPSON Preferred. Priority Technical Support. Toll-Free Phone Number. Security and Peace of Mind. On-Site Repair

User Guide. Portable Calibration Module

Agilent Technologies. Connectivity Guide. USB/LAN/GPIB Interfaces. Agilent Technologies

Instructions. RM3000 Rackmount Kit for TDS3000 Series Oscilloscopes

Macon Plastics, Inc. Solvent Weld SCH 40/80 DWV

SUPPLEMENT TO THE MDC-360C MANUAL MDC-361C. Film Deposition Controller. IPN Rev. C

Bluetooth USB User Guide. Revision A July 12, 2006 Part Number GC

TB-100 ControLynx Terminal Block

Omnitron Systems Technology, Inc. 1. iconverter. 19-Module Managed Power Chassis User s Manual

HP E2485A. Installation Guide. Memory Expansion Interface. Copyright Hewlett-Packard Company 1997 All Rights Reserved

OPERATOR S MANUAL FPD 50 WATT POWER SUPPLY. Flat Pack Single Output DC to DC Converter 50 Watt Output 24 Vdc Nominal Input

2 Welcome and Congratulations

EASON TECHNOLOGY. IO8 & IO24 Break-Out Module

Atlas. GPI Smart Matrix. User Manual

DOCKING STATION FOR THE APPLE 13 MacBook

DIO24 24 Bit Digital I/O Module for PC/104

Agilent Gb/s Serial BERT

Agilent CSV Export Utility

HP Worldwide Limited Warranty and Technical Support

Woolich Racing. USB ECU Interface User Guide

PDA-CAN Bridge. Version 1.0. LXNAV d.o.o. Kidričeva 24, 3000 Celje, Slovenia tel fax

MP3 Speaker USER GUIDE

VM4016 USER S MANUAL ANALOG COMPARATOR. P/N: Rev. September 5, VXI Technology, Inc.

Software User s Guide

Eclipse PX! Surround Panning Joystick Controller!!! !!!!!!!!!!!!!!!!!!! User Manual for Windows 7!

Errata. Agilent References in this manual. About this manual. Support for your product

Instruction Sheet. PCS Series. Fiber\Cable Management Spool

Carefree-Security. Installation and programming instructions 1050A. Owner s Manual

XS/SC26-2 Safety Controller. Quick Start Guide

TC1880 Series. 4/5/6/8 Channel RS-232 FIBER OPTIC MICRO MUX User's Manual

HDR-8X8-Plus. User Manual. HDMI 4K 8x8 Switch Matrix Connect up to eight different HDMI video sources to eight different outputs.

Surround Panner. for Avid Pro Tools HD. Users Manual

Protogate Freeway Lifecycle Support Plan (LSP)

XS/SC26-2 Safety Controller Quick Start Guide

Agilent 4155C Semiconductor Parameter Analyzer Agilent 4156C Precision Semiconductor Parameter Analyzer

Keysight M9393A PXIe Vector Signal Analyzer

Trimble R/5000 Series GPS Receivers Release Notes

VAR-EXT-CB8 Datasheet Camera Extension Board for VAR-DT8MCustomBoard & SPEAR-MX8CustomBoard V 1.x

SMA OneTouch ICS Environmental Monitoring System REMOTE MONITORING GUIDE

USB 3.0 Spectra

16/24 Port 10/100 Auto-Sensing Dual Speed Ethernet Switch

TD 600 Thermo-Fastprinter

VM1548 TTL I/O MODULE USER S MANUAL. P/N: Released June 7 th, VXI Technology, Inc.

USER S MANUAL ENTITY. Energy Meter TRINITY. Entity Operational Manual

Quick Start Setup Guide

4300 WINDFERN RD #100 - HOUSTON TX VOICE (713) FAX (713) web: IMPORTANT!!!

Process of Extended Warranty Certificate Generation

Digital Control Power Supply IT6720. User s Manual. Access for more products information

Quick Installation Guide TEW-AI07OB

User Guide. Portable Calibration Module

Instructions. TLA5Fxxx, TLA6Fxx, & TLA7Fxx PowerFlex Field Upgrade Kits for TLA5000, TLA600, and TLA700 Logic Analyzers

Asia Warranty Request Purchase from a Dow Corning Distributor

Keysight Add Source and Receiver Attenuators Upgrade Kit

Report Viewer Version 8.1 Getting Started Guide

Agilent Technologies E8482A/B Six/Three 1x4 RF Multiplexer Module User s Manual *E * Manual Part Number: E Printed in U.S.A.

