Operation Manual MODEL 3621 JENCO ELECTRONICS, LTD. Microcomputer Based ph/orp Controller MANUFACTURER OF PRECISION INSTRUMENTS

Similar documents
Cole-Parmer Instrument Company

Operation Manual MODEL 3631 JENCO ELECTRONICS, LTD. Microcomputer Based ph/orp Transmitter MANUFACTURER OF PRECISION INSTRUMENTS

ph/orp 300 (Cat. No ) MICROCOMPUTER BASED OPERATION MANUAL ph /ORP/ Temperature CONTROLLER Phone: , Fax:

Operation Manual TX10 1/8 DIN Microcomputer Based ph/orp Controller

Cole-Parmer Instrument Company

Operation Manual CX20 1/8 DIN Microcomputer Based Conductivity/Resistivity Controller

Operation Manual. MODEL ph6175. Microcomputer Based ph/orp/temperature Benchtop Meter

Operation Manual CX10 1/8 DIN Microcomputer Based Conductivity/Resistivity Controller

Operation Manual. Hand-held ph/orp/temperature Meter 6010M

Operation Manual. MODEL ph 630. Microcomputer Based ph&temperature Pocket Meter. ph630

JENCO ELECTRONICS, LTD. MANUFACTURER OF PRECISION INSTRUMENTS

JENCO INSTRUMENTS, INC. JENCO ELECTRONICS, LTD.

Operation Manual. Hand-held ph/mv/temperature Meter. ph6810

Operation Manual MODEL 2TX. 2-wire Isolated ph/orp Transmitter

Operating Instructions

Operation Manual. Hand-held Conductivity/Salinity/TDS/ Temperature Meter 3010M

JENCO ELECTRONICS,LTD. MANUFACTURER OF PRECISION INSTRUMENTS

JENCO ELECTRONICS, LTD. MANUFACTURER OF PRECISION INSTRUMENTS

OPERATION MANUAL MODEL TT-LP

Operation Manual. MODEL ph/ec80. Microcomputer Based ph/condctivity/tds/ Sanility & Temperature Pocket Meter. ph/ec80

Operation Manual. MODEL ph 618N. Microcomputer Based ph & Temperature Pocket Meter. ph618n

Operation Manual MODEL 3173/3173R. MICROCOMPUTER BASED Conductivity/Salinity/TDS &Temperature BENCH METER 3173/3173R

OPERATION MANUAL. ph 500 (Cat. No ) MICROCOMPUTER BASED ph / Temperature TRANSMITTER & CONTROLLER

Operations Manual EcoSense ph100

Operating Instructions

Operations Manual EcoSense ph100a

JENCO ELECTRONICS, LTD. MANUFACTURER OF PRECISION INSTRUMENTS

4801P ph/orp controller

WalkLAB ph mv/orptemperature

User Manual DOCUMENT #601119REF. EcoSense ph1000a. ph, mv (ORP), and Temperature Benchtop Meter USER MANUAL. English

Colorado Series. CR-10 Portable ph / mv / Temperature Meter Operation Manual

User's Guide. 1/16 th DIN ph Controller. Model 48PH2. Introduction

OPERATION MANUAL. JENCO MODEL 6312 CTB MICROCOMPUTER BASED Conductivity/TDS & Temperature TRANSMITTER & CONTROLLER

Global Water Instrumentation, Inc.


