DEPT on the U400. 2A. Enter the Macro In the experiment that contains the 13 C spectrum, enter: macro to set up a DEPT experiment in exp2

Similar documents
VNMRJ 4.2 INSTRUCTIONS: QANUC 500 FOR CHEMISTS

Instructions for 1 H-, 13 C-, 19 F-, and 31 P-Spectra on the Varian Mercury-Vx-300

PINMRF. Bruker AV-III / Avance DRX NMR Spectrometers running TopSpin Training Supplement for Advanced 1D NMR Spectroscopy

VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400 AND VARIAN VNMRS 500

Walkup NMR. Varian NMR Spectrometer Systems With VNMR 6.1C Software. Pub. No , Rev. A0800

c. Saturday and Sunday: 10 minute time blocks and unlimited time.

NMR Training VNMRJ 4.2A/VNMRS 400

NMR Users Guide Organic Chemistry Laboratory

VNMRJ 4.2 INSTRUCTIONS: VARIAN MERCURY 400

2 Dimensional NMR User s Brief Guide

PINMRF. Varian 300 MHz NMR Spectrometers Training Guide for Basic 1D NMR Spectroscopy

MNova Version Installation Instructions.

ADVANCED DATA PROCESSING

Gemini-300, Manual Operation (version 1.0, 4Feb95) 1

Using OPUS to Process Evolved Gas Data (8/12/15 edits highlighted)

Basic 1D Processing. Opening Saved Data. Start the VnmrJ software. Click File > Open.

UC DAVIS NMR FACILITY VNMRJ SHORT MANUAL FOR VARIAN/AGILENT NMR SPECTROMETERS

Delta V5 operation. Widetron Technologies Corp. Wu, Chung Ying. Progress through Synergy

MestReC Cheat Sheet. by Monika Ivancic, July 1 st 2005

North Carolina State University Department of Chemistry Varian NMR Training Manual

PROCESSING 2D SPECTRA USING VNMRJ JB Stothers NMR Facility Materials Science Addition 0216 Department of Chemistry Western University

Processing data with Bruker TopSpin

Frequently Asked Questions (FAQ)

TraceFinder Analysis Quick Reference Guide

Chromatography Software Training Materials. Contents

CH142 Spring Spectrophotometers with Vernier Data Acquisition Software

Varian NMR Instructions - 1D. Department of Chemistry NMR Facility University of Minnesota

1. Log in at the RESLOG terminal by entering your code and select the instrument you want to use.

PINMRF. Bruker Avance / AV NMR Spectrometers running TopSpin Training Supplement for Advanced 1D NMR Spectroscopy

Galaxie Photodiode Array Software

TopSpin QuickStart. Default State TopSpin should be running at all times. If this is not the case, you may need to Log In and Start TopSpin.

T 1 Relaxation Measurement: The Inversion-Recovery Experiment (Using IconNMR) Revised

NMR Spectroscopy with VnmrJ. University of Toronto, Department of Chemistry

MassHunter Personal Compound Database and Library Manager for Forensic Toxicology

How to analyze complex NMR signals through spin simulation/iteration

Agilent ChemStation Plus

[ WHITE PAPER ] System Policies and Privileges for Processing Data in Empower Software with Data Integrity in Mind INTRODUCTION

SequencePro Data Analysis Application. User Guide

ChromQuest 5.0 Quick Reference Guide

Quick Start Guide to nmrpipe Processing of 15 N-HSQC data

General Information for Automated Inova 500 MHz NMR (Indy)

15. HOMONUCLEAR COSY

Brief IconNMR and Topspin 3.5 User Guide for Bruker NMR Spectrometers Avance IIIHD 400MHz NMR with Autosampler in Chemistry room 93 DISCLAIMER

TraceFinder Analysis Quick Reference Guide

Retention Time Locking with the MSD Productivity ChemStation. Technical Overview. Introduction. When Should I Lock My Methods?