4300 WINDFERN RD #100 HOUSTON TX VOICE (713) FAX (713) web: IMPORTANT!!!

OPERATIONS MANUAL PCM-I/O48

Table Contents. Introduction Key Features...2. Getting Started Package Contents...3. Minimum System Requirements...3

EPSON PreferredSM Limited Warranty Program for the Epson Stylus

Advanced 4/8-Port DVI-D Matrix KVM

Contents HP E1361A 4X4 Relay Matrix Module User s Manual Edition 2

Customer Support: For more information or support, please visit or at Product Release Information...

Ludlum Lumic Data Logger Software Manual Version 1.1.xx

HP 87050A Option H42 Multiport Test Set

Instruction and Operation Manual. CRMS Contact Resistance Measurement System. Model Number Caution!

Secure Com Port Redirector User Guide

AMERICA S PREMIER EXERCISE EQUIPMENT RCD-347. Chin-Dip / Ab-Back / Push-Up Stand. TuffStuff Fitness Equipment, Inc. 83 3/4" 43 1/4" 44 3/4"

OPERATING AND SERVICE MANUAL. Universal Interface Device 47

Transcription:

EX7000 SCPI COMMANDS PROGRAMMER S MANUAL P/N: 82-0153-000 Released: 05 May 2015 Rev A VTI Instruments Corp. 2031 Main Street Irvine, CA 92614-6509 (949) 955-1894

Table of contents Support Resources... 4 Introduction... 5 Background... 5 Communication Protocol... Error! Bookmark not defined. There are 2 ways to communicate with EX7000 by using EX7000 SCPI commands, SCPI RAW and VXI-11.... Error! Bookmark not defined. SCPI RAW... 6 VXI-11... 6 Commands... 7 Syntax... 7 SCPI Commands... Error! Bookmark not defined. ROUTe:CLOSe (@<list>)... 8 ROUTe:CLOSe? (@<list>)... 8 ROUTe:OPEN (@<list>)... 8 ROUTe:OPEN:ALL... 8 ROUTe:CHANnel:VERify bool,(@<list>)... 8 ROUTe:CHANnel:VERify? (@<list>)... 8 ROUTe:CHANnel:VERify:POLarity [INVerted NORMal],(@<list>)... 9 ROUTe:CHANnel:VERify:POLarity? (@<list>)... 9 ROUTe:CHANnel:VERify:POSition:STATe? (@<list>)... 9 ROUTe:MODule:BUSY?... 9 ROUTe:MODule:WAIT... 9 1.1 Common Commands... 9 1.1.1 *IDN?... 9 1.1.2 SYSTem:ERRor?... 9 1.1.3 *STB?... 10 1.1.4 *RST... 10

CERTIFICATION VTI Instruments Corp. (VTI) certifies that this product met its published specifications at the time of shipment from the factory. WARRANTY The product referred to herein is warranted against defects in material and workmanship for a period of one year from the receipt date of the product at customer s facility. The sole and exclusive remedy for breach of any warranty concerning these goods shall be repair or replacement of defective parts, or a refund of the purchase price, to be determined at the option of VTI. VTI warrants that its software and firmware designated by VTI for use with a product will execute its programming when properly installed on that product. VTI does not however warrant that the operation of the product, or software, or firmware will be uninterrupted or error free. LIMITATION OF WARRANTY The warranty shall not apply to defects resulting from improper or inadequate maintenance by the buyer, buyer-supplied products or interfacing, unauthorized modification or misuse, operation outside the environmental specifications for the product, or improper site preparation or maintenance. VTI Instruments Corp. shall not be liable for injury to property other than the goods themselves. Other than the limited warranty stated above, VTI Instruments Corp. makes no other warranties, express or implied, with respect to the quality of product beyond the description of the goods on the face of the contract. VTI specifically disclaims the implied warranties of merchantability and fitness for a particular purpose. RESTRICTED RIGHTS LEGEND Use, duplication, or disclosure by the Government is subject to restrictions as set forth in subdivision (b)(3)(ii) of the Rights in Technical Data and Computer Software clause in DFARS 252.227-7013. VTI Instruments Corp. 2031 Main Street Irvine, CA 92614-6509 U.S.A

SUPPORT RESOURCES Support resources for this product are available on the Internet and at VTI Instruments customer support centers. VTI Instruments Corp. World Headquarters VTI Instruments Corp. 2031 Main Street Irvine, CA 92614-6509 Phone: (949) 955-1894 Fax: (949) 955-3041 VTI Instruments Cleveland Instrument Division 5425 Warner Road Suite 13 Valley View, OH 44125 Phone: (216) 447-8950 Fax: (216) 447-8951 VTI Instruments Lake Stevens Instrument Division 3216 Wetmore Avenue, Suite 1 Everett, WA 98201 Phone: (949) 955-1894 Fax: (949) 955-3041 VTI Instruments, Pvt. Ltd. Bangalore Instrument Division #75/76, Millers Road Bangalore 560 002 India Phone: +91 80 4040 7900 Fax : +91 80 4170 0200 Asia Support Phone: +852 9177 6127 Technical Support Phone: (949) 955-1894 Fax: (949) 955-3041 E-mail: support@vtiinstruments.com Visit http://www.vtiinstruments.com for worldwide support sites and service plan information.