Microprocessor ph/orp Controller

Dissolved Oxygen Meter Kit DOK. Instruction Manual

ph, mv and Temperature Instrument

ph, ORP, and TDS ORP Continuous Monitor OPERATING INSTRUCTIONS 8. Warranty 9. Return of Goods

AquaShock ph Meter K

Instruction Manual. αlpha TDS /8 DIN Total Dissolved Solids Controller with Temperature display and Transmitter. 68X rev 0 08/2002

Water Quality Meter. Water Quality Meter Kit. Instruction Manual

ph/orp Meter SD Card real-time Datalogger

Industrial. Laboratory and Field. Accessories. Product Selection Guide. Analyzers 1 Controllers 2 Transmitters 3 Monitors 4

INSTRUCTION MANUAL. Portable ph and ph/ise Meters. Models 210Aplus 230Aplus 250Aplus 290Aplus

OPERATING MANUAL TCAT-2DF CONTROLLER PHYSITEMP INSTRUMENTS INC. 154 Huron Avenue Clifton NJ USA Tel: Fax:

24-hour Continuous Field Monitors

INSTRUCTION MANUAL ORP-METER MODEL COR-01

Signet ph/orp Transmitter * *

Datalogging ph Meter

Economical Rugged Series M1300K

HI HI N HI HI

TX ph/mv Transmitter/Controller

+GF+ SIGNET ph/orp Transmitter Instructions

INSTRUCTION MANUAL. ph/mv CONTROLLER/RECORDER

HI HI N HI HI

DFI 7000 DIGITAL FORCE INDICATOR USER S GUIDE 1.0 INTRODUCTION 2.0 DESCRIPTION. 2.1 Configuration. 2.2 Specifications

Datalogging Conductivity/ TDS Meter. Instruction Manual

MP-2000 Dual Channel LVDT/RVDT Readout/Controller

Instruction Manual Temp JKT Temperature Meter Series

Professional Benchtop Conductivity meter BC3020

WalkLAB HC Professional Conductivity-TDS-Salinity Meter. Operations Manual 15,0,11,4,1,14,5,10,13,2,9,6,3,12,7,8

PH SD Card Datalogger. Instruction Manual

Thermo Scientific Orion Star A325 Portable ph/conductivity Meter. Instruction Sheet

Tempco Instruction Manual

Thermo Scientific Orion Star A214 Benchtop ph/ise Meter. Instruction Sheet

Bante 221 Portable ph/orp Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD

DDS-12DW Benchtop Conductivity Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD

MULTI- PARAMETER TESTER

Analyzer/Controller. ph/orp HART. Model 54e ph/orp ESSENTIAL INSTRUCTIONS WARNINGS RISK OF ELECTRICAL SHOCK

DPH230SD USER S MANUAL

User Guide. ExStik FL700. Fluoride Meter

Congratulations! Your new minichem-ph monitor is the latest in simple, reliable process monitoring instrumentation. With correct operation and

PS-5030 (5A) OPERATION MANUAL

ph Meter & Controller Model: NANO-PCD1

Type K/J Thermometers

HI ma Simulator

Global Water Instrumentation, Inc.

Thermo Scientific Orion Star A324 Portable ph/ise Meter. Instruction Sheet

HI 3220 HI 3221 HI 3222

FLS M9.06 PH/ORP MONITOR. safety instructions. packing List

PS-1030 (30V) OPERATION MANUAL

ph 500 & mv 600 Series

AquaController Apex Probe Module 2

Instruction Manual HI 9033 HI Waterproof Multi-Range EC and TDS Meters for Field Applications.

Instruments and Capabilities (Panel Mount and Standalone) By: Thornton Leading Pure Water Analytics

HI ma Simulator

Model 901-PH ph-mv-temp. Meter Handbook Version : 2.1 : 25-Mar-99

Platinum RTD Thermometer MODEL NO

Contents. 1. Installation

Manual No Revision MODEL GLI-9835 HAND-HELD/PORTABLE CONDUCTIVITY/TDS METER

Thermo Scientific Orion Star A326 Portable ph/rdo/do Meter. Instruction Sheet

Global Water globalw.com. Global Water. Instrumentation, Inc.

Platinum RTD Thermometer MODEL NO

Oyster Meter ph-conductivity-tds-salinity-orp(mv)

INSTRUCTION MANUAL. Digivac 760 Series. Models 100Led P / P / 29.9" 100Led 760 / SixVac 801 / 760 / 2c 801 Micro

Instruction Manual. ph Monitor ph/mv/ºc meter

PIECAL 322 Automated Thermocouple Calibrator Operating Instructions. Product Description. Practical Instrument Electronics

FLS M9.06. ph/orp MONITOR

Operation Manual. Concorde 600 Power Supply. *This instrument is intended for laboratory use only.