QUANTAX EDS SYSTEM SOP

Diffusion Ordered Spectroscopy (DOSY) Overview BRUKER last edit 5/14/12

Reference Manual MarkerView Software Reference Manual Revision: February, 2010

TOCSY 15. Goto. Introduction Pulse Sequence Diagram AVANCE User s Guide Bruker 167

Instruction manual for T3DS calculator software. Analyzer for terahertz spectra and imaging data. Release 2.4

Agilent G6854 MassHunter Personal Pesticide Database

Standard Operating Procedure for the Horiba FluroMax-4

To Create Multiple Columns in a Word Document

Short instructions for the basic operation of the AVANCE II 400 MHz Bruker NMR Using ICON-NMR (PS751)

Tutorial 7: Automated Peak Picking in Skyline

VNMRJ 2.1B Quick overview and BioPack features

Evex NanoAnalysis EDS Acquisition

IMSERC NMR MANUAL 02: Basic Processing of Varian 1D NMR Data

Agilent EZChrom Elite. PDA Analysis

Part 1: Calculating amplitude spectra and seismic wavelets

Fluorimeter User s Booklet

NMR Experiment Procedure

IconNMR Operating Procedures

JASCO CANVAS PROGRAM OPERATION MANUAL

Copyright 2008 Casa Software Ltd.

Vanderbilt Small Molecule NMR Center

Bruker Topspin NMR Training Manual

Routine 1 H and 13 C NMR Data Acquisition Basic Topspin Processing Guide For PSC and NSC Bruker 400 NMR Spectrometer

Bruker Avance 400 MHz Instructions

Alpy 600 Mixed Mode Calibration

Open Access Manual for Superusers

UniPoint System Software User s Guide

Troubleshooting Acquisition Related Problems

Thermo Xcalibur Getting Started (Quantitative Analysis)

NMR FACILITY DEPARTMENT OF CHEMISTRY University of Washington XWINNMR 3.5 TRAINING NOTES. 08/8/2005 rajan paranji

NMR Experiment Procedure

MATERIALS PLUS Segmentation Measurement

Getting Started. Environmental Analysis Software

SPECTRAL IMAGING VIEWER SOFTWARE MANUAL

Acquiring Data in VnmrJ 4.2A

NMR Course /13/ :31:47 PM Emory University

TopSpin. Multiplet Analysis Tutorial

ME scope Application Note 18 Simulated Multiple Shaker Burst Random Modal Test

Experiment 5: Exploring Resolution, Signal, and Noise using an FTIR CH3400: Instrumental Analysis, Plymouth State University, Fall 2013

EQNMR Facility Rules of Use

Vnmrj 3.1 New and improved features

Working with Spreadsheets in Target

Eft Spectrometer Guides

Data Acquisition with CP-2002/2003 Micro-GC Control

Varian Solution NMR Procedure

Position Table Creation

NMR INSTRUMENT INSTRUCTIONS: Safety and Sample Preparation

Microsoft Word 2007 on Windows

Multivariate Calibration Quick Guide

Smaart v8.3 Release Overview

SIMPLE TEXT LAYOUT FOR COREL DRAW. When you start Corel Draw, you will see the following welcome screen.

Varian Solution NMR Automation Procedure (VNMRS Machines running VNMRJ 4.2 under Red Hat Enterprise 5.1)

SELECTION OF A MULTIVARIATE CALIBRATION METHOD

Model FP-6500 Spectrofluorometer Instruction Manual. FP-6500 for Windows

DOING MORE WITH WORD: MICROSOFT OFFICE 2013

Transcription:

UVU444 DEPT on the U400 FL-26JUL94CD PRELIMINARY INFORMATION 1) The DEPT (Distortionless Enhancement by Polarization Transfer) experiment is frequently used to identify the types of carbons (i.e., C, CH, CH 2, and CH 3 ) in a molecule. Depending on the setting of the parameter mult, a different DEPT spectrum (or spectra if mult is arrayed), and consequently, different results can be obtained: mult=0.5 DEPT-45 (gives all protonated carbons) mult=1.0 DEPT-90 (gives CH's only) mult=1.5 DEPT-135 (gives CH's and CH 3 's up and CH 2 's down) Quaternary carbon signals are suppressed in the DEPT experiment, and therefore, can be easily identified upon comparison of any DEPT spectrum with the regular 13 C spectrum. Routinely, only a DEPT-135 is needed to identify all the carbons since CH's and CH 3 's usually have quite different chemical shifts. Occasionally, a DEPT-90 may be needed to distinguish CH's from CH 3 's. The DEPT experiment can also be set up to acquire a set of four spectra (mult=0.5,1,1,1.5 for one DEPT-45, two DEPT-90 and one DEPT-135) for use in automatic spectral analysis and/or editing (see details in Part 4 below). Finally, the DEPT experiment is somewhat more sensitive than the APT or normal 13 C experiment. In addition, the setting of the parameter d1 is governed by the T1's of the proton nucleus instead of the carbon nucleus. This could be advantageous if your carbons' T1's are longer than those of the protons, which is the case for most molecules. 2) pw90 and pp: The default pw90 and pp are usually good enough for this experiment, unless the physical properties of your sample are very different from those of the standard due to, for example, the presence of ionic or paramagnetic materials or the use of a very different solvent. Talk to MSL staff if you are not sure if you need to calibrate these parameters for your sample or don't know how. If new values are determined, remember to change the parameter pw90 before setting up the DEPT experiment, and change the parameter pp before starting the DEPT experiment. 3) d1: By default, d1 is set to 1.5 seconds. In general, d1 is set to 1 to 2 times the value of T1 for the PROTONS of interest. 4) nt: By default, nt is set to half the value used in your regular 13 C spectrum due to its higher sensitivity. 1. COLLECT A 13 C SPECTRUM If you have already collected a 13 C spectrum previously, load the data into an experiment (e.g., exp1) and process it as usual. Note that when set up this way, the DEPT spectrum may have to be phased manually. If no 13 C spectrum has been acquired, acquire one and phase and reference the spectrum as usual. 2. SET UP THE DEPT EXPERIMENT dodept INTRODUCTION: dodept is an interactive macro that will take you through the setup of a DEPT experiment. In this handout, the dialog will appear in this font, and the symbol signifies that a response is required. Most messages will appear in the Status Window, while inputs are entered via the Input Window. Window arrangement on the left side of the screen is, from top to bottom, Status, Input, Menu, Graphics, and Text. 2A. Enter the Macro In the experiment that contains the 13 C spectrum, enter: dodept <rtn> macro to set up a DEPT experiment in exp2 NOTE: If you wish to do the DEPT in a different experiment, enter dodept(exp#), where exp# is the experiment number for the DEPT. Experiment number must be between 1 and 9, excluding 5.

If the DEPT exp# is the same as the current experiment, the following dialog will appear: Do dept in current experiment? (y/n) If answered no: Enter dept exp#: Then the following menu will be displayed in the text window at the bottom of the screen: DEPT MENU 1) DEPT-135 2) DEPT-90 3) DEPT-135 and DEPT-90 4) DEPT for spectral editing 2B. Choose a Setup Enter your choice of setup at the following prompt in the Input Window: Enter your choice (1-4) from the menu below: If 1 is chosen, a DEPT-135 experiment will be set up (i.e., mult=1.5). If 2 is chosen, a DEPT-90 experiment will be set up (i.e., mult=1). If 3 is chosen, an array will be set up for a DEPT-135 and a DEPT-90 (i.e., mult=1.5,1). If 4 is chosen, an array will be set up to collect a set of four spectra: one DEPT-45, two DEPT-90, and one DEPT-135 (i.e., mult=0.5,1,1,1.5). This setup is recommended for spectral editing and/or analysis only, because the data will take four times as long to acquire as that in choice #1 and 2 and twice as long as choice #3. By default, d1 will be set to 1.5 seconds and nt to half of that of the regular 13 C spectrum. However, you may change these parameters by answering the following question accordingly: Do you wish to change nt and/or d1? (y/n) If answered yes, enter the new values at the following prompts: nt= d1= The dodept macro will terminate at this point. Be sure to check the parameters before starting the experiment. 2C. Start the Experiment You may start the DEPT experiment with the au or ga command: au start DEPT experiment and process the data automatically with the deptproc macro when acquisition is complete. NOTE: If the experiment is terminated with aa or sa (see below), no data processing will occur. ga start DEPT experiment and do a wft when acquisition is complete. NOTE: Once the experiment is started, it may be aborted by clicking the Abort Acq button or typing the aa command if no data is to be retained, such as when you discover that a mistake has been made in the setup of the experiment, or when it has become apparent that the experiment is useless. If data retention is desired, such as when sufficient signal-to-noise has been obtained or when you have run out of time, you should use the sa 2