SECTION 1 Introduction Background The intent of the programmer s manual is to describe the EX7000 SCPI commands and to introduce its concepts, structure, and capabilities to software and test application. The reader is expected to be familiar with SCPI commands. Understanding the architecture and concept of SCPI commands will significantly help the reader follow the design of the EX7000 SCPI commands described in this manual. Glossary This guide provides instruction for programming the EX7000 series only. Throughout this document, the following terms will be used: EX7000 RDB an LXI-based radio frequency (RF) switch system which is part of VTI Instruments Next Generation line of products an acronym for Relay Driver Board; hardware that interfaces between the controller and the actual microwave switches on an EX7000. Basic Concepts 1) The EX7000 SCPI commands are designed to work with the EX7000 platforms only. 2) An EX7000 platform provides a generic collection of relay drive lines and reset lines intended to drive RF switch coils. The EX7000 SCPI commands allow the user to control coils individually.

SECTION 2 Connection SCPI RAW The basic operation of the SCPI socket is: Connect to TCP port 5025 on the device. Repeat: Send a command or query terminated by a newline. If a query, read query result terminated by a newline. Close the socket. There is only one connection at a time allowed on the socket. When a connection is made, the listener is closed. This allows exclusive access to the device and provides an inherent locking mechanism. To select the SCPI raw interface, use a TCPIP::<hostname>::5025::SOCKET resource string, where <host> is the hostname or the IP address of the EX7000 instrument. VXI-11 The vxi-11 interface acts as a multiplexer to the SCPI socket. Many clients can connect at the same time using vxi-11. All requests will be handled in the order received. If the client uses vxi-11 locking, it will have exclusive access. When a client is using the SCPI socket, the VXI-11 interface will behave as if a VXI -11 client has obtained the lock. To select the VXI -11 interface, use a TCPIP::< hostname >::INSTR resource string, where <host> is the hostname or the IP address of the EX7000 instrument.

Commands Syntax Multiple commands and queries may be given in the same message separated by ; like SCPI. The leading : when using multiple commands is optional. On the other hand, case does not matter. Both long and short forms of a command are accepted and treated in the same way as in SCPI. Multiple queries in the same message are separated by ; like SCPI. And any whitespace in the commands is ignored. Most of the following commands take a list of coils as a parameter. The format of a coil is: K{rdbnumber}_{coilnumber} or R{rdbnumber}_{resetnumber} with the following rules The {rdbnumber} ranges from 1-8 to indicate which RDB is being specified The {coilnumber} ranges from 1-72 to indicate which coil is being specified The {resetnumber} ranges from 1-12 to indicate which reset line is being specified The list is comma separated. Ranges are colon separated. Ranges must be the same type. (All coils or all resets) Ranges can be increasing or decreasing and cross RDB boundaries. Example: Coils 1,3,4,5 on RDB 2: (@ K1_1, k1_3:k1_5) The example above would close the first coil and the third to fifth coils in RDB 2.

EX7000 Commands EX7000 SCPI commands provide a set of commands to close and open as well as verify the state of the drive line and the reset line on the RDB installed in the EX7000 instrument. Please note that the lower case in the command can be omitted. For example, ROUTe:CLOSe can be recognized by ROUT:CLOS and both syntax achieve the same goal. ROUTe:CLOSe (@<list>) Close drive lines or reset lines specified in <list>. When multiple ROUT:CLOS (or ROUT:OPEN) commands are given in the same message, they must have a ROUT:MOD:WAIT command between them. ROUTe:CLOSe? (@<list>) Return the open or closed state of the drive lines or reset line specified in <list>. The values 0 indicates open and 1 indicates closed. The drive value or the confidence value is used depending on the Verify state of each line. ROUTe:OPEN (@<list>) Open drive lines or reset lines specified in <list>. When multiple ROUT:OPEN (or ROUT:CLOS) commands are given in the same message, they must have a ROUT:MOD:WAIT command between them. ROUTe:OPEN:ALL Open all drive lines. ROUTe:CHANnel:VERify bool,(@<list>) Configure ROUT:CLOS? to return the indicator state instead of the driven state. ROUTe:CHANnel:VERify? (@<list>) Returns the Verify mode of the coils in <list>.