Transcription:

Operation Manual MODEL 3621 Microcomputer Based ph/orp Controller JENCO ELECTRONICS, LTD. MANUFACTURER OF PRECISION INSTRUMENTS

CONTENTS GENERALINTRODUCTION)))))))))))))))..)))...2 INITIALINSPECTION))...))))))...2 USINGTHEJENCOMODEL3621))...3 A. Mounting procedure...3 B. Front panel...5 C. LCD screen...7 D. Rear connectors...9 E. Measure mode...10 F. Settingmode...11 G. phcalibration mode...17 H. RmV calibrationmode...20 I. Controlling the relays...21 ERRORDISPLAYANDTROUBLESHOOTING)))))..)))...23 phbuffers)))))..)))...27 SPECIFICATIONS)))))..)))))))))))))))...28 WARRANTY)))))..)))...30 1

GENERAL INTRODUCTION Thank you for selecting the JENCO Model 3621. The 3621 ph/orp Controller is a rugged microprocessor based instrument assembled in a watertight 1/8 DIN case, designed for use in laboratories and process control applications. The system displays ph or ORP or Temperature and relay status in one large LCD screen. The model 3621 microprocessor performs a self-diagnostic routine every time you turn on the unit providing you with basic information on the stability of the instrument. The model 3621 is equipped with 2 control relays. All control relays are programmable and hysteresis driven. INITIAL INSPECTION Carefully unpack the unit and accessories. Inspect for damages made in shipment. If any damage is found, notify your Jenco representative immediately. All packing materials should be saved until satisfactory operation is confirmed. 2

USING THE JENCO MODEL 3621 ERVIEW A. Mounting Procedure PANEL CUTOUT DIMENSIONS ± ± 0.813 0.032 92.0 0.000 3.622" 0.000 ± ± 45.0 1.772" 0.61 0.00 0.024 0.000 1. Make a cutout on any panel, with a thickness of 1/16 inch (1.5mm) to 3/8 inch (9.5mm). Refer to DRAWING 1. DRAWING 1 DRAWING 2 Mounting Brackets 2. Remove the mounting assembly from the controller and insert the controller into the cutout. Refer to DRAWING 2. 3

Panel DRAWING 3 Mounting brackets 3. Replace the mounting bracket assembly onto the controller and secure the controller to the mounting panel. Refer to DRAWING 3. Note : If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. 4

B. Front Panel The front panel consists of a 4-digit LCD display and 4 keys. ph/orp CONTROLLER 3621 HI1 ph MODE STAND SLOPE 7.00 ATC 1. [MODE]key: 1a. In the Measure mode, this key will switch the display in sequence from ph, Temperature, ORP absolute mv, ORP relative mv and back to ph again. 1b. In the Calibration/Setting mode, pressing this key for three seconds will move you back to the previous parameter in the case when recalibration / resetting is required. 5

2. [UP]key: 2a. In the Calibration mode, pressing this key will show the next possible option. In the Setting mode, pressing this key will show the next possible option and increases the numeral increment. 2b. In the Measure mode, pressing this key and [] key at the same time, the unit will enter thecalibrationmode. 3. [DOWN]key: 3a. In the Calibration mode, pressing this key will show the next possible option. In the Setting mode, pressing this key will show the next possible option and decreases the numeral increment. 3b. In the Measure mode, pressing this key and [] key at the same time, the unit will enter thesettingmode. 4. []key: In any mode where the user can change the settings, pressing this key will save the new settings. If no change has been made then pressing this key will just move 6

the user to the next setting. C. LCD screen 1 2 3 4 5 6 HI1 HI2 LOW1 LOW2 C ph STAND SLOPE RmV 6.86 7.00 ATC MAN 7 8 9 10 11 12 13 1. Major LCD display. 2. This icon will be displayed if the meter is in thecalibration/settingmode. 3. HI1 & LOW1 These icons, when displayed, indicate relay action and relay number. 7