command to stop the acquisition. At this point, you should process the data manually because automation will not work when the experiment is terminated before its completion. When the experiment is finished, save the data and proceed with its processing and plotting as described below. 3. PROCESSING AND DISPLAYING NOTE: If a regular 13 C spectrum is acquired together with the DEPT data, it is most likely that no further phasing of the DEPT spectra will be necessary. In that case, you should use the deptproc macro to process the data. On the other hand, if the 13 C spectrum is collected at another time, the DEPT spectra will probably require some phasing, either auto or manual, depending on the type of data. If your data contains a DEPT-45 (preferred) or a DEPT-90 spectrum (may or may not work), you may use the macro deptpaph for processing. Otherwise, use deptproc to process the data and phase the spectrum manually if necessary. * * * * * A. To process and display DEPT data without autophasing: deptproc <rtn> B. To process and display DEPT data with autophasing: deptpaph <rtn> C. To display DEPT data without any processing: deptds <rtn> 4. PLOTTING A. deptplot for data acquired with menu item #1-3 above. It will plot DEPT data, with or without the regular 13 C spectrum on the same page. See examples in Fig. 1-2. deptplot(c13exp#) <rtn> C13exp# is the experiment number that contains the 13 C spectrum - - - - - - - - - - - - - - - - -< >- - - - - - - - - - - - - - - - - If no argument is provided with the deptplot macro, the following dialog will appear: Plot without 13C spectrum? (y/n) If answered yes, the macro will terminate here and only the DEPT spectrum (or spectra) will be plotted (see example in Fig. 1). If answered no, the dialog will continue: Enter exp# containing 13C spectrum: - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - Finally, you are provided with the option to plot the text: Plot with text? Then the plotting will begin. B. adeptpp for data acquired with menu item #4 only. It will plot the edited spectra and print the analysis table, OR plot the unedited spectra. See examples in Fig. 3-4. adeptpp <rtn> enter the plotting macro Then enter your choice of plotting option at the following prompt: Plot 1)edited or 2)unedited spectra? 3

If option 1 is chosen, spectral editing will be performed and the resulting four new spectra and a spectral analysis table will be plotted and printed respectively. The edited spectra will contain, from bottom to top, all protonated carbons, CH carbons, CH 2 carbons, and CH 3 carbons respectively. If option 2 is chosen, the unedited spectra will be plotted (i.e., from bottom to top, DEPT-45, -90, and -135). You will also have a choice of plotting with the parameters, text only, or none at all: Plot with 1)parameters, 2)text only, or 3)none? NOTE: If option 1 is chosen, wc (width of chart) will be adjusted accordingly to accommodate the parameters. Next, the macro will determine if the data have been edited, and if it has, the data will be retransformed. - - - - - - - - - - - - - - - - -< >- - - - - - - - - - - - - - - - - If you choose plotting option 1, the threshold will be adjusted and a line list will be printed in the text window. The number of lines found will also be shown in the status window. Check the threshold setting by answering the following question accordingly (NOTE: only protonated carbons should be and is counted!): Is the number of peaks picked correct? If you answer no to the above question, enter 1 if too many peaks are picked or 2 if too few: 1)increase or 2)decrease the threshold? The threshold will then be increased or decreased by 20% and the above dialog will be repeated until the correct number of lines are picked. Then, spectral editing and analysis will be performed on the data and the analysis table will be printed. - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - Finally, the DEPT spectra, edited or not, will be plotted. C. adeptplot for data acquired with menu item #4 only. It will plot the edited spectra and the spectrum analysis table, all on the same page. See example in Fig. 5. adeptplot <rtn> enter the plotting macro First, the macro will determine if the data have been edited, and if it has, the data will be retransformed. Then, the threshold will be adjusted and a line list will be printed in the text window. The number of lines found will also be shown in the status window. Check the threshold setting by answering the following question accordingly (NOTE: only protonated carbons should be and is counted!): Is the number of peaks picked correct? If answered no, enter 1 if too many peaks are picked or 2 if too few: 1)increase or 2)decrease the threshold? The threshold will then be increased or decreased by 20% and the above dialog will be repeated until the correct number of lines are picked. Then, spectral editing and analysis will be performed on the data, and the resulting edited spectra and analysis table will be plotted on the same page. D. adeptprint for data acquired with menu item #4 only. Prints spectrum analysis table. After the macro is entered, it will determine if the data has been edited. If it has, the results will be printed. If not, a dialog similar to the one for adeptplot will appear, and at the end, a spectral analysis table will be printed. 4

Fig. 1. deptplot: DEPT-135, with 13 C spectrum Fig. 2. deptplot: DEPT-135 and DEPT-90, with 13 C spectrum Fig. 3. adeptpp: edited spectra, with parameters (spectrum analysis table on next page) Fig. 4. adeptpp: unedited spectra, with text only Fig. 5. adeptplot: typical output, with text 5