ROUTe:CHANnel:VERify:POLarity [INVerted NORMal],(@<list>) Configure the polarity confidence values in the <list> to be inverted, or as-is. ROUTe:CHANnel:VERify:POLarity? (@<list>) Returns the confidence polarity setting of the coils in <list>. A 1 means the confidence value is inverted, and a 0 means the confidence value as-is. ROUTe:CHANnel:VERify:POSition:STATe? (@<list>) If Verify is set for a coil, return the driven state. If Verify is not set, return the confidence value with polarity applied.. ROUTe:MODule:BUSY? Return 1 if relays are settling or 0 if not. May wait an extra 100ms before returning.. ROUTe:MODule:WAIT Waits for relays to debounce. Waits at least 100ms after debouncing. Common Commands A subset of SCPI Common commands that are supported by EX7000. 1.1.1 *IDN? Provides the standard Vendor,Model,Serial,Revision response. 1.1.2 SYSTem:ERRor? Returns the next error in the error queue. Possible errors: 0, No error -102, Syntax error; Unknown command: [<erroneous text>] -350, Queue overflow

-400, rdb out of range -401, coil out of range -402, Mixed Reset lines and Coil lines in range -403, range crosses rdb boundaries 1.1.3 *STB? Always returns 0. 1.1.4 *RST Performs a Reset on the device.

SECTION 3 Application Examples The following section provides some practical examples of how the EX7000 SCPI commands can be used together to meet certain user requirements. Situations will be provided and an appropriate programming example will follow. Close, open and verify a single relay //Close the 3 rd relay on the 2 nd RDB. ROUT:CLOSE(@K2_3); //Verify that the 3 rd relay on the 2 nd RDB is closed. A 1 should be returned. ROUT:CLOSE?(@K2_3); //Open the 3 rd relay on the 2 nd RDB. ROUT:OPEN(@K2_3); //Verify that the 3 rd relay on the 2 nd RDB is open. A 0 should be returned. ROUT:CLOSE?(@K2_3); Close, open and verify couple relays //Close the 3 rd relay on the 2 nd RDB, the 10 th relay on the 1 st RDB and the 5 th relays on the 3 RDB. ROUT:CLOSE(@K2_3, K1_10, K3_5); //Verify that the 3 rd relay on the 2 nd RDB, the 10 th relay on the 1 st RDB and the 5 th relays on the 3 RDB are all closed. 1,1,1 should be returned. ROUT:CLOSE? (@K2_3, K1_10, K3_5); //Open the 3 rd relay on the 2 nd RDB, the 10 th relay on the 1 st RDB and the 5 th relays on the 3 RDB. ROUT:OPEN(@K2_3, K1_10, K3_5); // Verify that the 3 rd relay on the 2 nd RDB, the 10 th relay on the 1 st RDB and the 5 th relays on the 3 RDB are all open. 0,0,0 should be returned. ROUT:CLOSE?(@K2_3, K1_10, K3_5); Close, open and verify a series of relays

//Close the first 5 relays on the 1 st RDB. ROUT:CLOSE(@K1_1:K1_5); //Verify the first 5 relays on the 1 st RDB are all closed. 1,1,1,1,1 should be returned. ROUT:CLOSE?(@K1_1:K1_5); //Open the first 5 relays on the 1 st RDB. ROUT:OPEN(@K1_1:K1_5); //Verify the first 5 relays on the 1 st RDB are all open. 0,0,0,0,0 should be returned. ROUT:CLOSE?(@K1_1:K1_5); Close multiple relays, open all relays at once and verify the state of the relays //Close the 1 st relay on the 1 st RDB. ROUT:CLOSE(@K1_1); //Wait until all the relays are settle. This command is required between multiple close or open command. ROUT:MOD:WAIT; //Close the 2 nd, 3 rd, 4 th, 5 th relay on the 1st RDB. ROUT:CLOSE(@K1_2,K1_3,K1_4,K1_5); //Wait until all the relays are settle. This command is required between multiple close or open command. ROUT:MOD:WAIT; //Close the 6 th to 10 th relays on the 1 st RDB. ROUT:CLOSE(@K1_6:K1_10); //Wait until all the relays are settle. This command is required between multiple close or open command. ROUT:MOD:WAIT; //Open all the relays on all the RDB. ROUT:OPEN:ALL; //Verify all the closed relays are open. 0,0,0,0,0,0,0,0,0,0 should be returned. ROUT:CLOSE?(@K1_1,K1_2,K1_3,K1_4,K1_5,K1_6:K1_10);