4. HI2 & LOW2 These icons, when displayed, indicate relay action and relay number. 5. Temperature and unit display. 6. ph Unit indicator. 7. STAND This icon will blink before Buffer 1 calibration. The icon will stay on while Buffer 1 is being calibrated. 8. SLOPE This icon will blink before Buffer 2 calibration. The icon will stay on while Buffer 2 is being calibrated. 9. 6.86 The 6.86 buffer group: 6.86, 4.00, 9.18. 10. 7.00 The 7.00 buffer group: 7.00, 4.01, 10.01. 11. ATC This icon will be displayed when a temperature probe is connected. 12. MAN This icon will be displayed when a temperature probe is not connected. 13. RmV Unit indicator. 8

D. Rear connectors INPUT RFF +V -V GND 10k TH Pt 1000 L N E 100~230VAC Rel 1 Rel2 1. Connect the AC line to the rear of the instrument. The model 3621 can be used with 100~240V AC at 50/60 HZ. Make sure the EARTH connector is connected to the earth lead of the AC power line. 2. Connect the proper load to the output relays. Make sure that the load does not exceed the relay rating, 5 Amp at 115VAC and 2.5 Amp at 230 VAC. 9

3. V+ (5VDC) and V- (5VDC) output to provide excitation voltage for ph/orp pre-amplifier only. Note : (1) Make sure that the power is unplugged before wiring your probes, relay etc. (2) Make sure you connect the AC power cord to the correct AC terminals. Connecting incorrectly may damage the unit permanently. E. Measure mode Turning on the unit will always display the Measure mode. This instrument is designed to provide 4 distinct measurements: ph MODE C MODE 7.00 ATC ATC mv MODE MODE RmV 10

1. ph The degree of acidity or alkalinity of the solution. 2. Temperature Current temperature of the solution. 3. ORPmV Ameasurement of absolute ORP mv. 4. ORP RmV A measurement of relative ORP mv. The offset value at the RmV calibration will be added to the ORP absolute value to display the ORP relative value. Note : ORPrelative valuerange:orpabsolutevalue-1000mvtoorpabsolutevalue+1000mv Pressing[MODE] key in the Measure mode will cycle the display from the four modes above. F. Setting mode Pressing[DOWN] key and[] key at the same time, the meter will enter into the Setting mode. 1. Temperature compensation select: 11

+ ATC ATC MAN ph buffer 1 select Pressing [UP] key or [DOWN] key in this screen will cycle the display from 01 (Thermistor: 10k ohm) 02 (Resistor: PT1000) 03 (Manual) modes above. Select the preferred temperature compensation mode, press[] key to save, and enter the next setting screen. 2. ph buffer 1 select: Pressing[UP] key or[down] key in this screen will cycle the display from 7.00 6.86 buffer above. Select the preferred buffer, press[] key to save, and enter the next setting screen. 12

Temperature compensation select ph ph 7.00 6.86 ph buffer 2 select 3. ph buffer 2 select: Buffer 1 7.00 Buffer 2 6.86 ph ph ph ph 7.00 7.00 6.86 6.86 Relay control parameters select Relay control parameters select 13

Pressing [UP] key or [DOWN] key in this screen will cycle the display from 4.01 10.01 (or 4.00 9.18) buffer above. Select the preferred buffer, press[] key to save, and enter the next setting screen. Note : The ph buffer 2 is either 4.01 or 10.01 if select 7.00 buffer at ph buffer 1 select screen. The ph buffer 2 is either 4.00 or 9.18 if select 6.86 buffer at ph buffer 1 select screen. 4. Relay control parameters select: Buffer 2 select ph mv RmV Relay 1 HI/LOW select 14

Pressing [UP] key or [DOWN] key in this screen will cycle the display from ph ORP mv ORP RmV modes above. Select the preferred mode, press[] key to save, and enter the next setting screen. Note : The relay will work with the choice of parameters. 5. Relay1High/Lowselect: Relay control paramelers select HI1 LOW1 Relay 1 set point value 15 Pressing [UP] key or [DOWN] key in this screen will cycle the display from HI LOW modes above. Select the preferred mode, press [] key to save, and enter the next setting screen.

Note : See I. Controlling the relays page. 6. Relay 1 set point value: Relay 1 HI/LOW select ph Relay 1 hysteresis value Pressing [UP] key or [DOWN] key in this screen to adjust the value, press [] key to save, and enter the next setting screen. 7. Relay 1 hysteresis value: Relay 1 set point value ph Pressing [UP] key or [DOWN] key in this screen to adjust the value, press [] key to save, and enter the next setting screen. Relay 2 HI/LOW select 16

8. Relay2High/Lowselect: Same as Relay 1 work way select 9. Relay 2 set point value: Same as Relay 1 set point value 10. Relay 2 hysteresis value: Same as Relay 1 hysteresis value G. ph Calibration mode The model 3621 uses 2-point calibration for ph. The first point must be 6.86/7.00, and the second point can either be 4.00/4.01 or 9.18/ 10.01. In the ph Measure mode, pressing [UP] key and [] key at the same time to allow the meter to go to the phcalibrationmode. 17

1. Buffer 1(STAND) calibration: HI2 ph + ph ph STAND SLOPE 7.00 STAND 7.00 STAND 7.00 Buffer 2 calibration Enter into the ph calibration mode, the STAND icon will flash, the unit is ready to be standardized at the first buffer. Rinse the ph and ATC/Temp probes in distilled water and immerse them in the first buffer solution (either 7.00 or 6.86). Allow temperature reading to stabilize, then press key to calibrate. The ph icon will flash until the unit detects a stable reading. Once the unit calibrates the first point, the SLOPE icon will flash. The unit is ready to be sloped at the second buffer. 18

2. Buffer 2(SLOPE) calibration: Buffer 1 calibration ph ph Measure mode STAND SLOPE 7.00 STAND SLOPE 7.00 19 Rinse the ph and ATC/Temp probe in distilled water and immerse them in the second buffer solution (either 4.00/4.01 or 9.18/10.01). Allow temperature reading to stabilize, then press key to calibrate. The ph icon will flash until the unit detects a stable reading. Once the unit calibrates the second point and the unit will automatically exit the calibration mode and goes to the ph Measure mode. Dual point calibration is complete. Note : In the Setting mode (1. Temperature compensation select), select 03 (Manual temperature compensation mode) if no temperature probe is being used. Press the[up] key or[down] key in themanualtemperature compensation mode to adjust the value to that of the test solution temperature.

Then calibrate buffer 1 and buffer 2. H. RmV Calibration mode + or RmV RmV RmV RmV normal mode The model 3621 uses 1-point calibration for RmV. In the RmV Measure mode, pressing [UP] key and [] key at the same time, the meter will enter into RmV calibration mode. Rinse the ORP probe in distilled water and immerse it in the ORP standard solution, then press [UP] or [DOWN] key to adjust the ORP value to that the ORP standard. Press [] key to save. The unit beeps to indicate a successful calibration. Calibration is now complete and the unit will automatically switch to the ORP relative mvmeasuremode. 20

Note : When the ORP absolute mv reading is off, recalibrate RmV value. I. Controlling the relays 1. Isolation voltage: The maximum isolation voltage of the relay output contacts is 1500 VDC. The voltage differential between the relay output contacts and the load should not exceed 1500 VDC. 2. Output load: The current through the relay output contacts should not exceed 5 Amp at 115 VAC and 2.5 Amp at 230 VAC in order not to cause permanent damage to the relay contacts. This rating is specified for resistive loads only. 3. Relay action, relay set point and hysteresis value: Relay Action Effective RELAY-ON Set Point Effective RELAY-OFF Set Point HI S.P. + (1/2 H.V) S.P. (1/2 H.V) LOW S.P. (1/2 H.V) S.P. + (1/2 H.V.) S.P. = Relay Set point H.V. = Hysteresis value (Dead Band) 21

If the relay action is set to HIGH, the relay will turn ON at (Set Point +1/2 Hysteresis ), and will turnoff at (Set Point -1/2 Hysteresis ). If the relay action is set to LOW, the relay will turn ON at (Set Point -1/2 Hysteresis ), and will turnoff at (Set Point +1/2 Hysteresis ). There are two Independent relays the user can bind to theph,absolutemv or RELATIVE mv mode. The user can only bind the two relays to one reading mode at a time. The user can change this anytime by changing option at the setting mode. Note : 1. The ideal set point range for ph is 0.00 to 16.00 ph. 2. The ideal set point range for absolute mv is 2000 to 2000 mv. 3. The ideal set point range of for relative mv is -3000 to 3000 mv. 22

ERROR DISPLAY AND TROUBLESHOOTING ERVIEW ph/orp Temperature Possible cause(s) Display Mode Display Display [Action(s)] "OvEr" -10.0~120.0 C ph measure mode ph > 16.00pH [Recalibrate] Undr -10.0~120.0 C ph measure mode ph < -2.00pH [Recalibrate] "OvEr" "OvEr" ph measure mode a. Temperature > 120.0 C. [Bring buffer solution to lower temperature.] [Replace temperature probe.] b. No temperature sensor. [Adjust the manual temperature to -10~120 C.] 23

"OvEr" Undr ph measure mode a. Temperature < -10.0 C. [Bring buffer solution to higher temperature.] [Replace temperature probe.] b. No temperature sensor. [Adjust the manual temperature to -10~120 C.] "OvEr" Temperature Absolute ORP mv Absolute ORP mv > +1999 mv reading or relative ORP [Bring solution to a lower ORP reading] "Undr" Temperature Absolute ORP mv Absolute ORP mv < -1999 mv reading or relative ORP [Bring solution to a higher ORP reading] 24

ORP "OvEr" Absolute ORP mv a. Temperature > 120.0 C. reading or relative ORP [Bring buffer solution to lower temperature.] [Replace temperature probe.] b. No temperature sensor. [Adjust the manual temperature to -10~120 C.] ORP "Uder" Absolute ORP mv a. Temperature < -10.0 C. reading or relative ORP [Bring buffer solution to higher temperature.] [Replace temperature probe.] b. No temperature sensor. [Adjust the manual temperature to -10~120 C.] 25

"OvEr" 0~60 C a. ph calibration mode-buffer 1 (STAND)-7.00 b. ph calibration mode-buffer 1 (STAND)-6.86 c. ph calibration mode-buffer 2 (SLOPE) a. Offset < -100 or > 100 mv b. Offset < -91.7 or > 108.3 mv c. Slope > ideal mv by 30% [Use a new buffer solution.] [Replace electrode.] 26

ph BUFFERS The temperature characteristics of ph calibration buffers ph4.00, ph4.01, ph6.86, ph7.00, ph9.18 & ph10.01 are stored inside the instrument. The buffers used to calibrate the instrument must exhibit the same temperature characteristics as the stored values. C 4.00 6.86 9.18 4.01 7.00 10.01 0 4.01 6.98 9.46 4.01 7.11 10.32 5 4.00 6.95 9.39 4.01 7.08 10.25 10 4.00 6.92 9.33 4.00 7.06 10.18 15 4.00 6.90 9.28 4.00 7.03 10.12 20 4.00 6.88 9.23 4.00 7.01 10.06 25 4.00 6.86 9.18 4.01 7.00 10.01 30 4.01 6.85 9.14 4.01 6.98 9.97 35 4.02 6.84 9.10 4.02 6.98 9.93 27

40 4.03 6.84 9.07 4.03 6.97 9.89 45 4.04 6.83 9.04 4.04 6.97 9.86 50 4.06 6.83 9.02 4.06 6.97 9.83 55 4.07 6.83 8.99 4.08 6.97 9.80 60 4.09 6.84 8.97 4.10 6.98 9.78 Note : The actual reading of the instrument can differ from the values shown by ±0.01 ph. SPECIFICATIONS Mode Range Resolution Accuracy ph -2.00 to 16.00 ph 0.01 ph ±0.1% ± 1 digit ORP Absolute mv -1999 to 1999 mv 1 mv ±0.1% ± 1 digit ORP Relative mv -2999 to 2999 mv 1 mv ±0.1% ± 1 digit Temperature -10.0 to 120.0 C 0.1 C ±0.3 C ph: 28

Recognized ph buffers US (4.01, 7.00, 10.01) or NIST (4.00, 6.86, 9.18) ph Temperature compensation Manual/Auto -10.0 C to 120.0 C ph Buffer Temperature range 0.0 C to 60.0 C ph Electrode Offset recognition ±100 mv at ph 7.00 +108.3 mv/-91.7 mv at ph 6.86 ph Electrode Slope recognition ±30% at ph 4.00, 4.01, 9.18, 10.01 Input impedance >10 12 Calibration end point sensing Yes Temperature: Temperature sensor Thermistor: 10k ohm at 25 C, (User selectable) Resistor ( PT1000) or Manual Controller: Control type Two ON/OFF control Relay output(resistive load only) 5A at 115VAC or 2.5A at 220VAC 29

General: Keys Audio feedback in all keys Power: 100VAC to 240VAC, 50/60Hz Ambient Temperature range 0.0 to 50.0 C Case IP65, 1/8DIN case, depth 90mm Weight 290 g WARRANTY Jenco warrants this product to be free from significant deviations in material and workmanship for a period of 1 year from date of purchase. If repair or adjustment is necessary and has not been the result of abuse or misuse, within the year period, please return-freight-prepaid and the correction of the defect will be made free of charge. If you purchased the item from our Jenco distributors and it is under warranty, please contact them to notify us of the situation.jenco Service Department alone will determine if the product problem is due to deviations or customer misuse. Out-of-warranty products will be repaired on a charge basis. 30

RETURN OF ITEMS Authorization must be obtained from one of our representatives before returning items for any reason. When applying for authorization, have the model and serial number handy, including data regarding the reason for return. For your protection, items must be carefully packed to prevent damage in shipment and insured against possible damage or loss. Jenco will not be responsible for damage resulting from careless or insufficient packing. A fee will be charged on all authorized returns. Note : Jenco reserves the right to make improvements in design, construction and appearance of our products without notice. 31

Jenco Instruments, Inc. 7968 Arjons Drive, Suite C San Diego, CA 92126 USA TEL: 858-578-2828 FAX: 858-578-2886 E-Mail: jencoinfo@jencoi.com; sales@jencoi.com Website: www.jencoi.com Jenco Electronics Inc. 1F., NO. 11, Lane 370, Sec. 6, Zhongxiao E. Rd. Nangang Dist., Taipei, Taiwan TEL: 886-2-2782-3226 FAX: 886-2-2782-3234 Shanghai Jenco Instruments, Ltd. 18 Wang Dong Zhong Road Sijing Town, Songjiang Shanghai, China TEL: 86-021-5761-9599 FAX: 86-021-5761-9598 E-Mail: jencos@jenco.com.cn Website: www.jenco.com.cn 32

英文说明书 机型 :3621 编制人 : 汪蕾 日期 :2012-03-17 文件编号 :K3621-01 修改人 : 汪蕾 日期 :2012-08-27 版本 :01 核准人 : 日期 : 33